Professional Documents
Culture Documents
Myoglobin Test System: Product Code: 3275-300
Myoglobin Test System: Product Code: 3275-300
B. Myoglobin Tracer Reagent - 13 ml/vial - Icon E Samples may be refrigerated at 2-8°C for a maximum period of
first added to a streptavidin coated well. Biotinylated monoclonal One (1) vial containing enzyme labeled affinity purified two (2) days. If the specimen(s) cannot be assayed within this
and enzyme labeled antibodies (directed against distinct and antibody and biotin labeled monoclonal mouse anti-myoglobin time, the sample(s) may be stored at temperatures of -20°C for up
different epitopes of Myoglobin are added and the reactants IgG in buffer, dye, and preservative. Store at 2-8°C. to 30 days. Avoid repetitive freezing and thawing. When assayed
mixed. Reaction between the various Myoglobin antibodies and C. Light Reaction Wells – 96 wells – Icon ⇓ in duplicate, 0.050ml (50µl) of the specimen is required.
native Myoglobin forms a sandwich complex that binds with the One 96-well white microplate coated with streptavidin and
streptavidin coated to the well. packaged in an aluminum bag with a drying agent. Store at 7.0 QUALITY CONTROL
Reagent (fill)
8. Accurate and precise pipetting, as well as following the exact TABLE 3 B) 1 (13ml) 2 (13ml)
software should be ascertained. time and temperature requirements prescribed are essential. Between Assay Precision (Values in ng/ml) C) 1 plate 2 plates
Any deviation from Monobind IFU may yield inaccurate results.
Example 1 SAMPLE N X σ CV D) 1 (20ml) 1 (20ml)
9. All applicable national standards, regulations and laws,
Sample Well Mean Value including, but not limited to, good laboratory procedures, must Low 10 41.5 2.7 6.5% E) 1 (7ml) 2 (7ml)
RLU(A) Medium 10 131.2 7.2 5.5%
I.D. Number RLU (B) (ng/ml) be strictly followed to ensure compliance and proper device F) 1 (7ml) 2 (7ml)
A1 82 usage. High 10 238.9 10.1 4.2%
Cal A 83 0 *As measured in ten experiments in duplicate over ten days.
B1 84 10. It is important to calibrate all the equipment e.g. Pipettes,
C1 2028 Readers, Washers and/or the automated instruments used
Cal B 2099 10 with this device, and to perform routine preventative 14.2 Sensitivity
D1 2169 The sensitivity (detection limit) was ascertained by determining
maintenance.
E1 8559 the variability of the 0 ng/ml serum calibrator and using the 2σ
Cal C 8530 25 11. Risk Analysis- as required by CE Mark IVD Directive 98/79/EC
F1 8501 - for this and other devices, made by Monobind, can be (95% certainty) statistic to calculate the minimum dose. The assay
G1 21088 requested via email from Monobind@monobind.com. sensitivity was found to be 0.03 ng/ml.
Cal D 21289 50
H1 21289
12.2 Interpretation 14.3 Specificity
A2 55695 The cross reactivity of the Myoglobin AccuLite® CLIA test system
Cal E 55845 150 1. Measurements and interpretation of results must be
B2 55995 to selected substances was evaluated by adding the interfering
performed by a skilled individual or trained professional.
C2 100457 2. Laboratory results alone are only one aspect for determining substance(s) to a serum matrix at the following concentration(s).
Cal F 100000 400
D2 99543 patient care and should not be the sole basis for therapy, The antibody system used did not detect any hemoglobin, CK-MB,
E2 24418 particularly if the results conflict with other determinants. TnI or FABP when tested at very high concentrations. The
Ctrl 1 24845 58.1 presence of lipemia (25 mg/ml); hemoglobin (4.0 mg/ml) and
F2 25273 3. The reagents for the test system have been formulated to
eliminate maximal interference; however, potential interaction bilirubin (2.5 mg/ml) did not affect the assay precision.
G2 59709
Ctrl 2 60424 165.2 between rare serum specimens and test reagents can cause
H2 61138 14.4 Accuracy
erroneous results. Heterophilic antibodies often cause these
A3 17353 The Myoglobin AccuLite® CLIA test system was compared with a
Patient 1 17564 42.3 interactions and have been known to be problems for all kinds
B3 17775 of immunoassays (Boscato LM, Stuart MC. ‘Heterophilic predicate ELISA assay. Biological specimens from population
(symptomatic and asymptomatic) were used. (The values ranged
antibodies: a problem for all immunoassays’ Clin. Chem.
* The data presented in Example 1 and Figure 1 is for illustration 1988:3427-33). For diagnostic purposes, the results from this from N/D – 185 ng/ml). The total number of such specimens was
only and should not be used in lieu of a dose response curve 85. The data obtained is displayed in Table 4.
assay should be in combination with clinical examination,
prepared with each assay. In addition, the RLUs of the calibrators patient history and all other clinical findings.
have been normalized to 100,000 RLUs for the F calibrator TABLE 4
4. For valid test results, adequate controls and other parameters
(greatest light output). This conversion eliminates differences Mean Least Square Coefficient
must be within the listed ranges and assay requirements.
cause by efficiency of the various instruments that can be used to Method (x) Regression Analysis Coefficient
5. If test kits are altered, such as by mixing parts of different kits,
measure light output. The conversion varies between instruments which could produce false test results, or if results are Monobind (y) 23.42 y= 1.02(x) + -1.04 0.983
and should be established for each instrument before using it as a incorrectly interpreted, Monobind shall have no liability. Reference (x) 24.23
factor. 6. If computer controlled data reduction is used to interpret the
results of the test, it is imperative that the predicted values for Only slight amounts of bias between the Myoglobin AccuLite®
Figure 1 the calibrators fall within 10% of the assigned concentrations. CLIA test system and the reference method are indicated by the
150000 closeness of the mean values. The least square regression
13.0 EXPECTED VALUES equation and correlation coefficient indicates excellent method
100000 agreement.
Myoglobin values are consistently the same in plasma and serum.
14.5 High Dose Hook-Effect:
RLUs
50000 However, a serum sample that has been quickly separated from
The test will not be affected by Myoglobin concentrations up to
the red cells is preferred. Myoglobin levels are higher in trained
0 10,000ng/ml in serum. However, samples expected to be over
athletes or people who are used to a daily regimen of strenuous
400ng/ml should be diluted 1:10 and 1:100 in normal pooled
0 100 200 300 400 500 exercise.
human serum and the normal pool assayed along side to obtain a
Myoglobin Concentration ng/ml
Based on the clinical data gathered by Monobind in concordance base value. The base value and dilution factor should be taken
with the published literature the following ranges have been into account to get the corrected concentration of Myoglobin in the
assigned. These ranges should be used as guidelines only: sample.
11.0 Q.C. PARAMETERS
TABLE 1 15.0 REFERENCES
In order for the assay results to be considered valid the
Adult (Normal) < 96 ng/ml
following criteria should be met: 1. Apple Fred S, Christenson RH, Valdes RJ, Andriak AB, Duh
Show-Hong, Feng YJ, Saeed AJ, Johnson Nancy J, Koplen