7-1 Maintenance Part Before cleaning the air filter of the controller, be sure
CAUTION to turn off the power. Mistakenly touching high voltage
Be sure that an engineer with specialized electrical areas may cause an electric shock.
WARNING
knowledge performs the maintenance described in
this chapter. Mistakenly touching high voltage areas
may cause an electric shock. If the MD-X1000/1500 Series is used with a dirty air
filter, the temperature inside the marking unit rises,
weakening the laser power or causing malfunction.
The table below lists the replacement parts for the MD-X1000/1500 NOTICE
Series.
After removing the air filter, never insert metal chips
or foreign material into the ventilation port. Doing so
may lead to a breakdown of the unit.
Part name Recommended replacement timing
MD-X air filter Replace or clean the air filter when dust or dirt
(OP-87888) has built up on the filter surface. 1. Turn the key-operated switch of the controller to the "OFF" position
Replace the air filter if torn or damaged. to turn off the power.
MD-X marking Replace the fan of the marking unit when it
unit fan stops operation.
(OP-87889) 2. Loosen the eight screws on the upper and lower front panels.
Time-lag fuse Replace the fuse when blown.
Rating : 250 V, 10 A, Time-lag fuse
Recommended fuse :
0218010.MXP,Littelfuse,Inc.
Fuse holder
Maintenance
3. Remove the filter holder and the filter from the panel.
Be sure to use replacement parts specified by
KEYENCE. Failure to use the parts specified by
KEYENCE can result in damage to the unit. Filter holder
NOTICE
Clean the filter periodically. If the filter is clogged,
the internal temperature rises, causing the
MD-X1000/1500 Series to malfunction.
Air filter
7-2 Maintenance
When the window surface of the marking unit becomes dirty, the laser
beam transmittance decreases, resulting in faint marking or the
absence of marking.
Periodically wipe the window gently with a cloth dampened with
acetone or ethanol to remove dirt.
Window
Handle the air filter carefully to prevent it from being
NOTICE damaged.
Never allow it to dry in direct sunlight.
4. Remove the air filter. Clean it with neutral detergent, and allow it to
air dry in the shade.
5. Reverse the above procedures (Steps 1-3) to attach the dried air
filter to the controller.
2. Remove the two screws (M4 x 8) securing the fan connector cover
and the four screws (M4 x 35) securing the fan at the rear of the
marking unit and remove the following:
Fan connector cover
Fan guard
Marking unit
Fan
Fan connector
Fan guard
Fan connector
cover
3. Disconnect the fan connector and remove the fan from the marking
unit.
NOTICE
Tighten the screws with a tightening torque of 0.8
Nm.
CAUTION
Do not forget to attach the fan guard. Otherwise, the
rotating fan may cause injury.