SIMPLY HP

Sect.4 - Page 1 OF 9
Doc.000539
SIMPLY HP

4. ADJUSTMENTS
4.1 INTRODUCTION
The radiological system is entirely regulated in the factory: it is therefore only necessary to
install it to be able to use it.
All the information contained in this SECTION 5 and in SECTION 6 should be used when an Extraordinary
Maintenance operation becomes necessary.

ATTENTION – In the case of intervention, it is essential for the adjustment


! operations to be carried out in the sequence indicated.

Table 5.1 indicates the Potentiometers which do not require any further
adjustments since they have already been adjusted and sealed in the factory.

TABLE 4.1

CARD POT. DESCRIPTION


S15 PT2 Manual setting of the Filament Current
S 17 (MONOBLOC) P1
Return of the kV +
P2
P3
Return of the kV -
P4
S 19 P1 Maximum current
P2 Frequency setting
IB02 PT1 Maximum exposure time anticlockwise.

IB04 PT2 Maximum Battery Voltage

PT3 Micro Battery Reading setting

Sect. 4 - Page 2 OF 9
Doc.000539
SIMPLY HP
4.2 SET-UP PROCEDURE
In the following parts of this chapter, reference is made to the SET-UP PROCEDURE of the system.
To access this mode, only work from the UNIT console.
Selection of Set-up is made when the system is turned on, as illustrated below.

• Turn the Unit on by pressing the ON and mAs decrease keys at the same time .
• The Set-up screen will appear. Press the Set-up Symbol
• Now enter password 1441 on the alphanumerical keyboard which will appear and you will
enter the General Set-up screen .

X-RAY CONTROL NOT ENABLED


80 KV 400mA 80 KV 200mA SET 40 KV

1470 1330 416


kV Adjustment
mA Adjustment 80 KV 100mA 80 KV 50mA SET 120 KV

1280 1180 1202


50 KV 400mA 50 KV 200mA mAs adj

1545 1405 1202

4.3 KV ADJUSTMENTS
Ser kV ADJ to40 kV
The SET kV voltage is supplied by the microprocessor and must be adjusted
according to the ratio kv/20 ( 1V=20 kV). 2,00 V
• Press the “Set 40 kV” field. + -

• Connect the probes of a digital multimeter in Vdc on the S 19/B card CP1-10 CP1-2

between CP1-2(-) and CP1-10(+); Set kV ADJ to 110 kV


• Using the mAs increase and decrease keys, adjust the value read on
6,00 V
the digital multimeter until you obtain a value of 2Vdc.
• + -
Press the “Set 120 kV” field.
CP1-10 CP1-2
• Connect the probes of a digital multimeter in Vdc on the S 19/B card
between CP1-2(-) and CP1-10(+);
• Using the mAs increase and decrease keys, adjust the value read on
the digital multimeter until you obtain a value of 6Vdc. mAs

Sect.4 - Page 3 OF 9
Doc.000539
SIMPLY HP

4.4 MAXIMUM FILAMENT CURRENT ADJUSTMENT


4.4.1 FILAMENT CURRENT ADJUSTMENT

ATTENTION !
By FILAMENT CURRENT, the current passing through the primary
winding of the power supply transformer of the filament itself is meant
(COOLIDGE TRANSFORMER).

A) MEASUREMENTS

To measure the FILAMENT CURRENT, connect the probes of a digital multimeter in AmA
in series between S15-CP1/2
S15 CP1/2 (+) and the green wire of the electric cable of filament.
To measure the anode mA in RADIOGRAPHY, connect the probe of an oscilloscope
between test points TP6(+) and TP11(-
TP11(-) of the S 19/B card. Correspondence in
reading is 1V=50mA, whereas between test points TP3(+) and TP11(-TP11(-) of the S 19/B
card to display the kV wave form. Correspondence in reading is 1V=20kV ;

B) ADJUSTMENT OF THE WIDE FOCUS CURRENT


Activate the SET UP procedure;
Select “50 KV 400mA” ;
Adjust the set value of the “D DAC” by means of the mAs selection keys until you
obtain a filament current,
current displayed on the digital voltmeter, which is the same as
the one indicated in table 4.4.

(E.g. for 50 kV - 400 mA - Typical DAC 1465)


1V=50mA
Command X-Ray and check that the anode mA are 400;
400 if necessary re-adjust the SET
-
UP until you obtain the desired value; +

TP6 TP11

Select “80 KV 100mA”. 1V=20kV

Adjust the set value of the “DDAC”, again by means of the mAs selection keys, until + -

you obtain a filament current,


current displayed on the digital voltmeter, which is the same TP3 TP11

as the one indicated in table 5.4.

(E.g. for 80 kV - 100 mA - Typical DAC 1280)


mAs
Command X-Ray and check that the anode mA are 100 mA;
mA if necessary re-adjust the
SET UP until you obtain the desired value;

Select “80 KV 400mA”


Adjust the set value of the “D
DAC”, again by means of the mAs selection keys until you
obtain a filament current,
current displayed on the digital voltmeter, which is the same as
the one indicated in table 5.4.

(E.g. for 80 kV - 400 mA - Typical DAC 1545).


Command X-Ray and check that the anode mA are 400 mA.
mA If necessary re-adjust the
SET UP until you obtain the desired value.

Sect. 4 - Page 4 OF 9
Doc.000539
SIMPLY HP
C) ADJUSTMENT OF THE NARROW FOCUS CURRENT
Activate the SET UP procedure ;
Select “40 KV 125mA ”
Adjust the set value of the “D DAC” by means of the mAs selection keys until you
obtain a filament current,
current displayed on the digital voltmeter, which is the same as
the one indicated in table 5.4.

(E.g. for 40 kV - 125 mA - Typical DAC 1295)

Command X-Ray and check that the anode mA are 125; 125 if necessary re-adjust the SET
UP until you obtain the desired value ;
Select “80 KV 50mA”.
Adjust the set value of the “D
DAC”, again by means of the mAs selection keys until you
obtain a filament current,
current displayed on the digital voltmeter, which is the same as
the one indicated in table 5.4.
1V=50mA

+ -
(E.g. for 80 kV - 50 mA - Typical DAC 1170 ) TP6 TP11

Command X-Ray and check that the anode mA are 50 mA; if necessary re-adjust the
SET UP until you obtain the desired value; 1V=20kV
Select “80 KV 125mA” -
+
Adjust the set value of the “D DAC” by means of the mAs selection keys until you TP3 TP11
obtain a filament current,
current displayed on the digital voltmeter, which is the same as
the one indicated in table 5.4.

(E.g. for 80 kV - 125 mA - Typical DAC 1295)

Command X-Ray and check that the anode mA are 125;


125 if necessary re-adjust the
SET UP until you obtain the desired value .

Table 4.4 Dac Range


Adjustments Settings Minimum Maximum Sample
40 kV 300 600 415
Set kV
120 kV 1000 1500 1240
40kv 400 mA 800 1800 1465
Set mA WF 32 KW 80kv 100 mA 700 1600 1280
80kv 400 mA 600 1700 1545
40kv 125 mA 700 1700 1405
Set mA NF 11 KW 80kv 50 mA 600 1700 1180
80kv 125 mA 500 1700 1330
Set mAs 20 mAs 400 600 500

Sect.4 - Page 5 OF 9
Doc.000539
SIMPLY HP

X-RAY CONTROL NOT ENABLED


80 KV 400mA 80 KV 125mA SET 40 KV

1470 1270 416


80 KV 100mA 80 KV 50mA SET 120 KV

1280 1170 1202


50 KV 400mA 50 KV 125mA mAs adj

1545 1295 1202

Can 3 1V=100 mA

Can 1 1V=20 kV

OSCILLOSC0PES

FIGURE
IGURE 4.4.1 - ADJUSTMENT OF NF CURRENT 40KV 400MA

Sect. 4 - Page 6 OF 9
Doc.000539
SIMPLY HP

4.5 MAS ADJUSTMENT

A) MEASUREMENT
Remove the bridge between terminals CP3-
CP3-1 and CP3-
CP3-3 of the S19/Bcard.
Connect a mAs meter between the above terminals.

B) PROCEDURE
50mAs

Activate the SET UP procedure;


Select “MAS ADJ”; CP3-1 CP3-3

Command X-Ray;
On the instrument, check that the mAs value is really the one set (50
50 mAs – Typ. DAC 0480);
0480
if necessary re-adjust the SET UP until you obtain the desired value;
exit SET UP. mAs

4.6 READING BATTERY VOLTAGE


Activate the Set Up procedure, the battery voltage value read by the microprocessor is displayed on the
main screen.
screen

A) MEASUREMENTS
To measure the true voltage in the Battery,
Battery connect a digital Multimeter with RMS reading
between the XT2-C(+) XT2-C(-) of the IB04 card.

ATTENTION !
350 Vdc voltage present – take the necessary precautions!

Check that the maximum charge value does not exceed 355 Vdc.
Should it exceed this value, contact the technical service.

Sect.4 - Page 7 OF 9
Doc.000539
SIMPLY HP
4.7 CHECKING RADIOGRAPHIC PARAMETERS
After having completed the operations indicated in the previous paragraphs, it is
necessary to check the values of the proportional signals at the kV and mA which arrive
at the INVERTER. The kV signal tolerance is ± 5%, whereas for the mA it is ± 15%.
15%

ATTENTION!
If the kV value is incorrect, it is necessary to repeat
the SET kV adjustment.

ATTENTION!
For the following checks, an oscilloscope with two channels is required.

4.7.1 CHECKING KV and MA IN RADIOGRAPHY


Connect the probe of the oscilloscope as indicated below:

• CHANNEL A:
A TP3(+) - TP11(-
TP11(-) of the S 19/B card (kV
kV AT THE TUBE);
• TIME BASE: 20 ms
Ch. A
If necessary, connect the probe of the oscilloscope as follows : + -

TP3 TP11

• CHANNEL B:
B TP6(+) - TP11(-
TP11(-) of the S 19/B card (mA) ;
• TIME BASE: 20 ms Ch. B
+ -

Set the minimum value of the kV (50 mA ;


50 kV – 400 mA) TP6 TP11

Command X-Ray and check that the values read on the oscilloscope correspond with
those given in TABLE 5.6.1;
5.6.1
Repeat the previous operations setting the kV values at 80 kV and 100 kV and 120kV.
120kV

TABLE 4.7.1 WIDE FOCUS RADIOGRAPHIC PARAMETERS 32 KW


kV mAs mA Tempi (s)
Range of acceptance
Range of acceptance
Set S 19: TP3 - TP11 Set Theo-
eo-retical Theo-
heo-retical
(+/- 10%)
[1V = 20 kV]
40 38 ÷ 42 20 17.5 ÷ 22.5 400 0.50
80 76 ÷ 84 20 17.5 ÷ 22.5 400 0.50
100 96÷104 20 17.5 ÷ 22.5 320 0.62
120 96÷104 20 17.5 ÷ 22.5 200 0.100

TABLE 4.7.1 NARROW FOCUS RADIOGRAPHIC PARAMETERS 11 KW


kV mAs mA Tempi (s)
Range of acceptance
Range of acceptance
Set S 19: TP3 - TP11 Set Theo-
eo-retical Theo-
eo-retical
(+/- 10%)
[1V = 20 kV]
40 38 ÷ 42 20 17.5 ÷ 22.5 200 0,100
80 76 ÷ 84 20 17.5 ÷ 22.5 100 0,200
100 96÷104 20 17.5 ÷ 22.5 80 0,250
120 96÷104 20 17.5 ÷ 22.5 63 0,315

Sect. 4 - Page 8 OF 9
Doc.000539
SIMPLY HP

4.8 COLLIMATOR CENTRING


For any centring of the Collimator, please refer to the TECHNICAL MANUAL of the Collimator, enclosed
herewith under SECTION 8 BASIC COMPONENT DOCUMENTATION.