HF Series Generators

Configuration

SECTION 4 X-RAY TUBE DATA

The following table lists several common X-ray tubes and their corresponding
number. If a specific tube is not listed, tube specifications are given to enable
you chose a similar tube type. If none of the listed tubes are satisfactory, contact
your generator supplier to obtain special software.

Table 4-1
X-ray Tube Numbers

POWER RATINGS
TUBE TUBE CODE FOCAL
MODEL KHU
NUMBER (ID) SPOT LS (kW) HS (kW)

001 046 EUREKA RAD 8 1.0 / 2.0 24 / 51 -- 150

002 034 EUREKA RAD 14 0.3 / 1.2 4 / 43 7 / 73 286

003 036 EUREKA RAD 14 0.6 / 1.2 16 / 43 27 / 73 286

004 037 EUREKA RAD 16 1.0 / 2.0 34 / 72 52 / 110 300

005 038 EUREKA RAD 21 0.3 / 1.2 5 / 65 9 / 110 300

006 040 EUREKA RAD 21 0.6 / 1.2 21 / 65 36 / 98 300

007 043 EUREKA RAD 60 0.3 / 1.2 6 / 60 10 / 98 400

008 044 EUREKA RAD 60 0.6 / 1.2 21 / 59 34 / 94 400

009 045 EUREKA RAD 74 0.6 / 1.5 18 / 49 -- 200

010 047 EUREKA RAD 92 0.6 / 1.2 26 / 61 40 / 90 600

011 023 COMET XSTAR 8 1.0 / 2.0 22 / 42 -- 150

012 021 COMET XSTAR 14 0.6 / 1.2 16 / 44 27 / 74 300

013 022 COMET XSTAR 74 0.6 / 1.5 18 / 51 -- 200

014 132 COMET XSTAR WHIS 0.3 / 1.0 4 / 39 -- 140

015 013 COMET DI-7 1.0 / 2.0 22 / 46 -- 140

016 007 COMET DI-101 0.6 / 1.0 23 / 50 38 / 82 300

017 006 COMET DI-101 0.6 / 1.3 23 / 71 39 / 105 300

018 051 GE-CGR MN 640 1.0 / 1.8 23 / 46 -- 150

019 055 GE-CGR MSN 742 0.6 / 1.2 11 / 30 21 / 57 150

020 062 GE MAXIRAY-100 0.6 / 1.25 18 / 55 31 / 100 350

021 064 GE MAXIRAY-75 0.6 / 1.5 12 / 37 21 / 62 300

022 092 VARIAN A-192 0.6 / 1.2 25 / 63 40 / 96 300

023 094 VARIAN A-292 0.6 / 1.2 25 / 62 39 / 96 400

024 079 PHILIPS SRO-25-50 0.6 / 1.0 15 / 32 27 / 55 314

025 077 PHILIPS RO-17-50 0.6 / 1.3 19 / 55 -- 325

026 294 VARIAN A277 0.6 / 1.0 31 / 59 48 / 88 400

027

Note .-- Power Ratings are for 60 Hz. To calculate Power Ratings for 50 Hz multiply the values by 0.91