Professional Documents
Culture Documents
Astm F640-2012
Astm F640-2012
Astm F640-2012
--`,,```,,,,````-`-`,,`,,`,`,,`---
priate safety and health practices and determine the applica- density as measured by a densitometer; in this test method the
bility of regulatory limitations prior to use. expected range is 0.50 to 1.50.
2. Referenced Documents 3.2.5 optical density difference, n—the difference in optical
2.1 ASTM Standards: 2 density units between two regions or objects in an image,
B209 Specification for Aluminum and Aluminum-Alloy reported to at least two digits to the right of the decimal point.
Sheet and Plate 3.2.6 pixel intensity, n—the grayscale level of a pixel
D3182 Practice for Rubber—Materials, Equipment, and Pro- between 0 and 255, as determined by the digital analysis
cedures for Mixing Standard Compounds and Preparing program.
Standard Vulcanized Sheets
E94 Guide for Radiographic Examination 3.2.7 pixel intensity difference, n—the difference in gray-
scale level between two regions or objects in an image,
reported to within the significance capability of the digital
1
These test methods are under the jurisdiction of ASTM Committee F04 on analysis program.
Medical and Surgical Materials and Devices and are the direct responsibility of
Subcommittee F04.15 on Material Test Methods. 3.2.8 user-defined standard, n—a comparison standard se-
Current edition approved Dec. 15, 2012. Published January 2013. Originally
approved in 1979. Last previous edition approved in 2007 as F640 – 07. DOI:
lected by the user.
10.1520/F0640-12. 3.2.8.1 Discussion—This standard may be an existing medi-
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or cal product or a material in a particular form, it may be a
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Standards volume information, refer to the standard’s Document Summary page on commercially available standard, or it may be one developed
the ASTM website. by the user.
--`,,```,,,,````-`-`,,`,,`,`,,`---
6.3.3 Metal, Plastic, or Composite—A metal, plastic, or under the body mimic. As appropriate, the effect of the clinical
composite material of appropriate dimensions may be used. X-ray table shall also be included with use of an appropriate
For example, a 10.0 mm or 15.0 6 0.15 mm thick aluminum material of thickness similar to that used clinically.
sheet might be appropriate. The aluminum sheet shall be
≥99 % in purity, or type 1100 or purer, in accordance with 9.2 X-Ray Exposure—Complete X-ray exposure using con-
Specification B209. ditions typical of those used in the X-ray diagnosis of humans,
6.3.4 Phantom—An apparatus that mimics a portion of the the product, and for the particular area of interest.
body may be used; Note that this apparatus may be as complex 9.2.1 If using film, the exposure shall be of such duration
as a manufactured torso with appropriate densities representing that an optical density of 0.8 to 1.2 is obtained for the
all portions of the anatomy within the torso, or may be as background.
simple defined thickness of water. 9.3 Film Development (If Used)—Develop the X-ray film in
6.3.5 Step Wedge—A step wedge may be used as a user- accordance with the manufacturer’s instructions. If a digital
defined standard, if it has the requisite thickness steps. analysis method will be used, convert the developed film
6.4 Rubber Blankets—Blankets incorporating X-ray absorb- image(s) to digital format using an appropriate digital scanning
ers may be used to mask the image area not covered by the or photographic method.
body mimic material (this prevents undercutting). Lead sheets 9.4 Qualitative Analysis—Visually compare the image(s) of
may also be used for masking. the test specimen and the user-defined standard to the back-
6.5 Back-Scatter Protection, as described in Guide E94, or ground on the film or in the digital image (whether original or
as appropriate with the specific X-ray imaging system. converted from film).
6.6 Densitometer (if used)—The densitometer shall be ca- 9.5 Quantitative Analysis:
pable of measuring the optical density over the range of 0.0 to 9.5.1 Measurement of Optical Density (Film):
APPENDIX
(Nonmandatory Information)
X1. RATIONALE
X1.1 General—The initial development of this standard was preferred test method (of the three described). The standard
based on the need to determine the location in the body of specification for the medical devices may specify materials and
plastic parts of small diameter. This led to a proposed require- values for some parameters of the test method, such as for the
ment for an optical density contrast measured on the medical body mimic and the X-ray test voltage
device. However, radiopacity is a property of many types of
medical devices and its required numerical level is influenced X1.3 Background Density—It has been proposed that, for
by many variables, some of the principal being: type of many applications, an optical density contrast of 0.10 is
material, size, thickness, and configuration; part of body or satisfactory. For larger pieces and thin sections of the human
circulatory system; and X-ray energy applied during the body, a contrast of 0.05 optical density units may be adequate.
procedure.
X1.2 Specification Values—This standard does not have any
requirements for minimum levels of radiopacity, nor is there a
--`,,```,,,,````-`-`,,`,,`,`,,`---
Committee F04 has identified the location of selected changes to this standard since the last issue (F640 – 07)
that may impact the use of this standard. (Approved Dec. 15, 2012.)
(1) Method A was removed and methods B and C were (4) The calibration standard for the digital analysis (standard
combined. that creates a clear and completely opaque areas) was removed.
(2) The location of the test specimen in the X-ray image was (5) The step wedge was identified as a user-defined standard.
modified to allow it to sit at or near the middle versus only the (6) The X-ray exposure procedure was clarified to indicate
middle. completion of X-ray exposure using conditions typical of those
(3) The following was added to the procedure: As appropriate, used in the X-ray diagnosis of humans, the product and for the
the effect of the clinical X-ray table shall also be included with particular area of interest.
use of an appropriate material of thickness similar to that used (7) A requirement to report the material and thickness repre-
clinically. senting the X-ray table, when used, was added.
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the ASTM website (www.astm.org/
COPYRIGHT/).
--`,,```,,,,````-`-`,,`,,`,`,,`---