Boletín Shims para Mini Fast Scan SBAC7813
Boletín Shims para Mini Fast Scan SBAC7813
2015-10-23
Announcement
Introducing the Mini Fast Scan (MFS) Shim Tool TLUNAC4164 to determine the
quantity and location of shims required to correct the laser path through the fold mirror
during adjustment or replacement of the optical assembly.
The MFS Shim tool should be used during the Fold Mirror adjustment, when the path of
the laser cannot be corrected solely using the Fold Mirror Alignment Tool TL-5726
(1F6569) or (7H5593). Follow the instructions in the Adjustments and Replacement
Guide.
The MFS Shim Tool, has been designed to work with MFS assemblies SP9J0118 and
SPAB5798 installed in CR 825,850,950,975, Classic, Elite. It will not work with CR 500.
Order TLUNAC4164
Parts
The service p/n for Galvo and MFS assembly have changed. The new service parts will
include a set of adhesive backed plastic shims. Galvos and MFS without shims will no
longer be available.
Spare shims are available in Shim Kit, SPAD0144. It is recommended that FE’s retain
any unused shims in their trunk stock.
Background
When replacing a Galvo or Mini Fast Scan assembly (MFS), the beam might be very
close to either the top edge, bottom edge or clipped by the edges of the fold mirror.
The MFS Shim tool shim can be used to “shim” the MFS assembly in the reader to
enable proper centering of the beam on the fold mirror. Galvos are NOT shimmed.
The tool will indicate whether shims are needed, and if so, whether one or two are
required. Shimming will be applied to the top mount or to the two bottom mounts of the
MFS casting.
NOTE: A maximum of two (2) shims are applied per mount of the MFS casting. Shims
should not be applied to the Galvo.
The tool should be mounted on the same holes as the fold mirror. The larger holes and
slots in the tool are used to mount the tool to the holes in the upper casting for all the
readers with the large aluminum casting assembly. (all readers except Classic/Elite) The
smaller holes and slots in the tool are used for Classic / Elite readers.
Always follow the complete set of instructions in Adjustments and Replacement Guides.