Professional Documents
Culture Documents
Service Manual
Service Manual
SERVICE MANUAL
Endomed 482U
SERVICE MANUAL
Copyright:
Enraf-Nonius B.V.
P.O.Box 810
2600 AV Delft
The Netherlands
CONTENTS (contd.)
1 INTRODUCTION
This manual has been written for the technicians involved in the service of the
Endomed 482U. Service can be carried out by the service organization of the supplier or by
any other technician authorized by Enraf-Nonius B.V.
The manufacturer will not be held responsible for the results of maintenance or repairs by
unauthorized persons.
Enraf-Nonius B.V. is further referred to as Enraf-Nonius
Safe execution of the procedures in this manual requires technical experience and general
knowledge of safety precautions.
Warnings, Cautions and Notes have been used throughout the manual to bring special
matters to the immediate attention of the reader.
The sequence of steps in a procedure may also be important from the point of view of
personal safety and prevention of damage, therefore never change the sequence of pro-
cedural steps or alter a procedure.
This manual and the information herein are copyright property of Enraf-Nonius B.V.
(Delft, The Netherlands).
Enraf-Nonius B.V. disclaims any responsibility for personal injury and/or damage to
equipment caused by:
The following materials have been used for construction of the Endomed 482U:
Please dispose off properly in case of replacement according to the local regulations.
From the point of view of safety and product liability the following data must be registered for
each unit:
The distributor as well as their dealer and/or service organization if any, is responsible for
these records. They must be able to provide these data when requested.
A configuration file gives the serial numbers and/or batch numbers of main sub-assemblies
and/or important parts per unit. The parts listed in a configuration file are defined as
registered parts. Such are for example parts as transformers, PC Boards, key-boards and
software. The registered parts of the Endomed 482U are marked with a "Y" for "Yes" in the
spare parts list in appendix A (column C).
Enraf-Nonius has the original configuration file of each unit. Any possible changes in the
original configuration of the supplied units must be filed and the distributor must be able to
provide these data. These changes can simply be stored in the service file of the specific
unit. Part, serial or batch number(s) of the new parts have to be mentioned.
In case of a claim we expect the distributor to report the claim and the actual configuration
data of the equipment in question.
The documentation set for the Endomed 482U also includes operating instructions and a
therapy book. The operating instructions manual is a recommended item for all service
engineers.
1.1.6 Installation
1.1.7 Modifications
This publication could include technical inaccuracies or typographical errors. Changes are
periodically made to the information herein; these changes will be published as Service Info's
in appendix C of this Service Manual and are found on the Partner site of Enraf-Nonius
(password required). Enraf-Nonius internet site: http://www.Enraf-Nonius.com
In order to continuously improve the quality of our products, Enraf-Nonius requests her
service organizations to report any structural problem which may be discovered. When the
same problem regularly occurs or when safety problems occur, we would like you to send us
a product complaint report. The report must contain the name, the type of the product, a
detailed description of the problem, the number of the same problems, quantity of sold
equipment and your solution to the problem (if any). The product complaint report can be
found as an enclosure in the Service Reference book.
Enraf-Nonius B.V.
P.O. Box 810
2600 AV Delft
The Netherlands
1.2 GENERAL
The Endomed 482U is a microprocessor controlled unit for application of Electro Therapy
(E.T.). The unit is equipped with a large LCD display with around it the operating keys for a
clear view of the parameter settings. The power required for the unit is supplied by a
separate power supply. An optional (internal) battery makes it possible to apply treatment
to a patient independent from the mains supply. The light weight and compact design makes
the unit also suitable for treating patients at home.
Therapeutic possibilities
The unit provides two-channel Electro Therapy which can be used independently or in
combination:
Note: The Endomed 482 units are available in different types / configurations.
This service manual contains technical information regarding the Endomed 482U
unit, type number 1497.965 (and 1497.964 - USA version).
For the ‘older’ Endomed 482, which is designed as one unit from the Endomed 481
/ 482 / 484 series units, type number 1497.961 (1497.962 - USA version and
1497.963 - UK version), see service manual part number 1497.771.
Another unit is the Endomed 482e, type number 1497.960 which is an Endomed
482U based unit with limited output current types. Electronic circuits used for high
voltage and micro-current are NOT installed this PC Board.
The table below gives a comparison of the available therapy waveforms found on the
Endomed 482U and Endomed 482e units.
High voltage X -
Micro-current X -
Russian stimulation X X
Electro therapy can be applied, depending on the selected current type, for muscle
stimulation and pain control by stimulation of nerve tissue. As the output characteristic of the
electro therapy channel can be switched, at choice, between Constant Current (CC) or
Constant Voltage (CV), the Endomed 482U is also suitable for dynamic electrode
applications.
An important feature of the Endomed 482U is that the output channels Electro Therapy (E.T)
can be used independently from each other. Treatments can be applied separately as well
as linked (treatment parameters and treatment time). When applying therapy using one E.T
output channel the second E.T channel remains available for independent use.
The Endomed 482U offers the possibility of storing frequently used settings (maximum 11) in
the units memory. This allows the user to create a file of preferred therapy modes and
parameter selections. The most frequently used settings can be stored as program number
0 (zero). This program is automatically selected after switching on.
Another feature of the Endomed 482U is the possibility of using pre-programmed therapy
protocols. For a description of these therapy protocols see the Therapy Protocols guide.
1.3.1 General
1.3.2 Classification
Marking : CE 4)
* in some combinations of pulse width, frequency and spectrum the output current is limited
to 90 mA according the IEC 601-2-10.
High voltage
Alternating time : 5 - 50 s.
Ramp up and ramp down : 0.1 - 1 s.
Micro-current
Amplitude : 10 A - 1 mA.
Output voltage range : 200 V.
Faradic currents
Frequency : 8000 Hz
Duty cycle : 95%
Amplitude : 0 - 40 mA (at a patient resistance up to 500 ohm)
hold time 1 - 60 s.
ramp-down time 0 - 9 s.
pause time (interval) 1 - 60 s.
Amplitude : 0 - 100 mA (at a patient resistance up to 500 ohm)
Dipole vector field with automatic vector function (four pole interferential therapy)
2 DESCRIPTION
Because of the modular construction of the Endomed 482U the description is on block
diagram level.
- Application module (also referred to as main PC Board. This PC Board is located on the
chassis of the Endomed 482U).
- User interface module (also referred to as user interface PC Board. This board is located
on the inside of the top cover of the Endomed 482U).
The function of the application module is to supply the output currents (E.T.).
The user interface module communicates between the application module and all user
controls (selection keys, E.T amplitude control, and parameter control) and the LCD display.
The user interface module is used in combination with a custom LCD display.
For the repair and maintenance procedures of the Endomed 482U the following test equip-
ment and accessories are required:
It is assumed that a complete set of precision engineering tools is available. For the service
of the Endomed 482U the following additional tools are required:
- Special communication cable for flashing the embedded software, part number 0416.801.
- A standard Personal Computer (PC) running Windows® 95, 98, Me, XP or NT* with one
free serial (COM) port.
- Memory flash software (embedded software for Endomed 482U), part number 0497.7xx
(7xx = version number). See spare parts list.
* Windows® 95, 98, Me, XP and NT are registered trade marks of Microsoft Corporation.
4 MAINTENANCE
4.1 GENERAL
4.2 CHECKLIST
The following procedures must be carried out during an inspection and/or after every repair:
4.2.1 Modifications, if any, according to the "Service Info" sheets. See appendix C of this manual
for further information.
4.2.2 Visual inspection. Check the electrical wiring for safety and check that all components are
properly fastened. Check that the fuses have the specified ratings (see par. 1.3.1).
- Push the fourth and the fifth switches from the top (items B+ C) at the left side of the
control panel and keep down simultaneously, during switching on (see appendix B,
fig. 14)
The value which is displayed at the bottom line of the display is the software version number
(for example: 1.0).
Note: Updating the software of the processor is done according to the procedure found at
par. 4.5.6.
Connect the Endomed 482U by means of the power supply to the mains supply and switch
on the equipment.
The Endomed 482U will carry out the so-called self test for approx. 10 seconds.
At the end the buzzer sounds (otherwise see par. 4.4.1.2). It is assumed that when the
selftest is passed all functions of the Endomed 482U are okay.
Push the third and fourth switches (items A+ B) at the left side of the control panel and
keep down simultaneously, during switching on (see appendix B, fig. 14).
Check that no graphics or segments fail in the graphic LCD display, see the display at
appendix B, fig. 14 (otherwise see par. 4.4.1.3).
- Check the switches one by one; each time the buzzer sounds (otherwise see par.
4.4.1.5).
- Check the parameter control encoders one by one; at each step of the encoder the
buzzer sounds (otherwise see par. 4.4.1.5).
To check the output currents of the Endomed 482U use a load resistor of 500 6 or 10 k6 if
requested and connect the electro therapy output to an oscilloscope.
Note: The figures 2 to 12 will give an indication of the output signal. These images can
be slightly different from the actual output signal due to component tolerances and
unit calibrations and/or adjustments.
Equipment settings
Mean frequency : 2 kHz
Stimulation frequency : 0 Hz
Frequency modulation : 0 Hz
Output mode : CV
Output current settings : slowly increase the intensity control until the display reads 1 mA
Output load : R = 500 $
For output waveform see appendix B, fig. 2. (Otherwise see par. 4.4.1.8).
* Also see the note about the output signals at par. 4.3.3.
Equipment settings : increase the intensity control until the display reads 3 mA
Output load : R = 10 k$
For output waveform see appendix B, fig. 3. (Otherwise see par. 4.4.1.8).
* Also see the note about the output signals at par. 4.3.3.
Equipment settings
Mean frequency : 2 kHz
Stimulation frequency : 0 Hz
Frequency modulation : 0 Hz
Output mode : CV
Output current settings : increase the intensity control until the display reads 50 mA
Output load : R = 500 $
For output waveform see appendix B, fig. 4. (Otherwise see par. 4.4.1.8).
* Also see the note about the output signals at par. 4.3.3.
Oscilloscope settings : 100 s/div. 50 V/div. (Or 5 V/div. using a 10:1 measuring probe).
For output waveform see appendix B, fig. 5. (Otherwise see par. 4.4.1.8).
* Also see the note about the output signals at par. 4.3.3.
Disconnect the 10 k$ load resistor and check that the display reads less than 5mA.
Oscilloscope settings : 100 s/div. 100 V/div. (Or 10V/div. using a 10:1 measuring probe)
Equipment settings
Mean frequency : 2 kHz
Stimulation frequency : 0 Hz
Frequency modulation : 0 Hz
CC Mode test:
Check that the image on the oscilloscope looks like the waveform at appendix B, fig. 6.
* Also see the note about the output signals at par. 4.3.3.
Equipment settings
Phase duration : 20 s
Stimulation frequency : 100 Hz
Modulation frequency : 0 Hz
Burst frequency : 0 Hz
Output mode : CC
Output load : R = 500 $
Output current settings : increase the intensity control until the display reads 140 mA.
For output waveform see appendix B, fig. 7. (Otherwise see par. 4.4.1.8).
* Also see the note about the output signals at par. 4.3.3.
Equipment settings
Output mode : CC
Output load : R = 500 $
Output current settings : increase the intensity control until the display reads 70 mA.
For output waveform see appendix B, fig. 8. (Otherwise see par. 4.4.1.8).
* Also see the note about the output signals at par. 4.3.3.
Warning: When testing the high voltage output of the Endomed 482U the voltage at
the connections of the load resistor could be up to approx. 450 V. Take
precautions that these connections cannot be touched.
Oscilloscope settings : 20 s/div. 100 V/div. (Or 10 V/div. using a 10:1 measuring probe).
Equipment settings
High voltage : 400 V.
Stimulation frequency : 20 Hz
Modulation frequency : 0 Hz
Burst frequency : 0 Hz
Output load : R = (open circuit)
For output waveform see appendix B, fig. 9. (Otherwise see par. 4.4.1.8).
* Also see the note about the output signals at par. 4.3.3.
Equipment settings
High voltage : 5 V.
Stimulation frequency : 20 Hz
Modulation frequency : 0 Hz
Burst frequency : 0 Hz
Output load : R = (open circuit)
For output waveform see appendix B, fig. 10. (Otherwise see par. 4.4.1.8).
* Also see the note about the output signals at par. 4.3.3.
Equipment settings
Micro current
Stimulation frequency : 20 Hz
Mode : Not Alternating.
Output load : R = 10 k6 1%.
Output current settings : 1 mA.
Reversed polarity.
For output waveform see appendix B, fig. 11. (Otherwise see par. 4.4.1.8).
* Also see the note about the output signals at par. 4.3.3.
Equipment settings
Micro current
Stimulation frequency : 20 Hz
Mode : Not Alternating.
Output load : R = 10 k6 1%.
Output current settings : 0.1 mA.
Normal polarity.
For output waveform see appendix B, fig. 12. (Otherwise see par. 4.4.1.8).
* Also see the note about the output signals at par. 4.3.3.
Check that the output current of channel 1 increases and/or decreases when the remote
control is used.
Check that the output current for channel 2 increases and/or decreases when the remote
control is used.
Check the remote control, if the remote control does not function properly see par. 4.4.1.9.
4.3.5 Program memory test (Perform only when memory read/write does not function properly).
Testing the program memory can be done if a problem occurs during operation of the unit.
See the operating manual for details about storing parameters and programs into the
program memory of the unit. When program read/write fails see par. 4.4.1.11.
4.3.6 Mains adapter (Perform only when mains adapter does not function properly)
The output of the mains adapter is 15 VDC with a maximum current of 3.3 A. Do not open
the mains adapter, there are no serviceable parts inside. Replace when defective.
4.3.7 Battery charging circuit (Perform only when charger does not function properly).
For testing the battery charging circuit an electronic testing device is required.
The schematic diagram of the testing device is found below:
Switch off the unit and disconnect the mains adapter. Open the battery compartment. If a
battery is installed remove the battery from the housing and disconnect the battery. Connect
the testing device to the battery connections. Connect the + connection to the red wire, the
black wire to the - connection and connect a digital multimeter to the testing device.
Check that the battery charger circuit passes the test, if not, see par. 4.4.1.10.
The earth leakage current and patient leakage current are applicable to the combination of
the adapter-Endomed 482U unit combination. To determine the leakage currents it is
sufficient to test the mains adapter only.
Check the earth leakage current and the patient leakage current in normal condition (NC) as
well as in the single fault condition (SFC), see the testing diagram at appendix B fig. 13.
Check that the unit passes the safety test, if not, see par. 4.4.1.12.
3. Check/replace wiring
4. Replace item(s) specified in error
code list
4.4.1.8 No output or incorrect output E.T. 1. Check/replace E.T. output socket (CN6)
2. Check/replace flatcable between CN1
(main PCB) and CN3 (conn. PCB)
3. Replace main PC Board
The Endomed 482U is continuously checked for correct operation by the software. If
malfunctioning is detected an error message is displayed. A selftest or runtime error is
indicated by an error code at the centre bottom line of the display.
0 No error
001 ... 099 System selftest / safety circuits / powersupply error
100 ... 199 Electro therapy channel 1 error
200 ... 299 Electro therapy channel 2 error
022 / 500 Unit not calibrated / EEPROM calibration error
997 ... 999 System busy, special remote operations (programming of the
processor and related operations)
If an errorcode is displayed, switch off the unit and remove all connected accessories. Wait
approx. 15 sec. before switching on the unit again. When the errorcode is displayed again
46 Battery / charger voltage too high - Check/replace mains adapter / optional battery
(runtime error) - Check/replace battery if installed
59 Battery / charger voltage too high - Check/replace mains adapter / optional battery
(selftest error) - Check/replace battery if installed
4.5.1 General
When handling static sensitive devices such as the PC Boards of the Endomed 482U the
following precautions should be observed:
After every repair a functional test must be performed (see par. 4.3).
Warning: Inside the Endomed 482U high voltages are present, beware of electrical shock.
Do not touch components or heat sinks when the cover is removed.
Removal: Switch off the unit, remove the mains adapter from the mains and disconnect
the cable from the Endomed 482U supply connector. Also remove the cable
of the remote control if applicable. Place the unit upside-down. Remove the
screws of the battery compartment cover. Remove the battery if present,
see par. 4.5.5. Remove the two screws inside the battery compartment
which secures the top cover to the housing. Remove the two screws at the
output connector housing and remove the housing. Disconnect the
connector from CN1 at user interface (can be reached via the battery
compartment). Carefully put the unit on its feet. Carefully tilt the cover from
the chassis.
Caution: Be careful not to damage the flatcable of the user interface when mounting
the cover to the chassis.
Removal: See par. 4.5.2 for removal of the cover. Disconnect the power supply wiring
from CN8, remove the ON/OFF switch/battery connector CN9. Remove the
flatcable from connector CN7. Remove the five phillips screws. Remove the
main PC Board.
Note: Replace the main PC Board with the correct type (Endomed 482U), see the
note at par. 1.2. Also see Appendix B fig. 19 and PC Board specification.
Caution: Check that the fuse has the correct value (see par. 1.3.1)
Removal: See par. 4.5.2 for removal of the cover. Remove the parameter and the
amplitude control knobs. Place the cover upside-down and remove the
mounting plate from the inside of the cover. Remove the user interface
PC Board.
Installation: Installation of the user interface PC Board is in reverse order of the removal
procedure.
Removal: Switch off the unit, remove the mains adapter from the mains and disconnect
the cable from the Endomed 482U supply connector. Place the unit upside-
down. Remove the screws of the battery compartment cover. Slide-out the
battery. Disconnect the red wire from the positive connection and the black
wire from the negative connection of the battery. Remove the battery.
Requirements for flashing the Endomed 482U unit, see par. 3.2.
The communication cable which is used is not just a cable. Inside the 9-pins RS232 SUB-D
connector a small printed circuit board is built in (with MAX232A chip), that converts the TTL
levels (0 - 5 Volts) of the Endomed 482U into standard RS232 levels (-12 and +12 Volt).
Note: The xx in the filename will indicate the software version number of the
program.
The PC must run under Windows® 95, 98, Me, XP or NT* operating system. Windows® 2000
will probably function, but has not been verified.
Plug the special serial cable to the remote control connector of the Endomed 482U and in a
free serial RS232 (COM) port of the PC.
Run the flashing program 04977xx.EXE from a diskette or copy the two files (see above) to a
random directory on you harddisk and start it from there.
After you have started the flash program, you should read the instructions carefully before
you press the “FLASH Endomed 482U” button. Not following these instructions, may result
in a faulty Endomed 482U printed circuit board.
It is very important to strictly follow the instructions below in the correct order. Not following
these instruction may result in a faulty unit.
Instructions:
- Switch the PC on and wait until Windows® 95, 98, Me XP or NT has been started up
completely. Be sure that the Endomed 482U is not connected to the PC while it is
switched on. During start up, Windows tries to detect devices which are connected to the
COM ports. This may disturb synchronisation of the communication between the PC and
the Endomed 482U
- Close all non relevant Windows applications. Other running Windows applications may
disturb the flashing process of the unit. When this happens the unit may cause problems.
See the HELP FILE of the flashing program for more information.
- Connect the Endomed 482U to the PC by means of the special communication cable, see
chapter 3.2.
- Select the right COM port (COM1, COM2, COM3 or COM4) and set a proper baudrate
(When your PC is powerful, select 57600 baud. If this fails, select a lower baudrate ). A
baudrate of 19200 should always function.
- Switch on the Endomed 482U unit and wait for the self-test to be completed.
- Press the “FLASH Endomed 482U” button. The flash processor inside the Endomed
482U will be erased and then re-programmed.
- Wait until the message appears that the flashing/programming process has been
completed successfully.
- Disconnect the Endomed 482U from the PC and switch off Endomed 482U unit.
- Switch on the Endomed 482U unit and check that the unit passes the self-test.
When you get an error message, it is very important not to switch off the Endomed 482U.
Read the trouble shooting section of the Help file of the flashing program for more details.
The Endomed 482U and its mains adapter charger are free of mechanical and/or electrical
adjustments.
This chapter contains information on the storage and transport of the Endomed 482U and the
storage conditions of the spare parts.
5.1 STORAGE
The Endomed 482U and any accessories must always be stored in their original packing.
For the spare parts of the Endomed 482U the above mentioned requirements also apply.
5.2 TRANSPORT
The Endomed 482U and any accessories should always be transported in its original
packing. This packing withstands normal shocks and vibration caused by transport.
During transport and short term storage the temperature should be between - 40 °C and
+ 70 °C. The absolute temperature is not too important, however, rapid fluctuations in
temperature must be avoided. The unit should be able to have time to adapt to the
environmental conditions before operating. This to avoid condense inside the unit which
could cause problems with safety of the unit and errors in the electronic circuits inside the
unit.
6 SPARE PARTS
This chapter deals with placing orders for spare parts (refer to appendix A) as well as the
listing of the actual spare parts (appendix A, spare parts list).
6.1 ORDERING
Spare parts can only be ordered when the following data are supplied:
- Part number
- Description
- Quantity
- Shipping details (Incoterms); normal or rapid delivery
The required data are mentioned in appendix A. For the actual spare parts programme
please refer to the spare parts catalogue which is published every year.
Enraf-Nonius only supplies parts mentioned in appendix A. Standard parts such as screws,
nuts, cleaning agents and the like are not supplied by Enraf-Nonius; these parts must be
purchased locally.
Please send your order to your contact person at the sales department of Enraf-Nonius After
receipt of an order you will receive a confirmation. Always check the ordering data.
The procedure for 'normal delivery' is the standard procedure for processing your spare parts
orders. This procedure is suitable for spare parts orders which do not require rapid delivery
such as stock replenishment orders
If you have an agreement with Enraf-Nonius about fixed delivery dates, the spare parts will
be sent together with the next delivery, unless otherwise agreed.
If spare parts for an urgent customer repair are not on local stock, rapid delivery can be
requested from Enraf-Nonius The DHL-costs will be charged. Note that rapid delivery is
subject to stock availability.
Printed circuit boards are only obtainable as spare part in their latest version. Consult the
technical documentation for instructions on how to modify the latest version of the PC Board
for use in older versions of the apparatus.
Warranty claims must be provided with the Enraf-Nonius-invoice number and with the type
and serial number of the equipment in question. Enraf-Nonius may ask you to return the
(defective) spare parts. The warranty claim must be presented to Enraf-Nonius by means of
the "Authorization Request for Return Shipment (R-form)". Refer to the return shipment
procedure as per par. 6.5.
See the R-form and select “warranty” at REASON FOR RETURN SHIPMENT (par. 2)
Defective PC Boards can be offered to Enraf-Nonius for repair. The repair order must be
presented to Enraf-Nonius by means of the "Authorization Request for Return Shipment
(R-form)". Refer to the return shipment procedure as per par. 6.5.
See the R-form and select “repair” at REASON FOR RETURN SHIPMENT (par. 2)
In most cases ‘repair’ of a unit has become an activity of replacing (exchanging) the boards
and returning the defective board to the original manufacturer (Enraf-Nonius in The
Netherlands).
Enraf-Nonius has created a system in which defective parts / PC Boards can be exchanged
instead of repaired. The exchange order can be presented to Enraf-Nonius by means of the
"Authorization Request for Return Shipment (R-form)". Refer to the return shipment
procedure as per par. 6.5.
See the R-form and select “exchange PCB / US head” at REASON FOR RETURN
SHIPMENT (par. 2)
1. Fill in the "Authorization Request for Return Shipment" form and send this form by fax to
your contact person at Enraf-Nonius in Delft.
2. Within two working days you will receive the form back again by fax. You will notice that
the R(eturn) number has been filled in; this means that your return shipment has been
registered by Enraf-Nonius
4. Pack up the goods together with the "Authorization Request for Return Shipment" form
(or a photo copy). Make sure that all goods are properly packed to prevent damage
during transport. When a return shipment consists of more than one item, pack them
separately.
5. Use suitable packing materials to ensure that the shipped goods will not damage.
After the repair order has been carried out the goods will be returned as soon as possible.
If you have an agreement with Enraf-Nonius about fixed delivery dates, the repaired goods
will be sent together with the next delivery.
The spare parts list on the next page shows all available spare parts.
The various columns are explained hereafter:
R1: A number in this column indicates a suggested quantity of spare parts for
users (Quantity of equipment in use: 1 - 3).
R2: A number in this column indicates the minimum recommended quantity for
spare parts stock at the local service organization (Quantity of equipment in
use: 1 - 50).
Part no.: The part number column shows the code for ordering the part from
Enraf-Nonius.
Description 1: This column gives the general description of the spare part.
Description 2: This column gives the technical specification of the spare part.
APPENDIX B DRAWINGS
Figure 2 Output signal test (2-pole int). Figure 3 Output signal test (2-pole int).
Figure 4 Output signal test (2-pole int). Figure 5 Output signal test (2-pole int).
Figure 6 Output signal test (2-pole int). Figure 7 Output signal test (asym-biph).
-+
Figure 8 Output signal test (low freq DF). Figure 9 Output signal test (high voltage)
Figure 10 Output signal test (high voltage). Figure 11 Output signal test (micro curr.).
PC Board specification
Description
Electro-therapy (E.T.) application module for Endomed 482U unit.
Product code(s)
0497652 Part no of complete assy (printed on label on PC Board)
2557139 Part no of PC Board only (printed on PC Board)
0497819 Equivalent service (spare part) part no (not printed on PC Board; used for ordering only)
Software
0497775 In flash ROM version 1.00
Service
Circuit adjustments: No
Serviceable parts for FSE: Fuse F1
Connector CN6
Buzzer
Repairable by manufacturer: Yes
Application / compatibility
Can be used for Endomed 482U units only.
NOTE: This PC Board is not suitable for Endomed 482 and/or Endomed 483e units.
Notes
PC Board specification
Description
User interface module for Endomed 482U.
Product code(s)
NA. Part no of complete assy (printed on label on PC Board)
NA. Part no of PC Board only (printed on PC Board)
0497812 Equivalent service (spare part) part no (not printed on PC Board; used for ordering only)
Software
NA.
Service
Circuit adjustments: No
Serviceable parts for FSE: S2, S3
Repairable by manufacturer: Yes
Application / compatibility
Can be used for Endomed 482U and Sonopuls 492 units.
Notes
C.1 GENERAL
The technical information in the service manuals of Enraf-Nonius is up to date from the date
of issue.
The necessary additional information (of any kind) will be provided in the following way:
Once every year Enraf-Nonius publishes a Service Info which lists all the Service Info's
published so far.
Once every year Enraf-Nonius publishes a Service Info which lists all the published service
manuals. Note that from that moment on these manuals contain all the Service Info's
published so far. The relevant technical information will then be bound in this appendix.
When a supplement or a revision is made to the manual, a "document history file" is added to
this chapter. This file will give information about the date of issue of supplements or
revisions.
Revisions: