Pyrolysis Gas Chromatography of Polyethylene

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 6

Fh~ HEWLETT

!.:~ PACKARD

Pyrolysis-GC/MS/IR
Analysis of
Polyethylene
Application Note 228-97
Gas Chromatography
June 1989

Author Introduction (the gas is more viscous. causing


decreased flow with fixed head
Matthew S. Klee Pyrolysis gas chromatography is pressure), standard pressure-
Hewlett-Packard Co. a useful tool for characterizing controlled systems suffer both
Avondale Division polymers. Polymers which are losses in efficiency and increases
not volatile enough for standard in analysis time. The HP 5890
Abstract gas chromatographic analysis can SERIES II used in this study had a
be thermally cleaved into smaller, programmable pressure controller
Pyrolysis gas chromatogra- more volatile fragments which
phy was used to demon- which maintained constant flow
can then be separated and ana- rate during the run by automati-
strate the effect of pyrolysis lyzed. The resulting chromato-
temperature on distribution cally increasing the column head
grams yield characteristic finger- pressure.
of pyrolysis fragments of prints relating to the composition
polyethylene. Cryogenic of the polymer, structural infor- In addition, rapid independent
cooling of the on-column mation s uch as branching and temperature progTamming of the
injector port and oven defects, and information on
(HP 5890 SERIES II) al- cool injector, after pyrolysates
additives. have been trapped, ensured sharp
lowed chromatographic
injection profiles of solute bands.
resolution of pyrolysates Since many fragments are pro-
down to ethene. Detection duced during the pyrolysis proc- Selective detection of pyrolysate
and identification of pyroly- ess, the ability to provide high peaks is very useful for accurate
sates was accomplished resolution separation of these characterization of the starting
with an MSD and IRD con- frag·ments is very important. polymer, the mechanisms of its
nected in parallel. Pro- Capillary column chromatography thermal degradation. and identifi-
grammable pressure control with low dead-volume connections cation of additives and residual
allowed fast analysis while and cryogenic cooling of both the solvents. Therefore. the combina-
maintaining chromato- column and injector ensure tion of a capillary gas chromato-
graphic efficiency. maximum resolution of pyroly- graph with both mass spectral and
sates. Since the flow rate through infrared detectors can be extremely
a capillary column decreases with effective for polymer pyrolysis
increasing oven tempPrature studies.
Experimental Polyethylene was acquired from
Chemical Data Systems, Oxford,
Table 1. Instrumentation and Conditions PA. Pyrolyses were carried out at
700°C and 1000°C using a plati-
Gas Chromatograph: HP 5890 SERIES II
Cold On-Column Injector with Programmable Pressure num coil pyrolysis probe. Sample
and Temperature Control amounts are stated in the caption
Cooling: Cryo Blast, C02 Cryogenic (Oven and Injector) of figure 1 and were less than
Mass Selective Detector: Hewlett-Packard 5970 1 mg.
Mass Range: 19-250 (0 to 8 min)
37-400 (8 min to END) Pyrolyses were carried out in a
EMV: + 0 Vrelative heated interface which is supplied
Transfer Line: 325°C
Data System: Hewlett-Packard 9836C (Rev 3.1.1) for use with HP 5890 GCs (CDS,
Oxford, PA). The standard inter-
Infrared Detector: Hewlett-Packard 5965A
Resolution: 8 em -• (4000 to 700 em -•)
face needle assembly was too large
Transfer A,B: 300°C to fit into the on-column injector
Cell: 300°C and was replaced with a 28-gauge
Data System: Hewlett-Packard 59970C (Rev 1.0) needle via a 1/8-1/16 in. Swagelok
Column: 30m Series 530 JJ., 2.65 JJ.m HP-5 (Part No.19095J-123)
union and graphite!Vespel ferrule.
He Flow: 5 psi@ 0°C, Constant Flow Mode The pyrolyzer interface was used
Oven Program: ooc (1 min) -t275°C (2 min) @ 12°C/min without the standard quartz liner
Injector Program: - 10°C (0.5 min)-t 300°C (1 min) @ 120°C/min and was fed with an auxiliary
Pyrolyzer: Chemical Data Systems (Oxford, PA) Model122
supply of helium carrier gas. This
Interface: 275°C a uxiliary helium was connected to
Probe: Coil an event-controlled solenoid (in
Ramp: Off (ballistic) t he Purge B position) such that it
Interval: 10 sec
could be turned off and on within
Auxiliary He Supply a time program. As long as the
Pressure: 10-12 psi auxiliary pressure supplied to the
OFF Time: 0.5 min
interface exceeds that of the
programmable pressure controller
(PPC), it overrides the PPC and
supplies all of chromatographic
system's carrier gas flow through
the interface. As soon as the sole-
noid turns off the g-as flow to the
interface, th e PPC resumes con-
trol of the column head pressure
and no more helium flows through
the pyrolysis interface.

The inlet of the analytical column


was connected to the on-column
injector and the outlet to one side
of a 1/16-in. zero dead-volume
Swagelok union. To the other end
of t he union , two additional
capillary columns were connected
via a two-holed gTaphiLe!Vespel
ferrule: a sh ort piece of 0.32 mm
HP-5 coated capillary column was
used for connection to the IRD,
and a 2 m x 0.10 mm id blank col-
umn was used for connection to
the MSD.
2
Results and Discussion
1.4E+7 i (A)
The pyrolysis of polyethylene
l . 2E+71
yields three major homologous
series of linear hydrocarbons:
alkanes, alkenes, and dienes (1). ::::::! 1000 c
The ratio of the three compound 6.0E+6

types is dependent on the ~.0E+6 i

pyrolysis temperature, tempera- 2.0E+6


1

0
ture ramp rate, and final hold ..c
time. "c
~
.lJ
a:
!. 4E:+7 j
(B)
Figure 1 compares the total ion } . 2E:+7 1

chromatograms (mass spectral) l. 0E+7~


for 1000°C and 700°C pyrolyses 8.0E+6 1
700 c

I~ J
of polyethylene. Figure 2 is the
total reconstructed chromato-
gram (infrared), acquired simul-
taneously with the mass spectral
::::1~/'-~,
2.0E+6 l '+-1'1
i
ij~,,~~~J~,.-4- -t.A-,J.!l.~-.1!.,.-., !...4-.JlWJl~~C22
.l\....!l-l

information. Cryogenic cooling 2 4 6 8. 10


T1me
12 11
Cmtn. )
16 19 20 22 24 26

of the injector and column al-


lowed chromatographic separa- Figure 1. Total ion chromatograms (mass spectral) for the pyrolysis of
0.35 mg of polyethylene at 1000 °C (A), and 0.45 mg at 700°C (B).
tion of pyrolysis fragments down Numbers above the clusters indicate the length of the carbon
to ethene, as was verified by both chain of the hydrocarbons in the cluster.
mass spectral and infrared
library searches.

Chromatographic efficiency was


maintained through both detec-
tors. Peaks are symmetrical and ~ 2500
peak widths correlate well be-
tween detectors. This is in part 10l2'2 c [ 2000

due to the constant flow opera- ::- 1500


r
tion of the programmable pres- l_l-1-LJLJ_L~~~-;: 1 000
sure controller. The column
head pressure was automatically ~500
IJ

increased during the run from "c0 La


the starting pressure of 5 psi at Q_

0 °C, to about 7.8 psi at the "" ::25 00


ending temperature of 275 °C.
Q:
•r 2000
7 .lO C'
.
: 1 500
c2.'
t 1000
Iv-'L~~00
10 15 25
Tlme <m 1 n . )

Figure 2. Total reconstructed clu·omatograms (infrared) cht·omatograms


acquired simultaneously with the mass spectral data of f igure 1.
FJ/'ff4 HEWLETT
~~PACKARD

References For mm·e in fot·matio n , call your


loca l Hewlett- Packat·d Sa les Office
1. T. P . Wampler, R. J. Levy, or neare st Regional Of fice:
Analyst, 3 (1986) 1065-1067. Un ited States:
Eas t e t·n Region
(30 1) 670-4300
Midwestern Region
(312) 255-9800
Southern Region
(404) 955-1500
Western Region
(8 18) 505-5600
Canada
(41 6) 678-94 30
Austl'ia / Eastern Europe and
Yugos lavia
(02-22) 25-00- 0
Be lgium
(02) 76 1-31 -11
Denmark
(02 ) 81-66-40
Finland
(0358) 0887 21
France
(1) 69 82 60 60
Ge rmany
(6 9) 5000 60
Italy
(02) 92- 36- 91
Ne th erla nds
(2 0) 547 69 II
No n vay
(2) 24-60-90
Spain
(0 1) 637-00- 11
Swe d en
(8) 7 50-20-00
Swit ze rla nd
(057) 31-21-11
U nited K in gdom
(0734) 784774
M idd le East and Africa Operations
(022) 780-71-11
Fa t· Eas t
852- 5-848-77 77
Japan
81-422-55-957,1
Aust r al ia
61-2-888-444
La tin Amel'ica
55- 11 -421-1311
E lsewh e t·e in the \Vorld
(415) 857-5027

Copyrig ht (c) 1989


Hewle tt -Pacl<at·d Compa ny
P t·inte d in LS A 6 /89
43-5952-02 11
Each cluster of peaks corre-
sponds to a one-carbon incre- 1. dlene
ment over the preceding cluster. 2. alkene
With the 5% phenyl-methyl 3. alkane
silicone <HP-5) column used in
G:I!IE+Gl
(A) 2
these experiments, the order of
elution within each cluster was S.I!IE+Gi

diene, alkene, and then alkane, 4.1!1E+6 1


1000 c
with each containing the same 3.0E+6 3
number of carbons. '
2.€1€:+6

The most probable thermal I .I!IE+G


fragment which is produced from •0
c 0.0E:+0
the pyrolysis of polyethylene is a ,c
~

6.1!1E+6
linear alkene with a terminal :1
.<>
a: C,o (B) c,,
double bond. This is born out by S.l!lE+6 c,2
c,3
the size of the center peak in 4. 0E+6

each cluster of the chromato- 3. 0E+6


700 c
grams in figures 1 and 2 and by
2 . 0E+6
both mass and infrared spectra.
The next most probable fragment
~ !~
1.0E+6

type is either the saturated


alkane or a terminal diene,
0.0E+0 /IJ
13 .1!1
A
1 4 .0
A
IS .1!1 16. 0 I 7 ,1!1
Ttme Cm tn.)
depending on the pyrolysis
conditions. Figure 3 is an ex- Figure 3. Expanded total ion chromatograms for t he 1000°C (A) and
panded view of the total ion 700°C (B) pyrolysis, showing the change in relative heights of
chromatograms for 700°C and the first and third peaks in each cluster as a function of
1000°C pyrolyses, and shows pyrolysis temperature.
that the diene peaks are favored
(higher) over the corresponding
alkane peaks after 1000°C
pyrolysis. For the 700°C
pyrolysis, the alkane fragments
are favored (higher peaks) over
the corresponding diene peaks.

4
The length of the hydrocarbon
chain is also dependent on
pyrolysis temperature. The last (C)
significant cluster of peaks in the
1000 oc pyrolysis is C 18, whereas
in the 700°C pyrolysis, the
maximum chain length is at least
C22 (thG analysis was stopped
before elution of all pyrolysates).
The distribution of the pyrolysis
ae I (B)
&9
fragments for the 700°C t:NE ~

pyrolysis is also shifted to longer


chain length, as is evidenced by ;: }; \ ~ ;i ~ a_t
e~------~~------.-~~~~~
the flatter pyrolysis fingerprint. 10Qiil
101!10
WAVENUMBER
201i11!1
l t~- 1 l

These results are expected, since


higher thermal energy breaks IS
more molecular bonds, leading to 10
DIENE (A)
smaller fragments and more
doubly unsaturated hydrocar-
bons. 30Q0 2000
WAVENUMBER ( «•" 1 l

Figure 4 shows infrared spectra


Figure 4. Infrared spectra corresponding to the first (A), second (B), and
for the three peaks in the chro- third (C) peaks in the chromatographic cluster around 12.2 min
matographic cluster around of the 1000°C pyrolysis.
12.2 min. The relative heights of
the major absorption bands due
to the terminal alkene woups
(3084 em -1 and 914 em · 1) can be 1612!12!0
seen to increase in significance 1412!00 2867 cm-1
going from alkane to alkene to 1212!00
10000
diene.
8000
6000
Figure 5 compares selected- 4000
wavelength chromatogTams 200 0
of the -C-H absorption at e~~~~--~~~~~----~~~~~~~~--~~
:0. 0 12.0 J4.e
2867 em · 1 to the =C-H absorp- Time Cm1 n . )

tion at 3084 em - 1 for the 1000°C


pyrolysis. At 3084 em · 1, the
height of the first peak relative
to the second peak in each
Ql
{/)
c
0
::::l 3084 em-!

::::l
cluster is higher than in the a.
{/J

2867 cm·1 chromatogTam since n:" I


the ratio of double bond to single 0~ AJl .A .A
bonds is twice as large. 12.0 14.0

Figure 5. Selected-ion chromatograms from 8 to 15 min in the 1000 °C


pyrogram, comparing the relative h e ights of infra1·ed a bsor·
bances at 2867 em · ' (-C-H) and 3084 em · ' ( = C· H).

You might also like