Professional Documents
Culture Documents
The Innovative Versatile Transport Incubator Works With Any Environment Efficiently
The Innovative Versatile Transport Incubator Works With Any Environment Efficiently
nice 3000
Orchid
nice 3000
An infant's smile is
infinitely precious.
ACCESSORY SHELF
Auxillary tray, Shelf for other additional equipment
as Pulse Oximeter, Respirator with Blender, Infusion
Pump, Oxygen Analyzer.
Double wall Standard
Double wall Hood to decrease radiant heat loss
Specifications Battery specifications
Electronics : Micro Processor Based Design with Incubator : Single Battery (Standard),
self Type :12V, DC, 42Ah Gel-Type Battery
test function Battery Rating : Incubator maintains a differential of
Modes of Operation : Skin Servo and Air Servo. 23 0C between ambient and set point
Hood : Double wall (standard) for 3Hrs - 4Hrs on 42Ah battery
at full heater power.
Temperature Probe Life expectancy : 200 complete charge/discharge
Skin Probe : Thermistor Based Cycles minimum
Air Probe Skin Probe/Air probe : Thermistor Based Charge Time : 10 hours per battery from full
Accuracy : +/-0.2 0 C discharge.
Probe Interchangeability : +/-0.2 0 C Safety Alarms
Temperature Calibration : No need of calibration High temperature : If skin > 0.5 0 C from the set temp
Skin low : If skin < 1 0 C from the set temp
Thermoregulation Air temperature high : If skin > 0.5 0 C from the set temp
Skin mode control range : 30-37.5 0 C over ride 37.6 - 390 C Air low : If skin < 3 0 C from the set temp
Air mode control range : 23-390 C Sensor fail : If sensor fails
Skin / Air Mode Low Battery : If DC < 10.5 0 V DC
Temperature resolution : 0.10 C Heater fail, Power fail, Air Flow fail, >39 0 C Air/Skin temperature
Skin / Air display reading range : 20-50 0C
Temperature rise time : 100 C within 45 minutes. While Silence/Reset : Silences the audible portion of
ambient Temperature is 250 C alarms for 20 minutes except power
0
Temperature Variability :<-1.00 C fail, sensor fail, heater fail,
Temperature overshoot :<- 2.00 C Air flow fail.
Temperature uniformity : <1.0
- C Electrical
Air velocity over the bed : <0.3m/sec
- AC Power Requirements : 230 V / 50 Hz.
DC Power : 11-13 V/220 VA (max)
O2 Cylider mounting : The tank mount permits mounting Observation Lamp : CFL lamp 10 cm above mattress
gas cylinders with diameter of External battery supply : 12V, 10Ah
up to(12 cms) up to (86 cms) in length. Fuse : F3AL/215V
Humidity Pad : Sponge Physical characteristic
IV stand : Height adjustable Stainless steel Dimension with Stand : 96.3 cm - 132 cm(H) X 55 cm (W)
Examination Light : LED with high intensity X 110 cm (L)
Optional Features : Pressure Regulator, Flow meter , Dimension without Stand : 58 cm(H) X 49 cm (W) X 97 cm (L)
Cylinder and Oxygen driven Slow suction Weight with Stand : 90 kg
O2 Concentration range : 21% to 58% minimum Weight without Stand : 71.4 kg
Humidity capacity : 50% to 70% Distance from Bed to
Noise Level : < 60 dBA Vertical hood : 30 cm
IV Pole Maximum Load : 1.5 kg
Performance Characteristics Mattress Load : 5 Kg
Humidity Pad : Holds 400 ml sterile water with no Accessory Tray : 3 Kg
significant spillage for upto
450 tilt in either direction. Collapsible Stand
Air filter : Removes >99% of air borne particle Material : Stainless Steel
greater than 0.5 micron diameter. Mobility : 4 Nos. 6 inches castors (2 brakes)
Relative Humidity : 50% to 70% for 10-12 hours using Collapsible : Yes
humidity pad Gas spring : 4 nos.
Auto Loading Stretcher (Optional) : for different type of Ambulance
Control Displays IEC 60601-1 Specification
Battery Condition Status : 4 LED indicators Type of Protection against Electric shock : Class 1
Power Mode : Illuminates AC, DC Degree of Protection against Electric shock : Type BF(skin probe)
74-00-005, Rev 02
Heater Power : 4 LED indicates heater power : Type B(mattress)
25% to 100% Mode of Operation : Continuous
Skin Temperature 0 C : Displays infant temperature Protection against Hazardous of Explosion : Not Protected
Air Temperature 0 C : Displays incubator air temperature Protection against Increases of Liquid : Not Protected
Air set Temp : Air set required temperature Product Classification : Class IIb
Skin set Temp : Skin set required temperature Environmental
Operating Temperature : 10 0C to 30 0C.
Alarm indicators : Air low, Air high, Skin low, Skin high, Storage Temperature : - 10 0C to 60 0C.
Sensor fail, Heater fail, Air flow, Operating Humidity : 15 to 90% RH, Non condensing.
Power fail, Low level battery, Storage Humidity : 50 to 90% RH, Non condensing.
Air/skin > 39 0C. Altitude : Sea level to 1.9 miles (3 kms).
Test & Approval
Quality : ISO 13485:2016
Electrical safety : IEC 60601-1, IEC 60601-1-2
: IEC 60601-2-20, IEC 60601-1-10