Technical Bulletin 08-011 1600 Upgrade Kits: 1 Initial Information
Technical Bulletin 08-011 1600 Upgrade Kits: 1 Initial Information
1 Initial Information
Product: 1600 Systems
Technical Bulletin Number: 08-011
Priority: High
Category: Serious
Issue Date: October 28, 2008
Units Affected: 1600 stands manufactured before April 2007
Deadline Date: Not Applicable
This is a mandatory safety upgrade. Please ensure all personnel working on IDC
equipment are notified about this important bulletin.
Please complete the attached Fax Back Sheet and return it to Imaging Dynamics by
fax to 1.403.251.1771.
If you have any questions or concerns, please call the IDC Help Desk, toll-free 1-866-
XPLORER (1-866-975-6737) or e-mail service@imagingdynamics.com.
This bulletin can also be downloaded through the IDC tracking site under the project of
IDC-Service Technical Bulletins.
Imaging Dynamics Company • October 28, 2008 • Technical Bulletin: 08-011 v1.0
2 • Revision History
2 Revision History
3 Issue Description
This bulletin is to provide follow-up information to “Technical Bulletin 08-009 Stand Safety
Upgrade” distributed in May 2008 (see attached). IDC identified a potential safety issue on
1600 stands manufactured prior to April 2007. Although the safety risk has been determined
to be minimal, there is a possibility that the 1600 stand may experience inadvertent movement
of the cross-arm assembly.
4 Resolution Description
Upgrade kits are needed to correct this safety issue and prevent any potential safety hazards.
The upgrade kits have been tested and are ready for distribution shortly. IDC will be sending
you all the required kits for your 1600 stands to your main office for distribution. The order
will specify the serial number of your 1600 stand and will reference the appropriate part
number of the upgrade kit. If you have any questions regarding your order, please call the IDC
Technical Support team at 1-866-975-6737.
6 Safety Considerations
Until the upgrade kits have been installed, the systems can be used with caution. Operators
must ensure that they and their patients are not in the path of the moving system. If any
inadvertent movement occurs, the EMERGENCY STOP SWITCH must be used immediately.
In case of failure, this should prevent any personal injury.
NOTE: If the emergency stop switch is pressed and the system is reset, proceed with caution.
If the stand continues to move inadvertently, please contact the IDC Technical
Support team at 1-866-975-6737.
7 Work Instructions
Work instructions for the installation of the upgrade kits are located on the CD that will be
Imaging Dynamics Company • October 28, 2008 • Technical Bulletin: 08-011 v1.0
Work Instructions • 3
provided in the upgrade kits. The instructions will need to be accessed and printed prior to
initiating the upgrade. If you have any questions regarding the work instructions please call
the IDC Technical Support team at 1-866-975-6737.
Upon completion of the upgrade installation, please complete the attached form and
fax to IDC.
Imaging Dynamics Company • October 28, 2008 • Technical Bulletin: 08-011 v1.0
4 • Contact
8 Contact
IDC Corporate Office
Imaging Dynamics Company, Ltd.
151, 2340 Pegasus Way N.E.
Calgary, Alberta, Canada T2E 8M5
Tel: 403.251.9939
Fax: 403.251.1771
IDC Help Desk, toll-free 1-866-975-6737
E-mail service@imagingdynamics.com.
www.imagingdynamics.com
Imaging Dynamics Company • October 28, 2008 • Technical Bulletin: 08-011 v1.0
FAX BACK RECEIPT
Please fax this form back to IDC to acknowledge receipt of Technical Bulletin 08-011 1600
Upgrade Kits and to confirm implementation of this Service Technical Bulletin.
Fax to: Imaging Dynamics
Address: Suite 151, 2340 Pegasus Way NE
Fax #: 1-403-251-1771
Attention: Service Administrator
Date Implemented:
From (Company):
From (Name):
Fax Number: Phone Number:
Comments
TECHNICAL BULLETIN 08-009
1600 STAND SAFETY UPGRADE
1 Initial Information
Product: 1600 systems
Technical Bulletin Number: 08-009
Priority: Critical
Category: Serious
Issue Date: May 22, 2008
Units Affected: 1600 Stands manufactured before April 2007
Deadline Date: N/A
Please ensure all personnel working on IDC equipment are notified about this
important bulletin.
Please complete the attached Fax Back Sheet and return it to Imaging Dynamics by
fax to 1.403.251.1771.
If you have any questions or concerns, please call the IDC Help Desk, toll-free 1-866-975-
6737 or e-mail service@imagingdynamics.com.
This bulletin can also be downloaded through the IDC tracking site under the project of
IDC-Service Technical Bulletins.
Imaging Dynamics Company • May 23, 2008 • Technical Bulletin: 08-009 v2.0
2 • Issue Description
2 Issue Description
This bulletin is to inform you that we have identified a potential safety issue on the 1600
Stands that were manufactured prior to April 2007. Although the safety risk is quite minimal,
there is a possibility that the 1600 Stand may experience inadvertent movement of the cross-
arm assembly.
NOTE: In the event of inadvertent movement, the emergency stop switch will still operate.
3 Resolution Description
To correct this safety issue and prevent any potential safety hazards, 1600 Stands will need to
be upgraded. The upgrade includes a new security device and software that acts as a backup
safety to ensure that inadvertent movements of the 1600 Stand are prevented. Different
upgrade kits will be provided depending on the serial number of the stand:
Upgrade Kit 1 - 1600 Stands with serial number 10014 up to and including 10253
Upgrade Kit 2 - 1600 Stands with serial number 10254 up to and including 10532
Upgrade Kit 3 - 1600 Stands with serial number 10533 up to and including 10664
NOTE: 1600 Stands with a serial number of 10664 and higher are not affected and do not
require an upgrade.
5 Safety Considerations
Until the upgrade kits have been installed, the systems can be used with caution. Operators
must ensure that they and their patient are not in the path of the moving system. If any
inadvertent movement occurs, the EMERGENCY STOP SWITCH must be used immediately.
In case of failure, this should prevent any personal injury.
NOTE: If the emergency stop switch is pressed and the system is reset, proceed with caution.
If the stand continues to move inadvertently, please contact the IDC Technical
Support team at 1-866-975-6737.
6 Work Instructions
Installation instructions will be provided with the upgrade kits.
Imaging Dynamics Company • May 23, 2008 • Technical Bulletin: 08-009 v2.0
Contact • 3
7 Contact
North American, European & Indian Contact
Imaging Dynamics Company
151, 2340 Pegasus Way, NE
Calgary, Alberta
T2E 8M5, Canada
IDC Help Desk, toll-free1-866-975-6737
E-mail service@imagingdynamics.com.
Fax 1-403-251-1771
Asian Contact
IDC Beijing Rep. Office
Room 1007A China Resource
No.8 Jianguomenwai Bei Avenue
Beijing 100005, China
Tel 00 86 10 85192196
Fax 00 86 10 85192198
Imaging Dynamics Company • May 23, 2008 • Technical Bulletin: 08-009 v2.0
FAX BACK RECEIPT
Please fax this form back to IDC to acknowledge receipt of Technical Bulletin 08-009 1600
Stand Safety Upgrade and to confirm implementation of this Service Technical Bulletin.
Fax to: Imaging Dynamics
Address: Suite 151, 2340 Pegasus Way NE
Fax #: 1-403-251-1771
Attention: Service Administrator
Date Implemented:
From (Company):
From (Name):
Fax Number: Phone Number:
Comments