Professional Documents
Culture Documents
(Care and Use Manual) : Sugar-Pak I Column
(Care and Use Manual) : Sugar-Pak I Column
Sugar-Pak I Column
CONTENTS I. INTRODUCTION
The Waters™ Sugar-Pak™ Column is used for the analysis
I. INTRODUCTION
of sugar products and process streams in beet, cane, and
II. PREPARATION FOR OPERATION starch hydrolysis processing plants; and for incoming quality
a. Column installation inspection of commercial sugar products. Glucose, fructose,
b. Mobile-phase requirements maltose, and maltotriose can be separated from the higher
c. Equilibration oligomers found in typical corn syrups. The course of alcoholic
fermentations can be followed by monitoring the reduction of
III. CARE AND USE
fermentable sugars and the production of alcohol.
a. Precautions
b. Storage considerations The column can be used to separate many monosaccharides,
(more than 72 hours without use) polyols, and alcohols. Disaccharides and larger sugars are also
c. Starting up after a shutdown separated, mainly according to molecular weight. Many other
d. Overnight shutdown useful separations can also be accomplished (e.g., fruit juices,
(less than 72 hours without use) non-nutritive sweeteners containing sorbitol or mannitol, cell,
e. Starting up after an overnight shutdown and cell hydrolyzates, etc.) where a variety of monosaccharides
or sugar alcohols require separation and analysis.
IV. SAMPLE PREPARATION
The 300 x 6.5 mm (I.D.) Sugar-Pak I Column is packed
a. Pre-injection filtration
with a microparticulate cation-exchange gel in calcium
b. Chemical cleanup of samples
form. To obtain results comparable to Waters performance
c. Pre-injection cleanup of proteins and lipids
specifications, certain procedures regarding storage,
d. In-line cleanup of proteins and lipids
handling, and operation should be followed carefully,
e. Cleanup of polysacchacrides
and as explained in this manual.
f. Removal of salts and acids
g. In-line cleanup for samples
not prone to hydrolysis
VII. TROUBLESHOOTING
1
[ CARE AND USE MANUAL ]
3. Bring the mobile phase to its boiling point for a few minutes
Tube Ferrule
just before use, but maintain the temperature between
End must be
straight and smooth 70–90 °C during use. This practice ensures that gases
to achieve maximum
column efficiency (especially CO 2) do not redissolve in the mobile phase.
It also prevents the growth of microorganisms.
Critical distance to be determined by
Compression screw or nut each application (union, column fitting, etc.)
4. Keep the mobile-phase reservoir clean and covered and
supply freshly prepared mobile phase every 24 hours.
Sugar-Pak I Column 2
[ CARE AND USE MANUAL ]
c. Equilibration a. Precautions
To ensure that a new column is fully equilibrated with calcium Normal recommended pressure should not exceed: 2000 psi.
prior to analysis, the following procedure is recommended.
Maximum flow rate: 0.6 mL/min (at or above 70 °C).
1. Install the column with the normal direction of flow ■ DO NOT use calcium chloride, calcium nitrate, or any other
reversed (arrow on label pointing to column inlet tubing).
acidic calcium salt in the mobile phase, as this can corrode
A backflush valve may also be used.
the column and damage the packing.
2. Back flush the column with at least 100 mL of 0.001 M ■ Maximum mobile-phase organic content is 5% V/V. Small
calcium EDTA solution at 90 °C using a flow rate of amounts of acetonitrile, ethanol, methanol, and isopropanol
0.5 mL/min. A concentration of 500 mg/L of calcium in the sample will not effect column performance.
EDTA approximates 0.001 M.
■ Column temperatures should not exceed 95 °C.
3. Reverse the flow direction again (back to the normal Generally, high temperatures provide greater resolution,
direction) and flush the column with the stable baseline but little improvement occurs between 90 °C and 95 °C.
indicating that the column is ready for use. Temperatures below 70 °C do not provide adequate
resolution of many sugars due to the separation of
anomers. For quantification of ethanol, a column
III. CARE AND USE temperature of 75 °C should be used.
Liquid chromatography columns have a finite life influenced by
■ Column temperature changes can be rapid without
their care and use, number of injections, sample and solvent
adverse effects. If the column temperature is below 60 °C,
cleanliness, frequency of solvent changeover, and handling
a maximum flow rate of 0.2 mL/min should be used. Higher
and storage procedures (among other factors).
flow rates may produce excessive backpressure due to the
If a change is observed in the:
higher viscosity of water at lower temperatures.
■ retention of a particular compound,
■ Reverse flow direction through the column (after every
■ resolution between two compounds, or 5 liters of mobile phase). This procedure will prolong
■ peak shape column life significantly. Flow rate changes should be
made slowly and not exceed 0.5 mL/min2 (0.5 mL/min
take immediate steps to determine the reason for the
per minute). It is advisable to change the flow rate in
changes. Until you have made this determination, you must
steps of 0.1 mL/min and wait until the backpressure
not rely upon the results of any separation using the column.
stabilizes before the next step change.
Follow generally accepted procedures for quality control and
methods development when using these columns. ■ The pressure limit setting on the solvent delivery system
Note: Before running the first analysis on your new column, perform the test should be set to a few hundred psi above the normal
sample separation given in the test conditions section. working pressure to avoid any excessive pressure buildup
Before using your new Sugar-Pak I Column for the first time, it while the instrument is unattended.
is recommended that the following procedure be observed: ■ If sugars not prone to inversion are analyzed (e.g.,
1. Connect the column (refer to Section II a). hydrolyzates), it is not essential to use calcium EDTA in
the mobile phase. Pure water can be used but it should
2. Set the flow rate at 0.2 mL/min until the column
be filtered and degassed as recommended. Column
temperature reaches 70 °C.
regeneration and flow direction changes should be
3. Once the column temperature reaches 70 °C, increase carried out regularly for optimum column performance.
the flow rate to 0.6 mL/min in steps of 0.1 mL/min.
Wait for the backpressure to stabilize between each
0.1 mL/min increase.
Sugar-Pak I Column 3
[ CARE AND USE MANUAL ]
e. Cleanup of polysacchacrides
Note: Since C18 -guard columns will adsorb some neutral polysaccharides,
they should not be used for the analysis of such samples.
Sugar-Pak I Column 4
[ CARE AND USE MANUAL ]
Note: Do not use a guard column containing the mixed bed resin for sucrose
analysis, as this will cause severe inversion of the sucrose (or other sugars
prone to inversion). Only use the mixed bed resin in the guard column when
100% pure water is used as the mobile phase. Any calcium EDTA in the
mobile phase will saturate the column. Any other type of guard column that
exceeds the bed volume of the Waters Guard Column (about 0.2 mL) is not
recommended for use with Sugar-Pak I Columns. The increased bed volume
will cause excessive band spreading and loss of resolution.
Sugar-Pak I Column 5
[ CARE AND USE MANUAL ]
VII. TROUBLESHOOTING Allow 20 minutes after the column has reached its working
The Sugar-Pak I Column is tightly packed with a resin in a temperature before bringing the column up to its working
swollen form. If good sample preparation is practiced, most flow rate.
problems that arise will relate to the resin packing. The resin Refer to column regeneration and other cleanup methods
packing is delicate and can be forced into the end frits of the explained in this manual for actual procedures. Remember
column by pump pulsations due to faulty check valves or to switch your column to its original flow direction and allow
by sudden or excessive pressure buildup. Should pressure proper warm-up time before continuing with further analyses.
buildup occur, check for trapped particulates in the in-line
Caution: The resin in Sugar-Pak I Columns is tightly packed
filter or guard column.
and elastic. If the column end frits are removed, the packing
If this is not the problem, it is possible that the resin has may expand out of the column. Thus, opening the end fittings
partially blocked the outlet frit. Should this occur, shut of the column to inspect or clean the end frits (in an ultrasonic
down the system. Reverse the direction of flow through the bath) should be done only as a last resort.
column and begin pumping the mobile phase at 0.1 mL/min.
Waters Corporation
34 Maple Street
Milford, MA 01757 U.S.A.
Waters, Sugar-Pak, Guard-Pak, Sep-Pak, and The Science of What’s Possible are trademarks
of Waters Corporation. All other trademarks are the property of their respective owners.
T: 1 508 478 2000
F: 1 508 872 1990
©2020 Waters Corporation. Produced in the U.S.A. January 2020 WAT085454 Rev D IH-PDF www.waters.com