ALTL
Alanine Aminotransferase Enzymes
Order information
REF CONTENT Analyzer(s) on which cobas c pack(s) can be used
English The rate of the NADH oxidation is directly proportional to the catalytic ALT
activity. It is determined by measuring the decrease in absorbance at
System information 340 nm.
Test ALTL, test ID 0‑495
Reagents - working solutions
Intended use
In vitro test for the quantitative determination of the catalytic activity of ALT R1 TRIS buffer: 224 mmol/L, pH 7.3 (37 °C); L‑alanine: 1120 mmol/L;
(EC 2.6.1.2; L‑alanine: 2‑oxoglutarate aminotransferase) in human serum albumin (bovine): 0.25 %; LDH (microorganisms): ≥ 45 µkat/L;
and plasma on COBAS INTEGRA systems. stabilizers; preservative
This method sheet describes the application for ALT without pyridoxal SR 2‑Oxoglutarate: 94 mmol/L; NADH: ≥ 1.7 mmol/L; preservative
phosphate activation (test ALTL, 0‑495). The application for ALT activated
with pyridoxal phosphate is described in the method sheet Alanine R1 is in position B and SR is in position C.
Aminotransferase Pyridoxal Phosphate Activated . Precautions and warnings
Summary1,2 Pay attention to all precautions and warnings listed in
The enzyme alanine aminotransferase (ALT) has been widely reported as Section 1 / Introduction of this Method Manual.
present in a variety of tissues. The major source of ALT is the liver, which Reagent handling
has led to the measurement of ALT activity for the diagnosis of hepatic
diseases. Elevated serum ALT is found in hepatitis, cirrhosis, obstructive Ready for use
jaundice, carcinoma of the liver, and chronic alcohol abuse. ALT is only Storage and stability
slightly elevated in patients who have an uncomplicated myocardial
infarction. Shelf life at 2‑8 °C See expiration date on
Although both serum aspartate aminotransferase (AST) and ALT become cobas c pack label
elevated whenever disease processes affect liver cell integrity, ALT is the COBAS INTEGRA 400 plus system
more liver-specific enzyme. Moreover, elevations of ALT activity persist
longer than elevations of AST activity. On-board in use at 10‑15 °C 12 weeks
In patients with vitamin B6 deficiency, serum aminotransferase activity may COBAS INTEGRA 800 system
be decreased. The apparent reduction in aminotransferase activity may be On-board in use at 8 °C 12 weeks
related to decreased pyridoxal phosphate, the prosthetic group for
aminotransferases, resulting in an increase in the ratio of apoenzyme to Specimen collection and preparation
holoenzyme. For specimen collection and preparation only use suitable tubes or
Test principle collection containers.
Method according to the International Federation of Clinical Chemistry Only the specimens listed below were tested and found acceptable.
(IFCC), but without pyridoxal-5'-phosphate.3,4 Serum (free from hemolysis)
ALT catalyzes the reaction between L‑alanine and 2‑oxoglutarate. The Plasma (free from hemolysis): Li‑heparin or EDTA plasma. Do not use other
pyruvate formed is reduced by NADH in a reaction catalyzed by lactate anticoagulants.
dehydrogenase (LDH) to form L‑lactate and NAD+. Nonhemolyzed serum is the specimen of choice.
The sample types listed were tested with a selection of sample collection
ALT tubes that were commercially available at the time of testing, i.e. not all
L‑alanine + 2‑oxoglutarate pyruvate + L‑glutamate available tubes of all manufacturers were tested. Sample collection systems
from various manufacturers may contain differing materials which could
LDH affect the test results in some cases. When processing samples in primary
pyruvate + NADH + H+ L‑lactate + NAD+ tubes (sample collection systems), follow the instructions of the tube
manufacturer.
Centrifuge samples containing precipitates before performing the assay.
ALTL
Alanine Aminotransferase Enzymes
ALTL
Alanine Aminotransferase Enzymes
Lower limits of measurement 2 Moss DW, Henderson AR, Kachmar JF. Enzymes. In: Tietz NW, ed.
Lower detection limit of the test: Fundamentals of Clinical Chemistry, 3rd ed. Philadelphia, PA: WB
2 U/L (0.03 µkat/L) Saunders 1987;346-421.
The lower detection limit represents the lowest measurable analyte level 3 Bergmeyer HU, Hørder M, Rej R. Approved recommendation (1985) on
that can be distinguished from zero. It is calculated as the value lying 3 IFCC methods for the measurement of catalytic concentration of
standard deviations above that of a zero sample (zero sample + 3 SD, enzymes. Part 3. IFCC method for alanine aminotransferase. J Clin
repeatability, n = 30). Chem Clin Biochem 1986;24:481-495.
Expected values12 4 ECCLS. Determination of the catalytic activity concentration in serum of
Acc. to the optimized standard method (comparable to the IFCC method L-alanine aminotransferase (EC 2.6.1.2, ALAT). Klin Chem Mitt
without pyridoxal phosphate activation13): 1989;20:204-211.
5 Heins M, Heil W, Withold W. Storage of Serum or Whole Blood
Males up to 41 U/L (up to 0.685 µkat/L) Samples? Effect of Time and Temperature on 22 Serum Analytes. Eur
Females up to 33 U/L (up to 0.551 µkat/L) J Clin Chem Clin Biochem 1995;33:231-238.
Calculated values: a factor of 1.85 is used for the conversion from 25 °C to 6 Data on file at Roche Diagnostics.
37 °C.14 7 Schumann G, Bonora R, Ceriotti F, et al. IFCC Primary Reference
Each laboratory should investigate the transferability of the expected values Procedures for the Measurement of Catalytic Activity Concentrations of
to its own patient population and if necessary determine its own reference Enzymes at 37 °C – Part 4. Reference Procedure for the Measurement
ranges. of Catalytic Activity Concentration of Alanine Aminotransferase. Clin
Chem Lab Med 2002;40(7):718-724.
Specific performance data
8 Glick MR, Ryder KW, Jackson SA. Graphical Comparisons of
Representative performance data on the COBAS INTEGRA analyzers are Interferences in Clinical Chemistry Instrumentation.
given below. Results obtained in individual laboratories may differ. Clin Chem 1986;32:470-475.
Precision 9 Breuer J. Report on the Symposium “Drug effects in Clinical Chemistry
Precision was determined using human samples and controls in an internal Methods”. Eur J Clin Chem Clin Biochem 1996;34:385-386.
protocol with repeatability and intermediate precision (2 aliquots per run,
2 runs per day, 20 days). The following results were obtained: 10 Sonntag O, Scholer A. Drug interference in clinical chemistry:
recommendation of drugs and their concentrations to be used in drug
Level 1 Level 2 interference studies. Ann Clin Biochem 2001;38:376-385.
11 Bakker AJ, Mücke M. Gammopathy interference in clinical chemistry
Mean 34.9 U/L 132 U/L assays: mechanisms, detection and prevention.
(0.583 µkat/L) (2.20 µkat/L) Clin Chem Lab Med 2007;45(9):1240-1243.
12 Thefeld W, Hoffmeister H, Busch EW, et al. Referenzwerte für die
CV repeatability 1.0 % 0.5 % Bestimmungen der Transaminasen GOT und GPT sowie der
CV intermediate precision 1.5 % 1.9 % alkalischen Phosphatase im Serum mit optimierten Standardmethoden.
Dtsch Med Wschr 1974;99(8):343-351.
Method comparison 13 Klein G, Lehmann P, Michel E, et al. Vergleich der IFCC-Methoden für
ALT values for human serum and plasma samples obtained on a ALAT, ASAT und GGT bei 37 °C mit den eingeführten
COBAS INTEGRA 700 analyzer using the COBAS INTEGRA Alanine Standardmethoden bei 25 °C und 37 °C. Lab Med 1994;18:403-404.
Aminotransferase reagent ALTL (y) were compared with those determined 14 Zawta B, Klein G, Bablok W. Temperaturumrechnung in der klinischen
using commercially available reagents for ALT on a COBAS INTEGRA Enzymologie? Klin Lab 1994;40:23-32.
analyzer (x) and on an alternative manufacturer's clinical chemistry system
(x). Samples were measured in duplicate. Sample size (n) represents all 15 Bablok W, Passing H, Bender R, et al. A general regression procedure
replicates. for method transformation. Application of linear regression procedures
for method comparison studies in clinical chemistry, Part III. J Clin
COBAS INTEGRA analyzer Chem Clin Biochem 1988 Nov;26(11):783-790.
Sample size (n) 236 A point (period/stop) is always used in this Method Sheet as the decimal
separator to mark the border between the integral and the fractional parts of
Correlation coefficient (r) 1.00 a decimal numeral. Separators for thousands are not used.
(rs) 0.998 Symbols
Linear regression y = 1.068x - 2.31 U/L Roche Diagnostics uses the following symbols and signs in addition to
those listed in the ISO 15223‑1 standard.
Passing/Bablok15 y = 1.055x - 1.82 U/L
CONTENT Contents of kit
Alternative system
Volume after reconstitution or mixing
Sample size (n) 224
Correlation coefficient (r) 1.00 FOR US CUSTOMERS ONLY: LIMITED WARRANTY
(rs) 0.995 Roche Diagnostics warrants that this product will meet the specifications
stated in the labeling when used in accordance with such labeling and will
Linear regression y = 0.983x + 0.772 U/L
be free from defects in material and workmanship until the expiration date
Passing/Bablok15 y = 0.982x + 0.856 U/L printed on the label. THIS LIMITED WARRANTY IS IN LIEU OF ANY
The sample activities were between 3.7 and 496 U/L (0.062 and OTHER WARRANTY, EXPRESS OR IMPLIED, INCLUDING ANY IMPLIED
8.27 µkat/L). WARRANTY OF MERCHANTABILITY OR FITNESS FOR PARTICULAR
References PURPOSE. IN NO EVENT SHALL ROCHE DIAGNOSTICS BE LIABLE
1 Sherwin JE. Liver function. In: Kaplan LA, Pesce AJ, eds. Clinical FOR INCIDENTAL, INDIRECT, SPECIAL OR CONSEQUENTIAL
Chemistry, theory, analysis, and correlation. St. Louis: Mosby DAMAGES.
1984;420-438. COBAS, COBAS C, COBAS INTEGRA, PRECINORM, PRECIPATH and PRECICONTROL are trademarks of
Roche.