8 Electrical spare parts

8.6 Software & board version references


This table explains which units, programs, boards and softwares are compatible and permitted. In case of
mischoice your unit does not work correctly.

Ortho Ortho TMJ/ Digipan Ortho ID Interface board & CPU PAL Software
Trans Ortho Zone (ICD25) version version

No No No No 60123 & 1.0.0, 60123 & 1.1.0 1.2.01-16


60166 & 2.0.0, 60166 & 2.1.0
Yes 60123 & 1.1.0 1.2.06
60166 & 2.1.0
60123 & 1.0.0, 60123 & 1.1.0 1.2.07 - 1.2.16
60166 & 2.0.0, 60166 & 2.1.0
Yes -----> 60123 & 1.0.0 1.2.07- 1.2.16
60166 & 2.0.0
Yes No No 60123 & 1.0.0, 60123 & 1.1.0 1.2.06 - 1.2.16
60166 & 2.0.0, 60166 & 2.1.0
Yes 60123 & 1.1.0 1.2.06
60166 & 2.1.0
60123 & 1.0.0, 60123 & 1.1.0 1.2.07 -1.2.16
60166 & 2.0.0, 60166 & 2.1.0
Yes ----> 60123 & 1.0.0 1.2.07 - 1.2.16
60166 & 2.0.0
Yes ----> No ----> 60166 & 2.0.0 1.2.07 - 1.2.16
60166 & 2.1.0
Yes -----> 60166 & 2.0.0 1.2.07 - 1.2.16
OPTION MUST HAVE
DIGIPAN => Software v. 1.2.07 or 1.2.14 and PAL v. 1.0.0 or
2.0.0
Ortho ID => Software v. 1.2.06 and PAL v. 1.1.0 or 2.1.0
Software v. 1.2.07 and PAL v. 1.0.0 or 1.1.0 0r
2.0.0 or 2.1.0.
CPU v.1.0-1.2 and cable 68999 or CPU v 1.3 or
higher
Interface board 60166 => PAL v. 2.0.0 or 2.1.0
Ortho Zone or Ortho TMJ => Software v. 1.2.06, 1.2.07 or 1.2.14 all CPU
versions
Ortho Trans => Interface board 60166 and Software v. 1.2.07 -
1.2.16
(and PAL v. 2.0.0 or 2.1.0). All CPU versions.