SCHMIDT® Flow Sensor

SS 20.515
SS 20.516

Product description
Flow sensor with integrated
transducer for flow
measurement of air, especially
designed for the monitoring of
the laminar airflow in
cleanroom installations.

Optimized sensor head


resulting in widely independent
direction characteristics.
Cleanroom suitable design
according to GMP standard.

Special mounting adaptors


allow mounting in nearly any
cleanroom ceiling system.

Product advantages Applications SS 20.515 SS 20.516


– Design according to GMP – Laminar flow monitoring in For disinfection with H2O2 Sensor tip with protective
– Independent of flow cleanrooms and sterile coating made of two-
direction bottling machines component polyurethane
– High precision – Flow monitoring in the resine in polyacrylate /
– Long term stable signal production of polyurethane basis.
– Calibration certificate pharmaceuticals, For disinfection with alcohol.
available semiconductors, food and
– Easy mounting optical products
– Flow control in isolators,
filter-fan-units, flowboxes
Form 1014/04/06/1000/2 · Artikel-Nr. 500 989.02 · Technische Änderungen vorbehalten – Data subject to change

Mounting versions
1 2 3 4 5

Type 1 Type 2 Type 3 Type 4 Type 5


For mounting in ceilings For mounting in ceiling For mounting in ceiling For welding in stainless For mounting under the
or frames with thickness frames with existing frames with thickness of steel ceilings. ceiling with two screws
from 1 to 22 mm. opening with PG21 21 to 40 mm, especially For pressure-tight M6, when it is not
Opening Ø 26 mm winding (e.g. sprinkler for ceiling frames made mounting. possible to countersink in
necessary (mounting with opening in ceiling of hollow profiles. frame or ceiling.
Bitte wenden – Turn over

counternut) or a thread frames). Openings with Ø 26 mm Opening Ø 15 mm for


M25 x 1.5 in ceiling and Ø 28.5 mm are cable necessary and
frame. necessary. 2 threads M6.

Scope of delivery
– threaded bush M25 – threaded bush M25 – threaded bush M25 – welding bush – flange bush
(stainless steel) (stainless steel) (stainless steel) (stainless steel) (stainless steel)
– counternut – Thread adaptor – shank nut
M25 x 1.5 on PG21

SCHMIDT Technology GmbH · Feldbergstraße 1 · D-78112 St. Georgen · Germany


Telefon +49 (0) 77 24/899-0 · Telefax +49 (0) 77 24/899-101 · info@schmidttechnology.de · www.schmidttechnology.de
SCHMIDT® Flow Sensor
SS 20.515
SS 20.516

Technical data Mounting example


Measuring parameter standard velocity wN of air relative to
standard conditons 20 °C and 1013.25 hPa
Measuring range (wN) 0 ... 1 m/s
0 ... 2.5 m/s
Lower detection limit 0.06 m/s
Measurement inaccuracy ± (0.06 m/s + 10 % of measurement
value) for measuring range 1 m/s
± (0.15 m/s + 6 % of measurement value)
for measuring range 2.5 m/s
Repeatability 2 % of measurement value
Response time (t90) 1 s (0 to 5 m/s transient)
Operating temperature Sensor below a laminar flow unit mounted in the ceiling frame
- Sensor - 20 °C ... + 85 °C
- Electronics 0 °C ... 60 °C Accessories
Humidity range 0...95 % rel. humidity (rH) Fixing bracket for electronics 301 045
Humidity dependency negligible to 30 °C ISO calibration certificate 300 814
max. ± 5 % of measuring range in humidity
range referred on 55 % rH at 80 °C Ordering information
Pressure Range 700...1300 hPa SS 20.515 Art. Nr. 301 020 – XYZAB
Temperature gradient 2 K/min at wN = 5 m/s SS 20.516 Art. Nr. 301 025 – XYZAB
X = Sensor length 1 = 300 x 300 mm bended sensor tube
Analog output 0...10 V (ceiling distance H x throat depth L)
0...20 mA 2 = 300 mm straight sensor tube
4...20 mA (optionally) Y = Measuring range 3 = 2.5 m/s; 5 = 1 m/s
Load resistance Z = Output 1 = 0…10 V; 2 = 0...20 mA; 3 = 4...20 mA
- Voltage output ≥ 10 kΩ A = Connection 3 = plug-in
- Current output ≤ 300 Ω B = Mounting 1 = threaded bush M25 x 1.5
2 = threaded bush PG21
Supply voltage 24 VDC ± 20 % 3 = threaded bush with shank nut
Current consumption max. 103 mA (without load) 4 = welding bush
- Starting current max. 160 mA for 2 s 5 = flange bush, tight
Electrical connection plug-in connector, 4-pin
Dimensions
with female connector with screw 79,5
terminals for cables 3 x 0,56 mm2, 10,5 36,5

unshielded (cable not included)


Ø 35
29,7

Cable lengths
Ø 11,5
78 min.

Ø 67,3

Ø9
- Between sensor
and electronics 4 m fixed length Ø 18
L=4m
M20 x 1,5
42,15

- Voltage output 15 m max. ceiling SW 32


- Current output 100 m max.
11,5

SW 30
H

SW 22
Mounting
- Sensor mounted in supplied mounting bush Ø 19,5
52,5

can be adjusted under any rotation angle Ø9


- Electronics centre screw G ½ with nut
H

Dimensions
L
- Electronic housing Ø 68 mm x 81 mm aluminium anodized
- Sensor tip Ø 9 mm x 41 mm PBT glass enforced
Flow

- Sensor tube Ø 9 mm stainless steel 1.4571


Ø9

height H 300 mm (angled design)


length L 300 mm 9,3
wall

Weight ca. 300 g 78 min. L


M20 x 1,5

SW 22

11,5
Ø 11,5

Protection type IP 65
Flow
Ø 18

Ø 8,5
Ø9

29,7 42,15 9,3