2D730-154E

2. TROUBLESHOOTING

2.1 System Functions

(1) ISO-TRANS unit/power unit

The AC input is supplied to the power supply unit through the ISO-TRANS unit. To the
peripheral units, the power is supplied via the external output (isolation). The power unit
outputs ±5 V, ±12 V, and VH for the analog system and +5 V and +12 V for the digital system.

(2) T&R PWB

(a) PB-CNN : Connects the transducer connector and the PULSER PWB by the PWB.

(b) PULSER : Generates the ultrasound transmission high-voltage pulse and amplifies
the received ultrasound signals by the control data from the RECEIVER
PWB.

(c) RECEIVER : Performs phasing/addition and other signal processing for the received
signals after they are amplified in the PULSER PWB and then outputs
the echo signals to the DIGITAL PWB. This PWB has the MTX-SW, R-
DELAY, ECHO-FILTER, and RECEIVER T/R CONT functions.

(3) DIGITAL PWB

The DIGITAL PWB consists of a single PWB.


This PWB has the DSC-IO, DSC-FM, CPU, PANEL-IF, RPG, and STC functions.

(4) Keyboard

The keyboard is controlled by the CPU. Data transmission/reception is performed by


interrupting the CPU using the key switch control IC, gain, IP, or freeze switch.
The LED is turned ON/OFF according to the status signal from the IO PWB.
A buzzer is generated when an interrupt is transmitted to the CPU.

(5) ECG unit (option)

The output of the ECG amplifier is sent to the DIGITAL PWB by the PPM only in M mode.
The ECG output is added to the echo signals on the DIGITAL PWB and the ECG waveform is
displayed on M mode images.

(6) RS-232C

The system is provided with an RS-232C-IF channel for data transmission/reception with an
external computer.

(7) Transducer selector (option)

After this option is installed, two transducers can be connected to the system (only one
transducer can be connected in the standard configuration).

(8) Image memory (option)

A minimum of 32 frames can be read. Cine loop playback and frame-feed playback are
possible.

(9) Filing unit (option)

This option can record and file image data electronically.

2-1
No. 2D730-154E

2.2 System Block Diagram

MONITOR

MONITOR
I/O-ASSY

FOOT SW
BSM31-5669

VCR OUT

PRINTER

RS-232C
VCR IN

ECG
OUT

LAN
BSM31-5670 BSM31-5670

FAN FAN
CN6 CN5 CN4 CN3 CN2 CN1

I/O_CN10

I/O_PCB I/O_PCB
CN10
CN8 CN9

I/O

CN11 CN7 : LABEL

BSM31-5628 BSM31-5662 BSM31-5676


LCD BSM31-5630 DIGITAL FILING CONT
CNN1(FC)
LCD-BACKLIGHT
CN1
BSM31-5664
INVERTER

BSM31-5677
MEMORY-CARD CNN1(FC)
I/O_PCB CN02 FILING CD-R
CN1
CNN1(CDR) BSM31-5677-02
CN01

BSM31-5632
CN6
LCD-HV FILING CABLE
CNN1(CDR)
BSM31-5663
BSM31-5677-01
BSM31-5629 IM
FILING CD-R
LCD-MAIN CN101
TFT
CN101

BSM31-5660
CN5
POWER-SW-ASSY
CN1,2,3 PM30-33703 Gr1
LCD-UNIT PM30-33703 Gr2
T&R-PWB-ASSY
UPCON_CN8 CN3,4,5 CN6,7

BSM31-5668 TRANS
UPCON_CN4

BM31-5648
CN8 DT-CNN CN8,9
#POWER-UNIT
BSM31-5631 CN10 CN11 CN1 GND
LCD-UPCONV.
CN4

CN2 CN3 DTCNN,CN1


DTCNN_CN11
BM31-5646
BM31-5647

BSM31-5633 UPCON_CN3 DTCNN_CN11 PM30-33712 Gr1 #ISO-ASSY


PM30-33713 Gr1
LCD-CABLE
UPCON_CN2 OP-ASSY

2-2
No. 2D730-154E

2.3 Troubleshooting Related to Power Supply

The test program installed in the system operates on the premise that the power supply unit,
operating panel, monitor, and software are functioning normally. This subsection describes the
troubleshooting procedures related to the power supply and problems that cannot be checked by
the test program.

(1) Push-type breakers are used for the power unit and the ISO-TRANS unit. These breakers
are set to OFF by pushing the knob.

(2) A short-circuit protection circuit is installed in the output (secondary side) of the power unit. If
this circuit detects an abnormality in the output of the power unit, it turns OFF all of the power
supplies and stops the internal fans.

(3) When troubleshooting is completed, wait for at least 5 minutes before turning the system
power ON again.

<<Troubleshooting flow for the power unit>>

The system power cannot be


turned ON.
No image is displayed.

Use a tester or other suitable instrument for check.

Check whether primary N Ensure that primary power is


power is input. output.

Disconnect the transducer, ECG unit


connection cable, and footswitch from
the system.

Turn the system power ON. Note: If an abnormal smell is noted, stop this troubleshooting
procedure and identify the source of the abnormal smell.

An image is
Rotate the displayed.
monitor brightness VR to Set the brightness VR at the
check whether an image is appropriate position. 1
displayed.

No image is displayed.

Set the brightness VR to maximum.

Check The LED is lit.


whether the LED on the +5 V is supplied normally.
operating panel is lit. The monitor or Digital PWB is defective. 2

The LED is not lit.

The fans are


Check whether rotating. +12 V is supplied normally.
the fans of the I/O panel The monitor, Digital PWB, or the operating 3
are rotating. panel is defective.

The fans are not rotating

2-3
No. 2D730-154E

Check whether
the power of peripheral units
can be turned ON.
The breaker
is not OFF.
The power switch, internal
Check the breaker of
AC cable, or ISO-TRANS
the ISO-TRANS unit.
is defective.

The breaker is OFF.

Turn the system power OFF and then turn


ON the breaker of the ISO-TRANS unit.

If the breaker is OFF, wait for at least 5 minutes


before turning the system power ON again.
Turn the system power OFF and remove
the rear cover and I/O panel.

The breaker is OFF.


Check the breaker
of the power unit.

The breaker is not OFF. Turn the system power OFF and remove
the rear cover and I/O panel.

If the breaker is OFF, wait for at least 5 minutes


before turning the system power ON again.

Remove the power unit from the


system.

Check for foreign There is a foreign object.


objects in the power unit connector
on the DT-CNN PWB or in the output
connector on the power
unit.

There is no foreign object. Remove the foreign object.

Reinstall the power unit in the system.

The fans do
Check whether the not rotate.
The DT-CNN PWB or the power unit
internal fans rotate. is defective. 4

The fans rotate.


Note: The power unit is normal.

Turn the system power OFF.

Connect the operating panel connection


cable to the DT-CNN PWB.

Reinstall the power unit in the system.

2-4
No. 2D730-154E

Turn the system power ON.

The fans do
Check whether not rotate. The operating panel is defective.
the fans in the power Replace it with one that functions normally
unit rotate. and check system operations.

The fans rotate.

The LED is
Check whether not lit. The operating panel is defective.
the LED on the operating Replace it with one that functions normally
panel is lit. and check system operations.

The LED is lit.

Turn the system power OFF.

Connect the Digital PWB to the DT-CNN PWB.

Turn the system power ON.

The fans do
Check
not rotate. The Digital PWB is defective.
whether the fans in the
Replace it with one that functions normally
power unit rotate.
and check system operations.

The fans rotate.

Turn the system power OFF.

Connect the monitor connection cable to


the DT-CNN PWB.

Turn the system power ON.

The fans do
Check
not rotate. The monitor is defective.
whether the fans in the
Replace it with one that functions normally
power unit rotate.
and check system operations.

The fans rotate.

No ultrasound image
Check is displayed.
whether an ultrasound The monitor is defective.
image is displayed on the Replace it with one that functions normally
monitor. and check system operations.

An ultrasound image
is displayed.

2-5
No. 2D730-154E

Turn the system power OFF.

Connect the T&R ASSY to the DT-CNN PWB.

Turn the system power ON.

The fans do
Check The T&R ASSY is defective.
not rotate.
whether the fans in the Replace it with one that functions normally
power unit rotate. and check system operations.

The fans rotate.

Turn the system power OFF.

Reinstall the I/O panel.


If the filing unit is installed, remove it.

Turn the system power ON.

Check The fans do


not rotate. The I/O panel is defective.
whether the fans in the
Replace it with one that functions normally
power unit rotate.
and check system operations.

The fans rotate.

Turn the system power OFF.

Reinstall the filing unit.


If the CD-R unit is installed, disconnect the
CD-R unit connection cable.

Turn the system power ON.

Check The fans do


not rotate. The filing unit is defective.
whether the fans in the Replace it with one that functions normally
power unit rotate. and check system operations.

The fans rotate.

2-6
No. 2D730-154E

Turn the system power OFF.

Connect the CD-R unit to the filing unit.

Turn the system power ON.

Check The fans do


not rotate. The CD-R unit is defective.
whether the fans in the Replace it with one that functions normally
power unit rotate. and check system operations.

The fans rotate.


1

Turn the system power OFF.

Connect a transducer to the transducer


connector on the system.

Turn the system power ON.

Check The fans do


not rotate. The transducer is defective.
whether the fans in the Replace it with one that functions normally
power unit rotate. and check system operations.

The fans rotate.

Turn the system power OFF.

Connect the footswitch to the system.

Turn the system power ON.

Check The fans do


not rotate. The footswitch is defective.
whether the fans in the Replace it with one that functions normally
power unit rotate. and check system operations.

The fans rotate.

Turn the system power OFF.

2-7
No. 2D730-154E

Connect the ECG unit to the system.

Turn the system power ON.

Check The fans do


not rotate. The ECG unit is defective.
whether the fans in the Replace it with one that functions normally
power unit rotate. and check system operations.

The fans rotate.

Turn the system power OFF.

Observe system operations for a while.

2-8
No. 2D730-154E

Check The fans do not rotate.


whether the fans in the
I/O panel rotate.

The fans rotate.

Set the system to external video input.

An image is
Rotate the displayed. Replace the Digital PWB and then the
monitor brightness VR to
check whether an image monitor with ones that function normally
is displayed. and check system operations.

No image is displayed.

Replace the monitor and then the Digital


PWB with ones that function normally
and check system operations.

There is a problem
in the +12-V system.

Turn the system power OFF and remove


the rear cover and I/O panel.

Turn the system power ON.

An image is
Rotate the displayed. The I/O panel (fan) is defective.
monitor brightness VR to
check whether an image Replace it with one that functions normally
is displayed. and check system operations.

No image is displayed.

Set the brightness VR to maximum.

Turn the system power OFF, disconnect the monitor cable from
the DT-CNN PWB, and reinstall the I/O panel in the system.

Turn the system power ON.

The fans do
Check
not rotate. The monitor (or the monitor cable) is defective.
whether the fans in the
Replace it with one that functions normally
I/O panel rotate.
and check system operations.

The fans rotate.

Replace the Digital PWB and then the I/O panel with ones
that function normally and check system operations.

2-9
No. 2D730-154E

Turn the system power OFF and then replace


the power unit with one that functions normally.

Turn the system power ON.

Check The fans do


not rotate. The DT-CNN PWB is defective.
whether the fans in the Replace it with one that functions normally
power unit rotate. and check system operations.

The fans rotate.

The power unit is defective.


Replace it with one that functions normally
and check system operations.

2-10 *