Inserting/Removing Batteries: Compact Water Quality Meter Laquatwin-Ec-11, Laquatwin-Ec-22, Laquatwin-Ec-33
Inserting/Removing Batteries: Compact Water Quality Meter Laquatwin-Ec-11, Laquatwin-Ec-22, Laquatwin-Ec-33
Inserting/Removing Batteries: Compact Water Quality Meter Laquatwin-Ec-11, Laquatwin-Ec-22, Laquatwin-Ec-33
the measured value and actual value of the stan- 1 Measurement cell 1. Power OFF the meter.
dard solution in the measurement of the same 2
2. Lift the sensor tongue
Instruction Manual (Operation) standard solution as the one used for the calibra- 2 Protection cover
tip and slide the sensor
COMPACT WATER QUALITY METER tion. Temperature during the calibration and mea- 3 Lithium batteries
1 a little away from the
surement is the same. The error of standard 4 5 8 4 MEAS switch meter.
LAQUAtwin-EC-11, LAQUAtwin-EC-22,
solutions and rounding error (1 digit) are not 5 ON/OFF switch
LAQUAtwin-EC-33 included. 6 3. Pull out the sensor all the way from the
*2 The life period if the meter is used in the backlight 6 CAL switch meter.
Specifications off mode. If the backlight is used, battery life will 3 7 Waterproof gasket Inserting/removing batteries
shorten. 8 Strap eyelet
Model LAQUAtwin-EC- 11 22 33
*3 IP67: no failure when immersed in water at a Inserting the batteries
Minimum sample volume 0.12 mL depth or 1 meter for 30 minutes. 7 1. Power OFF the meter.
Range and resolution (valid digits) Please note that the meter can not be used 2. Slide both batteries into
Conductivity underwater the battery case as
0 to 199 S/cm: 1 S/cm Note
shown.
200 to 1999 S/cm: 1 S/cm Items in package Press the switches 0.5 seconds or more unless Be sure to use two Battery
2.00 to 19.99 mS/cm: 0.01 mS/cm otherwise specified.
Items Quantity CR2032 batteries, and put Battery clip
20.0 to 199.9 mS/cm: 0.1 mS/cm them with the plus sides
TDS 0.0 to 99.9 ppm: 0.1 ppm Sensor S070 1 (+) upwards.
100 to 999 ppm: 1 ppm Initial Setup
Meter 1 Removing the batteries
1000 to 9990 ppm: 10 ppm
Calibration Up to 2 points
Storage case 1 Attaching/detaching the sensor 1. Power OFF the meter.
Up to 3 points Batteries CR2032 2 2. Use a ball-point pen or
Attaching the sensor
Standard solution 1413 μS/cm 1 other tool to pry the batter-
Accuracy*1 1. Power OFF the meter.
12.88 mS/cm 1 ies out from the clips as
2% full scale (for each range) 2. Confirm that the waterproofing gasket is shown.
2% full scale (0 to 19.99 mS/cm) Pipette 1
clean and undamaged.
5% full scale (20.0 to 199.9 mS/cm) Conditioning solution 1 Electrode conditioning
Temperature display 0 to 50.0C Instruction manual (Operation) 1 3. Slide the sensor onto Waterproof
the meter so that gasket Note
Target Electrical conductivity Instruction manual (Before use) 1 catch "A" on the back
Measurement 2 electrode bipolar AC Before using the sensor for the first time or
of the meter fits into
principle Titanium coated with Platinum black Consumable parts sold separately after several days of disuse, perform electrode
hole "a" on the sen-
Display Custom (monochrome) digital LCD with A conditioning.
backlight Items Specifications Part No. sor tongue as shown. a
Perform calibration after electrode condition-
Operating 5 to 40C, 85% or less relative humidity Sensor S070, COND 3200459672 Note ing.
environment (no condensation)
Power CR2032 batteries (2)
Standard 514-22, 1413 μS/cm 3999960110 Be careful not to twist the waterproof gasket.
solution 1. Place some drops of the conditioning solu-
514-23, 12.88 mS/cm 3999960111
Battery life Approx. 400 h continuous operation*2 tion into the measurement cell.
Conditioning 514-20 3999960114
Outer dimen- 164 29 20 mm (excluding projections), 2. Wait 10 min before use.
sions/mass Approx. 45 g (excluding batteries)
solution
There is no need to switch the meter ON.
Main function Temperature compensation (2%/C fixed), 3. Clean the flat sensor with running water.
waterproof*3, auto stable/auto hold, auto-
matic power OFF
CODE: GZ0000472457
March, 2017
LAQUAtwin series (EC) Operation 1 2017 HORIBA Advanced Techno Co., Ltd.
Basic Operation Multi-point calibration Measurement When the measured value is stable, stops
blinking and the displayed value is locked with
Power ON 1. Open the protection cover and place some Sample setting and displayed simultaneously.
drops of the standard solution into the 1. Open the protection cover and put some
1. Press and hold the ON/ 3. Document the displayed value.
measurement cell. drops of sample into the measurement cell.
OFF switch.
Rinsing the sensor with the standard solution 4. Press the MEAS switch.
The power is switched ON, 2. Close the protection cover.
beforehand will provide a more accurate cali- The auto hold function is deactivated and
and the meter model num- Measurement mode
bration as it will reduce sample crossover con- disappears.
ber is displayed on the LCD.
tamination. The auto stable (AS) mode and the auto hold Be sure to perform this step before starting the
Power OFF 2. Close the protection cover and press the (AH) mode can be selected. Refer to " Mea- next measurement. Or, you may mistake the
1. Press and hold the ON/OFF switch. CAL switch. surement mode change" (page 4) for the opera- displayed hold value for the next measured
The power is switched OFF. and blink, and the calibration value is tion to set the measurement mode. value.
Precaution on sample setting displayed. Auto stable (AS) mode
Note
Place an appropriate amount of a sample or stan- Protection cover This is the default setting. appears when
dard solution into the measurement cell without the measured value meets the stability criteria. If a measured value is out of the specified
trapping bubbles inside. If not, the measurement If the value changes, disappears. measurement range, "Or" is displayed for
may be inaccurate. 1. Confirm that the meter is in the measure- upper range and "Ur" is displayed for under
ment mode, and place a sample on the sen- range.
Appropriate Too little sor. Ambient air may cause the measured values
to fluctuate. To reduce environmental interfer-
After the calibration is complete, and When the read value meets ence, close the protection cover.
Cell Electrode stop blinking and the measured value is dis- the stability criteria, When you have a problem with the calibration
Cross-section of the measurement cell played. appears and the reading is or measurement, refer to frequently asked
The calibration value at 25C is displayed for 1 locked. questions.
Calibration s and the display returns to the measurement 2. Document the displayed value when
Calibration is required before measurement. mode automatically. appears.
Use standard solution within the measurement 3. Open the protection cover and remove the If the read value does not meet the stability cri- Measurement display change
range in the specifications. standard solution. Then remove moisture teria, disappears and the reading changes Measurement display change is available on
Tip on the sensor by gently dabbing with a soft with time. LAQUAtwin-EC-22 and LAQUAtwin-EC-33.
Calibration values are saved even if the meter tissue. The display mode switches as follows by press-
Auto hold (AH) mode ing the MEAS switch in the AS mode.
is switched OFF. This completes the 1st point calibration. appears when the measured value meets LAQUAtwin-EC-22:
Calibration value is rewritten if calibration is 4. To perform 2nd point calibration, repeat the stability criteria. The reading then locks Between conductivity and temperature alter-
repeated using the same standard solution. steps 1. to 3. and will not change until the MEAS switch is nately
Calibration error pressed for the next measurement. LAQUAtwin-pH-33:
Calibration points 1. Confirm that the meter is in the measure- Among conductivity, TDS, and temperature
If blinks and Er4 (error dis-
The number of calibration points is dependent on play) appears, the calibration ment mode, and place a sample on the sen-
the meter model. has failed. sor.
LAQUAtwin-EC-11: Perform electrode conditioning.
Up to two-point calibration at 1413 μS/cm and 2. Press the MEAS switch.
Check that the correct standard solution is used,
12.88 mS/cm and repeat calibration after cleaning the sensor. The auto hold function is acti-
LAQUAtwin-EC-22 and LAQUAtwin-EC-33: If the calibration repeatedly fails when using the vated.
Up to three-point calibration at 1413 μS/cm, correct standard solution(s), the sensor may blinks until the mea-
12.88 mS/cm, and 111.8 mS/cm have deteriorated. Replace the sensor with new sured value has stabilized.
one.