Inserting/Removing Batteries: Compact Water Quality Meter Laquatwin-Ec-11, Laquatwin-Ec-22, Laquatwin-Ec-33

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 4

*1 Accuracy is the closeness of agreement between Part Names  Detaching the sensor

the measured value and actual value of the stan- 1 Measurement cell 1. Power OFF the meter.
dard solution in the measurement of the same 2
2. Lift the sensor tongue
Instruction Manual (Operation) standard solution as the one used for the calibra- 2 Protection cover
tip and slide the sensor
COMPACT WATER QUALITY METER tion. Temperature during the calibration and mea- 3 Lithium batteries
1 a little away from the
surement is the same. The error of standard 4 5 8 4 MEAS switch meter.
LAQUAtwin-EC-11, LAQUAtwin-EC-22,
solutions and rounding error (1 digit) are not 5 ON/OFF switch
LAQUAtwin-EC-33 included. 6 3. Pull out the sensor all the way from the
*2 The life period if the meter is used in the backlight 6 CAL switch meter.
Specifications off mode. If the backlight is used, battery life will 3 7 Waterproof gasket  Inserting/removing batteries
shorten. 8 Strap eyelet
Model LAQUAtwin-EC- 11 22 33
*3 IP67: no failure when immersed in water at a  Inserting the batteries
Minimum sample volume 0.12 mL    depth or 1 meter for 30 minutes. 7 1. Power OFF the meter.
Range and resolution (valid digits) Please note that the meter can not be used 2. Slide both batteries into
Conductivity underwater the battery case as
0 to 199 S/cm: 1 S/cm Note
shown.
200 to 1999 S/cm: 1 S/cm     Items in package Press the switches 0.5 seconds or more unless Be sure to use two Battery
2.00 to 19.99 mS/cm: 0.01 mS/cm otherwise specified.
Items Quantity CR2032 batteries, and put Battery clip
20.0 to 199.9 mS/cm: 0.1 mS/cm   them with the plus sides
TDS 0.0 to 99.9 ppm: 0.1 ppm Sensor S070 1 (+) upwards.
100 to 999 ppm: 1 ppm  Initial Setup
Meter 1  Removing the batteries
1000 to 9990 ppm: 10 ppm
Calibration Up to 2 points 
Storage case 1  Attaching/detaching the sensor 1. Power OFF the meter.
Up to 3 points   Batteries CR2032 2 2. Use a ball-point pen or
 Attaching the sensor
Standard solution 1413 μS/cm 1 other tool to pry the batter-
Accuracy*1 1. Power OFF the meter.
12.88 mS/cm 1 ies out from the clips as
2% full scale (for each range)  2. Confirm that the waterproofing gasket is shown.
2% full scale (0 to 19.99 mS/cm) Pipette 1
  clean and undamaged.
5% full scale (20.0 to 199.9 mS/cm) Conditioning solution 1  Electrode conditioning
Temperature display 0 to 50.0C   Instruction manual (Operation) 1 3. Slide the sensor onto Waterproof
the meter so that gasket Note
Target Electrical conductivity Instruction manual (Before use) 1 catch "A" on the back
Measurement 2 electrode bipolar AC  Before using the sensor for the first time or
of the meter fits into
principle Titanium coated with Platinum black  Consumable parts sold separately after several days of disuse, perform electrode
hole "a" on the sen-
Display Custom (monochrome) digital LCD with A conditioning.
backlight Items Specifications Part No. sor tongue as shown. a
 Perform calibration after electrode condition-
Operating 5 to 40C, 85% or less relative humidity Sensor S070, COND 3200459672 Note ing.
environment (no condensation)
Power CR2032 batteries (2)
Standard 514-22, 1413 μS/cm 3999960110 Be careful not to twist the waterproof gasket.
solution 1. Place some drops of the conditioning solu-
514-23, 12.88 mS/cm 3999960111
Battery life Approx. 400 h continuous operation*2 tion into the measurement cell.
Conditioning 514-20 3999960114
Outer dimen- 164  29  20 mm (excluding projections), 2. Wait 10 min before use.
sions/mass Approx. 45 g (excluding batteries)
solution
There is no need to switch the meter ON.
Main function Temperature compensation (2%/C fixed), 3. Clean the flat sensor with running water.
waterproof*3, auto stable/auto hold, auto-
matic power OFF

CODE: GZ0000472457
March, 2017
LAQUAtwin series (EC) Operation 1  2017 HORIBA Advanced Techno Co., Ltd.
Basic Operation  Multi-point calibration Measurement When the measured value is stable, stops
blinking and the displayed value is locked with
 Power ON 1. Open the protection cover and place some  Sample setting and displayed simultaneously.
drops of the standard solution into the 1. Open the protection cover and put some
1. Press and hold the ON/ 3. Document the displayed value.
measurement cell. drops of sample into the measurement cell.
OFF switch.
Rinsing the sensor with the standard solution 4. Press the MEAS switch.
The power is switched ON, 2. Close the protection cover.
beforehand will provide a more accurate cali- The auto hold function is deactivated and
and the meter model num-  Measurement mode
bration as it will reduce sample crossover con- disappears.
ber is displayed on the LCD.
tamination. The auto stable (AS) mode and the auto hold Be sure to perform this step before starting the
 Power OFF 2. Close the protection cover and press the (AH) mode can be selected. Refer to "  Mea- next measurement. Or, you may mistake the
1. Press and hold the ON/OFF switch. CAL switch. surement mode change" (page 4) for the opera- displayed hold value for the next measured
The power is switched OFF. and blink, and the calibration value is tion to set the measurement mode. value.
 Precaution on sample setting displayed.  Auto stable (AS) mode
Note
Place an appropriate amount of a sample or stan- Protection cover This is the default setting. appears when
dard solution into the measurement cell without the measured value meets the stability criteria.  If a measured value is out of the specified
trapping bubbles inside. If not, the measurement If the value changes, disappears. measurement range, "Or" is displayed for
may be inaccurate. 1. Confirm that the meter is in the measure- upper range and "Ur" is displayed for under
ment mode, and place a sample on the sen- range.
Appropriate Too little sor.  Ambient air may cause the measured values
to fluctuate. To reduce environmental interfer-
After the calibration is complete, and When the read value meets ence, close the protection cover.
Cell Electrode stop blinking and the measured value is dis- the stability criteria,  When you have a problem with the calibration
Cross-section of the measurement cell played. appears and the reading is or measurement, refer to frequently asked
The calibration value at 25C is displayed for 1 locked. questions.
Calibration s and the display returns to the measurement 2. Document the displayed value when
Calibration is required before measurement. mode automatically. appears.
Use standard solution within the measurement 3. Open the protection cover and remove the If the read value does not meet the stability cri-  Measurement display change
range in the specifications. standard solution. Then remove moisture teria, disappears and the reading changes Measurement display change is available on
Tip on the sensor by gently dabbing with a soft with time. LAQUAtwin-EC-22 and LAQUAtwin-EC-33.
 Calibration values are saved even if the meter tissue. The display mode switches as follows by press-
 Auto hold (AH) mode ing the MEAS switch in the AS mode.
is switched OFF. This completes the 1st point calibration. appears when the measured value meets  LAQUAtwin-EC-22:
 Calibration value is rewritten if calibration is 4. To perform 2nd point calibration, repeat the stability criteria. The reading then locks Between conductivity and temperature alter-
repeated using the same standard solution. steps 1. to 3. and will not change until the MEAS switch is nately
 Calibration error pressed for the next measurement.  LAQUAtwin-pH-33:
 Calibration points 1. Confirm that the meter is in the measure- Among conductivity, TDS, and temperature
If blinks and Er4 (error dis-
The number of calibration points is dependent on play) appears, the calibration ment mode, and place a sample on the sen-
the meter model. has failed. sor.
 LAQUAtwin-EC-11: Perform electrode conditioning.
Up to two-point calibration at 1413 μS/cm and 2. Press the MEAS switch.
Check that the correct standard solution is used,
12.88 mS/cm and repeat calibration after cleaning the sensor. The auto hold function is acti-
 LAQUAtwin-EC-22 and LAQUAtwin-EC-33: If the calibration repeatedly fails when using the vated.
Up to three-point calibration at 1413 μS/cm, correct standard solution(s), the sensor may blinks until the mea-
12.88 mS/cm, and 111.8 mS/cm have deteriorated. Replace the sensor with new sured value has stabilized.
one.

LAQUAtwin series (EC) Operation 2 CODE: GZ0000472457


Maintenance If Er4 (error display) appears, the adjustment Appendix Question Answer
has failed. Retry the above steps increasing
 Storage the time spent on the step 1.  Frequently asked questions Er4 is dis- Please note that if you press the
1. Clean the sensor with tap water. If the adjustment repeatedly fails, the sensor played during CAL switch in mV or temperature
may have deteriorated. Replace the sensor Question Answer the calibra- display mode, Er4 is displayed. This
2. Dab gently with soft tissue or cloth to
remove moisture on the sensor and meter. with new one. How can I Perform 2-point calibration. If cali- tion is because there is no calibration
 Initializing calibration data check the bration error occurs, the sensor has facility available for these modes.
Note Initialize calibration in the following cases. sensor's con- deteriorated. Replace the sensor. Er1 is dis- The internal IC in the meter may be
Especially be sure to treat the sensor gently to  To delete the calibration data dition? played soon defective. Perform meter initializa-
prevent damaging it.  If the number of points for the last calibration is power ON. tion.
Can I mea- This meter cannot measure a sam-
uncertain. sure high or ple with temperatures outside the If Er1 is still displayed after the initial-
3. Close the protection cover before storing  After the sensor is replaced. ization, the internal IC in the meter is
low tempera- meter's operating temperature range
the meter. 1. Press and hold the CAL and ON/OFF ture sam- (5C to 40C). defective. Replace the meter with a
 Temperature sensor adjustment switches for over 3 seconds when the ples? The difference between the sample new one (the meter cannot be
meter is switched OFF to Initialize calibra- temperature and ambient tempera- repaired).
Temperature sensor adjustment is available on
tion. ture increases the measurement Er2 is dis- The internal IC in the meter is defec-
LAQUAtwin-EC-22 and LAQUAtwin-EC-33.
To perform accurate measurement with correc- After a moment of all segment indication, the error. Perform measurement after played right tive. Replace the meter with a new
tion for temperature effects, follow the steps software version is displayed. the sample reaches the ambient after power one (the meter cannot be repaired).
below. Normally this is not necessary. And then, the display temperature. ON.
1. Ready a reference thermometer, and allow changes as shown right.
The mea- If lights steadily in AH mode, the Er3 is dis- The internal IC in the meter is defec-
the meter and reference thermometer to 2. Press the CAL switch. sured value measured value is locked.
All calibration data is reset. played right tive. Replace the meter with a new
reach to room temperature. does not Press the MEAS switch to unlock the after power one (the meter cannot be repaired).
2. Set the display mode to temperature refer- When the initialization of cali- change after value.
bration data is complete, End ON.
ring to "  Measurement display change" changing the If the value does not change after
(page 2). appears. sample. unlocking, the sensor may be dam-
The meter automatically switches OFF.
3. Press the CAL switch. aged. Replace the sensor.
 Initializing the settings
The meter displays the setting screen for tar- All setup choices are erased. The meter is reset “Or” or “Ur” The measured value may be out of
get temperature. to the factory default values. blinks in the specified measurement range.
4. Press the MEAS switch to adjust the dis- 1. Press and hold the MEAS, CAL and ON/OFF value mea- Measure a standard solution to
played temperature on the meter to match switches for over 3 seconds when the surement. check, and if “Or” or “Ur” still blinks,
the temperature indicated by the reference meter is switched OFF to enter the initial- replace the sensor.
thermometer. ization. blinks The measured temperature is not
Pressing the MEAS switch increases the dis- After a moment of all segment indication, the during mea- within the specified operating tem-
played temperature. After the displayed tem- software version is displayed. surement. perature (5C to 40C).
perature reaches 40°C, it returns to 5°C. And then, the display If the ambient temperature is within
5. Press the CAL switch again to apply the changes as shown right. the specified range and blinks,
displayed value to the adjustment. 2. Press the CAL switch. replace the sensor.
The adjustment starts. The adjusted value All calibration data is reset. The meter Check that the batteries are inserted
blinks with and displayed. When the initialization of set- does not properly. If the battery voltage is low,
After the adjustment is complete, the adjusted tings is complete, End power ON. replace them both with new ones at
value stops blinking with MEAS and dis- appears. the same time.
played. The meter automatically switches OFF.

LAQUAtwin series (EC) Operation 3 CODE: GZ0000472457


Setup Mode  Factor setting (Only LAQUAtwin-EC-33 with the TDS method set to FACt)
The setup mode allows the user to customize the meter to his specific needs. This step is bypassed on LAQUAtwin-EC-11 and LAQUAtwin-EC-22, and when the TDS method is set to
To enter the setup mode, press and hold the MEAS and ON/OFF switches for over 3 seconds when the 442, En, or NACL on LAQUAtwin-EC-33.
meter is switched OFF. All the LCD segments appear and then the meter enters the setup mode. * The setting range is from F0.4 to F1.0.
* In this setting, pressing the MEAS switch
Tip
increases the displayed value. After the dis-
 To have the changes apply, you need to go through the entire steps from “Setup mode entry” to played value reaches F1.0, it returns to
“Setup completion” shown below. To leave a setting as it is, just press CAL switch in the setting. F0.5 (default)
F0.4.
 To exit the setup mode with no change of settings, press the ON/OFF switch earlier than pressing
CAL switch in the last step but one, or the “Backlight setting” step.  Measurement mode change
Note
The measurement mode can be switched.
 Setup mode entry The AH (auto hold) mode is applied only to
conductivity measurement.
ON/OFF switch
* Measurement display change is available in
MEAS switch AS mode* (default) AH mode
the AS mode. Refer to "  Measurement
All segment Software ver. CAL switch display change" (page 2).
indication (2 s) indication (1 s)  Backlight setting
The backlight can be switched to ON or OFF.
 Unit setting
The display units can be changed.

Backlight ON Backlight OFF


(default)
S/cm (default) S/m  Setup completion

 TDS method setting (Only LAQUAtwin-EC-33)


The TDS method can be selected from the following options only on LAQUAtwin-EC-33.
 FACt: KCl with factor adjustable from 0.4 to 1.0 (default 0.5)
END indication (1 s) Powered OFF
 442: Myron L 442 non-linear standard curve
 En: European environmental standard non-linear curve
 NACL: Non linear salinity curve
This step is bypassed on LAQUAtwin-EC-11 and LAQUAtwin-EC-22.

Custom factor 442 standard EN 27888 TDS (NaCl)


31, Miyanonishi-cho, Kisshoin Minami-ku, Kyoto,
(default)
601-8306, Japan
http://www.horiba-adt.jp
For any questions regarding this product, please contact
your local agency, or inquire from the following website.
http://global.horiba.com/contact_e/index.htm

LAQUAtwin series (EC) Operation 4 CODE: GZ0000472457