Professional Documents
Culture Documents
ELEC4810: Introduction To Biosensors and Bioinstrumentation: Lecture Notes - Set #8
ELEC4810: Introduction To Biosensors and Bioinstrumentation: Lecture Notes - Set #8
ELEC4810: Introduction To Biosensors and Bioinstrumentation: Lecture Notes - Set #8
ELEC4810
Textbook: Chapter 8, 2
Chapter 3
Measurements of Blood Pressure and Blood Flow
Main Topics:
• Present the principles of hemodynamic pressure measurements
of human cardiovascular system.
• Discuss the interpretation of the results of current blood
pressure and flow measurement techniques.
2
Noninvasive measurement of blood flow: Ultrasonic Flowmeter
• Measure time-varying flow of blood.
• Penetrate the tissue → making the non-invasive monitoring practical.
3
Noninvasive measurement of blood flow: Ultrasonic Flowmeter
Doppler Ultrasonic Flowmeters:
Doppler effect
4
Noninvasive measurement of blood flow: Ultrasonic Flowmeter
Doppler Ultrasonic Flowmeters:
Doppler Geometry
• In most Doppler sensors both the transmitter (or transducer) and receiver are
stationary, and they illuminate a moving target to measure its velocity.
• Radar speed guns, like other types of radar, consist of a radio transmitter and
receiver. They send out a radio signal in a narrow beam, then receive the same signal
back after it bounces off the target object.
• Due to a phenomenon called the Doppler effect, if the object is moving toward or
away from the gun, the frequency of the reflected radio waves when they come back is
5
different from the transmitted waves.
Noninvasive measurement of blood flow: Ultrasonic Flowmeter
Doppler Ultrasonic Flowmeters:
Doppler effect
6
Noninvasive measurement of blood flow: Ultrasonic Flowmeter
Human Vocal Sound Production
• The vocal folds (vocal cords) are composed of twin infoldings of mucous membrane
stretched horizontally, from back to front, across the larynx.
• Sound is the generalized name given to “acoustic waves” that is basically a waveform of energy
that is produced by some form of a mechanical vibration.
• Diaphragm action pushes air from the lungs through the elastic vocal folds to producing audible
vibrations or sound. The process is called phonation.
• The acoustic waves have frequencies ranging from 1Hz up to many tens of thousands of Hz with
7
the upper limit of human hearing being around the 20 kHz, (20,000Hz) range.
Noninvasive measurement of blood flow: Ultrasonic Flowmeter
Rarefaction Compression
(low pressure) (high pressure)
𝑽𝑽
Wavelength 𝝀𝝀 = → spatial period of a wave.
𝒇𝒇
8
Noninvasive measurement of blood flow: Ultrasonic Flowmeter
The transducer:
• Generation of ultrasound wave: a piezoelectric material converts power from electric
to acoustic form.
Piezoelectric Effect is the ability of certain materials to generate an electric charge
in response to applied mechanical stress.
𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇
𝐹𝐹𝐹𝐹𝐹𝐹𝐹𝐹𝐹𝐹 = 0 𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶
𝑉𝑉 = 0 𝑉𝑉 → " − "
𝑉𝑉 → " + "
9
Noninvasive measurement of blood flow: Ultrasonic Flowmeter
The transducer:
• Generation of ultrasound wave: a piezoelectric material converts power from electric
to acoustic form.
Piezoelectric Effect is the ability of certain materials to generate an electric charge
in response to applied mechanical stress.
𝐴𝐴
𝐶𝐶 = 𝜀𝜀
𝑑𝑑
𝑄𝑄
𝐶𝐶 =
𝑉𝑉
𝑑𝑑 � 𝑄𝑄
𝑉𝑉 =
𝜀𝜀𝜀𝜀
𝑽𝑽+
𝑽𝑽−
Contracts
𝑽𝑽+ Expands
11
Noninvasive measurement of blood flow: Ultrasonic Flowmeter
The transducer:
• Generation of ultrasound wave: a piezoelectric material converts power from electric
to acoustic form.
Driving signal
+ + + + + +
0
− − − − − −
"+: 𝒓𝒓𝒓𝒓𝒓𝒓𝒓𝒓𝒓𝒓𝒓𝒓𝒓𝒓𝒓𝒓𝒓𝒓𝒓𝒓𝒓𝒓𝒓
"−: 𝒄𝒄𝒄𝒄𝒄𝒄𝒄𝒄𝒄𝒄𝒄𝒄𝒄𝒄𝒄𝒄𝒄𝒄𝒄𝒄𝒄𝒄𝒄
A realistic transducer
12
Noninvasive measurement of blood flow: Ultrasonic Flowmeter
The ultrasound sensor or receiver:
• Generation of voltage signal when pressure is applied: a piezoelectric material converts
power from acoustic to electric form.
"𝒄𝒄𝒄𝒄𝒄𝒄𝒄𝒄𝒄𝒄𝒄𝒄𝒄𝒄𝒄𝒄𝒄𝒄𝒄𝒄𝒄𝒄: − "
Received signal
"𝒓𝒓𝒓𝒓𝒓𝒓𝒓𝒓𝒓𝒓𝒓𝒓𝒓𝒓𝒓𝒓𝒓𝒓𝒓𝒓𝒓𝒓: + "
13
Noninvasive measurement of blood flow: Ultrasonic Flowmeter
The transducer:
• A piezoelectric material converts power from electric to acoustic form.
• Lead zirconate titanate, a crystal with highest conversion efficiency can be molded into
any shape by melting. For maximal efficiency, the crystal is one-half wavelength thick.
• The watery gel must be used to fill the gap between the transducer and tissue surface to
reduce the reflection. ← matching of acoustic impedance.
14
Noninvasive measurement of blood flow: Ultrasonic Flowmeter
Speed of sound:
• ~340 m/s in air at sea level
• ~1,500 m/s in soft tissue
• ~3,500 m/s in hard tissue
15
Noninvasive measurement of blood flow: Ultrasonic Flowmeter
The transducer:
• A piezoelectric material converts power from electric to acoustic form.
• Near field distance:
𝑽𝑽
where wavelength 𝝀𝝀 = 𝒇𝒇 → special period of a wave.
16
Noninvasive measurement of blood flow: Ultrasonic Flowmeter
Phased array ultrasonics:
• Phased array ultrasonics (PA) is an advanced method of ultrasonic testing that has applications in medical imaging.
• The PA probe consists of many small ultrasonic transducers, each of which can be pulsed independently.
• By varying the timing, for instance by making the pulse from each transducer progressively delayed going up the
line, a pattern of constructive interference is set up that results in radiating a quasi-plane ultrasonic beam at a set
angle depending on the progressive time delay.
• In other words, by changing the progressive time delay the beam can be steered electronically. It can be swept like a
search-light through the tissue or object being examined, and the data from multiple beams are put together to
make a visual image showing a slice through the object.
17
Noninvasive measurement of blood flow: Ultrasonic Flowmeter
Doppler Ultrasonic Flowmeters:
Doppler effect
18
Noninvasive measurement of blood flow: Ultrasonic Flowmeter
Doppler Ultrasonic Flowmeters:
Doppler effect:
• When a sound source moves relative to an observer, a frequency shift can be noted consisting
of an apparent shift to a higher frequency as the source approaches to the observer and an
apparent shift to a lower frequency as it moves away.
19
Noninvasive measurement of blood flow: Ultrasonic Flowmeter
Doppler Ultrasonic Flowmeters:
Doppler effect:
• The stationary observer hears a frequency:
𝑐𝑐
𝑓𝑓 = 𝑐𝑐
𝜆𝜆 𝜆𝜆 =
𝑓𝑓
• The crest velocity observed by the person on the
platform traveling at a velocity V, however , will be
c - V. Furthermore, the wavelength of the sound ,
unaffected by the platform, remains the same.
Therefore, the person hears a frequency fm of:
𝑐𝑐 − 𝑉𝑉
𝑓𝑓𝑚𝑚 =
𝜆𝜆
• Taking the ratio of these two equations gives
𝑓𝑓𝑚𝑚 𝑉𝑉 𝑉𝑉
=1− 𝑜𝑜𝑜𝑜 𝑓𝑓𝑚𝑚 = 1 − 𝑓𝑓
𝑓𝑓 𝑐𝑐 𝑐𝑐
𝑉𝑉 𝑐𝑐
𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷 𝑠𝑠𝑠𝑠𝑠𝑠𝑠𝑠𝑠: 𝑓𝑓𝐷𝐷 = 𝑓𝑓 − 𝑓𝑓𝑚𝑚 = � 𝑓𝑓 𝑎𝑎𝑎𝑎𝑎𝑎 𝑉𝑉 = � 𝑓𝑓𝐷𝐷 Velocity can be measured!
20 𝑐𝑐 𝑓𝑓
Noninvasive measurement of blood flow: Ultrasonic Flowmeter
Doppler Ultrasonic Flowmeters:
Doppler effect:
Sound wave: c, fs , λ
Source Receiver
𝑉𝑉𝑆𝑆 𝑉𝑉𝑅𝑅
𝑐𝑐 − 𝑉𝑉𝑟𝑟
𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅 𝑓𝑓𝑓𝑓𝑓𝑓𝑓𝑓𝑓𝑓𝑓𝑓𝑓𝑓𝑓𝑓𝑓𝑓: 𝑓𝑓𝑟𝑟 = � 𝑓𝑓𝑠𝑠 (Detailed derivations are provided in handout.)
𝑐𝑐 − 𝑉𝑉𝑠𝑠
𝑉𝑉𝑠𝑠 − 𝑉𝑉𝑟𝑟
𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷 𝑠𝑠𝑠𝑠𝑠𝑠𝑠𝑠𝑠: 𝑓𝑓𝐷𝐷 = 𝑓𝑓𝑟𝑟 − 𝑓𝑓𝑠𝑠 = � 𝑓𝑓𝑠𝑠
𝑐𝑐 − 𝑉𝑉𝑠𝑠
• Radar speed guns, like other types of radar, consist of a radio transmitter and
receiver. They send out a radio signal in a narrow beam, then receive the same signal
back after it bounces off the target object.
• Due to a phenomenon called the Doppler effect, if the object is moving toward or
away from the gun, the frequency of the reflected radio waves when they come back is
22
different from the transmitted waves.
Noninvasive measurement of blood flow: Ultrasonic Flowmeter
Doppler Ultrasonic Flowmeters:
Doppler Geometry
• In most Doppler sensors both the transmitter (or transducer) and receiver are
stationary, and they illuminate a moving target to measure its velocity.
• A general Doppler measurement geometry is shown below:
What is the relationship between Doppler shift (𝑓𝑓𝐷𝐷 = 𝑓𝑓𝑟𝑟 −𝑓𝑓𝑠𝑠 ) and the target velocity, V?
23
Noninvasive measurement of blood flow: Ultrasonic Flowmeter
Doppler Ultrasonic Flowmeters:
Doppler Geometry
• A simplest Doppler measurement geometry: 𝜃𝜃𝑡𝑡 = 𝜃𝜃𝑟𝑟 = 0
Target
Source
Receiver
𝑉𝑉𝑖𝑖
Source Receiver
−𝑉𝑉𝑖𝑖 𝑉𝑉𝑖𝑖
24
Noninvasive measurement of blood flow: Ultrasonic Flowmeter
Doppler Ultrasonic Flowmeters:
Doppler Geometry
• A simplest Doppler measurement geometry: 𝜃𝜃𝑡𝑡 = 𝜃𝜃𝑟𝑟 = 0
Target
Source
Receiver
𝑉𝑉𝑖𝑖
2𝑉𝑉𝑖𝑖 𝑐𝑐
𝑓𝑓𝐷𝐷 = − � 𝑓𝑓𝑠𝑠 𝑉𝑉𝑖𝑖 = −𝑓𝑓𝐷𝐷
𝑐𝑐 + 𝑉𝑉𝑖𝑖 2𝑓𝑓
25
Noninvasive measurement of blood flow: Ultrasonic Flowmeter
Doppler Ultrasonic Flowmeters:
Doppler Geometry
• A realistic Doppler measurement geometry: 𝜃𝜃𝑡𝑡 = 𝜃𝜃𝑟𝑟 = 𝜃𝜃
Source
Receiver 𝜃𝜃
𝑐𝑐
𝑉𝑉𝑖𝑖 𝑐𝑐𝑐𝑐𝑐𝑐𝜃𝜃 = −𝑓𝑓𝐷𝐷
2𝑓𝑓
𝑐𝑐
𝑉𝑉𝑖𝑖 = −𝑓𝑓𝐷𝐷
2𝑓𝑓𝑓𝑓𝑓𝑓𝑓𝑓𝜃𝜃
26
Noninvasive measurement of blood flow: Ultrasonic Flowmeter
Doppler Ultrasonic Flowmeters:
Doppler Geometry
• A realistic Doppler measurement geometry: 𝜃𝜃𝑡𝑡 ≠ 𝜃𝜃𝑟𝑟
𝑉𝑉𝑖𝑖
Source Receiver
𝑉𝑉𝑠𝑠 − 𝑉𝑉𝑟𝑟 𝑐𝑐
𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷 𝑠𝑠𝑠𝑠𝑠𝑠𝑠𝑠𝑠: 𝑓𝑓𝐷𝐷 = 𝑓𝑓𝑟𝑟 − 𝑓𝑓𝑠𝑠 = � 𝑓𝑓𝑠𝑠 𝑉𝑉𝑖𝑖 = −𝑓𝑓𝐷𝐷
𝑐𝑐 − 𝑉𝑉𝑠𝑠 𝑓𝑓 � 𝑐𝑐𝑐𝑐𝑐𝑐𝜃𝜃𝑠𝑠 + 𝑐𝑐𝑐𝑐𝑐𝑐𝜃𝜃𝑟𝑟
27
Noninvasive measurement of blood flow: Ultrasonic Flowmeter
Doppler Ultrasonic Flowmeters: a simple version
cosωst × cosωrt = 1/2[cos(ωs t + ωrt ) + cos(ωst − ωrt )]
= 1/2 cos(ωs + ωr)t + 1/2 cos(ωs − ωr )t
≈ 1/2 cos 2ωs t + 1/2 cos 2π fD t ( f >> fD )
𝑓𝑓𝑠𝑠 > 𝑀𝑀𝑀𝑀𝑀𝑀 𝑓𝑓𝐷𝐷 ~ 𝐻𝐻𝐻𝐻
cosωst × cosωrt
cosωst
cosωst
and
cosωrt
𝑐𝑐
𝑉𝑉𝑖𝑖 = −𝑓𝑓𝐷𝐷
28 𝑓𝑓 � 𝑐𝑐𝑐𝑐𝑐𝑐𝜃𝜃𝑠𝑠 + 𝑐𝑐𝑐𝑐𝑐𝑐𝜃𝜃𝑟𝑟
Noninvasive measurement of blood flow: Ultrasonic Flowmeter
Doppler Ultrasonic Flowmeters: a simple version
cosωst × cosωrt = 1/2[cos(ωs t + ωrt ) + cos(ωst − ωrt )] = 1/2 cos(ωs + ωr)t + 1/2
cos(ωs − ωr )t ≈ 1/2 cos 2ωs t + 1/2 cos 2π fD t ( f >> fD )
𝑐𝑐 𝑐𝑐
𝑉𝑉𝑖𝑖 = −𝑓𝑓𝐷𝐷 𝑉𝑉𝑖𝑖 = −𝑓𝑓𝐷𝐷
𝑓𝑓 � 𝑐𝑐𝑐𝑐𝑐𝑐𝜃𝜃𝑠𝑠 + 𝑐𝑐𝑐𝑐𝑐𝑐𝜃𝜃𝑟𝑟 2𝑓𝑓𝑓𝑓𝑓𝑓𝑓𝑓𝜃𝜃
Problems:
• Tumbling of cells and local velocities resulting from turbulence cause different Doppler-
frequency shifts.
• Velocity profiles are rarely blunt, with all cells moving at the same velocity. Rather, cells
move at different velocities, producing different shifts of the Doppler frequency.
• A given cell remains within the beam-intersection volume for a short time. Thus the signal
received from one cell is a pure frequency multiplied by some time-gate function, yielding a
band of frequencies.
• Acoustic energy traveling within the main beam, but at angles to the beam axis, plus energy
in the side lobes, causes different Doppler-frequency shifts due to an effective change in θ.
• The instrument can not provide the direction of blood flow.
29
Noninvasive measurement of blood flow: Ultrasonic Flowmeter
Doppler Ultrasonic Flowmeters: Quadrature-phase Detection
Trigonometric functions:
cos(ω s t + ϕ ) cos(ωr t ) =
1
[cos(ωst + ϕ + ωr t ) + cos(ωst + ϕ − ωr t )]
2
= [cos(ω s t + ϕ + ω r t ) + cos((ω s − ω r )t + ϕ )]
1
2
=
1
[cos(ω s t + ϕ + ωr t ) + cos(ωd t + ϕ )]
2
where ωd = 2πf d = ω s − ωr , ω s , ωr >> ωd
● If the blood cells move away from the ultrasound: ωd > 0 , there is
a +90° phase difference between cosine and sine channels.
31
Noninvasive measurement of blood flow: Ultrasonic Flowmeter
Doppler Ultrasonic Flowmeters: Quadrature-phase Detection
Low-pass filtered cosine channel (solid line) and sine channel (dash line):
𝑐𝑐𝑐𝑐𝑐𝑐𝜔𝜔𝜔𝜔
𝑐𝑐𝑐𝑐𝑐𝑐𝜔𝜔𝑟𝑟 𝑡𝑡 × 𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐
𝑠𝑠𝑠𝑠𝑠𝑠𝜔𝜔𝜔𝜔 𝑐𝑐𝑐𝑐𝑐𝑐𝜔𝜔𝐷𝐷 𝑡𝑡
𝑐𝑐𝑐𝑐𝑐𝑐𝜔𝜔𝑟𝑟 𝑡𝑡
𝑠𝑠𝑠𝑠𝑠𝑠𝜔𝜔𝐷𝐷 𝑡𝑡
𝑐𝑐𝑐𝑐𝑐𝑐𝜔𝜔𝑟𝑟 𝑡𝑡 × 𝑠𝑠𝑠𝑠𝑠𝑠𝑠𝑠𝑠𝑠
34
Noninvasive measurement of blood flow: Ultrasonic Flowmeter
Doppler Ultrasonic Flowmeters: Quadrature-phase Detection
“1”
𝑐𝑐𝑐𝑐𝑐𝑐𝜔𝜔𝐷𝐷 𝑡𝑡
“0” “−”
“0” “+”
“1”
“1”
𝑠𝑠𝑠𝑠𝑠𝑠𝜔𝜔𝐷𝐷 𝑡𝑡
“0”
cosine channel
• Laser Doppler Flowmetry (LDF) is a non-invasive method to estimate the blood perfusion in
the microcirculation.
• It measures average blood flow in skin capillaries.
Sensory receptors
𝑐𝑐
𝑉𝑉𝑖𝑖 = −𝑓𝑓𝐷𝐷
𝑓𝑓 � 𝑐𝑐𝑐𝑐𝑐𝑐𝜃𝜃𝑠𝑠 + 𝑐𝑐𝑐𝑐𝑐𝑐𝜃𝜃𝑟𝑟
𝑐𝑐
𝑉𝑉𝑖𝑖 = −𝑓𝑓𝐷𝐷 Capillaries of different sizes and orientations
2𝑓𝑓𝑓𝑓𝑓𝑓𝑓𝑓𝜃𝜃
36 → θ is NOT a constant.
Noninvasive measurement of blood flow: Laser Doppler Flowmeter
𝑐𝑐
𝑉𝑉𝑖𝑖 = −𝑓𝑓𝐷𝐷
𝑓𝑓 � 𝑐𝑐𝑐𝑐𝑐𝑐𝜃𝜃𝑠𝑠 + 𝑐𝑐𝑐𝑐𝑐𝑐𝜃𝜃𝑟𝑟
Distribution Distribution
𝑉𝑉𝑖𝑖
𝑓𝑓𝐷𝐷 = −𝑓𝑓 � � 𝑐𝑐𝑐𝑐𝑐𝑐𝜃𝜃𝑠𝑠 + 𝑐𝑐𝑐𝑐𝑐𝑐𝜃𝜃𝑟𝑟
𝑐𝑐
Distribution
𝑉𝑉𝑖𝑖
𝑓𝑓𝐷𝐷 = −𝑓𝑓 � � 𝑐𝑐𝑐𝑐𝑐𝑐𝜃𝜃𝑠𝑠 + 𝑐𝑐𝑐𝑐𝑐𝑐𝜃𝜃𝑟𝑟
38 𝑐𝑐
Noninvasive measurement of blood flow: Laser Doppler Flowmeter
39
Noninvasive measurement of blood flow: Laser Doppler Flowmeter
Accurate assessment of burn depth at an early stage is crucial to avoid unnecessary surgery or potential
hypertrophic scaring. It is clinically proven, that the status of the skin microcirculation can be used to estimate
burn depth. Superficial dermal burns will show a higher skin perfusion as compared to normal skin, whereas
the perfusion is compromised in deeper dermal burns.
40
Noninvasive measurement of blood flow: Laser Doppler Flowmeter
You can do it: System configuration of blood flow sensor using a smartphone.
41
Pressure Transducers (refer to Chapter 2)
Principles:
• A diaphragm pressure transducer is used for low pressure measurement.
• Pressure-deflection relationship of a circular diaphragm
P0
P> P0
Circular Diaphragm:
• radius: r 3(1−𝜇𝜇2 )𝑟𝑟 4
𝑥𝑥𝑑𝑑 = 𝑃𝑃 circular diaphragm (small 𝑥𝑥𝑑𝑑 )
16𝐸𝐸𝑤𝑤 3
• thickness: w
where
E = Young's modulus (N/ m2)
How to measure xd ? 𝜇𝜇 = Poisson's ratio (=0.33 for most metals)
42
Pressure Transducers (refer to Chapter 2)
Material properties:
• E = Young's modulus (N/ m2);
• 𝜇𝜇 = Poisson's ratio (=0.33 for most metals)
A: cross-section area
F: force
𝐹𝐹
𝐹𝐹 ∆𝑙𝑙 𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆
𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆 = ; 𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆 = 𝑌𝑌𝑌𝑌𝑌𝑌𝑌𝑌𝑔𝑔′ 𝑠𝑠 𝑚𝑚𝑚𝑚𝑚𝑚𝑚𝑚𝑚𝑚𝑚𝑚𝑚𝑚 𝐸𝐸 = = 𝐴𝐴
𝐴𝐴 𝑙𝑙 𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆 ∆𝑙𝑙
𝑙𝑙
2
𝐷𝐷
∆ 𝜋𝜋 ∆𝐷𝐷 ∆𝐴𝐴
∆𝐴𝐴 2 2∆𝐷𝐷
= 2 = 𝑃𝑃𝑃𝑃𝑃𝑃𝑃𝑃𝑃𝑃𝑃𝑃𝑛𝑛′ 𝑠𝑠 𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟𝑟 𝜇𝜇 = − 𝐷𝐷 = − 𝐴𝐴
𝐴𝐴 𝐷𝐷 𝐷𝐷 ∆𝑙𝑙 ∆𝑙𝑙
𝜋𝜋 2 2�
𝑙𝑙 𝑙𝑙
43
Pressure Transducers
A small displacement measurement: Resistance Sensors
• Measures displacement by determining the change in resistance of a conductor by
its deformation under load.
∆𝑅𝑅/𝑅𝑅 ∆𝜌𝜌/𝜌𝜌
𝐺𝐺 = = 1 + 2𝜇𝜇 +
∆𝑙𝑙/𝑙𝑙 ∆𝑙𝑙/𝑙𝑙
45
Pressure Transducers
A small displacement measurement: Resistance Sensors
46
Pressure Transducers
A small displacement measurement: Resistance Sensors
• Measures displacement by determining the change in resistance of a conductor by
its deformation under load → measurement of small change in electrical resistance
A B
48
Pressure Transducers
A small displacement measurement: Resistance Sensors
• Integrated semiconductor diaphragm pressure transducer
Compressing
𝑃𝑃 = 𝑃𝑃1 − 𝑃𝑃2
Stretching
Compressing
𝑉𝑉𝑜𝑜
𝑃𝑃 = 𝑃𝑃1 − 𝑃𝑃2 ∝
𝑉𝑉𝑖𝑖
𝑉𝑉𝑜𝑜
𝑜𝑜𝑜𝑜 𝑃𝑃 = 𝐶𝐶 �
𝑉𝑉𝑖𝑖
Proportional constant
50
Pressure Transducers
Piezoelectric Pressure Sensors
Piezoelectric Effect is the ability of certain materials to generate an electric charge
in response to applied mechanical stress.
𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇
𝐹𝐹𝐹𝐹𝐹𝐹𝐹𝐹𝐹𝐹 = 0 𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶
𝑉𝑉 = 0 𝑉𝑉 → " − "
𝑉𝑉 → " + "
51
Pressure Transducers
Piezoelectric Pressure Sensors
𝑑𝑑𝑑𝑑 𝑑𝑑𝑑𝑑
𝑖𝑖 = = 𝑘𝑘𝑘𝑘
𝑑𝑑𝑑𝑑 𝑑𝑑𝑑𝑑
• Note that this current is proportional to the time rate of change of the applied pressure.
• If pressure does not change, i = 0 → can only be used to sense “ac” pressure.
52