Download as pdf or txt
Download as pdf or txt
You are on page 1of 6

[ CARE AND USE MANUAL ]

Sugar-Pak I Column

CONTENTS I. INTRODUCTION
The Waters™ Sugar-Pak™ Column is used for the analysis
I. INTRODUCTION
of sugar products and process streams in beet, cane, and
II. PREPARATION FOR OPERATION starch hydrolysis processing plants; and for incoming quality
a. Column installation inspection of commercial sugar products. Glucose, fructose,
b. Mobile-phase requirements maltose, and maltotriose can be separated from the higher
c. Equilibration oligomers found in typical corn syrups. The course of alcoholic
fermentations can be followed by monitoring the reduction of
III. CARE AND USE
fermentable sugars and the production of alcohol.
a. Precautions
b. Storage considerations The column can be used to separate many monosaccharides,
(more than 72 hours without use) polyols, and alcohols. Disaccharides and larger sugars are also
c. Starting up after a shutdown separated, mainly according to molecular weight. Many other
d. Overnight shutdown useful separations can also be accomplished (e.g., fruit juices,
(less than 72 hours without use) non-nutritive sweeteners containing sorbitol or mannitol, cell,
e. Starting up after an overnight shutdown and cell hydrolyzates, etc.) where a variety of monosaccharides
or sugar alcohols require separation and analysis.
IV. SAMPLE PREPARATION
The 300 x 6.5 mm (I.D.) Sugar-Pak I Column is packed
a. Pre-injection filtration
with a microparticulate cation-exchange gel in calcium
b. Chemical cleanup of samples
form. To obtain results comparable to Waters performance
c. Pre-injection cleanup of proteins and lipids
specifications, certain procedures regarding storage,
d. In-line cleanup of proteins and lipids
handling, and operation should be followed carefully,
e. Cleanup of polysacchacrides
and as explained in this manual.
f. Removal of salts and acids
g. In-line cleanup for samples
not prone to hydrolysis

V. COLUMN REGENERATION PROCEDURE

VI. TEST CONDITIONS

VII. TROUBLESHOOTING

1
[ CARE AND USE MANUAL ]

II. PREPARATION FOR OPERATION b. Mobile-phase requirements


This column is packed with a calcium-loaded resin. Hydrogen or other cations
a. Column installation can replace the calcium and cause inversion on the column of sugars such as
sucrose, which are prone to inversion. It is recommended that a small amount
Remove the end plugs from the steel column with a 5/16-inch of calcium be used in the mobile phase to maintain the equilibrium and prevent
open-end wrench and save them for storage when the column inversion. The recommended mobile phase is deionized, bacteria-free water
containing approximately 0.0001 M calcium EDTA* (50 mg/L).
is removed from the system. The column outlet is indicated * Calcium EDTA has several common names:
by an arrow on the label (showing the direction solvent – Calcium di-sodium (ethylene dinitrilo) tetra-acetic acid
should flow). Tighten the fittings to turn past finger tight. – Calcium di-sodium ethylene diamine tetraacetate
DO NOT OVERTIGHTEN – this will damage the fitting seat. – Calcium di-sodium edentate
A properly prepared and assembled compression fitting in ■ Water should be deionized to greater than 2 megohms
good condition is all that is required. resistivity. It is essential that the water used is free of
polyvalent cations, particularly transition and heavy
If tube cutting is required to connect a new column or
metals (e.g., iron).
to improve the end connections on your existing fittings,
follow these steps: ■ Remove bacteria and other particulates from the water
just before use by vacuum filtration.
1. Use a file with a cutting edge to scribe the circumference
of the tube at the desired break. ■ Although the mobile phase will be partially degassed by
vacuum filtration, it is highly desirable that it be thoroughly
2. Grasp the tube on both sides of the scribe mark with
degassed by the following procedure:
cloth-covered or smooth-faced pliers (to prevent marring
the tube surface) and gently work the tube back-and-forth 1. Place the mobile phase in an Erlenmeyer flask on a
until it separates. stirrer/hot plate.

2. Cover the mouth of the flask with aluminum foil to


minimize evaporation.

3. Bring the mobile phase to its boiling point for a few minutes
Tube Ferrule
just before use, but maintain the temperature between
End must be
straight and smooth 70–90 °C during use. This practice ensures that gases
to achieve maximum
column efficiency (especially CO 2) do not redissolve in the mobile phase.
It also prevents the growth of microorganisms.
Critical distance to be determined by
Compression screw or nut each application (union, column fitting, etc.)
4. Keep the mobile-phase reservoir clean and covered and
supply freshly prepared mobile phase every 24 hours.

■ If the system is to be used continually for more than


Figure 1. Ferrule and compression assembly.
one day, place the mobile phase in an Erlenmeyer flask
on a stirrer/hot plate. Cover the mouth of the flask with
3. Inspect the tube at the break for burrs. File the outer edges
aluminum foil to minimize evaporation. Bring the mobile
at an angle to the tube opening. Do not file flat across the
phase to its boiling point for a few minutes just before use,
open tube as this might cause plugging or uneven flow
but maintain the temperature between 70–90 °C during
delivery. Assemble as shown.
use. This practice ensures that gases (especially CO 2) do
4. Slide the compression fitting over the tube, followed not redissolve in the mobile phase. It also prevents the
by the ferrule (large end of the taper first). growth of microorganisms.
5. Seat the ferrule by tightly mating the assembly to the ■ Keep the mobile phase reservoir clean and covered and
fitting seat in which it will be used. An improperly supply freshly prepared mobile phase every 24 hours.
positioned ferrule can form unwanted dead volume
which could result in unintentional sample mixing.
Note: Attach a union in place of the column and flush the lines free
of microparticulates before attaching the column.

Sugar-Pak I Column 2
[ CARE AND USE MANUAL ]

c. Equilibration a. Precautions
To ensure that a new column is fully equilibrated with calcium Normal recommended pressure should not exceed: 2000 psi.
prior to analysis, the following procedure is recommended.
Maximum flow rate: 0.6 mL/min (at or above 70 °C).
1. Install the column with the normal direction of flow ■ DO NOT use calcium chloride, calcium nitrate, or any other
reversed (arrow on label pointing to column inlet tubing).
acidic calcium salt in the mobile phase, as this can corrode
A backflush valve may also be used.
the column and damage the packing.
2. Back flush the column with at least 100 mL of 0.001 M ■ Maximum mobile-phase organic content is 5% V/V. Small
calcium EDTA solution at 90 °C using a flow rate of amounts of acetonitrile, ethanol, methanol, and isopropanol
0.5 mL/min. A concentration of 500 mg/L of calcium in the sample will not effect column performance.
EDTA approximates 0.001 M.
■ Column temperatures should not exceed 95 °C.
3. Reverse the flow direction again (back to the normal Generally, high temperatures provide greater resolution,
direction) and flush the column with the stable baseline but little improvement occurs between 90 °C and 95 °C.
indicating that the column is ready for use. Temperatures below 70 °C do not provide adequate
resolution of many sugars due to the separation of
anomers. For quantification of ethanol, a column
III. CARE AND USE temperature of 75 °C should be used.
Liquid chromatography columns have a finite life influenced by
■ Column temperature changes can be rapid without
their care and use, number of injections, sample and solvent
adverse effects. If the column temperature is below 60 °C,
cleanliness, frequency of solvent changeover, and handling
a maximum flow rate of 0.2 mL/min should be used. Higher
and storage procedures (among other factors).
flow rates may produce excessive backpressure due to the
If a change is observed in the:
higher viscosity of water at lower temperatures.
■ retention of a particular compound,
■ Reverse flow direction through the column (after every
■ resolution between two compounds, or 5 liters of mobile phase). This procedure will prolong
■ peak shape column life significantly. Flow rate changes should be
made slowly and not exceed 0.5 mL/min2 (0.5 mL/min
take immediate steps to determine the reason for the
per minute). It is advisable to change the flow rate in
changes. Until you have made this determination, you must
steps of 0.1 mL/min and wait until the backpressure
not rely upon the results of any separation using the column.
stabilizes before the next step change.
Follow generally accepted procedures for quality control and
methods development when using these columns. ■ The pressure limit setting on the solvent delivery system
Note: Before running the first analysis on your new column, perform the test should be set to a few hundred psi above the normal
sample separation given in the test conditions section. working pressure to avoid any excessive pressure buildup
Before using your new Sugar-Pak I Column for the first time, it while the instrument is unattended.
is recommended that the following procedure be observed: ■ If sugars not prone to inversion are analyzed (e.g.,
1. Connect the column (refer to Section II a). hydrolyzates), it is not essential to use calcium EDTA in
the mobile phase. Pure water can be used but it should
2. Set the flow rate at 0.2 mL/min until the column
be filtered and degassed as recommended. Column
temperature reaches 70 °C.
regeneration and flow direction changes should be
3. Once the column temperature reaches 70 °C, increase carried out regularly for optimum column performance.
the flow rate to 0.6 mL/min in steps of 0.1 mL/min.
Wait for the backpressure to stabilize between each
0.1 mL/min increase.

Sugar-Pak I Column 3
[ CARE AND USE MANUAL ]

b. Storage considerations IV. SAMPLE PREPARATION


(more than 72 hours without use)
■ Turn off the column heater and stop the pump. a. Pre-injection filtration
Filter all samples through a 0.45 µm membrane filter just
■ When the column is cool, remove it from the system and
before injection to remove particulates.
replace the endcaps originally supplied with the column.
The column should not be allowed to dry out during storage. b. Chemical cleanup of samples
The normal mobile phase is suitable as a storage solvent in Substances such as proteins, lipids, polysaccharides, salts
the column. Refrigerate the column at 5 °C (not frozen) to and acids, which could contaminate your column, should
minimize any microbial growth. be removed from samples using the suggested techniques
■ The end fittings in the system should be joined by a outlined in sections d through h.
union after the column is removed. Flush the entire
liquid chromatography system first with water, then c. Pre-injection cleanup of proteins and lipids
methanol, and store it in methanol. Proteins, or lipids such as fats and oils, can contaminate the
column and cannot be removed by washing with organic
c. Starting up after a shutdown solvents. If samples are contaminated with large amounts of
Note: Make sure all methanol is flushed from the system with water before lipid, the excess can be removed by liquid-liquid extraction
the column is reinstalled.
into a solvent such as chloroform.
When starting up a system containing a cold column, turn
on the column heater and start the mobile phase flowing at Residual lipophilic compounds can be removed using a
no more than 0.2 mL/min. Allow the instrument to stabilize Sep-Pak™ C18 Cartridge as explained in the following procedure:
under these conditions for at least 20 minutes after the 1. Fill a 10 mL syringe with 5 mL of methanol and slowly
column reaches working temperature before running the flush the methanol through the cartridge to condition
system under usual working conditions. the Sep-Pak C18 packing.

2. Wash the methanol from the cartridge with two 5 mL


d. Overnight shutdown
portions of pure water.
(less than 72 hours without use)
The column may be left in the system with the flow rate 3. Blow excess water from the cartridge with the syringe.
reduced to 0.2 mL/min. Recirculate the solvent by placing 4. Slowly pass the sample through the Sep-Pak C18 Cartridge.
the effluent line in the supply flask. If it is undesirable to
a. Reject the first 2 mL.
leave the column heater on, shut off the pump and the
column heater and allow them to cool down. b. Collect the next few milliliters for subsequent analysis.

e. Starting up after an overnight shutdown d. In-line cleanup of proteins and lipids


Turn on the column heater and start the mobile phase Protect your column from small amounts of lipophilic material
flowing at no more than 0.1 to 0.2 mL/min. Continue at by using the Waters Guard-Pak™ Holder and Guard-Pak C 18
this flow rate for at least 20 minutes after the column has Inserts. As the Guard-Pak Insert becomes contaminated with
reached its working temperature before setting normal lipophilic sample components, it can be simply replaced and
operating conditions for analysis. the old one discarded.

e. Cleanup of polysacchacrides
Note: Since C18 -guard columns will adsorb some neutral polysaccharides,
they should not be used for the analysis of such samples.

Neutral polysaccharides will pass through the Sugar-Pak


I Column. For some samples, such as starch hydrolyzates,
the analysis of such polysaccharides is desirable and the
use of guard columns containing resins (as described under
“Removal of Salts and Acids”) is recommended. These resins
remove most residual proteins and lipids, but allow neutral
polysaccharides to pass through.

Sugar-Pak I Column 4
[ CARE AND USE MANUAL ]

f. Removal of salts and acids V. COLUMN REGENERATION PROCEDURE


Salts and organic acids present in crude samples can The need to regenerate your Sugar-Pak I Column will vary
co-elute with sugars and interfere with the analysis. according to the relative purity of your samples. The cruder
Typically, salts of strong acids such as sodium chloride, the samples (particularly those containing heavy or transition
elute near the void volume. Salts of weak acids, such as metals or organic acids), the more frequent regeneration will
lactate and acetate, elute later. be required due to inversion of the column.
For some samples of low MW sugars, Waters Sep-Pak You can test for inversion by injecting sucrose. If a tail
Alumina Type A Cartridges are suitable for removing is produced on the sucrose peak, the column should be
salts and acids. retreated as follows:
The recommended procedure is as follows: ■ Make up 500 mL of regeneration solution (500 mg
1. Pass 5 mL of pure water through the Sep-Pak Alumina calcium EDTA/L) and allow it to run through the column
Type A Cartridge. overnight at 0.5 mL/min at 90 °C. Make sure the
direction of flow is reversed before starting regeneration
2. Place 10 mL (maximum) of sample in the syringe and
and returned to normal following regeneration.
slowly pass the sample through the cartridge.
When flow is reversed it is important to maintain solvent
a. Reject the first 5 mL.
temperature in order to ensure proper cleaning. Using cold
b. Collect the next few milliliters of sample for injection. solvent will slow the cleaning process.

Make sure this procedure is suitable for quantitative recovery


for your particular samples.
VI. TEST CONDITIONS
g. In-line cleanup for samples Every column is thoroughly tested and passes strict
not prone to hydrolysis manufacturing specifications. It is recommended that
Use a Waters Guard Column between the column and the you perform a test sample injection, like the one below,
injector to remove salts and acids from the sample. Pack prior to your first analysis. Record the results along with
the guard column with a mixture of strong cation resins (in all relevant instrument settings. These results can then be
the H form) and strong anion resins (in the OH form) with
+ - used for comparison throughout the lifetime of the column
a maximum particle size of 200/400 mesh. These resins are to monitor equipment performance, sample variation, and
available from a number of sources (e.g., CG120 and CG400 column condition.
resins made by Rohm and Haas), but have to be converted
into the appropriate ionic forms just before use. In particular,
the anion-exchange resins are unstable in the OH- form 5.80 DP3 Maltoriose
during prolonged storage. Repack the guard column with
4.22

6.82 DP2 Maltose


fresh resin when necessary. 9.80 Glucose

Note: Do not use a guard column containing the mixed bed resin for sucrose
analysis, as this will cause severe inversion of the sucrose (or other sugars
prone to inversion). Only use the mixed bed resin in the guard column when
100% pure water is used as the mobile phase. Any calcium EDTA in the
mobile phase will saturate the column. Any other type of guard column that
exceeds the bed volume of the Waters Guard Column (about 0.2 mL) is not
recommended for use with Sugar-Pak I Columns. The increased bed volume
will cause excessive band spreading and loss of resolution.

Test sample: 42 DE corn syrup


Flow rate: 0.5 mL/min
Temp.: 90 °C
Mobile phase: Deionized water

Figure 2: A typical chromatogram of sugar standards analyzed with a Sugar-


Pak I Column. Note the high resolution of glucose, maltose, and maltotriose.
Glucose should elute in less than 10 minutes. Maximum backpressure should
be less than 2000 psi.

Sugar-Pak I Column 5
[ CARE AND USE MANUAL ]

VII. TROUBLESHOOTING Allow 20 minutes after the column has reached its working
The Sugar-Pak I Column is tightly packed with a resin in a temperature before bringing the column up to its working
swollen form. If good sample preparation is practiced, most flow rate.
problems that arise will relate to the resin packing. The resin Refer to column regeneration and other cleanup methods
packing is delicate and can be forced into the end frits of the explained in this manual for actual procedures. Remember
column by pump pulsations due to faulty check valves or to switch your column to its original flow direction and allow
by sudden or excessive pressure buildup. Should pressure proper warm-up time before continuing with further analyses.
buildup occur, check for trapped particulates in the in-line
Caution: The resin in Sugar-Pak I Columns is tightly packed
filter or guard column.
and elastic. If the column end frits are removed, the packing
If this is not the problem, it is possible that the resin has may expand out of the column. Thus, opening the end fittings
partially blocked the outlet frit. Should this occur, shut of the column to inspect or clean the end frits (in an ultrasonic
down the system. Reverse the direction of flow through the bath) should be done only as a last resort.
column and begin pumping the mobile phase at 0.1 mL/min.

Table 1. Typical column problems and solutions


Problem Cause Solution
Replace the in-line filter, or shut off the pump. When inlet
Excess pressure In-line filters plugged pressure has dropped to zero, disconnect the in-line filter from
buildup with particulates the column and backflush the prefilter ONLY with mobile phase
at 9 mL/min until the particulates are backflushed out.
Column inlet frit plugged Always use an in-line filter or guard column to prevent this. Reverse
with particulates column to try to wash out particulates at normal flow rate.
Exit frit plugged Reverse flow of mobile phase. Check pump and system
Fluctuating
with resin operation before proceeding.
backpressure
Gas in solvent Check degassing procedure.
Faulty pump Inspect pump check valves and carry out “Ramp Test”
operation (consult the operator’s manual for your pump).
Peak “tailing” H+ ions or heavy metals Follow column regeneration procedure using 500 mg calcium
on sucrose (e.g, Fe) on column EDTA per liter as described.
Reverse column flow direction. This may allow the packing
Band broadening Void in column
bed to reform, thus improving peak shape.
“Spurious” peaks, Elution of salts
Check sample cleanup procedures.
not due to sugars and/or acids
Fittings in bad state Tighten fittings properly. DO NOT OVERTIGHTEN.
Leaking mobile phase
of repair Replace worn fittings.
Variations in flow
Variable elution times Check system for faults, leaks, especially pulsations from pump.
rate or temperature

Waters Corporation
34 Maple Street
Milford, MA 01757 U.S.A.
Waters, Sugar-Pak, Guard-Pak, Sep-Pak, and The Science of What’s Possible are trademarks
of Waters Corporation. All other trademarks are the property of their respective owners.
T: 1 508 478 2000
F: 1 508 872 1990
©2020 Waters Corporation. Produced in the U.S.A. January 2020 WAT085454 Rev D IH-PDF www.waters.com

You might also like