Professional Documents
Culture Documents
Application Note P-2: Subject: Basics of Voltammetry and Polarography
Application Note P-2: Subject: Basics of Voltammetry and Polarography
analytical purposes.
20
18
II 16
14
12
Cu rrent (u A)
E 10
4
2 – 5 mV/sec I
0
0.1 0.0 -0.1 -0.2 -0.3 -0.4 -0.5 -0.6 -0.7 -0.8 -0.9 -1.0 -1.1 -1.2 -1.3 -1.4
Ei Potential (V)
t
FIGURE 5: Polarograms of (I) 1M hydrochloric acid alone
and (II) 1M hydrochloric acid containing 0.5 mM cadmium
FIGURE 4: Programming waveform for dc polarography. ion.
The current oscillations in Figure 5 are due to the Pulse Polarographic Techniques
changing area of the mercury drop during the drop life.
This is a characteristic of the “conventional” DME, but
The techniques of normal pulse and differential pulse
not the Model 303A Static Mercury Drop Electrode. This
polarography have largely displaced dc polarography for
undesirable feature can be remedied by using sampled dc
analytical measurements. The pulse waveforms are
polarography, in which the current is only measured for
designed to enhance the Faradaic current relative to the
the last several milliseconds of the drop life. This
charging current, leading to significantly improved
technique results in a smooth polarogram that has the
detection limits. With the pulsed techniques the drop
same shape as a dc polarogram. The drop time in
time must be controlled, i.e., the mercury drop is
sampled dc polarography must be mechanically
dislodged by mechanically tapping the capillary. The
controlled.
drop knocker must be incorporated into the electrode
system. The drop time is variable and is set on the
The analytical detection limit for dc and sampled dc
polarographic analyzer. The instrument can then
polarography is about 1 ppm. The net current that is
synchronize the drop time with the waveforms described
measured by these techniques is actually the sum of two
below.
components, the Faradaic current and the charging, or
capacitance current. Faradaic current is due to the
electron transfer at the electrode surface and is I. Normal Pulse Polarography
proportional to the concentration of the material being
oxidized or reduced. Charging current is due to the The programming waveform used for normal pulse
polarography is shown in Figure 6. The DME is held at
C urrent (uA )
DC
current comes from two sources. The first is the
additional current, which must be passed to charge the
double layer capacitance of the electrode to the new 2
As a result of these current decay profiles after the FIGURE 7: Normal pulse and dc polarograms for iron in
potential is stepped, the charging or capacitance current ammonium tartrate buffer, pH 9.
decays very rapidly to nearly zero but the Faradaic
current (the current that is proportional to concentration)
decays more slowly. The current is measured about 40 II. Differential Pulse Polarography
msec after the pulse is applied, at which time the current
is almost purely Faradaic. A normal pulse versus a dc
A still greater improvement in sensitivity is made by
polarogram for Fe (III) in 0.2M ammonium tartrate
using the programming waveform shown in Figure 8.
buffer, pH 9, is shown in Figure 7. An improvement in
The technique using this waveform is called differential
C urrent (uA )
current. The differential pulse polarogram is thus a plot
of current difference versus applied potential. The use of 0.5
the pulse minimizes the effect of charging current, as in
0.4
normal pulse polarography. When both potentials, i.e., Classical DC
Polarography
the potential before and the potential after the pulse is 0.3
applied, lie either before or after the rising portion of the
0.2
polarographic wave (see Figure 5), not change in the
measured Faradaic current will be observed. However, 0.1
when at least one of these potentials is on the rising
0.0
portion of the polarographic wave, a significant Faradaic 0.0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1.0
current will be measured using the differential pulse Potential (V)
technique. The measurement of current difference gives
the differential pulse polarogram a peak shape which is FIGURE 9: Differential pulse polarography.
analogous to the derivative of the dc polarogram. A
comparison of a differential pulse and a dc polarogram is
shown in Figure 9. Notice the slope of the dc polarogram The pulse height is a parameter that can be varied in
baseline versus the baseline of the differential pulse differential pulse polarography to improve the sensitivity.
polarogram. With differential pulse polarography, the As the pulse height increases (Figure 10) the peak
peak current is a quantitative measure of concentration potential increases in an almost linear manner. As this
and peak potential is analogous to E1/2. Notice, however, polarogram of Fe (III) and Mn (II) indicates, the pulse
that peak potential and E1/2 are not identical because the height also causes resolution to decrease and quantitation
finite pulse height that is used results in a differential becomes more difficult for closely spaced peaks. One
instead of a true derivative. can manipulate sensitivity with pulse height, making the
differential pulse technique a very potent analytical tool.
I vs E
50
Pulse Height
A = 3 mV E
16.7 ms B = 10 mV
C = 25 mV
D 40
D = 50 mV
E = 100 mV
0.5 – 5s
2
PULSE REPETITION PERIOD 1
C 30
Current (uA)
56.7 ms
B 20
A
I cap
I farad
10
CURRENT
40 ms 16.7 ms
0
TIME -1.1 -1.2 -1.3 -1.4 -1.5 -1.6 -1.7 -1.8 -1.9 -2.0
Potential (V)
FIGURE 8: Differential pulse excitation waveform and FIGURE 10: Effect of pulse height on peak height and
resulting current-time behavior. resolution. 1: 20 ppm Fe (III) 2: 20 ppm Mn (III).
1600
polarographic techniques can be seen in Figure 11. The
Peak
polarograms were obtained on a solution of 1 ppm Pb2+ 1 -0.350 V
1400
1177 nA
and Cd2+ in 0.1M HNO3. The current range, drop time 2 -0.452 V
and scan rate were the same for the four curves. The 15070 nA
1200
pulse height for the differential pulse polarogram was 50
mV. 1000
C urrent (nA )
I vs E
0.20
800
0.18
Differential 600
Pulse 0.16
400
0.14
0.12 200
C urrent (uA )
0.10
0
-0.2 -0.3 -0.4 -0.5 -0.6
Normal 0.08
Pulse Potential (V)
0.06
Sampled DC
FIGURE 12: Analysis of a solder bath for Pb and Sn (II).
0.04 Supporting electrolyte: acetate buffer, pH 4.5.
DC
0.02
0.00
-0.1 -0.2 -0.3 -0.4 -0.5 -0.6 -0.7 -0.8 -0.9
Chemical Pretreatment
Potential (V)
Figure 13 illustrates the determination of 1 ppm Cu (II) I
FIGURE 11: Comparison of polarographic modes. 1 ppm the presence of several thousand ppm Fe (III). Since Fe
Pb2+ and Cd2+ in 0.1M HNO3. (III) is usually reduced at potentials more positive than
Cu (II), most electrolytes present problems when the
solution contains a large excess of Fe (III) compared to
It is also possible to distinguish between Faradaic and Cu (II). The reduction current due to Fe (III) swamps the
capacitive currents in other ways, such as ac and square current form the Cu (II) reduction. However, by reducing
wave polarography, but these will not be discussed Fe (III) to Fe (II) with hydroxylamine (hydrochloride or
here.11,12 sulfate), the analysis becomes trivial because the Fe (II) is
no reduced before the Cu (II) reduction potential.
ANALYSIS BY DIFFERENTIAL PULSE
POLAROGRAPHY Extraction
A different sample preparation technique is illustrated by
Direct Analysis the work done on Levamisole, a veterinary
pharmaceutical compound. This analysis uses an
To illustrate the different approaches to polarography
extraction technique to remove and concentrate the
analysis, several examples will be presented. The vast
Levamisole form a milk matrix. After extraction form
majority of polarographic analyses involve a simple
milk and evaporation of the solvent, the residue was taken
addition of the supporting electrolyte to the sample,
up in 0.1 M tetramethylammonium iodide and the
followed by the polarographic scan. For example, the
polarographic scan was performed. The peak in Figure
analysis of a lead-tin solder plating bath is performed in
14 is for 30 ppb Levamisole in milk.
this way. The species of interest are lead and stannous
(Sn2+) ion. The analysis is carried out by adding 100 µL
of the plating bath with a micropipette to 10 mL of an
acetate buffer, pH 4.5. The polarogram obtained is
shown in Figure 12.
D
difficult to carry out directly. The analysis is easily
performed, however, by formation of the Al (III) –
1.4
Solochrome Violet RS complex. This dye reduces at
about -0.45 V in the same buffer. Figure 15 shows that
C 1.2
one may determine levels of aluminum as low as 50 ppb
in this manner. This method has been applied to
C urrent (uA )
1.0 aluminum analyses in such diverse matrices as deodorants
B
and explosives.
0.8
I vs E
1200
0.6
1100
1000
0.4 2
A 900
0.2 800
-0.1 -0.2 -0.3 -0.4 -0.5
C urrent (uA )
Potential (V) 700
600
FIGURE 13: Differential pulse polarography of copper in
1 500
the presence of iron saturated hydroxylamine HCL. A:
Blank; B: Sample; C: Sample + 1 ppm Cu2+; D: Sample + 2 400
ppm Cu2+.
300
I vs E
200
100
100
90 -0.2 -0.3 -0.4 -0.5 -0.6 -0.7 -0.8
Potential (V)
80
FIGURE 15: Al-SVRS complex in 0.1M acetate buffer by
70
differential pulse polarography. 1: 50 ppb; 2: 125 ppb.
60
C urrent (uA )
50
Supporting Electrolyte
40
The choice of supporting electrolyte may sometimes be
30
the most important factor in a successful polarographic
1 2 20 analysis. The determination of maleic and fumaric acids,
for example, is impossible in acidic solutions because the
10
peak potentials of the two isomers are nearly identical. In
0 a phosphate buffer containing 1 M NH4Cl at pH 8.2, these
-1.1 -1.2 -1.3 -1.4 -1.5 -1.6 -1.7 -1.8
two isomers have well-separate peak potentials and
Potential (V)
analysis is straightforward as shown in Figure 16.
FIGURE 14: Analysis of levamisole in milk by differential
pulse polarography. 1: milk control; 2: milk control + 30
ppb levamisole.
Derivatization
Sample preparation can also involve conversion of a non-
reducible species to one that is reducible. The
5000 I vs E
Peak 10
1 -1.288 V
1326 nA 1
2 -1.524 V
9
4000
4390 nA
8
7
3000
C urrent (nA )
6
C urrent (uA )
5
2000
3
1000 3
2
2
1
0
-1.1 -1.2 -1.3 -1.4 -1.5 -1.6 -1.7
0
Potential (V) -0.7 -0.6 -0.5 -0.4 -0.3 -0.2 -0.1
Potential (V)
FIGURE 16: Differential pulse polarography of maleic acid
(1) and fumaric acid (2) in malic acid. FIGURE 17: Sulfide, thiosulfate and sulfite in 0.025M
NaOH and 0.025M acetate buffer. 1: 6.4 ppm S2-; 2: 16 ppm
SO32-; 3: S2O32-.
pH Effects I vs E
4
The effect of pH in an analysis can be very important and
adjustment of solution pH can often be used to analyze
several species in a single solution. A mixture of S2-,
S2O32- and SO32- is often found in paper mill process C
3
liquors. In a 0.025 M NaOH solution, only S2- shows a
differential pulse peak, as shown in Figure 17. After
C urrent (uA )
adding acetic acid to bring the sample to pH 5, the S2- is
eliminated form the solution as H2S while deaerating with 2
1
Complexation A
600
80
60
400
40
to examine the kinetics of electrode processes. Cyclic
20
voltammetry is not used for analytical determinations.
200
0
The peak height of a linear sweep voltammogram is
0 200 400 600 800 1000 1200
ppb proportional to concentration. The analytical sensitivity
0
0 2 4 6 8 10 12
of linear sweep is approximately 1 ppm. The peak
current is proportional to the square root of the scan rate.
-200
ppm I vs E
16 F
FIGURE 19: Chromium (III) peak current vs.
concentration, differential pulse polarography.
14
E
12
ANODIC AND CATHODIC LIMITS
D
10
C urrent (uA )
6
However, since mercury is itself oxidized at potentials B
between 0 and +0.3 V vs. SCE, analytes that react 4
positively of these potentials cannot be determined due to
A
the excessively high current form the mercury oxidation. 2
In these instances, another electrode material must be
used. Alternative electrodes include platinum, gold, 0
0.5 0.6 0.7 0.8 0.9 1.0
graphite and glass carbon. Glassy carbon is generally the
Potential (V)
best choice because of its inertness, hardness, nonporosity
and ability to be polished to a very smooth finish. FIGURE 20: Linear sweep voltammetry of butylated
hydroxyanisole in 0.12 M H2SO4 in 1:1 ethanol/benzene. A:
The cathodic, i.e., negative, limit in the polarogram is Blank; B: 8.26 ppm; C: 16.4 ppm; D: 24.5 ppm; E: 32.6
determined by the reduction of cations in the supporting ppm; F: 40.5 ppm.
FIGURE 21: Relative sensitivities of polarographic 11. A. J. Bard, “Electroanalytical Chemistry”, Marcel
techniques. Dekker, Inc.; New York (1966).
18. Mercury
Instrument mercury, purified by triple distillation, is
available form Bethlehem Apparatus Co. In., Front
and Depot Streets, Hellertown, PA 18055.