3.3 IM PWB
(1) Functions
This PWB is used to record ultrasound images and performs loop playback or frame advance
playback of such recorded images on the monitor of the main unit. This PWB is optional.
Performs IIR frame correlation when data is written (recorded) into RAM consisting of
image memory. The IIR correlation circuit is used to pass through the ROM into which
correlation coefficients are written. Correlation uses data from the ADC and the
previous frame data. This function can specify the same coefficient as that for frame
correlation of FM from DIGITAL by the host.
Generates signals for memory address count timing and memory write control timing
with respect to the sampling endpoint, sampling speed, combination focus point, etc.
specified by the host.
Preserves each of the recording, frame advance playback, loop playback, and CPU
access modes specified by the host and generates the frame numbers from them. It
also generates control signals for each section.
3-17
No. 2D730-154E
Serves as the interface through which the control register and status reading register
are accessed from the host CPU. The data bus width must be 16 bits and the address
must be composed of 16 lines. Address decoding is performed inside the image
memory board.
Consists of registers which store the IIR and LIP utilization patterns specified by the
control host.
from to
"DIGITAL" "DIGITAL"
8 F/F 8 F/F 6 8 B/W MEMORY 8 8 F/F
IIR LIP
ADCCK
6 DRAM
synchronization Read timing
16Mbit *2 synchronization
(ADCCK 15 MHz)
Write timing
synchronization
Data BUF
CPU
ACCESS Address
CONT BUF
MODE
FRAME
NUMBER
CONTROL
Control register