Professional Documents
Culture Documents
Maintenance, Care, and Service: ACUSON Cypress™ Operator's Manual
Maintenance, Care, and Service: ACUSON Cypress™ Operator's Manual
and Service
11
The proper maintenance, care, and service of your Cypress system will help
provide proper and safe operation. Carefully follow the steps described in this
section.
This chapter includes information about:
Cleaning the system
Maintaining the system
Electrostatic discharge protection
Fuse replacement
If you need additional help, call the Help Desk at: 1-800-888-SIEM.
LENS ACTUATOR
(INSIDE THE DRIVE)
Maintenance
Preventive Maintenance
Basic and regular preventive maintenance can help prevent product failures.
Follow these practices:
Conduct periodic and thorough cleaning.
Avoid spilling liquids or gel onto the control panel or display screen.
Power down when not using the system.
Follow the guidelines when transporting the system (see “Transporting the
System” on page 1-12).
Provide good air flow around the system to allow proper cooling.
Insure that the fan is operating properly. Do not block the fan or vents.
Allow space for the transducer and external connectors so there is not a
forced bend on any of the cables.
Check the ECG cable for cuts or breaks. Replace the cable if there are prob-
lems. New cables can be ordered from Siemens.
Transducer Maintenance
Regularly check transducers for signs of wear or damage. Routinely inspect
your transducers for the following problems:
Cracks in case
Cracks on transducer face
Cuts or gouges on any part of the transducer, including transducer face,
case, cable, and connector
Buckling or bulging of the lens material on the transducer face
Damage to the transducer connector
NOTE: Always remove the MO disk from the MO drive when transporting the
Cypress system.
Frequency of Maintenance
Table 58 provides a guide to the frequency of maintenance procedures. If you
use your system heavily it is advisable to increase the frequency. Safety-related
inspections should be conducted frequently.
Troubleshooting
No Power
If you turn the Cypress system ON and it does not boot up properly:
If you hear the fan and the keyboard is lit, but the screen doesn’t light up, call
your Siemens representative.
If the fan doesn’t come on and the keyboard is not lit:
1. Check to make sure the power cable is plugged in.
2. Check the fuse. See “Fuse Replacement” on page 11-8.
LCD Screen
Fuse Replacement
For 110 V and 220 V applications the Cypress system uses two fuses. The fuse
rating is: 2.5A SB-250 V, 3AG or 5 x 20 mm.
WARNING Before starting this procedure, make sure the power is OFF.
WARNING Risk of fire. Replace fuse as marked.
1. Unplug the power cable from the back of the Cypress system (the door won’t
open if the power cord is connected).
2. Using a flat-nosed screwdriver, open the module door by inserting the screw-
driver at the top of the module and pulling down.
3. Use a screwdriver to pull the red fuse module out.
4. Remove the old fuses from the sides of the module.
5. Insert the new fuses in the sides of the module.
6. Hold the fuses in place with your hand and push the module back in the
Cypress system until it snaps into place.
7. Close the module door. If the door doesn’t close properly, then the red fuse
module is not seated properly.
8. Re-connect the power cable and turn the system ON to make sure the fuses
are working.
INSERT THE
SCREWDRIVER HERE TO
OPEN THE MODULE
FUSE MODULE
(RED)
–o
POWER CORD
Trackball Maintenance
If the performance of the Trackball becomes sluggish, you may need to clean
the Trackball and internal surface of the ball-cage.
NOTE Before cleaning, check the Trackball Speed settings in the Setup
menu to make sure the current settings are set to your preferences.
WARNING Before cleaning the Trackball and internal ball-cage, make sure
the Cypress system power is OFF.
RETAINING RING
PAPER
CLIP
HOLES
TRACKBALL
NOTE If you have a problem completing step 2, insert the end of a paper
clip into one of the two small holes located at the top and bottom of
the retaining ring and turn counterclockwise.
3. Place your hand over the Trackball and lift the keyboard until the Trackball
falls into your hand.
4. Clean the ball with the lint-free cloth by rubbing it carefully. Remove any oil or
dust and set the ball aside until it is time to re-assemble the device. Do NOT
use alcohol, soap, or any other cleaning material on the ball.
BALL
BEARINGS ONCE THE BALL IS
(THREE) REMOVED, CAREFULLY
CLEAN THE
FOLLOWING:
THE INSIDE SURFACE
BALL BEARINGS (3)
LENS
OPTICAL LENS
5. Blow carefully into the ball-cage to dislodge dust and lint that may have
accumulated on the surface or on the lens.
6. Dip the lint-free swab in alcohol. Use the alcohol sparingly; it should not sat-
urate the swab. If necessary, reduce the amount of alcohol on the swab by
applying it to the lint-free cloth or by squeezing the swab tip with tweezers.
You may use any method for reducing the amount of alcohol on the swab as
long as the swab and alcohol are not contaminated.
7. Once the proper amount of alcohol is on the swab, lightly rub the swab on
the inside surface of the ball-cage to remove any contaminants. Look for a
build-up of dirt on the ball bearings. Inspect the surface for cleanliness as
you go. After cleaning, be sure that fluff from the swab is not left on the sur-
face of the ball-cage or optical lens.
8. When all areas have dried completely, replace the previously cleaned ball.
9. Replace the retaining ring.