Drager-Cato-Manual Service
Drager-Cato-Manual Service
Drager-Cato-Manual Service
Anesthesia Ventilator
Technical
Documentation
I
0
8050
PM
er
Dr g
D
Copyright by Dräger Medical AG & Co. KGaA, Lübeck, Federal Republic of Germany
The warranty and liability conditions of the general terms and conditions for business transac-
tions of Dräger Medical AG & Co. KGaA are not extended by this technical documentation.
Safety Regulations:
Reference is hereby made to the observance of the relevant safety provisions, such as the
Medical Equipment Ordinance (Medizingeräteverordnung), the Pressure Container Ordi-
nance (Druckbehälterverordnung), the Technical Rules for Pressurised Gases (Technische
Regeln Druckgase) or the Occupational Health and Safety Provisions (Unfallverhütungsvor-
schriften).
Insofar as reference is made to laws, regulations or standards, these are based on the legal
system of the Federal Republic of Germany.
General
1 Intended Use 8
2 Technical Data 19
Functional Description
1 Basic Unit 28
7 Anaesthesia ventilator 57
8 PM 8050 cd 103
9.1 Messages in the ventilation system during the self-test ..................................... 123
10 A-Cone 139
Test List
6 Divan 149
7 Self-test 150
16 ORC 157
18 PM 8050 cd 159
24 Vapor 164
Changes
Appendix
1 Intended Use
Cato is intended for inhalation anaesthesia in semi-closed systems both for mechanical
ventilation and for manual ventilation and spontaneous breathing. The low flow technique
used in semi-closed systems allows for a reduced consumption of gas and anaesthetic
agents.
Cato can be used with all usual inhalation anaesthetics (Enflurane, Isoflurane, Halothane,
Sevoflurane and Desflurane) by changing the anaesthetic vaporizer. Explosive anaesthetics
such as ether, cyclopropane or similar must not be used.
Any use of the equipment requires full understanding and strict observation of the relevant
Instructions for Use. The equipment may only be used for the purposes specified herein.
1.1.2 Maintenance
The equipment must be inspected and serviced by experts at a regular 6-month interval (and
a record kept). We recommend obtaining a service contract with DrägerService. Repairs may
only be carried out by DrägerService. Use only original Dräger parts for maintenance.
Carry out regular safety checks on the equipment as specified in the certification required by
the MedGV regulations.
1.1.4 Accessories
Only use accessories listed in the certification required by the MedGV regulations. Other
accessories may only be used if they are certified as safe and suitable for use with the
equipment.
The equipment may only be used in rooms with mains supply installations complying with
national safety standards (such as VDE 0107 in Germany). Follow the regulations laid down
in DIN IEC 601 / VDE 0750 for electrically powered medical equipment.
The equipment has not been approved for use in explosion-hazard areas.
For electrical connection with equipment not listed in the instructions for use consult the
respective manufacturer and/or an expert.
GBK5170XXXA01.fm 14.06.01
The liability for the proper function of the equipment is irrevocably transferred to the owner or
the operator if the equipment is serviced or repaired by personnel not employed or
authorized by DrägerService and/or if the equipment is used in a manner not conforming to
its intended use.
Dräger Medical AG & Co. KGaA cannot be held responsible for damage caused by non-
compliance with the above recommendations. The warranty and liability provisions of the
Terms of Sale and Delivery of Dräger Medical AG & Co. KGaA are likewise not modified by
the above recommendations.
For internal use only. Copyright reserved.
GBK5170XXXA01.fm 24.01.02
The equipment or parts thereof must be cleaned and disinfected before any maintenance
work is carried out and before it is sent in for repairs.
PM8050cd
0 I
Dräger
Anaesthesia ventilator
(Divan)
0 I
D
automatic
protocol (PC)
Drucker
(im Frame Maker)
Printer
For internal use only. Copyright reserved.
GBK5170XXXA01.fm 14.06.01
Sp0 sensor
2
Temperature sensor
PM8050cd
Anaesthesia ventilator
(switched)
Power supply unit
Mains
PM8050cd
Water trap
Microbial filter
Anaesthesia ventilator
Patient
connection
Microbial filter
Anaestetic gas
scavenging
Reservoir bag
installed inside
to be installed during setup
Prior to use, test the unit according to the check list attached to the equipment.
(MedGV 6 Section 4).
Using the relevant Instructions for Use (for the unit and additional devices) adapt the check
list to the type and the configuration of the unit by deleting and/or adding items as
applicable.
The check list covers the tests to be carried out on Cato before each start-up.
Enter the type identification and the serial number of the unit. Tick off the results of the tests
on the ACTUAL column. Do not forget to enter the date and your signature.
If the equipment is in order, the test described will be completed successfully. Should the
sequence be changed, follow the individual instructions.
Key :
Activities to be performed
Item to be checked
Acknowledge
on airway monitor
Acknowledge
on ventilator
Breathing system
Close flow control valves
O2 -Flush
Hoses, reservoir bag, absorber, humidifier, measurement
connections, valve disc on inspiratory and expiratory end Close off O2 sensor connection.
and fresh-gas hose checked for completeness and tightness? Actuate O2 flush lever
Power supply
Y-piece open?
Mount O2 sensor
Leak IPPV = .... mL
Calibration
Ventilator power on test
PM8050cd and ventilator
Standby
Where tolerances are stated both as percentages and absolute values, the higher value shall
apply.
Identification: The serial number and the part number of Cato and
the type approval code are shown on the rating plate
(nameplate) which is attached to the rear panel of the
trolley. The scope of the equipment can be identified
from the part number.
Weight: 160 kg
Dräger
1480
470 785
700
For internal use only. Copyright reserved.
GBK5170XXXA01.fm 14.06.01
Absorber volume: 1.5 liters of soda lime Drägersorb 800 absorb approx.
150 liters of CO2.
Approximate service period: 6 hours
Pressure control valve: Adjustable from 5 to 70 mbar, 15% of the set value
2.1.4 Ventilator:
TIP:TI: 0 to 60% ± 5%
Resolution: 1 mbar
2.2.2 O2 Measurement
Solution: 1 % by vol.
Zero drift: within the accuracy specified above, without time limit.
Zero drift: within the accuracy specified above, without time limit.
Accuracy (adults):
in the range 70 to 100% SpO2 ± 2% SpO2
in the range 50 to 70% SpO2 ± 3% SpO2
not specified in the range 0 to 50% SpO2
Accuracy (neonates):
in the range 70 to 95% SpO2 ± 3% SpO2
not specified in the range 0 to 70% SpO2
not specified in the range 95 to 100% SpO2
Acoustic pulse signal: 1 tone generated for each pulsation, the tone pitch is
GBK5170XXXA01.fm 14.06.01
Protocol (printer):
Analog output:
CO2: 0 to 10 kPa - 0 to 10 V
Signal
Ground
Noise emission: max. 56dB (A) (in accordance with free field
measurement over reflecting surface)
Protection classes:
The PM 8050 cd has been tested and found to comply with the radio interference
suppression Order 1046 published in the official gazette of the German Federal Post Office
(Deutsche Bundespost).
For internal use only. Copyright reserved.
GBK5170XXXA01.fm 14.06.01
1 Basic Unit
Configuration / Variants
Cato is designed as a 3-gas unit (O2, N2O and AIR) and comes with ORC low flow and low
flow flowmeters.
The unit can be retrofitted with an optional compressed gas cylinder set. This set consists of
two 11-liter cylinders (O2 and N2O). The high-pressure reducers (200 bar/ 5 bar for O2 and
60 bar/ 5 bar for N2O are located immediately below the cylinders. The cylinder pressure
can be read on the manometer of the respective pressure reducer. The cylinders are
connected to the unit by high-pressure coiled tubes. O2 and N2O is supplied to the
mechanical gas dosage through non-return valves.
O2, N2O and AIR of 2.7 to 5.5 bar is supplied via three ISO connectors on the rear panel of
the unit. The central gas supply pressures are indicated on manometers which are located
behind the flowmeter block.
Additional pressure reducers decrease the O2 and N2O pressures to 3 bar and the AIR
pressure to 1.5 bar. The respective gas is delivered to the flowmeter block depending on
For internal use only. Copyright reserved.
which operating mode has been selected. After flowing through the flowmeter block, the gas
mixture flows into the Vapor unit where it is enriched with Halothane, Enflurane, Isoflurane,
Sevoflurane or Desflurane. The fresh gas is then available at the fresh-gas outlet.
GBF5170XXXT01.fm 14.06.01
Druckversorgung Divan
Pressure supply Divan
Behälter O2-Flush
Frischgas
Container
Fresh gas
Ejector 11 12 Alarmventil
Alarmvalve Meßröhrenblock
10 Flowmeter unit
Injector
Umschalter
(bei AIR-Ausfall)
13
Switch
(with Air failure)
Absaugung
Aspirator
14
O2/Fl stat.
15
Steuerventil
Control valve 2,3 bar
Flaschenanschluß stat.
5 bar 16
Cylinder connection
N2O-Sperre
N2O-Cutoff
Functional Description
O2/ZV
ZV-Anschluß
S-Connection Alle Versorgungsdrücke vorhanden
All pressure available
Page 32
AIR/ZV
N2O/ZV
1.1 Integrated Pneumatics
The integrated pneumatics replaces the conventional mechanical gas dosage. It covers the
function of the following individual components:
• check valves;
• N2O shutoff;
• control valve;
• alarm valve;
• mean pressure reducer
selector switch for Divan gas supply and secretion suction system in the event of AIR failure
threshold value switch for the selector switch
Function
An aluminium block is divided into three levels. Each gas (O2, N2O or AIR) is assigned to a
specific level. The pneumatic connection of the individual elements is ensured by channels
within the block. This method saves approx. 50% of plug-in connections compared with a
mechanical gas dosage system. The valves are designed as compact components. Short
switching paths and lack of diaphragms reduce the rate of wear and increase reliability and
life of the assembly.
For internal use only. Copyright reserved.
GBF5170XXXT01.fm 14.06.01
The numbers and characters specified in the gas flow diagrams and in the following Figures
are engraved into the integrated pneumatics.
O2 deficiency alarm
Gas connections 2
at
at
For internal use only. Copyright reserved.
O2 pressure reducer
at
GBF5170XXXT01.fm 14.06.01
Valve elements 1
For internal use only. Copyright reserved.
alarm valve
Valve elements 2
The check valves prevent any gas exchange between the cylinders and the central gas supply
as well as a loss of gas when the connections are open.
* * check valves
*
For internal use only. Copyright reserved.
GBF5170XXXT01.fm 14.06.01
The alarm valve is an integral part of the O2 deficiency alarm. In the event of O2 failure the
whistle in the O2 deficiency alarm is activated.
Alarm valve
Injector
Container
control pressure
When the gas supply is open the container is supplied with gas through the check valve. The
existing O2 pressure closes the alarm valve. If the O2 pressure drops below 1.7 bar the alarm
valve opens. The O2 stored in the container flows through the injector to the whistle. The
injector draws ambient air and extends the alarm time. The audible alarm sounds for approx.
7 secs.
For internal use only. Copyright reserved.
GBF5170XXXT01.fm 14.06.01
O2 and N2O are present at a pressure of 3bar. The AIR pressure is reduced to 1.5 bar. The
3-bar pressure is necessary for the function of the ORC which is located after the control
valve (AIR) and the N2O shutoff.
Flowmeter
block and
ORC
control valve
N2O shutoff
The control valve and the N2O shutoff are supplied with pressure via the O2 line. The control
valve closes. AIR cannot be supplied. The N2O shutoff is opened by the existing pressure.
O2 and N2O are available at the dosage valves of the flowmeter block. The N2O shutoff has
two functions. First, it is used as a changeover valve. Second, if the O2 supply fails or if the
minimum operating pressure drops below approx. 2 bar the spring will close the valve and
interrupt the N2O dosage.
For internal use only. Copyright reserved.
GBF5170XXXT01.fm 14.06.01
The control valve and the N2O shutoff are not triggered. Venting occurs via the gas selector.
The control valve opens in this position. The N2O shutoff closes and prevents a dosage of
N2O. O2 and AIR are available at the dosage valves of the flowmeter block.
Flowmeter
block and
ORC
control valve
N2O shutoff
For internal use only. Copyright reserved.
GBF5170XXXT01.fm 14.06.01
Caution! :
Up to 0.8 L/min of N2O can be supplied over the bypass in the low flow range.
Dosage
check valve
The O2 and N2O flow is controlled by spindle valves on the flowmeter block. Restrictors
located at the inlet ports of the flowmeters generate a flow resistance. The resistance level
varies with the position of the spindle valve and the restrictor values. The resistance exert a
force on the diaphragm of the ORC. As a result thereof the O2 pressure opens the ORC
valve and the N2O pressure closes it. The restrictors and the spring force are designed such
that a minimum O2 concentration of 25% by vol. is ensured. The restrictors limit the maximum
O2 flow to approx. 9 L/min.
For internal use only. Copyright reserved.
Bypass:
A minimum O2 flow of 0.3 to 0.8 L/min is required to open the N2O branch. Working in the
low-flow range is made more difficult by the opening behaviour of the ORC since mixed gas
dosage is not possible in the lower flow range. A non-adjustable restrictor of 0.8 L/min in the
GBF5170XXXT01.fm 14.06.01
N2O branch solves this problem. The safety function of the ORC is disabled in this flow
range.
• It makes sure that a vital oxygen concentration is present in the fresh gas supplied to the
patient.
• It closes the N2O branch in the event of oxygen shortage.
Measurement unit
Vapor
O2 Air N2O
O2
Air
N2O
S-ORC
The spindle valves on the flowmeter block meter the O2 and N2O. Resistors located in front
of the flowmeters’ inlet ports in the S-ORC generate a backpressure while the gas flows. The
backpressure level varies with the opening degree of the spindle valves and the value of the
resistors. The backpressure exerts a pressure on the diaphragm of the S-ORC causing the
O2 pressure to open the S-ORC valve and the N2O pressure to close the S-ORC valve. The
resistors and the spring energy are dimensioned such that a minimum concentration of 21%
by vol. O2 is ensured. The resistors limit the maximum O2 flow to approx. 9 L/min.
For internal use only. Copyright reserved.
GBF5170XXXT01.fm 14.06.01
4 X 12 83081
R 20311
N2O
E
A
E
A
M 32941 O2
D Vent hole
M 30 307
M 30 306
Section C-D
Reject diaphragms which were already fitted. Do not re-use those diaphragms.
A
N2O
A E
O2
Note: Do not confuse the connections of the flowmeter block to the S-ORC valve, otherwise
the system will not function.
For internal use only. Copyright reserved.
GBF5170XXXT01.fm 14.06.01
• Adjust the spindles (M 30306) for O2 and N2O such that they protrude about 1 mm from
the valve element M 30307.
• Set the distance between the top edge of the valve element (M 32941) and the
enclosure to about 10 mm.
Given a supply pressure of 5 ±0.1 bar and a supply rate of 20 ±1 L/min, the pressure
regulators for the N2O and O2 supply are adjusted dynamically to the following
backpressures:
The leak test must be carried out with the S-ORC valve in place.
• Fine-adjust the S-ORC on the N2O side using the spindle M 30306.
• Adjust the concentration to 25% by vol. O2,with an O2 flow of 1 L/min. The spindle
M 30306 on the O2 side remains unchanged.
• If the concentrations for the other O2 flows are not within tolerance, connect a
manometer to the inlet port of the O2 side of the S-ORC (E) and check whether the O2
control pressure is at an O2 flow of 200 mL/min is 25 +1 mbar.
• If the O2 control pressure is not 25 +1 mbar, adjust the value with the spindle M 30306
on the O2 side.
• After adjusting the O2 control pressure, check the concentration again at 1 L/min O2. If
necessary, correct the value with the N2O spindle.
For internal use only. Copyright reserved.
GBF5170XXXT01.fm 14.06.01
After adjusting/re-adjusting the S-ORC, the concentration test ist done with an O2 analyzer.
If the adjustments on the S-ORC have not been changed, the check can be done with the
test certificate. In this case, the flow values can be used for testing. When replacing the
entire S-ORC, testing according to the test certificate is also satisfactory.
The secretion suction system (including the secretion containers) is installed on one side of
the trolley. During normal operation the secretion suction system is powered by an adjustable
compressed-air ejector. In the event of AIR failure the ejector can be powered with O2.
D
For internal use only. Copyright reserved.
GBF5170XXXT01.fm 14.06.01
The power supply unit provides the dc voltages required for the Divan and supplies the
PM 8050cd with the mains voltage via an isolating transformer. The isolating transformer is
connected to the power switch.
Schematic
Non-swichted Swtiched
Mains voltage selector switch
Isolating transformer
D C B A Power input: 100,120,127,230,240 Vac
Socket strip Power output: 230 Vac
S1 T1
Isolated power
K2 Si 5 supply
Wi 1 (monitoring)
Si 4
X2
Si 3 Wi 2
T2 Mains relay
Si 2 Si 1 (230 bis 240 Vac )
X5
PCB Power Supply X1
K2 (DIVAN )
82 01 667
Autotransformer
24V + 5V + 15V -15 V
Power input: 100,120,127 Vac
Power output: 240 Vac X2
X1 X6
The mains voltage flows to the non-switched socket D and, through fuses Si1 and Si2, to the
central power switch on the front panel of Cato. The contactor K2 remains in the rest position
(contacts K2 are open) as long as the power switch is open. This means, the whole power
supply unit is de-energized (except for the above mentioned socket D).
The contactor K2 is supplied with mains voltage via the autotransformer and closes. The
isolating transformer T1 is supplied with voltage via the closed contacts of K2, Si3, Si4 and
the mains voltage selector switch. The mains voltage selector switch has four levels. Three
levels switch the pick-offs of the isolating transformer according to the position of the switch
and the mains voltage applied. The fourth level is used to supply the autotransformer T2 with
the voltage required to energize the contactor K2.
The electrically isolated output of the isolating transformer is protected by the fuse Si5. It
supplies the PM8050 cd via X2 and the power supply PCB (Divan) via X5. The output X6 of
the voltage supply provides Divan with the following voltages:
• + 24 V
• +5V
• + 15 V
• – 15 V
For internal use only. Copyright reserved.
GBF5170XXXT02.fm 14.06.01
4 x gr/yl
X2
gr/yl
gr/yl
bl bk
br bl br bl
bk 39 39 bk
biscuit connectors T2
rd wh 35 35 bk 240 V
bk bl
bk br A1 bk K2 ( Mains relay )
wh 40 31 rd 127 V
bl 27 wh 110 V
br T3 K2 K2 14 23 bl 100 V
A2
bl 13 0V bl bl
bl 0V
bl br br bl
For internal use only. Copyright reserved.
2 1 3 4
X1 ( Power switch )
GBF5170XXXT02.fm 14.06.01
Warning
Hazardous voltages are present. To avoid the risk of electric shock and
danger to personal health always unplug the power cord before opening
the equipment.
Power supply
PCB
Mains voltage
selector
K2: contactor
T2: autotransformer
The Power Supply PCB transforms the mains voltage 230 VAC into dc voltages required for
Divan.
U2
F1A
18 36 196
U3
F1A
18 36 196
Pin 10 15 ± 0,45 V
U3
Pin 11 GND U3
Mains
T4A
18 35 122
6 5 4 3 2 1
X1
SL X5
For internal use only. Copyright reserved.
L1 from
N Isolating
Mains voltage
transformer T1
Umains = 230 Vac -15% / +10%
50/60 Hz 49 to 61 Hz
GBF5170XXXT02.fm 14.06.01
U1 U2 U3 U4
Voltage (V) 5.1 15 15 24
Continuous current (A) 4 0.5 0.5 6
Peak current (A) — — — 8
protection
Short-circuit proof yes yes yes yes
+5Vsense
U1=+5 ± 0,15V
Powerfail:
The Powerfail signal is made available by optocouplers and open collector outputs.
The ventilator in the Cato system is an electronically controlled piston cylinder system (13)
which is driven by an electric motor. The ventilator has a compact breathing system (1-9)
which can be removed separately.
4 13
1
5
12 BAG POP-OFF
7
2
6
MAN
MAN
9
10
8
12
11
9. Plug to seal the Y-piece during the front panel leak test and during the self-test
10. Front panel
11. Absorber
12. Breathing bag connection (BAG)
GBF5170XXXT03.fm 14.06.01
17 18 1 2 3 4
5
16
6
15
7
14 13 12 11 10 9 8
The electronics converts the values set by the operator into control signals for the drive motor
and the control solenoid valves. The control valve exerts a pressure (control pressure) on the
pneumatic diaphragm valves in the compact breathing system in order to generate the
desired ventilation mode. The respective position of the piston in the cylinder is determined
by an incremental encoder and the value is channeled back to the control unit.
Ventilator Control
control unit
settings
Computer 1 Computer 2
Control Monitoring
1 2
from 1 from 2
Motor
decoder
For internal use only. Copyright reserved.
GBF5170XXXT03.fm 14.06.01
Ventilator Monitoring
settings
Computer 1 Computer 2
Control + Monitoring Monitoring
1 2 3 4
from 1 from 2
Motor
decoder
Release lever
Piston C1 B1
Pressure Pressure measuring
connection connection
to roller diaphragm airway pressure
The function of the piston cylinder unit is to deliver the fresh gas supplied from the flowmeter
block and stored in the breathing bag to the patient at a defined volume, pressure and speed.
During expiration the piston fills with the gas expired by the patient and with the fresh gas
stored in the breathing bag. During inspiration, this piston volume is administered to the
patient by an accurate stroke. Before reaching the patient it flows through a CO2 absorber.
To ensure this gas flow, the control system switches the valves of the breathing system
accordingly.
The most important elements of the piston cylinder unit are a piston which is moved by a
drive unit located in the rear part of the ventilator and a double roller diaphragm which is filled
with gas during operation and which functions as a sliding bearing. The double roller
diaphragm moves the piston along the inside walls of the cylinder. The front cylinder part
separates the ambient atmosphere from the respiratory gas of the patient. Several interfaces
are located on the underside facing the ventilator. The airway pressure is measured via a
pressure measurement connection (B1) and a pressure sensor. The pressure connection
(C1) provides the double roller diaphragm with the required positive pressure of approx.
For internal use only. Copyright reserved.
160 mbar. The patient gas is supplied to the compact breathing system via a large lateral
outlet port.
GBF5170XXXT03.fm 14.06.01
The piston cylinder unit should be checked and serviced at regular intervals since the roller
diaphragm may become leaky as a result of ageing and frequent sterilization. The steps to
replace the roller diaphragm half on the patient side are as follows:
Step 1:
Remove the piston cylinder unit from the ventilator while in the standby mode.
Step 2:
Release the piston cylinder unit, pull straight up and remove.
Step 3:
Loosen the 2 quick-release locks (1) with a 6-mm wrench and remove the cylinder head (2).
Step 4:
Grasp the piston and unscrew the central piston screw using the 6-mm wrench and remove
the upper part of the piston (3).
Step 5:
Replace the roller diaphragm half (4) on the patient side against a new one. To ensure a tight
seal, the embossed manufacturing clock must point towards the upper part of the piston (3).
Step 6:
Use the reverse order for reassembly. Make sure neither kinks nor folds are formed during
installation of the roller diaphragm. The complete roller diaphragm must form a typical U-fold
at both ends.
Step 7:
Carry out the Cato power-on test to check the function and the tightness of the seal of the
roller diaphragm. The roller diaphragm must unroll without forming folds.
Note:
It is not advisable that the roller diaphragm half on the equipment side, ie. the non-
contaminated part, be replaced by the user. It should rather be replaced by DrägerService
since its safe function can only be ensured if various screw locking agents and special tools
(tightening torques) are used.
For internal use only. Copyright reserved.
GBF5170XXXT03.fm 14.06.01
Legend
1. Quick-release locks
2. Cylinder head
3. Upper part of the piston
4. Roller diaphragm half
5. Manufacturing clock
11 10
For internal use only. Copyright reserved.
GBF5170XXXT03.fm 14.06.01
The fan (2) and the drive unit (1) are located at the rear part of the ventilator. To access the
drive unit for repair purposes, it is necessary to remove the complete ventilator from the Cato
basic unit.
13 1 3
6
4
7
12
8
5
9
11 10
Legend
TIP:
If the piston cylinder unit cannot be removed easily it is possible to have the removal position
readjusted by DrägerService through software manipulation.
For internal use only. Copyright reserved.
GBF5170XXXT03.fm 14.06.01
The compact breathing system consists of a control valve plate (1) with the diaphragm valves
which are pneumatically triggered by the solenoid valves. The inspiratory and the expiratory
valves (7) responsible for the respiratory phase and the valve craters are located in the
respiratory valve base (2). The APL valve (Airway Pressure Limiting valve) (8) for pressure
control of the airway pressure during manual ventilation is located in the respiratory gas
branch. A check valve (3) and an emergency ventilation valve (4) are integrated in the airway.
The flow sensor element (9) of the PM 8050 cd is found at the connection of respiratory
hoses (12). The fresh gas from the flowmeter block flows into the breathing system via the
fresh gas terminal (15) and is stored in the breathing bag (16). In the inspiratory phase, it is
administered to the patient through the inspiratory valve (7) by a piston stroke. In the
expiratory phase, the gas flows through the expiratory valve (7) and the absorber element (11)
into the CO2 absorber container (10). The absorber element makes sure the expiratory gas
flows evenly through the entire surface of the soda lime in absorber.
7 9 12 16 12 11 10
5 13 4 3 14 15
Legend:
The two large lip seals on the underside of the respiratory gas base to the absorber element
(11) and to the absorber container (10) should be regularly checked for contamination or
damage by absorber lime residues. Otherwise major leaks may occur at this point which will
be detected during the self-test.
The heater’s contact surface (approx. 5 x 4 cm) for indirect heating of respiratory gas is
located immediately under the built-in compact breathing system. The heater is permanently
on during operation. In case of a malfunction, a spring-loaded contact switch causes the
heater to be immediately switched off. The heater electronics is mounted immediately under
the contact surface. The maximum heating temperature is set such that the respiratory gas
temperature at the Y-piece does not exceed 30 °C.
Note:
After switching on the master gas switch, compressed air is delivered via a filter to the
pressure regulator which reduces the gas pressure to a constant value of 1.5 bar
(backpressure). The gas is delivered to Dos2 and Dos4 (0.5 L/min). Dosage Dos4 vents the
gas flow to the PEEP valve via the 2.0 L/min dosage and the solenoid valve MY8. A pressure
of 160 mbar is generated for the roller diaphragm by means of 2.0 L dosage and dosage
Dos2. The safety valve 250 mbar discharges at a pressure of approx. 250 mbar. The gas is
delivered to the double roller diaphragm of the piston cylinder unit via the solenoid valve MY9
at the interface C1.
Divan
87 mbar
2,7 - 6 bar
MY 8 Dos 2
E
Dos 3
MY 4 P PE 3
PE 2 PE 1
E E
P P
MY 9 MY 1 MY 3 MY 2
Cu PM 8050cd
V1 V6 V3 V2 V7
E A- Gas
Y1 Insp. O2
Fresh gas CO2 aspirator
(PM 8050cd)
Ab- POP-OFF Y7
sorber
Reser-
Motor drive C1 B1
voir PEEP
bag V
Y3 Y2
Exp. Y6
Piston-Cylinder
unit Spt. Man.
KZE APL
5-70 mbar
Compact Breathing
check valves Auxiliary air valve System BREASY
scavenging
For internal use only. Copyright reserved.
system
GBF5170XXXT03.fm 14.06.01
The piston chamber, however, must be filled until the piston is pushed as far as it will go.
Divan
87 mbar
2,7 - 6 bar
MY 8 Dos 2
E
Dos 3 MY 4 P PE 3
PE 2 PE 1
E E
P P
MY 9 MY 1 MY 3 MY 2
Cu PM 8050cd
V1 V6 V3 V2 V7
E A- Gas
Y1 Insp. O2
Fresh gas CO2 aspirator
(PM 8050cd)
Ab- POP-OFF Y7
sorber
Reser-
Motor drive C1 B1
voir PEEP
bag V
Y3 Y2
Exp. Y6
Piston-Cylinder
unit Spt. Man.
KZE APL
5-70 mbar
Compact Breathing
check valves Auxiliary air valve System BREASY
scavenging
system
For internal use only. Copyright reserved.
GBF5170XXXT03.fm 14.06.01
The solenoid valve MY8 is activated electrically and pneumatically closed. No additional gas
flows into the control system (pressure free); consequently no control pressures can be built
up for the valves.
The solenoid valve MY8 is de-energized, thus being switched so that gas can pass through.
The gas flows to the solenoid valves MY9, MY1, MY3, MY2 and MY4 (as described under
Section 5.3.1 „Device de-energized”) .
IPPV Inspiration
Valve Y1 is supplied with pressure via solenoid valve MY1 thus blocking the fresh-gas flow
inside the system. The motor drive of the piston cylinder unit sets the piston in motion. The
content of the cylinder is pressed into the compact breathing system. Fresh gas is delivered
to the patient via the absorber and the inspiratory valve. Gas is also present at valve Y7. The
blocking effect of the expiratory valve is created by a PEEP valve which is set to 40 mbar. The
piston pressure is measured at connection B1 by the pressure transducer PE2. PE1
measures the inspiratory pressure of the patient. Inspiratory O2 measurement is carried out
by the usual oxygen sensor.
Divan
87 mbar
2,7 - 6 bar MY 8 Dos 2
E
Dos 3
MY 4 P PE 3
160 mbar
87 mbar
PE 2 PE 1
E E
P P
MY 9 MY 1 MY 3 MY 2
Cu PM 8050cd
V1 V6 V3 V2 V7
E A- Gas
Y1 Insp. O2
Fresh gas CO2 aspirator
(PM 8050cd)
Ab- POP-OFF Y7
sorber
Reser-
Motor drive C1 B1 voir
bag PEEP V
Y3 Y2
Exp. Y6
Piston-Cylinder
unit
Spt. Man.
KZE
APL
For internal use only. Copyright reserved.
5-70 mbar
scavenging
system
GBF5170XXXT03.fm 14.06.01
At the beginning of the expiration, valve Y1 is vented via solenoid valve MY1. Gas will flow
from the compact breathing system into the breathing bag during the first moments, since
the pressure in the compact breathing system is often higher than in the breathing bag. At
the same time, the PEEP valve is set to the desired pressure. The piston returns to its initial
position after a delay time of 70 ms. The volume required to fill the piston is taken both from
the air expired by the patient as well as from the air in the breathing bag. Should the volume
be too low, the message „Fresh gas low“ will be displayed and the system will react
accordingly.
PE 2 PE 1
E E
P P
MY 9 MY 1 MY 3 MY 2
Cu PM 8050cd
V1 V6 V3 V2 V7
E
Y1 O2 A- Gas
Insp.
Fresh gas CO2 aspirator
(PM 8050cd)
POP-OFF Y7
Ab-
Reser- sorber
C1 B1 voir
Motor drive PEEP
bag V
Y3 Y2
Exp. Y6
Piston-Cylinder
unit Spt. Man.
KZE APL
5-70 mbar
Compact Breathing
check valves Auxiliary air valve
System BREASY
scavenging
system
For internal use only. Copyright reserved.
GBF5170XXXT03.fm 14.06.01
The inspiration sequence is the same as described in „IPPV Inspiration”. The PEEP valve is
set to 40 mbar.
The PEEP valve switches over to the desired pressure at the beginning of the expiration
phase. The PEEP pressure is present only in the patient and in the device up to the PEEP
valve or inspiratory valve. The pressure in the piston space is less than 2 mbar at the end of
the expiration phase. Should gas loss occur in the patient area due to a sudden leakage, it is
no longer possible to build up the PEEP pressure for this breathing stroke. In the event that
a great amount of gas is lost, the message „Fresh gas low“ will be displayed during the next
breathing stroke. A deviation between PEEP actual value and PEEP target value will occur in
the PEEP function as a result of tolerances in the electrical and pneumatic elements. This
deviation is compensated by means of a software control loop. A target/actual-value
comparison is carried out after each breathing stroke in order to correct the next stroke
adequately.
For internal use only. Copyright reserved.
GBF5170XXXT03.fm 14.06.01
Should a pressure of approx. 2 mbar build up at PE2 at the end of the expiration phase, this
pressure is recorded during the first ventilation stroke. During the next stroke, valve Y1 opens
as usual and the PEEP valve is set to the actual value. The piston starts working after a time
delay of 70 ms. Therefore, the solenoid valve MY2 is switched to vent position. The excess
fresh gas is delivered to the anaesthetic gas scavenging line via valve Y2 and the non-return
valve.
Divan
87 mbar
2,7 - 6 bar
MY 8 Dos 2
E
Dos 3 MY 4 P PE 3
160 mbar
87 mbar
PE 2 PE 1
E E
P P
MY 9 MY 1 MY 3 MY 2
Cu PM 8050cd
V1 V6 V3 V2 V7
E
Y1 O2 A- Gas
Fresh gas Insp.
CO2 aspirator
(PM 8050cd)
Ab- POP-OFF Y7
Reser- sorber
Motor drive C1 B1 voir
bag PEEP V
Y3 Y2
Exp. Y6
Piston-Cylinder
unit Man.
Spt.
KZE APL
5-70 mbar
scavenging
Cough
For internal use only. Copyright reserved.
If, during inspiration, the patient produces a backpressure (cough) then the pressure is
recorded by the pressure transducer PE1. If the pressure is 5% above the set Pmax value, the
device switches over to expiration.
GBF5170XXXT03.fm 14.06.01
Fresh gas supply is activated in the position Man./Spont. (Y1 opened, MY1 set to venting
mode). The APL valve is switched on via valve Y3 (MY3 vented). Valves Y2 and Y7 are de-
activated. The piston is moved forward and held in this position.
Manual
During manual ventilation, the lever of the APL valve is set to „MAN“ position. Now it is
possible to carry out manual ventilation using the breathing bag. Pressure limitation of
5-70 mbar can be adjusted at the APL valve. The excess gas flows to the anaesthetic gas
scavenging line via the APL valve.
Inspiration
PE 2 PE 1
E E
P P
MY 9 MY 1 MY 3 MY 2
Cu PM 8050cd
V1 V6 V3 V2 V7
E
Y1 O2 A- Gas
Insp.
Fresh gas CO2 aspirator
(PM 8050cd)
Ab- POP-OFF Y7
Reser- sorber
C1 B1 voir
Motor drive
bag PEEP V
Y3 Y2
Exp. Y6
Piston-Cylinder
unit Spt. Man.
KZE APL
5-70 mbar
scavenging
system
For internal use only. Copyright reserved.
GBF5170XXXT03.fm 14.06.01
PE 2 PE 1
E E
P P
MY 9 MY 1 MY 3 MY 2
Cu PM 8050cd
V1 V6 V3 V2 V7
E
Y1 O2 A- Gas
Insp. CO2 aspirator
Fresh gas
(PM 8050cd)
POP-OFF Y7
Ab-
Reser- sorber
C1 B1 voir
Motor drive
bag PEEP V
Y3 Y2
Exp. Y6
Piston-Cylinder
unit Man.
Spt.
KZE APL
5-70 mbar
scavenging
system
Spontaneous
Spontaneous breathing of the patient is made possible by setting the APL valve to „SPONT“.
The air exhaled by the patient is fed to the anaesthetic gas scavenging line via the non-return
valve.
Result: a pressure is built up which ensures that the breathing bag is sufficiently filled for the
inspiration phase. In case of gas deficiency in the breathing bag, additional gas is supplied
via the auxiliary air valve.
For internal use only. Copyright reserved.
GBF5170XXXT03.fm 14.06.01
PE 2 PE 1
E E
P P
MY 9 MY 1 MY 3 MY 2
Cu PM 8050cd
V1 V6 V3 V2 V7
E A- Gas
Y1 Insp. O2
Fresh gas CO2 aspirator
(PM 8050cd)
Ab- POP-OFF Y7
Reser- sorber
C1 B1 voir
Motor drive bag PEEP V
Y3 Y2
Exp. Y6
Piston-Cylinder
unit Man.
Spt.
KZE
APL
5-70 mbar
Compact Breathing
check valves Auxiliary air valve
System BREASY
scavenging
system
Expiration
PE 2 PE 1
E E
P P
MY 9 MY 1 MY 3 MY 2
Cu PM 8050cd
V1 V6 V3 V2 V7
E
Y1 O2 A- Gas
Insp. CO2 aspirator
Fresh gas
(PM 8050cd)
Ab- POP-OFF Y7
Reser- sorber
voir
Motor drive C1 B1
bag PEEP V
Y3 Y2
Exp. Y6
For internal use only. Copyright reserved.
Piston-Cylinder
unit Man.
Spt.
KZE
APL
5-70 mbar
System BREASY
scavenging
system
Safety Valve
Should a pressure build up in the system which is higher than the control pressure, the
diaphragm in the valve Y7 is pushed away from the crater. The gas is released via the non-
return valve.
Emergency Venting
The POP-OFF valve is used as an emergency venting device. If the POP-OFF valve is
pressed, the 87-mbar control pressure will be instantly vented. As a result thereof, the
emergency venting Y7 opens. The gas in the compact breathing system flows to the
anaesthetic gas scavenging line via the non-return valve.
PE 2 PE 1
E E
P P
MY 9 MY 1 MY 3 MY 2
Cu PM 8050cd
V1 V6 V3 V2 V7
E A- Gas
Y1 Insp. O2
Fresh gas CO2 aspirator
(PM 8050cd)
POP-OFF Y7
Ab-
Reser- sorber
C1 B1 voir
Motor drive bag PEEP V
Y3 Y2
Exp. Y6
Piston-Cylinder
unit Man.
Spt.
KZE
APL
5-70 mbar
Compact Breathing
check valves Auxiliary air valve
System BREASY
scavenging
system
For internal use only. Copyright reserved.
GBF5170XXXT03.fm 14.06.01
A leak test can be carried out by pressing the button „Test”. The roller diaphragm is
pressurized via MY9. The valves Y1, Y2, Y3 and Y7 are closed. The piston is used to maintain
a certain pressure (standard: 30 mbar) over a period of approx. 15 s. The leak is calculated
from the piston displacement and the value is displayed. The compliance is also calculated.
Divan
87 mbar
2,7 - 6 bar
MY 8 Dos 2
E
Dos 3 MY 4 P PE 3
PE 2 PE 1
E E
P P
MY 9 MY 1 MY 3 MY 2
Cu PM 8050cd
V1 V6 V3 V2 V7
E
Y1 Insp. O2 A- Gas
Fresh gas CO2 aspirator
(PM 8050cd)
Ab- POP-OFF Y7
sorber
Reser-
Motor drive C1 B1 voir
PEEP V
bag
Y3 Y2
Exp. Y6
Piston-Cylinder
unit
KZE Spt. Man.
APL
5-70 mbar
Compact Breathing
check valves Auxiliary air valve
System BREASY
scavenging
system
For internal use only. Copyright reserved.
GBF5170XXXT03.fm 14.06.01
IPPV Ventilation
Pressure
Druck
(mbar)
( mbar )
P max
PEEP
0 Time
T IP Zeit ((s)
s)
TI TE
1 / f IPPV
TIP : TI = 0%
P (mbar)
IPPV
Pmax = 50
VT = 800 ml
f = 12 1/mn
TI : TE = 1 : 2
t (s)
TIP : TI = 20%
P (mbar)
t (s)
TIP : TI = 40%
P (mbar)
t (s)
TIP : TI = 60%
P (mbar)
For internal use only. Copyright reserved.
t (s)
GBF5170XXXT03.fm 14.06.01
The sequence of inspiration is the same as described under IPPV. Expiration also takes
place as described under IPPV. However, the time between two ventilation strokes is
prolonged. During this time, the patient can breath fresh gas spontaneously via valve Y1. The
spontaneous breathing time can be calculated as follows:
The time window for synchronization is dependent on the spontaneous breathing time.
Time Window
Ttrig (s)
2,5
1,5
0,5
0,1
For internal use only. Copyright reserved.
A detailed description of the more important messages is given in the following pages. The
remaining messages are described in the Instructions for Use.
This message is displayed, if during expiration phase in the mechanical ventilation mode, the
piston moves to the vacuum pressure limit. Vacuum pressure means -3 mbar at P1 or -8 mbar
at P2. The message is displayed once per respiratory phase. It is accompanied by an audible
alarm if there is no connection to PM 8050 cd.
In test block 8 (IPPV leak test) this message may result from three different conditions:
• A pressure of approx. 30 mbar is generated during the leak test. This pressure is
measured by P1 and stored as reference pressure. If, during the leak test, the pressure
at PE1 exceeds the reference pressure + 10 mbar, the message “fresh gas closed” will
be displayed. If the unit is equipped with a measuring gas return line, this message will
also be displayed if the Andros bench of the PM 8050 cd is performing a calibration
during the leak test. Explanation: The ambient air drawn for the calibration flows into the
breathing system of Divan via the measuring gas return line. This leads to an increased
pressure and the message „fresh gas closed“ is displayed.
• If the pressures between PE1 and PE2 deviate by more than 10 mbar, the message
„Fresh gas closed“ is displayed instead of „P-sensor error“.
• If a relatively large leak has occured then the message „fresh gas closed“ may also be
displayed if more than half of the piston stroke is necessary to maintain the pressure.
This message is displayed if the test pressure can not be generated during a leak test before
half of the piston stroke has been carried out. It is accompanied by an audible alarm and
For internal use only. Copyright reserved.
must be confirmed.
GBF5170XXXT03.fm 14.06.01
This message is displayed if the VT setting at the front panel has been changed to values
< 200 mL. The message is displayed only once when reaching the < 200 mL range. It must
be confirmed and is accompanied by an audible alarm.
A test stroke is activated every 10 min during mechanical ventilation, in order to check
subsystem 1 for leakage. (Subsystem 1: ventilation hoses, patient connection , inspiratory
valve). At he end of the inspiratory pause there is a pause of approx. 1 s during which the
pressure drop in subsystem 1 is measured. The message is displayed if the pressure drop is
> 15%. The maximum test pressure is 30 mbar. The message is accompanied by an audible
alarm and must be confirmed. If the leak is > 500 mL/min during the leak test and the user
accepts the leak then this message will be suppressed during operation. The message
„subsystem 1“ can also be displayed after the leak test in the self-test routine.
t t + 1s
Pressure Limit
This message is displayed if , during inspiration in the mechanical ventilation mode, Divan
detects a pressure P1 which is > the Pmax adjusted on the front panel. The message is
displayed once per ventilation stroke. It is not accompanied by an audible alarm. (For special
characteristics refer to "P-Sensor Error")
This message is displayed if Divan detects that the pressure P1 at the end of the expiration is
10 mbar higher than the set PEEP value. The message is displayed once per ventilation
stroke and is accompanied by an audible alarm if there is no connection to the PM 8050 cd.
For internal use only. Copyright reserved.
GBF5170XXXT03.fm 14.06.01
This message is displayed if, during inspiration in the mechanical ventilation mode, a
pressure difference of > 10 mbar is detected between P1 and P2. A set PEEP is taken into
consideration. The current piston stroke is interrupted immediately and the expiraton is
activated. In order to avoid errant messages due to sudden pressure peaks caused by the
patient, the message „Pressure limit“ is displayed after first occurrence of this fault. The
message „P-Sensor Error“ is displayed with every succeeding continuously occurring error.
The message is displayed once per ventilation stroke and is accompanied by an audible
alarm if there is no connection to PM 8050 cd.
This message is displayed if, during operation, the control pressure P3 at the 87-mbar safety
valve and the control pressure P4 at the dosage Dos3 fall below 40 mbar. (Refer to Note).
The message remains on the display until one of the control pressures exceeds 40 mbar. It
is not displayed in the Standby mode. The message is accompanied by an audible alarm if
there is no connection to PM 8050 cd.
Note
There are two ways of generating a control pressure. In normal operating mode, the control
pressure is generated via that 87-mbar safety valve (control pressure P3). Apart from this,
the control pressure is generated during the self-test or in case of failure by switching from
MY4 to Dos3 (control pressure P4).
The self-test checks whether the piston cylinder unit was inserted correctly. For this
purpose, the piston generates a small stroke starting at the removal position. If a low vacuum
is generated by this piston stroke, then the device assumes that the piston cylinder unit was
inserted correctly. If not, the message „remove piston“ will be displayed. The piston will
return to its initial position, after removing the piston and confirming the message.
It is not possible to generate a vacuum by the small piston stroke if the absorber container
has not been mounted or if the inspiratory valve plate has not been inserted (major leak). The
message „leak in breathing system“ is displayed. After confirming the message, the piston
moves to a position where it can be removed easily. The message „remove piston“ is then
displayed.
This message is displayed if a leak test was carried out successfully. The message must be
For internal use only. Copyright reserved.
confirmed if a leak > 175 mL/min has occurred. The message is accompanied by an audible
alarm.
GBF5170XXXT03.fm 14.06.01
This message is displayed if the test pressure has been generated but the measured leak
rate is too high and can not be displayed. The message is also displayed if the front reversal
point for ventilation has been exceeded during measurement. The message is accompanied
by an audible alarm and must be confirmed.
For internal use only. Copyright reserved.
GBF5170XXXT03.fm 14.06.01
General Description
1rd Approximation
1. Annäherung
2nd Approximation
2. Annäherung
3rd Approximation
3. Annäherung
4th Approximation
4. Annäherung
1. Measurement of device compliance, including hose compliance, during the self-test and
using the „Test“ key. It is determined which volume ∆Vdevice is necessary in order to
achieve a pressure of 20 mbar (∆Pdevice).
Pressure
Druck
(mbar)
(mbar)
P2
P
P2
PEEP
0
TimeZeit
(s)(s)
For internal use only. Copyright reserved.
P = P2 - P1
If the device compliance and the ∆P is known, it is possible to calculate which volume ∆V was
consumed by the device. This amount is not administered to the patient.
GBF5170XXXT03.fm 14.06.01
∆V= Cdevice x ∆P
VT corrected = V T + ∆V
4. The same procedure is repeated during the second and all successive strokes. If the
deviation is below 2%, the correction procedure is stopped. A maximum of 6
approximation steps is carried out (1 cycle VT + 6 approximation steps = 7 respiratory
cycles). The ventilation strokes up to the 20th respiratory cycle, inclusive, will be
performed using the corrected V T which was found to be the best one. Then the
correction value is re-checked for the next 20 respiratory cycles. Should a deviation of
> 5% occur, then the next correction procedure is initiated.
Examples
Assume that the device has calculated a device compliance of Cdevice = 4 mL/mbar during
the self-test. The volume administered to the patient shall be VT1 = 600 mL. The volume of
600 mL is supplied to the system (breathing system, hose, patient) with the first stroke.
V T1 600 mL
Ctotal = _________ =_________ = 20 mL/mbar
∆ P1 30 mbar
Cpat. = 16 mL/mbar
Vpat.1 = 480 mL
There is an error of
For internal use only. Copyright reserved.
V T2 = 720 mL
The newly calculated VT2 is applied with the second stroke and the following system pressure
is built up:
VT2 720 mL
∆P2 = _________ =_________ = 36 mbar
Ctotal 20 mL/mbar
In order to correct this error, the device increases the volume to be applied by Verror2.
The newly calculated volume VT2 is applied with the third stroke resulting in a system pressrue
of
V T3 744 mL
∆P2 = _________ =_________ = 37.2 mbar
Ctotal 20 mL/mbar
Vpat.3 = 595.2 mL
Now the calculated error is below 2% and therefore the device stops the VT approximation
procedure for the moment
GBF5170XXXT03.fm 14.06.01
The voltages required for electronic control are supplied from a separate power supply unit.
External data exchange with the operator is carried out via the front panel. Data exchange
with with the PM 8050 cd takes place via the signal line/RS232 interface.
Legend
PCB Incremental PCB Pulse
Shaft encoder Fan 20
Encoder Generator PCB CPU1 (Standard2) = number of cores
X 16 X1 X4 = plugged in
12
3 PCB ADDA
4 2 = soldered
2 X4
= screw connection
Cable harness 2
(Divan) Cable PCB CPU 2
harness 5 (Monitoring)
Motor (Divan)
X1
d
19 X1 ar Operator panel
Cable harness 1 r bo
Cable harness 1 he
8201647
(Divan) 40 X2 ot
11/93
V
PCB Heater
X2 B
(compact
II X1
1 PC
breathing
3 40 X5
system) 16 50 12
III 15
X3
X2
0
Illuminated foil 19 X3
(mech. gas I X 18
IV
2 10 X5
dosage) X10
X4 2
0 X15
X5
9 X 32 31 30 PCB Frontplate
Power pack
IV X X X9
(Cicero EM) 2
(Cato)
V III 4
X6
Functional Description
4
Cable harness 2
Cable- X7 X8
harness 5 X7 X8
PCB Cable harness 4
Optocouplers 7 2 6
PCB PCB
PCB CICERO Pulse generator
Alphanumeric Display
Switching auxiliary
contacts contacts PM 8060/ 3/2-way
Peep valve
Main power valves
PM 8050cd
switch DIVAN
Interconnection Diagramm
Page 93
7.12 Electronics Components
Central Electronics
The principal printed circuit board of the system electronics is the Motherboard PCB.
The CPU Standard 2 PBC and the CPU Monitoring PBC are the two brain cells of the
ventilator and use the concept of mutual data checking.
The CPU Standard 2 PCB, in conjunction with the output/input board ADDA generates the
majority of the control and actual-value signals for the peripherals such as motor control,
valve control and Front Panel PCB. Almost all the control signals are used to produce a quit
(acknowledgement) signal in the corresponding peripheral hardware. This quit
(acknowledgement) signal is then read in parallel by the ADDA PCB and CPU Monitoring
PCB using parallel ports. The analog signals for actual values and pressures are likewise
picked up by both the ADDA PCB and the CPU Monitoring PCB. They do this independently
of one another with A/D converters.
In constrast to the ADDA PCB and the CPU Standard 2 PCB, the independent CPU
Monitoring PCB is responsible almost exclusively for the check function described above
and constantly exchanges data with the CPU Standard 2 PCB via communication ports for
the purpose of verifiying the quit and analog signals.
The Motherboard PCB is located on a base plate which also accomodates the solenoid
valves for the valve controller of the compact breathing system, the electronic PEEP valve and
other pneumatic elements. Also located on the Motherboard PCB are the pressure sensors
for the airway, piston and roller diaphragm pressures, the rechargeable battery and the horn.
Peripheral Electronics
The remaining peripherals, including the motor, are connected to the motherboard by the
wiring harness. The Front Panel PCB is connected up to the motherboard by a ribbon cable.
Peripheral PCBs:
The Motherboard PCB accomodates the three printed circuit boards: ADDA, CPU
Monitoring and CPU Standard 2.
Motor Controller
The task of the motor controller is to effect movement of the piston in the piston cylinder unit
with the correct time and amplitute by the motor and the gears. The control variable is the
position of the piston. The target value for the piston position is an output of the ADDA PCB
as the motor target value. The „travel“ is defined by impulses which are measured by the shaft
encoder. The Incremental Encoder PCB determines, from these impulses, the actual piston
position value (motor actual value) as an analog voltage.
PEEP Controller
The PEEP pressure in the expiratory airway is generated by a PEEP valve which is supplied
with a variable control pressure between 0 and 40 mbar. A plunger-coil system, an essential
part of the PEEP control pressure potentiometer, is driven by a current source, namely the
PEEP controller. The target current corresponds to the analog value, PEEP-target, specified
by the ADDA PCB. This is adjusted in the controller to the actual current, PEEP-actual.
The task of the supply voltage monitor is to constantly compare all operating voltages (+5 V, +
15 V, - 15 V, + 24 V) to corresponding ranges of voltage deviations. If a supply voltage
deviates from the given tolerance, the signals for emergency off, horn and motor
disconnection are immediately generated.
For internal use only. Copyright reserved.
Valve Controller
The solenoid valves can only be switched if the EMERGENCY OFF signal is not active and
GBF5170XXXT03.fm 14.06.01
thus the voltage supply is guaranteed. All eight possible valves are enabled by transistor
stages by signals. The signal acknowledgements are tapped at the collectors of the
switching transistors via single-state transistor amplifiers.
The horn can be actuated by the ADDA PCB, the Monitoring PCB and by the supply voltage
monitor. It is supplied with power from the + 15V voltage or from the 9 V rechargeable
battery.
The driver ICs for RXD/DCD/TXD/DTR signals are found on the Motherboard PCB. The
signals RXD and TXD are generated by the CPU Standard 2 PCB, while DCD and DTR
originate on the ADDA PCB. The charging circuit for the 9 V rechargeable battery is likewise
located on the Motherboard PCB. It ensures that the rechargeable battery is charged
correctly.
Fan Controller
In the case of a failure in the compressed air supply, Divan will operate on oxygen. The fan
inhibits any oxygen leaking into the electronic area of Divan. The function of the fan is tested.
Opening the upper cover of the Divan reveals a right-hand cover plate under which three
PCBs are located. All three are of the plug-in type, with the left one being the CPU
Monitoring PCB.
The task of the CPU Monitoring PCB is to make sure the anaesthesia ventilator is functioning
properly. For this purpose, all signals relevant to equipment safety - both analog and digital -
are read in via 2 PIOs and an 8-bit A/D converter and evaluated by the microprocessor by
various digital control lines. The microprocessor is capable of influencing the equipment
hardware, in each case, so as to make for a safe equipment status in the event of a fault.
The CPU 6802 of CPU Monitoring PCB works indepedently and in parallel with the ADDA
PCB/CPU Standard 2; it is however connected with the latter for the purposes of exchanging
information.
System errors and faults are detected by mutual checking of the measurement and control
data and result in shutdown.
The PCB-internal hardware is also checked and faults are recognized as long as the
microprocessor itself is not shut down. In such a situation, a watchdog - a monitoring logic - is
activated and generates not only a system reset, but also various emergency-off signals
For internal use only. Copyright reserved.
which shut down the motor and the valves and switch the Divan to Man./Spont.
The emergency-off signals and the horn switch-on are also generated by the software in the
event of a general fault.
GBF5170XXXT03.fm 14.06.01
Opening the upper cover of the reveals a right-hand cover under which three PCBs are
located. All are of the plug-in type, with the right-hand one being the CPU Standard 2 PCB.
ADDA PCB
Opening the upper cover of the reveals a right-hand cover under which three PCBs are
located. All are of the plug-in type, with the centre one being the ADDA PCB.
The name of this PCB is the abbreviation of the terms „A/D, D/A converter“. Several parallel
I/O ports are located on the PCB in addition to these elements. The signals supplied by the
sensors and controllers are converted on the ADDA PCB in such a manner that they can be
processed and evaluated by the microprocessor of the CPU Standard 2 PCB. The PCB also
serves to switch the valves, normalize the pressure signal of the airway pressure sensors for
the analog output and to generate the target values for the motor and PEEP controllers.
The Front Panel PCB is located behind the touch-sensitive controls/display unit and serves
as the main board for the Display and Alphanumeric PCBs. This assembly forms the interface
between keypad and display (Display PCB, LP Alphanumeric) and the microprocessor on
CPU Standard 2 PCB. The data coming from the processor is transferred to the Display
PCB in such a manner that it can be indicated on the 7-segment LED displays and on the
bargraph. Furthermore, the control signals for the alphanumeric display are generated here.
The key matrix of the touch-sensitive controls is made accessible to the processor by a
keypad encoder. The Front Panel PCB also features the actuator and the driver stages for the
key LEDs, as well as the evaluation circuitry for the incremental encoder.
The Incremental Encoder PCB is located - viewed from the front - behind the piston cylinder
unit in the area of the drive unit. It is mounted on the right-hand carrier plate above the drive
belt. The Pulse Generator PCB is located immediately at the motor of the drive and carries
out control functions with an incremental encoder. The Incremental Encoder PCB is used to
transfer the signals received from the incremental encoder attached to the piston drive unit
and serves to determine the position of the piston. The piston receives the absolute position
during the power-on test. The Incremental Encoder PCB supplies the output signal for the
actual value of the piston position.
For internal use only. Copyright reserved.
GBF5170XXXT03.fm 14.06.01
The Heater PCB is located under the heating plate of the compact breathing system. It is
accessible from below. The heater is switched ON/OFF by the contact pressure of the
compact breathing system.
Note:
The heater is operated with dc voltage. Polarity reversal can lead to destruction of the PCB.
1. Remove five hex socket-head screws from the rear panel. Firmly hold the rear panel
when removing the sixth hex socket-head screw. Risk of injury.
1 2
For internal use only. Copyright reserved.
GBF5170XXXT03.fm 14.06.01
4. Disconnect the GND connection to Divan on cover 1. Disconnect also the 3-pin heater
connector, 18-pin voltage connector and the 15-pin RS 232 connector (PM).
1
GND
5. Disconnect the hoses for pressure supply to Divan (1), for the measuring gas return (2),
and for the airway pressure sensor (3).
1 black/
white
2
For internal use only. Copyright reserved.
3
GBF5170XXXT03.fm 14.06.01
7. Remove the breathing system and the piston cylinder unit from Divan.
D
For internal use only. Copyright reserved.
Reinstalling Divan
2. Remove the five hex socket-head screws in the front protection guard and one hex
socket-head screw on the right-hand rail. Lift up and remove the protection guard.
3. Remove three screws on the front. Remove front from the mounte. Observe antistatic
precautions. Disconnect the 50-pin connector from the electronics module.
For internal use only. Copyright reserved.
GBF5170XXXT03.fm 14.06.01
5. Screw on safety valve mount as shown in the Figure. This is very important when Divan is
to operate with the front part folded down.
6. Remove two hex socket-head screws. Remove the electronics cover. Remove detached
board mount.
For internal use only. Copyright reserved.
L
GBF5170XXXT03.fm 14.06.01
1 2
8. Remove hex socket-head screws 1 and 2. Firmly hold cover when loosening screw 3.
1 2
3
9. Fold down the cover with the motherboard an the printed circuit boards. Caution:
Observe antistatic precautions.
For internal use only. Copyright reserved.
GBF5170XXXT03.fm 14.06.01
H/L-signal
ambient air
Debug interface
RS 232
BUS
RS 232
RS 232
O2 side stream serial data
+ 5V
Anaesthetic gas +12V
CO2 Andros RS 232 ±15V Power pack
MV bank
N2O
11/93
Addressbus
Data
Control
O2 side stream
BUS
Flow
PCB BUS PCB Front
Temp
Functional Description
Measured (location 5)
O2 main stream Value
(location 2)
O2 side stream serial bit streams
EL-Display
PCB Motherboard H/L-signal
Keyboard
Pressure Analog
Pressure Module H/L-signal
Page 105
There are four measurement channels. Two channels are used for the conventional
measurement using an O2 sensor capsule. One of the remaining two channels is provided for
inspiratory and expiratory O2 measurement in the sidestream (intake, optical measurement
bench). Different amplification factors (v = 23.5 or v = 11.75) allow a multiplexer to detect any
short circuit between the two measurement channels. The measurement cell provides a
voltage between 9.3 mV and 21.6 mV at 21% by vol. and a voltage between 45 mV and 105
mV at 100% by volume. A sensor test is carried out after starting the unit. For this purpose, a
cell is loaded via a resistor (1kOhm) and the response of the second cell is checked
(Teledyne test).
Temperature measurement
An NTC sensor is used to measure the respiratory gas temperature. This sensor has the
following specifications:
The signals of the NTCs are processed via a temperature hybrid. It provides an output voltage
(1.4333 V at 30°C and 0.7535 V at 41°C) which is supplied to a multiplexer.
Pressure measurement
The pressure module is located on the motherboard. It generates a pressure signal which is
transmitted to a multiplexer on the Measured Value PCB. The pressure module provides a
linear output voltage between 0.110 V and 2.450 V for a pressure range between -20 mbar
and 100 mbar. The operating voltage of the pressure module is +5 V. The pressure module
consists of a differential pressure sensor which measures against ambient air. This
compensates for potential drifts. DrägerService performs a zero calibration at 6-month
intervals during the inspection procedure.
For internal use only. Copyright reserved.
GBF5170XXXT04.fm 14.06.01
A Spirolog sensor is used for measurement of the flow. It works according to the principle of
the hot-wire flowmeter. The sensor forms a bridge between two resistors R94 and R86. The
operation amplifier (OP) N7 forms a closed-loop control system with the transistor V17 which
keeps the temperature of the hot branch at a constant level (125 °C at a flow = 0). If heat of
the hot wire is dissipated by a certain flow, the resistance in the hot wire will change. This will
detune the jumper. As a result, the control increases the current and compensates for the
heat loss.
A working resistor with a narrow tolerance is located in the hot branch of the bridge. The
voltage decrease in this branch (caused by the heating current) is a non-linear measure for
the flow. Amplified by a factor of 4, the voltage is used to adjust the sensor adjustment;
amplified by a factor of 2 it is used as a measurement signal for the flow. The cold branch of
the bridge is used for temperature compensation. The Spirolog sensor is subject to
production tolerances and ageing tolerances and must therefore be calibrated. Calibration is
carried out with a D/A converter which is controlled via a microprocessor. The D/A converter
applies a voltage to the resistor R94 which is proportional to the measurement voltage. This
resistor is connected in parallel with the cold wire and makes sure the total resistance in the
cold branch is variable. Impurities on the platinum wires can be eliminated by annealing the
wires. Annealing is achieved by detuning the jumper in the cold branch such that a high
limited current is allowed to flow through the wires. A N2O compensation can be performed
by a zero calibration with air.
V17
R68-71
N6
cold hot
FI-Low
N8
N7
FI-High
cleaning N9
R94 R86
For internal use only. Copyright reserved.
N10
D/A
N11
GBF5170XXXT04.fm 14.06.01
PM 8050 cd and Divan are switched ON/OFF with the main power switch.
When switching the system on using the main power switch, PM 8050 cd will automatically
initiate a self-test (cold start).
PM 8050 cd
Version 1.0
26.03.92
Self test
For internal use only. Copyright reserved.
GBF5170XXXT04.fm 14.06.01
screen keys
measured-value
graphics section section
The alarms of the ventilator are displayed on PM 8050 cd during communication with Divan.
This is to improve perception of the alarm messages.
VENT INOP
Priority 25 (alarm)
VENT INOP
Priority 25 (alarm)
VENT INOP
Priority 25 (alarm)
PRESS LIMIT
Priority 11 (caution) / 7 (advisory)
FRESH GAS?
Priority 11 (caution)
The following configuration screen appears at the end of the power-on test.
Calibrate sensors!
Test ventilator!
The anaesthetic gases available as well as messages such as „Vapor not present“ or „Vapor
not identified“ are listed in the „anaesth. gas“ column. Only three Vapors (halothane,
enflurane, isoflurane) are presently identified automatically. Once a specific Vapor has been
identified automatically, it is no longer possible to change the anaesthetic manually. Question
marks in the display prompt the user to calibrate or to carry out a ventilator test.
The mode of operation of Divan is transmitted every 3 seconds and each time the operating
mode is changed. This allows PM 8050 cd to identify which alarm limits have to be activated.
When changing from IPPV to Man/Spont, for example, the lower pressure monitoring
(disconnection), the upper O2 monitoring and the lower anaesthetic concentration
monitoring are automatically deactivated. These alarm limits are reactivated when changing
from Man/Spont to IPPV. In both cases the changes are indicated by a flashing mode
display and by an audible alarm. No acknowledgement is required.
For internal use only. Copyright reserved.
GBF5170XXXT04.fm 14.06.01
If fSIMV is < 6/min then Divan will switch over to the man/spont. mode.
If fSIMV is > 6/min then Divan will switch over to the IPPV mode.
Standby mode:
If Divan enters the standby mode, PM 8050CD will do the same.
...Standby...
PM 8050 cd
Version 1.0
26.03.92 Cancel
trend
Config.
MAN
Manual ventilation:ventilator set to SPONT
No manual ventilation is possible in the standby mode. Ventilation with the manual bag is only
possible after selecting the man/spont. mode!
GBF5170XXXT04.fm 14.06.01
• the main power switch and the power switch of PM 8050 cd are switched on,
• the voltage supply to Divan is available, and
• the mains voltage fails simultaneously.
Each unit generates its own power failure alarm. If the power fails for a longer period, the
alarm remains active for a maximum of 30 seconds (± 10 seconds). A new power failure alarm
will be generated only after removing the cause for the first power failure. This can be
achieved by switching the main power switch ON/OFF, or if the system voltage is available
again.
• the main power switch is switched OFF/ON, or if a power failure lasts more than 120
seconds (± 5 seconds). In this case, PM 8050 cd and Divan will start the boot and the
self-test routine.
• the power failure lasts less than 120 seconds (± 5 seconds). The most recent screen
content is displayed after a warm start. The picture setup on PM 8050 cd starts approx.
20 seconds after that of Divan.
Different conditions may occur after a power failure as both units have tolerances in the
power-on behaviour.
PM 8050 cd detects that the communication to Divan has been interrupted. It tries to re-
establish communication and to log onto Divan again. If Divan is ready-to-receive,
communication will be established during a cold start. In this case, it is not necessary to
calibrate the sensors. After completion or interruption of the Divan self-test, Divan will
synchronize the modes of operation of the two units.
GBF5170XXXT04.fm 14.06.01
After a cold start or warm start, Divan suppresses the „communication interrupted“ message
for 60 seconds. After approx. PM 8050 cd logs onto Divan after approx. 30 to 50 seconds.
Divan transmits its current mode of operation to PM 8050 cd thereby interrupting the power-
on test of the monitor. The sensor do not need to be calibrated.
For internal use only. Copyright reserved.
GBF5170XXXT04.fm 14.06.01
The optical measurement system (Andros bench 4610) consists of the measurement bench
and the relevant electronics. Furthermore, a pump module is required for delivery of the
sample gas, while a changeover valve is required for zeroing.
Block diagram:
Andros bench
from
patient
optical
valve measurement electronics I/O PCB
ambient air system
Power
pump supply
unit
outlet port
The sample gas (e.g. 200 mL/min) is drawn by the pump of the Andros bench. The gas flows
through a T-piece with filter and the sampling hose to the connection of the water trap. From
there it flows into a filter unit. The major portion of the gas (180 mL/mins) flows through the
Goretex diaphragm which removes the humidity. This gas portion is supplied to the cuvette
and is measured there. The remaining 20 mL/min are used to transport the collected liquid
into specific containers.
The advisory message „CO2 Line” on PM 8050 cd points out that the filter unit might be
contaminated. In this case, it is necesssary to replace the filter unit (see Instructions for Use
under „Replacing the filter unit”).
A contamination of the filter unit is also perceptible by a discoloration (black) of the Goretex
For internal use only. Copyright reserved.
diaphragm which is clearly visible from below. The drawn patient gas can be fed back into the
circle system via the optional measurement gas return line.
GBF5170XXXT04.fm 14.06.01
Measurement: The pump module supplies the sample gas to the measurement bench
(switchable from 60 to 200 mL/min). The gas is analyzed by using the infrared absorption
technique. For this purpose, different filters (reference filter, anasthetic gas filter, N2O filter,
CO2 filter and black filters) are positioned in the beam path by a filter wheel.
measured
values
filter wheel
emitter 5 filters cuvette detector
For internal use only. Copyright reserved.
The associated electronics uses these signals to generate the values for CO2, N2O and the
anaesthetic gases (halothane, enflurane, isoflurane, as well as desflurane and sevoflurane).
These values are transmitted for each single ventilation to the I/O PCB via the RS-232
interface. From the I/O PCB, they are then transmitted to the CPU PCB. To achieve an
accurate CO2 measurement, it is necessary to perform an O2 correction. This is achieved by
GBF5170XXXT04.fm 14.06.01
transferring the O2 values of PM 8050 cd to the measurement bench. Only +12V are
required as voltage supply.
The power supply unit provides the necessary supply voltages for PM 8050 cd:
• +5 V
• +15 V
• +12 V
• -15 V
The PM 8050 cd power supply unit is electrically isolated from the Cato power supply. If the
operating voltage of the Cato changes (mains voltage), no changes must be implemented in
the power supply unit of PM 8050 cd. The power supply unit can be exchanged as a
complete assembly.
The CPU PCB is the host computer of the PM 8050 cd and performs the central tasks. The
PCB provides the bus for communication with the other PCBs. The CPU PCB accomodates
one Motorola 68000 microprocessor which can be switched between 8 and 16 MHz. The
memory consists of a 256-kB battery-backed RAM, a 256-kB unbacked RAM and a 2-MB
EPROM. Furthermore, the PCB contains a real-time clock with integrated battery-backed
2K x 8-bit RAM.
The Measured Value PCB has the task to measure the signals „airway pressure, expiratory
flow, inspiratory and expiratory oxygen concentration, and respiratory gas temperature“ and
to use them to compute and monitor important parameters (e.g. Peak, PEEP, Plat, Vt, AMV). It
consists of the measured value computer and the circuit elements for determination of the
parameters mentioned above. The printing module is not located on the Measured Value
PCB but on the motherboard. The Measured Value PCB has a Z-80 microprocessor which
communicates with the CPU PCB via a dual port RAM (1k x 8 bit). The Z-80 calculates the
flow, the pressure, the temperature, and the O2 values. The values are made available to the
Z-80 via a multiplexer, a Sample&Hold level, and a downstream A/D converter.
The SPO2 PCB is not part of the basic equipment but is available as an option. It determines
the parameters oxygen saturation and pulse rate. It communicates the I/O PCB via a serial
interface.
GBF5170XXXT04.fm 14.06.01
The Front PCB is responsible for the screen output, the generation of acoustic and optical
alarms, as well as for keyboard scanning. The PCB accomodates a graphics processor
(TMS 34010) which drives the EL display (EL = electroluminescent).
The I/O PCB is used for communication of the CPU PCB with peripheral devices such as
optical measurement bench, Divan (via Cicero PCB), SpO2 module, two serial interfaces and
one analog interface. The analog interface and the external RS-232 interfaces are electrically
isolated (2.5 kV) from the remaining equipment electronics through optocouplers. In
addition, the I/O PCB provides the functions for horn control, battery charge, temperature
monitoring and +5 V monitoring (digital).
PM 8050 cd communicates with Divan via the RS 232 interface. The modes of operation, the
settings, the compliance and the alarm messages are transmitted in co-ordination with Divan.
The electrical signal isolation of 2.5 kV from the equipment and other interfaces is ensured
by optocouplers. The voltage for the interface is supplied from Divan. If PM 8050 cd is
connected to Divan, then the contacts X5 pin 8 and pin 15 are bridged. A MOSFET (see
circuit diagram of Cicero PCB) is blocked. The main power switch is connected in series with
the internal power switch of PM 8050 cd. Since the internal power switch is always switched
on , PM 8050 cd can only be switched on/off via the main power switch. In the event of a
power failure, the position of this switch determines whether the power failure logic of the
I/O PCB trips an alarm. If both units are not interconnected the bridge between the contacts
X5 pin 8 and pin 15 is not available. The MOSFET becomes conductive via the voltage
TVBAT 2 thus causing the internal power switch to become active. In the event of a power
failure, its switching condition determines whether the power failure logic of the I/O PCB
trips an alarm. The three reflection light barriers are clocked with the signal VAPOUT. The
Vapor logic identifies the type of Vapor and transmits this information to the CPU PCB via the
I/O PCB.
Display
The display is mounted on the front frame. It is an EL display (EL = electroluminescent). The
respective electronics board is connected with the EL display. The display consists of a
640 x 400 electrode array. Light is emitted if an electrode is triggered with ac voltage via a
lines/columns combination. The electronics board provides the power supply, high voltage
driver and controller modules. The operating voltages for the display are + 5 V and + 12 V.
For internal use only. Copyright reserved.
GBF5170XXXT04.fm 14.06.01
0 I
D
2. The rear of the monitor is secured by screws to the trolley. Remove the two cheesehead
screws.
3. Note the position of the hoses for later reinstallation. Disconnect the hoses from the gas
intake ports 1 and 2.
1
2
P
For internal use only. Copyright reserved.
4. Remove the monitor including all other connections from the mount. Disconnect the
sensor connections, where appropriate.
GBF5170XXXT04.fm 14.06.01
Use the reverse order to mount the PM 8050 cd. Follow these precautions:
• Make sure the hoses of the water trap do not get caught between the rear panel of the
monitor and the monitor mount.
• The power cable is placed in a sheet-metal bracket which is located on the rear of the
monitor mount. Inser the monitor halfway into the mount before mounting the power
cable.
For internal use only. Copyright reserved.
GBF5170XXXT04.fm 14.06.01
Message is cancelled
after calibration. The
self-test is continued
or the machine
switches to
GBF5170XXXT05.fm 14.06.01
»Man./Spont.« mode.
external humidifier.
Y-piece closed? Dialogue during the self-test. Close the Y-piece to
continue the self-test.
GBF5170XXXT05.fm 14.06.01
test«.
GBF5170XXXT05.fm 14.06.01
PULS SPO2 < !!! The pulse rate has exceeded upper
alarm limit.
SPO2 < !!! The oxygen saturation is below the Check ventilation.
lower alarm limit. Check the O2
concentration of the
fresh gas.
PAW > !!! The upper alarm limit for airway Check the hose
pressure has been exceeded, system on the
ventilation hose buckled, stenosis. anaesthesia unit.
Pressure limits set too high. Correct the pressure
Pressure limitation valve limits on the
disconnected. anaesthesia unit.
Switch on pressure
limitation on the
anaesthesia unit.
PM 8050cd INOP !!! Internal system fault. The machine is Switch off the
not operational. machine.
Call DrägerService.
For internal use only. Copyright reserved.
GBF5170XXXT05.fm 14.06.01
10.1 General
In the anaesthesia systems Cato and Cicero, the fresh gas flows to the patient through a
fresh gas hose and through the patient system. The A-cone is an additional fresh-gas outlet.
The A-cone and the fresh-gas outlet to the patient system is changed over via a valve.
Vaporizer
Patient system
10.2 Subassemblies
The changeover valve is toggled between the positions “patient system” and “A-cone” using
the button. If the A-cone mode is selected, the button lamp will come on. The actual position
of the changeover valve is determined by a proximity switch.
If a warm start of the Cato or Cicero occurs, e.g. due to a short-term power interruption, the
changeover valve will remain in the position selected last. In the event of a cold start, the
changeover valve will always select the position “patient system”. If the Divan switches over to
“IPPV”, the changeover valve will also select the position “patient system”.
The switching logic and the changeover valve control are located on the A-cone PCB. The A-
cone PCB also generates the 24 V operating voltage required for switching the changeover
valve.
The operating voltage of the A-cone PCB is 5 V. It is available at the Divan/Monitor cable
harness.
Voltage
generation
24 V
CAUTION:
Electrostatic discharge can damage electronic components. Use an ESD
protection mat and a wrist strap when handling electronic modules.
X1
Pin Assignment
7 +5 V
2 GND
C36
Pin Assignment
Plus (+) +24 V (generated on the A-cone PCB)
Minus (–) GND2
For internal use only. Copyright reserved.
GBF5170XXXT07.fm 14.06.01
X1
Pin Assignment
1 not assigned
2 GND
3 GND
4 Status (signal) (during a cold start, “0” for a short period)
5 Coding pin
6 Man/Spont (signal), 1 = IPPV
7 +5 V
X2
Pin Assignment
1 Lamp (GND)
2 GND
3 GND
4 Coding pin
5 Button (TastR) (button not pressed –> pin 3 and 5 connected)
6 Button (TastA) (button not pressed –> pin 2 and 6 open)
7 Lamp (+)
X3
Pin Assignment
1 GND
2 Valve position (0 = patient system, 1 = A-cone)
3 +24 V
For internal use only. Copyright reserved.
4 GND2 (+24 V)
5 Magvent2 (bistable solenoid valve)
6 Coding pin
GBF5170XXXT07.fm 14.06.01
X2
1
D5
N4
+ C37
C36
+
X1
Check
No No Divan-Monitor
5 V operating
Divan in the cable harness, check
voltage present
IPPV mode? fuse on Divan
(at X1)?
basic board.
Yes Yes
A-cone deactivated
in the IPPV mode. 24 V operating No
Replace
voltage present A-cone PCB.
(at C36)?
Yes
Open circuit in
No cable harness or
IPPV and status
signal set to "1"? possibly in the Divan
basic board.
Yes
Do button
contacts close and No
open? (Check ohmic Replace button.
resistance).
Yes
Button lamp No
OK? (Check ohmic
resistance, if Replace lamp.
necessary).
Yes
For internal use only. Copyright reserved.
Replace
changeover Yes Changeover No
valve (check valve po- valve control OK? Replace
sition if necessary). (Short pulse). A-cone PCB
GBF5170XXXT07.fm 14.06.01
I
0
8050
PM
er
Dr g
D
Copyright reserved. Reproduction allowed for non-commercial purposes only.
GBK5170XXXTL1.fm
This Test List does not replace inspections and servicing by Dräger Medical AG & Co. KGaA
This Test List does not replace inspections and servicing by Dräger Medical AG & Co. KGaA
3 Visual inspection
This Test List does not replace inspections and servicing by Dräger Medical AG & Co. KGaA
Compressed
air hose
Secretion suction
Shut-off valve
tubing
Contol valve
Manometer
Bacterial filter
Connecting
hose
Container cap
Open the AIR and the O2 gas supply. Open the waste
gas check valve.
Copyright reserved. Reproduction allowed for non-commercial purposes only.
This Test List does not replace inspections and servicing by Dräger Medical AG & Co. KGaA
MAN
MAN
increases slowly.
6 Divan
This Test List does not replace inspections and servicing by Dräger Medical AG & Co. KGaA
PM 8050 cd
Version 1.0
26.03.92
Self test
volu-
No anesth. other meter
gas
default values
Calibrate sensors!
Test ventilator!
GBK5170XXXTL1.fm
This Test List does not replace inspections and servicing by Dräger Medical AG & Co. KGaA
Acknowledge!
GBK5170XXXTL1.fm
This Test List does not replace inspections and servicing by Dräger Medical AG & Co. KGaA
This Test List does not replace inspections and servicing by Dräger Medical AG & Co. KGaA
LEDs
8.1 Basic power unit Serial no
O2
AIR
N 2O
GBK5170XXXTL1.fm
This Test List does not replace inspections and servicing by Dräger Medical AG & Co. KGaA
This Test List does not replace inspections and servicing by Dräger Medical AG & Co. KGaA
12 Alarm Valve
• N2O = 1 L/min.
This Test List does not replace inspections and servicing by Dräger Medical AG & Co. KGaA
Vapor is disconnected.
• Connect 1-bar manometer at the mixed gas con-
nection.
• Close the AIR and the N2O systems.
• Build up and maintain a pressure of
P = 100 mbar with the O2 control valve.
Vapor is connected. This test is carried out during the Divan self-test
sequence and is diplayed under „MAN leak“.
Copyright reserved. Reproduction allowed for non-commercial purposes only.
GBK5170XXXTL1.fm
This Test List does not replace inspections and servicing by Dräger Medical AG & Co. KGaA
16 ORC
O2/L/min N2O/L/min
0 0.6 − 0.8
Copyright reserved. Reproduction allowed for non-commercial purposes only.
This Test List does not replace inspections and servicing by Dräger Medical AG & Co. KGaA
O2/L/min N2O/L/min
0 0.6 − 0.8
0.8 1.8 − 2.4
1.5 3.3 − 4.6
3.0 6.8 − 9.0
16.2 S-ORC
O2/L/min N2O/L/min
0 0
0.2 +0.07/−0.05 N2O starts flowing
0.5 1.2 − 1.8
3.0 6.0 − 10
17 Vapor Identification
Copyright reserved. Reproduction allowed for non-commercial purposes only.
This Test List does not replace inspections and servicing by Dräger Medical AG & Co. KGaA
2 3
Interface connections:
• Sensor connections
• Flow sensor (1)
• O2 sensor (2)
• Temperature sensor (optional) (3)
• SpO2 sensor (optional) (4)
1 4
• Clean the fan mat / replace annually.
This Test List does not replace inspections and servicing by Dräger Medical AG & Co. KGaA
Settings:
• Connect a lung simulator, a training thorax or the
breathing bag to the Y-piece.
• Adjust the fresh gas.
• Press the IPPV key.
• Press the rotary knob.
Settings:
• V T = 0.6 L
• fIPPV = 12/min
• 25 mbar
• TI:TE = 1:1.7
• TIP:TI = 10%
Copyright reserved. Reproduction allowed for non-commercial purposes only.
• PEEP = 0 mbar
This Test List does not replace inspections and servicing by Dräger Medical AG & Co. KGaA
First display:
Copyright reserved. Reproduction allowed for non-commercial purposes only.
Second display:
IPPV leak = xx mL
This Test List does not replace inspections and servicing by Dräger Medical AG & Co. KGaA
This Test List does not replace inspections and servicing by Dräger Medical AG & Co. KGaA
Fuse-links:
Fuse-links: 4 x 2 AT
Fuse-links:
1 X 4 AT 110 V
1 X 2 AT 230 V
GBK5170XXXTL1.fm
This Test List does not replace inspections and servicing by Dräger Medical AG & Co. KGaA
23.5.1 PM 8050 cd
24 Vapor
26 Test performed
Date: _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ Name: _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _
GBK5170XXXTL1.fm
This Test List does not replace inspections and servicing by Dräger Medical AG & Co. KGaA
The O2 sensor must be replaced if the „use-by“ date has expired or O2 calibration is no
longer possible.
• Remove the O2 sensor housing from the anaesthesia ventilator.
• Unscrew the O2 sensor housing.
• Remove the O2 sensor.
Note:
The cooling air filter must be cleaned or replaced every 4 weeks. It must be replaced
minimally after 1 year.
• Remove the cooling air filter from the mount.
• Replace the cooling air filter or wash it in warm water with a small amount of detergent.
Dry well.
• Insert the cooling air filter into the mount avoiding folds.
P
For internal use only. Copyright reserved.
GBK5170XXXV01.fm 14.06.01
The NiCd battery ensures that the acoustic alarm sounds in the event of a power failure. The
battery must be replaced every 3 years.
Replacement procedure:
Caution !
4. The battery is located in a mount at the front of the PCB Motherboard. Remove the
battery from the mount. Insert the new battery. Make sure that the polarity is correct: plus
of the battery on plus of the mount.
Note:
Danger:
The NiCd battery ensures that the acoustic alarm sounds in the event of a power failure. The
battery must be replaced every 3 years.
Replacement procedure:
0 I
D
4. The monitor is secured by 2 screws to the rear of the trolley. Remove the 2 screws.
5. Note the arrangement of the water trap hoses for reinstallation. Disconnect the hoses
from the gas intake sockets (1+2).
For internal use only. Copyright reserved.
GBK5170XXXV01.fm 14.06.01
6. Turn the mains switch to the OFF position. Disconnect the remaining connections.
Remove the PM 8050 cd from the mount.
5
1
2
1 = O2 sensor
2 = Flow sensor
3 = Temperature sensor (optional)
4 = SpO2 sensor (optional)
5 = Port for anaesthetic gas return line or scavenging line
6 = Paw pressure measuring line
7 = Power cable
8 = Vapor identification
9 = Ventilator
Caution !
For internal use only. Copyright reserved.
LP CPU 1
LP Meßwert
LP SpO2 (Option)
P LP I/O
LP Front
LP Cicero 6
For internal use only. Copyright reserved.
GBK5170XXXV01.fm 14.06.01
Note:
Danger:
The bacterial filter must be replaced every 6 months. The pressure measuring line must be
replaced when damaged or minimally after 1 year.
1
For internal use only. Copyright reserved.
GBK5170XXXV01.fm 14.06.01
The piston cylinder unit should be checked and serviced at regular intervals since the roller
diaphragm may become leaky as a result of ageing and frequent sterilization. The steps to
replace the roller diaphragm half on the patient side are as follows:
Step 1:
Remove the piston cylinder unit from the ventilator while in the standby mode.
Step 2:
Release the piston cylinder unit, pull straight up and remove.
Step 3:
Loosen the 2 quick-release locks (1) with a 6-mm wrench and remove the cylinder head (2).
Step 4:
Grasp the piston and unscrew the central piston screw using the 6-mm wrench and remove
the upper part of the piston (3).
Step 5:
Replace the roller diaphragm half (4) on the patient side against a new one. To ensure a tight
seal, the embossed manufacturing clock must point towards the upper part of the piston (3).
Step 6:
Use the reverse order for reassembly. Make sure neither kinks nor folds are formed during
installation of the roller diaphragm. The complete roller diaphragm must form a typical U-fold
at both ends.
Step 7:
Carry out the Cato power-on test to check the function and the tight seal of the roller
diaphragm. The roller diaphragm must unroll without forming folds.
Note: It is not advisable that the roller diaphragm half on the equipment side, ie. the non-
contaminated part, be replaced by the user. It should rather be replaced by DrägerService
since its safe function can only be ensured if various screw locking agents and special tools
(tightening torques) are used.
For internal use only. Copyright reserved.
GBK5170XXXV01.fm 14.06.01
Legend
1. Quick-release locks
2. Cylinder head
3. Upper part of the piston
4. Roller diaphragm half
5. Manufacturing clock
For internal use only. Copyright reserved.
GBK5170XXXV01.fm 14.06.01
1 Type of Changes
Bitte hier für Inhaltsverz. klicken/ Please, press here for table of contens
Sach-Nr. Benennung
Part No. Description
M32040 CATO
CATO
M33010 GRUNDGERAET
CATO BASIC UNIT
6 KABELMONTAGE M32304
CABLE MOUNTING
12 MESSROEHRENBLOCK(02,AIR,N20) ( M33602
VORDERANSICHT)
MEASUR.TUBES BLOCK(O2,AIR,N2O) (FRONT VIEW)
18 O2-ANSCHLUSS-SCHLAUCH 1M M32651
O2-CONN.HOSE 1M,75 (DIN-DIN)
20 MESSGASRUECKFUEHRUNG M32595
MEASURING GAS FEED-BACK
23 ATEM-SCHLAUCHSET M30146
HOSE SET VENTILATION
24 OVALROHR-GELENKARM M34080
ARTICULATED ARM, CPL.
25 KOLBENZYLINDEREINHEIT M29575
CLOSED SYSTEM
26 FLASCHENPAKET 2M85056
SECRETION JAR SET
27 ABSPERRB.ABSAUGEEJEKTOR 2M85032
LOCKABLE ASPIR.EJECT.OXY.C.AIR
31 GRIFF,KOMPLETT M33012
HANDLE, CPL.
32 BASISPLATTE,VOLLST. M31973
BASE PLATE, CPL.
HALTER M29152
SUPPORT
KOMPAKT-ATEMSYSTEM M29880
COMPACT BREATHING SYSTEM
KOMPAKTATEMSYSTEM M32900
COMPACT BREATHING SYSTEM
ABLAGESCHALE 8408343
TRAY EVITA 2
RUESTSATZ-SCHIENE 8602090
MODIFICATION KIT PANEL
DIVAN US AF00360
DIVAN US
Datei/File: 5170000.001
2a DIVAN US AF00360
DIVAN US
3 LEERGEHAEUSE 8H M32965
EMPTY CASE 8H
4 ANSCHLUSSGEHAEUSE M32960
CONNECTION BOX
5 KLAMMER M24155
CLAMP
Datei/File: 5170000.002
2 HALTER M31963
SUPPORT
3 SENSORHALTER 2M19248
SENSOR HOLDER
4 WATERLOCK 6870511
DW - WASSERFALLE
WATER LOCK M
DW - WATERLOCK
4a WAL-12ER-SET 6870567
FÜR DW - WASSERFALLE
OHNE ABBILDUNG
WATERTRAPS SET OF 12
FOR DW - WATERLOCK
WITHOUT ILLUSTRATION
5a WASSERFALLE 8600571
OHNE ABBILDUNG
WATER SEPARATOR
WITHOUT ILLUSTRATION
5b AUFKLEBER 8600618
OHNE ABBILDUNG
ADHESIVE LABEL
WITHOUT ILLUSTRATION
6- 6b GELENKARMHALTER 8601400
SET-HINGED-BRACKET-HOLDER
6a BUCHSE M30295
SOCKET
Datei/File: 5170000.003
3 BOLZEN 8302380
BOLT
Datei/File: 5170000.004
1 PLATTE M29110
PLATE
5 PROFIL M29336
PROFILE
6 ABDECKKAPPE M29202
CAP
Datei/File: 5170000.005
Druckdatum: 30.10.2000 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division Dräger
Ersatzartikelliste 5170.000 Ausgabe/Edition
Spare parts list 27.10.00
CATO Seite/Page 16 von 74
CATO
2 STUTZEN 2M12754
CONNECTION PIECE
KABELMONTAGE Bild/Picture 6
CABLE MOUNTING
Datei/File: 5170000.006
7 KABELBINDER 8712007
CABLE CLIP (2,4X92)
8 PU-SCHLAUCH,1,5 8600513
PU-HOSE 1,5
9 KABELBINDER 8712065
CABLE CLIP
14 SCHUTZHUELSE 8402123
PROTECTING SLEEVE
Datei/File: 5170000.007
1 STECKANSCHLUSS M27265
PLUG-IN SYSTEM,HOUSING PART
4 UMSCHALTER M31775
SWITCH
5 TUELLE M25072
SOCKET
6 STIMMPFEIFE M25073
TUNING HAMMER
7 VAPORERKENNUNGSSENSOR M30340
EINFACH
VAPOR IDENTIFICATION SENSOR
SINGLE
8 VAPORERKENNUNGSSENSOR M32834
DOPPELT
SENSOR
DOUBLE
Datei/File: 5170000.008
3 O-RING U04314
O-RING SEAL
Datei/File: 5170000.009
9 WINKELSTECKANSCHLUSS M30935
OHNE ABBILDUNG
ANGLE CONNECTION
WITHOUT ILLUSTRATION
Datei/File: 5170000.010
1 SCHRAUBE 2M17338
SCREW
2 RUNDSCHNURRING M19241
TOROIDAL SEALING RING
3 SIEBEINSATZ M19238
MESH BOTTOM
4 STEUERBLOCK M32041
CONTROL BLOCK
5a O-RING R50514
O-RING SEAL
6 HALTER 2M17304
SUPPORT
8 STECKANSCHLUSS M30952
PLUG-TYPE CONNECTION
10 INJEKTOR M25062
INJECTOR
11 WINKEL-STECKANSCHLUSS M30953
ANGLE CONNECTION
17 DICHTRING R15422
OHNE ABBILDUNG
PACKING RING
Datei/File: 5170000.011
1 HALTENOCKEN M28935
FIXING CAM
2 O-RING U04314
O-RING SEAL
3 FEDER A10594
OHNE ABBILDUNG
SPRING
WITHOUT ILLUSTRATION
Datei/File: 5170000.012
Datei/File: 5170000.013
Datei/File: 5170000.014
6a KAPPE,WEISS M25146
OHNE ABBILDUNG
CAP,WHITE
WITHOUT ILLUSTRATION
7a FEINREGELVENTIL-EINBAUSATZ M32917
OHNE ABBILDUNG
PRECISION CONTROL VALVE KIT
WITHOUT ILLUSTRATION
10 MANOMETER M33402
PRESSURE GAUGE
12 SICHERHEITSVENTIL M21683
SAFETY VALVE
13 O-RING E20274
FÜR RÖHRCHEN M30299
OHNE ABBILDUNG
O-RING SEAL
FOR TUBE M30299
WITHOUT ILLUSTRATION
15 ISOLIERKASTEN M30293
INSULATING BOX
16 EINSCHRAUBANSCHLUSS M28816
OHNE ABBILDUNG
SCREW-IN CONNECTION
WITHOUT ILLUSTRATION
17 S-ORC M32940
RELIEF VALVE
18 SCHEIBE M30297
PANE
19 O-RING M19657
FÜR MESSRÖHREN
O-RING SEAL
FOR METER TUBES
20 FEDERLAGER M28705
SPRING HANGER
21 FEDER M12928
SPRING
22 STECKANSCHLUSS M30952
PLUG-TYPE CONNECTION
25 O-RING RM07823
O-RING SEAL
26 O-RING M30308
O-RING SEAL
28 WIDERSTAND O2 M27543
(25+26) RESISTOR O2
33 WINKELSTECKANSCHLUSS M30935
ANGLE CONNECTION
Datei/File: 5170000.015
Datei/File: 5170000.016
1 STEUERBLOCK M32041
CONTROL BLOCK
6 O-RING R50514
O-RING SEAL
7 DRUCKREGLER M34751
PRESSURE REGULATOR
8 SCHWELLWERTSCHALTER M32456
THRESHOLD SWITCH
9 3/2-WEGEVENTIL M31919
3/2-PORT DISTRIBUTING VALVE
9a O-RING M34832
FÜR 3/2 WEGE-VENTIL
OHNE ABBILDUNG
O-RING SEAL
FOR 3/2-PORT DISTRIBUTING VALVE
WITHOUT ILLUSTRATION
10 STECKANSCHLUSS M30952
PLUG-TYPE CONNECTION
11 WINKEL-STECKANSCHLUSS M30953
ANGLE CONNECTION
12 RUECKSCHLAGVENTIL M31918
OHNE ABBILDUNG
NONRETURN VALVE
WITHOUT ILLUSTRATION
Datei/File: 5170000.017
Druckdatum: 30.10.2000 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division Dräger
Ersatzartikelliste 5170.000 Ausgabe/Edition
Spare parts list 27.10.00
CATO Seite/Page 41 von 74
CATO
4 DICHTRING M05128
PACKING RING
O2-ANSCHLUSS-SCHLAUCH 1M Bild/Picture 18
O2-CONN.HOSE 1M,75 (DIN-DIN)
Datei/File: 5170000.018
Druckdatum: 30.10.2000 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division Dräger
Ersatzartikelliste 5170.000 Ausgabe/Edition
Spare parts list 27.10.00
CATO Seite/Page 43 von 74
CATO
4 DICHTRING M07152
PACKING RING
Datei/File: 5170000.019
Druckdatum: 30.10.2000 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division Dräger
Ersatzartikelliste 5170.000 Ausgabe/Edition
Spare parts list 27.10.00
CATO Seite/Page 45 von 74
CATO
4 DICHTRING M07152
PACKING RING
MESSGASRUECKFUEHRUNG Bild/Picture 20
MEASURING GAS FEED-BACK
Datei/File: 5170000.020
Druckdatum: 30.10.2000 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division Dräger
Ersatzartikelliste 5170.000 Ausgabe/Edition
Spare parts list 27.10.00
CATO Seite/Page 47 von 74
CATO
1 BAKTERIENFILTER 8402868
BACTERIA FILTER
2 LUER-LOCK.MAENNL. 8600214
LUER-LOCK, MALE - PM8020
3 GEHAEUSE M31848
HOUSING
4 O-RING 2M10296
O-RING SEAL
5 KAPPE M21482
CAP
Datei/File: 5170000.021
Druckdatum: 30.10.2000 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division Dräger
Ersatzartikelliste 5170.000 Ausgabe/Edition
Spare parts list 27.10.00
CATO Seite/Page 49 von 74
CATO
4 SCHNALLRIEMEN M19101
BUCKLE STRAP
Datei/File: 5170000.022
Druckdatum: 30.10.2000 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division Dräger
Ersatzartikelliste 5170.000 Ausgabe/Edition
Spare parts list 27.10.00
CATO Seite/Page 51 von 74
CATO
1 BEUTEL 2120992
REPLACEMENT BAG INFANT
4 HALTERING 2120178
RETAINING RING / CLAMP
14 BEUTEL 2120968
REPLACEMENT BAG BABY
15 BEUTEL 2120119
BAG RESUSCITATOR
ATEM-SCHLAUCHSET Bild/Picture 23
HOSE SET VENTILATION
Datei/File: 5170000.023
Druckdatum: 30.10.2000 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division Dräger
Ersatzartikelliste 5170.000 Ausgabe/Edition
Spare parts list 27.10.00
CATO Seite/Page 53 von 74
CATO
OVALROHR-GELENKARM Bild/Picture 24
ARTICULATED ARM, CPL.
Datei/File: 5170000.024
1 SENSORHALTER M34081
SENSOR HOLDER
2 KABELFUEHRUNG M34082
CABLE BUSHING
3 KLEMMHEBEL 8602177
CLAMPING LEVER
6 KLEMMHUELSE M5 8602179
SPLIT TAPER SOCKET M5
7 KLEMMHUELSE 8602210
SPLIT TAPER SOCKET
8 SCHRAUBE 8602178
SCREW
KOLBENZYLINDEREINHEIT Bild/Picture 25
CLOSED SYSTEM
Datei/File: 5170000.025
3 O-RING M31019
O-RING SEAL
4 ZYLINDER M29587
CYLINDER
5a ROLLMEMBRAN M29664
DIAPHRAGM
5b ZYLINDEREINHEIT M29688
CYLINDER UNIT
6 KOLBENKOPF M29960
OHNE ABBILDUNG
PISTON TOP
WITHOUT ILLUSTRATION
6a DICHTUNG M27721 5
OHNE ABBILDUNG
GASKET
WITHOUT ILLUSTRATION
7 KOLBENSTANGE M32891
PISTON ROD
8 ZYLINDER M29586
CYLINDER
9 ZUGANKER M31652
TIE ROD
10 LIPPENDICHTUNG 8407628
LIP SEAL
11 FEDER 8400496
OHNE ABBILDUNG
SPRING
WITHOUT ILLUSTRATION
12 KOLBENHALTER M29602
OHNE ABBILDUNG
PISTON SUPPORT
Druckdatum: 30.10.2000 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division Dräger
Ersatzartikelliste 5170.000 Ausgabe/Edition
Spare parts list 27.10.00
CATO Seite/Page 58 von 74
CATO
13 ARRETIERUNGSBLECH M29603
OHNE ABBILDUNG
LOCKING DISK
WITHOUT ILLUSTRATION
14 REPARATURSET AF00358
BEINHALTET M29590 UND M29605
OHNE ABBILDUNG
REPAIR SET
COMPRISED M29590 AND M29605
WITHOUT ILLUSTRATION
15 RAT-KOLBENZYLINDEREINHEIT M30869
REP.EXCH.CLOSED SYSTEM (M29575
FLASCHENPAKET Bild/Picture 26
SECRETION JAR SET
Datei/File: 5170000.026
1 WINKELTUELLE 2M19063
ANGULAR PORCELAIN BUSH
2 SEKRETFLASCHENDECKEL 2M19065
COVER FOR SECRETION BOTTLE
3 HUELSE 2M19064
BUSH
4 SCHWIMMER M26007
FLOAT
5 MUFFE M26008
BUSHING
6 ABSAUGESCHLAUCH M25780
ASPIRATION HOSE
7 SEKRETSCHAUGLAS M07582
SECRETION CONTROL GLASS
8 BEHAELTER 2M85594
RECEPTACLE
9 TRAGEGESTELL 2M85380
SUPPORTING FRAME
ABSPERRB.ABSAUGEEJEKTOR Bild/Picture 27
LOCKABLE ASPIR.EJECT.OXY.C.AIR
Datei/File: 5170000.027
1 ABSPERRVENTIL 2M85031
STOP VALVE,CPL.
2 ABSAUGEEJEKTOR 2M85019
ASPIRATION EJECTOR
6 UEBERWURFSCHRAUBE M09088
MALE FITTING FOR PIPES
7 SCHALTER,KPL. 2M85609
SWITCH,CPL.
8 DICHTRING M07142
OHNE ABBILDUNG
PACKING RING
WITHOUT ILLUSTRATION
9 DICHTRING M07143
OHNE ABBILDUNG
PACKING RING
WITHOUT ILLUSTRATION
10 DICHTRING M07144
OHNE ABBILDUNG
PACKING RING
WITHOUT ILLUSTRATION
12 SIEBEINSATZ D11521
OHNE ABBILDUNG
MESH BOTTOM
WITHOUT ILLUSTRATION
14 DICHTRING-SET D20215
OHNE ABBILDUNG
SET OF PACKING RINGS
WITHOUT ILLUSTRATION
15 MANOMETER,ES/MS 7278090
PRESSURE GAUGE
16 SCHALLDAEMPFER 2M85034
SOUND ABSORBER
17 ABSPERRB.ABSAUGEEJEKTOR 2M85622
OHNE ABBILDUNG
LOCKABLE ASPIR.EJECT.3BAR
WITHOUT ILLUSTRATION
Datei/File: 5170000.028
1 STECKANSCHLUSS M30952
OHNE ABBILDUNG
PLUG-TYPE CONNECTION
WITHOUT ILLUSTRATION
3 O-RING D18400
OHNE ABBILDUNG
O-RING SEAL
WITHOUT ILLUSTRATION
4 O-RING E20274
O-RING SEAL
6 SICHERHEITSVENTIL 8602028
OHNE ABBILDUNG
SAFETY VALVE
WITHOUT ILLUSTRATION
7 SCHLAUCHHALTER M34160
OHNE ABBILDUNG
HOSE SUPPORT
WITHOUT ILLUSTRATION
9 LEITUNG,UGR. M17934
OHNE ABBILDUNG
LINE
WITHOUT ILLUSTRATION
10 RING R10664
OHNE ABBILDUNG
RING
WITHOUT ILLUSTRATION
11 SCHLAUCHSCHELLE RM08365
OHNE ABBILDUNG
HOSE CLIP
WITHOUT ILLUSTRATION
Datei/File: 5170000.029
Druckdatum: 30.10.2000 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division Dräger
Ersatzartikelliste 5170.000 Ausgabe/Edition
Spare parts list 27.10.00
CATO Seite/Page 68 von 74
CATO
1 HALTER 2M17304
SUPPORT
2 SCHRAUBE 2M17288
SCREW
3 SCHRAUBE 2M17338
SCREW
4 BLECH 2M17305
SHEET METAL
5 KLINKE 2M17303
CATCH
Datei/File: 5170000.030
Druckdatum: 30.10.2000 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division Dräger
Ersatzartikelliste 5170.000 Ausgabe/Edition
Spare parts list 27.10.00
CATO Seite/Page 70 von 74
CATO
3 HALTER M31963
SUPPORT
GRIFF,KOMPLETT Bild/Picture 31
HANDLE, CPL.
Datei/File: 5170000.031
2 FOLIENTASTATUR,VOLLST. 8201346
KEY PAD, LEFT
3 FOLIENTASTATUR,VOLLST. 8201336
KEY PAD, RIGHT
BASISPLATTE,VOLLST. Bild/Picture 32
BASE PLATE, CPL.
Datei/File: 5170000.032
Druckdatum: 30.10.2000 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division Dräger
Ersatzartikelliste 5170.000 Ausgabe/Edition
Spare parts list 27.10.00
CATO Seite/Page 74 von 74
CATO
1 HALTER 2M17304
SUPPORT
2 KAPPE M27256
CAP
3 KAPPE,WEISS M25146
CAP,WHITE
4 DICHTRING R15422
PACKING RING
6 SCHRAUBE 2M17338
SCREW
Sach-Nr. Benennung
Part No. Description
M31880 DIVAN
DIVAN
2 DIVAN M31880
DIVAN
3 DIVAN M31880
DIVAN
4 FRONTPLATTE M29680
FRONT PANEL
5 HEIZUNG M29973
HEATING
6 LUEFTER M31533
FAN
7 SCHNITTSTELLE-BREASY M31690
INTERFACE - BREASY
8 ANSCHLUSSPLATTE M31691
CONNECTING PLATE
9 ANTRIEBSEINHEIT M31461
DRIVE ASSEMBLY
10 ANTRIEBSEINHEIT M31461
DRIVE ASSEMBLY
11 ANTRIEBSEINHEIT M31461
DRIVE ASSEMBLY
12 KOLBENSTANGEN-VERRIEGELUNG M31535
PISTON ROD LOCK
13 ELEKTRONIK-PNEUMATIK M31881
ELECTRONIC PNEUMATIC
14 ELEKTRONIK-PNEUMATIK M31881
ELECTRONIC PNEUMATIC
15 ELEKTRONIK-PNEUMATIK M31881
ELECTRONIC PNEUMATIC
16 PNEUMATIK M31688
PNEUMATIC
17 PNEUMATIK M31688
PNEUMATIC
18 PNEUMATIKBLOCK M31687
PNEUMATIC BLOCK
DIVAN Bild/Picture 1
DIVAN
Datei/File: 5170010.001
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5170.010 Ausgabe/Edition
Spare parts list 12.12.97
DIVAN Seite/Page 6 von 54
DIVAN
DIVAN Bild/Picture 2
DIVAN
Datei/File: 5170010.002
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5170.010 Ausgabe/Edition
Spare parts list 12.12.97
DIVAN Seite/Page 7 von 54
DIVAN
DIVAN Bild/Picture 3
DIVAN
Datei/File: 5170010.003
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5170.010 Ausgabe/Edition
Spare parts list 12.12.97
DIVAN Seite/Page 8 von 54
DIVAN
1 GEHAEUSE,VOLLST. M31882
HOUSING, CPL.
3 DECKEL,VOLLST. M32681
COVER, CPL.
6 ABDECKUNG M29636
COVER
8 TYPENSCHILD 8303877
TYPE PLATE
9 ABDECKBLECH M29541
COVER PLATE
11 SCHIENE,KOMPLETT M34157
RAIL, CPL.
14 BLECH M29942
SHEET METAL
16 ABSTANDSHALTER M29941
DISTANCE SPREADER
20 KREUZHALTERUNG 8404154
CROSS SUPPORT
22 SCHACHTABDECKUNG M29638
SHAFT COVER
25 WINKEL 8300391
ANGLE
31 KAPPE M27256
CAP
32 SCHLAUCHHALTER M34160
HOSE SUPPORT
33 KLOTZ M34158
BLOCK
FRONTPLATTE Bild/Picture 4
FRONT PANEL
Datei/File: 5170010.004
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5170.010 Ausgabe/Edition
Spare parts list 12.12.97
DIVAN Seite/Page 11 von 54
DIVAN
1 DREHKNOPF M29954
CONTROL KNOB
2 BEDIENFELD M29557
CONTROL PANEL
2a BEDIENFELD M34770
(PCV-Divan)
CONTROL PANEL
(PCV-Divan)
4 BUCHSE M29784
SOCKET
7 ABSTANDSHALTER 8305279
SPACER
10 FIBERRING D04766 50
FIBRE RING
12 IMPULSGEBER,VOLLST. 8200957
PULSE GENERATOR,CPL.
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5170.010 Ausgabe/Edition
Spare parts list 12.12.97
DIVAN Seite/Page 12 von 54
DIVAN
14 WINKEL M29569
ELBOW
16 BLECH M30719
SHEET METAL
HEIZUNG Bild/Picture 5
HEATING
Datei/File: 5170010.005
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5170.010 Ausgabe/Edition
Spare parts list 12.12.97
DIVAN Seite/Page 14 von 54
DIVAN
1 ABSAUGSTUTZEN,UGR. M30533
ASPIRATION SOCKET
2 O-RING M29902 5
O-RING SEAL
3 DICHTUNG M27721 5
GASKET
4 GRUNDPLATTE,UGR. M30537
BASE PLATE
5 HEIZPLATTE,UGR. M30862
HOT PLATE
7 VLIES M30529
MAT
8 TRAEGERPLATTE M30524
SUPPORT PANEL
10 FEDER R21629 5
SPRING
11 DISTANZBOLZEN 8200641
DISTANCE BOLT
15 BLECH M30538
SHEET METAL
18 KAPPE M30688
CAP
19 FEDER R24863
SPRING
20 RIEGEL M30534
BOLT
21 BODENBLECH M30532
BASE PLATE
22 BODENBLECH M30523
BASE PLATE
23 ISOLATIONS-FILM M31679
INSULATION FILM
26 ISOLATIONS-FILM M31679
INSULATION FILM
LUEFTER Bild/Picture 6
FAN
Datei/File: 5170010.006
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5170.010 Ausgabe/Edition
Spare parts list 12.12.97
DIVAN Seite/Page 17 von 54
DIVAN
3 AMP-STIFT 8301954
AMP-PIN
SCHNITTSTELLE-BREASY Bild/Picture 7
INTERFACE - BREASY
Datei/File: 5170010.007
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5170.010 Ausgabe/Edition
Spare parts list 12.12.97
DIVAN Seite/Page 19 von 54
DIVAN
1 ANSCHLUSSPLATTE M31691
CONNECTING PLATE
3A DICHTUNG 8404065 5
GASKET
4 GEWINDEZAPFEN M29940
THREADED PLUG
6 MITNEHMER M29545
DRIVER
7 KUPPLUNGSSTIFT 1299859
COUPLING PIN
8 BUCHSE M29601
SOCKET
9 WELLE M29546
SHAFT
10 SCHEIBE 8402392 10
WASHER
11 HEBEL M29536
LEVER
14 SCHLAUCHKLEMME M31681
HOSE CLAMP
ANSCHLUSSPLATTE Bild/Picture 8
CONNECTING PLATE
Datei/File: 5170010.008
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5170.010 Ausgabe/Edition
Spare parts list 12.12.97
DIVAN Seite/Page 22 von 54
DIVAN
2 ANSCHLUSSPLATTE M31811
CONNECTING PLATE
3 O-RING M32804 10
O-RING SEAL
4 TUELLE 8400964 10
SOCKET
5 SCHLAUCHVERSCHRAUBUNG 8402032 5
THREADED HOSE COUPLING
ANTRIEBSEINHEIT Bild/Picture 9
DRIVE ASSEMBLY
Datei/File: 5170010.009
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5170.010 Ausgabe/Edition
Spare parts list 12.12.97
DIVAN Seite/Page 24 von 54
DIVAN
ANTRIEBSEINHEIT Bild/Picture 10
DRIVE ASSEMBLY
Datei/File: 5170010.010
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5170.010 Ausgabe/Edition
Spare parts list 12.12.97
DIVAN Seite/Page 25 von 54
DIVAN
ANTRIEBSEINHEIT Bild/Picture 11
DRIVE ASSEMBLY
Datei/File: 5170010.011
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5170.010 Ausgabe/Edition
Spare parts list 12.12.97
DIVAN Seite/Page 26 von 54
DIVAN
1 SEITENTEIL M31307
SIDE PANEL
2 SEITENTEIL M31308
SIDE PANEL
3 ABSTANDSBOLZEN M29503
DISTANCE BOLT
7 ZAHNRIEMEN M31499
TOOTHED BELT
8 ANTRIEBSABDECKUNG M29574
DRIVE COVER
9 FUEHRUNG M29699
GUIDE
12 ZAHNRIEMEN M31498
TOOTHED BELT
13 KOLBENSTANGEN-VERRIEGELUNG M31535
PISTON ROD LOCK
15 WELLE,VOLLST. M34384
SHAFT, CPL.
16 WELLE,UGR. M31458
SHAFT
18 ANTRIEBSSCHEIBE M31496
DRIVE WHEEL
20 WINKEL M30805
ELBOW
21 V-GABELLICHTSCHRANKE NG 1810294
ohne Abbildung
V LIGHT BARRIER
without illustration
27 DREHGEBER,KOMPL. M34380
SHAFT ENCODER, CPL.
KOLBENSTANGEN-VERRIEGELUNG Bild/Picture 12
PISTON ROD LOCK
Datei/File: 5170010.012
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5170.010 Ausgabe/Edition
Spare parts list 12.12.97
DIVAN Seite/Page 29 von 54
DIVAN
1 MITNEHMER M31507
DRIVER
3 HEBEL,UGR. M29625
LEVER
4 DRUCKBOLZEN M29621
PRESSURE BOLT
6 KRALLE M29624
CLUTCH
8 SICHERUNGSBLECH M29663
SAFETY PLATE
10 SPANNSTUECK M29627
MOUNTING PIECE
ELEKTRONIK-PNEUMATIK Bild/Picture 13
ELECTRONIC PNEUMATIC
Datei/File: 5170010.013
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5170.010 Ausgabe/Edition
Spare parts list 12.12.97
DIVAN Seite/Page 31 von 54
DIVAN
ELEKTRONIK-PNEUMATIK Bild/Picture 14
ELECTRONIC PNEUMATIC
Datei/File: 5170010.014
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5170.010 Ausgabe/Edition
Spare parts list 12.12.97
DIVAN Seite/Page 32 von 54
DIVAN
ELEKTRONIK-PNEUMATIK Bild/Picture 15
ELECTRONIC PNEUMATIC
Datei/File: 5170010.015
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5170.010 Ausgabe/Edition
Spare parts list 12.12.97
DIVAN Seite/Page 33 von 54
DIVAN
1 PNEUMATIK M31688
PNEUMATIC
2 KABELSCHELLE 2M14128
CABLE CLAMP
8 KABELBAUM 5 NG 8200686
CABLE HARNESS 5
11 IC MC 6802 1809156
INTEGRATED CIRCUIT MC 6802
12 KREUZHALTERUNG 8404154
CROSS SUPPORT
21 FIBERRING D04766 50
FIBRE RING
29 NIPPEL 8301823
NIPPLE
PNEUMATIK Bild/Picture 16
PNEUMATIC
Datei/File: 5170010.016
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5170.010 Ausgabe/Edition
Spare parts list 12.12.97
DIVAN Seite/Page 36 von 54
DIVAN
PNEUMATIK Bild/Picture 17
PNEUMATIC
Datei/File: 5170010.017
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5170.010 Ausgabe/Edition
Spare parts list 12.12.97
DIVAN Seite/Page 37 von 54
DIVAN
1 EINSCHUBBLECH M31686
PLUG-IN PLATE
5 KABELDURCHFUEHRUNG CH17949
CABLE BUSHING
6 BUCHSE 8408312
SOCKET
7 T-TUELLE 6800187 5
T-SOCKET
9 PNEUMATIKBLOCK M31687
PNEUMATIC BLOCK
10 GIMETALL-LAGER 2M06157 5
RUBBER METAL CONNECTION
14 3/2-WEGE-VENTIL M29566
3/2-PORT DISTRIBUTING VALVE
15 SCHWENKVERSCHRAUBUNG 8402034
SWIVELLING SCREW-FITTING
16 KABELBAUM 6 8201597
CABLE HARNESS 6
18 REDUZIERSTUECK M22286 5
REDUCER
19 T-STUECK 8401083
T-PIECE
21 PLATTE M31689
PLATE
23 PLATTE M30552
PLATE
27 BUCHSE 8301453
SOCKET
28 HALTEBLECH M30684
SUPPORT PLATE
29 SICHERHEITSVENTIL M31821
SAFETY VALVE
30 BUEGEL CH13155
CLIP
33 SICHERHEITSVENTIL M31822
SAFETY VALVE
35 PEEP-VENTIL 8406516
PEEP VALVE
40 HALTER M31684
SUPPORT
PNEUMATIKBLOCK Bild/Picture 18
PNEUMATIC BLOCK
Datei/File: 5170010.018
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5170.010 Ausgabe/Edition
Spare parts list 12.12.97
DIVAN Seite/Page 41 von 54
DIVAN
2 PNEUMATIKBLOCK M31682
PNEUMATIC BLOCK
3 TUELLE 8400964 10
SOCKET
4 SCHLAUCHVERSCHRAUBUNG 8402032 5
THREADED HOSE COUPLING
5 O-RING M32804 10
O-RING SEAL
Datei/File: 5170010.019
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5170.010 Ausgabe/Edition
Spare parts list 12.12.97
DIVAN Seite/Page 43 von 54
DIVAN
1 BOLZEN M28790
BOLT
2 RUNDSCHNURRING M12165 10
TOROIDAL SEALING RING
3 SCHEIBE M28791
DISC
4 TUELLE M28785
SOCKET
4A TUELLE M29930
SOCKET
SICHERHEITSVENTIL Bild/Picture 20
SAFETY VALVE
Datei/File: 5170010.020
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5170.010 Ausgabe/Edition
Spare parts list 12.12.97
DIVAN Seite/Page 45 von 54
DIVAN
1 VENTILGEHAEUSE M31820
VALVE HOUSING
3 FEDER 8403504
SPRING
4 EINSTELLSCHRAUBE M31819
ADJUSTING SCREW
5 MUTTER 8300656
NUT
Datei/File: 5170010.021
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5170.010 Ausgabe/Edition
Spare parts list 12.12.97
DIVAN Seite/Page 47 von 54
DIVAN
2 ABDECKBLECH M32682
COVER PLATE
3 SCHLITZMUTTER T04896
SLOTTED ROUND NUT
6 HALTERUNG M29938
SUPPORT
7 HALTERUNG M29937
SUPPORT
8 ABDECKUNG M30950
COVER
9 BLECH M29311
SHEET METAL
11 SCHARNIER M29632
HINGE
13 SCHARNIERBEFESTIGUNG M29633
HINGE FIXING
17 MAGNETVERSCHLUSS 2M10046
MAGNETIC LOCK
Datei/File: 5170010.022
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5170.010 Ausgabe/Edition
Spare parts list 12.12.97
DIVAN Seite/Page 49 von 54
DIVAN
Datei/File: 5170010.023
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5170.010 Ausgabe/Edition
Spare parts list 12.12.97
DIVAN Seite/Page 50 von 54
DIVAN
Datei/File: 5170010.024
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5170.010 Ausgabe/Edition
Spare parts list 12.12.97
DIVAN Seite/Page 51 von 54
DIVAN
1 GEHAEUSE,UGR. M29535
HOUSING
2 DRUCKANSCHLUSS,UGR. M29698
PRESSURE CONNECTION
3 DICHTUNG 8404065 5
GASKET
5 LUEFTERBLECH M29544
FAN SHEET METAL
10 HALTER,UGR. M29573
SUPPORT
11 HALTER,UGR. M29572
SUPPORT
14 SCHIENE M29654
RAIL
15 VERSCHLUSS-STOPFEN G60455 10
VENT PLUG
17 ROHR M29653
TUBE
18 SCHEIBE M29565 2
WASHER
20 DURCHFUEHRUNG NG 2M12831
DUCT
22 ABDECKBLECH M31837
COVER PLATE
24 BODENBLECH M32904
BASE PLATE
25 BODENBLECH M31836
BASE PLATE
26 ANTRIEBSEINHEIT M31461
DRIVE ASSEMBLY
27 ELEKTRONIK-PNEUMATIK M31881
ELECTRONIC-PNEUTATIC
28 SCHNITTSTELLE-BREASY M31690
INTERFACE - BREASY
29 FUEHRUNG M29552
CONDUCTING
33 PUFFER 6800478 20
BUFFER
34 KABELSCHELLE 2M14128
CABLE CLAMP
35 ANSCHLUSSNIPPEL 8404506
CONNECTION NIPPLE
36 BLECH M29542
SHEET METAL
37 KABELDURCHFUEHRUNG CH17949
CABLE BUSHING
39 KAPPE M27256
CAP
40 GEWINDEBOLZEN M31817
HINGE BOLT
43 HEIZUNG M29973
HEATING
44 DOSIERUNG 8403795
DOSAGE
47 WINKEL M29951
ELBOW
48 GEHAEUSE M27548
HOUSING
49 SCHLAUCHVERSCHRAUBUNG 8402032 5
THREADED HOSE COUPLING
50 REDUZIERNIPPEL 8402582
REDUCTING NIPPLE
51 DRUCKREGLER 8402745
PRESSURE REGULATOR
52 SIEBEINSATZ D02316 5
MESH BOTTOM
53 FILTRATIONSVORSATZ 8405241 5
FILTERING ADAPTER
54 DICHTRING U01346 10
PACKING RING
55 EINSCHRAUBANSCHLUSS M28816
SCREW-IN CONNECTION
60 SCHRAUBE M30871
SCREW
61 KABEL M30654
CABLE
62 WASSERFALLE,UGR. M30723
WATER TRAP
63 MESSANSCHLUSS,UGR. M30725
MEASURING CONNECTION
Sach-Nr. Benennung
Part No. Description
8201650 SPANNUNGSVERSORGUNG,VOLLST.
POWER SUPPLY,CPL.
2 SPANNUNGSVERSORGUNG,VOLLST. 8201650
POWER SUPPLY, CPL.
3 SPANNUNGSVERSORGUNG,VOLLST. 8201650
POWER SUPPLY, CPL.
SPANNUNGSVERSORGUNG,VOLLST. Bild/Picture 1
POWER SUPPLY, CPL.
Datei/File: 5170011.001
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5170.011 Ausgabe/Edition
Spare parts list 14.05.96
SPANNUNGSVERSORGUNG, VOLLST. Seite/Page 5 von 8
POWER SUPPLY, CPL.
SPANNUNGSVERSORGUNG,VOLLST. Bild/Picture 2
POWER SUPPLY, CPL.
Datei/File: 5170011.002
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5170.011 Ausgabe/Edition
Spare parts list 14.05.96
SPANNUNGSVERSORGUNG, VOLLST. Seite/Page 6 von 8
POWER SUPPLY, CPL.
SPANNUNGSVERSORGUNG,VOLLST. Bild/Picture 3
POWER SUPPLY, CPL.
Datei/File: 5170011.003
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5170.011 Ausgabe/Edition
Spare parts list 14.05.96
SPANNUNGSVERSORGUNG, VOLLST. Seite/Page 7 von 8
POWER SUPPLY, CPL.
3 NETZKABELDURCHFUEHRUNG 2M18642
MAINS CABLE BUSHING
4 NETZLEITUNG, 6M 2M13126
SUPPLY MAIN, 6M
6 SICHERUNGSHALTER 8300611
FUSE HOLDER
10 SICHERUNGSHALTER 8300611
FUSE HOLDER
12 NETZTRAFO 8201666
MAINS TRANSFORMER
13 RUECKWAND 8201638
BACK PANEL
22 RS ABSCHALTVERZOEGER(CICERO EM 8602093
MODIF.KIT DECELERAT.(CICERO EM
Sach-Nr. Benennung
Part No. Description
M29880 KOMPAKT-ATEMSYSTEM
COMPACT BREATHING SYSTEM
2 ATEMGASBLOCK,KOMPL. M30429
BREATHING GAS BLOCK, CPL.
3 ATEMGASBLOCK,KOMPL. M30429
BREATHING GAS BLOCK, CPL.
4 ATEMGASBLOCK,KOMPL. M30429
BREATHING GAS BLOCK, CPL.
8 APL-VENTIL M31381
APL-VALVE
9 DRUCKKNOPFVENTIL M30433
VALVE
11 ANSCHLUSS M31674
CONNECTION
13 APL-VENTIL M32499
APL-VALVE
KOMPAKT-ATEMSYSTEM Bild/Picture 1
BREATHING SYSTEM CICERO
Datei/File: 51311.001
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.1 Ausgabe/Edition
Spare parts list 10.02.98
KOMPAKT-ATEMSYSTEM Seite/Page 5 von 30
COMPACT BREATHING SYSTEM
1 STEUERBLOCK,KOMPL. M30392
CONTROL BLOCK,COMPL.
2 ATEMGASBLOCK,KOMPL. M30429
BREATHING GAS BLOCK, COMPL.
4 6-KT.SCHRB.DREH.DWN622-B6X100 1640623
HEXAG.SCR.DRIVER DWN622-B6X100
ATEMGASBLOCK,KOMPL. Bild/Picture 2
BREATHING GAS BLOCK, CPL.
Datei/File: 51311.002
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.1 Ausgabe/Edition
Spare parts list 10.02.98
KOMPAKT-ATEMSYSTEM Seite/Page 7 von 30
COMPACT BREATHING SYSTEM
ATEMGASBLOCK,KOMPL. Bild/Picture 3
BREATHING GAS BLOCK, CPL.
Datei/File: 51311.003
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.1 Ausgabe/Edition
Spare parts list 10.02.98
KOMPAKT-ATEMSYSTEM Seite/Page 8 von 30
COMPACT BREATHING SYSTEM
ATEMGASBLOCK,KOMPL. Bild/Picture 4
BREATHING GAS BLOCK, CPL.
Datei/File: 51311.004
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.1 Ausgabe/Edition
Spare parts list 10.02.98
KOMPAKT-ATEMSYSTEM Seite/Page 9 von 30
COMPACT BREATHING SYSTEM
3 APL-VENTIL M30411
APL-VALVE
3 APL-VENTIL M31381
APL-VALVE
3 APL-VENTIL M32499
APL-VALVE
4 UEBERWURFMUTTER M30453
UNION NUT
5 SCHAUGLAS M09230
CONTROL GLASS
6 KAPPE M21482
CAP
8 ZENTRIERSTIFT M30398
CENTERING PIN
9 RUECKSCHLAGVENTIL M32458
NONRETURN VALVE
11 ZUGANKER,UGR. M30428
TIE ROD
13 FEDER E20215
SPRING
14 KONTERSCHRAUBE M30404
COUNTER SCREW
16 ATEMGASBLOCK M30444
BREATHING GAS BLOCK
16 ATEMGASBLOCK M31619
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.1 Ausgabe/Edition
Spare parts list 10.02.98
KOMPAKT-ATEMSYSTEM Seite/Page 10 von 30
COMPACT BREATHING SYSTEM
18 SCHRAUBE 1321668
SCREW
19 DICHTRING M09231
PACKING RING
21 VENTILTELLER M23225 3
VALVE DISK
22 O-RING 2M08777
O-RING SEAL
23 ZUSATZLUFTVENTIL M32459
ADDITIONAL AIR VALVE
24 FUEHRUNGSBLECH,LINKS M30395
GUIDE PLATE, LEFT
26 FUEHRUNGSBLECH,RECHTS M30394
GUIDE PLATE, RIGHT
27 O-RING M30405
O-RING SEAL
29 GASFUEHRUNG,UGR. M30416
GAS FLOW CONTROL
30 ATEMBEUTELANSCHLUSS,UGR. M30426
BREATHING BAG CONNECTION
30 ANSCHLUSS M31674
CONNECTION
31 GEWINDEEINSATZ M30399
THREADED INSERT
33 SENSORGEHAEUSE M30396
SENSOR HOUSING
36 ABSORBERDECKEL,VOLLST. M30460
ABSORBER TOP, CPL.
Datei/File: 51311.005
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.1 Ausgabe/Edition
Spare parts list 10.02.98
KOMPAKT-ATEMSYSTEM Seite/Page 13 von 30
COMPACT BREATHING SYSTEM
2 ABSORBERTOPF M29994
ABSORBER POT
3 DICHTRING M30563
PACKING RING
4 KUGELKETTE,UGR. M30562
BEAD CHAIN
5 LINS.ZYL.SCHRB.AM4X6DIN85-A4 1315838 20
OVAL HEAD SCREW AM4X6DIN85-A4
Datei/File: 51311.006
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.1 Ausgabe/Edition
Spare parts list 10.02.98
KOMPAKT-ATEMSYSTEM Seite/Page 15 von 30
COMPACT BREATHING SYSTEM
1 STEUERBLOCK M30379
CONTROL BLOCK
1 STEUERBLOCK M31613
CONTROL BLOCK
2 VERSCHLUSSOBERTEIL M29648
SEAL TOP
3 TELLERFEDER M29056
PLATE SPRING
4 VERSCHLUSSBOLZEN M29644
LOCKING BOLT
6 GRIFF M30382
HANDLE
7 GEWINDEBUCHSE M30380
THREADED BOLT
8 DRUCKKNOPFVENTIL M30433
PUSH-BUTTON VALVE
10 RUNDSCHNURRING D16558
TOROIDAL SEALING RING
11 RING M30383
RING
12 MEMBRAN,UGR. M30462 2
DIAPHRAGM
13 BUCHSE M30384
SOCKET
14 TELLER 8407797
PLATE
15 DICHTSCHEIBE 8407979
SEALING WASHER
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.1 Ausgabe/Edition
Spare parts list 10.02.98
KOMPAKT-ATEMSYSTEM Seite/Page 16 von 30
COMPACT BREATHING SYSTEM
17 STIFT M31614
PIN
19 O-RING R22467
O-RING SEAL
20 SCHRAUBE M30385
SCREW
21 VENTILPLATTE M30381
VALVE REED
23 O-RING M30430
O-RING SEAL
24 O-RING R50438
O-RING SEAL
26 O-RING 2M10296
O-RING SEAL
27 O-RING M29384
O-RING SEAL
Datei/File: 51311.007
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.1 Ausgabe/Edition
Spare parts list 10.02.98
KOMPAKT-ATEMSYSTEM Seite/Page 18 von 30
COMPACT BREATHING SYSTEM
1 ABSORBERDECKEL M30459
ABSORBER TOP
2 O-RING R50313
O-RING SEAL
3 RUNDSCHNURRING R18352
TOROIDAL SEALING RING
6 LIPPENDICHTUNG M30455
LIP SEAL
7 DICHTRING M30456
PACKING RING
APL-VENTIL Bild/Picture 8
APL-VALVE
Datei/File: 51311.009
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.1 Ausgabe/Edition
Spare parts list 10.02.98
KOMPAKT-ATEMSYSTEM Seite/Page 20 von 30
COMPACT BREATHING SYSTEM
1 UEBERWURFMUTTER M30448
UNION NUT
2 HEBEL M31460
LEVER
4 DREHKNOPF M31384
CONTROL KNOB
5 DICHTRING NG M13459
PACKING RING
6 KAPPE M30594
CAP
7 SPINDEL M31304
SPINDLE
8 TELLER M31383
PLATE
9 SICHERUNGSRING NG M31305
CIRCLIP
10 FEDER NG M31385
SPRING
11 WELLE M31301
SHAFT
12 SICHERUNGSSCHEIBE M31303
LOCK WASHER
13 FEDER NG M30657
SPRING
14 DAEMPFUNGSELEMENT M31372
DAMPING UNIT
15 VENTILTELLER M31382
VALVE DISK
DRUCKKNOPFVENTIL Bild/Picture 9
VALVE
Datei/File: 51311.010
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.1 Ausgabe/Edition
Spare parts list 10.02.98
KOMPAKT-ATEMSYSTEM Seite/Page 22 von 30
COMPACT BREATHING SYSTEM
1 DRUCKKNOPFSTIFT M30432
PUSH-BUTTON PIN
2 FEDER V06499
SPRING
3 VENTILGEHAEUSE M30431
VALVE HOUSING
4 O-RING M30430
O-RING SEAL
5 O-RING R50438
O-RING SEAL
Datei/File: 51311.011
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.1 Ausgabe/Edition
Spare parts list 10.02.98
KOMPAKT-ATEMSYSTEM Seite/Page 24 von 30
COMPACT BREATHING SYSTEM
1 GEHAEUSE,ATEMBEUTELANSCHLUSS M30421
HOUSING,BREATH. BAG CONNECTION
2 O-RING R50121
O-RING SEAL
3 FRISCHGASANSCHLUSS NG M30422
CONNECTION
5 O-RING M30405
O-RING SEAL
6 INSPIRATIONSTUELLE NG M30424
INSPIRATION SOCKET
7 ISO-ANSCHLUSS NG M30423
ISO CONNECTION
8 O-RING R26807
O-RING SEAL
ANSCHLUSS Bild/Picture 11
CONNECTION
Datei/File: 51311.012
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.1 Ausgabe/Edition
Spare parts list 10.02.98
KOMPAKT-ATEMSYSTEM Seite/Page 26 von 30
COMPACT BREATHING SYSTEM
1 GEHAEUSE M31564
HOUSING
2 INSPIRATIONSTUELLE M31673
INSPIRATION SOCKET
3 ISO-ANSCHLUSS NG M30423
ISO CONNECTION
Datei/File: 51311.013
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.1 Ausgabe/Edition
Spare parts list 10.02.98
KOMPAKT-ATEMSYSTEM Seite/Page 28 von 30
COMPACT BREATHING SYSTEM
1 SCHLIESSRING TV 2M06548
CLOSING RING
2 TUELLE M30417
SOCKET
3 UEBERWURFMUTTER M30418
UNION NUT
4 O-RING 2M08777
O-RING SEAL
APL-VENTIL Bild/Picture 13
APL-VALVE
Datei/File: 51311.014
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.1 Ausgabe/Edition
Spare parts list 10.02.98
KOMPAKT-ATEMSYSTEM Seite/Page 30 von 30
COMPACT BREATHING SYSTEM
1 UEBERWURFMUTTER M30448
UNION NUT
2 HEBEL M31460
LEVER
4 DREHKNOPF M32882
CONTROL KNOB
5 DICHTRING NG M13459
PACKING RING
6 KAPPE M30594
CAP
7 SPINDEL M31304
SPINDLE
8 TELLER M31383
PLATE
9 SICHERUNGSRING M34474
SEALING RING
10 FEDER NG M31385
SPRING
11 WELLE M32885
SHAFT
12 SICHERUNGSSCHEIBE M31303
LOCK WASHER
13 FEDER NG M30657
SPRING
14 DAEMPFUNGSELEMENT M31372
DAMPING UNIT
15 VENTILTELLER M32498
VALVE DISK
16 O-RING R22467
O-RING SEAL
Sach-Nr. Benennung
Part No. Description
M29500 DIVAN
DIVAN
2 GEHAEUSE,VOLLST. M31840
HOUSING, COMPL.
3 GEHAEUSE,VOLLST. M31840
HOUSING, COMPL.
4 GEHAEUSE,VOLLST. M31840
HOUSING, COMPL.
5 HEIZUNG M29973
HEATING
6 LUEFTER M29676
FAN
7 LUEFTER M31533
FAN
8 SCHNITTSTELLE-BREASY M29658
INTERFACE - BREASY
9 SCHNITTSTELLE-BREASY M31690
INTERFACE - BREASY
10 ANSCHLUSSPLATTE M31691
CONNECTING PLATE
11 ANTRIEBSEINHEIT M29980
DRIVE ASSEMBLY
12 ANTRIEBSEINHEIT M31461
DEVICE ASSEMBLY
13 ANTRIEBSEINHEIT M31461
DRIVE ASSEMBLY
14 ANTRIEBSEINHEIT M31461
DRIVE ASSEMBLY
15 KOLBENSTANGENVERRIEGELUNG M29620
PISTON ROD LOCK
16 KOLBENSTANGEN-VERRIEGELUNG M31535
PISTON ROD LOCK
17 ELEKTRONIK-PNEUMATIK M29660
ELECTRONIC-PNEUMATIC
18 ELEKTRONIK-PNEUMATIK M29660
ELECTRONIC-PNEUMATIC
19 ELEKTRONIK-PNEUMATIK M29660
ELECTRONIC-PNEUMATIC
21 ELEKTRONIK-PNEUMATIK M31839
ELECTRONIC-PNEUMATIC
22 ELEKTRONIK-PNEUMATIK M31839
ELECTRONIC-PNEUMATIC
23 PNEUMATIK M29670
PNEUMATIC
24 PNEUMATIK M29670
PNEUMATIC
25 PNEUMATIK M29670
PNEUMATIC
26 PNEUMATIK M31688
PNEUMATIC
27 PNEUMATIK M31688
PNEUMATIC
28 PNEUMATIKBLOCK M31687
PNEUMATIC BLOCK
29 TUELLE,VOLLST. M30777
SOCKET,CPL.
30 SICHERHEITSVENTIL M29672
SAFETY VALVE
31 SICHERHEITSVENTIL M31822
SAFETY VALVE
32 DECKEL,VOLLST. M29630
COVER, CPL.
33 FRONTPLATTE M29680
FRONT PANEL
DIVAN Bild/Picture 1
DIVAN
Datei/File: 51312.001
1 GEHAEUSE,VOLLST. M29690
Schnittstelle M29694 vorhanden
HOUSING COMPLETE
interface M29694 is existing
1 GEHAEUSE,VOLLST. M31840
Schnittstelle M 29694 entfällt
HOUSING, COMPL.
interface M29694 is cancelled
3 DECKEL,VOLLST. M29630
COVER, CPL.
4 FRONTPLATTE M29680
FRONT PANEL
6 ABDECKUNG M29636
COVER
8 TYPENSCHILD 8303877
TYPE PLATE
9 ABDECKBLECH M29541
COVER PLATE
11 SCHIENE,KOMPLETT M34157
RAIL, CPL.
14 BLECH M29942
SHEET METAL
16 ABSTANDSHALTER M29941
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 8 von 79
DIVAN
20 KREUZHALTERUNG 8404154
CROSS SUPPORT
22 SCHACHTABDECKUNG M29638
SHAFT COVER
30 KLOTZ M34158
BLOCK
31 SCHLAUCHHALTER M34160
HOSE SUPPORT
GEHAEUSE,VOLLST. Bild/Picture 2
HOUSING, COMPL.
Datei/File: 51312.004
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 10 von 79
DIVAN
GEHAEUSE,VOLLST. Bild/Picture 3
HOUSING, COMPL.
Datei/File: 51312.005
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 11 von 79
DIVAN
GEHAEUSE,VOLLST. Bild/Picture 4
HOUSING, COMPL.
Datei/File: 51312.006
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 12 von 79
DIVAN
1 GEHAEUSE,UGR. M29535
HOUSING
2 DRUCKANSCHLUSS,UGR. M29698
PRESSURE CONNECTION
3 DICHTUNG 8404065 5
GASKET
5 LUEFTERBLECH M29544
FAN SHEET METAL
10 HALTER,UGR. M29573
SUPPORT
11 HALTER,UGR. M29572
SUPPORT
14 SCHIENE M29654
RAIL
15 VERSCHLUSS-STOPFEN G60455 10
VENT PLUG
17 ROHR M29653
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 13 von 79
DIVAN
18 SCHEIBE M29565 2
WASHER
19 TRENNBLECH,UGR. M29695
SEPARATING PLATE
20 DURCHFUEHRUNG NG 2M12831
DUCT
22 ABDECKBLECH M31837
COVER PLATE
24 BODENBLECH M32904
BASE PLATE
25 BODENBLECH M31836
BASE PLATE
26 ANTRIEBSEINHEIT M31461
Zahnriemen ist schwarz
DRIVE ASSEMBLY
toothed belt is black
27 ELEKTRONIK-PNEUMATIK M31869
ELECTRONIC-PNEUMATIC
27 RAT-ELEKTRONIK-PNEUMAT.(M31869 M31868
REP.EXCH.ELECTRON.PNEUM(M31869
28 SCHNITTSTELLE-BREASY M31690
Schnittstelle M29694 entfällt
INTERFACE - BREASY
interface M29694 is cancelled
29 FUEHRUNG M29552
CONDUCTING
33 PUFFER 6800478 20
BUFFER
34 KABELSCHELLE 2M14128
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 14 von 79
DIVAN
35 ANSCHLUSSNIPPEL 8404506
CONNECTION NIPPLE
36 BLECH M29542
SHEET METAL
37 KABELDURCHFUEHRUNG CH17949
CABLE BUSHING
39 KAPPE M27256
CAP
40 GEWINDEBOLZEN M31817
HINGE BOLT
43 HEIZUNG M29973
HEATING
44 DOSIERUNG 8403795
DOSAGE
47 WINKEL M29951
ELBOW
48 GEHAEUSE M27548
HOUSING
49 SCHLAUCHVERSCHRAUBUNG 8402032 5
THREADED HOSE COUPLING
50 REDUZIERNIPPEL 8402582
REDUCTING NIPPLE
51 DRUCKREGLER M29561
PRESSURE REGULATOR
52 SIEBEINSATZ D02316 5
MESH BOTTOM
53 FILTRATIONSVORSATZ 8405241 5
FILTERING ADAPTER
55 EINSCHRAUBANSCHLUSS M28816
SCREW-IN CONNECTION
60 SCHRAUBE M30871
SCREW
61 KABEL M30654
CABLE
62 WASSERFALLE,UGR. M30723
WATER TRAP
63 MESSANSCHLUSS,UGR. M30725
MEASURING CONNECTION
HEIZUNG Bild/Picture 5
HEATING
Datei/File: 51312.007
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 17 von 79
DIVAN
1 ABSAUGSTUTZEN,UGR. M30533
ASPIRATION SOCKET
2 O-RING M29902 5
O-RING SEAL
3 DICHTUNG M27721 5
GASKET
4 GRUNDPLATTE,UGR. M30537
BASE PLATE
5 HEIZPLATTE,UGR. M30862
HOT PLATE
7 VLIES M30529
XYZ
8 TRAEGERPLATTE M30524
SUPPORT PANEL
10 FEDER R21629 5
SPRING
11 DISTANZBOLZEN 8200641
DISTANCE BOLT
15 BLECH M30538
SHEET METAL
18 KAPPE M30688
CAP
19 FEDER R24863
SPRING
20 RIEGEL M30534
BOLT
21 BODENBLECH M30532
BASE PLATE
22 BODENBLECH M30523
BASE PLATE
23 ISOLATIONS-FILM M31679
INSULATION FILM
LUEFTER Bild/Picture 6
FAN
Datei/File: 51312.008
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 20 von 79
DIVAN
3 AMP-STIFT 8301954
AMP-PIN
LUEFTER Bild/Picture 7
FAN
Datei/File: 51312.009
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 22 von 79
DIVAN
3 AMP-STIFT 8301954
AMP-PIN
SCHNITTSTELLE-BREASY Bild/Picture 8
INTERFACE - BREASY
Datei/File: 51312.010
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 24 von 79
DIVAN
3 DICHTUNG 8404065 5
GASKET
4 GEWINDEZAPFEN M29940
THREADED PLUG
6 MITNEHMER M29545
DRIVER
7 KUPPLUNGSSTIFT 1299859
COUPLING PIN
8 BUCHSE M29601
SOCKET
9 WELLE M29546
SHAFT
10 SCHEIBE 8402392 10
WASHER
11 HEBEL M29536
LEVER
12 STIFT M29691
PIN
SCHNITTSTELLE-BREASY Bild/Picture 9
INTERFACE - BREASY
Datei/File: 51312.011
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 26 von 79
DIVAN
1 ANSCHLUSSPLATTE M31691
CONNECTING PLATE
3a DICHTUNG 8404065 5
GASKET
4 GEWINDEZAPFEN M29940
THREADED PLUG
6 MITNEHMER M29545
DRIVER
7 KUPPLUNGSSTIFT 1299859
COUPLING PIN
8 BUCHSE M29601
SOCKET
9 WELLE M29546
SHAFT
10 SCHEIBE 8402392 10
WASHER
12 STIFT M29691
PIN
14 SCHLAUCHKLEMME M31681
HOSE CLAMP
ANSCHLUSSPLATTE Bild/Picture 10
CONNECTING PLATE
Datei/File: 51312.012
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 29 von 79
DIVAN
2 ANSCHLUSSPLATTE M31811
CONNECTING PLATE
3 O-RING M32804 10
O-RING SEAL
4 TUELLE 8400964 10
SOCKET
5 SCHLAUCHVERSCHRAUBUNG 8402032 5
THREADED HOSE COUPLING
ANTRIEBSEINHEIT Bild/Picture 11
DRIVE ASSEMBLY
Datei/File: 51312.013
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 31 von 79
DIVAN
1 SEITENTEIL M31307
SIDE PANEL
2 SEITENTEIL M31308
SIDE PANEL
3 ABSTANDSBOLZEN M29503
DISTANCE BOLT
7 ZAHNRIEMEN 2 M29527
TOOTHED BELT 2
8 ANTRIEBSABDECKUNG M29574
DRIVE COVER
9 FUEHRUNG M29699
GUIDE
15 KUGELLAGERAUFNAHME M29963
BALL BEARING RECEPTACLE
16 WELLE M29969
SHAFT
18 HALTER M29970
SUPPORT
21 ZAHNSCHEIBE M29519
LIFTER DISC
23 KUGELLAGERAUFNAHME M29964
BALL BEARING RECEPTACLE
25 WELLE M29967
SHAFT
26 ZAHNSCHEIBE M29968
LIFTER DISC
27 MOTOR M29515
MOTOR
28 ANTRIEBSSCHEIBE M30804
DRIVING DISC
30 WINKEL M30805
ELBOW
35 KOLBENSTANGENVERRIEGELUNG M29620
PISTON ROD LOCK
38 DREHGEBER M30683
SHAFT ENCODER
39 HAUBE M30686
HOOD
40 SET-DREHGEBER M30720
SET SHAFT ENCODER
41 KABELBAUM 8200995
CABLE HARNESS
ANTRIEBSEINHEIT Bild/Picture 12
DEVICE ASSEMBLY
Datei/File: 51312.014
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 35 von 79
DIVAN
ANTRIEBSEINHEIT Bild/Picture 13
DRIVE ASSEMBLY
Datei/File: 51312.015
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 36 von 79
DIVAN
ANTRIEBSEINHEIT Bild/Picture 14
DRIVE ASSEMBLY
Datei/File: 51312.016
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 37 von 79
DIVAN
1 SEITENTEIL M31307
SIDE PANEL
2 SEITENTEIL M31308
SIDE PANEL
3 ABSTANDSBOLZEN M29503
DISTANCE BOLT
7 ZAHNRIEMEN M31499
TOOTHED BELT
8 ANTRIEBSABDECKUNG M29574
DRIVE COVER
9 FUEHRUNG M29699
GUIDE
12 ZAHNRIEMEN M31498
TOOTHED BELT
13 KOLBENSTANGEN-VERRIEGELUNG M31535
PISTON ROD LOCK
15 KUGELLAGERAUFNAHME I M31456
BALL BEARING SEAT I
16 GEHAEUSE,ROHTEIL M24383
HOUSING, UNMACHINED PART
21 ZAHNSCHEIBE M31497
TOOTH LOCK WASHER
23 KUGELLAGERAUFNAHME M31457
BALL BEARING SEAT
25 WELLE M31452
SHAFT
26 ZAHNSCHEIBE M31453
TOOTH LOCK WASHER
27 MOTOR M29515
MOTOR
28 ANTRIEBSSCHEIBE M31496
DRIVE WHEEL
30 WINKEL M30805
ELBOW
38 DISTANZSTUECK M34180
SPACER
39 DREHMOMENTSTUETZE M34381
TORQUE SUPPORT
40 TUELLE M34382
SOCKET
KOLBENSTANGENVERRIEGELUNG Bild/Picture 15
PISTON ROD LOCK
Datei/File: 51312.017
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 41 von 79
DIVAN
1 MITNEHMER M29626
DRIVER
3 HEBEL,UGR. M29625
LEVER
5 DRUCKBOLZEN M29621
PRESSURE BOLT
6 KRALLE M29624
CLUTCH
8 SICHERUNGSBLECH M29663
SAFETY PLATE
10 SPANNSTUECK M29627
MOUNTING PIECE
KOLBENSTANGEN-VERRIEGELUNG Bild/Picture 16
PISTON ROD LOCK
Datei/File: 51312.018
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 43 von 79
DIVAN
1 MITNEHMER M31507
DRIVER
3 HEBEL,UGR. M29625
LEVER
4 DRUCKBOLZEN M29621
PRESSURE BOLT
6 KRALLE M29624
CLUTCH
8 SICHERUNGSBLECH M29663
SAFETY PLATE
10 SPANNSTUECK M29627
MOUNTING PIECE
ELEKTRONIK-PNEUMATIK Bild/Picture 17
ELECTRONIC-PNEUMATIC
Datei/File: 51312.019
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 45 von 79
DIVAN
ELEKTRONIK-PNEUMATIK Bild/Picture 18
ELECTRONIC-PNEUMATIC
Datei/File: 51312.020
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 46 von 79
DIVAN
ELEKTRONIK-PNEUMATIK Bild/Picture 19
ELECTRONIC-PNEUMATIC
Datei/File: 51312.021
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 47 von 79
DIVAN
1 PNEUMATIK M29670
PNEUMATIC
2 KABELSCHELLE 2M14128
CABLE CLAMP
8 KABELBAUM 5 NG 8200686
CABLE HARNESS 5
9 KABELBAUM 1 8200676
CABLE HARNESS 1
11 IC MC 6802 1809156
INTEGRATED CIRCUIT MC 6802
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 48 von 79
DIVAN
14 STOPFEN NG 8404842
PLUG
15 CRIMP-KONTAKT 8302233
CRIMP CONTACT
21 FIBERRING D04766 50
FIBRE RING
ABSTANDSHALTER 8305279
ohne Abbildung
SPACER
without illustration
ELEKTRONIK-PNEUMATIK Bild/Picture 20
ELECTRONIC- PNEUMATIC
Datei/File: 51312.022
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 51 von 79
DIVAN
ELEKTRONIK-PNEUMATIK Bild/Picture 21
ELECTRONIC-PNEUMATIC
Datei/File: 51312.023
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 52 von 79
DIVAN
ELEKTRONIK-PNEUMATIK Bild/Picture 22
ELECTRONIC-PNEUMATIC
Datei/File: 51312.024
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 53 von 79
DIVAN
1 PNEUMATIK M31688
PNEUMATIC
2 KABELSCHELLE 2M14128
CABLE CLAMP
8 KABELBAUM 5 NG 8200686
CABLE HARNESS 5
9 KABELBAUM 1 8200676
CABLE HARNESS 1
11 IC MC 6802 1809156
INTEGRATED CIRCUIT MC 6802
12 KREUZHALTERUNG 8404154
CROSS SUPPORT
14 STOPFEN NG 8404842
PLUG
15 CRIMP-KONTAKT 8302233
CRIMP CONTACT
21 FIBERRING D04766 50
FIBRE RING
30 ABSTANDSHALTER 8305279
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 55 von 79
DIVAN
PNEUMATIK Bild/Picture 23
PNEUMATIC
Datei/File: 51312.025
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 57 von 79
DIVAN
PNEUMATIK Bild/Picture 24
PNEUMATIC
Datei/File: 51312.026
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 58 von 79
DIVAN
PNEUMATIK Bild/Picture 25
PNEUMATIC
Datei/File: 51312.027
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 59 von 79
DIVAN
1 EINSCHUBBLECH M30714 3
PLUG-IN PLATE
3 WINKEL,UGR. M29543 3
ELBOW
4 PLATTE,UGR. M29693
PANEL
5 FEDER B E20342
SPRING
6 SCHRAUBE M29319
SCREW
9 KABELDURCHFUEHRUNG CH17949
CABLE BUSHING
10 BUCHSE 8408312
SOCKET
11 T-TUELLE 6800187 5
T-SOCKET
13 PNEUMATIKBLOCK,UGR. M29521
PNEUMATIC BLOCK
14 GIMETALL-LAGER 2M06157 5
RUBBER METAL CONNECTION
18 3/2-WEGE-VENTIL M29566
3/2-PORT DISTRIBUTING VALVE
19 L-SCHWENKANSCHLUSS G12065
L-SWIVEL CONNECTION
21 TUELLE M29927
SOCKET
25 REDUZIERSTUECK M22286 5
REDUCER
26 WINKELSTECKANSCHLUSS NG M29933
ANGLE CONNECTION
27 T-STUECK 8401083
T-PIECE
28 STOPFEN NG 8404842
PLUG
29 PLATTE M30530
XYZ
30 PLATTE M30552
XYZ
35 BUCHSE 8301453
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 61 von 79
DIVAN
36 HALTEBLECH M30684
SUPPORT PLATE
38 BUEGEL CH13155
CLIP
41 SICHERHEITSVENTIL M31821
250 mbar
SAFETY VALVE
250 mbar
42 SICHERHEITSVENTIL M31822
82 mbar
SAFETY VALVE
82 mbar
43 PEEP-VENTIL 8406516
PEEP VALVE
PNEUMATIK Bild/Picture 26
PNEUMATIC
Datei/File: 51312.028
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 63 von 79
DIVAN
PNEUMATIK Bild/Picture 27
PNEUMATIC
Datei/File: 51312.029
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 64 von 79
DIVAN
1 EINSCHUBBLECH M31686
PLUG-IN PLATE
5 KABELDURCHFUEHRUNG CH17949
CABLE BUSHING
6 BUCHSE 8408312
SOCKET
7 T-TUELLE 6800187 5
T-SOCKET
8 PNEUMATIKBLOCK M31687
PNEUMATIC BLOCK
10 GIMETALL-LAGER 2M06157 5
RUBBER METAL CONNECTION
14 3/2-WEGE-VENTIL M29566
3/2-PORT DISTRIBUTING VALVE
15 SCHWENKVERSCHRAUBUNG 8402034
SWIVELLING SCREW-FITTING
16 KABELBAUM 6 8201597
CABLE HARNESS 6
18 REDUZIERSTUECK M22286 5
REDUCER
20 PLATTE M30530
XYZ
21 PLATTE M31689
XYZ
23 PLATTE M30552
XYZ
27 BUCHSE 8301453
SOCKET
28 HALTEBLECH M30684
SUPPORT PLATE
30 BUEGEL CH13155
CLIP
31 LINSENSCHRB.DIN7985-M3X6-A2-H 1340727 20
OVAL HEAD SCR.DIN7985-M3X6-A2
33 SICHERHEITSVENTIL M31822
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 66 von 79
DIVAN
34 SICHERHEITSVENTIL M31821
SAFETY VALVE
35 PEEP-VENTIL 8406516
PEEP VALVE
40 HALTER M31684
SUPPORT
PNEUMATIKBLOCK Bild/Picture 28
PNEUMATIC BLOCK
Datei/File: 51312.030
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 68 von 79
DIVAN
2 PNEUMATIKBLOCK M31682
PNEUMATIC BLOCK
3 TUELLE 8400964 10
SOCKET
4 SCHLAUCHVERSCHRAUBUNG 8402032 5
THREADED HOSE COUPLING
5 O-RING M32804 10
O-RING SEAL
TUELLE,VOLLST. Bild/Picture 29
SOCKET,CPL.
Datei/File: 51312.031
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 70 von 79
DIVAN
1 BOLZEN M28790
BOLT
2 RUNDSCHNURRING M12165 10
TOROIDAL SEALING RING
3 SCHEIBE M28791
DISC
4 TUELLE M28785
SOCKET
4a TUELLE M29930
SOCKET
SICHERHEITSVENTIL Bild/Picture 30
SAFETY VALVE
Datei/File: 51312.032
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 72 von 79
DIVAN
1 VENTILGEHAEUSE M30948
VALVE HOUSING
3 FEDER 8403504
SPRING
4 EINSTELLSCHRAUBE M30567
ADJUSTING SCREW
SICHERHEITSVENTIL Bild/Picture 31
SAFETY VALVE
Datei/File: 51312.033
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 74 von 79
DIVAN
1 VENTILGEHAEUSE M31820
VALVE HOUSING
3 FEDER 8403504
SPRING
4 EINSTELLSCHRAUBE M31819
ADJUSTING SCREW
5 MUTTER 8300656
NUT
DECKEL,VOLLST. Bild/Picture 32
COVER, CPL.
Datei/File: 51312.034
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 76 von 79
DIVAN
2 ABDECKBLECH,UGR. M29635
COVER PLATE
3 SCHLITZMUTTER T04896
SLOTTED ROUND NUT
6 HALTERUNG M29938
SUPPORT
7 HALTERUNG M29937
SUPPORT
8 ABDECKUNG M30950
COVER
9 BLECH M29311
SHEET METAL
11 SCHARNIER M29632
HINGE
13 SCHARNIERBEFESTIGUNG M29633
HINGE FIXING
17 MAGNETVERSCHLUSS 2M10046
MAGNETIC LOCK
FRONTPLATTE Bild/Picture 33
FRONT PANEL
Datei/File: 51312.035
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 5131.2 Ausgabe/Edition
Spare parts list 17.12.96
DIVAN Seite/Page 78 von 79
DIVAN
1 DREHKNOPF M29954
CONTROL KNOB
2 BEDIENFELD M29557
CONTROL PANEL
4 BUCHSE M29784
SOCKET
7 ABSTANDSHALTER 8305279
SPACER
10 FIBERRING D04766 50
FIBRE RING
11 LINSENSCHRB.DIN7985-M3X6-A2-H 1340727 20
OVAL HEAD SCR.DIN7985-M3X6-A2
12 IMPULSGEBER,VOLLST. 8200957
PULSE GENERATOR,CPL.
DREHGEBER 8500088
ohne Abbildung
SHAFT ENCODER
without illustration
13 WINKEL M29568
ELBOW
14 WINKEL M29569
ELBOW
16 BLECH M30719
SHEET METAL
Sach-Nr. Benennung
Part No. Description
8600440 PM 8050 GRUNDGERAET
PM 8050 BASIC UNIT
8600678 PM8050-CATO
PM8050-CATO
2 Frontplatte,PM 8050 CD
Frontpanel, PM8050 CD
3 Frontplatte, Rückansicht
Frontpanel, Rearview
4 Innenansicht
Insideview
5 Rückansicht
Rearview
6 Zubehör
Accessories
Datei/File: 6494300.001
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 6494.300 Ausgabe/Edition
Spare parts list 03.02.98
PM 8050 / PM 8050 CD Seite/Page 6 von 18
PM 8050 / PM 8050 CD
1 WASSERFALLE 8600571
WATER SEPARATOR
2 WASSERABSCHEIDER 8600570
WATER SEPARATOR
3 AUFKLEBER 8600618
ADHESIVE LABEL
4 TOPF 8600227
POT
6 FRONTPLATTE 8600426
FRONTPANEL
7 DREHKNOPF M29954
CONTROL KNOB
Datei/File: 6494300.002
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 6494.300 Ausgabe/Edition
Spare parts list 03.02.98
PM 8050 / PM 8050 CD Seite/Page 8 von 18
PM 8050 / PM 8050 CD
2 FRONTPLATTE 8600426
FRONTPANEL
3 DREHKNOPF M29954
CONTROL KNOB
Datei/File: 6494300.003
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 6494.300 Ausgabe/Edition
Spare parts list 03.02.98
PM 8050 / PM 8050 CD Seite/Page 10 von 18
PM 8050 / PM 8050 CD
LINSENSCHRB.DIN7985-M3X6-A2-H 1340727
OVAL HEAD SCR.DIN7985-M3X6-A2
6 KLEMMSICHERUNG 8100245
LOCKING RING
Innenansicht Bild/Picture 4
Insideview
Datei/File: 6494300.004
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 6494.300 Ausgabe/Edition
Spare parts list 03.02.98
PM 8050 / PM 8050 CD Seite/Page 12 von 18
PM 8050 / PM 8050 CD
6 FILTERMATTE 8305367
FILTER MAT
9 GASMESSMODUL 6870200
Ersetzt durch Rüstsatz IRIA 8601625
GAS METER MODULE
Replaced by upgrade kit IRIA 8601625
9 AGAS-MODUL 6870190
AGAS MODULE
SENSORKOPF,VOLLST. 6870180
SENSOR HEAD,CPL.
KUEVETTE,VOLLST. 6870183
nur für IRIA-Modul
BULB,CPL.
only for IRIA-Module
PUMPENPLATINE 6870050
nur für ANDROS-Pumpe
PUMP BOARD
only for ANDROS-Pump
MAGNETVENTIL,UGR. 6870108
ELECTROVALVE
Rückansicht Bild/Picture 5
Rearview
Datei/File: 6494300.005
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 6494.300 Ausgabe/Edition
Spare parts list 03.02.98
PM 8050 / PM 8050 CD Seite/Page 15 von 18
PM 8050 / PM 8050 CD
11 TUELLE 8600512
12 TUELLE 8600509
SOCKET
13 SCHRAUBE,VOLLST. 6850929
SCREW,CPL.
14 O2-KNOPFZELLENADAPTER 6850932
ROUND CELL ADAPTER O2
O2-SENSOR-S 6850930
O2-SENSOR-S
O-RING D15330
O-RING SEAL
15 ABDECKUNG 8600778 10
16 TUELLE 8600511
17 NC AKKU 9V 8301856
NC STORAGE BATTERY 9V
KABELBINDER 8712065
CABLE CLIP
Zubehör Bild/Picture 6
Accessories
Datei/File: 6494300.006
Druckdatum: 16.06.1999 Änderungen vorbehalten
Printed Subject to change without notice
DrägerService Medizintechnik
DrägerService Medical Division
Ersatzartikelliste 6494.300 Ausgabe/Edition
Spare parts list 03.02.98
PM 8050 / PM 8050 CD Seite/Page 18 von 18
PM 8050 / PM 8050 CD
2 FILTER 8600225
FILTER
5 DRUCKMESSLEITUNG,UGR. 8302841
PRESSURE GAUGE LINE
8 SPIROLOG-SENSOREN,5ER-SET 8403735
SPIROLOG SENSORS 5PCS SET
9 KOMBINATIONS-MESSANSCHLUSS M28833
COMBINED MEASURING ADAPTOR
10 KAPPE M21482
CAP
11 O2-ADAPTER 8405807
O2 ADAPTER
12 SENSOR-AUFNAHME 8407390
SENSOR SEAT
Moislinger Allee 53 – 55
D-23542 Lübeck
Federal Republic of Germany
5170 e
Subject to modification.
Will not be replaced in the event of modifications.