General Specifications: Model SC100 EXA100 Series Panel Mount Conductivity Analyzer
General Specifications: Model SC100 EXA100 Series Panel Mount Conductivity Analyzer
FEATURES
• 96 x 96 mm panel mount design for indoor use
• Easy access to routine maintenance mode
• Practical functions packed in a compact housing
• Large 4-digit display
• One analog output and two/four contact outputs as
standard
• Simple one touch automatic calibration
• Self-diagnostics (e.g., calibration failure, measuring
range failure, converter failure)
GS 12D11A01-01E-H
1st Edition
2
GENERAL SPECIFICATIONS Contact output
Contact Type: Relay contact output
1. Panel Mount Conductivity Converter Number of contacts: 2 or 4 outputs
Note: Should be specified when ordering.
Model: SC100 Contact action: On/Off
Measurement: Conductivity of a solution Contact functions: Selectable, High, low, high-high,
Measuring range: 0.05µS/cm to 200 mS/cm low-low, high-high/low-low limit alarms, FAIL
Alarm setting resolution: 0.000 µS/cm to 200.0 mS/cm
Indication Contact output hysteresis: 0 to 100% (configurable)
Display: Digital (LED) Contact output delay time: 0 to 200 seconds
Range: 0.000µS/cm to 200.0 mS/cm (configurable)
Indication: Conductivity reading, setting, status, Contact rating:
temperature (range: -10 to 110°C) In case of 2 contact outputs
Input signal S1: 240 VAC 3A or 30 VDC 3A (resistance load),
Conductivity input range: Form C (NC/NO/COM, 3 terminals)
Minimum: 5 µS/cmxK (K: cell constant) S2: 240 VAC 3A or 30 VDC 3A (resistance load),
Maximum: 50 mS/cmxK (K: cell constant) Form A (NO/COM, 2 terminals)
Temperature input: Pt1000/Pt100 In case of 4 contact outputs
Temperature input range: -10 to 110°C When 4 contact outputs specified
S1: 240 VAC 3A or 30 VDC 3A (resistance load),
Transmission signal output Form C (NC/NO/COM, 3 terminals)
Number of output points:1 output, conductivity reading only S2, S3, S4: 240 VAC 3A or 30 VDC 3A (resistance load),
Output signal: 4 to 20 mA DC, isolated Form A, shared common
Load resistance: 600 Ω or less Maximum load current on common is 3A.
Transmission signal range: Configurable within measuring Contact status:
range
Minimum span: 0 to 0.2 µS/cm
Terminal
Maximum span: 0 to 200 mS/cm NC
1
Note: Configurable range is defined by the sensor used. S1 NO
Maintenance output signal: Output hold "enabled/ 2
disabled" selectable
Hold output value: Last measured value/preset value COM 3
(2.0 to 20.8 mA) selectable
S2 4
Fail output signal: Downscale burnout (2 mA)
"enabled/disabled" selectable S3 5
S4 6
COM 7 F011E.eps
Function selected
Contact H, L, HH, LL, HH/LL limit alarms FAIL
Power off Power on Power off Power on
No alarm Alarm No alarm Alarm
S1 NO-COM Open Open Closed Open Closed Open
NC-COM Closed Closed Open Closed Open Closed
S2 Open Open Closed Open Closed Open
S3 when specified Open Open Closed Open Closed Open
S4 when specified Open Open Closed Open Closed Open
T01E.EPS
Note: When a contact is activated, the LED on display panel turns on.
4 S2 I2 14 14
Contact output 2
Contact output 3 5 S3 I1 15 16
Contact output 4 6 S4 16
7 COM2 T1 17 11
L 8 L T2 18 12
N 9 N mA() 19
Analog output
10 mA() 20
Ground to earth Shield
(grounding resistance 100 or less)
Ground to earth
Converter Rear View (grounding resistance 100 or less)
Dimensions
1. Panel Mount Conductivity Converter, SC100
91.6 Max.
125 Min.
1-10
Power supply terminal cover
145 Min.
11 111
92+0.8
0
EXASC100
+0.8
S/cm 92 0
91.6
111
S1 S3 mS/cm
96
S2 S4
CANCEL
SET CAL MEAS
ENT
96 Panel depth: 1 to 10
100
YOKOGAWA HEADQUARTERS YOKOGAWA CORPORATION OF AMERICA Yokogawa has an extensive sales and dis-
9-32, Nakacho 2-chome, 2 Dart Road
Musashinoshi Newnan GA 30265 tribution network.
Tokyo 180 United States
Japan Tel. (1)-770-253-7000 Please refer to the European web-site
Tel. (81)-422-52-5535 Fax (1)-770-251-2088
Fax (81)-422-55-1202 E-mail: info@yca.com (www.yokogawa-europe.com) to contact
E-mail: webinfo@mls.yokogawa.co.jp www.yca.com
www.yokogawa.com your nearest representative.
GS12D11A01-01E-H
GS 12C11A01-01E-H
Subject to change without notice Printed in the Netherlands, 01-405 (A) Q
Copyright©