System Inspection: Samsung Smart Digital Radiography Solution
System Inspection: Samsung Smart Digital Radiography Solution
SMART DIGITAL
RADIOGRAPHY SOLUTION
System Inspection
GF50
v2.0.0
GF50_SI_v2.0.0
Revision History
Version Date History
Hospital
Handler
[Product Information]
Product Name
Software Version
S/N
Symptom
None Main unit HDD DVD/CD
Stored Item
USB Photo Other ( )
None
Outsourced Product
Provide details if applicable ( )
[Administration]
* Please sign inside the box below if you agree to the following.
Data saved on memory devices, for example, hard disks, can be lost during tests and repairs.
Data saved on a product being repaired must be backed up by the customer.
Samsung will not be held responsible for any loss of data not backed up.
Samsung does not use or modify patient information, or provide it to third parties.
[Customer Signature]
GF50 | System Inspection General Safety
Table of Contents
GF50_SI_v2.0.0 5
General Safety GF50 | System Inspection
1. General Safety
Before installing or servicing this system, you must read and comply with all the safety instructions in
the "General Safety" manual.
6 GF50_SI_v2.0.0
GF50 | System Inspection System Inspection
2. System Inspection
2.1. Inspection
• The following explains the regular inspections of the system that need to be performed.
• Perform regular inspections as below in order to ensure the durability of the system, and perform
preventive maintenance for the safety of the system.
Open the ceiling top cover (It can be opened by removing 4 screws marked in the following figure of
the top cover) and manually move the ceiling up/down in a 15 mm interval. Check the spring balance
and wire condition and verify that the wire works properly and has not been cut.
• 4 screws marked below are removed.
Figure 2-1
Figure 2-2
• Push the patient table top from the front and side to see if it can hold weight properly while the
patient table is stopped.
GF50_SI_v2.0.0 7
System Inspection GF50 | System Inspection
• While pressing down the foot switch, move the table top to check the noise.
• Open the PCB cabinet of the generator cabinet and measure the voltage of the terminals.
• Visually check the cables in the generator cabinet and also check if they are connected properly.
• Check if the grids are bent or damaged. Check if there are any foreign objects in the captured X-
ray images.
• When mounting the grids onto the stand and the patient table, check that each unit detects them.
• Touch each part of the THU and check its functionality. Check the touching functionality after
rotating the THU, and check if the THU casing gets in the way. Also check the parts visually.
• Disconnect the HVG cable from the X-ray tube and check if there are any foreign objects or arched
marks on the HVG cable. If there are any foreign objects or arched marks on the HVG cable, wipe
them off with a clean cloth and put silicone gel on the surface.
• If there is a loud noise coming from the X-ray tube rotor, measure the noise level with a DB
measurement device and decide to whether to replace it or not. (80 db and above of noise 1 m
away from the anode)
8 GF50_SI_v2.0.0
GF50 | System Inspection System Inspection
• Press the emergency stop buttons in order and visually check their operation. Disengage the
emergency buttons and check if they have become detached.
• Check the input power at the power distribution panel using a multimeter. (A voltage drop should
be less than ± 10% at the R, S and T phases.)
• Check if a gear backlash occurs while operating the system, and adjust the gears if necessary. If
the gear backlash is severe it may cause a system error.
GF50_SI_v2.0.0 9
System Inspection GF50 | System Inspection
• Elevate the actuators of the patient table and check if there is any noise from them and if they are
operating properly.
• (Measure the noise without a patient. Noise level: less than 75 db)
• The noise is measured while the cover is installed.
• Check if all the screws are firmly tightened. If there are any screws missing, replace them with
spare ones.
• Check if there are any safety screws missing from the collimator.
• Check if the collimator is rubbing against anything and check for noise while rotating the collimator.
• Check if the light field is aligned with the X-ray field by adjusting the field.
10 GF50_SI_v2.0.0
GF50 | System Inspection System Inspection
Classific
No. Items Inspections and Details Inspection Period
ations
Before Every year or comply with your
1 DAP Calibration and record retention
use country's regulations
Before Every year or comply with your
2 AEC Calibration and record retention
use country's regulations
Before
3 Detector Functional tests and calibration Daily, Monthly and Annually
use
Before
4 HVG Functional tests and calibration Daily, Monthly and Annually
use
Sensor While in Check if sensors are functioning
5 While in use
function tests use properly.
While in Check if the lamp is functioning
6 Collimator While in use
use properly.
Operation of
Before Check if the emergency switch is
7 the emergency While in use
use functioning properly.
switch
[Vertical moving part of the tube
stand] Check the residual quantity and
Every year
color of lubricant, and wear status
(one-sided wear, brokenness)
[Horizontal moving part of the tube
stand] Check the residual quantity and
Every year
color of lubricant, and wear status
Lubricant on
(one-sided wear, brokenness)
moving parts Before
8 [Y-axis moving part of the tube stand]
of the tube use
Check the residual quantity and color
stand Every year
of lubricant, and wear status (one-
sided wear, brokenness)
[Rotation of the tube stand around the
horizontal axis]
Check the residual quantity and color Every year
of lubricant, and wear status (one-
sided wear, brokenness)
[Moving part of the receptor on the
patient table]
Check the residual quantity and color Every year
Lubricant on of lubricant, and wear status (one-
the moving Before sided wear, brokenness)
9
parts of the use [Vertical moving part of the patient
patient table table top]
Check the residual quantity and color Every year
of lubricant, and wear status (one-
sided wear, brokenness)
Lubricant on [Vertical moving part of the wall stand]
the moving Before Check the residual quantity and color
10 Every year
parts of the use of lubricant, and wear status (one-
wall stand sided wear, brokenness)
Table 2-1