Reagent Inspection and Analysis Report: Rayto Life and Analytical Sciences Co., LTD
Reagent Inspection and Analysis Report: Rayto Life and Analytical Sciences Co., LTD
4 Accuracy The relative deviation should not exceed ±10%. Result: 0.50%
☑Conformable □Inconformable
Result: Low-level: 0.92%
5 Repeatability Coefficient of Variation (CV) ≤4% High-level: 0.54%
☑Conformable □Inconformable
When testing the 0.65mmol/L calibrator or
Sensitivity for Result: 0.2217A
6 sample, at the optic path length of 10mm, the
analysis
absorbance difference (△A) should be ≥ 0.15A. ☑Conformable □Inconformable
Within the range of 0.02mmol/L ~ 1.50mmol/L
Linearity Result: 1.0000
7 (37℃), the correlation coefficient (r) should be ≥
range
0.990. ☑Conformable □Inconformable
Conclusion Conforming