Download as pdf or txt
Download as pdf or txt
You are on page 1of 7

focal point

BY M. W. BLADES
DEPARTMENTOF CHEMISTRY,
UNIVERSITY OF BRITISH COLUMBIA,
VANCOUVER, BRITISH COLUMBIA
V6T I ZI, CANADA

Atomic Mass
Spectrometry
INTRODUCTION used--the determination of the partial common; the discussion below will
or complete elemental content of a focus on these. Spark source mass
he use of atomic mass spectrom- sample. Common techniques that have spectrometry will not be included in

T etry has grown enormously dur-


ing the past two decades. Com-
pared with most optical methods,
been and are still being used for this
purpose include flame and furnace
atomic absorption spectroscopy and
the discussion because, although it was
a very important method for direct
solids analysis in the 1960s and 1970s,
atomic mass spectrometry offers lower plasma emission spectroscopy. To it has gradually been replaced by the
detection limits (ppt to ppq) and broad some extent, these "traditional" meth- methods discussed below. Also, ther-
elemental coverage, with the addition- ods are being displaced by the new and mal ionization mass spectrometry
al capabilities of providing isotopic powerful mass spectrometric analogs. (TIMS) and accelerator mass spec-
abundance information and internal Photons are out, ions are in! trometry (AMS) will not be included
standardization through isotope dilu- A generic atomic mass spectrometer although both are very sensitive meth-
tion. The growth in the use of atomic is depicted schematically in Fig. 1. ods for elemental analysis. There is an
mass spectrometry is directly related to Common to all the techniques is an at- excellent paper in this issue of Applied
the development of reliable instrumen- omization/ion source and a mass ana- Spectroscopy that provides a brief in-
t a t i o n - p r i m a r i l y commercial instru- lyzer. For some techniques (e.g., reso- troduction to TIMS and discusses its
ments for inductively coupled plasma nance ionization mass spectrometry) application to environmental monitor-
mass spectrometry (ICP-MS), glow an additional or supplementary ioniza- ing. 3
discharge mass spectrometry (GD- tion source may be used between the
LASER MICROPROBE MASS
MS), laser sampling mass spectrome- primary atomizer/ionizer and the mass
SPECTROMETRY
try, and secondary ion mass spectrome- analyzer to improve the ionization effi-
try (SIMS). ciency and, hence, lower the detection The unique characteristics of laser
Koppenaal ~ defines atomic mass limits or improve the selectivity. microprobe mass spectrometry
spectrometry as "the mass spectromet- Some atomization/ionization sources (LMMS) are the low detection limits,
ric measurement of atomic (as opposed that can be used for atomic mass spec- the potential for highly localized anal-
to molecular or polyatomic) ions, for trometry are outlined in the left-hand ysis, and the lack of a need for sample
the primary purpose o f elemental box in Fig. 1. The bottom three (in ital- pretreatment prior to analysis. Al-
and/or isotopic composition determi- ics) are potential ion sources but have though the method has been known
nations". The term inorganic mass not yet been reported as being success- since the early 1960s, 4 the most signif-
spectrometry is also sometimes used to fully coupled to a mass analyzer. The icant event in its evolution was the in-
refer to these same areas of interest. 2 common mass analyzers are listed in troduction of commercial equipment by
The essential distinction seems to be, the right-hand box in Fig. 1. Leybold-Heraeus some 15 years ago.
in a sense, territorial. Atomic mass In theory, any combination of source A schematic diagram of a "typical"
spectrometry can be thought of as and analyzer is possible, and many of time-of-flight/laser microprobe mass
being used primarily for the same pur- the possible combinations have been spectrometry (TOF-LMMS) instru-
pose for which "traditional" optical investigated. However, in practice, ment is provided in Fig. 2. For obtain-
spectrometric methods have been only a few of these combinations are ing inorganic mass spectra, the output

12A Volume48, Number11,1994


rials and the micro-analysis of inte-
grated circuits; for the characterization
Sample
Introduction
P Ion Source
k Mass
Analyzer k Ion detection
and Readout
I of asbestos fibers; and for the study of
environmental aerosols.
Because the sampling and ionization
step takes place in a very short time pe-
riod in LMMS, the mass analyzer must
Inductively Coupled Plasma (ICP) be capable of "capturing" the entire
Microwave Induced Plasma (MIP) mass range of interest without scan-
DC-Glow Discharge (GDL)
rf-GIow Discharge (rf-GDL) Magnetic Sector Mass Spectrometer ning. As a result, the most popular
Spark Source (SS) Quadrupole Mass Filter mass analyzer for LMMS has been the
Laser Ablation/Ionization Time-of-Flight Mass Spectrometer
Ion-beam Quadrupole Ion Trap time-of-flight system. There are, how-
Electrnspray Fourier Transform Mass Spectrometer ever, limitations to TOF spectrometers
rf-Capacitively Coupled Plasma (ocCCP)
Capacitive Microwave Plasma (CMP) for this application, one of which is re-
Direct Current Plasma (DCP)
lated to the limited mass resolution,
which prevents the unambiguous inter-
Fla. I. A generic atomic mass spectrometer. The bottom three ion sources, given in pretation of mass spectra. Fourier
italics, have not yet been reported as being successfully coupled to a mass analyzer. transform mass spectrometers (FTMS)
coupled with laser sampling offer far
higher mass resolution and accurate
of a pulsed, frequency-quadrupled ence is that a laser energy of 10 to 1000
mass assignment. Van Vaeck's re-
Nd:YAG laser (266 nm) is directed nJ is normally used for LDMS, yield-
search group at the University of
onto the sample and focused to a spot ing power densities at the surface of
Antwerp has developed a FTMS-
size of between 0.5 and 2/xm in diame- the sample of 2.12 X 107 to 2.12 × 108
W cm -2. LMMS system using externally gener-
ter. Pulse energies from 100/zJ to 20
ated ions injected into the FTMS cell. 7
mJ are used for ablation and provide ir- LMMS is used for a variety of inor-
radiances of between 2 × 1011 and 4 X They have used this instrument for in-
ganic applications. It has been used in
l 0 j4 W c m -2 at the sample. This tech-
organic microanalysis of mineralogi-
plant and animal biology studies for
nique is closely related to laser desorp- cal samples, thin films, and com-
the determination of the distribution of
tion mass spectrometry (LDMS), pounds deposited from solution. These
elements in biological tissue at the cel-
which is used to obtain mass spectra lular and subcellular level; for research authors also demonstrated the possibil-
from nonvolatile organic solids. 5'6 In in metallurgy, geochemistry, and soil ity of performing inorganic speciation
fact, the two techniques share identical science; for the determination of ionic with this instrument. Binary salts of
instrumentation. The essential differ- contaminants in semiconductor mate- MX, yielded ions such as (MX,)mM +
and ( M X n ) m X - , whereas oxy-salts like
MOX provided cluster ions built from
MO and X moieties. Another benefit of
FTMS-LMMS is that it can be used to
Sampling
Laser study the time of ion formation with re-
spect to the arrival of the laser pulse.
For example, Van Vleck et al. 7 have
shown that ion formation can take
place up to several hundred milisec-
onds after the arrival of the laser pulse.
This factor is one further shortcoming
Reflectron of the TOF spectrometer in this appli-
Time-of-Flight cation.
Mass Spectrometer Laser sampling mass analysis has
the potential for remote sampling ap-
plications and is being investigated for
deployment in space, either from sur-
face rovers or from satellites, for deter-
Ion Detector I mination of the elemental composition
bIon Optics-Ion Extraction of planetary bodies. De Young and
Situ s described an instrument in which
Sample a Nd:YAG laser was used to sample
and ionize a "lunar simulant" material.
The resulting ions were collected in an
Fla. 2. Schematic diagram of a TOF/laser microprobe mass spectrometer instrument. ion trap at the distance of 11.1 m away

APPLIEDSPECTROSCOPY 13A
focal point
from the ionization point and subse- The characteristics of LMMS that the periodic table are in the 0.001 to
quently analyzed by means of a TOF. A make it useful are its microsam- 0.05 ppb range, the exceptions being
paper in this issue discusses this inter- pling capability, low detection limits the halogens, P and S. For rapid quali-
esting space-age application of ( - - 1 0 -2° g), minimal sample prepara- tative analysis of aqueous samples,
LMMS.9 tion requirement, and ability to per- ICP-MS is unequaled. It can provide a
Laser ablation ion-trap mass spec- form molecular analysis. The biggest quick assessment of the elemental
trometry has been investigated by Gill downfall of the technique is the diffi- composition of a sample from Li
and Blades. 10They ablated and ionized culty of quantitation and sample-to- through U in a few seconds.
pin-shaped conductors and noncon- sample reproducibility. There is hope The vast majority of commercial
ductors placed inside the ion trap, that, when an adequate understanding ICP-MS instruments employ a quadru-
which was used as both a trap and a of the fundamental physics and chem- pole mass filter since these devices
mass spectrometer. The use of an ion- istry of the ablation process is attained, offer a good balance between cost and
trap mass spectrometer allows collec- the technique will mature into a routine performance. They have several limi-
tion and storage of ions generated by quantitative analytical tool. tations, however, the most severe
several laser pulses, thus offering the being the lack of simultaneous ion
INDUCTIVELY COUPLED
potential for enhancing sensitivity for monitoring and the relatively low reso-
P L A S M A MASS
trace elemental analysis. For example, lution. The resolution difficulty has
SPECTROMETRY
ions generated from successive firings been addressed primarily through the
of the laser may be stored and integrat- In his 1988 fundamental review, use of magnetic sector instruments. Si-
ed inside the trapping volume to en- Koppenaal listed 123 references to multaneous ion monitoring is impor-
hance the signal-to-noise ratio during plasma source mass spectrometry; in tant for accurate measurement of iso-
readout. The disadvantages of this 1990 there were 163 and in 1992 there topic abundances and for precise
technique include a restricted sample were 291.1'16'17 The majority of these application of isotope-dilution meth-
(pin) geometry, a limited dynamic papers dealt with inductively coupled odology. In a paper in this issue, War-
range resulting from space charge ef- plasma mass spectrometry (ICP-MS). ren et al. 21 approach the problem by
fects inside the trapping volume, and This "explosive growth ''18 is indicative splitting the ion beam and using two
possible sample matrix interferences. of the tremendous appeal of ICP-MS quadrupole mass filters. Another ap-
Additionally, sampling and ionization for elemental analysis. There have proach for simultaneous ion monitor-
are not independent since they result been several recent reviews of ICP- ing that has been explored is the use of
from the same laser pulse. M S , 18-2° and the detailed description of TOF mass spectrometers for ICP-MS.
A very useful variant of straightfor- ICP-MS already offered in these re- The TOF approach offers high
ward LMMS is the combination of views will not be repeated here. throughput (which counters its rela-
laser sampling and resonance ioniza- A schematic diagram of a "typical" tively low duty cycle) and simultane-
tion. 11.12 Resonance ionization has quadrupole ICP-MS instrument is pro- ous ion extraction from the ICP. 22 Duty
been used for many years for ultratrace vided in Fig. 3. Nebulized sample solu- factor in the ICP-TOF instrument is
atomic analysis. 13'14The ion yield effi- tion that is transported to the plasma is further improved by the use of orthog-
ciency for LMMS is typically in the often desolvated and is then atomized onal ion extraction, in which the slow-
range of 10 -2 to l 0 -3, and the sensitivi- and ionized in the ICE The resulting ly moving ion beam from the ICP is ac-
ty of LMMS is often limited by back- atomic ions are extracted from the cumulated and pulsed as a packet into a
ground signals from the matrix. By plasma tailflame into a vacuum enclo- TOF flight tube oriented perpendicu-
tuning a second laser to a resonance sure with the use of a combination of larly to it. Alternative approaches for
transition of desorbed neutrals, one can sampling and skimmer cones. The ions simultaneous ion monitoring are the
use resonance ionization to enhance are then focused and transmitted to the use of a magnetic sector instrument
the ionization efficiency for selected mass spectrometer. The mass-analyzed with an imaging detector 21 or the use of
species so that a detection limit of sub- ions are subsequently detected, and ei- a hybrid mass spectrometer. An inter-
ppb can be achieved. For even better ther they are counted individually or esting example of the latter type of in-
selectivity and sensitivity, multiphoton the ion current they produce is record- strument has been published by Bari-
ionization can be used. Alimpiev et ed in an analog fashion. naga and Koppenaal, 23 who have
al. 15have described the use of laser ab- The four most important attributes coupled a quadrupole mass filter with
lation combined with a two-color (1 + of ICP-MS that make it attractive to an ion-trap mass spectrometer. The
1) resonant multiphoton ionization analytical chemists are: quadrupole mass filter was used as a
(REMPI) technique to determine the bandpass or notch filter to remove ma-
1. Ability to accommodate solution
concentration of B, Al, Fe, and Cr at trix species, and the ion-trap was used
samples easily.
the ppb level in semiconductor sam- as the mass spectrometer. An advan-
2. Low detection limits.
ples. Under conditions of strong satu- tage of this instrument was an apparent
3. Broad elemental coverage.
ration of both transitions, the ioniza- reduction in spectral interferences
4. Isotopic analysis capability.
tion efficiency was estimated to be caused by polyatomic ions.
about unity. Detection limits for most elements in While its positive attributes are ob-

14A Volume 48, Number 11,1994


ly and, with an appropriate sample ma-
Ion Detector nipulation stage, can be used to obtain
spatially resolved elemental concen-
Quadrupole Mass Filter tration information from a sample.
Wang et al: discuss the application of
(LA) ICP-MS to spatially resolved mi-
croanalysis of biological tissue in this
issue. 37
Direct Injection Nebulization. Di-
rect injection nebulization (DIN) ICP-
MS 38 is a method whereby the nebuliz-
er is incorporated directly into the
plasma torch and transports nearly
I J ~ Ion Lenses 100% of the aspirated sample into the
plasma. While it has not received
Skimmer Cone widespread application, it does offer
"- SamplingCone the advantage of low dead volume for
applications where the ICP is coupled
ICP Torch with liquid chromatography and flow
injection.
ICP-MS has become a very impor-
FIe. 3. Schematicdiagram of a quadrupole-based inductively coupled plasma mass tant technique in atomic spectrometry,
spectrometer. and, while there are still some short-
comings, it has clearly joined the main-
vious, there are some negative aspects ization, and application of alternative stream of analytical methods. Some re-
to ICP-MS, and this concern is where methods for sample introduction. Pop- searchers have predicted that it will
much of the current research work in ular methods that have been extensive- completely displace optical emission
the area is being directed. Isobaric ly evaluated to date are itemized and absorption spectrometry, while
overlaps from molecular species below. others see it as being complementary to
formed in the plasma or, more likely, Electrothermal Vaporization, Elec- these techniques.
during the extraction stage are one of trothermal vaporization (ETV) ICP-
the most serious problems. The most M S 26-29 is a technique in which a G L O W D I S C H A R G E MASS
commonly encountered species are ox- graphite furnace is used to vaporize the SPECTROMETRY
ides of the various elements in the sam- sample prior to its transport to the plas-
ple and plasma gas. Their impact on ma. It provides microsampling capabil- As is the case for LMMS, one of the
the mass spectrum is a complex func- ity and a partial reduction of interfer- chief strengths of glow discharge mass
tion of solution composition, plasma ences from oxides since the solvent is spectrometry (GDMS) is the ability to
operating characteristics, and instru- removed prior to the atomization step. sample solids directly. Glow dis-
mentation used. Matrix interferences In this issue, Gray et al. discuss meth- charges have been used for atomic
arise from components in the sample ods for enhancing sensitivity and stabil- spectrometry for many years. Glow
that cause depressions or enhance- ity for (ETV)-ICP-MS. 3° discharge mass spectrometry was in-
ments in the measured signal. The Flow Injection. Flow injection vestigated in the 1960s 39-41 for the
most ubiquitous of these is the "mass- (FI) 31'32 also permits small samples to characterization of sputtered materials
dependent interference" caused by be introduced into the ICP and offers a and for the analysis of semiconductor
space charge effects in the extracted convenient and rapid means of au- materials using a radio-frequency glow
ion beam. 24'25 Space charge effects are tomating sample handling and pre- discharge. There are numerous reviews
caused by the mutual repulsion of ions analysis sample preparation chemistry. of glow discharges for analytical spec-
of like charge in the skimmed ion The coupling of flow injection with an troscopy and G D M S ; 42-47 all the details
beam. The effect is dependent on the ICP is relatively simple. The outlet of of operation and application certainly
total ion beam current and on the na- the FI apparatus is connected directly need not be repeated here. During the
ture of the ions in the beam; hence to the nebulizer inlet. A paper by Dad- mid-1980s there was a strong resur-
lighter ions, such as Li, are affected to farnnia and McLeod in this issue 33 dis- gence of interest in the technique as an
a greater extent than heavier ions. cusses the use of flow injection for pre- alternative to spark source mass spec-
The most common means of sample concentration of uranium in natural trometry (SSMS) for the following rea-
introduction in ICP-MS is using solu- waters. sons:
tion nebulization with a pneumatic L a s e r Ablation. Laser ablation
nebulizer. An active area of interest in (LA) ICP-MS 34-36 sampling offers the 1. The ion beam produced by GDMS
ICP-MS is the development, character- ability to sample solid materials direct- is not as erratic as that provided by

APPLIED SPECTROSCOPY 15A


focal point
spark source MS, thus providing solid) for most elements. Quadrupole ducting samples. Myers et al. describe
superior signal-to-noise ratios. mass filters were first used for GDMS a novel GDMS instrument utilizing a
2. The glow discharge consumes by Coburn and Kay 41 in 1971 and sub- TOF in this issue. 61
much more sample than the spark sequently by Harrison's group. 49'5°
source, thus improving the sam- Typically, the quadrupole mass filter S E C O N D A R Y ION M A S S
pling statistics. provides unit mass resolution with an SPECTROMETRY
3. The ion-energy spread produced abundance sensitivity (the leakage of Since the late 1960s, secondary ion
by the GD is much narrower than an intense peak into the adjacent chan- mass spectrometry (SIMS) has been
that from the spark source; hence, nel 1 amu away) of about 105 to 10 6. acknowledged as one of the most im-
quadruple mass filters can be used Detection limits are in the sub-ppm portant techniques for surface analysis
as the mass spectrometer. range (in the solid). A triple-quadru- and for elemental imaging. 62'63 The
4. The ionization of sample ions is pole GDMS system has also been con- strong points of SIMS are its outstand-
not as strongly influenced by the structed for studying the formation of ing sensitivity, which is in the ppb
matrix as it is for secondary ion polyatomic ions, but it appears to offer range (in the solid), and its extremely
mass spectrometry, resulting in a little commercial potential. The analyt- fine lateral-profiling resolution, which
sensitivity variation among the el- ical capabilities of quadrupole GDMS is on the order of 0.1 to 1/xm.
ements of only about one order of and magnetic sector double-focusing A schematic diagram of a "typical"
magnitude. 48 GDMS have been compared. 51 Raith et SIMS instrument is provided in Fig. 4.
Sampling in a glow discharge takes al. reported that (l) the magnetic sector For obtaining mass spectra with this
instrument was superior for ultratrace instrument, a tightly focused beam of
place when energetic support-gas ions,
analysis as a result of its ability to bet- primary ions, such as Cs +, Ga +, Rb +, O +
formed in the plasma glow adjacent to
ter handle isobaric overlaps; (2) the Ar +, or N +, is directed at the sample
the sample, are accelerated toward and
impact with the sample surface. The two instruments were comparable with surface at an energy in the range of 20
impinging ion either back-scatters respect to relative sensitivity factors; to 60 keV. The outermost atomic layers
from a surface atom or penetrates to a and (3) quadrupole instruments were are sputtered by these impacting ions;
depth of a few angstroms where its ki- superior for depth profiling because of these secondary ions are separated by
netic energy can cause surface atoms to their rapid scan capability. the mass spectrometer and measured.
be ejected, provided that the amount of The interest in glow discharge mass There are three types of SIMS mea-
energy which is transferred exceeds spectrometry is driven by the need for surements:
the binding energy of the surface a fast, sensitive, reliable, relatively in-
1. Dynamic SIMS uses a high-inten-
a t o m s - - a phenomenon called sputter- expensive method for the characteriza-
sity beam of O~ or Cs + to etch the
ing. The sputtered atoms diffuse into tion of metals and semiconductors. 52
surface at a very rapid rate to ob-
the plasma glow, where they are ion- For example, Hutton et al. describe the
tain concentration depth profiles
ized. The ions are extracted from the application of a quadrupole-based
at very high sensitivity. It is used
plasma discharge through an aperture glow discharge mass spectrometer for
primarily for bulk, high-sensitivi-
into the mass analyzer. the determination of trace impurities in
ty elemental analysis.
Glow discharges can be formed with "pure" copper. 53 GDMS is also ex-
2. Imaging SIMS utilizes a narrow
a direct current (dc) or radio-frequency tremely useful for thin-film depth pro-
beam of primary ions (Ga ÷ or In+),
(rf) power supply. The dc glow dis- filing.oEtCh rates of between 0.1 and typically < 1 0 /xm in diameter,
charge is formed between the sample, 1000 A/s are possible, and a depth res- and the beam is rastered across the
which operates as the cathode, and an olution as fine as 0.03/~m has been re-
sample to obtain a topographic
anode. There are a variety of geometric ported; s4 however, the physical size of image of the elemental distribu-
arrangements including the planar GD sources limits their utility for later- tion. It is used primarily to obtain
diode in which the anode and cathode al resolution. Applications of the tech- a "chemical image" of the sample.
are parallel plates, the hollow cathode, nique include the characterization of 3. Static SIMS utilizes a relatively
and the Grimm lamp, which is perhaps thin coatings on metals and laminated low intensity ion beam (Ar ÷) to
the most useful from a sampling stand- materials. GD-MS has also been used etch the sample surface at an ex-
point because the samples need only be for the analysis of compacted powders tremely slow rate; hence the tech-
sufficiently flat and large to form a and dried deposited solutions. A paper nique is used primarily for obtain-
vacuum seal. in this issue investigates the effect of ing chemical information about
Several instrumental configurations solution-deposited anions on ion the surface monolayer. Static
are utilized for GDMS. The first com- yield. 5s SIMS not only is useful for mea-
mercial instrument employed a reverse An area of current interest is the de- suring elemental concentrations
Nier-Johnson double-focusing geome- velopment and exploitation of rf- but also can determine molecular
try. This arrangement provided a typi- G D M S . 56-6° The motivation for this
structure.
cal resolution of 8000 for the mass work is that the use of an rf-GD en-
range from 1 to 280 amu with detection ables the sputtering of insulators, ce- Earliest applications of SIMS uti-
limits in the range of 1 to 10 ppb (in the ramics, and other electrically noncon- lized the dynamic mode for high-sensi-

16A Volume 48, Number 11, 1994


and Horlick in this issue introduces the
Ion Detector method and assesses it for use in atomic
spectrometry. 65
Mass
Spectrometer CONCLUSION
Research at universities, govern-
Ion Gun ~ Electron ment laboratories, and instrument
Flood Gun companies has led to the development
~Beam Pulsing /----., of many new ion sources and improved
mass analyzers during the past few
decades. The coupling of these devices
Focusing has provided the analyst with new and
Optics very powerful instrumentation for ele-
mental analysis. This short paper has
Raster Scanner attempted to outline the characteris-
tics, strengths, and limitations of some
currently popular atomic mass spectro-
FIG. 4. A schematic diagram of a secondary ion mass spectrometer. The electron metric methods: LMMS, ICP-MS, GD-
flood gun is used to maintain charge balance at the sample. MS, SIMS, and ES-MS. As these meth-
ods continue to be improved, detection
tivity trace analysis of the chemical that were originally in the solution. limits lowered, and matrix-effects
composition of surfaces; this is still an These ions can be analyzed with a mass overcome, the range of applications for
important application since the detec- spectrometer. In addition to its obvious which they are used will expand, and,
tion limits for many elements are at the simplicity (no plasma source or ion gun as a result, continued growth in atomic
ppb level. This application has been is necessary), the particular advantage mass spectrometry can be expected for
particularly useful for the measure- that electrospray has to offer is that it many years to come.
ment of dopant and impurity concen- can provide speciation information, Finally, on the front cover of this
trations in semiconductor materials. Of since with the proper conditions, the journal, the reader can find the familiar
course, the ability to do elemental chemical form of the element (ligands logo of the Society for Applied Spec-
imaging, at a resolution as fine as 0.1 and oxidation state) is retained by the t r o s c o p y - - a circle with a white light
/zm, is the major selling feature of gas-phase ions. This issue is becoming beam entering from the top being dis-
SIMS. For surface-science applica- increasingly important in environmen- persed into the familiar rainbow of col-
tions there are a couple of reasons why tal analytical chemistry; as a result, ES- ors punctuated with the letters hu. For
static-SIMS is preferable to LMMS MS could be poised to become a very generations, analytical spectroscopists
and GDMS. First, each area of the sur- important method. A paper by Agnes have understood what this m e a n s - -
face is probed only once; therefore re-
deposition of sputtered material is not
an issue. Second, it can be operated es-
sentially nondestructively.

F
E L E C T R O S P R A Y MASS
SPECTROMETRY
Elemental analysis utilizing electro-
spray mass spectroscopy (ES-MS) is a @ ? eee
new method lurking on the horizon.
Electrospray ionization is a means by / Conductive ~ ~'~1 (~ ~) c~, . . . .
which ions can be obtained directly To Mass
from solution samples. 64 Small droplets Spectrometer
@e@ ss * *
are formed at the tip of a capillary tube
held at a high voltage (3-4 kV). The @e e
very high electric field between the tip [ High Voltage
of the capillary and a counter electrode [ Power Supply _-4
causes the droplets to be enriched in
positive electrolyte ions (Fig. 5). As the I'I'
droplets that are formed evaporate, the
surface charge on the droplets increas-
es, eventually causing droplet fission FIG. 5. Schematicdiagram of the electrospray ion formation process in ES-MS (after
and leading to free ions of the species Kebarle and Tang64).

APPLIEDSPECTROSCOPY 17A
focal point
" a n a l y t i c a l s p e c t r o s c o p i s t s d o it w i t h 18. G.M. Hieftje and L. A. Norman, Adv. Mass lambous, and D. Hall, Mikrochim. Acta 1,
light". These days, however, many Spectrom. 12, 519 (1992). 275 (1987).
19. A. P. Bruins, Mass Spectrom. Rev. 10, 53 44. W. W. Harrison, J. Anal. At. Spectrom. 3,
s p e c t r o s c o p i s t s a r e n ' t d o i n g it w i t h (1991). 867 (1988).
light: t h e y are d o i n g it w i t h i o n s - - 20. R. S. Houk, S. C. K. Shum, and D. R. 45. W. W. Harrison, C. M. Barshick, J. A.
using mass spectrometers. There may Wiedern, Anal. Chim. Acta 250, 61 (1991). Klinger, P. H. Ratliff, and Y. Mei, Anal.
come a time when analytical atomic 21. A.R. Warren, L. A. Allen, H.-M. Pang, R. S. Chem. 62, 943A (1990).
Houk, and M. Janghorbani, Appl. Spec- 46. M. Saito, Bunseki 6, 406 (1985).
s p e c t r o s c o p i s t s w i l l l o o k at t h e l o g o o n
trosc. 48, 1360 (1994). 47. Y. Mei and R. K. Marcus, Trends Anal.
t h e c o v e r o f Applied Spectroscopy as 22. G. M. Hieftje, T. W. Burgoyne, D. P. Myers, Chem., TRAC 12, 86 (1993).
t h e y w o u l d at a h i s t o r i c a l a r t i f a c t in a G. Li, and P. Yang, 1994 Pittsburgh Confer- 48. W. Vieth and J. C. Huneke, Spectrochim.
museum. How quaint . . . they used ence and Exhibition, Chicago, Illinois, Acta 46, 137 (1991).
light! (1994), Paper No. 006. 49. B.L. Bentz, C. G. Bruhn, and W. W. Harri-
23. C. J. Barinaga and D. W. Koppenaal, Rapid son, Int. J. Mass Spectrom. Ion Phys. 8, 409
ACKNOWLEDGMENT Comm. Mass Spectrom. 8, 71 (1994). (1978).
24. S. D. Tanner, L. M. Cousins, and D. J. 50. W. W. Harrison and B. L. Benty, Trace
The author acknowledges funding from the Douglas, Appl. Spectrosc., 48, 1367 (1994). Elem. Anal. Chem. Med. Biol.-Proc. Int.
Natural Sciences and Engineering Research 25. S. D. Tanner, Spectrochim. Acta 47B, 809 Workshop, 1st, 285 (1980).
Council of Canada. (1992). 51. A. Raith, W. Vieth, and J. C. Hunecke, J.
26. C. J. Park and G. E. M. Hall, J. Anal. At. Anal. At. Spectrom. 7, 943 (1992).
1. D.W. Koppenaal, Anal. Chem. 60, I13R Spectrom. 3, 355 (1988). 52. J. E. Cantle, E. F. Hall, C. J. Shaw, and P. J.
(1988). 27. C. J. Park and G. E. M. Hall, J. Anal. At. Turner, Int. J. Mass Spectrom. Ion Phys. 46,
2. Inorganic Mass Spectrometry, Vol. 95, F. Spectrom. 2, 473 (1987). 11 (1983).
Adams, R. Gijbels, and R. van Grieken, 28. J. Wang, J. M. Carey, and J. A. Caruso, 53. R. C. Hutton and A. Raith, J. Anal. At.
Eds. (Wiley, New York, 1988). Spectrochim. Acta 49B, 193 (1994). Spectrom. 7, 623 (1992).
3. J.J. Stoffel(s), J. F. Wacker, J. M. Kelley, 29. D. C. Gregoire, J. Anal. At. Spectrom. 3, 54. A. Raith, R. Hutton, and J. Huneke, J. Anal.
L. A. Bond, R. A. Kiddy, and E P. Brauer, 309 (1988). At. Spectrom. 8, 867 (1993).
Appl. Spectrosc. 48, 1326 (1994). 30. D. J. Gray, S. Wang and R. Brown, Appl. 55. K. R. Hess, C. M. Barshick, D. C. Duck-
4. R.E. Honig and J. R. Woolston, Appl. Phys. Spectrosc. 48, 1316 (1994). worth, and D. H. Smith, Appl. Spectrosc.
Lett. 2, 138 (1963). 31. J. R. Dean, L. Ebdon, H. M. Crews, and R. 48, 1307 (1994).
5. L. Van Vaeck and R. Gijbels, Fresenius' J. C. Massey, J. Anal. At. Spectrom. 3, 349 56. C. Lazik and R. K. Marcus, Spectrochim.
Anal. Chem. 337, 755 (1990). (1988). Acta 47B, 1309 (1992).
6. L. Van Vaeck, J. Bennett, W. Lauwers, A. 32. C.W. McLeod, J. Anal. Atom. Spectrom. 2, 57. C. Lazik and R. K. Marcus, Spectrochim.
Vertes, and R. Gijbels, Microchim. Acta 549 (1987). Acta 48B, 1673 (1993).
111, 283 (1990). 33. S. Dadfarnia and C. W. McLeod, Appl. 58. S. A. McLuckey, G. L. Glish, D. C. Duck-
7. L. Vanvaeck, W. Vanroy, H. Struyf, F. Spectrosc. 48, 1331 (1994). worth, and R. K. Marcus, Anal. Chem. 64,
Adams, and P. Caravatti, Rapid Commun. 34. E. R. Denoyer, K. J. Fredeen, and J. W. 1606 (1992).
Mass Spectrom. 7, 323 (1993). Hager, Anal. Chem. 63, 445A (1991). 59. C. Lazik and R. K. Marcus, Spectrochim.
8. R . J . De Young and W. Situ, Appl. Phys. 35. B. Fryer, S. Jackson, and H. Longerich, Acta 48B, 863 (1983).
Lett. 64, 2943 (1994). Chem. Geol. 109, 1 (1993). 60. J. Shick, R. Charles, A. Raith, and R. K.
9. R.J. De Young and W. Situ, Appl. Spec- 36. S. Jackson, H. Longerich, G. Dunning, and Marcus, J. Anal. At. Spectrom. 8, 1043
trosc. 48, 1297 (1994). B. Fryer, Can. Mineralog. 30, 1049 (1992). (1993).
10. C. Gill andM. Blades, J. Anal. Atom. Spec- 37. S. Wang, R. Brown, and D. J. Gray, Appl. 61. D.P. Meyers, M. J. Heinz, P. P. Mahoney, G.
trom. 8, 261 (1993). Spectrosc. 48, 1321 (1994). Li, and G. M. Hieftje, Appl. Spectrosc. 48,
11. N. S. Nogar and R. C. Estler, Anal. Chem. 38. D. R. Wiederin, R. E. Smyczek, and R. S. 1337 (1994).
64, 465 (1992). Houk, Anal. Chem. 63, 1626 (1991). 62. A. Benninghoven, B. Hagenhoff, and E.
12. T.-Y. Hung and C.-S. Su, Rev. Sci. Instr. 63, 39. J. Sosniak, J. Vac. Sci. Technol. 4, 87 Niehuis, Anal. Chem. 65, 630A (1993).
5299 (1992). (1967). 63. J. C. Vickerman, Chem. in Britain 969,
13. H. Rimke, G. Herrmann, and M. Mang, 40. R. Bes, J. Mahenc, G. Lacoste, and A. (1987).
Mikrochim. Acta 3, 223 (1989). Savall, Method. Phys. Anal. 7, 40 (1971). 64. P. Kebarle and L. Tang, Anal. Chem. 65,
14. D. H. Smith, J. P. Young, and R. W. Shaw, 41. J. W. Coburn and E. Kay, Appl. Phys. Lett. 972A (1993).
Mass Spectrom. Rev. 8, 345 (1989). 19, 350 (1971). 65. G.R. Agnes, I. I. Stewart, and G. Horlick,
15. S.S. Alimpiev, M. E. Belov, and S. M. Niki- 42. W.W. Harrison, K. R. Hess, R. K. Marcus, Appl. Spectrosc. 48, 1347 (1994).
forov, Anal. Chem. 65, 3194 (1993). and F. L. King, Springer Ser. Chem. Phys.
16. D. W. Koppenaal, Anal. Chem. 62, 303R (Second. Ion Mass Spectrom., SIMS 5) 44,
(1990). 75 (1986).
17. D. W. Koppenaal, Anal. Chem. 64, 320R 43. N.E. Sanderson, E. Hall, J. Clark, P. Chara-
(1992).

18A Volume 48, Number 11,1994

You might also like