No.

2D730-154E

8. DISPOSAL

8.1 Disassembly into Units

Refer to the service manual 2D730-502E for the procedure for disassembling the system into units.

• Personnel required : 2
• Mass of the system main unit : Approximately 45 kg

CAUTION: Be sure to use nippers and not a knife or cutter to cut cable bands.
Knives or cutters may slip and cause hand injuries.

Standard time
Step Tools required Precautions
required
0 Preparation Disconnect the power plug of the
system and lock the casters before
starting the work.
1 Removing the 3 minutes M3 Phillips screwdriver 1. The following units are heavy.
peripheral units Take care when handling these
units.
<Reference> Mass of the peripheral
units
B/W printer : UP-D895 : 3.4 kg
2 Removing the 5 minutes M3 Phillips screwdriver, 1. The backlight of the LCD monitor
monitor M6 Allen wrench contains mercury, which is harmful.
Disposal of the LCD should be
entrusted to a specialized
disposal/recycling company.
2. If the LCD is subjected to an
impact, the LCD may be damaged.
Take care not to drop the LCD
monitor.
3. The cable sheaths are made of
PVC. Wash your hands thoroughly
after handling the cables.
3 Removing the 10 minutes M4 Phillips screwdriver,
exterior covers M3 Phillips screwdriver
and rear handle
4 Removing the 5 minutes M4 Phillips screwdriver, 1. The cover becomes unstable when
peripheral unit M6 Allen wrench the screws are removed. Hold the
mounting plate cover while removing the rack.
5 Removing the 5 minutes M4 Phillips screwdriver 1. The operating panel weighs 2.5 kg.
operating panel
2. The cable sheaths are made of
PVC. Wash your hands thoroughly
after handling the cables.

8-1
No. 2D730-154E

Standard time
Step Tools required Precautions
required
6 Removing the 3 minutes M3 Phillips screwdriver 1. The cable sheaths are made of
ECG unit (option) PVC. Wash your hands thoroughly
after handling the cables.
2. To prevent the ECG unit from
falling, hold it while removing the
last bolt.
7 Removing the 5 minutes M4 Phillips screwdriver, 1. Remove the air filter.
PWB rack M3 Phillips screwdriver
2. The PWB rack (with circuit boards)
weighs approx. 7 kg.
3. The cable sheaths are made of
PVC. Wash your hands thoroughly
after handling the cables.
8 Removing the 15 minutes M3 Phillips screwdriver 1. The power supply unit weighs
power supply unit approx. 3 kg.
2. The cable sheaths are made of
PVC. Wash your hands thoroughly
after handling the cables.
9 Removing the 10 minutes M12 wrench 1. The frame weighs approx. 10 kg on
casters its own.
2. The casters may swing under their
own weights. Take care not to
allow your fingers to be caught by a
caster.
10 Removing the 10 minutes M4 Phillips screwdriver 1. The transformer weighs approx.
transformer 11 kg.
2. The cable sheaths are made of
PVC. Wash your hands thoroughly
after handling the cables.
11 Removing the I/O 5 minutes M3 Phillips screwdriver 1. The cable sheaths are made of
panel PVC. Wash your hands thoroughly
after handling the cables.
2. Hold the I/O panel when removing
the last screw to prevent the I/O
panel from falling.

8-2
No. 2D730-154E

8.2 Disposal and Recycling

No. Unit Precautions Items Material


1 Optional 1. If storage media are left in recording B/W printer
peripheral units units, request your Toshiba Medical VCR
Systems Corporation service Filing unit
representative to take appropriate DVD
measures. CD-R (figure 1)
HDD (figure 1)
2. Confirm with your Toshiba Medical
IASSIST controller
Systems Corporation service
Footswitch
representative that the hard disk
Gel warmer
has been erased.
TEE hanger
3. Entrust the disposal of recording etc.
units to a specialized
disposal/recycling company.
2 Monitor 1. Entrust disposal of the LCD and Resin cover >ABS<
circuit boards to a specialized
disposal/recycling company. Tilt mechanism Iron

2. The screws securing the back cover Shield case Iron


are designed so that they do not fall Mounting bracket Iron
out. Pull out the bushing to remove
the screws from the back cover. Screw Iron

3 Exterior cover 1. Two brass insert nuts are used in Resin cover >ABC<
Handle the front cover and a brass insert
nut is used in the rear cover. In the Bottom cover Aluminum die
case of separate disposal/recycling, casting material
remove them from the covers. Cable hook >PPE<
2. Remove the name labels. Mounting bracket Iron
Cap >PVC<
Handle Aluminum die
casting material
Screw, bolt Iron
Leaf spring, Stainless steel
machined screw
4 Peripheral unit Mounting plate Iron
mounting plate
Screw Iron
5 Operating panel 1. Entrust disposal of the keyboard, Resin cover >ABS<
speaker, and circuit boards to a
specialized disposal/recycling Knob >ABS<
company. Key top >PC<
2. Three brass insert nuts are used in (transparent)
the top cover. If separate Trackball >POM<
disposal/recycling is required,
remove them from the covers. Metal plate Iron
Screw, bolt Iron
6 ECG unit 1. Entrust disposal of the ECG unit to
(option) a specialized disposal/recycling
company.

8-3
No. 2D730-154E

No. Unit Precautions Items Material


7 PWB rack 1. Entrust disposal of the circuit Chassis Iron
boards to a specialized
disposal/recycling company. PWB guide rail >PA66<

2. The lithium battery used on the Metal plate Iron


Digital PWB contains harmful Screw Iron
substances. It can be removed
from the holder on the board.
Entrust disposal of the lithium
battery to a specialized
disposal/recycling company and
inform them that it contains harmful
substances.
3. The gasket, shield, cushion, and
cables are affixed using double-
sided tape. In case of separate
disposal/recycling, remove them
from the rack.
8 Power supply 1. Entrust disposal of the transformer, Chassis Iron
unit DC power supply, AC power supply,
line filter, and circuit boards to a Metal plate Iron
specialized disposal/recycling Air filter Stainless steel
company.
Screw Iron
Braided shield Stainless steel
9 Frame 1. The casters may swing under their Frame Iron
Casters own weights. Take care not to
allow your fingers to be caught by a Base Aluminum die
caster. casting material

2. Entrust disposal of the casters to a Screw Iron


specialized disposal/recycling Caster cover >PA66<
company.

10 Transformer 1. Entrust disposal of the transformer Chassis Iron


and circuit boards to a specialized
disposal/recycling company. Screw Iron

11 I/O PWB 1. Entrust disposal of the circuit Metal plate Iron


boards to a specialized
disposal/recycling company. Screw Iron

2. The gasket and shield are affixed


using double-sided tape. If
separate disposal/recycling is
required, remove them from the
panel.

8-4 E