Thermo Scientific Dionex Chromeleon: Annex To Operating Instructions
Thermo Scientific Dionex Chromeleon: Annex To Operating Instructions
Chromeleon
Operational Qualification/
Performance Qualification
for
Ultimate 3000 RSLCnano
Nano, Cap and MIC Systems
Annex to
Operating Instructions
Version 1.00
Revision: A
Date: May 2017
Doc.: Annex-A_RSLCnano_OQ_PQ_OI_V1_00.docx
Annex to OQ and PQ Operating Instructions
Contents
1 How to Use This Manual ................................................................................... 3
2 Process .............................................................................................................. 3
2.1 Required Materials .......................................................................................................................3
2.2 Performing the Checks ................................................................................................................3
3 Procedures ........................................................................................................ 4
3.1 Support of Thermo Scientific Dionex UltiMate 3000 RSLCnano ProFlow Flow Meter ..........4
3.2 Thermo Scientific Dionex UltiMate VWD-3x00 detector - Linearity with MIC flow cell .........4
3.2.1 Performing the Check .............................................................................................................4
2 Process
2.1 Required Materials
Part number 6720.0360 includes the RSLCnano OQ/PQ operating instructions, set of 8 caffeine
samples, 1 µL sample loop with nanoViper connections and connection tubing.
The parts (depending on the configuration) necessary to perform the OQ/PQ procedure, which are not
included in the OQ/PQ kit, are described in the RSLCnano OQ/PQ operating instructions.
The sample positions used in the sequence templates were changed to match with the ampoules in
the standards kit: The content of the 10 µg/mL standard ampoule has to be filled in only one vial that is
TM TM
placed in Thermo Scientific UltiMate WPS autosampler position RA4 (instead of RA4 and RB2).
Additionally, the content of the 40 µg/mL and the 60 µg/mL standard ampoules have to be filled in vials
and placed at the WPS autosampler positions RA6 and RA7 respectively. Thus, the following sample
positions are used with the new templates (differences to the operating instruction are marked bold):
Table 2-1 Sample Positions in the WPS Autosampler Rack
3 Procedures
3.1 Support of Thermo Scientific UltiMate 3000 RSLCnano ProFlow
Flow Meter
TM TM
Thermo Scientific Dionex Chromeleon RSLCnano Nan/Cap/Mic OQ/PQ supports the UltiMate
TM
RSLCnano ProFlow flow meter (since version 4.4 distributed with CM 6.80 SR15b).
The OQ/PQ is performed and evaluated against the specifications for the NAN (nano) configuration as
described in the RSLCnano OQ/PQ operating instructions, version 1.0.