Professional Documents
Culture Documents
3180 Oximeter: Service Manual
3180 Oximeter: Service Manual
3180 Oximeter: Service Manual
Service Manual
98
%SpO2
OFF
96
80
BPM
200
30
- English
Catalog Number 1905
Version 5, June 2008
© 2008 Smiths Medical family of companies. All rights reserved.
Table of Contents
Table of Contents
BCI and the Smiths design mark are trademarks of the Smiths Medical family of companies. The symbol ®
indicates the trademark is registered in the U.S. Patent and Trademark Office and certain other countries. All
other names and marks mentioned are the trade names, trademarks or service marks of their respective owners.
3180 Service Manual i
Table of Contents
Revision History
Proprietary Notice
Information contained in this document is copyrighted by Smiths Medical PM, Inc. and may not be duplicated in
full or part by any person without prior written approval of Smiths Medical PM, Inc. Its purpose is to provide the
user with adequately detailed documentation to efficiently install, operate, maintain, and order spare parts for
the device supplied. All information contained in this document is believed to be current and accurate as of the
date of publication or revision, but does not constitute a warranty.
Warranty
Limited Warranty
Smiths Medical PM, Inc. (“Seller”) warrants to the original purchaser that the Product, not including accessories,
shall be free from defects in material and workmanship under normal use, if used in accordance with its labeling,
for two years from the date of shipment to the original purchaser.
Seller warrants to the original purchaser that the reusable oximeter sensors supplied as accessories, shall be free
from defects in materials and workmanship under normal use, if used in accordance with its labeling, for one
year from the date of shipment to the original purchaser (USA only).
Disclaimer of Warranties
THE FOREGOING EXPRESS WARRANTY, AS CONDITIONED AND LIMITED, IS IN LIEU OF AND EXCLUDES ALL
OTHER WARRANTIES WHETHER EXPRESS OR IMPLIED, BY OPERATION OF LAW OR OTHERWISE, INCLUDING
BUT NOT LIMITED TO, ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR
PURPOSE.
Seller disclaims responsibility of the suitability of the Product for any particular medical treatment or for any
medical complications resulting from the use of the Product. This disclaimer is dictated by the many elements
which are beyond Seller’s control, such as diagnosis of patient, conditions under which the Product may be used,
handling of the Product after it leaves Seller’s possession, execution of recommended instructions for use and
others.
Conditions of Warranty
This warranty is void if the Product has been altered, misused, damaged by neglect or accident, not properly
maintained or recharged, or repaired by persons not authorized by Seller. Misuse includes, but is not limited to,
use not in compliance with the labeling or use with accessories not manufactured by Seller. This warranty does
not cover normal wear and tear and maintenance items.
Limitation of Remedies
The original purchaser’s exclusive remedy shall be, at Seller’s sole option, the repair or replacement of the
Product. THIS IS THE EXCLUSIVE REMEDY. In no event will Seller’s liability arising out of any cause
whatsoever (whether such cause is based on contract, negligence, strict liability, tort or otherwise) exceed
the price of the Product and in no event shall Seller be responsible for consequential, incidental, or
special damages of any kind or nature whatsoever, including but not limited to, lost business, revenues,
and profits.
Warranty Procedure
To obtain warranty service in the USA, you must request a Customer Service Report (CSR) number from Technical
Service. Reference the CSR number when returning your Product, freight and insurance prepaid, to:
Seller will not be responsible for unauthorized returns or for loss or damage to the Product during the return
shipment. The repaired or replaced Product will be shipped, freight prepaid, to Purchaser.
To obtain warranty information outside of the USA, contact your local distributor.
Keep all original packing material, including foam inserts. If you need to ship the device, use only the original
packaging material, including inserts. Box and inserts should be in original condition. If original shipping
material in good condition is not available, it should be purchased from Smiths Medical PM, Inc.
Damages occurred in transit in other than original shipping containers are the responsibility of the shipper. All
costs incurred returning devices for repair are the responsibility of the shipper.
CE Notice
Marking by the symbol 2 indicates compliance of this device to the Medical Device Directive 93/42/EEC.
Authorized Representative (as defined by the Medical Device Directive):
Smiths Medical International Ltd. Phone: (44) 1923 246434
Colonial Way, Watford, Herts, Fax: (44) 1923 240273
WD24 4LG, UK
Australian Representative:
Smiths Medical Australasia Pty. Ltd. Tel: +61 (0) 7 3340 1300
61 Brandl Street, Eight Mile Plains,
QLD 4113, Australia
Chapter 1: Introduction
• Parts Lists & Circuit Diagrams Contains detailed parts lists and circuit diagrams for the oximeter.
Definition of Symbols
symbol definition
p Type CF equipment
J Date of manufacture
1 Non AP device
6 Caution: Federal (USA) law restricts this device to sale by or on the order of a physician.
x ON/OFF key
a Wave/Trend key
b Menu/Enter key
X AC Power LED
symbol definition
E Input voltage
S Alternating Current
0 Loud speaker
M Battery
w Fuse
F Print
Disposal (EU Countries)
Under the Waste Electrical and Electronic Equipment (WEEE) Directive 2006/96/EC and
implementing regulations, all devices and service items within the scope of the Directive
Z purchased new after August 13, 2005 must be sent for recycling when ultimately becoming
waste. Devices and items must not be disposed of with general waste.
Collect
separately If purchased before that date, they may also be sent for recycling if being replaced on a one-
for-one, like-for-like basis (this varies depending on the country). Recycling instructions to
customers using Smiths Medical products are published on the internet at:
http://www.smiths-medical.com/recycle
Disposal (other countries)
When disposing of this device, its batteries or any of its accessories, ensure that any negative
impact on the environment is minimized. Contact your local waste disposal service and
use local recycling or disposal schemes. Separate any other parts of the equipment where
Y
arrangements can be made for their recovery; either by recycling or energy recovery. The
main batteries are potentially harmful and will require separate disposal according to
manufacturer’s instructions or local regulations.
Note: If applicable, EU, national or local regulations concerning waste disposal must take
precedence over the above advice.
keyword definition
WARNING! Tells you about something that could hurt the patient or hurt the operator.
CAUTION! Tells you about something that could damage the monitor.
NOTE! Tells you other important information.
Warnings
WARNING! This monitor is not for home use.
WARNING! ELECTRICAL SHOCK HAZARD when cover is removed. Unit is not user serviceable.
7 Refer servicing to qualified personnel.
WARNING! Risk of fire. Replace fuses as marked. Refer servicing, including replacement of fuses to
qualified personnel.
WARNING! Patient safety can be compromised by the use of a power cord not supplied by Smiths
Medical PM, Inc. Use only the power cord included with your monitor, or approved by Smiths
Medical PM, Inc.
WARNING! Do not use this device in the presence of magnetic resonance imaging (MR or MRI)
equipment.
WARNING! Operation of this device may be adversely affected in the presence of conducted transients
or strong EM or RF sources, such as electrosurgery and electrocautery equipment, x-rays, and
high intensity infrared radiation.
WARNING! Operation of this device may be adversely affected in the presence of Computed Tomography
(CT) equipment.
WARNING! Any monitor that has been dropped or damaged, should be inspected by qualified service
personnel, prior to use, to insure proper operation.
WARNING! If the accuracy of any measurement is in question, verify the patient’s vital sign(s) by an
alternative method and then check the monitor for proper functioning.
WARNING! In the event that earth ground integrity is lost, the performance of this device and/or other
devices nearby may be affected due to excessive RF emissions.
WARNING! This device must be used in conjunction with clinical signs and symptoms. This device is only
intended to be an adjunct in patient assessment.
WARNING! This device is intended for use by persons trained in professional health care or those who
have access to the oversight of a professional health care provider. The operator must be
thoroughly familiar with the information in this manual before using the device.
WARNING! It is the operator’s responsibility to set alarm limits appropriately for each individual patient
and to verify proper alarm and functional operation before leaving patient unattended.
WARNING! Prolonged use or the patient’s condition may require changing the sensor site periodically.
Change sensor site and check skin integrity, circulatory status, and correct alignment at least
every 4 hours.
WARNING! When attaching SpO2 sensors with Microfoam® tape, do not stretch the tape or attach the
tape too tightly. Tape applied too tightly may cause inaccurate readings and blisters on the
patient’s skin (lack of skin respiration, not heat, causes the blisters).
WARNING! Incorrectly applied sensors may give inaccurate readings. 2 Refer to the sensor insert for
proper application instructions.
WARNING! Use only SpO2 sensors supplied with, or specifically intended for use with, this device.
WARNING! SpO2 measurements may be adversely affected in the presence of high ambient light. If
necessary, shield the sensor area (with a surgical towel, for example).
WARNING! Dyes introduced into the bloodstream, such as methylene blue, indocyanine green, indigo
carmine, patent blue V (PBV), and fluorescein, may adversely affect the accuracy of the SpO2
reading
WARNING! Any condition that restricts blood flow, such as use of a blood pressure cuff or extremes in
systemic vascular resistance, may cause an inability to determine accurate pulse and SpO2
readings.
WARNING! Optical cross-talk can occur when two or more sensors are placed in close proximity. It can be
eliminated by covering each site with opaque material.
WARNING! Remove fingernail polish or false fingernails before applying SpO2 sensors. Fingernail polish
or false fingernails may cause inaccurate SpO2 readings.
WARNING! Tissue damage may result from overexposure to sensor light during photodynamic therapy
with agents such as verteporphin, porfimer sodium and meta-tetrahydroxyphenylchlorin
(mTHPC). Change the sensor site at least every hour and observe for signs of tissue damage.
More frequent sensor site changes or inspections may be indicated depending upon the
photodynamic agent used, agent dose, skin condition, total exposure time or other factors.
Use multiple sensor sites.
WARNING! Using a damaged sensor may cause inaccurate readings, possibly resulting in patient injury
or death. Inspect each sensor. If a sensor appears damaged, do not use it. Use another sensor
or contact your authorized repair center for help.
WARNING! Using a damaged patient cable may cause inaccurate readings, possibly resulting in injury or
death. Inspect the patient cable. If the patient cable appears damaged, do not use it. Contact
your authorized repair center for help.
WARNING! If any of the integrity checks fail, do not attempt to monitor the patient. Use another sensor
or patient cable, or contact the equipment dealer for help.
WARNING! When connecting this monitor to any instrument, verify proper operation before clinical use.
Refer to the instrument’s user manual for full instructions. Accessory equipment connected to
the monitor’s data interface must be certified according to the respective IEC standards, i.e.,
IEC 60950 for data-processing equipment or IEC 60601-1 for electromedical equipment. All
combinations of equipment must be in compliance with IEC 60601-1-1 systems requirements.
Anyone connecting additional equipment to the signal input port or signal output port
configures a medical system, and, therefore, is responsible that the system complies with the
requirements of the system standard IEC 60601-1-1.
WARNING! IEC 60950 approved equipment including the MCP8850B printer must be placed outside of
the “patient environment.” The patient environment is defined as an area 1.5m (4.92 feet)
from the patient.
WARNING! Do not place the monitor in the patient’s bed or crib. Do not place the monitor on the floor.
WARNING! Failure to place the monitor away from the patient may allow the patient to turn off, reset, or
damage the monitor, possibly resulting in the patient not being monitored. Make sure the
patient cannot reach the monitor from their bed or crib.
WARNING! Failure to carefully route the cable from the sensor to the monitor may allow the patient to
become entangled in the cable, possibly resulting in patient strangulation. Route the cable
in a way that will prevent the patient from becoming entangled in the cable. If necessary, use
tape to secure the cable.
WARNING! Under certain clinical conditions, pulse oximeters may display dashes if unable to display
SpO2 and/or pulse rate values. Under these conditions, pulse oximeters may also display
erroneous values. These conditions include, but are not limited to: patient motion, low
perfusion, cardiac arrhythmias, high or low pulse rates or a combination of the above
conditions. Failure of the clinician to recognize the effects of these conditions on pulse
oximeter readings may result in patient injury.
3180 Service Manual 1-5
Chapter 1: Introduction
Cautions
CAUTION! Ensure the device’s AC rating is correct for the AC voltage at your installation site before using
the monitor. The AC rating is set by a switch located on the bottom of the monitor. If the rating
is not correct, set the switch to the correct setting.
CAUTION! Do not autoclave, ehtylene oxide sterilize, or immerse the monitor in liquid. Do not allow
water or any other liquid to spill onto the monitor. Unplug the AC power cord from the
monitor before cleaning or disinfecting the monitor. Evidence that liquid has been allowed to
enter the monitor voids the warranty.
CAUTION! Should the device become wet, wipe off all moisture and allow sufficient time for drying
before operating.
CAUTION! Do not autoclave, ethylene oxide sterilize, or immerse the sensors in liquid. Unplug the sensor
from the monitor before cleaning or disinfecting.
CAUTION! The monitor should be operated from its internal power source if the integrity of the
protective earth conductor is in doubt.
CAUTION! Follow local governing ordinances and recycling instructions regarding disposal and recycling
of device components.
CAUTION! Pressing front panel keys with sharp or pointed instruments may permanently damage the
keypad. Press front panel keys only with your finger.
CAUTION! Blocking the ventilation holes on the monitor’s rear panel can prevent air circulation inside
the monitor. Leave an air gap behind the monitor to allow the air to circulate through the
ventilation holes.
CAUTION! Device batteries are not user replaceable. This unit may contain a Lead Acid and/or NiMH
battery. Disposal of such batteries should be conducted in accordance to local or federal
guidelines. Smiths Medical PM, Inc. can not dispose of these batteries.
CAUTION! Chemicals used in some cleaning agents may cause brittleness of plastic parts. Follow
cleaning instructions in this manual.
Notes
NOTE! Obstructions or dirt on the sensor’s red light or detector may cause a sensor failure. Make sure
there are no obstructions and the sensor is clean.
NOTE! If an operator changes the low SpO2 alarm limit to a value less than 80, and a power down - power
up sequence takes place, a minimum value of 85 takes the place of the operator entered value.
NOTE! The pitch of the pulse beep is determined by the SpO2 value. The higher the SpO2 value, the
higher the pulse beep pitch, and vice versa.
NOTE! Performance and safety test data are available upon request.
NOTE! High priority alarms can be tested by setting the alarm limits outside the patient’s measurements.
Be sure to restore alarm limits after testing.
System Context
This document describes the operational electronic hardware for the BCI® 3180 main board, 71952B1. This
document is meant to supplement the system schematics 71952S1-1-8 (Please see 71950SRS, 3180 Oximeter
Requirements Summary and 71960SDR, Software Design Requirements, for additional information on the
monitor and it’s hardware design).
Power Entry
The BCI® 3180 takes power directly from a line cord. There is no external “brick”. The line input is fused and
routed through an EMI filter consisting of inductor L4 and capacitor C6. TR1 is a transorb that limits voltage
spikes coming in from the line.
The voltage selector switch SW3 is used to select between 110VAC and 220VAC operation. This switch is
accessible by the user through the bottom of the case of the 3180. The transformer T1, bridge D1 and the
capacitors form a simple unregulated power supply capable of delivering 24VDC at 1.0A.
Battery Charger
The BCI® 3180 uses an internal battery to provide power during line interruptions. The circuit has been designed
to support two different battery technologies: Lead Acid and NiMH. The board is built up to specifically support
one of the two battery technologies.
Both battery charger designs use the LT1074 (U1) to deliver charge current to the battery and power to the
system. The battery (-) is tied to system ground through a resister (R26). When the charger is active the voltage
drop across this resister is used to control the charge current to the battery independent of the current drawn
from the supply to power the oximeter.
For a Lead Acid system the charge current is controlled with U3B by regulating a 360mV drop across R26. During
normal charging R26 is in parallel with R22\R28 for a charge current of 480mA. The peak battery voltage is
detected by the comparitor at U8A. When the peak voltage occurs this amp is latched low with diode D6. This
then turns off Q4 and releases the R22\R28 resisters from the sense line. The charge current drops to a trickle
value of 4.4mA. Q3 is used to generate a CHARGE signal to the processor. Q20 and D19 limit the supply voltage
to U8A so that U8, Q3 and Q4 are not damaged during high line conditions.
The NiMH system substitutes the Maxim MAX713 for the descrete design and provides the same control of
charge current via R26. This chip also provides temperature compensation of the the charge cycle using T3, T4
and a thermister located inside the NiMH battery pack.
The battery disconnect switch SW1 is used to positively disconnect the battery from the system during shipment
and long storage intervals.
CPU Core
The system clock for the BCI® 3180 is generated by X1, U32 and R97. This 18.432Mhz signal is fed into the
Cypress Semiconductor PPL (U86). This chip provides a frequency modulated output that has a center frequency
of 18.432Mhz but is spread out over a 25Khz band to limit EMI spikes. Both clocks are routed to the Altera
EPM7064 FPGA (U14). The processor gets the modulated clock and the oximeter PIC processor gets a divided
down version of the unmodulated clock. A separate modulated clock is provided to the LCD display at P12. In
this way the oximeter timing is precise while EMI is reduced.
The FPGA contains all of the “glue” logic in the system as well as an audio DAC for controlling the audio
volume. The Zilog ZS180 8 bit microprocessor (U8) contains several internal peripheral elements including
two timers, two UARTS, an MMU, two DMA channels and a Synchronous Serial Port. This is supplemented with
up to 512Kbytes of ROM (U9) and 128Kbytes of RAM (U10). The buffer at U18 and header P1 provide the link
between the processor and the LCD display. The signals on header P8 are used to reprogram the FPGA during
development or service updates. The configuration switch SW2 is not currently used.
Oximeter Section
The Oximeter front end circuitry is controlled by the Microchip PIC16F873 (U85). Connector P11 is used to
reprogram the PIC on the board during development. The following signals are used by the PIC to control the
oximeter hardware and to communicate with the Z180 host processor.
SIGNAL DESCRIPTION
AD-DO 3 wire serial interface for analog to digital converter U25.
/AD-CS
ADCLK Clocks data into digital potentiometer chip U24 and from A\D converter chip U25.
POT-LD POT-LD is used to select digital potentiometer chip U24.
POT-DI Serial Data Input to digital potentiometer chip U24
/AMB-LATCH Signal informing processor about excess of ambient light.
BASE Keeps integrator U23B from saturating during data acquisitions
CAP-GND Signal used to short blocking capacitor C76 to ground.
FET-ON Used to control MOSFET H-bridge Q13 and Q14, which powers probe LED’s.
INTGRAT, Signals used to enable integrator U23B.
IR-DRV Used to control MOSFET H-bridge Q13 and Q14, which powers probe LED’s.
PRB-DET PRB-DET is used to inform processor if probe is plugged in.
/PRB-FAULT Signal used to inform processor about probe cable problem.
RED-DRV Used to control MOSFET H-bridge Q13 and Q14, which powers probe LED’s
RST-INT Resets integrator U23B by shorting capacitor C85.
OX-CLK 3 wire synchronous serial interface for communication to the Z180.
OX-DO
OX-DI
/OX-IRQ Interrupt line to Z180 used to start data comm. cycle
The Analog to Digital Converter (ADC) U25 is a 12 bit, successive approximation sampling ADC. The ADC +IN
signal input comes from TP14. VREF is used here as a pseudo ground. The ADC is configured as a differential
input converter with the -IN input connected to VREF. This allows the +IN signal to range from 2.5V to 5V.
The ADC transfers data with a 3 wire serial interface. AD-CLK synchronizes the data transfer with each bit being
transmitted on the falling AD-CLK edge and captured on the rising AD-CLK edge. /AD-CS must be held low
when the ADC transmits data.
LED Drive circuitry is used to turn on and off the Oximetry probe LED’s, precisely controlling the current through
them. The second half of the variable potentiometer U24 is used to generate voltage between 0 and 2.5VDC.
U24 is controlled, serially, by the PIC processor. Operational amplifier U27B together with transistor Q16 create
a constant current sink. The current is proportional to the voltage generated by variable potentiometer. The
H-bridge Q13-Q14 is used to activate either LED, red or infrared, inside the Oximeter probe. The following table
describes the states of the H-bridge:
The “Open circuit” state is used to check for a possible probe cable fault. In case of probe cable fault, one of the
LED wires can short to the ground shield causing high current through the probe LED. To prevent this, before
turning any LED on, the processor checks the cable by “floating” the H-bridge. If the cable is shorted, CON3
and CON2 connections are pulled low, and comparator U28B generates /PRB_FAULT signal, which causes the
processor to shut down the LED excitation cycle and generate the appropriate message.
The differential photo amplifier formed by U21 and U20B, converts the photodetector’s current output to a
voltage at TP11 (VAMB). Amplifier U20A offsets the signal at TP11 allowing a wider signal range for the negative-
going pulses at TP12.
Comparator U28A is used to inform the microprocessor if U20B is saturated by an excess of ambient light. VAMB
is passed through blocking capacitor C76 to remove the signal’s DC component. The signal is then buffered
and amplified by U23A. Gain is defined by the first channel of digital potentiometer U24. The output of U23A is
routed to the integrator-filter-offset circuitry U23B, controlled by analog switch U22. The output of the integrator
is then routed to the ADC for measurement.
• The waveforms are measured with the finger probe attached to the oximeter with finger or piece of paper
in probe.
TP13: TP6:
A= 0.2 TB= 0.4ms A= 1 V TB= 0.4ms
0V 0V
TP1: TP2:
A= 1 V TB= 0.4ms A= 1 V TB= 0.4ms
0V 0V
TP3:
A= 1 V TB= 0.4ms
0V
• Language: Selects the language for all displayed text: English, French, Spanish, German, Italian, or
Portuguese.
• Trend Sample: This selects the basic sample time for trend data record storage. The sample time
determines the maximum number of records stored and consequently the maximum storage time.
“4 sec/12 hrs” means 4 second sample time for 12 hours maximum storage. “12 sec/36 hrs” means 12
second sample time for 36 hours maximum storage. “30 sec/90 hrs” means 30 second sample time for
90 hours maximum storage.
NOTE: If the Trend Sample time is changed, the monitor will turn off after quitting the Service menu. This
forces the monitor to reconfigure memory at power up and the trends will be cleared.
• Default?: “NO” means the monitor will NOT reset to factory default settings. “YES” means the monitor
will reset to factory default settings as shown below:
Alarms:
SpO2 High: Off
SpO2 Low: 85
Pulse High: 150
Pulse Low: 45
Alarm Volume: 3
Oximeter:
Pulse Volume: 3
Pleth Speed: 32
Averaging: 8/8
Backlight: On
Trend Scales:
SpO2: 50-100
Pulse: 0-100
Trend Sample Interval: 12 secs
Trend Time Span: 30 minutes
Print Format: IR Patient Data
Print Time: 60 secs
Language: English
Date Format: USA
• Alarm Audio: “Both” means alarms can be silenced indefinitely by pressing and holding the ALARM
AUDIO OFF/PAUSE key for 3 seconds, and alarms can be silenced for 2 minutes by pressing the ALARM
AUDIO OFF/PAUSE key momentarily. “Off” means 2-minute alarm silence is not allowed. In this case, only
permanent alarm silence is available. “Pause” means permanent alarm silence is not allowed. In this
case, only 2-minute alarm silence is available.
3180 Service Manual 3-1
Chapter 3: Field Service Menu
Chapter 4: Troubleshooting
Appendix
Australian Representative:
Smiths Medical Australasia Pty. Ltd. Tel: +61 (0) 7 3340 1300
61 Brandl Street, Eight Mile Plains,
QLD 4113, Australia
H Manufactured By
Smiths Medical PM, Inc.
N7W22025 Johnson Drive
Waukesha, WI 53186-1856 USA