SAMSUNG

SMART DIGITAL
RADIOGRAPHY SOLUTION

System Inspection
GF50
v2.0.0

GF50_SI_v2.0.0
Revision History
Version Date History

Rev.0 2013.08.22 • Initial release of document

Rev.1 2016.10.26 • Autorun applications added.


Patient Privacy Policy
Treatment of Patients' Personal Information
One of Samsung's social responsibilities is to recognize the importance of patient privacy, and handle
and protect the personal information provided by patients in an appropriate manner.
Patient information stored in repaired and demo products should be handled as follows.
• Patient Information: IDs, names, DOBs, and images are information used to identify patients.
• Patient Information Management: Personal information provided by patients must be protected
during repairs and demonstrations. Do not use, distribute, or lose patient information.

Management of Patient Information on Demo and Trade Products


• Patient information stored on a product during the course of its use must be handled as the patient
desires. (delete or backup)
• When a demo or trade product is collected, patient information must be deleted if so requested by
the patient, to prevent its unintended distribution.

Management of Patient Information on Repaired Hospital Products


• Before removing the product from hospital premises for repairs, all patient information should be
backed up to an external HDD or memory device.
• Before removing the product from hospital premises, each patient should sign the patient privacy
agreement.
Request for Repairs and Patient Privacy Agreement
Date: MM/DD/YY
[Customer Information]

Hospital

Handler

[Product Information]

Product Name
Software Version
S/N
Symptom
 None  Main unit  HDD  DVD/CD
Stored Item
 USB  Photo  Other ( )

 None
Outsourced Product
 Provide details if applicable ( )

[Administration]

Handling date MM/DD/YY


Customer Service Representative

* Please sign inside the box below if you agree to the following.
Data saved on memory devices, for example, hard disks, can be lost during tests and repairs.
Data saved on a product being repaired must be backed up by the customer.
Samsung will not be held responsible for any loss of data not backed up.
Samsung does not use or modify patient information, or provide it to third parties.

I entrust my product to ( ) to be repaired and agree to the above policies.

Date: MM/DD/YY Name: (Signature)

[Customer Signature]
GF50 | System Inspection General Safety

Table of Contents

1. General Safety ..................................................................................................... 6


2. System Inspection .............................................................................................. 7
2.1.Inspection ................................................................................................................................... 7
2.1.1. Tube stand & Wall stand wire condition .................................................................... 7
2.1.2. Noise and wear of the patient table magnet.............................................................. 7
2.1.3. Inspection of the internal power of the generator cabinet ......................................... 8
2.1.4. Inspection of the Grids .............................................................................................. 8
2.1.5. Inspection of the Touching Functionality of the THU ................................................. 8
2.1.6. Inspection of the X-ray Tube ..................................................................................... 8
2.1.7. Inspection of the Emergency Stop Buttons ............................................................... 9
2.1.8. Voltage Rise or Drop ................................................................................................. 9
2.1.9. Inspection of Gear Mesh ........................................................................................... 9
2.1.10. Inspection of patient table elevation ........................................................................ 10
2.1.11. Inspection of the Screws ......................................................................................... 10
2.1.12. Inspection of the Collimator ..................................................................................... 10
2.1.13. Inspection of the foot switch .................................................................................... 10
2.2.Key Inspections and Details .................................................................................................... 11

GF50_SI_v2.0.0 5
General Safety GF50 | System Inspection

1. General Safety
Before installing or servicing this system, you must read and comply with all the safety instructions in
the "General Safety" manual.

6 GF50_SI_v2.0.0
GF50 | System Inspection System Inspection

2. System Inspection

2.1. Inspection

• The following explains the regular inspections of the system that need to be performed.

• Perform regular inspections as below in order to ensure the durability of the system, and perform
preventive maintenance for the safety of the system.

2.1.1. Tube stand & Wall stand wire condition

Open the ceiling top cover (It can be opened by removing 4 screws marked in the following figure of
the top cover) and manually move the ceiling up/down in a 15 mm interval. Check the spring balance
and wire condition and verify that the wire works properly and has not been cut.
• 4 screws marked below are removed.

Figure 2-1

Figure 2-2

2.1.2. Noise and wear of the patient table magnet

• Push the patient table top from the front and side to see if it can hold weight properly while the
patient table is stopped.

GF50_SI_v2.0.0 7
System Inspection GF50 | System Inspection

• While pressing down the foot switch, move the table top to check the noise.

2.1.3. Inspection of the internal power of the generator cabinet

• Open the PCB cabinet of the generator cabinet and measure the voltage of the terminals.
• Visually check the cables in the generator cabinet and also check if they are connected properly.

2.1.4. Inspection of the Grids

• Check if the grids are bent or damaged. Check if there are any foreign objects in the captured X-
ray images.
• When mounting the grids onto the stand and the patient table, check that each unit detects them.

2.1.5. Inspection of the Touching Functionality of the THU

• Touch each part of the THU and check its functionality. Check the touching functionality after
rotating the THU, and check if the THU casing gets in the way. Also check the parts visually.

2.1.6. Inspection of the X-ray Tube

• Disconnect the HVG cable from the X-ray tube and check if there are any foreign objects or arched
marks on the HVG cable. If there are any foreign objects or arched marks on the HVG cable, wipe
them off with a clean cloth and put silicone gel on the surface.
• If there is a loud noise coming from the X-ray tube rotor, measure the noise level with a DB
measurement device and decide to whether to replace it or not. (80 db and above of noise 1 m
away from the anode)

8 GF50_SI_v2.0.0
GF50 | System Inspection System Inspection

2.1.7. Inspection of the Emergency Stop Buttons

• Press the emergency stop buttons in order and visually check their operation. Disengage the
emergency buttons and check if they have become detached.

2.1.8. Voltage Rise or Drop

• Check the input power at the power distribution panel using a multimeter. (A voltage drop should
be less than ± 10% at the R, S and T phases.)

2.1.9. Inspection of Gear Mesh

• Check if a gear backlash occurs while operating the system, and adjust the gears if necessary. If
the gear backlash is severe it may cause a system error.

GF50_SI_v2.0.0 9
System Inspection GF50 | System Inspection

2.1.10. Inspection of patient table elevation

• Elevate the actuators of the patient table and check if there is any noise from them and if they are
operating properly.
• (Measure the noise without a patient. Noise level: less than 75 db)
• The noise is measured while the cover is installed.

2.1.11. Inspection of the Screws

• Check if all the screws are firmly tightened. If there are any screws missing, replace them with
spare ones.

2.1.12. Inspection of the Collimator

• Check if there are any safety screws missing from the collimator.
• Check if the collimator is rubbing against anything and check for noise while rotating the collimator.
• Check if the light field is aligned with the X-ray field by adjusting the field.

2.1.13. Inspection of the foot switch

• Check the contact of the foot switch by pressing the pedals.

10 GF50_SI_v2.0.0
GF50 | System Inspection System Inspection

2.2. Key Inspections and Details

Classific
No. Items Inspections and Details Inspection Period
ations
Before Every year or comply with your
1 DAP Calibration and record retention
use country's regulations
Before Every year or comply with your
2 AEC Calibration and record retention
use country's regulations
Before
3 Detector Functional tests and calibration Daily, Monthly and Annually
use
Before
4 HVG Functional tests and calibration Daily, Monthly and Annually
use
Sensor While in Check if sensors are functioning
5 While in use
function tests use properly.
While in Check if the lamp is functioning
6 Collimator While in use
use properly.
Operation of
Before Check if the emergency switch is
7 the emergency While in use
use functioning properly.
switch
[Vertical moving part of the tube
stand] Check the residual quantity and
Every year
color of lubricant, and wear status
(one-sided wear, brokenness)
[Horizontal moving part of the tube
stand] Check the residual quantity and
Every year
color of lubricant, and wear status
Lubricant on
(one-sided wear, brokenness)
moving parts Before
8 [Y-axis moving part of the tube stand]
of the tube use
Check the residual quantity and color
stand Every year
of lubricant, and wear status (one-
sided wear, brokenness)
[Rotation of the tube stand around the
horizontal axis]
Check the residual quantity and color Every year
of lubricant, and wear status (one-
sided wear, brokenness)
[Moving part of the receptor on the
patient table]
Check the residual quantity and color Every year
Lubricant on of lubricant, and wear status (one-
the moving Before sided wear, brokenness)
9
parts of the use [Vertical moving part of the patient
patient table table top]
Check the residual quantity and color Every year
of lubricant, and wear status (one-
sided wear, brokenness)
Lubricant on [Vertical moving part of the wall stand]
the moving Before Check the residual quantity and color
10 Every year
parts of the use of lubricant, and wear status (one-
wall stand sided wear, brokenness)

Table 2-1

GF50_SI_v2.0.0 11
This document is the technical property of
Samsung Electronics Co Ltd. Using this
document without prior permission or for
purposes other than repairs of the products of
Samsung Electronics without approval may be
subject to legal actions.

Samsung Electronics Co, Ltd. All rights reserved