HS30

Data Sheet
V1.00

Oct. 19, 2018


Rev03

Samsung Medison CO., Ltd., 145 Pangyoyeok-ro, Bundang-gu, Seongnam-si, Gyeonggi-do, South Korea 13530
TEL:(82)2-2194-1508 FAX:(82)-32-778-3199
www.samsungmedison.com
Business Development Group
www.samsungmedison.com

Table of Contents

1. SPECIFICATION SUMMARY ........................................................................................................... 1

Physical Specifications ........................................................................................................................................ 1


Imaging Modes ................................................................................................................................................... 1
Focusing .............................................................................................................................................................. 1
Probe connections .............................................................................................................................................. 1
Monitor ............................................................................................................................................................... 1
ECG...................................................................................................................................................................... 1
Image Storage ..................................................................................................................................................... 1
Rear Panel input/output connections ................................................................................................................. 1
Auxiliary .............................................................................................................................................................. 1
User Interface ..................................................................................................................................................... 1
Electrical Parameters .......................................................................................................................................... 2
Pressure Limits .................................................................................................................................................... 2
Humidity Limits ................................................................................................................................................... 2
Temperature Limits............................................................................................................................................. 2

2. GENERAL SPECIFICATION ............................................................................................................. 3

Physical Specification .......................................................................................................................................... 3


Console Design .................................................................................................................................................... 3
Main Monitor ...................................................................................................................................................... 3
Control Panel ...................................................................................................................................................... 3
PC ........................................................................................................................................................................ 3
Electrical Specifications....................................................................................................................................... 3

3. SYSTEM SPECIFICATION ................................................................................................................ 4

Applications ........................................................................................................................................................ 4
Presets ................................................................................................................................................................ 4
Operation Mode ................................................................................................................................................. 4
Display Mode ...................................................................................................................................................... 4
Transducer Types ................................................................................................................................................ 5
System Standard Features .................................................................................................................................. 5
System Options ................................................................................................................................................... 5

CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
www.samsungmedison.com

Display ................................................................................................................................................................. 6
Language ............................................................................................................................................................. 6
Operating Environment ...................................................................................................................................... 6

4. PROCESSING ....................................................................................................................................... 7

Data Processing................................................................................................................................................... 7
Pre-Processing .................................................................................................................................................... 7
POST-Processing ................................................................................................................................................. 8

5. CONNECTIVITY .................................................................................................................................. 9

DICOM ................................................................................................................................................................. 9
IHE ....................................................................................................................................................................... 9
Peripheral interface ............................................................................................................................................ 9

6. SCANNING PARAMETERS............................................................................................................ 10

2D mode............................................................................................................................................................ 10
M Mode ............................................................................................................................................................ 10
Color Mode ....................................................................................................................................................... 10
PWD Mode ........................................................................................................................................................ 11
CWD Mode ........................................................................................................................................................ 11
PD Mode ........................................................................................................................................................... 11
ElastoScan Mode............................................................................................................................................... 12

7. TRANSDUCERS ............................................................................................................................... 13

Linear ................................................................................................................................................................ 13
Convex .............................................................................................................................................................. 13
Endocavity......................................................................................................................................................... 13
Phased Array ..................................................................................................................................................... 14
Pencil................................................................................................................................................................. 14

8. MEASUREMENT.............................................................................................................................. 15

Caliper ............................................................................................................................................................... 15
Abdomen .......................................................................................................................................................... 15
Cardiac .............................................................................................................................................................. 15
Carotid .............................................................................................................................................................. 16

CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
www.samsungmedison.com

UE Artery ........................................................................................................................................................... 16
UE Vein.............................................................................................................................................................. 16
LE Artery............................................................................................................................................................ 16
LE Vein .............................................................................................................................................................. 16
Gynecology ....................................................................................................................................................... 17
Obstetrics .......................................................................................................................................................... 17
Fetal Heart ........................................................................................................................................................ 17
Urology ............................................................................................................................................................. 17
Breast ................................................................................................................................................................ 17
MSK ................................................................................................................................................................... 18
Thyroid .............................................................................................................................................................. 18
Testicle .............................................................................................................................................................. 18
Superficial ......................................................................................................................................................... 18
Pediatric ............................................................................................................................................................ 18

9. SAFETY / EMC ................................................................................................................................. 19

Classifications.................................................................................................................................................... 19
Applied standards ............................................................................................................................................. 19
Acoustic output management .......................................................................................................................... 19
Anti-virus solution ............................................................................................................................................. 19

CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
www.samsungmedison.com

MONITOR
SPECIFICATION SUMMARY
 Main Monitor
PHYSICAL SPECIFICATIONS - Resolution: 1,920 x 1,080

- 21.5 Inch LED Monitor


 Height: 1,393 mm (with monitor)

 Width: 519 mm
ECG
 Depth: 664 mm
 USB Type (Type CF)
 Weight: Approx. 47.8 Kg (without accessories)

 Weight: Approx. 52 Kg (with Safe Working Load)


IMAGE STORAGE

IMAGING MODES  Maximum 45,000 Frames for Cine memory

 Maximum 14,000 Lines for Loop memory


 2D-Mode
 Image filing system
 M-Mode

 Color M-Mode
REAR PANEL INPUT/OUTPUT CONNECTIONS
 Anatomical Mode
 Audio Output Port (Right/Left)
 Color Doppler Mode
 VGA monitor
 Pulsed Wave (PW) Spectral Doppler Mode
 S-Video Output
 Continuous Wave (CW) Doppler Mode
 LAN
 Tissue Doppler Imaging (TDI) Mode
 USB Port
 Tissue Doppler Wave (TDW) Mode
 HDMI Output
 Power Doppler (PD) Mode

 ElastoScan Mode AUXILIARY


 Dual Mode
 DVD Multi-Drive
 Quad Mode
 Digital B/W Video Printer
 Combined Mode
 Digital Color Video Printer
 Simultaneous Mode
 USB Printer
 Zoom Mode
 DVD Recorder
 S-Flow Mode
 Foot switch (IPX8)
FOCUSING  USB Flash Memory Media

 USB HDD
 Transmit focusing, maximum of eight points (four points
 USB ECG
simultaneously selectable)
 Monitor
 Digital dynamic receive focusing (continuous)

USER INTERFACE
PROBE CONNECTIONS
 English, French, Italian, German, Spanish, Russian,
 2 or 3 Probe Connectors (Optional)
Chinese, Portuguese (Brazilian)
 CW Probe Connector (Optional)

HS30 DATA SHEET V1.00 Rev01 Page 1


www.samsungmedison.com

ELECTRICAL PARAMETERS

 100-240 VAC, 620 VA, 50/60 Hz

PRESSURE LIMITS

 Operating: 700 hPa to 1,060 hPa

 Storage: 700 hPa to 1,060 hPa

HUMIDITY LIMITS

 Operating: 30 % to 75 %

 Storage & Shipping: 20 % to 90 %

TEMPERATURE LIMITS

 Operating: 10 °C to 35 °C

 Storage & Shipping: -25 °C to 60 °C

HS30 DATA SHEET V1.00 Rev01 Page 2


www.samsungmedison.com

CONTROL PANEL
GENERAL SPECIFICATION
 Alpha-Numeric KBD
PHYSICAL SPECIFICATION  Analog TGC

 6 User Keys
 Height: 1,393 mm (with monitor)
 Tri-Status backlit
 Width: 519 mm
 4 Probe Holders
 Depth: 664 mm

 Weight:
PC
- Approx. 47.8 Kg (without accessories)
 Main Processer: AMD Bald Eagle RX-425BB
- Approx. 52 Kg (with safe working load)
 Main Memory: 8 GB
CONSOLE DESIGN  Built-in HDD : 500GB

 2 or 3 Active Probe Ports (Optional)


ELECTRICAL SPECIFICATIONS
 4 Swivel Wheel Cart Based Type
 Frequency: 50/60 Hz
 Rotation&Tilt Monitor Arm
 Voltage: 100 ~ 240 VAC
 Built-in Printer Storages
 Power Consumption: Max. 620 VA with Peripherals
 Ergonomic Operation Panel
 Heat Dissipation: 2,729.7 [BTU/h]
 Alpha-Numeric Keyboard
 System Noise: 40 dBA
 Analog TGC
 Built-in Equipotential Circuit
 Trackball

 Probe Holder / Gel Holder

 Front Handles

 Integrated PC Module

 Integrated HDD

 Windows Embedded Standard 7

 Gel Warmer (Optional)

 ECG (Optional)

MAIN MONITOR

 21.5 Inches High Resolution LED Monitor

 Resolution: 1,920 x 1,080 (16:9)

 Number of Color: 16.7 M

 Brightness Adjustment

 Interactive Dynamic Software Menu

 Articulated Monitor Arm

- Rotation: +/- 135°


- Tilt: -10° / 75°

HS30 DATA SHEET V1.00 Rev01 Page 3


www.samsungmedison.com

SYSTEM SPECIFICATION  Testicle

 TCD

APPLICATIONS  Uterus

 Venous
 Abdomen
st
 1 Trimester
 Cardiac
nd
 2 Trimester
 Gynecology
rd
 3 Trimester
 MSK

 Obstetrics OPERATION MODE


 Pediatric
 B-Mode (2D)
 Small Parts
 Color Doppler Mode (C)
 Urology
 Pulse Wave Doppler (PWD)
 Vascular
 Continuous Wave Doppler(CWD): Steered / Static

PRESETS  Power Doppler Mode (PD)

 S-Flow Mode
 Abdomen
 M-Mode (M)
 Adult Echo
 Anatomical M Mode
 Adnexa
 Single/Dual/Quad Mode
 Aorta
 TDI/TDW
 Aortic Arch
 ElastoScan Mode
 Arterial

 Bladder DISPLAY MODE


 Bowel
 Dual Mode
 Breast
- B+B, B+B/C, B+B/PD, B+B/S-Flow
 Carotid
- ElastoScan + ElastoScan
 Deep
 Dual Live Mode
 Fetal Heart
- B+B, B+B/C, B+B/PD, B+B/S-Flow
 General
- B+ElastoScan
 Neo Head
 Real-Time Triplex Mode (Simultaneous Mode)
 NT
- B+C+PW, B+PD+PW, B+S-Flow+PW, B+TDI+TDW
 Ped Abd
 Duplex, Triplex Mode
 Ped Echo
- B+C, B+M, B+PW, B+PD, B+S-Flow, B+CW,
 Ped Hip
B+C+PW, B+C+CW, B+C+M, B+ElastoScan, B+TDI,
 Prostate
B+TDW
 Renal
 Quad Mode
 Spine
- Combinations of B/B, B/C, B/PD and B/S-Flow,
 Superficial
ElastoScan
 Thyroid

HS30 DATA SHEET V1.00 Rev01 Page 4


www.samsungmedison.com

 Zoom Mode  Post Processing

- Write Zoom / Read Zoom / Pen zoom/ Panning  Digital Calipers / Measurement

 Needle Mate+  Cine Memory

 Panoramic - Capacity: 500 MB

 Trapezoid - Cine loop: Max. 14,000 Lines

- Image storage: Max. 45,000 Frames


TRANSDUCER TYPES
 QuickScan
 Linear Array: LN5-12, L5-12/50  Report Package
 Curved Array: C2-5, C2-8  Body Marker
 Endo-Cavity: EVN4-9, ER4-9  System Boot up: Max. 150 Sec
 Micro-Convex Array: CF4-9  Probe Change: 2-3 Sec
 Phased Array: PN2-4  User Programmable Preset : Over 30 Presets
 Pencil: DP2B  User Programmable Key: 6 Keys

 SonoView
SYSTEM STANDARD FEATURES
 Data Backup / Restore
 Hybrid Full Digital Beam-forming  Image Exporting and Importing
 Frequency Range: 1 ~ 18MHz  PW Velocity Range: 0.1cm/s ~ 8.8m/s
 Displayed Imaging Depth (Probe dependent)  CW Velocity Range: 1cm/s ~ 19.3m/s
- Minimum Depth of Field: 2cm
SYSTEM OPTIONS
- Maximum Depth of Field: 38cm

 Number of Focal Points: 1 ~ 4  Auto IMT+


 Transmission Focal Zone Position selection  Cardiac Measurement
- 1 ~ 8 Focal Points Selectable  CW Function
 (Probe and Application dependent)  DICOM 3.0
 Continuous Dynamic Receive Focus / Aperture  ElastoScan
 Multi-frequency / Wideband Technology  EZ Exam+
 Frequency Compounding (FSI)  EZ Assist
 ClearVision  Mobile Export
 256 Shades of Gray  Needle Mate+
 System Internal Dynamic Range: 256  Panoramic
 Maximum Frame Rate  MultiVision
- 2,000 fps (Hz)  Strain+
 Maximum Color Frame Rate  ECG (AHA / IEC)
- 400 fps (Hz)  Foot Switch
 Image Reverse: Right/Left, Up/Down  Gel Warmer
 Image Rotation: 90⁰, 180⁰, 270⁰  Printer Tray (Large / Small)
 Pre Processing  2P Connector PSA

HS30 DATA SHEET V1.00 Rev01 Page 5


www.samsungmedison.com

 2P Connector PSA (with Pencil Probe Port)  Frequency

 3P Connector PSA  Gain

 3P Connector PSA (with Pencil Probe Port)  Dynamic Range

 Map
DISPLAY
 Frame Average
 Application  Power
 Preset  ClearVision Index
 Mode  MultiVision Index
 Date: 3 types (Selectable)  Gray Bar
- YYYY-MM-DD  Acoustic Index: TIs, TIb, Tic
- MM-DD-YYYY  Mechanical Index: MI
- DD-MM-YYYY  Caliper & Measurement Result
 Time: 2 types (Selectable)  Indicator
- 24 hours  Pointer
- 12 hours  Body Marker
 Patient (General Information)  ROI Position / ROI Size
- Patient ID  Wall filter
- Patient Name (First, Middle & Last)  Zoom / Panning
- Gender: Female, Male, Other  Biopsy Guide Line (Probe dependent)
- Birth / Age
LANGUAGE
- Accession Number

- Diag. Physician  Display Language

- Ref. Physician - English, French, Italian, German, Spanish, Russian,

- Operator Chinese, Portuguese (Brazilian)

- Indication  Input Language

- Study Information - English, French, German, Russian, Nordic (Norwegian,

- E-mail Finnish, Swedish, Danish)

 Gestational Age: LMP/EDD/GA


OPERATING ENVIRONMENT
 Institute

 Operator  Temperature: 10°C ~ 35°C

 Probe Name  Humidity: 30% to 75%

 Probe Orientation  Pressure: 700 ~ 1,060 hPa

 Depth / Width

 Focal Zone

 Focal Number

 TGC Line
 FPS (Hz)

HS30 DATA SHEET V1.00 Rev01 Page 6


www.samsungmedison.com

PROCESSING - Write Zoom

- MultiVision (Probe Dependent)

DATA PROCESSING - Beam Steering (Probe Dependent)

- Trapezoid (Probe Dependent)


 System Processing Channel: 286,720
- Free Angle Plane
 Raw Data Image Analysis
 PW Mode
 Cine
- Filter
- Function: save / review / play / stop / pause / export /
- Frequency
Trim Start / Trim End
- Gain
 Clipboard: displays thumbnail images of the acquired
- Power
data for the current exam
- PRF (Scale)
 Enlarged Preview of the image
- Sample Volume Angle
 Image Archive / Connectivity
- Sample Volume Position
 Image format: AVI, MPEG, JPEG, BMP, TIFF, DICOM
 CW Mode
 Image Viewer (Sonoview)
- Sample Rate
 Measurements, Calculations and Annotations on CINE
- Filter
Playback
- Gain
 Number of Image Storage (built-in SSD): max.
- Power
350,000 images (RAW format)
- Sample Volume Angle
 Image Preview
- Sample Volume Position
 Cine Image Preview
 Color Doppler / Power Doppler mode
 Recalling Image from the Clipboard
- Filter
 Scrolling Timeline Memory
- Frame Average
 Start and End Frame Selections for Loop Playback
- Frequency
PRE-PROCESSING - Gain

- Line Density
 B/M-Mode
- Power
- Dynamic Range
- PRF (Scale)
- Frame Average
- Smoothing
- Frequency
- Sensitivity
- Gain
- Steer Angle
- Harmonic (Probe dependent)
 ElastoScan Mode
- Pulse Inversion Harmonic (Probe dependent)
- Frame Average
- Line Density
- Frequency
- Power
- Line Density
- Reject
- Scan Area

- TGC

HS30 DATA SHEET V1.00 Rev01 Page 7


www.samsungmedison.com

POST-PROCESSING - Blending Level

- Enhancement
 B-Mode

- Chroma Map

- Gray Map

- Image Size

- Read Zoom

- ClearVision

- Sweep Speed

 M-Mode

- Chroma Map

- M Mode Map

- Read Zoom
- Sweep Speed

 PW / CW Mode

- Base line

- Chroma Map

- Doppler Map

- Invert

- Read Zoom

- Sound

- Trace Direction

- Trace Method

 Color Doppler / Power Doppler Mode

- Balance

- Baseline

- Chroma Map

- Color Map

- Hide Color

- Invert

- Read Zoom

 3D Mode

- Freedhand 3D

 ElastoScan Mode

- E-Gain

- Contrast
- Color Map

- Alpha Blending

HS30 DATA SHEET V1.00 Rev01 Page 8


www.samsungmedison.com

CONNECTIVITY - Digital BW Video Printer: Sony UP-D897, Sony UP-

D898MD, Sony UP-X898MD, Mitsubishi P95DE, Mitsubishi

DICOM P95DW, Mitsubishi P95D

- Digital Color Video Printer: Sony UP-D25MD,


 DICOM 3.0
Mitsubishi CP30DW
 DICOM Media
- USB Line Printer: Samsung CLP-620NDK, ML-2950
 DICOM Performed Procedure Step (PPS)

 DICOM Print

 DICOM Storage

 DICOM Storage Commitment (SC)

 DICOM Structured Reporting (SR)

 DICOM Verification

 DICOM Worklist
 Gray Scale Converting

 Multi Frame

 Single Frame

 Transfer Mode

- Send after acquisition

- Send on end exam

- Send manually

 VOI LUT Setup

IHE

 Scheduled Workflow (SWF)

 Patient Information Reconciliation (PIR)

 Portable Data for Imaging (PDI)

 Evidence Documents (ED)

PERIPHERAL INTERFACE

 Audio out L/R

 D-SUB output

 S-Video output

 HDMI output

 USB 2.0 (6 ports)

 Ethernet 10/100/1000BASE-T
 Foot Switch: USB 2.0 (IPX 8)
DVD Recorder: LG GP60NB50 - Recording only

 Printers

HS30 DATA SHEET V1.00 Rev01 Page 9


www.samsungmedison.com

 Scan Area: 40 ~ 100%


SCANNING PARAMETERS
 Zoom

- Read Zoom: 100 ~ 800 %


2D MODE
- Write Zoom
 Angle Steering (Linear probes only)
 Panning
- LN5-12: -8, -4, 0, 4, 8°
 Free Angle Plane
- L5-12/50: -12, -7, 0, 7, 12°

 Chroma Map: Off, 1 ~ 11 M MODE


 Cine Play: On, Off
 Chroma Map: Off, 1 ~ 11
 Cine Speed: 6, 12, 25, 50, 100, 150, 200, 300
 Display format
 Depth:
- M-mode only
- Convex: 5~38cm
- Up/down, Side by side
- Micro Convex: 3~18cm
- Size: 50/50, 70/30, 30/70
- Linear: 2~14cm
 Dynamic Range: 30 ~ 256 (Step 2)
- Endo: 3~18cm
 Gain: 0 ~ 100
- Phased: 5~30cm
 M Mode Map: 1 ~ 12
 Dual Live
 Power: 2 ~ 100
 Dynamic Range: 30 ~ 256 (Step 2)
 QuickScan: On, Update, Off
 Flip: L/R, U/D
 Sweep Speed
 Focus Number: 1 ~ 4
 Color M
 Frequency Compounding
 Anatomical M
 Frequency: 3 ~ 5 steps (Probe Dependent)

- Pen, Gen1, Gen2, Res1, Res2 COLOR MODE

 Gain: 0 ~ 100  Balance: 0 ~ 16


 Gray Map: 1 ~ 12  Baseline: -8 ~ 8
 Harmonic: On, Off  Color Map: 1 ~ 12
 Image Size: 70 ~ 100%  Line Density: Low, Medium, High
 Line Density: Low, Medium, High  Dual Live: On, Off
 Number of TGC Level: 8  Sensitivity: 0 ~ 5
 Frame Average: 0 ~ 9  Frame Average: 0 ~ 5
 Power: 2 ~ 100  Frequency: 2 steps
 Pulse Inversion Harmonic: On, Off (Probe dependent)  Gain: 0 ~ 100
 QuickScan: On, Update, Off  Hide Color: On, Off
 Reject Level: 0 ~ 30  Invert: On, off
 MultiVision Index: Off, Low, Medium, High  Power: 2 ~ 100
 ClearVision Index: Off, 1 ~ 5  PRF: 0.1kHz ~ 19.5kHz (Probe dependent)
 Trapezoid: On, Off (Linear Probes only)  Sensitivity: 0 ~ 5

HS30 DATA SHEET V1.00 Rev01 Page 10


www.samsungmedison.com

 Smoothing: 0 ~ 7  Display Format: Up/down, Side by side, Doppler Only

 Steer Angle: -8, -4, 0, 4, 8°  Display Size: 70/30, 50/50, 30/70

 Velocity  Doppler Map: 1 ~ 12

 Filter: 1 ~ 4  Dynamic Range: 30 ~ 256 (Step2)

 Vel + Variance Map  Gain: 0 ~ 100

 Invert: On, Off


PWD MODE
 Power: 2 ~100
 Auto Calc: Off, Live, Frozen  Sample Rate: 1.8kHz ~ 57kHz (probe dependent)
 Base Line: -8 ~ 8  QuickScan: On, update, Off
 Chroma Map: Off, 1 ~ 11  Sound: 0 ~ 100
 Display format: Up/down, Side by side, Doppler Only  Angle Correction: -80° ~ 80°
 Display Size: 70/30, 50/50, 30/70  SV Position Control
 Doppler Map: 1 ~ 12  Quick Angle: -60°, 0°, 60°
 Dynamic Range: 30 ~ 256 (Step 2)  Sweep Speed: 15 ~ 117 mm/s
 Frequency: 2 Steps (Probe dependent)  Trace
 Gain: 0 ~ 100 - Method: Off, Mean, Max
 Invert: On, Off - Direction: Both, Above, Below
 Power: 2 ~100  Filter: 1 ~ 4
 PRF: 1.0 ~ 22.5 kHz (Probe dependent)
PD MODE
 QuickScan: On, Update, Off

 Simultaneous: On, Off  Balance: 0 ~ 16 step

 Sound: 0 ~ 100  Color Map: 1 ~ 12

 Angle Correction: -80° ~ 80°  Line Density: Low, Medium, High

 SV Position control  Dual Live: On, Off

 SV Size: 0.5 ~ 25mm  Filter: 1 ~ 4

 Quick Angle: -60°, 0°, 60°  Frame Average: 0 ~ 5 step

 Sweep Speed: 15 ~ 117 mm/s  Frequency: 2 steps (probe dependent)

 Trace  Gain: 0 ~ 100

- Method: Off, Mean, Max  Hide Color

- Trace Direction: Both, Above, Below  Invert: On, Off (S-Flow™ only)

 Update  Power: 2 ~ 100

 Filter: 1 ~ 4  PRF: 0.1 ~ 19.5 kHz (Probe dependent)

 Sensitivity: 0 ~ 5
CWD MODE
 Smoothing: 0 ~ 5
 Auto Calc.: Off, Live, Frozen  Steer Angle: -8, -4, 0, 4, 8
 Base line: -8 ~ 8  Filter: 1 ~ 4
 Chroma Map: Off, 1 ~ 11

HS30 DATA SHEET V1.00 Rev01 Page 11


www.samsungmedison.com

ELASTOSCAN MODE

 Line Density: Low, Medium, High

 Invert: On, Off

 Dual Live: On, Off

 Frequency

 Gain: 0 ~ 100

 Contrast: 0 ~ 100

 Frame Average: 0 ~ 100

 Color Map: 1 ~ 5

 Alpha Blending: On, Off

 Blending Level: 0 ~ 100

 Enhancement: 0 ~ 100

HS30 DATA SHEET V1.00 Rev01 Page 12


www.samsungmedison.com

C2-5
TRANSDUCERS
 Center Frequency: 3.35 MHz

 Band Width : 2 ~ 5 MHz


LINEAR
 Radius of curvature :39.64 mm
LN5-12
 Field of view : 75°
 Center Frequency: 8 MHz  Number of elements : 128
 Band Width : 5 ~ 12 MHz  Biopsy Guide : Available
 Radius of curvature : Flat  Application : Abdomen, MSK, OB, GYN, Pediatric, Vascular,
 Field of view : 38.1 mm Urology
 Number of elements : 128  Safety Class: BF
 Biopsy Guide : Available5
CF4-9
 Application : Abdomen, MSK, Small Parts, Vascular, OB,
 Center frequency : 5.65MHz
GYN, Pediatric
 Band Width: 4 ~ 9 MHz
 Safety Class: BF
 Radius of curvature : 14 mm
L5-12/50  Field of view : 92 °

 Center Frequency: 7.3 MHz  Number of elements : 128

 Band Width : 5 ~ 12 MHz  Biopsy Guide : Not available

 Radius of curvature : Flat  Application : Abdomen, MSK, OB, GYN, Pediatric, Vascular,

 Field of view : 52 mm Urology

 Number of elements : 128  Safety Class: BF

 Biopsy Guide : Available


ENDOCAVITY
 Application : Abdomen, MSK, Small Parts, Vascular, OB,

GYN, Pediatric EVN4-9

 Safety Class: BF  Center frequency : 6.65MHz

 Band Width: 4 ~ 9 MHz


CONVEX  Radius of curvature : 10.073 mm

C2-8  Field of view : 148.092 °

 Center Frequency: 4.7 MHz  Number of elements : 128

 Band Width : 2 ~ 8 MHz  Biopsy Guide : Available

 Radius of curvature : 51.071 mm  Application : OB, GYN, Urology

 Field of view : 68.176°  Safety Class: BF

 Number of elements : 128


ER4-9
 Biopsy Guide : Available
 Center frequency : 6.65MHz
 Application : Abdomen, MSK, OB, GYN, Pediatric, Vascular,
 Band Width: 4 ~ 9 MHz
Urology
 Radius of curvature : 10.073 mm
 Safety Class: BF
 Field of view : 148.092 °

HS30 DATA SHEET V1.00 Rev01 Page 13


www.samsungmedison.com

 Number of elements : 128

 Biopsy Guide : Available

 Application : OB, GYN, Urology

 Safety Class: BF

PHASED ARRAY

PN2-4
 Center frequency : 2.6MHz

 Band Width: 2 ~ 4 MHz

 Radius of curvature : Flat

 Field of view : 90 °

 Number of elements : 64

 Biopsy Guide : Not available

 Application : Abdomen, Cardiac, Vascular, Pediatric

 Safety Class: BF

PENCIL

DP2B
 Center frequency : 2.0MHz

 Application : Cardiac, Vascular

 Safety Class : BF

HS30 DATA SHEET V1.00 Rev01 Page 14


www.samsungmedison.com

 Time (M, Doppler)


MEASUREMENT
 Velocity

 Acceleration
 Caliper
 RI
 Abdomen
 Volume Flow (Diameter)
 Cardiac
 Volume Flow (Area)
 Vascular
 Auto Trace
 Gynecology
 Manual Trace
 Obstetrics
 Limited Trace
 Fetal Heart

 Urology
ABDOMEN
 MSK
 Gallbladder
 Small Parts
 Pancreas
 Pediatric
 Bowel
CALIPER  Kidney Vol. (Right / Light)

 Liver
 2D Distance
 Spleen
 M Distance
 Aorta
 2D Trace
 RA (Right / Left)
 2D Trace length
 Seg. A (Right / Left)
 Doppler Manual Trace
 Arc. A (Right / Left)
 Doppler Limited Trace
 Celiac A
 2 Lines Angle
 Splenic A
 3 Points Angle
 Hepatic A (C / R / L)
 Ellipse (Area / Circumference)
 Hepatic V (R / M / L)
 Spline
 Portal V (R /M / L)
 Open Spline
 SMA
 Closed Spline
 IMA
 %Stenosis (Diameter)
 IVC
 %Stenosis (Area)
 IMV
 1 Distance Volume
 SMV
 2 Distance Volume

 3 Distance Volume
CARDIAC
 Ellipse Volume
 LV (2D)
 Ellipse + Distance Volume
 LV Vol. (Simpson)
 Disk Volume
 LV Vol. (A/L)
 Slope
 LV Vol. (Bullet)
 Heart Rate (M, Doppler)

HS30 DATA SHEET V1.00 Rev01 Page 15


www.samsungmedison.com

 LV Mass UE VEIN
 RV (2D)
 Internal Jugular V (Right / Left)
 Aorta
 Innominate V (Right / Left)
 LA
 Subclavian V (Right / Left)
 LA Vol. (Simpson)
 Axillary V (Right / Left)
 RA
 Brachial V (Right / Left)
 LVOT
 Cephalic V (Right / Left)
 RVOT
 Basilic V (Right / Left)
 AV
 Radial V (Right / Left)
 MV
 Ulnar (Right / Left)
 TV

 PV LE ARTERY
 Shunt
 CIA (Left / Right)
 IVC
 IIA (Left / Right)
 Tei Index
 EIA (Left / Right)
 Plum. Vein
 CFA (Left / Right)
 Hepatic Vein
 SFA (Left / Right)
 Tissue Doppler
 DFA (Left / Right)
 Qp/Qs
 Popliteal A (Left / Right)
 LV (M)
 ATA (Left / Right)
 RV (M)
 PTA (Left / Right)

CAROTID  Peroneal A (Left / Right)

 DPA (Left / Right)


 Subclavian A (Right / Left)
 MPA (Left / Right)
 CCA (Right / Left/Prox./Mid./Dist)
 LPA (Left / Right)
 Bulb (Right / Left)
 Metatarsal A (Left / Right)
 ICA (Right / Left/Prox./Mid./Dist)
 Digital A (Left / Right)
 ECA (Right / Left)

 Vertebral A (Right / Left) LE VEIN

UE ARTERY  CIV (Left / Right)

 IIV (Left / Right)


 Subclavian A (Right / Left)
 EIV (Left / Right)
 Axillary A (Right / Left)
 CFV (Left / Right)
 Brachial A (Right / Left)
 PFV (Left / Right)
 Radial A (Right / Left)
 SFV (Left / Right)
 Ulnar A (Right / Left)
 GSV (Left / Right)
 SPA (Right / Left)
 Popliteal V (Left / Right)

HS30 DATA SHEET V1.00 Rev01 Page 16


www.samsungmedison.com

 LSV (Left / Right)  Uterine A (Right / Left)

 ATV (Left / Right)  Placenta A

 PTV (Left / Right)  Fetal Carotid (Right / Left)

 Peroneal V (Left / Right)  Fetal Aorta

 MPV (Left / Right)  Renal A (Right / Left)

 LPV (Left / Right)  Duct Venosus

 Metatarsal V (Left / Right)  Fetal HR

 Digital V (Left / Right)  PLI

GYNECOLOGY FETAL HEART

 Uterus  LV Vol. (Simpson)

 Cervix  2D Echo

 Cyst (Right / Left)  CTAR

 Ovary (Right / Left)  MPA

 Follicles (Right / Left / 1 ~ 12)  Duct Artriosus

 Mass 1 ~ 3  IVC

 Ovarian A (Right / Left)  Duct Venosus

 Uterine A (Right / Left)  Asc Aorta

 Pericystic Flow  Dsc Aorta

 Endometrial Flow  MV

 Endo. Polyp  TV

 Ovarian Mass (Right / Left)  PLI

 Uterine Tumor 1 ~ 3  TEI

 Cervical Tumor  Fetal HR

 Ectopic Pregnancy  M Echo

OBSTETRICS UROLOGY

 Fetal Biometry  WG Prostate

 Fetal Cranium  T-Zone Vol

 Fetal Long Bone  Bladder Vol.

 Fetal others  Residual Vol

 AFI  Renal Vol. (Right / Left)

 CTAR
BREAST
 Maternal Others

 Ratio  Mass 1 ~ 10 (Right / Left)

 Umbilical Artery  Breast Flow (Right / Left)

 Mid Cereb A

HS30 DATA SHEET V1.00 Rev01 Page 17


www.samsungmedison.com

MSK

 Shoulder (Right / Left)

 Wrist (Right / Left)

 Knee (Right / Left)

 Ankle (Right / Left)

THYROID

 Thyroid Vol. (Right / Left)

 Thyroid Flow (Right / Left)

 Mass 1 ~ 5 (Right / Left)

TESTICLE

 Testis Vol. (Right / Left)

 Epididymis (Right / Left)

 Testis Flow (Right / Left)

 Mass 1 ~ 5 (Right / Left)

SUPERFICIAL

 Superficial Vol (Right / Left)

 Superficial Flow (Right / Left)

 Mass 1 ~ 5 (Right / Left)

PEDIATRIC

 Hip Angle (Right / Left)

HS30 DATA SHEET V1.00 Rev01 Page 18


www.samsungmedison.com

- NEMA/AIUM UD-3: 2004


SAFETY / EMC
ACOUSTIC OUTPUT MANAGEMENT
CLASSIFICATIONS
 User selectable, transducer and scanning mode
 SAFETY
dependent
- Type of protection against electrical shock: Class I
 Dedicated Output Display on the system monitor display
- Degree of protection against electrical shock (Patient
of output acoustic
Connection): Type BF or CF Applied Part
 Power level, as well as thermal and mechanical indices:
 EMC
 PWR – Output Power level. Range: From 2 % of maximum
- RF Emission CISPR 11: Class A
output
 Degree of protection against harmful ingress of water:
 Level is increased by 2% in each step.
Ordinary equipment (All of applied parts (IPX7) except for
 Mechanical Index (MI): 0.01~1.90 Range
ECG, and Foot switch (IPX8))
 Thermal Index (TI): 0.01~6.00 Range
 RoHS Compliant
- TIC – Thermal Index, Bone at Surface
 WEEE Compliant
- TIB – Thermal Index, Bone at Focus

- TIS – Thermal Index, Soft Tissue


APPLIED STANDARDS

 Safety ANTI-VIRUS SOLUTION


- IEC/EN 60601-1: Ed. 3.0:2005 + A1:2012
 Disable USB Autorun Feature
- IEC/EN 60601-1: Ed. 2.0:1988 + A1:1991 + A2:1995
- Executable applications in USB stick are never launched
- IEC/EN 60601-1-1: Ed. 2.0:2000
- Prevent autorun virus through USB stick Dedicated
- IEC/EN 60601-1-2: Ed. 2.0:2001 + A1:2004 + Ed.
Output Display on the system monitor display of output
3.0:2007
acoustic
- IEC/EN 60601-1-4: Ed. 1.1:1996 + A1:1999
 Block Network Port (Except DICOM communication port)
- IEC/EN 60601-1-6: Ed. 2.0:2006 + Ed. 3.0:2010
- Ultrasound Machine allow only DICOM data through
- IEC/EN 60601-2-37: 2001 + A1:2004 + A2:2005 + Ed.
DICOM port
2.0:2007
- The network data of other network ports are rejected
- IEC/EN 60601-2-37: Ed. 1.0:2001 + A1:2004 + A2:2005
by Windows firewall
- IEC/EN 62366: Ed. 1.0:2007
 Prohibit user from accessing windows application (such
 Biocompatibility
as Explorer)
- ISO/EN 10993-1: 2009
- Impossible to execute applications which is not allowed
 Labeling
- Impossible to access internet web pages
- EN 980: 2008
 Windows Defender
- EN 1041: 2008
 Avast
- ISO 15223-1: 2012
SAMSUNG MEDISON Business Development Group
 NEMA/AIUM Phone: +82-2-2194-1508
Fax: +82-2-563-6528
- NEMA/AIUM UD-2: 2004 E-Mail:td.hwang@samsungmedison.com
© 2018 SAMSUNG MEDISON Ltd – All rights reserved