Professional Documents
Culture Documents
Se-601c Manual de Servicio Ingles V1.4
Se-601c Manual de Servicio Ingles V1.4
Se-601c Manual de Servicio Ingles V1.4
Copyright
© Copyright EDAN INSTRUMENTS, INC. 2009. All rights reserved.
Statement
Edan Instruments, Inc. (hereinafter called EDAN) makes no warranty of any kind with regard to
this material, including, but not limited to the implied warranties of merchantability and fitness
for a particular purpose. EDAN assumes no responsibility for any error that may appear in this
document, or for incidental or consequential damage in connection with the furnishing,
performance or use of this material.
All information contained in this publication is believed to be correct. EDAN shall not be liable
for errors contained herein nor for incidental or consequential damages in connection with the
furnishing, performance, or use of this material. This company does not assume any liability
arising out of any infringements of patents or other rights of third parties.
Product Information
Product Name: Electrocardiograph
The electrical installation of the relevant room complies with national standards, and
I
WARNING : This device is not intended for treatment.
Upon request, EDAN may provide, with compensation, necessary circuit diagrams, and other
information to help qualified technician to maintain and repair some parts, which EDAN may
define as user serviceable.
WARNING
A WARNING label advises against certain actions or situations that could result in personal
injury or death.
CAUTION
A CAUTION label advises against actions or situations that could damage equipment, produce
inaccurate data, or invalidate a procedure.
Revision History
Date ECO# Version Description
2009.01 V1.0 1st edition
II
Table of Contents
Chapter 1 Warranty and Service............................................................1
Chapter 2 Safety Information.................................................................4
2.1 List of Symbols............................................................................................................. 4
2.2 Warning and Caution................................................................................................... 6
Chapter 3 Overview ................................................................................9
3.1 Environment and Power Supply Requirements ...................................................... 9
3.2 System Diagnostics Window ................................................................................... 10
Chapter 4 Work Theory Introduction...................................................16
4.1 Basic Theory of ECG Operation.............................................................................. 16
4.2 Composition of ECG ................................................................................................. 16
4.3 Lead and Electrode ................................................................................................... 18
4.4 Hardware Design Principle Description ................................................................. 19
4.4.1 Mainboard description .................................................................................... 20
4.4.2 ECG Board Description .................................................................................. 21
4.4.3 PS900G Power Supply Module Description ............................................... 22
4.4.4 Key Board Description.................................................................................... 23
Chapter 5 Functional Verification ........................................................24
5.1 System Tests .............................................................................................................. 24
5.2 Application System Tests.......................................................................................... 24
5.3 Safety Tests ................................................................................................................ 26
Chapter 6 Troubleshooting ..................................................................27
Chapter 7 Modules’ Malfunction Verification......................................31
7.1 Verifying the Main Board .......................................................................................... 31
7.2 Verifying the ECG Board .......................................................................................... 33
7.3 Verifying the Keyboard.............................................................................................. 35
Chapter 8 Disassembling the Monitor.................................................36
8.1 Disassembly Procedures.......................................................................................... 36
8.2 Internal Boards and Interfaces ................................................................................ 42
Chapter 9 Field Replaceable Parts ......................................................49
Chapter 10 Cleaning .............................................................................50
10.1 Cleaning .................................................................................................................... 50
10.1.1 Cleaning the Main Unit and Patient Cable ................................................ 50
10.1.2 Clean the Electrodes .................................................................................... 50
10.1.3 Clean the Print Head .................................................................................... 50
10.2 Disinfection ............................................................................................................... 51
Chapter 11 Maintenance.......................................................................52
11.1 Recharge and Replacement of Battery ................................................................ 52
11.2 Recording Paper ...................................................................................................... 53
III
11.3 Maintenance of Main Unit, Patient Cable & Electrodes..................................... 53
IV
SE-601 series 6-Channel Electrocardiograph Service Manual
Standard Service
EDAN provides a one-year-warranty for the warranted products (accessories are included). The
warranty period begins on the date the products are shipped to customers. If customer promptly
notifies EDAN of customer’s warranty claim hereunder, EDAN will either repair, adjust or replace
(with new or exchange replacement parts) the EDAN’s product. EDAN warrants that any service it
provides to customers will be performed by trained individuals in a workmanlike manner.
Limitation of Warranty
Direct, indirect or final damage and delay caused by the following situations for which EDAN are
not responsible may void the warranty:
Service Procedure
Fill in the SCF with detailed information including: Model Name, Serial Number (SN) and
Problem Phenomena.
EDAN should not have any obligation to take over the case without this information. The form
can be downloaded at: http://www.edan.com.cn or obtained from EDAN’s Service Department.
Once service department receives the fully filled SCF, EDAN’s engineer will offer a solution in
three working days. EDAN will follow out the case based on the following two conditions:
Within Warranty:
-1-
SE-601 series 6-Channel Electrocardiograph Service Manual
i) After receiving the Return Material Authorization (RMA) form from EDAN service
department, the customer sends EDAN the defective parts and informs the shipment tracking
number. Then we will dispatch new part(s) to your confirmed address with confirmed shipping
invoice.
ii) The customer signs the Declaration Form and sends it back by email or fax. This form is
legally certificated to make sure the customer or end-user will return the defective parts to
EDAN on time. We will, at this option, dispatch the replace one(s) with confirmed shipping
invoice.
NOTE:
(1) Both Return Material Authorization Form and Declaration Form are offered by
EDAN service department once the SCF is confirmed by service engineer.
(2) Customer is responsible for freight & insurance charges when the equipment
is shipped to EDAN for service including custom charges. EDAN is
responsible for the freight, insurance & custom charges from EDAN to
customer.
Out of Warranty:
After receiving the RMA from service department, customer sends defective parts to EDAN in
advance. We will analyze the problems and discuss with customer about either repairing or
replacing the part(s). Once the maintenance fee is invoiced and paid, we will make sure to
dispatch good part(s) to confirmed address.
NOTE: Customer is responsible for any freight & insurance charge for the
returned product.
Before the shipment of the materials, customer must obtain a RMA form from our service
department, in which the RMA number, description of returning parts and shipping instruction
are included. The RMA number should be indicated on the outside of the shipping container.
-2-
SE-601 series 6-Channel Electrocardiograph Service Manual
NOTE: EDAN should not have any obligation to end-user or customer who
returns the goods without the notification by EDAN’s service
department. The sender takes the whole responsibility of accounted fee.
² Disassemble the parts with anti-static facility. Do not touch the parts with naked hand.
² Pack the parts safely before returning.
² Put the RMA number on the parcel.
² Describe the returned parts. The total value on the invoice should be less than USD100, and
note on the invoice as “sample, no commercial value”.
² Confirm the invoice with EDAN before shipment.
² Send back the parts after EDAN’s confirmation.
Contact Information
If you have any question about maintenance, technical specifications or malfunctions of devices, do
not hesitate to contact us.
E-mail: support@edan.com.cn
-3-
SE-601 series 6-Channel Electrocardiograph Service Manual
External output
External input
Potential equalization
Mains supply
USB port
Battery indicator
Delete key
Enter key
Esc key
-4-
SE-601 series 6-Channel Electrocardiograph Service Manual
Space key
Shift key
Fn key
MODE key
RESET key
1mV/COPY key
START/STOP key
Recycle
Part Number
Serial Number
Date of Manufacture
Manufacturer
-5-
SE-601 series 6-Channel Electrocardiograph Service Manual
Warning and Caution messages must be observed. To avoid the possibility of injury, observe the
following precautions during the operation of the instrument.
WARNING :
Z Only qualified service engineers can install this equipment. And only service engineers
authorized by the manufacturer can open the shell.
Z Accessory equipment connected to the analog and digital interfaces must be certified
according to the respective IEC/EN standards (e.g. IEC/EN 60950 for data processing
equipment and IEC/EN 60601-1 for medical equipment). Furthermore all
configurations shall comply with the valid version of the standard IEC/EN 60601-1-1.
-6-
SE-601 series 6-Channel Electrocardiograph Service Manual
Therefore anybody, who connects additional equipment to the signal input connector
or output connector to configure a medical system, must make sure that it complies
with the requirements of the valid version of the system standard IEC/EN 60601-1-1. If
in doubt, consult our technical service department or your local distributor.
Z The summation of leakage current should never exceed leakage current limits while
several other units are used at the same time.
Z Improper operation may cause the battery to be hot, ignited or exploded, and it may
lead to the declination of battery’s capacity. It is necessary to read the user manual
carefully and pay more attention to warning messages.
Z Only qualified service engineer authorized by the manufacturer can open the battery
compartment and replace the battery, and the battery of same model and specification
provided by manufacturer should be used.
Z Danger of explosion -- Do not reverse the anode and cathode when connecting the
battery.
Z If using electrode gel with reusable electrodes during defibrillation, ECG recovery will
take longer than 10 seconds. EDAN recommends the use of disposable electrodes at
all times.
Z When leakage or foul smell is found, stop using the battery immediately. If your skin or
cloth comes into contact with the leakage liquid, cleanse it with clean water at once. If
the leakage liquid splashes into your eyes, do not wipe them. Irrigate them with clean
water first and go to see a doctor immediately.
Z The device and accessories are to be disposed of according to local regulations after
their useful lives. Alternatively, they can be returned to the dealer or the manufacturer
for recycling or proper disposal.
Z Only when the device is off can the battery be Installed or removed.
CAUTION :
-7-
SE-601 series 6-Channel Electrocardiograph Service Manual
Z Ruptured fuse must only be replaced with that of the same type and rating as the
original.
Z The device and reusable accessories could be sent back to the manufacturer for
recycling or proper disposal after their useful lives.
Z Federal (U.S.) law restricts this device to sale by or on the order of a physician.
Z Turn off the power before cleaning and disinfection. If mains supply used, the power
cord should also be drugged out of the outlet, and prevent the detergent from seeping
into the equipment.
Z Do not immerse the unit or patient cable into liquid under any circumstances.
Z Do not clean the unit and accessories with abrasive fabric and avoid scratching the
electrodes.
Z Any remainder of detergent should be removed from the unit and patient cable after
cleaning.
Z Do not use chloric disinfectant such as chloride and sodium hypochlorite etc.
-8-
SE-601 series 6-Channel Electrocardiograph Service Manual
Chapter 3 Overview
The service manual is a reference for periodic preventive maintenance and corrective service
procedures for the SE-601 series (SE-601A/B/C) 6-channel Electrocardiograph.
1) Mains Supply:
Rated input voltage: 100V~240V
Rated frequency: 50Hz/60Hz
Rated input power: 70VA
2) Built-in Rechargeable Lithium Battery Pack:
-9-
SE-601 series 6-Channel Electrocardiograph Service Manual
This System Diagnostics Window is password protected and only the manufacturer authorized
technician can access it.
(1) Entering the System Diagnostics Window
In the System Setup window, press Shift+F2 on the keyboard to pop up the Enter Password
interface. (See Figure 3-1) After entering the right password and pressing Enter key, the System
Diagnostics window will appear. (See Figure 3-2)
In the System Diagnostics Window, press / / / function key
and Tab key to move the cursor to a certain sub-item, and then press Enter key to get onto the test
interface of the sub-item.
System
Diagnostics
Window
Sub-item
-10-
SE-601 series 6-Channel Electrocardiograph Service Manual
-11-
SE-601 series 6-Channel Electrocardiograph Service Manual
-12-
SE-601 series 6-Channel Electrocardiograph Service Manual
-13-
SE-601 series 6-Channel Electrocardiograph Service Manual
-14-
SE-601 series 6-Channel Electrocardiograph Service Manual
Note: During the course of testing, the user should not press the ESC key to exit.
Note: The Net Test function and External Test function are reserved.
-15-
SE-601 series 6-Channel Electrocardiograph Service Manual
The heart is a power organ of the body blood circulation. Before the systole or diastole, a cardiac
impulse happens in the heart muscle, and a faint bioelectric current signal is thus generated. The
bioelectric current signal is transmitted through the whole body, and the electric potential difference
is generated on the different skin surfaces because of the different distances from the heart.
The cardiogram is an amplified record of the electric potential distribution on the body skin surface.
The electric potential difference is sampled by the electrodes, and is amplified and processed by the
electrocardiograph, and at last it is recorded on the paper. From it the doctor can detect if the heart
has a disease. The Electrocardiograph is just a kind of medical electrical equipment intended for
producing cardiograms for diagnostic purposes with 10 associated electrodes at certain specific
locations on the body. They only record the heart's electrical activity. They do not produce any
electricity of their own. The test does not hurt and has no known side effects. It does not require any
preparation (except for possibly shaving chest hair to get a better recording). The recording itself
takes only a few seconds.
There has been more than one hundred years since the electrocardiograph was applied in clinical
diagnosis. It is an important measure in clinical diagnosis of heart-disease. This instrument has been
equipped in almost every hospital and clinic.
The standard 12-lead electrocardiogram is a representation of the heart's electrical activity recorded
from electrodes on the body surface. The normal ECG wave consists of P wave, QRS complex, ST
segment, T wave and U wave. In the following figure, the x-axis is time and y-axis is voltage. When
the recording speed is 25mm/s and the sensitivity is 10mm/mV, one little grid in x-axis represents
0.04 second, and one little grid in y-axis represents 0.1mV.
This diagram illustrates ECG waves and intervals as well as standard time and voltage measures on
the ECG paper.
-16-
SE-601 series 6-Channel Electrocardiograph Service Manual
-17-
SE-601 series 6-Channel Electrocardiograph Service Manual
-18-
SE-601 series 6-Channel Electrocardiograph Service Manual
-19-
SE-601 series 6-Channel Electrocardiograph Service Manual
-20-
SE-601 series 6-Channel Electrocardiograph Service Manual
-21-
SE-601 series 6-Channel Electrocardiograph Service Manual
-22-
SE-601 series 6-Channel Electrocardiograph Service Manual
Key board is shown in the figure below. When a key is pressed down, the interrupt logic of key
matrix emits an interrupt signal. After CPU detects the interrupt signal, it will read the keys by line
scan and sequence scan, and send values to the main control board through the serial port.
-23-
SE-601 series 6-Channel Electrocardiograph Service Manual
This section describes the procedure of a complete functional test to support recommended
preventive-maintenance schedules.
Functional verification does not require opening the device case.
WARNING : Only qualified service personnel should perform a full functional checkout
procedure.
Whenever the electrocardiograph is serviced or problems are suspected, the manufacturer
recommends a Functional Check.
1. Equipment Requirement
Equipment Model
ECG emulator FLUKE MPS450
-24-
SE-601 series 6-Channel Electrocardiograph Service Manual
lead cables inserted into the ECG emulator, the electrocardiograph will accurately identify the
disconnection and connection of the corresponding lead cables.
When pulling out several lead cables simultaneously, the electrocardiograph will give the hint
information of lead off circularly.
4. Keys tests
Press every key on the keyboard to test if they work as described in the user manual.
7. Recording tests
1) Set the following values:
The heart rate output from ECG emulator is 80bpm, the work mode of electrocardiograph is auto,
the speed is 25mm/s, the AC Filter is off, the Lowpass Filter is 150Hz and all the items on the
Record Info setup interface of the System Setup window are on. Install the recording paper and
press START/STOP key to begin to record. Examine whether the recording content is integrated and
consistent with the information displayed on the LCD screen.
2) Make the ECG detector output the sine wave with 2mV and 10Hz. Press START/STOP key to
begin to record. Examine whether the recording content is integrated and consistent with the
information displayed on the LCD screen. The amplitude of ECG waveform of Lead II is 20±1mm.
3) The ECG record includes: Date and time, ID, Name, Sex, Age, Sensitivity, Paper speed, filter,
-25-
SE-601 series 6-Channel Electrocardiograph Service Manual
Lead, 1mV calibration mark, ECG waveform, Heart Rate etc. Meanwhile, before recording ECG
wave, 1mV Calibration mark and lead names will be recorded. Recorded characters and waveforms
are clear.
4) Open the casing of recorder and remove paper, the hint information of “No Paper” should be
displayed. Install the paper again and close the casing of recorder, the hint information of “No
Paper” should disappear.
1) Test the protection earth resistance according to IEC/EN 60601-1: Limit 0.1 ohm.
2) Test the earth leakage current according to IEC/EN 60601-1: Limit: NC 500 uA, SFC 1000uA.
3) Test the patient leakage current according to IEC/EN 60601-1: Limit: 10 uA (CF).
4) Test the patient leakage current under single fault condition with mains voltage on the applied
part according to IEC/EN 60601-1: Limit: 50uA (CF).
-26-
SE-601 series 6-Channel Electrocardiograph Service Manual
Chapter 6 Troubleshooting
1. System troubles
① Fuse failure.
① Change fuse.
Can not turn on the ② Power board or Main
② Change power board or Main
electrocardiograph with mains board failure.
board.
supply. ③ Silica gel keyboard
③ Change silica gel keyboard.
failure.
① Battery failure.
Can not turn on the ① Change battery.
② Main board failure.
electrocardiograph with ② Change Main board.
③ Silica gel keyboard
battery. ③ Change silica gel keyboard.
failure.
① Some keys are plunged
into the Silica gel
Can not turn into the system ①loosen these keys
keyboard
②Restart the electrocardiograph
② software fails to
function
① Strong interference of
electric networks. ① Inspect Power supply and
The electrocardiograph fails to ② Main board failure. grounded system.
function ③ Power supply or ② Change Main board.
plug-in unit on the master ③ Change or repair plug-in unit.
control board failure.
① The power supply or
Fuse is burned when turning on ① open the electrocardiograph
other component is
the electrocardiograph. for further examination.
short-circuited.
-27-
SE-601 series 6-Channel Electrocardiograph Service Manual
-28-
SE-601 series 6-Channel Electrocardiograph Service Manual
5. Parameter troubles
-29-
SE-601 series 6-Channel Electrocardiograph Service Manual
-30-
SE-601 series 6-Channel Electrocardiograph Service Manual
3) Using a multimeter, measure the voltage to earth of the test points listed below.
Item Test point Reference result
① T3 +22V±0.2V(for SE-601B)
+30V±0.2V(for SE-601C)
② T5 +3.3V±0.15V
③ T6 +12V±0.6V
④ T7 +8.3V±0.42V
⑤ T8 +5V±0.25V
⑥ T19 -9.28V±0.1V (for SE-601A)
⑦ T14 +2.0V±0.1V
⑧ T17 +8.1V±0.4V
⑨ GND
-31-
SE-601 series 6-Channel Electrocardiograph Service Manual
④ T7
⑤ T8 ③ T6 ①T3 ② T5 ⑦ T14
4) Compare the measurement results with the reference results in the list. If anyone of the results
exceeds the reference range, the main board defection is confirmed. Replacement of the main
board is recommended.
-32-
SE-601 series 6-Channel Electrocardiograph Service Manual
1) Make sure the malfunction of ECG measurement is not caused by other defects, such as bad
connection, defective electrodes or main board, etc.
4) Using a multimeter, measure the voltage to earth of the test points listed below.
Item Test point Reference result
① 5V 5.0V±0.15V
② -5V 5.0V±0.15V
③ 2.5V 2.5V±0.085V
④ -2.5V -2.5V±0.05V
⑤ 3.3V 3.3V±0.1V
⑥ 1V596 1.596V±0.014V
⑦ 1V1 1.2V±0.4V
⑧ -1V1 1.2V±0.4V
⑨ GND
-33-
SE-601 series 6-Channel Electrocardiograph Service Manual
③ 2.5V ④ -2.5V
① 5V
② -5V
⑥ 1V596 ⑨ GND
5) Compare the measurement results with the reference results in the list. If anyone of the results
exceeds the reference range, the ECG board defection is confirmed. Replacement of the ECG
board is recommended.
-34-
SE-601 series 6-Channel Electrocardiograph Service Manual
4) Using a multimeter, measure the voltage to earth of the test points listed below.
Item Test point Reference result
① T3V3 3.3±0.1V
② T+5V 5±0.5V
③ GND
② T+5V
③ GND
① T3V3
4) Compare the measurement results with the reference results in the list. If anyone of the results
exceeds the reference range, the keyboard defection is confirmed. Replacement of the keyboard is
recommended.
-35-
SE-601 series 6-Channel Electrocardiograph Service Manual
Remove the eight Machine Screws in the lower case assembled to open the device. But the upper
case assembly and lower case assembly can not be separated because there are signal connectors
between them. Therefore the user should be careful when disassembling the device, avoiding
destroying the signal connectors. Remove these signal connectors that can be dismantled directly
and move out the shield box in the lower case assembled to uncover the hidden signal connectors.
And then dismantle these hidden signal connectors to separate the upper and lower case as the
following figure shows.
Lower Case
Upper Case
-36-
SE-601 series 6-Channel Electrocardiograph Service Manual
The Electrocardiograph upper case consists of Keyboard Assembly, Recorder Case and Upper Case.
And the disassembly sequence is: Keyboard Assembly > Recorder Case -> Upper Case.
Remove the fifteen tapping screws, and then the key board and the silica gel keyboard can be
dismantled from the Upper Case, pull out the axis of rotation between the recorder case and the
Upper Case, and then the Recorder Case can be dismantled from the Upper Case, as the following
figure shows.
Recorder Case
Axis of Rotation
Upper Case
Silica Gel Keyboard
Tapping Screws Keyboard
-37-
SE-601 series 6-Channel Electrocardiograph Service Manual
Remove the two tapping screws and the two groups of Fixing Bar, and then the Recorder Rotation
Axis can be dismantled from Recorder Case, as the following figure shows.
Tapping Screws
Fixing Bar
-38-
SE-601 series 6-Channel Electrocardiograph Service Manual
Main Board
Handle
ECG Board
-39-
SE-601 series 6-Channel Electrocardiograph Service Manual
LCD Screen
LCD Screen
-40-
SE-601 series 6-Channel Electrocardiograph Service Manual
LCD Screen
-41-
SE-601 series 6-Channel Electrocardiograph Service Manual
Pin Description
2 GND
3, 4 Data transmitting
5 Vcc+5V
-42-
SE-601 series 6-Channel Electrocardiograph Service Manual
1
2
Pin Description
1 Battery supply
2 Battery detecting
3 GND
3、J3-- connected to Mainboard J17 power output
Pin Description
1, 2 +12V
3, 4, 7, 10, 11 GND
5, 6 +5V
8, 9 8.1V
-43-
SE-601 series 6-Channel Electrocardiograph Service Manual
Pin Description
1, 7, 9 GND
10 Turn on/off
5 Recharging Lock
3 Turnoff detecting
6 Mains supply
8 Battery measure
4 Battery detecting
-44-
SE-601 series 6-Channel Electrocardiograph Service Manual
1、J6—connected to Mainboard
Pin Description
1 Vcc+ 5V
2, 3,10 GND
4, 5 Data transmitting
6 Turn on/off
7 Mains supply
4. Mainboard
1 2 3 4 5 6
12 11 10 9
-45-
SE-601 series 6-Channel Electrocardiograph Service Manual
Pin Description
1 Vcc+ 5V
2, 3,10 GND
4, 5 Data transmitting
6 Turn on/off
7 Mains supply
Pin Description
3 Vcc+5V
2 GND
1 Vcc+20V
Pin Description
1,5,12,19,26,33 GND
28,29 Vdd+3V
Pin Description
2 GND
3, 4 Data transmitting
-46-
SE-601 series 6-Channel Electrocardiograph Service Manual
5 Vcc+5V
Pin Description
1, 7, 9 GND
10 Turn on/off
5 Recharging Lock
3 Turnoff detecting
6 Mains supply
8 Battery measure
4 Battery detecting
Pin Description
1, 2 +12V
3, 4, 7, 10, 11 GND
5, 6 +5V
8, 9 8.1V
Pin Description
1,9,10 GND
2 Vcc+5V
-47-
SE-601 series 6-Channel Electrocardiograph Service Manual
Pin Description
2,11,12,13,14,15,16,17,18,20, GND
1,24 Vdd+3V
3,4,5,28,29,30 VHH+8.1V
Pin Description
1 Vcc+5V
2 GND
Pin Description
1,2,3 Vcc+28V
4,5,6 GND
Pin Description
1,20 GND
2 Vcc+5V
12、J1—connected to Audio
Pin Description
2 GND(MCU control)
-48-
SE-601 series 6-Channel Electrocardiograph Service Manual
-49-
SE-601 series 6-Channel Electrocardiograph Service Manual
Chapter 10 Cleaning
10.1 Cleaning
CAUTION :
Turn off the power before cleaning and disinfection. Mains supply must be switched off
if it has been in use.
The surfaces of the main unit and patient cable can be wiped with a clean soft cloth damped in
soapy water or non-caustic neutral detergent. After that, remove detergent remainder with a
clean dry cloth.
Remove the remainder gel from the electrodes with a clean soft cloth first. Take the suction
bulb and the mental cup of chest electrodes apart, and take the clamp and the metal part of the
limb electrodes apart. Clean them in warm water and be sure there is no remainder gel. Dry the
electrodes with a clean dry cloth or air dry naturally.
Dirty and soiled thermal print head will deteriorate the record definition. So it should be
cleaned at least once a month regularly.
Open the recorder casing and remove the paper. Wipe the print head gently with a clean soft
cloth damped in 75% alcohol. For stubborn stain, soak it with a little alcohol first and wipe it
off with a clean soft cloth. After air dried, load the record paper and shut the casing of the
recorder.
CAUTION :
Prevent the detergent from seeping into the main unit while cleaning. Do not immerse
the unit or patient cable into liquid under any circumstances.
-50-
SE-601 series 6-Channel Electrocardiograph Service Manual
CAUTION :
Do not clean the unit and accessories with abrasive fabric and avoid scratching the
electrodes.
10.2 Disinfection
To avoid permanent damage to the equipment, disinfection can be performed only when it has
been considered as necessary according to your hospital’s regulations.
Before disinfection clean the equipment first. Then wipe the surfaces of the unit and patient cable
with hospital standard disinfectant.
CAUTION :
Do not use chloric disinfectant such as chloride and sodium hypochlorite etc.
-51-
SE-601 series 6-Channel Electrocardiograph Service Manual
Chapter 11 Maintenance
1) Capacity Identification
Current capacity of the rechargeable battery can be identified according to the battery symbol
in the top right corner on LCD screen.
: Full capacity
: Capacity is limited, and hint information “BAT WEAK” will be displayed on LCD
screen. Recharge should be taken into account.
: Battery is weak, and hint information “BAT WEAK” will be displayed on LCD screen.
The battery should be recharged immediately.
2) Recharge
6-channel electrocardiograph is equipped with recharge control circuit together with built-in
rechargeable lithium battery. When connect with the mains supply, the battery will be
recharged automatically. And then the battery recharge indicator lamp ( ) and the mains
supply indicator lamp ( ) will be lit at the same time. During the course of recharging, the
symbol “ ” will flash in the top right corner of LCD screen. When the capacity of battery is
full, the symbol will stop flashing, and the battery recharge indicator lamp ( ) will be black.
Because of the capacity consumption during storage and transport, the capacity of battery is
not full while using at the first time. Battery recharge should be considered before first usage.
3) Replacement
When the useful life of battery is over, or foul smell and leakage has been found, please
contact with manufacturer or local distributor for replacement of battery.
WARNING :
♦ Only qualified service engineer authorized by the manufacturer can open the
battery compartment and replace the battery. And the battery of same model
and specification provided by manufacturer must be used.
♦ Danger of explosion -- Do not reverse the anode and cathode when
connecting the battery.
♦ When the battery’s useful life is over, contact the manufacturer or local
distributor for disposal or dispose the battery according to local regulations.
-52-
SE-601 series 6-Channel Electrocardiograph Service Manual
Note: Recording paper provided by manufacturer should be used. Other paper may
shorten thermal print head’s life. And the deteriorated print head may lead to
illegible ECG record and block the advance of paper etc.
Storage requirements:
♦ Recording paper should be stored in dry, dark and cool area, avoiding excessive
temperature, humidity and sunshine.
♦ Do not put the paper under fluorescence for long time.
♦ Be sure that there is no polyvinyl chloride or other chemicals in the storage environment,
which will lead to color change of the paper.
♦ Do not overlap the recording paper long time, or else the ECG record may trans-print each
other.
The following safety checks should be performed at least every 24 months by a qualified
person who has adequate training, knowledge, and practical experience to perform these tests.
a) Inspect the equipment and accessories for mechanical and functional damage.
b) Inspect the safety related labels for legibility.
c) Inspect the fuse to verify compliance with rated current and breaking characteristics.
d) Verify the device functions properly as described in the instructions for use.
e) Test the protection earth resistance according to IEC/EN 60601-1: Limit 0.1ohm.
f) Test the earth leakage current according to IEC/EN 60601-1: Limit: NC 500uA, SFC
1000uA.
g) Test the patient leakage current according to IEC/EN 60601-1: Limit: 10uA (CF).
h) Test the patient leakage current under single fault condition with mains voltage on the
applied part according to IEC/EN 60601-1: Limit: 50uA (CF).
The leakage current should never exceed the limit. The data should be recorded in an
equipment log. If the device is not functioning properly or fails any of the above tests, the
device has to be repaired.
-53-
SE-601 series 6-Channel Electrocardiograph Service Manual
1) Main Unit
♦ Avoid excessive temperature, sunshine, humidity and dirt.
♦ Put on the dustproof coat after use and prevent from shaking violently when moving it to
another place.
♦ Prevent any liquid from seeping into the equipment, for it will affect the safety and
performance of electrocardiograph.
2) Patient Cable
♦ Integrity of patient cable, including main cable and lead wires, should be checked
regularly. And be sure that it is conductible.
♦ Do not drag or twist the patient cable with excessive stress while using. Hold the
connector plugs instead of the cable when connecting or disconnecting the patient cable.
♦ Align the patient cable to avoid twisting, knotting or crooking in closed angle while using.
♦ Store the lead wires in a bigger wheel to prevent any people from stumbling.
♦ Once damage or aging of the cable patient has been found, replace it with a new one
immediately.
3) Electrodes
♦ Electrodes must be cleansed after use and be sure there is no remainder gel on them.
♦ Keep the suction bulb of chest electrode from sunshine and excessive temperature.
♦ After long-term use, the surface of electrodes will be oxidized because of erosion and
other causes. By this time, electrodes should be replaced to achieve high-quality ECG.
-54-
EDAN INSTRUMENTS, INC.
EC REPRESENTATIVE