PRL - Allura - FD20C PREFERRED ROOM LAYOUT
PRL - Allura - FD20C PREFERRED ROOM LAYOUT
6'-0" 24'-10"
MG
ATY SP
PBK
TV
8'-9"
8'-9"
MP
MSA
MA
Clea/Table
Centerline
17'-3"
Head End MAV
Customer's
Physio DB
(PHY)
7'-0"
CY
Equipment Layout
Recommended Ceiling Height : 9' - 6 3/16 ", +3/8 / -0
(2900mm, +10mm / -0)
Philips
Equipment Designation Detail Sheet Equipment Designation Detail Sheet
MSA Angio Diagnost 5 with Pivot 787 375 DB Documentation Box - Quantity = 4 75 0
(Final location to be coordinated with customer (each)
MG Velara Generator 40E Cabinet (MG) 510 4780 and/or local Philips service.)
MP Peripherial 40E Cabinet 510 11950 ATY Exam Room Auxiliary Box 7 1.7
PBK PDU 4000/UPS (Installed By Contractor) 860 2450 TV Three LCD Monitor Suspension 585 512
MA Mains 40E Cabinet 710 5634 MAV Mavig Ceiling Track w/ Radiation Shield 167 350
Power Requirements
Velara with PDU 4000
Supply Configuration:
3 Phase, 3 Wire Power and Ground, Delta or Wye
3 Phase, 4 Wire Power with Neutral + Ground, Wye.
Philips Medical Systems North America Company, 710 Bridgeport Avenue, Shelton, Connecticut 06484 2
Equipment specifications are subject to change without notice. Documentation number 4535 983 01531. February, 1998