Adjustment: November 2012 XP Series S/M
Adjustment: November 2012 XP Series S/M
Adjustment: November 2012 XP Series S/M
(3) Press the Start Switch and verify WBC and RBC background counts fall into the ranges below:
WBC background value 0.30 x 103/μL
RBC background value 0.30 x 104/μL
When the Gain Adjustment screen appears, W-MFV target box will be reversed. Enter the target
value.
(11) Close the Detector Cover, and press the Start Switch.
(12) The recount analysis sequence will be performed automatically three times and the three analysis
values (1 to 3), difference between the maximum and minimum (MAX-MIN), and compensation
ratio (RATIO%) will be calculated. (For approximately 2 minutes.)
(3) Enter the Service mode. (Enter C, 9, -, 0 into the sample number input box.) Hot start is also can be
enter the service mode.)
(4) Press the Start Switch and verify PLT background counts fall into the ranges below.
PLT background value ≤ 0.30 x 104/μL
201297
(2) When below screen appears, Adjust the VR6 on the PCB No. 2135 (KX-21) so that the WBC clog
level falls within the following range.
WBC Clog value:100.0±1.0
The system returns to the Ready screen in 30 seconds. If the adjustment has not been completed
when the system returns to Ready status, select WBC again.
(2) LED
LED Item Description
D1 POWER ON: Power is supplied.
OFF: Power is not supplied.
D6 RUN ON: CPUunder CPU operation.
OFF: Under CPU Reset Process
D7 FL BSY ON: under Writing on Flash Memory/ Deletion
OFF: Normal Operation.
D8 B START Flashing: Outside bus cycle is operating.
ON/OFF: Inside unit access or HALT
D11 LINK ON: Connected to network.
OFF: Not connected to network.
D12 RD Flashing: Packet being received.
OFF: Packet not being received.
D13 TD Flashing: Packet being sent.
OFF: Packet not being sent.
TP Item TP Description
TP1 5V (5V for logic) TP7 PGND (GND for 8V)
TP2 G12 (GND for 12V) TP8
1.9V (1.9V for logic)
TP3 GND (GND for logic) TP9 I-RST (RESET input terminal for ICE)
TP4 12V (12V Power Input) TP10 33MHz (bus clock)
TP5 3.3V (3.3V for logic) TP11 1Hz (terminal for calendar operation
check)
TP6 (8V for embedded printer)
Note) Adjustment volume for LCD contrast is located at the bottom of Panel Keyboard.
(6) Verify that this drill bit can penetrate through SRV (3 pieces: SRV Fixed Valve (L), Rotor Valve and
SRV Fixed Valve (R)) at the SRV alignment hole.
(5) Verify the pipette tip is placed in the gap "A". (View from "B" is easy to verify.)(View from "B" is easy
to verify.)