Professional Documents
Culture Documents
Astm E2847-14
Astm E2847-14
W7軸肌lγo〃Aふ
Designa置jon: E2847 - 14
TTds §tandard lS issued under the暁d designatlOn E2847; the nllmber immed】ately followlng the desjgnation lndlCa'es the year of
Or!gmal adop血01・言n the case of re‘,lSion` 'he year of las置revisIOn. A number in parentheses indicates the year of 】ast reapproval A
SuPersCrip[ epsi】on (g)一一}djcates an ed-tOrlal change since ‘he last rcvI§ion or reapprova】.
1. Scope
2. Referenced Documents
l・1 Thjs test methOd covers electronic instmments intended
2.1 ASTM SIandaI擁:2
for measureme-1' of teInPerature by detecting the intensity of
E344 Teminology Relating to Thermometry and Hydrom-
thel’mal 「adiation excha一一ged between the subject of measu.e_
et「y
men書aild lhe scnsor.
E1256 “fesI Methods for Radiation Themometers (Sing】e
1・2 The devices covered by [】lis test method are l.窮rred to Waveband Type)
as jnfrared the】・mOmeterS in thjs document. E2758Guide for Selection and Use of Wideband, Low
Tさmperature Infrared Themometers
l.3 The infrared thermometers covered in this test method
a「e instruments that are i-1tended to measul.e temPeratureS
3・耽rminology
bcIow 1000OC, meaSurC thcrmal radiation ovcr a widc band-
3.1 D帥nitioJ” Qf 7Z棚串やec擁c ro 7yzis fta履けd:
Widlh in lhe in血cd roglOn’and are dirccトrending ln ‘empera-
3.1・1 ca句, bottom, ′十十he pol.tion of the cavity radjatjon
SOurCe foming the end of the cavity.
l"4 This guidc covers best practice in ca看ibrating infrared
3工i・i Di踊S高n-The cavity bottom is the primary area
lhcrmome‘ers・ It addresses conccms that will help lhe user
Where∴an jn」血ed 'hemome'er bejng calibrated measures
PCrfom more accurate calibralions. It also llrOVides a struc加e 重adiation.
如Calculation of uncertaintics∴and rcpo血g of calibralion
3.l.2 ca↓砂radiation JO肌e, ”-a COnCaVe Shal)ed geonト
l●esu11s to include uncer'桝in'y.
etry approximating a perfect b庇kbody of contro-1ed tempera-
1.5 De(ai】s on the design a11d construction of infrared
ture and defined emissivity used for calibration of radiation
Ihemometers are not covered in this test method. thermometers.
1.6 This tes白ne‘hod does not cover infrared thermometry 3・l・2.1 Disc′`諒on-A cavity radiation sourcc is a subset of
above iOOOOC. It does not address the use o白1a叶OWband thermal radialion sources.
jn缶ared theI.mO】TleterS Or i一一什al.ed themoIneterS that do not 3.l-2.2 Disf咽ion」hノbc a∴CaVi‘y radiation source of
indicate temperature directly. Praclical va】ue for calibra‘ion, at leas1 90 % oJ the fic温-Of-View
Of a radiatio掘themome{er is expected lo be incidcnl on the
].7 The ‘′a】ues stated in SI units are to be regarded as the
CaVity bot10m. In addi‘ion, the ratio of lhe lenglh of thc cavity
Sta一一dard・ The values given in parentheses are for informa(ion
VCrSL'S thc cavily diamctcr is cxpectcd lo be grcater than or
Only.
e甲al to 5:1.
l.8 77,e Val〃C,‘唖a‘ed扉′初pomd 4‘ni,'川re ’c, be rega7加。
3工3 cat句′ Walls', "-the inside surfaces of the concave
a"’'胸nda肋77’e一′a/胸given jn pa′のt/都e.’GI“7∽〃樹naIica/
Shape forming a cavity radiation source.
COnVer高めn▲‖O S’l l‘n涼Jhat c”叩,1短deケ妙手j痢mation only
and a手e ILOt COnSideied JtandaI宏 3.1.4 c鵬tO′ne7; 7トthe individual or institution to whom the
Copy「ight ③ ASTM Internat10nal’100 Ba「r Ha「bor Dnve・ PO Box C700' West Con§hohocken, PA 1942B-2959 United §ta(es
l
山はo附0則0耽唯職Oし06高
二 言二.高 年′当寺 半 FY岬S.A.S.
在暮雪し′義 cop,A
士軸」鴎」錆】隠す鞘章董藷 cON丁RoLADA
4齢 且2847 - 14
3.l.7 feId-(申,ie男f?-a uSually circular, flat surface of a intendcd to provide the user with a consistent mcthod, Whilc
measured ol事c=r。m Which the radiation lhermome(er re- rcmaining Hexib]e in thc choice of ca]ibrzltjon equipmenl・ I白s
ccivcs radiation. (l)3 understood lha=he uncerlainty ob庇Lined depends in la∫ge Parl
3. 1.7. 1 Di,ぐC〃S-rioIトMany handheld infrared themometers upon the apparatus and instrumentation used' Therefore, Since
manufacturers include distance-tO-Size ratio (D:S) in their this guide is not prescriptive in approach, it provides deしailed
SPeCificalions. Distance-tO-Size ratio relates to lhe following instruction in uncertainty evaluation to accommodaIe the
Physical situation: at a given distance (D), the infrared ther- Variety of apparatus and instrumen†alion that Inay be em-
3.1.10 p研きe,肝-a PrOCeSS that uses adry gas to remove the the calibration can be reproduced in a se arate laboratory,
SOurCC Will be∴a CaVity source having an emissivily close Lo radiation source. the oplical charactcristics of [he infrared
unity (1.00). Howcver, becausc the field-Ofnyiew of somc thcrmomcter and the detcctor and filter characteristics of the
infrared thermometers is larger lhan typical blackbody cavily
aperturcs, a 】argc-arCa flatp】atc sourcc may bc used重br lhcsc
TAB」E I Components of uncertajn重y
calibrations. In either case, the traceab獲e measuremenl o白he
109,1
Mary喜and or the National Research Cou重ICil (NRC) in Ottawa,
 ̄「est Methods E1256
0】lt紬io, Canada.
Append寂X3
X2.3
1014.1
一The ho購ace mⅢTlbel‘S in pa「entheses 「efe「 to a l st of 「efe「ences a=he end of
10.15,2
this standa「d.
丁 年言∴∴ノ 主〔′
鵬的刷け0削o眺眺鵬0し00調
軽高車丁子も 書Y岬S.A.s.
t(、事鱒穣暮
COPIA
∵年報雷門粛電 cO軸丁韓○○▲議
句齢 且2847 - 14
jn丑arcd山crmomcler, lhc∴COnlribulion o=hcsc uncertaintics will resu虹n a smaller unccrtainty. A smaller conical angle @
may changc significanlly in the overall uncertainty budgel. Wi】1 also resu旧n a sma=er uncertainty.
provides∴a∴SOu一・Ce Of radiation血at has∴a mOre Predictable 1ivity will resul白n a sma11er uncerlainly due to leInPerature
en-issivity. However, the flatp】ate source can usually be made
gradients in this region.
less expensive]y, and ca一一be made with aL diameter large
7.1.3.3 The walls of the cavity source can be treated in
enough tO Calibrate infrared the∫mOmeterS With ]ow dis[ance to
severa] differcnt ways. A painted or ceramic surface wiIl
size ratios (D:S).
genera11y result in higher emissivity than an oxidized metal
7.1.2 Idcally, Ihe sizc oJ lhe Ihermal radialion sourcc should
surface. By ‘he same measure an OXidized metこ一I sしIrface wi】l
bc∴SPeCificd by the infrored thcrmometer manufacturer" In
generally result in higher ehissivity than a non-OXidizcd metal
many cases, this infomation may not be available・ In lhese
surface. In some cases亘may be imposstble to paln=he cavity
cases a field-Of-View Iest should be completed as discusscd in
source surface. This is especially true at high temperatures.
E1256. The porlion of signal incidcnt on lhe in血ared themom-
7.1,3.4 The e価ective cmissivity ol thc cavity source shall be
etcr that does not comc from lhe source shou】d he accountcd
calculated to detel・mine the radiance teml〕eratbre of the cavity.
for in thc unccrtainty budgct.
Calculation of e鉦新ive emissivity is beyond the scope of this
7.1.3 Caviり, So初でe:
slan a∫d. Detcrminalion of eifeclive emissivity can bc math-
7工3.1 A cavity source can be constructed in several shapes
ematically calculated or mOdeled.
as shown in Fig・ 1. In general, a high length-tO-diameter ratio
(L:D) or radius-tO-diameter ralio (R:D) in the spherical case 7.1.4 Fla印Iate Sou′t;e・●
丁静⊥
丁D⊥
ヽ
TD⊥
TD⊥
同園 Sph e「e
鵬的議了O則0唯椛鵬oしo6高
手打‡騨宣告 3
FYH S細A.S.
.露細見.移軽で、
星章で嘗㌍ ’∴
copiA
CoNT韓○○ADA
‥ 1最善を ` j / ̄
綱か 宣2847 _ 14
7.1"4.1 A皿atplate source is∴a dcvicc that consisIs of a 7.3.1 The thermal radi加ion sourcc sh租ll be calibrzltCd wilh a
Painlcd circular or rcctangular plate. ’rhe cmissivity is like工y to transfer sねndard lraceable lo a na'ional metroIogical institutc
be lcss well dcfincd than with a cavity source. This can be SuCh as lhe Nalional Ins血Ic of Standards and耽chnoIogy
Partially overcomc by performing a radiometric tran垂r (SCe (NIST) or Nalional Research Council (NRC). If a refercnce
Scheme II in 7・3.7) to the flatplate sourcc. However, the 血crmometor (radiomctric or contact) is uscd during thc cali-
radiometric transfer sh。uld bc carried o山Wjth an jnst「ument bration o白he unit-under-1est, thjs serves as thc calibrali。n O「
OPCrating over a similar spcctral band as thc in血arcd thermom- thc radiation sourcc. In this casc. thc rcfercncc thcrmomcIcr
eter under test. Shall have a calibration traceable ↓o a mtional me‘rological
7工4.2 A cavity source is the preferred radiometric source institute.
for infrared themometer calibrations. Tlle CaVi†y source has
7.3.2 This calibration of the thermal radiation source may
two main advantages over a flatplate source. First’the cavity
take place in the calibration laboratory, Or it may be done by a
SOurCe has better defined emis§ivity and an emissivity much
third party calibralion 】aboratory. The ill↑erval o白hese checks
CIoser to unity due to i†s geometric shape. Second, along wi'h
is dete重・mined by the calib‘.a書ion labOratO賞・y. The d輔related to
the emissvity being c萱oser to unily, the eifects or re伸ected
the calibration interva=s paft of the calib】・ation mlCe11ainties
temperature are lessened. Tempe裏・ature unifomity on the flat-
for the infrared tilermOmete]. Caljbration.
P】ate source may be more of a concem as wel]. However, a
7.3.3 Regardless of whether a cavity soし一rCe O「 a鯛atplate
flatplate source has a main advantage ove賞・ a CaVity source. The
SOurCe is used, there are two approaches to calibrating the
temperature cont「O]led flatplate surface can be muc=arger
SOurCe: COntaCt Calibration (Hg. 3, Scheme J) and l.adiometric
than a tyl,ical cavity source opening, allowillg for much
CalibraLtion (Fig, 3, Scheme H). (3)
Sma]]el’D:S ratios (greater frold-Of-View).
7.3.4 In軸g. 3 the arrows show the path oftraceability t。 the
7.2 ApeI●他控:
International System of Units (SI) thr。ugh a national metro-
7.2.1 An additiona] ape11ure may nOt be needed for a】】
logica] institute (NMl). The reference radiation source is the
Ca]ibrations. An ape血re js typical】y used to contro] scatter. If
Cavity source or bIackb。dy s。urCe uSed to ca皿rate the infrared
used, the aperture shou】d be temperature-COntrO】led or reflec-
thcrmomctcr. In Schcmc I, it is shown that the △T mcasurc-
tive. An aperture shouId be used if recommended by the mcnt md the emissivity correction shz111 be addcd inlo thc
infrared thermometer mamlfacturcr. If an aperture is uscd for
lemperature calculation. The △T mcasuremen白s bascd on the
Calibrafron・ this information should bc staled in [he rcporしof
di餓3renCe in temperalurc betwecn thc refercnce thcmomcter
Calibration. The inIormation lhal shall be included is the
and lhc cavity walls. Thc emissivity corrcction is based on thc
aperlure dis離しnCe, lhc aperture size, and thc measuring dis-
radiation sourcc nol having 'he same cmissivily as the infrarcd
1ance. A possib看e configuration for apcr‘ure use is shown in ng.
thcmomctcr’s cmissivity sctting. Thc symboI加rcfers to thc
2. WaVeleng瓜and bandwid山o† the lrans」tr radiation them0m_
7.2.2 In蹄基・ 2, dap∫ is thc apcrlure distancc. Thc mcasuring
eter and t重le in紐ared thermometer,
distance is shOWn by dm.ar
7.3,5 In either scheme, the transI料slandard sha看l be trace-
7.3 7tⅥ明雄?r StandaId.・ able to a national metrologica】 institute.
子癌上諒
娘軸e「mo州債劇
症精魂融胸高 う
機d洞鈎S
Å畔血鵬 囲圏
FIG, 2 Use of an Aperture fo「 a Caiib「ation
4 鵬的鵬†0則o唯調印鴫o! ● ●
霊書A重患鵬Y二番 FYH S.A.S.
無蓋鱈子言霊 COPIA
魚群A鴫竣義. ∴書直、謹・` ○○N丁慮oしみ
固寓丁由
也齢 巨2847 - 14
purge・
8. Preparation o書ApparatuS .
8. 1 ”坊働t'd 772er7nOmeter.・
tance along with the target size sha】l be recorded on the report ture. This setting should represen=he temperature of t】le
of calibra章ion. ambient surroundings facing the themal radiation source. The
9.4・1 Thc cmissivity sctting o」“ thc inharcd lhemOmCtCr thcrmometcr may have some othcr op‘ical alignmenl device
Should match the eInissivity or emissivity setting of the SuCh as lighトen血ing diodes tha' may be used as we11.
radiation source. M狐inlizing血e slgnal is the preferred mclhod.
9.4"2 Some inhared thermometers have a fixed emissivity 9・5・3.2 If using 】aser a】ignmel-t、 Cente「 †he 】aLSer On the
Setting alld some radiation sources have a fixed emi§Sivity. In Center Of thc radiation sourcc.
a ca§e Where both settings are fixed and are not equai, a 9.5.3"3 1f maximizing the signa], for calibration pojnts
mathematical a勘ustment sha11 be made. An examp賞e of such an above ambient・ the posjtion of the infrared themo】neter Sha】I
adiustme賞1t Can be found in X2.3. be a軸sted vertica]ly and ho「izontally to produce 】maXimum
9.4.3 The Preferred method is to adjust the infrared ther_ temperature whj】e aiso maintainjng t重一e ]jne of sight perpen-
momete】・ emissivity setting to the radiatio一一SOurCe・s emissivjty. dicular to the∴SOurCe. Thjs is皿stratcd in鴫. 5. In th。
Jf the radiation source receives a contact ca-ibration (Fjg. 3, CXanPle in Fig. 5・ the maximum tempcrature obscrved on lhe
Scheme I), thi§ emissivity wou]d be the emissivity of the infrared thermomeler.s readou白s 300.3OC. For ca]ibration
Surface. 1f the radjation source receives a radiometric caHbra_ POints below ambient‘ the ‘emPera山rc shall be minimizcd.
ti{)n (吋3’Scheme II)' t]一e emjssivi書y wouId be the emjssivity 9.5.3.4 ln cases whe「e the sjze o白he radiation source is
Setting of the transfer standard. If t】le emissivjty settjng of the much largcr than [hc ficld-Of-Vicw of thc radiation
i血ared thermomete】・ camot be set exactly to the e餌ective thcmometer, the ‘emperature may plalcau instead of rcach王ng
emissivity of the themal 】・adiation source言hen a correction a simple maximum or minimum. In such cases, fl dcflncd
may be made as is shown jn X2.3. Change in lempe「ature should bc obscrved whi宣c moving皿c
infrared themometer a】ong an axis. Then lhc inhared ther-
9・5 AligI硯ent.・
9.5.1 P均)aI・a子i個.・
mometer should be cenlered midway betwecn thcse 'wo points.
This shaI量be done for both axes. This is ilIustra‘cd in I帝6. In
9"5.1"1 If an addi-iona】 apertwre is used for the calibration,
this case, the in血ed lherlnOmeter is move即rom Side lo side.
ensure that the aperture is properly cmplaccd at血o specified
distance a§ Shown in rig. 2. I白he aperture is tempera[ure-
A plateau in the temperature readout of 300.3OC is observed. In
this case the user shal! observe a drop-O航=he temperature
COn(rolled’enSure tha=he apcrture is wilhin its specified
(emr)eratUrC !imi(S. readout of 3"OOC. This means the user should be Iooking for a
reading of 297.3OC. Points ・A’and ・B, indicate where this
9.5.1.2 In Fig. 4, thc mcasuring distancc is dcsignated by
`dJ The `Ⅹ, axis refers to the horizonta萱direction; the ・Y‘ axis drop-Off occurs. Poin' `C・ represents the midi)O血of ・A, and
`B’.
refers to the vertical direction; and the ・Z・ axis refers to the
壷 主 , 、
看鵬的軸げo則O o書籍惟鵬Oし0e鵬
書脅章魚農遺物守 を FYH S.A.S.
代書1聾喋墓 Cop霊A
一兎狐疑‡箪‡羊蹄悠揚 ○○N丁韓○○A議
左紗 岳2847 _ 14
閣上諒 d ⊥-諒
喜櫨冊e「肌o叩eくe「
重鼠冊を「州o問合鳴r
闇臆臆臆 d --÷融
軸t Pねte Measu「ing Distance
FiG・ 4 Caiibration Setup Showing Measuring Devices
lO,1 Overvie†t’:
Uncertainty in Measurement.”
いBO聞け0削○ ○さき惟鵬0し06j鼻
言争∴予言点上∴ l言 ′ 手・号 声Y岬S.A,S.
.善書A.露嶋Yう
COPIA
無毒てこ「二㍍ cO議了韓oしADA
氏壇船 担中華、了毘評善言
4齢 E2847 _ 14
曾▲S = emissivily o白he measured surface In Scheme I・ this is the一一nCertainty of the temperature reado
diation sollrCe
]0.5 So初でe E′n短il句′:
Tw = re鯖ected radiation temperature (walls)
10.;.1 Ca/ibration Scheme / - Fla印Iate S鋤’re-In lh
事I 二dclecIor lcmperature
CaSe, the e血ssivity of the surface is detemi11ed by some 。thl
」 O.2.2 Uncertainty due to Reflected Tbmperature" method such as Fourier-tranSform infrared or radiometr
10.2.2.1 Tb evaluate for reflec'ed tcmperature uncertainty, COmParison’Tbst methods to de↑ermine eInissivity are ou[linc
Eq =s di節erentiated to get Eq 2, This number is then used in in Guide E2758. Since the emissivi‘y for a specific surfa(
Eq 3 to ge=he temPerature meaSurement uncertainty due to COating may vary widely, thjs can cause a -arge amo…一=
Variance in the emissivity va】ue and ‘he res山ing ul-Certainl
reflected tempe農・ature. An example of this ca】cu]ation is shown
鵠幸吉 (2)
detem上ned by either modcling・ Ol・ by ‘he cmjssivity uncc
U証言吉葉(7・w)二鵠語り(Tw) (3) the emissivity may still vary over time or be dependent o
SPeCt!・aI bandwidth. However, these uncel.tainties are reduce
10.2.3"l Tb evaluate for source emissivjty uncertainty, Eq l 10.6.1 Reflected radiation is sometimes referred to as back
js di餌ercntiatcd 10 ge一】 q 4. This mmber is thcn uscd in t沌5 g賞・Ound radiation. The cause of its e飾ect js sometimes referre
to as background temperature. Tl-is ul-Certalnty is muc】l mOr
to get the uncertainty due to一・eflected temPerature. An example
Of a concem when ca]ibrating instmments with a flat可at
Of this calculation is shown in X2.4.
SOurCe than jt js with a cavjty source' It is especia一]y a concer
8 山田O紬的則o眺眺†肌o融
.時点」専一
外書緯既着 FY岬S.A.S葛
‥、∵章∵y- ..言 cOplA
cONTRO」▲D▲
綱か 且2847 - 14
10.7 Sの初Ce /1eat E抑/硯ge: for. Thc calcしIlations outlined in this stilndard ilrC bascd on zI
10.7.1 Sourcc hcal cxchange is ‘he uncertain‘y of the B初t?a〃 i,寝e〃裏ational des poid9 Ct meSl聴g (BIPM) document.
di鮒erence belween the∴SOurCe’s controI sensor or readout (4)
lcmpcra川re and thc source’s aclual surface lempera'ure. This 10.13.2 For measuring dislances grealer than l m, a mOdc]
unccr融nly is due to hcat flow betwcen the scnsor localion and Of slandard atmosphcre∴Shall be consulted. For meas皿ng
the sou「cc’s surface. I白hc flatplatc s(冊CC is calihrated wjth a djslances of l m (一l. less, lhe expanded皿ccr(alnty (k = 2) is
radiometrically, this uncerLainty is minimi∠ed. However, there norma11y O.0006.
Stil=s some uncerlainty since the heat flow may be di節erent 10"13.3 ’塙blじX2.] in X2.6 giYeS C租lculaled values 」t)r lhe
from time to lime. atmospheric abso町tion uncertainties at various temperatures.
to be exaggerated to get a true idea of this effect. the infrared thermometer readout.
10.15.2 Tt) Calculate display resoIution un6e】・tainty, take the
10.9 So蹄でe UI所)7γ扉か: 、
dispIay resolution and divide by two. This result has∴Zl
10.9. 1 Source皿iformity lS unCertall-ty due to temperature
rectangular distribution. Use standard practice to determine the
non-homogcncily on thc calibr租lor su可elCe. Sincc an in血ared
expanded uncertaiilty Of a rectallgula〇・ distribution. For
lhermOmelcr avcragc=he lcmperalures wi皿n its field-Of-
example, if an infrared thermometer has a display reso皿ion 。f
Vicw, this uneerlainly lS Calculatcd by considcring how much
O.1OC, then lhc rectangular dislribl】tion is ±0.05OC and the
lhc皿iIormity will czluSe fl differencc in measurcment between
expanded uncerlainty is O.058OC.
a mcasuremcnt of a small spol at the ccnler of lhe source and
a largcr spot c。rreSPOnding lo the in什arcd lhemometcr under 10-16 Sample Uncer/ain砂B職dget.・
uncertainties (k = 2).
10.1 1 Size-〆二Sou7℃e匂解でt:
are typiealiy ve「y low. Neverthe】ess, they should be accounted 1l.3 A料nlPle report is inc】uded in Appendix X4.
・主蒔浩二。 いさo請けO則0唯職場鵬0しo6i4
.電着鼻.霊 草玉子、 革Y物S.A.S.
操車や子羊㌻∴ 、 coplA
読(へ 自信、〔音手短言霊 C○○葛T教OLADA
の 良2847 - 14
12.1 A rccord system of all calibrations sha11 be kept. This uncertainty evaluation shaH follow the method shown in thc
APPENDIXES
(Nomlandatory Information)
XL1
Source
叫巧い山里豊 明呪い帰⋮
B B A B B A一 B A B A A
CaIibralion Temperature
Soll「Ce Emissivity
Reflected Ambienl Radiation
Sou「Ce Heat Exchange
Ambien書Conditi(〕nS
Source Unito「mity
Inf「a「ed Thermometer
Size-OトSource E償∈rot
Ambien( Tempera血re
Atmospherie Absorptton
Noise
D胆0∩ ∴∴∴∴∴∴∴∴∴
Comb]ned Expanded Uncertalnty (k=2)
Ⅹ2.1 Ceneral
X
2 2
2 ︹う
derivative of the Sakuma輸Hattori is shown in Eq X2.3. W】len PeraturC. A and B a重C dcrivcd from thc in血arcd thcrmomctcl.,s
doing the mathematics using t】leSe equatふons, Kelvins shaI! be SPeCtral rcsponsc. C is a scalar bascd on infrarc(=hemomeler
used for temperature rather than OC. gain and can bc considered as unity Ior this analysjs. Calcula-
tion for these constants is shown in Eq X2.4, Eq X2"6, Ijq X2ふ
(Ⅹ2.1) and Eq X2.7. Thc vzllue for △九is lhe inhared themomcter,s
CCntCr WaVCIcnglh buscd on zI SimpIc avcr!lgC Ol’thc in正arcd X2.4 Emissivity Uncertainty
lhcrmometcr’s high and low bandwidth limits.
X2.4.1 For this calculalion, Eq 4 and liq 5 are used.
X2.2.3 Fo「 example, the 8 to 14四一SPeCtral band is X2.4.3 Using Tat}1c X2‘l, S(Ts) = 0.0461160 and S打w) =
COnSidered. For血e 8 to 14甲m spectral band,品=】 l Hm and 0.0076966. These values are used in Eq 4 to give the value in
△九二6 tJm. Using X2.4 and X2.5, A=9’364 Hm and Bこ]78.4
し皿lK. ム
X2.2"4 1宅間e X2.] shows computed values for the 8 to 14 豊=尋(Tb・) - S(Tw)巨万封04剛- ・)・。07中
二0・0404415 (Ⅹ2.量0)
∂⊂s 事丁′ 、 0・0404415
U。 ( T,剛)
布U周= 0.脚30う968
〔)T
(X2.11)
X2.4.5 A graphical representation of a similar examp]e is
Shown in Fig, X2.=n with TS ranging froln -500C to lOOOOC.
In †his example, U(gs) = 0.0工
emissivity setling is used with a source at a different emissivity, X2.5.3 Use Eq 2 to calcu】ate ∂S(T,。一。fiS) I ∂S(Tw) as is shown
Eq l can be rewrilten as Eq X2.8. This assumesTD =Tw as the in Eq X2.12.
ambient temperature uncertain†y accoしmtS for lhis.
f)S(T鵬)_1-〔、 ]-0・95
○ ○こ「手痛-二0・052632 (X2・重2)
s(T,,吉S陣笠[S(T出Tw)] (X2.8) ∂S(T,-,)  ̄ ⊂.柄
O.00019945 K ̄ 1
s(7「剛)こ0・。202920十豊[0・0202920 - 0 0076966]
二0・0205572 (X2.9)
(X2、13)
X2.3.4 Tb detel’mille What the measured tempel・ature Should X2.5.5 A graphicaI representation of a s血i]ar examP】e is
be, uSe Eq X2.2. This resu】ts in a temperature of 374.43 K or Shown in rig, X2.2 wilh TS ranging from -50OC to 10000C. In
lO重.280C. this ex紺nple, U(Tw) = 1.O K.
. ∵γ∴)完 ・、 櫨0議†O則0唯物帥.
. ‡.了を.ゼ 門口 FY岬S,A.:
ふき埠耀こ争 Cop置▲
点膏喚 星畠までえ 申告書jィ葦 ●●こO議了ROし一
軸か 宣2847 _ 14
Co附軟調硝柳の陣場掩肌蹴雌貧血やi調京e馳債e種丁emp轡融職場
鋤丁臓かu鵬糊鳴血母町機雷ま30C,さ雪0費響き)
鼻 五 〇 0
虫き つ玉 王 8
ハU 代り 請)
罫纂請の筆意‖へ豊喜望房寝藍重賞塩素電駕篭a婆露
蹄∴∴的∴∴舵
慨 0 〇 .
偶 的0 細0 000 餌場 的0 0
餌場弼徳で馨珊露eratり鴫(OC)
FIG" X2・1 Emissivity uncertain置y Contribu置ion Io lbtal Uncertairty
X2.6 Atmospheric Absorption Uncertainty x2・6.3 App】y山ese血a 'o Eq Xユ14. The solution is shown
duc to atmosphc「ic absorption. An 8 to ]4いm i11fra「cd X2.6.4 Tbble X2・2 shows∴Caiculated unccrtainty due to
thermomelcr is calibraled using a 200OC (473,15 K) sollrCC. atmospheric ahsorp'ion k-r the 8 to 14 um spectral band.
Using Eq X2工S = 0.0461160, and Ijq 3"1.13∂SI∂T =
0.000305968 K-1.
、言寄 電場 ㌔′雪′ヽき主
山田o聞けo則O耽れ惟†ROしo6鵬
∴措置密議Y′ゞ FYH S.A.S.
.高車国事〇 cop書A
・〇・をや薄手∴ 串や
cO軸丁韓○○▲鵬
鍋か 巨2847 - 14
さ的請的細腕慣れの青筋農のま駒鴨屡競副調印i職長鵬彊S緬勘
Q嬉
帥丁巌微弛れ総瞭蚤涌重電了a強め篇穂のC,鑑幕億農場部
露)重富霜鍵董重量黒蜜竜巻雲量雷電書
(′
傭D 与
4つ つJ つふ
0 糊0 午的 馬的 8鋤 錐的
So舶耽墓丁g競pき胎軸隠‡oC)
FiG. X2,2 Reflected Temper露tu「e Unce「書ainty Cont「ibution to “lblaI Uncertainty
つ◆二㌔基準鋤京飢餓鎚手
鵬o議了o則○ ○調印的し0融
管Y録S.A.S置
膏,心.こ高Y胃
coplA
A主宰恵む」} cO軸でR○○A田A
4齢 宣2847 - 14
Ⅹ3.1 A poinl of di組cu宣ty血Oughout in delermining tem- ambien=emperalure do not need to be known。 A11 1hat is
Peratllre unCertai】1ty is tha=he value o」’Td is not k】10Wn tO the rcquired is tha=hese values a.rc constant during thc ‘wo
user. It is measured intemally by the instrument, but is not measurements. The deteclo「 1empcrature, Td, Can be calculatcd
displayed on lhe readout. The bcst assumption is that T,l Will bc 血on1 1hc Eq Xう」
the Same aS, Or C】ose to, ambient temperatl-re, and this may
・碑。β置S(11M榔,) 〇 ・1村部2S(1「鵬1音,2)
require some conditioning of the instrument before use. S(乃)二
〔値湘1  ̄ 〔]Nし卵職2 ●
Howcver・ there aLrC SOme Situations whcrc ensuring that Td =
is used for an extended period in front of a hot source (for Z完劇Sl - 7L,劇S2 = the two jl南ared themo】11eter 】・ead-
ex.anlp]e, during calibration at high temperatures), Where the ]ngS COrreSPOnding to 8INSTRl and
radiation fro重n the source may hea=he detector above ambie11t CINS丁R2, l“e叩eCtive】y, and
temperature. The sccond is wherc血e infrared lhemometcr is f高汀Rl and cINS7R2 = the emissivily seltings used
defined temperature, uSually above OOC. X3.5 F。r eXan叩Ie, tWO readings nf a b】ackb。dy using an 8
to 14 Hm thcmomctcr might be Tmふl = 14l.80C when thc
X3.2 Dependence on the detector temperature is remOVed if
inslrumenlal emissivily is <e=0 E-n、… = 1・ al-d Tmca,2 =
しhc instrumenlal emissivity is se=o l (1his js precludcd in fixed
219.4OC when cうnQtr2 = 0.5. Substiluling these values into Eq
Cmjssivity instruments). On thc othcr ha-重d, reiative]y hrge
X3.1 gives Td = 2l.50C.
errors in the estimate (guess) of the detector temperatlJre Can
usua】1y be toIerated without signjficant error in the reading. X3・6 Calculatio】lS Ol’1his sort a「e useful for delemining 'he
See賞了ig. X4.1
議薄鵡鵠青 鵬o鵬動議10唯州打的し06山
魚.筈持せま 下Y岬S.A.S.
雷電軍務奉 CDpiA
暗∴∵叩擬で柵馳種 cO競丁RO」A寄▲
句齢 各2847 - 14
XY王鏡駒闘簸押出書記頑o町
1之3今的副職Sぬ
き触S栂瞳織約号-烏789
竜!儀phc鵬; (霜雪)与与5-曇ら与写輪職こく粥5)与野づ縮丘
鼠蓋PQ餌の戸的患醐調O同声○減
車れr幽門沌i下れ駁剛寄珊成合『
鞠欝鴨N山部b散・: 187勘之宣裏
叩廟轍関る
翻b鵬賄晦d露y:
し圃間的A購U融恥駅裏NG
曾87軸艇銃。
館的S船艇之勤務・7890
臨調鵬l 郭博豊 胸創れ9ofUげ「 鵬a那加鶏 A燐制限 釦母鵬再説亀nd 劇胸囲靴
剛的緋鵬 鴫明細鵬 丁轟き「醐鵬r 静脚臓 脚 d軸触「 鄭鳩調e
四囲回国
(。C) (eC) (均 何問)
^U
0 鵬1 柵0青ヱ 紗5 討o 軸トp麻惚A搬m調 覚煽請負轍
O 一エ1∴一つふ
鮒 9諦8 100.轟 きア軍 駒 的叫劃禽技霊芝O同調 紺鍋肌魯車重
4糊 4i9.9写 俄0,8 27章 的 珊珂a胎8章糊爪的 鏡鵬湘低】重
線的 的虹エ ア的、5 6脚 鳴き き綱母A駒糊朋 剥き紬蛮
丁極細離れ鵬の曲Y合飴鵬のf細さ職印轟き脚y 〇両Yめく鵬鵬州$揮瞬G副中細渦的聞合的印面
丁船繕請p鍾研鎧函鵬画臨場的轟き胎t鵬きd都画壇轍鳩I撮曲融か副職競陛融鵬§錆i色癌書鈎0くれ竜一秘・
誤認幣苫議謹豊謹諒灘認諾謙語言語薬譜濡譜諜誌
面si調9節剛ゆA鵬朗a章馨鵬咋地軸鵬で0n書′0爛ap軸物嶋面章寄る細朋蝕汀が紬同調魯寝間弾雨「e d軸賄e郎200朋部的珊櫨滑
inf昭「錆湘亀「mo和償餅
肌儀鴫血吋疏at割れeのく;丁場的部隊ned欽叩d功蟻e岡部y庇暮し鵬§章塵鍬1帰陣舶U脳細胞i母胎的樹卸的紬競ぬ鴫的
豊嶋鱒浦疎め購総圃n義時予鵬鵬剃肥m鍋p聞醗両統葉蘭訓臆測柵削W織雑う一両e釣棚軸聡d e棚融曲聞紺削噂
蕊欝諜露盤糀盤謹謙語露語盤蓄盤菌態嵩嵩謹豊諾謙器露盤藍嵩鵬
00n餌s慮w軸鑓曲別協調健出師章的γ版e調製汁血面喝坤p随a駒n.
↑旧説a調印地場剛削章詳鴫顔照れぬ「d富の輔は嶺購b棺的請わぬ掲めける晦甑執隊褐帥箆的請容寄約〇胴冊箆同的y着繭的鴨Of
し脚部e南山a露d純的直〇千a co聞p購面e彊s睦pro轡輪同船手e醐車的雌湘軍i胴ed a削職印弧d調餓臨u「帥融抽蛾bi噂博細れ鵬
艇唯) o同m豊胸吋噂鴻鵬的両誌ぬb競艇能動単軸e Iab舶捌y漉掴踊脱ぐ鵬青鵬曲加的亀恐く睦u噂Ue記聞膿調的調捷「 to臆鵬曲
暗e賭場縄dれ智n糊剛r醐e融煽鞘説も噂か請は請は棚関数甜i蛮肌輔徴自動鵬推叩傭〇時下鴨的C読め噂〇年間割に両船A帥d鞠b
両地8 i8 「adio鵬e鵬(鱈丁袖W櫓7烏6写敦刑場鵬II)鵬鳴らn 81掴聞同繭敦細欺ぬrd" me鑓るC鎚bi時O手簡時A結∞舶破
く購丁的間職穐もらS癌合議e裏).
F漢G. X4.1 SampIe Report for an lnf「ared Thermometer
申子=‥.当主∴一 LABO鵬TO則O唯物印南OLO611
.◆ “● 〇〇 〇●●
.六十・:閏Y弓 管Y岬S.A.S.
子∵・こ待つ COpIA
玉髄宣言で・接す嬉きつ篇 cON丁京oしAロム
句齢 巨2847 - 14
諜誌驚鰐讃認諾誤認謀議驚謹叢驚議案
p畦湖d食ら職色緑藻龍ve朗教is§iv時宙i ・0請暗面職9削W欠き76銘・
欝調所的調蘭留調を劃事o請創髄の鵬紬購職場倣Cal肘陶龍馴販
推那肺腑馳階数,2eC蝕献納胎的同相時掬%
たく叩きp調e競紬d §掛れ鳴雨読
憎韓機軸風儀機能「同職寄劇 尊徳を諦咽の機 鴨紬的間隔o調勘鴫e
軸C S癒合随瓶 8線OA 鞄を閥浪隠朗at脚Sou「馳(軸年Pl轡te A) 宣ま1171之0拙
館C S離隔餅陀 8主禎A 抽意Pla章色鼠壷d治縄0れSour龍(酷くPlate B) 句26!2鎚書
薄手寄れ 鵜的 噛綱や鴎的軸S的場(亡av時Å) 的A
鞠舶 親之00 V呼陣軸色「爪o鈎岬は 裏01ま0I20棚
すね色事珊a‡X∴∴∴∴騨丁D 800 剛勘 卸221之鎚O
p撮ced曲調巨細雷電寄雪XC一概賂龍練熟那3重ね章ed訓陥り1の,抑1僻
の穏健腿統帥や聞1曾調創機能聡的白州y l.和宣0
D遍くe o青職場榔競登録印種陶をion豆u吋3,郭10
Dき暁o章c暮駒陶勧o動きJ函之.之0抑
D鵬da船p餅鎗膳的強eでき職叩魯兜重し母方郭豊0
勤馳融毎肌p餅ね踊eき寄yこJ宙洞A, S調駒,心S軸隔鴫鵜幽c黒自陣藩閥留節
丁l「出血調(心轡調費「きp鴨巾か山中両面働録鴫榊雷む曲調的∩し重責富強か購丁電工正己而占革細さ重賞重く相田)タロ
誌篭悪罵欝鵠認轟音轟触鵬岬鵬。鴫。鮎液胞細山轟。暮町
田G, X4,1 Sample Report fo「 an lnfrared Thermomete「 /虞mfroり
REFE RENCES
( l) 】EC/TS 62492工看ndustrial p「ocess control devices -Radiation瓜er- M., Bioembergen. P∴U調m:血中h〃電CIs ror calibration of radialion
mometers -Pa「t ] : Ttch11ical data for radiation t置漢ermOmetel’S, IEC, themometers bek’W the sil'Cr町im∴ OC’トWG5 working document
GeileVa. Switzerland, 2008. Aval】ab1e什om Inlernatjona】 Electrotech- CC丁WG508」13. BiPM. SC巾眼L I+劃軸i May 2In8. Availablc rrom
11ica獲Comn〕ission (IEC), 3, ruC de Varemb6, P.O. Box 131, CH-12】1 the B!PM webche. h叫訓を胃管-鵬Im-org.
GcllCVa 20. Switzer】and, http://www.iec.ch.
15) Saunders, P, MSL ’I轟hnical G面責七22 -Calibratjon of Low耽mpera-
(2) DcWitt, D. P, and Nutter, G. D.. eds.. “Theory and Practice of ture Jnfra∫ed ll噛mDiI_ MeaLSure重nenl Standards Laboratory of
Radiatioll Thermometry,’’Johll Wi!ey and Sons. New Yol’k, NY, 1988.
New Zea!and. 2(東部. hnpJ/msi.iri.c「i.nzltraining-and-reSOurCCS/
(到VDI 3511 Blatt 4.4 / Part 4・4, Tt:Chnische恥mpcraturemcssung
tcchnicaトguides,
Strahlungsthermometric Kalibrie則ng VOn StrahlungsthermOmCtem,
′]tml)erature meaSuremCnt in indしIStry Radiation thermometl’y Cali- (6) E‘;al〃a存on (Z/ Wc瓜録がIC"I DaIa臆G話de柄所c E準rc."ioIt∴0/
[朽ceIでく肌1申n ”ea∫llmれC爪JCGM 1 00, BIPM, S色vres, France, May
b「atioil Of radia亡ion thc「moine書e「S.
2008. Avajlable from the BIPM websjte, http’〃www.bipm.Org.
(4) Fischer,工Saundel’S, P, Sadli, M.. Battue11o. M., Park, C. W.. Yuan.
(7) U. S. Guide lo thc Expres§ion of Uncertainty in Measurement,
Z., Yoon, H., Li, W., Van der Ham, E., Sakuma, F., Yamnda, Y.,
Nationa重Conference of S書andards Laboratories. Boulder, CO, 1997.
Ba皿co. M.. Machin, G,, Fox, N.. HollandしJ., Ugur. S,, Matveyev.
貰彊帥鵬†舶i苑鋤 Å的附0刷0耽陸†帥しOe山
戸Y岬S.A.も.
き譲を琵魅Y孝
鼓甲即鵡嚢 C①P書Å
も」読離郷丁場蹄∴ co能で駐韓」A観▲
鴫野 巨2847 - 14
AS7勅Wemat/Onal lakes r]O POS柄bn伯apec面g the va胸Iy of any patent /匂hts asse万ed frうcomeCめn面伽any /tem menfわned
ln肋is sfandかd│ Users o柵is slanda′d are創press/y advised肋al dele仰ha的n o欄e va佃的y oI any such patenr /ゆts, and lhe高ek
Of /nhJngemenI oI such /ゆh吟a伯en折el画heir own伯sponsめ偽りt
77)is sねn佃rd is su切ecl lo rewsわn創any I緬e by Ihe伯印onsib/e lech面oaI comm鮎ee and musI be胎viewed ever画ve yea/S a∩d
存noI revISeα e肋er reappmved or w鮪drawr). ¥匂ur commen!s are /nV/led e肋er b「 rews/On Of肌js sIan〔ぬrd or foradt%na/ SIandaIds
and shouId be addressed lo AS7M /nIemafronaI Hea的uaners. Ybur commenls w,〃 rece’Ve careん/ COnS′dera面on at a meeI’ng O予めe
re印ons/ble technica/ COmm/撮ee. wh/Ch yOL/ may細end /i you /eeI m匂りのur COmmen(S have not IeOe/Ved a fair heanng J,Ou Shou/d
make your vIeWS known /O勅e ASTM Comm祐ee on SlanくねIds, a=he address show" be/OW
77)IS SねndaId /S OOpy[ゆhted by AS7勅/nIemalionaI JOO Ba〃 H自めor Dhve, PO Box O700, l侮S書Conshohocken, /24 J9428-2959,
Un/ted S削es. /ndIV/dua/伯p面ts作yng/e or muIlp/e copie?) oI mis sねndard may be obl観ined by conぬo面g AS7M a川he above
address or a1 610-832-9585 (bhone), 610-832-9555伯母Or ser而e@aslm.oIg (e-m負肌Or伽ough肋e AS7M websife
(WW肌aS面O働L Pe〃面ssbn 〃ghts to photocapy蛤e sぬnd加d m8y a/SO be secu′ed舟()m the Capy"ghI CIea個nce CenIe' 222
仔osewood D万ve, DanveIS. MA O 1923′ T軌@7切G46-2600:団p伽w脇copy〃ghr.com/
山容o議同軸o耽肝†押し的iA
胃Y岬S.
coPl
○○軸丁Re.. ∴● .へ