SOMATOM AR.

HP
CT

System

Installation and Setting Instructions


Upgrade AR.HP to AR.HP Spiral (40s/80s)

© Si em en s A G 1 9 95
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.

Register 10
C2 - 017.033.08.03.02 English 02.97
Replaced: C2 - 017.033.08.02.02
0-2 Revision

Chapter Page Rev.


0-4 all 01
0 all 02
1 all 02
2 all 02
4 all 02
0 all 03
2 all 03
4 all 03

SOMATOM AR.HP Register 10 C2 - 017.033.08 TD CT 1 Siemens AG


System Rev. 03 02.97 Medical Engineering
Contents 0-3

Page
1 _______General _______________________________________________________ 1 - 1

Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
Prerequisite . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
For upgrade SOMATOM AR.HP --> AR.SHP . . . . . . . . . . . . . . . . . . . .1-2
For upgrade SOMATOM AR.HP --> AR.SHP 80s . . . . . . . . . . . . . . . . . .1-2
Required tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3

2 _______Upgrade ______________________________________________________ 2 - 1

Preliminary Steps: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1


Gantry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Rotanx . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
D30 Master . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
Scan counter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
D31 ROT Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
D5 Rosy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
D3 DAB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
GEL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
D34 ( Key1, Key2 ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
D32 DAT_COL_RI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
Counterbalance weights . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
Installing the temperature sensor . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
Patient Table. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-9
IMS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 15
Start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 18
Software configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 18
Service/Adjust:. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 19
Service/Tabgen: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 20
Quality test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 20
Concluding steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 21
Attaching labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 21
Product Status Lists ( ES Lists ) . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 23
Test and Handing Over Certificate . . . . . . . . . . . . . . . . . . . . . . . . 2 - 30
Technical documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 30

3 _______Appendix _____________________________________________________ 3 - 1

4 _______Changes to previous version _____________________________________ 4 - 1

Siemens AG TD CT 1 C2 - 017.033.08 Register 10 SOMATOM AR.HP


Medical Engineering Rev. 03 02.97 System
0-4 Contents

Page
This page intentionally left blank.

SOMATOM AR.HP Register 10 C2 - 017.033.08 TD CT 1 Siemens AG


System Rev. 03 02.97 Medical Engineering
General 1 1-1

Safety 1

NOTE: Disconnect the CT system from the main power supply for the
entire upgrade.

When the upgrade is completed, connect all protective


conductors to the system covers.

Conventions 1

SOMATOM AR.HP with SPIRAL HP (40s) = AR.SHP


SOMATOM AR.HP with SPIRAL HP (80s) = AR.SHP 80s

Siemens AG TD CT 1 C2 - 017.033.08 Register 10 SOMATOM AR.HP


Medical Engineering Rev. 02 12.96 System
1-2 General

Prerequisite 1

For upgrade SOMATOM AR.HP --> AR.SHP 1

Upgrade kit SOMATOM AR.HP -> AR.SHP part-no.: 27 91 049 K1085

For upgrade SOMATOM AR.HP --> AR.SHP 80s 1

NOTE: This upgrade is possible only for systems with serial number
13108 to 13... .

Upgrade kit SOMATOM AR.HP -> AR.SHP part-no.: 27 91 049 K1085


ROTANX * part-no.: 62 07 059 X033E

*only necessary for systems with serial number 13108 to 13... and ROTANX with serial
number ≤ 3500 and revision level < 24.

SOMATOM AR.HP Register 10 C2 - 017.033.08 TD CT 1 Siemens AG


System Rev. 02 12.96 Medical Engineering
General 1-3

Required tools 1

• All-purpose tool kit


• Service box
• Lifting device
• Torque wrench 20Nm...100Nm
• 8mm Allen attachment for torque wrench
• 5mm Allen attachment
• 1/4 inch ratchet
• Open-ended wrench M30
• Socket wrench M8, M10
• CCR Head Phantom
• CCR Body Phantom

Siemens AG TD CT 1 C2 - 017.033.08 Register 10 SOMATOM AR.HP


Medical Engineering Rev. 02 12.96 System
1-4 General
This page intentionally left blank.

SOMATOM AR.HP Register 10 C2 - 017.033.08 TD CT 1 Siemens AG


System Rev. 02 12.96 Medical Engineering
Upgrade 2 2-1

Preliminary Steps: 2

NOTE Inform your Opsis site manager when the upgrade will be per-
formed so a system-specific password can be assigned for the
upgrade.
Therefore the system type must be changed and the new Host-ID
must be entered.

• Open the front gantry door.


• Raise the patient table to the highest position.
• Delete all images no longer needed and all raw data files on the store device.
• Repeat the following steps on your system for the remaining images, before upgrading
the host.
1. Export the data set files to the /usr/appl/data/disk1 directory with the
EXPORT function.
2. Delete the data sets you successfully copied into the export directory using
the FILE/DELETE function.
3. Repeat steps 1 and 2 until all data sets needed under SOMARIS/4 are copied to
the file system and the patient directory called up with PATIENT/SELECT shows
only patients with no images.
• Save the system tables under Service/Maintenance using Adjust-Save.
Never use this tables on SOMARIS/4 systems.
• Record the following system-specific and customer-specific data:
- Name of installation ( max. 3 )
- Device installation
- Camera configuration
- View direction
- PACS configuration ( if present )
( All information can be found under Service/Maintenance-Config.)
• Record the status of the hardware scan counter in the form found in the appendix. This
form should be signed by the customer before you begin with the upgrade.
• System/End/No Restart.
• Switch off system.

NOTE All replaced parts must be packaged and returned to:

Mr. Ryll, Siemens AG BMed, Dept. TD ER 12, Eltersdorf, Germany

Whenever possible, the original packaging material for the


upgrade kit should be used.

Siemens AG TD CT 1 C2 - 017.033.08 Register 10 SOMATOM AR.HP


Medical Engineering Rev. 03 02.97 System
2-2 Upgrade

Gantry 2

D STE
AR

M
BO 1

30 R
T 3

A
O D
R

INVERTER

D26 CPU BOARD D5 ROSY


D27 I/O BOARD
D3 DAB

SOMATOM AR.HP Register 10 C2 - 017.033.08 TD CT 1 Siemens AG


System Rev. 03 02.97 Medical Engineering
Upgrade 2-3

NOTE Before replacing the following boards and FW Proms, remove


the corresponding covers. You can attach them again after the
upgrade is completed.

Rotanx 2

Only for upgrade SOMATOM AR.HP --> AR.SHP 80 systems with serial number
13108 to 13....
• Replace the parts according to the instructions found in the ACS folder, Register 5,
“Replacement of parts”.

D30 Master 2

• Replace the old master (D30) with the new Master_SP (D30) .
• Replace the cover plate for the master with a new EMI cover.

Scan counter 2

• Replace the old hardware scan counter with the new scan second counter.
• Connect the cables according to the diagram. The colour of the cables is either black
and white or brown and blue.

black ( brown ) 1

white ( blue ) 2

Scan counter ( back view )

NOTE If the scan counter counts backward reverse the cables.

Siemens AG TD CT 1 C2 - 017.033.08 Register 10 SOMATOM AR.HP


Medical Engineering Rev. 03 02.97 System
2-4 Upgrade
D31 ROT Board 2

• Install the new ROT firmware.


• Mark the new revision level ’11’ on the AS-label with a waterproof pen.

D5 Rosy 2

• Replace the old ROSY (D5) and CDC


(D8) with the new ROSY_SP (D5).
• Remove the four bolts.
• Remove the bolt shown here.

Fig. 2

D3 DAB 2

• Jumper the new DAB 5 (D3) atJ12->5L2 and install it. Do not install the cover and the
cable clamp ( fig.11 ) yet.

SOMATOM AR.HP Register 10 C2 - 017.033.08 TD CT 1 Siemens AG


System Rev. 03 02.97 Medical Engineering
Upgrade 2-5
GEL 2

WARNING Hazardous Conditions:


Hazardous voltage levels may be present at the DC input of the
inverter after the system has been switched off. No voltage must
be present at connection terminals DC+ and DC- when working on
the inverter.

CAUTION The capacitors on the capacitor assembly of the inverter change


after the equipment is deenergized.
Do not touch any uninsulated part of the inverter or the conduc-
tors labeled INV+, INV-, DC+ and DC- at the terminal block on the
outside of the electronic unit before reading the caution labels on
the inverter and at the terminal block X224 on the outside of the
electronic unit.

• First remove the I/O board (D27) and the CPU board (D26).

• Remove the holder shown in fig.4. It will be replaced with one which is attached to the
inverter.
• Replace the old inverter with the new one according to the instructions found in the ACS
folder, Register 5, “Replacement of parts”.
• Install the new CONTROL-BOARD (D27).

D34 ( Key1, Key2 ) 2

• Install the new KEY firmware.


• Place the label with the firmware version beside the part number label on the shielding
plate.
• Mark the new revision level ’01’ on the AS-label with a waterproof pen.

Siemens AG TD CT 1 C2 - 017.033.08 Register 10 SOMATOM AR.HP


Medical Engineering Rev. 03 02.97 System
2-6 Upgrade
D32 DAT_COL_RI 2

• Jumper the new DAT_COL_RI (D32) at X29->1-2 and install it.

Counterbalance weights 2

This step is not necessary if the new inverter, part number 63 03 007, was already
installed before this upgrade.

NOTE It is not necessary to replace the detector.


Use the new self-locking nuts ( M10 ) included in the delivery vol-
ume to re-attach the counterbalance weights.

• Shift one thick lead plate from position 1 to


position 2.
• Install the 3 metal plates, 485g, included
Pos. 1
in the delivery volume at position 1.

Pos. 2

Fig. 5

SOMATOM AR.HP Register 10 C2 - 017.033.08 TD CT 1 Siemens AG


System Rev. 03 02.97 Medical Engineering
Upgrade 2-7
Installing the temperature sensor 2

• Spread the adhesive included with the


shipment on the knurled surface of the
sensor socket only.

Fig. 6

• Insert the sensor socket in the hole in the


detector and tap it in place with a soft
metal hammer.

Fig. 7

• Spread the heat-conducting paste


included with the shipment on the tip of
the sensor.

Fig. 8

Siemens AG TD CT 1 C2 - 017.033.08 Register 10 SOMATOM AR.HP


Medical Engineering Rev. 03 02.97 System
2-8 Upgrade

• Spread Loctite 221 on the threads of the


sensor. Insert the temperature sensor in
the sensor socket and tighten it manually.

Fig. 9

• Installing the sensor cable


- Plug the sensor cable into the temperature sensor.
- Route the sensor cable parallel to the analog cable W229. Ground the shielding at the
same cable clamp as the analog cable.
- Open the cable clamp ( fig.11 ).
- Replace the two spacers with the 3mm spacers included with the shipment.
- Cut the contact springs near the spacers as indicated in fig.10. Place them on the fixed
part of the cable clamp (fig.11 ).
- Close the cable clamp.
- Secure the sensor cable with cable ties.
- Connect the sensor cable at Rosy D5.X155.
• Install the cover plate.

SOMATOM AR.HP Register 10 C2 - 017.033.08 TD CT 1 Siemens AG


System Rev. 03 02.97 Medical Engineering
Upgrade 2-9

Patient Table 2

• Remove all cover plates and the bellows.

• Remove the six M6 screws and


disconnect the protective conductor from
the underside of the table.

Fig. 12

• Remove the right rear screw on the top of


the patient table and then remove the
upper table cover.

Fig. 13

• Remove the horizontal position


potentiometer by using the 1.5mm Allen
key included with the shipment to open
the two clamping screws (do not remove
the cable connections).

Fig. 14

Siemens AG TD CT 1 C2 - 017.033.08 Register 10 SOMATOM AR.HP


Medical Engineering Rev. 03 02.97 System
2 - 10 Upgrade

• Move the tabletop approximately 450mm


into the gantry, so that the belt clamp is
accessible from below through the
opening shown.

Fig. 15
2
• Loosen the belt on the tensioning device
- Loosen two screws (2) at the top and
two (2) at the bottom of the tensioning
carriage.
- Loosen the belt via the tensioning screw
(1).

Fig. 16
2

2 1 tension lock
2 tension pin
3 attachment screws
3 4 threaded pin

Fig. 17

SOMATOM AR.HP Register 10 C2 - 017.033.08 TD CT 1 Siemens AG


System Rev. 03 02.97 Medical Engineering
Upgrade 2 - 11

• Use a hammer to tap the tension pin


upward out of the belt clamp.
• Remove only both exterior screws
( 3/fig.17 ).

Fig. 18

• Move the tabletop completely toward the


gantry to access the belt clamp.
• Mark the position of the belt clamp.
• Open the belt clamp after removing the
center screw from above.
• Remove the toothed belt from the motor
only.

Fig. 19

• Removal of the motor


• Remove the screws from soldering
terminal X8 and then remove the terminal.
• Cut the motor cables X8.1/X8.2 and the
brake cables X8.3/X8.4 near the
soldering terminal.
• Disconnect the protective conductor from
the motor support and the motor.

Fig. 20

Siemens AG TD CT 1 C2 - 017.033.08 Register 10 SOMATOM AR.HP


Medical Engineering Rev. 03 02.97 System
2 - 12 Upgrade
• Disconnect the cable between D35/X2 and board D37/X6 located on the motor (the new
cable will be installed in a later step).
• Disconnect all cable ties at the motor support.
• Remove the motor support (secured with six screws) together with the motor and brake.

• Installation of the motor


X8
• Slide in the installation plate through the
elongated hole until it reaches the end
3
1 stop.
• Hang the new motor support with the
installation plates in the table from below.
• Slide out the installation plate through the
2 elongated hole until it reaches the end
stop and fix the plate with the two screws.
• Fix the motor support with five screws (the
sixth screw over the toothed gear is not
Fig. 21 needed).

• Solder the new connector with cables X8.1/X8.2 to soldering terminal X8.
• Connect the new connector M2.X1-X8 with the one coming from the motor (2/Fig. 21).
• Route the cable (1) as shown in Fig. 21.
• Secure the cable and the connector with cable ties (3/Fig. 21) and attach the label
M2.X1 close to the connector.
• Solder the brake cable to soldering terminal X8.3/X8.4.
• Secure soldering terminal X8 to the new motor support and attach the labels X8 and V1.
• Secure the protective conductors to the motor support and the motor.

• Replace the old D35 with the new


D35_SP.
Note: Attach the edge protector on the
lower EM protection cover of the D35
before installing the new D35_SP.

Fig. 22

SOMATOM AR.HP Register 10 C2 - 017.033.08 TD CT 1 Siemens AG


System Rev. 03 02.97 Medical Engineering
Upgrade 2 - 13

• Route the new cable from D35.X2 to the


table feed encoder B1/X1 parallel to the
existing cable harness.
• Secure the cable with cable ties.

Fig. 23 table feed encoder

• Secure the new toothed belt to the old


toothed belt with insulating tape.
• Pull the new toothed belt through using
the old toothed belt up to the motor and
then secure it manually to the final
position.
• Slide the belt clamp over it and secure it
with the center screw ( 3/fig.17 ).

Fig. 24

• Insert and turn the threaded pin M5*25 ( 4/fig.17 ) approximately 5 mm in one of the two
outer threaded holes of the tabletop to facilitate the installation.
• Place the belt clamp over it.
• Secure the belt clamp ( 3/fig.17 ) to the tabletop with a set screw M5*25. Do not tighten
the screw yet.
• Remove the threaded pin.
• Insert the second set screw M5*25 (do not tighten the screw yet) and push in the tension
pin.
• Tighten both screws M5*25.

Siemens AG TD CT 1 C2 - 017.033.08 Register 10 SOMATOM AR.HP


Medical Engineering Rev. 03 02.97 System
2 - 14 Upgrade

• Place the metal rod ( part of delivery ) in


the belt tensioning device ( pos.1 ).
• Tighten the belt to 180N-200N, using a
spring scale ( pos.2 ) to measure the
3 tension. Tighten the screw ( pos.3 ) with
the belt in this position.
• Tighten the tensioning screw ( 1/fig.16 )
until it touches the tightening block.
2 • Loosen screw 3.
1
• Check the belt tension again.

Fig. 25

• If the tensioning screw lifts away from the tightening block, set it so that it touches the
tightening block at the given belt tension.
• Lock the backing nut of the tensioning screw at the tensioning carriage.
• Tighten the three remaining screws.
• As a result of this adjustment the following values should be measured:
- Force to move tabletop (measured in STOP condition ): 50N - 140N
• Attach all table covers removed prior to the upgrade except for the right and the left side
covers, to allow later installation of labels and potentiometers.

Presetting the potentiometer R2


• Retract the tabletop completely.
• Turn the shaft of the potentiometer toward the "rear" end stop and then back again by
145° ±30° .

NOTE Ensure the end stop is correct, otherwise the potentiometer will
be turned too far.

SOMATOM AR.HP Register 10 C2 - 017.033.08 TD CT 1 Siemens AG


System Rev. 03 02.97 Medical Engineering
Upgrade 2 - 15

IMS 2

NOTE Observe EGB/ESD guidelines.

• Remove the cover and the upper front cover.


• Disconnect the plug connectors of the following cables coming from outside the IMS:

Cable From To connector


W1 Gantry D40.X3
W60 Monitor X2 ( AC power unit )
W65 Keyboard X16
W66 Monitor X14
W67 Camera ( analog ) X15
W68 Camera ( analog ) X13
W70 Camera ( digital ) X13
W71 Camera ( digital ) D66.X12
W73 Modem E52/RS232-A
W74 Network E52 Ethernet

• Disconnect and remove the following cables:

Cable From connector To connector


E52 RS232-B and
W80 D40.X1
E52 KEYBOARD/MOUSE
E52 X6 and
W81 D40.X4
E52 Audio input/output
W82 X16 D40.X2
W83 X17 ( do not remove ) D69.X3 ( disconnect only at X3 )
W84 X14 D62.X8
D64 BAS or
W87 X15
D66 X8
D64.X5 or
W88 X13
D66.X5
W90 E52 SCSI all devices

Siemens AG TD CT 1 C2 - 017.033.08 Register 10 SOMATOM AR.HP


Medical Engineering Rev. 03 02.97 System
2 - 16 Upgrade
• Remove the Option Pals from Palbank D121(on Siam board D25) except the
DYMAMIC Pal.
• Disconnect the E4A+/E4A- cables from the 5V power supply E4A.
• Remove the 4 attachment nuts for the slide-in module rack and the ground cable.
• Lift out the SMI slide-in module rack.
• Connect the 5V cables from the old SMI slide-in module rack, D115 5V/0V, to the new
SMI slide-in module rack.
• Place the new SMI slide-in module rack in the IMS container.
• Secure the new SMI slide-in module rack and the ground cable.
• Secure the E4A+/E4A- cables at the E4A power supply.
• Install the Option Pals removed earlier on D202 (CCI). The following locations are
available: I11 - I18, I31 - I37, I51,I53,I55,I57, however, I38 should remain reserved.
• Replace ’ Pal 40s’ with ’Pal 80s’ in the case of an AR.SHP 80s.
• Replace the old system disk SD3 ( E1 ) with the new 1GB disk.
• Remove bus terminators RP1, RP2, and RP3 from the streamer tape.
• Feed the Y cable W173 in the voltage supply cable for the tape drive ( E2 ) and connect
the open end of the cable to the supply cable for the backplane D117.
• Replace the old connection field
- Remove connectors X16, X20 (optional) from the old connection field and install them
on the new connection field.
- Install the new connection field.
• Install the following cables in the container:

Cable From connector To connector


D205/X20
W171(SCSI) X11
to all disk drives
W104(Ethernet) X12 D205/X21
W112(Camera) X13 D66/X5
W130/W131 X15 D66.X8
3xW130/W131 X14(R/G/B) D70/X7,X8,X9
Keyboard X16 D40/X2
W170(Modem) X21 D205/X22
W131 (D40-E52) D40.X1 D205/X24
D202/X178
W172 D40/X3
Gantry cable (W1 )
D205/X25
W132 D40.X4
D202/X20

SOMATOM AR.HP Register 10 C2 - 017.033.08 TD CT 1 Siemens AG


System Rev. 03 02.97 Medical Engineering
Upgrade 2 - 17

Cable From To connector


W1 Gantry W172
W65 Keyboard X16
W67 Camera ( analog ) X15
W68 Camera ( analog ) X13
W70 Camera ( digital ) X13
W71 Camera ( digital ) D66.X12
W73 Modem ( optional ) X21
W74 Network X12

• Connect cable W83 (SMDATA) to D69/X3.


• Replace the monochrome monitor with the color monitor.
- Remove the monochrome monitor including the mains cable.
- Route the mains cable as well as the two new 6 m coaxial cables with the old coaxial
cable from the IMS container (R/G/B, AC_Power_Unit.X2 ) to the new monitor.
• Install the new firmware and the three new keys in the keyboard and rearrange the
buttons to get the condition shown in fig.27.

Firmware

• Connect the potentiometer module to X3 of the keyboard.


• Attach the labels indicating the new function key assignments.
• Connect the desktop CDROM
- Connect the bus terminator to the desktop CDROM.
- Connect the data cable W75 to X11 of the connector field.
- Connect the mains cable to X5 (AC_Power_Unit/additional container).

Do not attach the IMS covers yet.

Siemens AG TD CT 1 C2 - 017.033.08 Register 10 SOMATOM AR.HP


Medical Engineering Rev. 03 02.97 System
2 - 18 Upgrade

Start-up 2

NOTE Use the documents for the SOMATOM AR.SP for all procedures.

• Test the protective conductor connections to all metal covers on the system.
- The test is performed using a protective conductor meter
( P/N 44 15 899 RV 090 ).
- The measurement results will be incorrect if the measurement cables are not unrolled
from the roll.
- The protective conductor resistance must be ≤100mΩ ( measured against the
PDS protective conductor connection ).
• Switch on the system.
System starts up automatically with System/Config.

NOTE Do not switch off the unit until Configuration of software under
System/Config. was done.

• Adjust the 5V/200A power supply in the IMS Container.


- Remove the right cover of the IMS Container
- Set the voltage to 5.2V -0.5% measured at the terminal studs.
• Attach the covers of the IMS container, then place the desktop CDROM on the container.
• Set the horizontal potentiometer R2 on the patient table:
- Move the tabletop to the “0” or end position manually.
- Set the voltage between 0.12V and 0.74V by turning the potentiometer shaft.
- Install the potentiometer at this setting and secure.
- Test the measurement results again and then move the tabletop toward the gantry
until it reaches the end stop.
- Measure the value again. At this position, it should be less than 9.9V.
• Attach the right table cover.

Software configuration 2

NOTE The software should be configured according to the release for


SOMARIS/4. The following points provide an overview.

• Configure software under System/Config:


1) General
2) Serial No
3) Host Cam
4) Device
5) Network
6) SMI
Reboot
• Definition of service key (with host-id and system serial number).

SOMATOM AR.HP Register 10 C2 - 017.033.08 TD CT 1 Siemens AG


System Rev. 03 02.97 Medical Engineering
Upgrade 2 - 19
• Restore old patient images using the following procedure:
(DO NOT INITIALIZE THE DATA DISK. )
After installing the new system disk and after having finished all installation steps you
can use FILE/COPY to copy all your data sets from source Unix into the STORE device
again.
• Install the reference images.
• Install all options previously existing on the system.
• Update the Tube History data under Service/Tube tool kit.

Service/Adjust: 2

1) PHS Horiz
2) PHS Vert
3) Getter

If the ROTANX was not replaced:

4) PSD Off
5) Focus Ali ( at this point the Plexiglas ring should be installed)
6) Voltage ( select checkbox ’NEUTRAL TABLES’ )
7) LightM

If the Rotanx was replaced to upgrade the system to SOMATOM AR.SHP 80s.

4) Z-Adjust
5) PSD Off
6) Det Align
7) Det Tilt (If the shift for DET ALIGN is <0.7mm, no DET TILT is necessary. If the shift
for DET ALIGN is >0.7mm, it is necessary to perform DET TILT and another DET ALIGN.)
8) Top Bottom
9) Focus Ali ( at this point the Plexiglas ring should be installed )
10)Voltage ( select checkbox ’NEUTRAL TABLES’ )
11)LightM

Siemens AG TD CT 1 C2 - 017.033.08 Register 10 SOMATOM AR.HP


Medical Engineering Rev. 03 02.97 System
2 - 20 Upgrade
Service/Tabgen: 2

1) Base Cal
2) Diff Cal
3) CCR
4) Beam Cent
5) Beam Hard
6) Voltage ( deselect checkbox ’NEUTRAL TABLES’)
7) Scaling

Quality test 2

• Perform the constancy test (Acceptance test in Germany/BRD ):


- Scan Setup/Quality/Constancy
• Store the system tables:
- System/Run/Save_Tables

SOMATOM AR.HP Register 10 C2 - 017.033.08 TD CT 1 Siemens AG


System Rev. 03 02.97 Medical Engineering
Upgrade 2 - 21

Concluding steps 2

Attaching labels 2

Gantry

• Attach the label with new part number


SOMATOM AR.
HP
Spiral
27 91 049 next to the existing label.
• Attach the label with the new product
name

SOMATOM AR.HP
Spiral

Fig. 28

PHS

• Attach the labels with "AS-label" and "ES-


Gantry label" below existing labels at position 2
2
behind the left table cover.
• Attach the left table cover.
• Attach the label with new part number
88 81 815 at position 1 next to the existing
label.

Fig. 29

Siemens AG TD CT 1 C2 - 017.033.08 Register 10 SOMATOM AR.HP


Medical Engineering Rev. 03 02.97 System
2 - 22 Upgrade
IMS

• Attach the label with part number


16 51 509 next to the existing label.
• Attach the label for identification in the
case of an AR.SHP 80s.

Fig. 30

Keyboard
• Attach the labels Kit Keyboard SHP and part number 37 94 513 K1072 next to the
existing labels on the back.

SOMATOM AR.HP Register 10 C2 - 017.033.08 TD CT 1 Siemens AG


System Rev. 03 02.97 Medical Engineering
Upgrade 2 - 23
Product Status Lists ( ES Lists ) 2

The upgrade kit includes a new ES (product status) list. All important parts and assem-
blies in the upgrade kit appear in this list.
After the upgrade is completed, the new ES list should be filed with the other system
papers.
In the existing ES lists for the system, place the label ”refer to ES list 27 91 049 K1085
Pos. ...” next to each position that will be replaced by a part with a new item number or
version status ( for firmware or software). The corresponding labels are included with the
upgrade kit. Prior to changing the existing ES lists, make copies of them and file these
copies with the other system papers.
All updating of the design status of these positions should be done only on the ES list.
Those positions not affected by the upgrade will be updated in the original ES list.

The following pages show examples of how ES lists should be modified. Modifications
( labels ) on the original ES lists for the system appear in brackets.

NOTE Do not compare your ES Lists with the following examples.

Siemens AG TD CT 1 C2 - 017.033.08 Register 10 SOMATOM AR.HP


Medical Engineering Rev. 03 02.97 System
2 - 24 Upgrade
Examples

SOMATOM AR.HP Register 10 C2 - 017.033.08 TD CT 1 Siemens AG


System Rev. 03 02.97 Medical Engineering
Upgrade 2 - 25

Siemens AG TD CT 1 C2 - 017.033.08 Register 10 SOMATOM AR.HP


Medical Engineering Rev. 03 02.97 System
2 - 26 Upgrade

SOMATOM AR.HP Register 10 C2 - 017.033.08 TD CT 1 Siemens AG


System Rev. 03 02.97 Medical Engineering
Upgrade 2 - 27

Siemens AG TD CT 1 C2 - 017.033.08 Register 10 SOMATOM AR.HP


Medical Engineering Rev. 03 02.97 System
2 - 28 Upgrade

SOMATOM AR.HP Register 10 C2 - 017.033.08 TD CT 1 Siemens AG


System Rev. 03 02.97 Medical Engineering
Upgrade 2 - 29

Siemens AG TD CT 1 C2 - 017.033.08 Register 10 SOMATOM AR.HP


Medical Engineering Rev. 03 02.97 System
2 - 30 Upgrade
Test and Handing Over Certificate 2

• Please fill out the ’Test and Handing Over Certificate’ and send it to Erlangen, TD CT 5.

Technical documentation 2

A new set of technical documentation for SOMATOM AR.SP will be delivered with the
upgrade kit. Please replace the old set for the SOMATOM AR.HP with this new set of
technical documentation.
Before doing so, remove the ’Generator wiring Diagram X039E’ from register 6, System
Manual SOMATOM AR.HP and place it in the new ACS Manual SOMATOM AR.SP, regis-
ter 6.
Remove the old set of technical documentation for SOMATOM AR.HP from customer site.

TD CT 1 / Lapunik
TD CT 2 / Gursch
SMS Iselin / Weiss Costa

SOMATOM AR.HP Register 10 C2 - 017.033.08 TD CT 1 Siemens AG


System Rev. 03 02.97 Medical Engineering
Appendix 3 3-1

Confirmation
Customer Name: .....................................................
Address: .....................................................
.....................................................
.....................................................
.....................................................

CT system SOMATOM AR
Serial no. ..................................
Date ..................................

The CT system indicated above was upgraded with the SPIRAL HP function.

During the upgrade, the CT scan counter was converted to scan seconds, beginning
with 0 seconds.
The previous scan counter status ( old ) was calculated at 2 seconds per scan. After the
upgrade, scan seconds will be used.

As a result, the scan counter status for the CT system indicated above is converted to
scan seconds as follows:

Scan counter status ( old ) ......................... Scans x 2 sec. =


......................... Scan seconds ( old )

The next time the CT X-ray tube is replaced, the scan counter status will be calculated as
follows:
Scan seconds ( old ) + Scan seconds ( new ) =
Scan seconds ( total )

Location / date: ........................................ / ..........................................

Signatures:
.......................................... ..........................................
( Customer ) ( Siemens Service
Representative)

Siemens AG TD CT 1 C2 - 017.033.08 Register 10 SOMATOM AR.HP


Medical Engineering Rev. 01 08.95 System
3-2 Appendix
This page intentionally left blank.

SOMATOM AR.HP Register 10 C2 - 017.033.08 TD CT 1 Siemens AG


System Rev. 01 08.95 Medical Engineering
Changes to previous version 4 4-1
• Chapter 2:
Fig.21 was changed.
’Route the cable (1) as shown in Fig.21’ was added.
- Pages affected by the change: 12

Fig.28 was changed.


- Pages affected by the change: 21

Siemens AG TD CT 1 C2 - 017.033.08 Register 10 SOMATOM AR.HP


Medical Engineering Rev. 03 02.97 System
4-2 Changes to previous version
This page intentionally left blank.

SOMATOM AR.HP Register 10 C2 - 017.033.08 TD CT 1 Siemens AG


System Rev. 03 02.97 Medical Engineering