DigiEye760 smart electrical system

Radiography product global technical support department


Service@mindray.com
Version number:V1.0
No.:DR2502

© 2011 Mindray Confidential


OTY-12002(3.0)
Contents
 1, DR25 product configuration
 2, Connection of equipments
 3, Synchronous control of FPD
 4, LCD control board
 5, Stand column control board
 6, Stand column power board
 7, 4143 power board
 8, Collimator
 9, Generator
 10, Connection of computer
 11, LED screen configuration machine

2 © 2011 Mindray
Confidential
Contents
 1, DR25 product configuration
 2, Connection of equipments
 3, Synchronous control of FPD
 4, LCD control board
 5, Stand column control board
 6, Stand column power board
 7, 4143 power board
 8, Collimator
 9, Generator
 10, Connection of computer
 11, LED screen configuration machine

3 © 2011 Mindray
Confidential
Review: DR25 mechanical system
4.Mechanical structure
1. Workstation

2. HMV-50 Generator 3.Hand switch &


Remote control

4 © 2011 Mindray
Confidential
1, DR25 product configuration
Tube Toshiba Siemens
 X-ray tube E7876X SDR150/30/50-1
Max Voltage 150kV 150kV
 Toshiba E7876X
Focus size 1.2/0.6 1.0/0.6
 Siemens SDR150/30/50-1
Target Angle 12° 12°
 Detector anode heat content 230kHU 330kHU

 Samsung Xmaru 1717G Inherent Filtration 1.3mmAl/75kV 1.0mmAl/80kV


Total filtration 1.8mmAl/75kV 2.0mmAl/80kV
 Trixell Pixium 4143R
Detector Samsung Xmaru 1717G Trixell Pixium 4143R
 Toshiba FDX 4343R
material Gd2O2S Gd2O2S
 Screen type a-Si(Amorphous Silicon) a-Si

 LCD screen (big) power DC:24V, 5V, -8V; 24VDC(Mindray)


supply AC:100V (Samsung)
 LED screen (small)
matrix 3072×3072 2775×2874
pixel size 0.143mm 0.148mm

5 © 2011 Mindray
Confidential
1, DR25 product configuration
 Generator
 Mindray MHV-50:
• Maximum power (50KAV)
• Tube voltage (40~150KV)
• Tube current (10~640mA)

 Collimator
 CX-02:
• Control mode (Auto)
• Max field (43×43cm @ SID=100)
• Min field(3×3cm @ SID=90)
• Filter (1.0 mmAl/ @70kV)

 Optional components
 Ion chamber :
 Patient table:MTAB-065
 Grid:1.3m(default),1.0m,1.8m
 DAP:Vacutec DAP

6 © 2011 Mindray
Confidential
Contents
 1, DR25 product configuration
 2, Connection of equipments
 3, Synchronous control of FPD
 4, LCD control board
 5, Stand column control board
 6, Stand column power board
 7, 4143 power board
 8, Collimator
 9, Generator
 10, Connection of computer
 11, LED screen configuration machine

7 © 2011 Mindray
Confidential
2, Connection of equipments-- general drawing
F U X20 Emt 1, Power distribution
up Dn
B D
X14 X15 X16 X17
X19

M C
X18
Carriage 2, Grounding system
3, Communication
M C P 4, Exposure Sync
X27 X28
X5
5, Motor drive
6, Signal feedback
EOS
X4
PDC
J9 J8 J15 J6 J4
J7 J13 J3 J14
HCB J1 J2 J1

Tuch LCD Left BB Right BB

ARC

HV+

HV-
Mid BB
Collimator
Stand
column X3
EOS
R BB J1
Detector Asa X8

HS
Monitor
J1
4143 PB
J2 Generator
network
synchronization FPD X10
J1
H M KB
power D S
Ion chamber BB J2 ARB
J8 J7
J3 J1/J2 DP
J5 J4
J9 WSCB J12 J7
J6 X6 Emt
L BB J1
X8
J15
MCB
J6 J7 J4 J5

WS J9 220VAC

P
27, 31
power
26, 30 Bucky
J1

interface
9, 14, 40, 45

COM1
C
network card

220VAV
USB network card

8 © 2011 Mindray
Confidential
2, Connection of equipments
F U X20 Emt 1, Power distribution
up Dn
B D

M C
Carriage 2, Grounding system

M C P
X27 X28
X5

PDC
J9 J8 J15 J6 J4
J7 J13 J3 J14
HCB J1 J2 J1

Tuch LCD Left BB Right BB


Mid BB
Collimator
Stand
column
R BB J1
Detector Asa X8

HS
Generator Monitor
network
synchronization FPD X10
J1
H M KB
power D S
Ion chamber BB J2 ARB
J8 J7
J1/J2 DP
J5
J9 WSCB J7
J6 X6 Emt
L BB J1
X8
J15
MCB
J6 J7 J4

WS J9 220VAC

P
27, 31
power J1 Bucky
interface
9, 14, 40, 45

COM1
C
network card

220VAV
USB network card

9 © 2011 Mindray
Confidential
2, Connection of equipments
F U X20 Emt
up Dn
B D

M C
Carriage
3, Communication
M C P 4, Exposure Sync
X27 X28
X5

PDC
J9 J8 J15 J6 J4
J7 J13 J3 J14
HCB J1 J2 J1

Tuch LCD Left BB Right BB

ARC

HV+

HV-
Mid BB
Collimator
Stand
column
R BB J1
Detector Asa X8

HS
Generator Monitor
network
synchronization FPD X10
J1
H M KB
power D S
Ion chamber BB J2 ARB
J8 J7
J3 J1/J2 DP
J5 J4
J9 WSCB J12 J7
J6 X6 Emt
L BB J1
X8
J15
MCB
J6 J7 J4 J5

WS J9

P
27, 31
power J1 Bucky
26, 30

interface
9, 14, 40, 45

COM1
C
network card

220VAV
USB network card

10 © 2011 Mindray
Confidential
2, Connection of equipments
F U X20 Emt
up Dn
B D
X14 X15 X16 X17
X19

M C
X18
Carriage

M C P
X27 X28
X5
5, Motor drive
6, Signal feedback
EOS
X4
PDC
J9 J8 J15 J6 J4
J7 J13 J3 J14
HCB J1 J2 J1

Tuch LCD Left BB Right BB


Mid BB
Collimator
Stand
column X3
EOS
R BB J1
Detector Asa X8

HS
Generator Monitor
network
synchronization FPD X10
J1
H M KB
power D S
Ion chamber BB J2 ARB
J8 J7
J3 J1/J2 DP
J5 J4
J9 WSCB J12 J7
J6 X6 Emt
L BB J1
X8
J15
MCB
J6 J7 J4 J5

WS J9

power Bucky P
interface

COM1
C
network card

220VAV
USB network card

11 © 2011 Mindray
Confidential
Contents
 1, DR25 product configuration
 2, Connection of equipments
 3, Synchronous control of FPD
 4, LCD control board
 5, Stand column control board
 6, Stand column power board
 7, 4143 power board
 8, Collimator
 9, Generator
 10, Connection of computer
 11, LED screen configuration machine

12 © 2011 Mindray
Confidential
3, Synchronous control of FPD
 Exposure synchronization:
Means the X-ray generating device (generator, tube) and the X-ray receiving device (flat-panel-
detector) works at the same time while exposing

 Synchronization control:
Measure is taken for various components working together in the exposure process, in order to mare
sure the controllable, accurate and efficient X-ray output.
Components involved include: handswitch, generator, tube, detector, PC and so on.

 Two kinds of sync control:


 Soft sync (digital signal), digital signal is the synchronization signal of detector.
 Hard sync (analog signal), the synchronization signals of detector are analog signals,
only to be recognized by the PC after AD conversion (the analog converse to digital signal.)

Detector ready Detector ready


(Digital signal) AD (Analog signal) Samsung
Window open converse Window open detector
(Digital signal) (Analog signal)
PC host
Detector ready
(Digital signal) Trixell
Window open detector
(Digital signal)

13 © 2011 Mindray
Confidential
3.1, Synchronous control of FPD--Trixell
 Trixell 4143R

6
Hand 1 exposing
Generator
switch Exposure command

2 5
COM 1 COM 1
GEN_READY Please expose
3
network
Det_READY

PC 4 Detector Open window


network
Windows open

14 © 2011 Mindray
Confidential
3.2, Synchronous control of FPD--Samsung
 Samsung 1717G
Synchronization cable
9 exposing

4
Det_ready

1
5
Hand Exposure command
Generator 6 Detector Open window
Switch
Window open

2 3 7 8
COM 1 COM 1 COM 1 COM 1
Gen_ready Det_ready Window open Please expose

PC Host

15 © 2011 Mindray
Confidential
3.2, Synchronous control of FPD--Samsung
 Samsung 1717 Synchronization cable

D28 D37 D103


(Exposing)
(First gear of HS) (Second gear of HS)

Generator

D105

D104 take down


the voltage

16 © 2011 Mindray
Confidential
3.3, Synchronous control of FPD-- port definition

Trixell
4143R

Synchronization cable

Samsung
1717G

Samsung
Power box

17 © 2011 Mindray
Confidential
3.4, Synchronous control of FPD-- indicator light
Trixell 4143R
on off

D1 24VDC normal supply Power abnormal

D2 working properly panel is not working

D3 Communication normal communication error


End image
D4 Start Image acquisition acquisition

D5 Window open Window off

Samsung 1717G

Power Exposure Working


Unused
(green) Indicate (blue) (yellow)
normal on off Flashing exposure Flashing
abnormal off / / On or Off

18 © 2011 Mindray
Confidential
Contents
 1, DR25 product configuration
 2, Connection of equipments
 3, Synchronous control of FPD
 4, LCD control board
 5, Stand column control board
 6, Stand column power board
 7, 4143 power board
 8, Collimator
 9, Generator
 10, Connection of computer
 11, LED screen configuration machine

19 © 2011 Mindray
Confidential
4, LCD control board
Main function:
 Motor control. Controlling the DC motor, which makes the telescopic up and down movement. (Collected
information from potentiometer);
 Position detection. The carriage on the Stationary rail position. (acquisition information by potentiometer);
 Brake control. On the carriage;
 Tube Head key processing
 Display information on the LCD screen and process signal of touch screen information;
 Serial communication with column control board and collimator;
 Detect Tube Head angle;
 Communication with the PC with network cable.

LCD control board (front) LCD control board( back)


20 © 2011 Mindray
Confidential
4.1, LCD control board--port
serial cable to theJ6 J10 , J12
CPU board socket
bottom of the J3
column X1 LCD screen
backlight cable.

connected to PC
J15
J9
Input signal from
carriage: SID ,
tube high, the
J14
LCD screen
maximum/minimu
LVDS cable.
m limit switch
from telescopic.

J13
to touch screen

J1
to left and right
key board J8
Device driver
J7 J4 J2 Connected with horizontal line: motor,
To collimator clutch, brake
Connected to stand key board J1
column power board J8
21 © 2011 Mindray
Confidential
4.2, LCD control board--pin definition

22 © 2011 Mindray
Confidential
4.3, LCD control board--indicator light
Flashing: Every 1S,means CPU board is
operating normally.
Lit or not lit: indicates abnormal operation CPU
board D1 red

+5V: light blue for


normal work
D29 blue

VCC: light blue for


normal work
D26 blue

+12V: light blue


for normal work
D27 blue

+24V: light blue


for normal work
D28 blue
23 © 2011 Mindray
Confidential
4.3, LCD control board--indicator light
Indicator Meaning Explanation
Flashes to indicate that the CPU board is operating normally; lit or not lit indicates
D1 RUN indicator
abnormal operation CPU board

D8 PCI Reset Indicator light Lit means PCI bus to enter the work state; not lit indicates abnormal PCI bus

D9 direction indicator of Telescopic motor Directional Indicator

Over-current detection indicator of Telescopic


D10 light lit means motor over current is detected; not lit indicates normal motor
cylinder motor

Remote control receiver module RF_VT signal light Lit means that the corresponding received signal; not lit indicates no corresponding
D11
indicator signal

Remote control receiver module RF_D signal light Lit means that the corresponding received signal; not lit indicates no corresponding
D12
indicator signal

Remote control receiver module RF_C signal light Lit means that the corresponding received signal; not lit indicates no corresponding
D13
indicator signal

Remote control receiver module RF_A signal light Lit means that the corresponding received signal; not lit indicates no corresponding
D14
indicator signal

Remote control receiver module RF_B signal Light lit means that the corresponding received signal; not lit indicates no corresponding
D15
indicator signal

D26 VCC Indicator Light lit means power-on, not lit indicates the power supply is off

D27 +12 V power indicator Light lit means power-on, not lit indicates the power supply is off

D28 +24 V power indicator Light lit means power-on, not lit indicates the power supply is off

D29 +5V power indicator Light lit means power-on, not lit indicates the power supply is off

D32 SSD Flash operation indicator Flashing: SSD is for operating; lit or not lit indicates no operation

Flashing indicates that the network is communicating; lit indicates no communication; not
D33 Ethernet port LED
lit indicates a circuit abnormality

Flashing indicates that the network is communicating; lit indicates no communication; not
D34 Ethernet port LEDs
lit indicates a circuit abnormality

D35 VDD LEDs Light lit means power-on, not lit indicates the power supply is off

24 © 2011 Mindray
Confidential
4.4, LCD control board--test point

Test Points Meaning Explanation


TP1 GND
TP2 GND
TP3 GND
TP4 rail_wiper
TP5 drawtube_wiper
TP6 incline_cs#
TP7 incline_mosi
TP8 incline_miso
TP9 incline_sclk
TP10 over_current_shutoff#
TP11 VCC
TP12 +5V_A
TP14 Vref

TP15 Vref

25 © 2011 Mindray
Confidential
Contents
 1, DR25 product configuration
 2, Connection of equipments
 3, Synchronous control of FPD
 4, LCD control board
 5, Stand column control board
 6, Stand column power board
 7, 4143 power board
 8, Collimator
 9, Generator
 10, Connection of computer
 11, LED screen configuration machine

26 © 2011 Mindray
Confidential
5, Stand column control board
Main function
 Communicate to tube head module;
 Receive signal: buttons, potentiometer and other switches;
 The motor drive of detector;
 Detector Handle switch control;
 Detector rotation brake control;

27 © 2011 Mindray
Confidential
5.1, Stand column control board--port

J1/J2 detector key board Power input, connected to Detector rotation Detector up-down
(rotation angle, backlight ) the column power supply brake motor and clutch
board J4
J3 J9 J12 J8

J4 J5 J6 J7

Received from the left, right Serial line to LCD To PC serial Detector up-
rotation key board and the control board communication line down potentiometer con
handle switch signal. (LED screen configuration) nected line.

28 © 2011 Mindray
Confidential
5.2, Stand column control board--pin definition

29 © 2011 Mindray
Confidential
5.3, Stand column control board--indicator light
D3 Detector rotation manually unlock
D4 Detector up-down manual unlocking D9 D10 communication status with the PC
D5 Detector up movement button serial port
D6 Detector down movement button D11 D12 communication status with
D7 collimator light on button D24 5V power indicator LCD head control board
D8 automatic tracing button D25 12V power indicator D21 3.3V power indicator

Error code D14 rotary brake control indicator D23 5V power


D1, D2 indicator D17 lifting clutch control indicator indicator D26: 24V input indicator

30 © 2011 Mindray
Confidential
5.3, Stand column control board--indicator light
Indicator Meaning Indicator Meaning

D36 Program configuration completed (lit) D11 Connect the serial port to send the data

D1 /D2 error code D12 Connect the serial port to receive the head
data

D3 panel rotary manual unlocking D9 Connect the PC serial port to send data

D4 panel rotary manual unlocking D10 Connect the PC serial port to send

D5 panel electric up button D14 Rotary brake control output

D6 panel electric down button D17 出Lift the clutch control output

D7 The beam limiter light field button D16 Lifting motor enable

D8 Automatically atrcking key D19 The direction of the lifting motor

D26 input indicator D23 5V power indicator (potentiometer)

D25 power indicator D21 power indicator

D24 5V12V power indicator

31 © 2011 Mindray
Confidential
5.4, Stand column control board-- test point

Test point Meaning Test point Meaning

TP2 24V power supply TP3 12Vpower supply

TP4 5Vpower supply TP5 5V power supply (potentiometer)

TP6 3.3Vpower supply TP7 1.2Vpower supply

TP1 Motor drive PWM

32 © 2011 Mindray
Confidential
Contents
 1, DR25 product configuration
 2, Connection of equipments
 3, Synchronous control of FPD
 4, LCD control board
 5, Stand column control board
 6, Stand column power board
 7, 4143 power board
 8, Collimator
 9, Generator
 10, Connection of computer
 11, LED screen configuration machine

33 © 2011 Mindray
Confidential
6, Stand column power board
Main function
 Input 220VAC, output 24VDC to stand column control board;
 Put 220VAC to the 4143 power supply board for Trixell detector;
 EOS control

34 © 2011 Mindray
Confidential
6.1, Stand column power board--port

To Bucky J1
interface board J1.
the input 220VAC.

J6
The 220VAC output.
To 4143 power
board for detector

J7, J5
Two EOS to stand
column and tube head

Output to J4
column control board
24V.

J9
Output to LCD control
board 24V.

35 © 2011 Mindray
Confidential
6.2, Stand column power board--pin definition

36 © 2011 Mindray
Confidential
Contents
 1, DR25 product configuration
 2, Connection of equipments
 3, Synchronous control of FPD
 4, LCD control board
 5, Stand column control board
 6, Stand column power board
 7, 4143 power board
 8, Collimator
 9, Generator
 10, Connection of computer
 11, LED screen configuration machine

37 © 2011 Mindray
Confidential
7, 4143 power board
Main function
The input of 220VAC into 24VDC power supply to Trixell detector

38 © 2011 Mindray
Confidential
Contents
 1, DR25 product configuration
 2, Connection of equipments
 3, Synchronous control of FPD
 4, LCD control board
 5, Stand column control board
 6, Stand column power board
 7, 4143 power board
 8, Collimator
 9, Generator
 10, Connection of computer
 11, LED screen configuration machine

39 © 2011 Mindray
Confidential
8, Collimator
CX-02 collimator
 CX-02 is Modified from CX- 01.
 Using high-brightness LED lamps replace the original lamp. It’s only used 24vDC, no need
the 30VAC, so there is no power board, only has an control board.
 Remove the LCD display screen.
 Remove the Copper filtration motor drive assembly, cancel the Copper filtration switching
button and Memory button.
 Remove the laser lamp.

40 © 2011 Mindray
Confidential
8.1, Collimator control board--port
Lead window position LED power LED lamp temperature Serial communication
signal output 5V. switch to LCD control board

J11 J9 J12 J8

Fan power
J6
output 24V

24VDC power J2
input

Lead window J4
stepper motor
control

41 © 2011 Mindray
Confidential
8.2, Collimator control board--pin definition

J12 J7
温度开关 2 1
3 2
3
4
5
6
J4 009-000966-00 转接接口 电机自带线材 步进电机
主控板 1 1 1
转接接口 009-000963-00 J6 2 4 4
1 1 1 LM LM(X)
风扇 3 6 6
2 X7 2 2
5 3 3
3 3 3 7 4 4
9 1 1
J5 6 FM FM(CU)
11 6
1 13 3 3
2 16 1 1
18 4 4
19 6 TM 6 TM(Y)
RS232 J8 20 3 3
1
2
转 3 传感器自带线
接 J11
机头控 4 1
制板 端
2
子 24V J2 3
光耦传感器Y
1 4
GND 2
J11
5
6
7
光耦传感器X
J9
V+ 8
1
LED V- 2

限束器整机连接图

42 © 2011 Mindray
Confidential
8.3, Collimator control board--indicator light
lights lit when there is
output of J9
D31

D5
+24VDC
power on

D6
The Light is on after FPGA is
configured
43 © 2011 Mindray
Confidential
8.3, Collimator control board--indicator light
Test point Meaning Indicator Meaning

TP1,TP2,TP4,TP8 GND D6 The Light is on after The FPGA is configured

TP21 1.2V D19 Light lit indicates The Master control


board issued orders to open view lamp power
supply board

TP19 3.3V D13 Light lit indicates 0.1mm CU zero trigger

TP20 5V D12 Light lit indicates 0.2mm CU zero trigger

TP18 24V D14 Light lit indicates collimator rotation mechanical


zero trigger

TP3 CAN_L signal D5 Light lit indicates 24V power is on

TP5 CAN_H signal D4 Light lit indicates 5V power is on

D3 Light lit indicates 3.3V power is on

44 © 2011 Mindray
Confidential
8.4, Collimator control board--test point

45 © 2011 Mindray
Confidential
Contents
 1, DR25 product configuration
 2, Connection of equipments
 3, Synchronous control of FPD
 4, LCD control board
 5, Stand column control board
 6, Stand column power board
 7, 4143 power board
 8, Collimator
 9, Generator
 10, Connection of computer
 11, LED screen configuration machine

46 © 2011 Mindray
Confidential
9, Generator
 Can match Two kinds of tube: Toshiba E7876X and Siemens SDR 150/30/50-1;
 Input Power:
voltage: The three-phase voltage: 380/400/415/440V ± 10%
Frequency: 50/60 ± 1HZ;
 Maximum power ≥ 50kW, maximum current ≥ 640mA, maximum output voltage ≥ 150kV;
 With switch button, you can control other subsystems power off;
 Can provide external AC power supply 220V ± 10% output power ≥ 200W;
 External power supply can provide DAP, 24V±10%, the output power of ≥ 12W;
 Straight connector with hand brake, used to control exposure;
 Enables stitching function;
 Support AEC mode exposure.

Mindray generator is compatible with a variety of Door


interlock switch and Warning light .

Warning light light cable port (Bucky shorting line Door


voltage interface board) interlock
220V~ 24, 28 23, 32 37, 38
110V~ 24, 36 23, 32 37, 38
24V- 24(+), 42(-) 23, 46 37, 38
12V- 24(+), 42(-) 23, 48 37, 38

47 © 2011 Mindray
Confidential
9, Generator--port

48 © 2011 Mindray
Confidential
9, Generator--pin definition
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27 N/A
28
Assistant 29
power 30
supply 31
Board 32
33

old new
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48

49 © 2011 Mindray
Confidential
9, Generator--pin definition

M
C
B

50 © 2011 Mindray
Confidential
9, Generator-- anode cable
Rotating anode winding:
When the rotation with low-speed, the pulse frequency is 60Hz;
When high-speed rotation, the pulse frequency 180Hz

main
shift

Rotating
COM tube
TH1
anode board
J2 TH2
GND

75 ℃ protection switch (normally closed):


When the temperature reaches 75 degrees, the
exposure is prohibited.
6 Pin cable Ground wire

51 © 2011 Mindray
Confidential
9, Generator-- ion chamber port

52 © 2011 Mindray
Confidential
Ionization chamber
 Ionization chamber type:
1) Solid ion chamber
2) Gas ion chamber

Data mode:
1) Analog mode
2) Digital mode

Square wave signal


Ramp (analog)
(analog)
Mode Integration Analog
1.0 Control signal Control signal
module ion chamber

Generator
Control signal
Mode Digital
2.0 Pulse signal (digital) ion chamber

Integration module:
1) Internal integration
2) External integration

53 © 2011 Mindray
Confidential
9, Generator -- Warning light and Door interlock switch

Generator
Generator

Lamp

interlock

Shorting
line

Warning light light cable port (Bucky shorting line Door


voltage interface board) interlock
220V~ 24, 28 23, 32 37, 38
110V~ 24, 36 23, 32 37, 38
24V- 24(+), 42(-) 23, 46 37, 38
12V- 24(+), 42(-) 23, 48 37, 38

54 © 2011 Mindray
Confidential
Contents
 1, DR25 product configuration
 2, Connection of equipments
 3, Synchronous control of FPD
 4, LCD control board
 5, Stand column control board
 6, Stand column power board
 7, 4143 power board
 8, Collimator
 9, Generator
 10, Connection of computer
 11, LED screen configuration machine

55 © 2011 Mindray
Confidential
10, Connection of computer

To generator(COM1)

To head control board


(USB network)

To detector
(integrated network)

To stand column control board(COM2)


Only for LED screen

To PACS/Printer
(independent network)

56 © 2011 Mindray
Confidential
10, Connection of computer

57 © 2011 Mindray
Confidential
10, Connection of computer--IP address settings

USB NIC 1
LCD control board
1 (222.222.222.222)
(222.222.222.223)
(255.255.255.0)

USB NIC 2
Remote diagnosis
2 ( whatever )
( whatever )
(255.255.255.0)

Independent NIC PACS/Printer


3 (192.168.37.101) (192.168.37.102)
(255.255.255.0) (255.255.255.0)

Integration NIC
Detector
(192.168.37.131)Samsung
4 (192.168. 37.130)Samsung
(192.168. 37.131)Trixell
(192.168. 37.130)Trixell
(255.255.255.0)
NIC: Network Interface Card
58 © 2011 Mindray
Confidential
Contents
 1, DR25 product configuration
 2, Connection of equipments
 3, Synchronous control of FPD
 4, LCD control board
 5, Stand column control board
 6, Stand column power board
 7, 4143 power board
 8, Collimator
 9, Generator
 10, Connection of computer
 11, LED screen configuration machine

59 © 2011 Mindray
Confidential
11, LED screen configuration machine

LED screen control board(back) LCD screen control board(front)


60 © 2011 Mindray
Confidential
11, LED screen configuration machine
F U X20 Emt 1, Power distribution
up Dn
B D
X14 X15 X16 X17
X19

M C
X18
Carriage 2, Grounding system
3, Communication
M C P 4, Exposure Sync
X27 X28
X5
5, Motor drive
6, Signal feedback
EOS
X4
PDC
J9 J8 J15 J6 J4
J7 J13 J3 J14
HCB J1 J2 J1

Tuch LCD Left BB Right BB

ARC

HV+

HV-
Mid BB
Collimator
Stand
column X3
EOS
R BB J1
Detector Asa X8

HS
Monitor
J1
4143 PB
J2 Generator
network
synchronization FPD X10
J1
H M KB
power D S
Ion chamber BB J2 ARB
J8 J7
J3 J1/J2 DP
J5 J4
J9 WSCB J12 J7
J6 X6 Emt
L BB J1
X8
J15
MCB
J6 J7 J4 J5

WS J9 220VAC

P
27, 31
power
26, 30 Bucky
J1

interface
9, 14, 40, 45

COM1
C
network card

220VAV
USB network card

61 © 2011 Mindray
Confidential
11, LED screen control board
Main function:
 Position control. Controlling the DC motor, which makes the telescopic up and down
movement. (Collected information from potentiometer);
 position detection of carriage on stationary rail (acquisition by potentiometer);
 brake control on the carriage;
 Tube Head key processing
 display information on the LED
 serial communication with column control board
 serial communication with collimator
 Tube Head angle detection;

LCD screen control board(front)


62 © 2011 Mindray
Confidential
11, LED screen control board
J4
24v power input,
from stand
column power
J6 board J8
serial cable to the
bottom of the
column X1 J7
To
collimator
J9
Input signal from J8
carriage: SID , device driver line:
tube high, the motor, clutch and
maximum/minimu brake
m limit switch
from telescopic.

J1
to left and right
key board

63 © 2011 Mindray
Confidential
11, LED screen control board--indicator light

D5 blue
+24VDC power
supply

D6 blue
+12VDC
power supply

D7 green
The over-current off
indicator light of the
motor

64 © 2011 Mindray
Confidential
11, LED screen control board--indicator light

Indicator color meaning


D5 blue +24V Input power indicator
D6 blue +12 Input power indicator
D7 Green over current of motor
D15 blue VCC power supply

D16/D17 white Automatically tracking the keypad backlight lamp


D18 Green NCEO,FPGA configuration completed
D19 red Software encryption dog lamp
D21 Green Motor direction indicator
D23 blue RF_VT
D24 blue VDD
D37/D38 white Move_Up
D39/D40 white Move_Down
D43 blue RF_D
D44 blue RF_C
D45 blue Rf receive module RF_A signal lamp
D46 blue RF_B

65 © 2011 Mindray
Confidential
11, LED screen control board--test point
Test point meaning explain
TP1 GND
TP2 Incline_sclk
TP3 Incline_miso
TP4 Incline_mosi
TP5 Incline_cs_n
TP6 Rail_wiper
TP7 Drawtube_wiper
TP8 Dectector_TX
TP9 Collimator_TX
TP10 Dectector_RX
TP11 Collimator_RX
TP12 Overcurrent_shutoff_n
TP13 GND
TP14 GND
TP15 GND
TP16 VCC

66 © 2011 Mindray
Confidential
Contents
 1, DR25 product configuration
 2, Connection of equipments
 3, Synchronous control of FPD
 4, LCD control board
 5, Stand column control board
 6, Stand column power board
 7, 4143 power board
 8, Collimator
 9, Generator
 10, Connection of computer
 11, LED screen configuration machine

67 © 2011 Mindray
Confidential
Version information
SN Writer Date Description
1.0 Benson 2-20-2014 DigiEye760 smart Training Material

68 © 2011 Mindray
Confidential
Thanks!