Professional Documents
Culture Documents
StressVue 3.00.04 Wireless Patient Module, Rev.B - English
StressVue 3.00.04 Wireless Patient Module, Rev.B - English
Safety Summary
Wireless Patient Module Equipment Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . v
Safety Symbols Marked on the Wireless Patient Module Packaging . . . . . . . . . . . . . . . . . . . . vi
Disposal Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vi
Conventions Used in the Instructions for Use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
Important Patient and Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
Warnings about the Wireless Patient Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
Additional Wireless Patient Module Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ix
Caution Statements for the Wireless Patient Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . x
Important Notes about the Wireless Patient Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . xi
FCC Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xii
Industry Canada Compliance Statement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xiii
The Wireless Patient Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xiii
Indications for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xiii
Appendix D. Translations
Symbol Description
Attention. See the StressVue Instructions for Use and other product
documentation for information.
CE mark.
v
Safety Summary
Symbol Description
Keep dry.
Fragile.
Disposal Information
The wireless modules use one alkaline battery and disposable monitoring electrodes. Disposal
must be in accordance with the following procedures:
Battery — applicable disposal or recycling standards
Electrodes — normal waste
WARNING Warning statements describe conditions or actions that may result in personal injury or
loss of life.
CAUTION Caution statements describe conditions or actions that may result in damage to equipment or
software.
The wireless patient module device transmits data reflecting a patient’s physiological
condition to a properly equipped system and when reviewed by a trained physician or
clinician can be useful in determining a diagnosis. However, the data should not be used
as a sole means for determining a patient’s diagnosis.
To maintain designed operator and patient safety, peripheral equipment and accessories
used that can come in direct patient contact must be in compliance with UL 2601-1,
IEC 60601-1, and IEC 60601-2-25.
FCC Warning (Part 15.21): Changes or modifications not expressly approved by the party
responsible for compliance could void the user’s authority to operate the equipment.
(United States only)
To maintain designed operator and patient safety, only use parts and accessories supplied
with the device and available through Philips Medical Systems.
To avoid the possibility of serious injury or death during patient defibrillation, do not
come into contact with device or patient cables. Additionally, proper placement of
defibrillator paddles in relation to the electrodes is required to minimize harm to the
patient.
Ensure that the electrodes or lead wires do not come in contact with any other conductive
materials (including earth-grounded materials), especially when connecting or
disconnecting electrodes to or from a patient.
WARNING Failure to follow these warnings could affect both patient and operator safety.
A possible explosion hazard exists. Do not use the device in the presence of flammable
anesthetics, or flammable mixtures with air, oxygen, or nitrous oxide.
In some countries, the use of the wireless patient module (2500 MHz version) has been
limited as follows:
Connecting multiple medical electrical instruments to the same patient may pose a safety
hazard due to the summation of leakage currents. Any combination of instruments should
be evaluated by local safety personnel before being put into service.
The channel setting of the wireless patient module, when used in the above listed
countries, must be set accordingly.
In the United States and Canada, use of the wireless patient module (608 MHz version) is
regulated by Federal Communications Commission (FCC) rules, Part 15.
– The wireless patient module must be used solely on the premises of health care
facilities (see Part 15, section 15.242a).
– A health care facility operating the wireless patient module must coordinate with the
directors of existing nearby TV stations and Radio Astronomy Observatories to ensure
compatible use. Minimum separation distances from such facilities may apply. It may
be necessary to obtain written authorization from such facilities prior to installation
and use of the transmitter (Part 15, section 15.242d,e).
– Under U.S. Federal Communications Commission (FCC) rules, authorized health care
providers must register their WMTS devices with an authorized Frequency
Coordinator designated by the FCC. The American Society for Healthcare
Engineering (ASHE) is the current designated Frequency Coordinator.
To ensure safety and prevent damage to the system, only connect the patient data cable to
the connector on the patient module.Do not connect the patient data cable into any other
connector or port on the StressVue computer.
– Pour garantir la sécurité et éviter d'endommager le système, brancher le câble de
données patient uniquement sur le connecteur du module patient. Ne pas le brancher
sur aucun autre connecteur ni sur le port de l'ordinateur de l'application StressVue.
– Per garantire una protezione adeguata ed evitare di danneggiare il sistema, collegare
il cavo dati paziente esclusivamente al connettore sul modulo paziente. Non collegare
il cavo dati paziente ad altri connettori o porte del computer di StressVue.
– Aus Sicherheitsgründen und zur Vermeidung von Schäden am System darf das
Patientendatenkabel nur in den Anschluss des Patientenmoduls eingesteckt werden.
Stecken Sie das Patientendatenkabel nicht in einen anderen Anschluss oder Eingang
des StressVue-Computers.
– Para garantizar la seguridad y evitar cualquier daño en el sistema, el cable del paciente
sólo puede conectarse al conector del módulo de paciente. No conecte el cable de
datos de paciente en ningún otro conector o puerto del StressVue.
To prevent possible damage to the keypad, do not use sharp or hard objects to depress
keys; only use fingertips.
Do not attempt to clean the device or patient cables by submersing into a liquid,
autoclaving, or steam cleaning.
Wipe the exterior surface of the device and patient cables with a non-alcohol sterilizing
disinfectant, then dry with a clean cloth.
Conductive parts of the patient cable, electrodes and associated Type CF connections,
including the neutral conductor of the patient cable and electrode, should not come into
contact with other conductive parts, including earth ground.
The wireless patient module and patient cable should be cleaned between each use.
Do not pull or stretch patient cables as this could result in mechanical and/or electrical
failures. Patient cables should be stored after forming them into a loose loop.
The wireless patient module only works with systems that are equipped with the
appropriate option.
No user-serviceable parts are inside. Any modification of this device may alter
defibrillator protection. Any modification to any part of this device is to be performed by
qualified service personnel only.
The following equipment may cause interference with the RF channel: microwave ovens,
diathermy units with LANs (spread spectrum), amateur radios, and government radar.
To prevent possible damage to the device during transport and storage (while in the
original packaging), the following environmental conditions must be adhered to:
Ambient Temperature Range:
-20°C to 65°C (-4°F to 149°F)
Relative Humidity Range:
5% to 95% (non-condensing)
Atmosphere Pressure:
700 hPa to 1060 hPa
Allow the device to stabilize within its intended operating environment for a minimum of
two hours prior to use. The allowable operating environment is as follows:
Ambient Temperature Range:
0°C to 45°C (32°F to 113°F)
Relative Humidity Range:
5% to 95% (non-condensing)
Atmosphere Pressure:
700 hPa to 1060 hPa
MEDICAL EQUIPMENT
WITH RESPECT TO ELECTRIC SHOCK, FIRE, AND
MECHANICAL HAZARDS ONLY, IN ACCORDANCE WITH
UL 2601-1, CAN/CSA C22.2 No. 601.1, IEC60601-1,
AND IEC60601-2-25.
5P35
To comply with WMTS, only use channels 0-63 with the wireless 608 module.
FCC Compliance
In the United States, use of the wireless patient module is regulated by Federal
Communications Commission (FCC).
FCC ID: HJR-X12-600-15
FCC ID: HJR-X12P-2500
These devices comply with Part 15 of the FCC rules.
Operation is subject to the following conditions:
This device may not cause harmful interference.
This device must accept any interference received, including interference that may cause
undesired operation.
The wireless 608 patient module must be used solely on the premises of health care
facilities (see Part 15, section 15.242a).
A health care facility operating the wireless 608 patient module must coordinate with the
directors of existing nearby TV stations and Radio Astronomy Observatories to ensure
compatible use. Minimum separation distances from such facilities may apply. It may be
necessary to obtain written authorization from such facilities prior to installation and use
of the wireless patient module (Part 15, section 15.242d,e).
Under U.S. Federal Communications Commission (FCC) rules, authorized health care
providers must register their WMTS devices with an authorized Frequency Coordinator
designated by the FCC. The American Society for Healthcare Engineering (ASHE) is the
current designated Frequency Coordinator.
1-1
Getting Started Using the Wireless Patient Module
LCD display
Figure 1-2 Wireless patient modules, rear view (without battery cover)
The transmitter will generally maintain a connection with the USB interface connector as long
as the devices are within approximately 45 feet (15 meters) of each other. This number is an
approximate figure; the distance may vary depending upon the local area network at your site,
any wireless networks, telemetry units in the area, and other associated RF usage.
Supported Channels
The Wireless 608 patient module, for use in North America, Australia, and China supports the
64 channels in the Wireless Medical Telemetry System (WMTS), which includes the 608-614
frequency band (channels 00-3F on the patient module). WMTS is the only frequency
spectrum designated exclusively for wireless medical telemetry systems.
The WMTS has been established by the FCC in response to concerns about interference
resulting from new digital television transmitters, low power television transmitters, and
greater use of Private Land Mobile Radio equipment. In addition, the FCC also appointed
ASHE (American Society for Healthcare Engineering) as the Frequency Coordinator for the
WMTS bands.
Hospitals deploying telemetry systems in the WMTS bands must register the device before the
equipment is operational. Registration is required to ensure interference-free operation. For
registration information, see Appendix C, “ASHE Registration.” This information is also
available in the StressVue Installation and Getting Started Guide.
The Wireless 2500 patient module, for use in regions other than the two listed above, can
operate in any of the 256 available channels.
The USB interface connector provides a convenient and simple interface to transmit data from
the patient module to the StressVue PC. This interface is used both for direct connect and
wireless transmitters. Figure 1-3 on page 1-5 illustrates the wireless option. For ordering
information, see the StressVue Instructions for Use.
The transmitter will generally maintain a connection with the USB interface connector as long
as the devices are within approximately 45 feet (15 meters) of each other. This number is an
approximate figure; the distance may vary depending upon the local area network at your site,
any wireless networks, telemetry units in the area, and other associated RF usage.
The interface connector does not require a battery or separate power cord; it draws power from
the StressVue PC, via the USB cable. The analog output, together with TTL, is standard on all
StressVue systems.
For additional information about the TTL and analog outputs, see Appendix C, “TTL and
Analog Export for Echo” in the StressVue Instructions for Use.
Unused
Rear view
TTL Analog
output output
USB connector
Antenna ECG
connectors
The wireless interface connector has two BNC-type antenna connectors for direct connection
of antennas. These antenna are used to provide diversity and increase signal throughput.
For details on attaching the antennas and connecting the interface connector to the PC, see the
StressVue Installation and Getting Started Guide. These steps are performed once, when
setting up the system.
CAUTION Be sure the connector on the patient cable is properly aligned with the connector on the patient
module, and that you push the connector straight into or pull it straight out of the patient module.
Inserting the connector at an angle may bend and damage the connector pins on the patient
module.
LA/L
RA/R
RL/N V4/C4
V5/C5
V1/C1 V6/C6
V2/C2
V3/C3
LL/F
Available Documentation
The following documentation is available with the wireless patient module and StressVue
system:
Wireless Patient Module Instructions for Use
This guide. Provides information about how to set up and maintain the wireless patient
module.
StressVue Instructions for Use
A reference that provides detailed information about StressVue functionality, including
how to configure the system, as well as run the StressVue application. This guide also
includes troubleshooting and maintenance information.
This manual is written for clinical professionals. They are expected to have a working
knowledge of medical procedures and medical terminology as required for monitoring
potential cardiac patients.
StressVue Installation and Getting Started Guide
Describes how to set up the StressVue hardware and software, and provides an overview
of StressVue components and operation.
StressVue System Cart Setup Guide
Describes how to set up the StressVue hardware and cart.
StressVue tutorial
HTML-based training program teaches you how to configure and use StressVue.
Menu item Menu items and button names appear in a bold no-serif font.
Button name
Example: Click Settings.
Field names and list Field names and list items appear in a no-serif font.
items Example: Select the appropriate format from the Format dropdown
list.
WARNING Warning statements describe conditions or actions that may result in injury to the
patient.
CAUTION Caution statements describe conditions or actions that may result in damage to equipment or
software.
1 Getting Started Using the Wireless Patient Module. This chapter. Provides
an overview of the StressVue system.
2 Configuring and Using the Wireless Patient Module. Describes all aspects of
the patient session, from getting started through the four phases of the test.
CAUTION Any changes or modifications not expressly approved by Philips Medical Systems can void your
authority to operate the equipment.
The Philips Wireless 608 and 2500 patient modules comply with part 15 of the FCC rules.
Operation is subject to the following conditions:
The device may not cause harmful interference.
The device must accept any interference received, including interference that may cause
undesired operation.
The device must be used solely on the premises of health care facilities (see Part 15,
section 15.242a).
A health care facility operating the wireless patient module must coordinate with the
directors of existing nearby TV stations and Radio Astronomy Observatories to ensure
compatible use. Minimum separation distances from such facilities may apply. It may be
necessary to obtain written authorization from such facilities prior to installation and use
of the wireless patient module (Part 15, section 15.242d,e).
Under U.S. Federal Communications Commission (FCC) rules, authorized health care
providers must register their WMTS devices with an authorized Frequency Coordinator
designated by the FCC. The American Society for Healthcare Engineering (ASHE) is the
current designated Frequency Coordinator.
Operating frequencies are 608.48-631.52 MHz or 2400.96-2452.96 MHz operating frequency
range.
This device is defibrillator-protected in compliance with AAMI standards and IEC 601-2-25.
2-1
Configuring and Using the Wireless Patient Module
LCD display
5 Replace the battery cover and slide it back until it snaps into place.
The wireless patient module turns itself on as soon as you insert a fresh battery.
Full battery
Battery at 75%
Battery at 50%
Battery at 25%
Low battery
If a battery with unknown voltage is inserted and the main menu does not appear, replace the
battery.
NOTES If the battery voltage is less than 1V, the patient module will not power on. Replace with a fresh
battery to continue operation.
When the battery voltage indicator shows 25%, we recommend you discard the battery and replace
it with a new one.
Keypad Buttons
The key pad buttons are used as follows. Note that, during transmission, you can press any key
to send a Call signal.
Up/Right arrow Used to display the menu and navigate through the
options.
In general, use the Up/Right arrow to display the menu, use the Down arrow to navigate through
the options, and the Enter key to select the option and save your settings.
Failing leads
Press Enter (Left arrow) to
Press the Up/Right
power off the patient
arrow to display
module
menu options
Menu Options
The patient module menu provides the following options:
Lead Check
Used to verify signal impedance. See “Checking Signal Impedance” on page 2-18.
Display ECG
Lets you scroll through each lead and examine the waveform. See “Inspecting Signal from
Individual Leads” on page 2-19.
Configure
Provides options for configuring the channel and cables used, as well as verifying patient
module information and selecting the display language. See “Configuring the Patient
Module for Use” on page 2-8.
Done
Press Enter to exit the menu and return to the main screen.
Figure 2-3 on page 2-8 shows a flow chart illustrating the menu options.
DONE
LEAD CHECK DISPLAY ECG CONFIGURE
Return to Main
menu
Select a channel Select 10-wire, Select language Software version Show real-time Return to
(page 2-9) 5-wire, 4-wire (page 2-11) information voltage Configure menu
(page 2-10) (page 2-12) (page 2-12)
2 Press the Down arrow to move through menu selections, until the Configure menu item is
displayed.
5 When the desired item is displayed, press the Enter key to select it.
7 Ensure that you configure the same channel on the StressVue system.
For details, see the section “Configuring Wireless Mode” in Chapter 5, “Configuring
StressVue,” of the StressVue Instructions for Use.
You can choose from three options. The first one to appear is 10-wire.
3 Press the Down arrow key to scroll through the three options.
4 When the desired option is displayed, press Enter to select it and return to the Configure
menu.
CAUTION Be sure the connector on the patient cable is properly aligned with the connector on the patient
module, and that you push the connector straight into or pull it straight out of the patient module.
Inserting the connector at an angle may bend and damage the connector pins on the patient
module.
Correct alignment
LA/L
RA/R
RL/N V4/C4
V5/C5
V1/C1 V6/C6
V2/C2
V3/C3
LL/F
3 Place the gel area of the electrode over the center of the prepared area, using the
positioning described in Table 2-2, below, and illustrated on the previous page; then press
the adhesive ring into place.
– Avoid pressing the center of the gel area; this might hinder conduction.
– Place the electrodes on flat, fleshy parts of the arms and legs, avoiding muscle. See
limb lead placement notes below.
– For female patients, place the precordial electrodes on top of the breast (not
underneath the breast tissue).
Lead placement is similar to standard 12-Lead ECG placement; however, limb leads are
modified, as follows:
– Right Arm and Left Arm leads should be placed close to the shoulders on the clavicle
bone, away from muscular areas to avoid muscle interference.
– The Right Leg lead is typically placed on the sternum midway between the arm leads
and V1 and V2 leads.
– The Left Leg lead should be placed on a rib, below V6 by about 2 fingers, in the lower
left area of the patient’s chest (avoiding flabby areas and the belt). This may need
adjustment depending on body habitus.
4 Ensure the electrodes are firmly attached.
A good test for firm electrode contact is to try to move it. If it moves easily, the electrode
connection is too loose. Do not allow the electrodes to move in any way.
5 Have the patient raise their arms over their head. This will help verify good lead
placement and no strain on the electrodes.
Review the following lead wire labeling and electrode placement information to
ensure a quality ECG.
Table 2-2 Leads and positioning
Between V4 and V6
RA/R
LA/L
RL/N
V1/C1 V4/C4
V5/C5
V2/C2 V6/C6
V3/C3
LL/F
To check impedances
1 With the main screen displayed, hold down the Up/Right arrow key for a few seconds to
access the menu.
The Lead Check menu item appears.
A graph depicting the impedance measured at the Right Arm (RA), Left Arm (LA), Left
Leg (LL), and V1 through V6 electrodes is displayed from left to right in vertical columns
on the screen. The higher the bar, the better the contact is between the skin and the
electrode.
A full-bar graph (6 bars) means optimal high quality and good electrode contact. For good
quality transmissions, the bars should be at least 4 bars high.
A low-bar graph means poor quality and high electrode impedance. Re-prep the skin, and,
if necessary, replace the electrode(s).
3 After verifying acceptable impedance levels, press Enter to return to the top-level menu.
5 After visually verifying each lead, press Enter to return to the top-level menu.
NOTE During a patient session, if a lead fails, the affected lead appears on the display. See Appendix B,
“Patient Module Messages and Channel Assignments” for details on lead failure messages.
Precautions
Before cleaning any equipment, remove the battery from the patient module.
Do not attempt to clean the device or patient cables by submersing into a liquid,
autoclaving, or steam cleaning.
Wipe the exterior surface of the device and patient cables with a non-alcohol sterilizing
disinfectant, and then dry with a clean cloth.
Conductive parts of the patient cable, electrodes, and associated Type CF connections,
including the neutral conductor of the patient cable and electrode, should not come into
contact with other conductive parts, including earth ground.
Do not pull or stretch patient cables as this could result in mechanical and/or electrical
failures. Patient cables should be stored after use by forming them into a loose loop.
Proper patient preparation is important for proper application of ECG electrodes and
operation of the device.
Patient cable exteriors should be checked for cracks or breakage prior to use.
3-1
Maintaining the Wireless Patient Module
As defined by IEC 60601-1 and IEC 60601-2-25, the device is classified as follows:
– Class IIa equipment
– Type CF applied parts
– Ordinary equipment
– Not suitable for use in the presence of flammable anesthetics
– Continuous operation
CAUTION Dispose of waste material according to local regulations and laws regarding medical and battery waste.
CAUTION Do not:
Use isopropyl alcohol
Autoclave the patient module, patient data cable/lead wires, or use ultrasonic cleaners
Immerse
Use abrasive materials
Wet the connectors
A-1
Specifications and Requirements
Feature Specifications
NOTE Operating range performance is 15 meters with no barriers. The manufacturer does not supply
accessories for direct cardiac applications.
RF Emissions CISPR 11 Group 2 The wireless patient module must emit electromagnetic
energy in order to perform its intended function. Nearby
electronic equipment may be affected.
RF Emissions CISPR 11 Class B The wireless patient module is suitable for use in all
establishments other than domestic and those directly
Harmonic Emissions not applicable connected to the public low-voltage power supply
IEC 61000-3-2 network that supplies buildings used for domestic
purposes.
Voltage fluctuations/ not applicable
flicker emissions
IEC 61000-3-3
⎡ 3.5 ⎤
d=⎢ P
⎣ 3Vrms ⎥⎦
⎡ 7 ⎤
d=⎢ ⎥ P 800 MHz to 2.5 GHz
⎣ 3V / m ⎦
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, am-
ateur radios, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the elec-
tromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the wireless patient module is used exceeds the applicable RF compliance level above, the
wireless patient module should be observed to verify normal operation. If abnormal performance is observed, additional mea-
sures may be necessary, such as reorienting or relocating the wireless patient module.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [3] V/m.
d = 1.2 P d = 2.3 P
0.01 0.1 m 0.2 m
1 1.2 m 2.3 m
10 4.0 m 7.0 m
NOTES For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
At 800 MHz, the separation distance for the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by the
absorption and reflection from structures, objects, and people.
Message Description/Solution
B-1
Patient Module Messages and Channel Assignments
Message Description/Solution
A combination of any of the One or more limb leads failed. Check whether the leads are
limb leads (e.g., RA/LL) off or the electrodes need to be replaced.
V1 V1 lead failure. Check whether the lead is off or the
electrode needs to be replaced.
V2 V2 lead failure. Check whether the lead is off or the
electrode needs to be replaced.
V3 V3 lead failure. Check whether the lead is off or the
electrode needs to be replaced.
V4 V4 lead failure. Check whether the lead is off or the
electrode needs to be replaced.
V5 V5 lead failure. Check whether the lead is off or the
electrode needs to be replaced.
V6 V6 lead failure. Check whether the lead is off or the
electrode needs to be replaced.
A combination of any of the One or more chest leads failed. Check whether the leads
chest leads (e.g., V1/V3) are off or the electrodes need to be replaced.
Channel Assignments
Table B-2 shows channel assignments for the Wireless 608 system; the table shows the UHF
TV channel occupying the same frequency range.
Table B-3 on page 4 shows the hexadecimal/decimal channel mapping for the 64 channels in
the WMTS band that is set aside by the FCC for medical wireless transmissions.
Table B-4 on page 5 shows channel assignments for the Wireless 2500 system; the table
shows the UHF TV channel occupying the same frequency range.
Table B-2 Wireless 608 channel assignments (FCC Part 15 compliant) (continued)
The following table shows channel numbers for the Wireless 608 system in hexadecimal and
the corresponding decimal number. The channels shown are the 64 channels in the WMTS
(medical) band.
00 = 0 0A = 10 14 = 20 1E = 30 28 = 40 2 = 50 3C = 60
01 = 1 0B = 11 15 = 21 1F = 31 29 = 41 33 = 51 3D = 61
02 = 2 0C = 12 16 = 22 20 = 32 2A = 42 34 = 52 3E = 62
03 = 3 0D = 13 17 = 23 21 = 33 2B = 43 35 = 53 3F = 63
04 = 4 0E = 14 18 = 24 22 = 34 2C = 44 36 = 54
05 = 5 0F = 15 19 = 25 23 = 35 2D = 45 37 = 55
06 = 6 10 = 16 1A = 26 24 = 36 2E = 46 38 = 56
07 = 7 11 = 17 1B = 27 25 = 37 2F = 47 39 = 57
08 = 8 12 = 18 1C = 28 26 = 38 30 = 48 3A = 58
09 = 9 13 = 19 1D = 29 27 = 39 31 = 49 3B = 59
The following table shows channel assignments for the Wireless 2500 system; the table shows
the UHF TV channel occupying the same frequency range.
Table B-4 Wireless 2500 channel assignments (FCC Part 15 compliant)
Table B-4 Wireless 2500 channel assignments (FCC Part 15 compliant) (continued)
Table B-4 Wireless 2500 channel assignments (FCC Part 15 compliant) (continued)
Troubleshooting Issues
When you encounter a problem working with the wireless patient module, address the issue in
the following order:
1 Review Table B-5, “Troubleshooting wireless patient module issues,” on page B-8 to see
whether your issue is addressed.
2 Review the Troubleshooting information in the StressVue Instructions for Use.
3 Contact your StressVue administrator.
4 Contact the Philips Response Center.
For phone numbers, see “Philips Response Center Contact Information” on page B-11.
The Philips Response Center can assist with product troubleshooting and provide
technical expertise to help with any issue with the StressVue system or patient modules.
For details, visit the following URL on the Web:
www.medical.philips.com/main/services/response_center
Packet errors/ signal Signal on the StressVue screen Review any messages in the upper right
drop shows artifact and/or square corner of the display. A lead failure
waves intermittently. message (Battery Low) indicates the
system is not receiving ECG signals.
Check that no 2.4 GHz or 608MHz
network units operate in the proximity of
the StressVue system.
Perform an RF Scan to locate the source
of the interference and to change to the
optimal channel for the wireless patient
module. See Chapter 5, “Configuring
StressVue” of the StressVue Instructions
for Use.
No waveform signal on With the patient (or simulator) Check the StressVue Miscellaneous
the StressVue screen. connected, no waveforms are settings, as well as the wireless patient
displayed on the StressVue module settings to ensure that the two
screen; only the square waveform units are set to the same channel.
and lead-fail indicator appear.
Check patient module for leads off
indication.
NOTE: If neither the signal nor the
squared waveform is present, click For details about the patient module settings,
Abort and start a new exam from see “Setting the Transmission Channel” on
the Main window. page 2-9.
For details about setting the receiving
channel settings on the StressVue PC, see
Chapter 5, “Configuring StressVue” of the
StressVue Instructions for Use.
Part number
Serial number
Have the model number, serial number, and other system information available when calling
with questions and service requests.
Manufacturer
Product Information
Date of Purchase / /
Serial Number
(under the battery in the
patient module)
Software Version
South America
Country Telephone Number
Europe
Country Telephone Number
Europe
Country Telephone Number
Asia Pacific
Country Telephone Number
C-1
ASHE Registration
Coordinating Frequencies
Before proceeding, have available the following information:
Registration confirmation and site access information from the registration process
described in the previous section.
County in which the facility using the devices is located.
Latitude and longitude that represents the center of the area where the transmitting devices
will be deployed. Comsearch, the third-party that provides the frequency coordination
services, can help provide this information through their website, www.comsearch.com.
Names of the clinical units using the wireless devices (for example, Diagnostic Cardiology
department, 4th floor, Building One, and so on).
Radius of deployment, expressed in meters. Imagine drawing a circle around the center of the
clinical unit that encloses/encompasses the unit. What is its radius?
Number of the highest floor on which a transmitting device will operate.
Number of transmitting devices in use; that is, the number of Philips wireless patient modules in
use.
Effective Radiating Power: < 1 mW
Frequency range to be used: Select all frequencies in the range of 608.0 through 632 MHz.
D-1
Index
A P T
AAMI electrode placement 2-15 patient cables technical support
attaching electrodes 2-15 approved cleaning solutions 3-2 how to contact B-8
cleaning 3-1, 3-2 system information log B-9
connecting to patient module 1-6 technical support, contacting B-11
C detaching from patient module 1-6
patient instructions
cables, cleaning 3-2 before taking ECG 2-14 W
cleaning patient interface module (PIM)
approved cleaning solutions 3-2 wireless transmitter
cleaning 3-2
cables 3-2 connecting to direct connect cable 1-6 safety iii-vii
electrodes 3-2 detaching from direct connect troubleshooting B-8
lead wires 3-2 cable 1-6
patient cables 3-1, 3-2 patient session
precautions 3-1 multiple patients 2-14
preparing patients 2-14
E multiple patients 2-14
skin preparation 2-15
ECGs preparing the skin 2-15
quality 2-14
electrodes
attaching to the patient 2-15
R
cleaning 3-2 response center, contacting B-11
placement 2-15
I S
safety
IEC electrode placement 2-15
wireless transmitter iii-vii
Instructions for Use
shaving. See patient
conventions used in 1-9
skin preparation before taking ECG 2-15
organization of 1-9
StressVue
cleaning 3-1
L maintenance 3-1
system information log B-9
lead wires, cleaning 3-2
log. See system information log
M
maintenance precautions 3-1
multiple patients, preparing 2-14
Index-1
Copyright © 2006
Koninklijke Philips Electronics N.V.
All rights are reserved. Permission is granted to copy
and distribute this document for educational purposes.
Printed in the U.S.A.
July 2008
Edition 2
453564103061
*453564103061*
*2*