2. TROUBLESHOOTING
The AC input is supplied to the power supply unit through the ISO-TRANS unit. To the
peripheral units, the power is supplied via the external output (isolation). The power unit
outputs ±5 V, ±12 V, and VH for the analog system and +5 V and +12 V for the digital system.
(a) PB-CNN : Connects the transducer connector and the PULSER PWB by the PWB.
(b) PULSER : Generates the ultrasound transmission high-voltage pulse and amplifies
the received ultrasound signals by the control data from the RECEIVER
PWB.
(c) RECEIVER : Performs phasing/addition and other signal processing for the received
signals after they are amplified in the PULSER PWB and then outputs
the echo signals to the DIGITAL PWB. This PWB has the MTX-SW, R-
DELAY, ECHO-FILTER, and RECEIVER T/R CONT functions.
(4) Keyboard
The output of the ECG amplifier is sent to the DIGITAL PWB by the PPM only in M mode.
The ECG output is added to the echo signals on the DIGITAL PWB and the ECG waveform is
displayed on M mode images.
(6) RS-232C
The system is provided with an RS-232C-IF channel for data transmission/reception with an
external computer.
After this option is installed, two transducers can be connected to the system (only one
transducer can be connected in the standard configuration).
A minimum of 32 frames can be read. Cine loop playback and frame-feed playback are
possible.
2-1
No. 2D730-154E
MONITOR
MONITOR
I/O-ASSY
FOOT SW
BSM31-5669
VCR OUT
PRINTER
RS-232C
VCR IN
ECG
OUT
LAN
BSM31-5670 BSM31-5670
FAN FAN
CN6 CN5 CN4 CN3 CN2 CN1
I/O_CN10
I/O_PCB I/O_PCB
CN10
CN8 CN9
I/O
BSM31-5677
MEMORY-CARD CNN1(FC)
I/O_PCB CN02 FILING CD-R
CN1
CNN1(CDR) BSM31-5677-02
CN01
BSM31-5632
CN6
LCD-HV FILING CABLE
CNN1(CDR)
BSM31-5663
BSM31-5677-01
BSM31-5629 IM
FILING CD-R
LCD-MAIN CN101
TFT
CN101
BSM31-5660
CN5
POWER-SW-ASSY
CN1,2,3 PM30-33703 Gr1
LCD-UNIT PM30-33703 Gr2
T&R-PWB-ASSY
UPCON_CN8 CN3,4,5 CN6,7
BSM31-5668 TRANS
UPCON_CN4
BM31-5648
CN8 DT-CNN CN8,9
#POWER-UNIT
BSM31-5631 CN10 CN11 CN1 GND
LCD-UPCONV.
CN4
2-2
No. 2D730-154E
The test program installed in the system operates on the premise that the power supply unit,
operating panel, monitor, and software are functioning normally. This subsection describes the
troubleshooting procedures related to the power supply and problems that cannot be checked by
the test program.
(1) Push-type breakers are used for the power unit and the ISO-TRANS unit. These breakers
are set to OFF by pushing the knob.
(2) A short-circuit protection circuit is installed in the output (secondary side) of the power unit. If
this circuit detects an abnormality in the output of the power unit, it turns OFF all of the power
supplies and stops the internal fans.
(3) When troubleshooting is completed, wait for at least 5 minutes before turning the system
power ON again.
Turn the system power ON. Note: If an abnormal smell is noted, stop this troubleshooting
procedure and identify the source of the abnormal smell.
An image is
Rotate the displayed.
monitor brightness VR to Set the brightness VR at the
check whether an image is appropriate position. 1
displayed.
No image is displayed.
2-3
No. 2D730-154E
Check whether
the power of peripheral units
can be turned ON.
The breaker
is not OFF.
The power switch, internal
Check the breaker of
AC cable, or ISO-TRANS
the ISO-TRANS unit.
is defective.
The breaker is not OFF. Turn the system power OFF and remove
the rear cover and I/O panel.
The fans do
Check whether the not rotate.
The DT-CNN PWB or the power unit
internal fans rotate. is defective. 4
2-4
No. 2D730-154E
The fans do
Check whether not rotate. The operating panel is defective.
the fans in the power Replace it with one that functions normally
unit rotate. and check system operations.
The LED is
Check whether not lit. The operating panel is defective.
the LED on the operating Replace it with one that functions normally
panel is lit. and check system operations.
The fans do
Check
not rotate. The Digital PWB is defective.
whether the fans in the
Replace it with one that functions normally
power unit rotate.
and check system operations.
The fans do
Check
not rotate. The monitor is defective.
whether the fans in the
Replace it with one that functions normally
power unit rotate.
and check system operations.
No ultrasound image
Check is displayed.
whether an ultrasound The monitor is defective.
image is displayed on the Replace it with one that functions normally
monitor. and check system operations.
An ultrasound image
is displayed.
2-5
No. 2D730-154E
The fans do
Check The T&R ASSY is defective.
not rotate.
whether the fans in the Replace it with one that functions normally
power unit rotate. and check system operations.
2-6
No. 2D730-154E
2-7
No. 2D730-154E
2-8
No. 2D730-154E
An image is
Rotate the displayed. Replace the Digital PWB and then the
monitor brightness VR to
check whether an image monitor with ones that function normally
is displayed. and check system operations.
No image is displayed.
There is a problem
in the +12-V system.
An image is
Rotate the displayed. The I/O panel (fan) is defective.
monitor brightness VR to
check whether an image Replace it with one that functions normally
is displayed. and check system operations.
No image is displayed.
Turn the system power OFF, disconnect the monitor cable from
the DT-CNN PWB, and reinstall the I/O panel in the system.
The fans do
Check
not rotate. The monitor (or the monitor cable) is defective.
whether the fans in the
Replace it with one that functions normally
I/O panel rotate.
and check system operations.
Replace the Digital PWB and then the I/O panel with ones
that function normally and check system operations.
2-9
No. 2D730-154E