Recommended Settings for Epic S for Various ENT Procedures 
 
 
By Dr. Yosef Krespi  
June 2014 
 
  
Procedure  Watts in CW  Tip  Suggestions 
Hemostasis;  Non‐contact ~2mm, Non 
‐ Control of Epistaxis  initiated Tip.  
Anterior septum; short tip 
anterior septum,  2‐3 Watts CW  300mic fiber 
Nasopharynx; long tip 
Angiomas,   Select the length of the fiber tip 
‐AV Malformations  according to anatomical site 
Cutting;  Contact, Tip will initiate in  Anterior septum; short tip 
Septal spur  tissue  Tonsil / Palate; long tip 
Uvulopalatoplasty  6‐7 Watts CW  400mic fiber  Higher wattages are needed 
Tonsil Ablation  Select the length of the fiber tip  in more fibrous/less 
Frenectomy  according to anatomical site  pigmented tissue 
Interstitial Ablation;  Contact, Penetrating Tissue, 
Firing laser, penetrate tissue 
Turbinate reduction  Initiate tip in tissue or ex‐vivo 
5‐6 Watts CW  and maintain firing until you 
Tongue reduction  400mic fiber 
exit the tissue 
Thyroid  Short or medium size fiber tip 
 
 