Professional Documents
Culture Documents
Lambda Technical Manual (MT Lambda)
Lambda Technical Manual (MT Lambda)
(MT LAMBDA)
LAMBDA
WARNING
Using and keeping X-Ray equipment and device must follow the local Regulations and national laws
concerning Medical X-Ray handling.
Note
According to CE 93/42 Directive for Medical Devices concerning product traceability, GENERAL
MEDICAL ITALIA as manufacturer, must be informed of any owner or installation address change.
MDD traceability is prerequisite to assure their safety and reliability over the time.
INDEX
Document status sheet pag. II
CHAPTER 2 - INSTALLATION
Device classification..................................................................................................pag. 2-3
Identification label .....................................................................................................pag. 2-4
Precaution for installation..........................................................................................pag. 2-7
Lambda dimensions and weight................................................................................pag. 2-8
Installation area.......................................................................................................pag. 2-10
Anti X shield and unit placement.............................................................................pag. 2-11
Unpacking ...............................................................................................................pag. 2-12
Ambient storage condition ......................................................................................pag. 2-13
Ambient operating condition ...................................................................................pag. 2-13
Installation check list ...............................................................................................pag. 2-14
Backside view main internal components location ..................................................pag. 2-15
C-arm components location ....................................................................................pag. 2-16
Serial ports..............................................................................................................pag. 2-17
Line voltage check ..................................................................................................pag. 2-18
Line frequency adaptation.......................................................................................pag. 2-18
Measuring line resistance .......................................................................................pag. 2-18
Protective measures ...............................................................................................pag. 2-18
Power line connection.............................................................................................pag. 2-20
Emergency cut off ...................................................................................................pag. 2-22
LCD contrast adjustment ........................................................................................pag. 2-23
Alarm messages language selection ......................................................................pag. 2-23
H.V. generator tank vent valve opening ..................................................................pag. 2-24
Date and time setting ..............................................................................................pag. 2-25
Maximum compression force setting.......................................................................pag. 2-26
Foot pedal connection.............................................................................................pag. 2-26
Door safety switch...................................................................................................pag. 2-27
External lamp ..........................................................................................................pag. 2-28
Exposure counter....................................................................................................pag. 2-28
Delay times between two exposures.......................................................................pag. 2-29
Service menu ..........................................................................................................pag. 2-30
Dip Switch and related functions.............................................................................pag. 2-31
Calibration................................................................................................................pag 2-32
GENERAL MEDICAL ITALIA CHAPTER 1 Page 5 of 236
LAMBDA TECHNICAL MANUAL
Unit cleaning ...........................................................................................................pag. 2-32
Packing instructions (UNI 9151)..............................................................................pag. 2-33
CHAPTER 1
UNIT DESCRIPTION
TECHNICAL SPECIFICATIONS OF MAMMOGRAPHY UNIT LAMBDAI ............................8
ELECTROMAGNETIC INFORMATIONS ACCORDING TO IEC 601-1-2 ..........................18
WEEE INFORMATIONS ACCORDING TO DIRECTIVE 2002/96/EC................................22
WARNINGS........................................................................................................................23
H.V.GENERATOR
Line voltage compensation AUTOMATIC
H.V. generator with kV closed loop and line
Feed forward compensation
Inverter Technology Current fed, Mosfet bridge with output
current limit capability and short circuit
protection
Inverter Frequency 50 kHz
Ripple Frequency/Amplitude 100 kHz < 2%
Output power 3500 W 35 kVx100 mA
(5000W optional)
kV range 20 / 35 kV
kV resolution (Manual & Auto mode) 0.5 kV
kV precision ±1%
kV repeatability ± 0,1%
kV rise time ≤1.5 ms from 0 to100%
kV display XX,X kV (3 digits)
Anode current max 100 mA (140mA optional)
mAs range Small focus: 1/200 mAs (from 20 to 30 kV)
1/180 mAs (from 31 to 35 kV)
Large focus 1/640 mAs (from 20 to 30 kV)
1/500 mAs (from 31 to 35 kV)
mAs values according to series R’20 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 13, 16, 20, 25,
32, 40, 50, 63, 80, 100, 130, 160, 180, 200,
250, 320, 400, 500, 640
mAs resolution (Automatic) 0,1 mAs
mAs display XXX,X mAs (4 digits)
Exposure Time Automatically selected in function of selected
mAs
Safety timer 10 sec
IMAGE QUALITY
Spatial resolution Conformity with:
“European Guidelines for quality assurance in
mammography screening”, third edition,
and with
”Recommended specifications” for Quality
assurance in mammography of American
College of Radiology
C-ARM
F.F.D. 65cm for standard and BYM C-ARM.
Optional (only available for BYM C-ARM) 65
to 70 cm variable with motorized movement
Rotation Manual ±180° with disc brake
Vertical movement with respect to Breast 65cm F.F.D. STANDARD C-ARM
support (C-ARM in vertical position) 55 to 130 cm
65cm F.F.D. BYM C-ARM
83 to 149 cm
65/70 cm variable F.F.D. BYM C-ARM
83 to 149 cm F.F.D. 65 cm
78 to 144 cm F.F.D. 70 cm
Patient protection Face shield with plastic screen for patient’s
face protection
C-arm rotation angle Showed on C-arm by mechanical goniometer.
Optional showed on display
COLLIMATOR
Light beam With automatic switch ON when operating
compression and electronic timer
Light intensity ≥ 150 lux
Light BEAM collimation accuracy according to IEC 601-1-3
Mirror with automatic OUT of FIELD function
Collimation plates (standard) 18X24cm, Ø 14cm
Collimation plates (optional) 24x30, 10x24
COMPRESSION SYSTEM
Compression Plate movement Manual or motor driven
Compression Plates (standard) 16x22 shifted, 12x8 straight,
Compression Plates 10x22 shifted, 24x30 shifted, 10x8 shifted,
(optional) 12X22 straight 16x22 shifted for biopsy
Free space available between 65cm F.F.D. STANDARD OR BYM C-ARM
Compression Plate and image receptor 39,5 cm with shifted Compression Plates
65/70 cm variable F.F.D. C-ARM
39,5cm /44,5cm
in Magnification Mode F.F.D.= 65cm
MAG. X2 = 140mm
MAG. X 1.8 = 180mm
MAG. X1,5 = 240mm
Compression Thickness on Display 0-80 mm in steps of 5 mm.
Compression Paddle Release after Selectable from control panel, Automatic or
exposure Manual for 2D Biopsy
Compression plate aluminium Less than 0.2 mm Al (0.135 mm Al@30 kV)
equivalence
MAGNIFICATION
Magnification ratio x1,5 / x2 (standard) with Bucky table
x1.8 with automatic focal spot selection,
cassette holder without grid and cassette
detector (optional)
POTTER BUCKY
Bucky factor (grid) 1,96
Ratio 5:1
Lines/cm 36
Contrast factor 1,47
Cassette size 18x24 standard
24x30 optional
Cassette compatibility All the most diffused models with window as:
Agfa, Dupont, Fuji,Kodak,3M
Cassette detector switch With alarm in 8 languages to avoid double
exposure or exposure without cassette
Top Cover Carbon fiber
Film marker Integrated with 9 brass labels
Test with NORMI Phantom Typical 3,5 balls
Other features Easily interchangeable with other accessories
without tools
Grid movement synchronized with X-RAY
beam
DOSE CALCULATOR
Data visualization On display, label printer and data memory
CONTROL TABLE
Technology Microprocessor controlled with unique safety
features exceeding IEC 601-1-4-, all
functions under active operator control
Display GRAPHIC LCD Display 240x128 dots
Alarm messages In 8 languages
Serial port exposition data for Easylabel Dedicated for Film Labelling device Easylabel
(200 characters printed), A.E.C. / tube
calibration and last 100 exposures memory
download.
Calibration and service Serial port For service laptop
Special Features Last 1300 exposure memory;
Tube Thermal Unit display and active
protection.
Technical display for self-test and defective
block identification, firmware release,
exposure counter and last exposure
time/date.
Error and faulsts data logger.
Statistics funtion Average dose, amount of exposure for every
kV value, amount of exposure in every test
technique
PROTECTIVE GLASS
Protective glass (for shield model Stand Alone full size 210cm (h) x 70 cm (w)
2SPDES and 2SPDES-F) (optional) >0,34mm Pb equivalent @35kV
Protective glass (for shield model Stand Alone half size 115cm (h) x 70 cm (w)
2SPDES-T) (optional) >0,34mm Pb equivalent @35kV
ENVIRONMENTAL CONDITIONS
Storage and delivery conditions (while temperature -20°C / + 70°C
packed) relative humidity 10% / 90%
barometric pressure 500 hPa/1060 hPa
Operating conditions temperature +10°C / + 40°C
relative humidity 30% / 75%
barometric pressure 700 hPa/1060 hPa
Protection degree according to standard IP 30
IEC 529
Heat dissipated in load condition of 35 KV 264 Kcal/h
500 mAs (1 shot every 5 minutes)
NOTE:
The manufacturer reserves the right to make further improvements while
keeping main features unchanged
Information: Fixed equipment or system cabling, which can not be removed by the user, is not listed.
This cabling is part of the system and was present at all EMC-measurements. Without this cabling
there is no complete functionality of the system.
Warning: The use of accessories, transducers and cables other than those specified, with the exception
of transducers and cables sold by the manufacturer of the equipment or system as replacement parts for
internal components, may result in increased emission or decreased immunity of the equipment or
system.
Voltage fluctuations/
flicker emissions
Not applicable
IEC 61000-3-3
Warning: The equipment or system should not be used adjacent to or stacked with other equipment
and that if adjacent or stacked use is necessary, the equipment or system should be observed to verify
normal operation in the configuration in which it will be used.
IEC 61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the LAMBDA is used exceeds the applicable RF compliance level above, the
LAMBDA should be observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the LAMBDA.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
that is present on the device, means that within the European Union the product must be taken to
separate collection at the product end-of life. Therefore, at the and of the life-cycle of the device, the
user should deliver the device to the proper collection facilities of the Electric and Electronic
Equipments. Alternatively, the user can return the device to the seller, on a one-to-one basis, as long as
he is buying a new one of equivalent type and that fulfills the same functions as the old one.
Disposing of the device separately avoids possible negative consequences for the environment and
health deriving from inappropriate disposal and enables the constituent materials to be recovered to
obtain significant savings in energy and resources.
Who disposes any Electric and Electronic Equipment, reporting the above symbol, as unsorted
municipal waste, instead of collecting it separately, incurs the administrative sanctions in accordance
with law.
WARNINGS
CHAPTER 2
INSTALLATION
1 - DEVICE CLASSIFICATION....................................................................................26
2 – IDENTIFICATION LABELS....................................................................................27
3 – PRECAUTION FOR INSTALLATION.....................................................................30
4 – LAMBDA DIMENSIONS AND WEIGHT .................................................................31
5 - INSTALLATION AREA ...........................................................................................33
6 – ANTI X SHIELD AND UNIT PLACEMENT .............................................................34
7- UNPACKING...........................................................................................................35
8 - AMBIENT STORAGE CONDITIONS (WHILE PACKED) ........................................36
9 - AMBIENT OPERATING CONDITIONS ..................................................................36
10 - INSTALLATION CHECK LIST FOR LAMBDA MAMMOGRAPHIC UNIT ..............37
11 - BACKSIDE VIEW MAIN INTERNAL COMPONENTS LOCATION........................38
12 - C-ARM COMPONENTS LOCATION ....................................................................39
13 – SERIAL PORTS...................................................................................................40
14 - LINE VOLTAGE CHECK ......................................................................................41
15 - LINE FREQUENCY ADAPTATION.......................................................................41
16 - MEASURING THE LINE RESISTANCE ...............................................................41
17 - PROTECTIVE MEASURES..................................................................................41
18 - POWER LINE CONNECTION ..............................................................................43
19 - EMERGENCY CUT OFF ......................................................................................45
20 - LCD CONTRAST ADJUSTMENT .........................................................................46
21 - ALARM MESSAGES LANGUAGE SELECTION ..................................................46
22 - H.V. GENERATOR TANK VENT VALVE OPENING.............................................47
23- DATE AND TIME SETTING...................................................................................48
24- MAXIMUM COMPRESSION FORCE SETTING....................................................49
25 - FOOT PEDAL CONNECTION ..............................................................................49
26 - DOOR SAFETY SWITCH .....................................................................................50
27 - EXTERNAL LAMP ................................................................................................51
28 – EXPOSURE COUNTER ......................................................................................51
29- DELAY TIMES BETWEEN TWO EXPOSURES ....................................................52
30 - SERVICE MENU ..................................................................................................53
GENERAL MEDICAL ITALIA CHAPTER 1 Page 24 of 236
LAMBDA TECHNICAL MANUAL
31 - DIP SWITCH AND RELATED FUNCTIONS .........................................................54
32 - CALIBRATION .....................................................................................................55
33 – UNIT CLEANING .................................................................................................55
34 - PACKING INSTRUCTIONS (UNI 9151)................................................................56
1 - DEVICE CLASSIFICATION
2 – IDENTIFICATION LABELS
The meaning of symbols appearing on the plate and on some components of the machine is
specified in the following table.
BASE PLATE
When carrying out the necessary support calculation, weight of patient and operator must be
taken into consideration.
773
700
2100
810
505
5 - INSTALLATION AREA
The delivered cable length from unit to the wall outlet is about 5m
7- UNPACKING
WARNING : Do not apply forces to c-arm when moving the unit out of the box and during
final location placement
Installation check list shall be started just after the device has been unpacked.
In case of risk of dew, once the equipment has been unpacked, wait for at least 24 hours
before applying voltage, thus enabling it to reach a thermal equilibrium with the surrounding
environment.
LAMBDA s/n_____________
INSTALLATION PLACE ________________________________________
INSTALLATION DATE ________________________________________
INSTALLATOR FIRM ________________________________________
RESPONSIBLE PERSON ________________________________________
SIGNATURE________________________________
13 – SERIAL PORTS
X-ray pushbutton
connection
Easylabel
connection
Mains
Pedals
Second pair of
pedals (option)
Note: the picture represents the serial ports of LAMBDA SE; the LAMBDA has the same
serial ports.
17 - PROTECTIVE MEASURES
It is very important that any intervention in the equipment will start with disconnecting it from the
mains by means of wall mounted circuit breaker. Before removing or inserting any of the printed
circuit board, switch OFF the equipment.
In the following list all dangerous parts are highlighted, read it carefully before to touch any internal
part.
For location of the listed parts refer to “Backside view main internal components location” and “C-
Arm components location” in the previous pages.
If the system is only switched OFF at the control panel, line voltage is still present at the
followings:
• Terminal and fuses of power line connection.
• Main contactor and emergency cut off buttons PB11A and PB11B
• Line filter
• Terminal M1
• Fuses F1, F2, F5
• Static Relay RY2
After shut down of the system, there may still be dangerous voltage up to 550Vdc on inverter
power supply. Voltage decreases to zero by means of bleeder resistors in not less than 5 minutes due
to capacitor size, energy stored is still dangerous if short to ground occurs when It’s less than 60Vdc
and till when capacitors are not fully discharged.
Printed Circuit Boards contain electrostatic highly sensitive components requiring particular
care in their handling (ground before making contact and place only a conductive surface).
To prevent accidental triggering of High Voltage and X-radiation output X-ray pushbutton has Dead
Man function by which H.V. generator is disconnected by CPU if push button is not intentionally
pressed.
For more safety during intervention:
1. DEMO MODE can be selected by means of DIP 3 SW1 PCB 01-170
2. Fuses F6, F9 can be removed and Inverter voltage monitor disabled by means of JP1 closed on
PCB 92-082-6. Vdc across C1,2,3,4, inverter and H.V. transformer will be zero after
discharging time.
WARNING:
The unit is supplied with single phase Mains input with :
phase / neutral / protective earth.
If available Mains is:
phase / phase / protective earth
the terminal block for neutral conductor connection must be replaced with a fuse holder and same fuse
rating of the pre existing one for phase conductor connection.
fuse
P
fuse
P
P fuse
For single-phase mains with Phase / Neutral connection, make sure that the neutral conductor of
the mains is connected to terminal N.
Protective Earth connection must be checked by qualified electrician with appropriate instrument and
procedure according to local applicable laws regulations.
Voltage selector is originally configured for 220Vac Mains. Others 230 or 240 must be configured
according to the following wiring diagram and pictures.
1 2 3 4 Language
OFF OFF OFF X Italian
ON OFF OFF X French
OFF ON OFF X English
ON ON OFF X German
OFF OFF ON X Spanish
ON OFF ON X Polish
OFF ON ON X Turkish
ON ON ON X Portuguese
DIPSW1
RV1
PCB 92-083
Vent valve
Unscrew vent plug of about one or two turns, and secure it with ring nut. If the vent plug is hard to
unscrew, it is already secured by the ring nut. So you have first to unscrew ring nut, then unscrew vent
plug an finally screw ring nut.
Vent plug
Ring nut
Activate the regulation mode by simultaneously pressing the alarm reset button and the decompression
mode selection button. Keep these buttons pressed during regulation.
The date and time can be selected using the following buttons:
day KV regulation
month mAs regulation
year film screen selection / AEC selection
minutes compression force regulation
hour optical density regulation
To connect the foot pedals, use the connections in the picture. If you have two pedals to connect, they
can be connected in parallel; wire both in the same connector.
The pedals have three wires to be used, the ground is not used. By means of a ohmmeter you can check
witch of the three wires of the pedals is the common wire, witch corresponds to the movement up and
witch corresponds to the movement down.
Then, connect the common wire the terminal M1-8. Connect the compression down to the terminal
M1-9, and the compression up to the M1-10.
Verify that pushing the compression up foot pedal, the compressor moves up, and that pushing the
compression down the light of the compression system switches on and the compressor moves down.
DOOR OPEN
On the other hand, opening the door during exposure will have no effect and no need to repeat the
exposure, thus delivering an extra dose to the patient.
27 - EXTERNAL LAMP
An output is provided to connect an external mains voltage lamp.
A set of two signals allows the state of the unit to be recognised (POWER ON / RADIATION ON).
28 – EXPOSURE COUNTER
The unit contains an electromechanical pulse counter to count all the shots taken. The pulse counter is
located on PCB 92-082.
C
+
U D
Pulse counters DO NOT necessarily record the same values, and cannot be synchronised.
30 - SERVICE MENU
If buttons film/screen selection and alarm reset are pushed at the same time, the Service Menu is
displayed:
SW2
DIP1 ON = C-arm position S/L/R for Digital Bym controlled
from control panel
DIP2 ON = PHTM9000 three fields detector enabled
DIP3 OFF = 96-153; ON = mirror magnet (for old units compatibility)
DIP4 Must remain ON (OFF = C-Arm brake for LAMBDA )
DIP5 ON = Printer S’PRINT I-LP1 for adhesive labels
SW1, DIP1 & DIP2: To enter Anode current calibration. See dedicated section in Chapter 3,
Filament.
SW1, DIP3: In DEMO mode all functions are active, except for the X-ray output, whose sequence is
simulated (only manual mode)
Demo mode can be used during an exhibition and during safety rotating anode calibration.
SW1, DIP4: For BYM C-arm this dip switch must be ON, for normal C-arm must be OFF, because it
influences the breast thickness value and the AGD values.
SW1, DIP5: Reduced Deata Plus with only one film/screen and one technique. It is only intended for
manual calibration when PC densitometer, sensitometer are not available. Optical density linearity is
not guaranteed as well fulfilment of applicable standard and Q.A protocols.
SW1, DIP6: To enable manual adjustment of O.D. for Reduced Deata Plus. Dip 6 is active only if DIP
5 is ON. Refer to related chapter of the manual.
SW1, DIP7: Calibration of compression system.
SW1, DIP8: ON to enable FILTROMAMM functionality. If FILTROMAMM is installed later, as
upgrade, dip 8 must be switched ON and Mo/Rh curves of AEC must be calibrated.
SW2, DIP1: When the device is configured with bym C-arm and digital bym, two detectors placed on
c-arm detect c-arm position.
If the device has no detectors, c-arm position can be detected from control panel after switching DIP1
SW2 = ON on PCB/01-170 board.
SW2, DIP2: must be set OFF only for PHTM6000.
SW2, DIP3: The configuration of this DIP Switch depends on the hardware of the mammography unit.
SW2, DIP4: For LAMBDA must remain ON. (OFF for LAMBDA E)
SW2, DIP5: OFF for Easylabel protocol ON for S’PRINT protocol
32 - CALIBRATION
Automatic exposure meter calibration (AEC) is the only necessary operation to do before use.
See chapter 3 of this manual.
33 – UNIT CLEANING
The unit can be cleaned using a soft cloth with neutral detergent.
Position the wood base (A) over the blocks (B) as in FIG. 1
Protect the unit by wrapping it with air bubble plastic foil, fix foil with adhesive tape avoiding to stick
it on painted surfaces.
Place two foam cushions on the wood base and lay down the unit
Place boxes with accessories and secure them with other foam cushions.
Nail the top wall cover and wrap the box with two belts as in FIG. 7
CHAPTER 3
THEORY & CALIBRATION
ROTATING ANODE STARTER .........................................................................................61
1 – CALIBRATION OF PROTECTION CIRCUIT .........................................................62
2 – CONTROL LED .....................................................................................................63
3 – FUNCTION CONFIGURATION .............................................................................64
4 – TROUBLE SHOOTING..........................................................................................65
FILAMENT..........................................................................................................................66
1 – FILAMENT POWER SUPPLY................................................................................67
2 – ANODE CURRENT CALIBRATION .......................................................................68
3 – CALIBRATION OF FILAMENT CURRENT PROTECTION ....................................71
4 – CALIBRATION OF FILAMENT VOLTAGE PROTECTION.....................................72
5 - ANODE CURRENT CHECK BY EXTERNAL DEVICE............................................72
6 – CHECK-LED..........................................................................................................73
7 – TROUBLE SHOOTING..........................................................................................73
HIGH VOLTAGE GENERATOR.........................................................................................75
1 – DESCRIPTION ......................................................................................................76
2 – KV CHECK AT TP1/TP2 PCB 94-132 ....................................................................78
3 – NON INVASIVE KV CHECK ..................................................................................79
4 – CHECKING AND ADJUSTMENT POINTS ............................................................80
5 – EXPOSURE TIME CONTROL ...............................................................................81
6 – ADJUSTMENT MAX OUTPUT CURRENT ............................................................81
7 – CONTROL LED .....................................................................................................81
8 – TROUBLE SHOOTING..........................................................................................82
AUTOMATIC EXPOSURE CONTROL DEATA PLUS........................................................83
1 - THINGS TO KNOW BEFORE STARTING..............................................................84
2 - PROCEDURE PREPARATION ..............................................................................88
3 - CALIBRATION PROCEDURE ................................................................................95
REDUCED DEATA PLUS ................................................................................................106
1 – REDUCED CALIBRATION PROCEDURE...........................................................107
2 - 18X24 AND 24X 30, GRIDLESS OR BUCKY TABLES .........................................107
DOSE CALCULATOR ......................................................................................................108
2 – CONTROL LED
3 – FUNCTION CONFIGURATION
3.1 – Enabling braking
3.2 – Braking time
3.3 – Disabling protection circuit
4 – TROUBLE SHOOTING
During the calibration phase, undesired protection circuit tripping could be an obstacle. If this happens,
it is advisable to exclude it by reducing, even to zero, the voltage on TP3 and TP4 through RV1 and
RV2.
In special cases, the protection circuit can be completely excluded by short-circuiting JP4.
During calibration make sure that the anode is rotating.
1.2 Calibration
During tube running measure Vdc between TP1 and GND (ground) , then adjust RV1 to have Vdc at
TP3 = 55% of Vdc at TP1.
V(TP3)0.55xV(TP1)
The voltage values of the above mentioned Test Points are shown on the calibration sheet in the
documentation accompanying the unit.
Repeat the operations with TP2, RV2 and TP4.
V(TP4)0.55xV(TP2)
2 – CONTROL LED
LED1 (yellow) bright during normal rotation phase (RUN)
LED2 (yellow) bright during braking or alarm (dark if brake is disabled)
LED3 (yellow) bright during braking (dark if brake is disabled)
LED5 (green) bright in stand by or rotation
LED6 (green) bright during rotation
3 – FUNCTION CONFIGURATION
4 – TROUBLE SHOOTING
If the display with technical information containing this message appears:
ROTOR SUPPLY
FAIL
Check that the LAMBD Acable has been correctly inserted on 92-074 and 01-170 boards.
Check fuses F1 and F3 on 92-074 board.
Check the calibration of the safety devices on 92-074 board.
Check the phase shifter capacitors.
The alarm message may appear in extreme and very rare conditions when the tube temperature has
reached the trip value of the protection thermo-switch located inside it.
FILAMENT
1 – FILAMENT POWER SUPPLY
6 – CHECK-LED
7 – TROUBLE SHOOTING
7.1 Filament power supply
7.2 Calibration fail
7.3 Low anode current
TUBE
LED5
TP5 L S
IFIL AM
RA
GL
EL
RY1 CURRENT FILAMENT
SENSE SAFETY
ERROR POWER RV6/7
1 AMPLI AMPLI
TP6
STK4044 10mA=1khz
5 GEN T2 RY4 TP4
SIN OUTSIN
7 VCA T1
3 XR2206 V
C 12 RESET
F
P
U 42 SRI
CLK DAC
RY3
TP8 R16
6 LD 8043 TP2
VREF VTP2=1V/10mA
IFIL MASMETER
LM331
16 PULSEK
14 PULSEA
8 FILAMENT SELECTION
On power supply board an I/f converter is connected to the CPU 01-170 for a very accurate mAs
integration.
The anode current related to each value kV for each focus depends on the characteristics of the tube
installed and it has to be read from the calibration report delivered with the unit.
To enter mAH current calibration set DIP1 of SW1 ( PCB 01-170) = ON.
To enter mAL current calibration set DIP2 of SW1 ( PCB 01-170) = ON.
In the mAs window of the display, the following parameters will appear:
♦ mAH or mAL, filament current If in Ampere
♦ mAs value selected before entering calibration mode
♦ after each exposure the expected anodic current value and effective Ia in mA delivered.
mAH If 4,710 A
90 Ia 90,1 mA
mAs 40
Over-current
Check if LED1 and LED4 switch OFF after pushing x-ray pushbutton.
In such a circumstance, increase voltage reference at TP1 by means of RV6 for 0.3 mm focus
or RV7 for 0.1mm focus.
Over-voltage
Check if LED1 and LED4 switch OFF after pushing x-ray push button.
In such circumstance increase voltage reference at TP3 by means of RV3 for both focuses.
I(A)
The passage of anode current is indicated by LED5, which during exposure lights up proportionally to
the anode current value.
6 – CHECK-LED
LED1 is inserted in the state signaling circuit of PCB. In normal operation conditions it is
illuminated.
LED2 is on with filament W (only for bimetallic tube option)
LED3 indicates the selection LARGE/SMALL focal spot and it is illuminated when SMALL
FOCAL SPOT is selected
LED4 indicates the state of the filament safety circuit, it is illuminated when filament is power
supplied
LED5 is inserted in the I/F LM331 converter output circuit, it is illuminated only during
exposure and its luminosity is directly proportional to anode current.
7 – TROUBLE SHOOTING
7.1 Filament power supply
The appearance of the technical screen with message:
a) if problem persists after having switched the Unit off and on again, verify the integrity of fuses
F1 F2 and of filaments by an ohmmeter.
b) if problem only arises during exposure, verify any possible intervention of Over-current and
Over-voltage safety circuits.
In case it is necessary to service on PCB it is advisable to replace filament with a car incandescent
lamp of 2x50W with two filaments in parallel.
This load is very similar to a tube filament and allows intervening avoiding risks for filament.
IN THESE CONDITIONS, HIGH VOLTAGE CIRCUIT IS OPEN. DO NOT EXECUTE ANY
EXPOSURE WITHOUT HAVING PRIORY DISABLED THE HIGH VOLTAGE
GENERATOR!!!
To disable the generator, close J1 located on the mother PCB 92-082-6, open fuses and wait until filter
capacitors are discharged.
After an exposure attempt in these conditions you will sound the alarm for insufficient anode current.
7 – CONTROL LED
8 – TROUBLE SHOOTING
1 – DESCRIPTION
High Voltage generator has positive polarity with respect to ground to supply X-ray tube with
grounded cathode.
VDC IN
HV CABLE
HV GENERATOR
25khz
BRIDGE
A
E7
F
E X
E
D T
R U
F
O E8
A B
R 50khz
W CHOPPER
Y E
A
R TP11
D E3
V
d
c
I
N
TP1 E4
P
U K
T
C
- + R16
O 50khz PWM 10ohm 1V=100mA 1V=10mA 100ohm
M
P
E
N
+ -
CURRENT FEEDBACK
S
A
T TP2
I
O
C KV0
FILAMENT CIRCUIT
N E6 O NOT SHOWN
N KV0
T KV1
P R
KV2
E5
KV3 KV7 CPU
WO KV FEEDBACK DAC KV4
KV5
M L CONTROL VOLTAGE KV6
E1 L KV7
E TP7
R TP5
EXPOSURE TIME
INVON
E? X RAY PB
50khz
TP6
The value of high tension required is set by the CPU through an 8-bit opto-insulated bus IC6/IC7/IC8.
It is important to point out that the ground return of the opto-insulated commands is physically
interlocked to the ray button. This means that the commands can be transferred from the CPU to the
high voltage feeder only when this is activated.
The PWM controller regulates the tension applied to the bridge that is necessary to obtain the kV value
set by the CPU.
A loop band width of about 3 kHz and the feedforward of the regulator ensure perfect stabilization of
the high voltage value and totally cancel the double frequency mains ripple from the output wave
form.
The characteristics described above allow for a perfect wave form without overshoot and with a kV
value that is totally independent of the anodic current value supplied to the tube.
The oscillogram shows the high tension form for 20 kV and 35 kV measured on TP1.
The kV wave form does not present overshoot in any operating state and the kV value is independent
of the current fed to the tube.
The current limitation of the chopper ensures perfect protection from current overloads at output and a
constant conductance of the kV rise time typical of 1.5 ms.
KV VTP1/2 (±1%)
20 kV 279 Vdc
21 kV 293 Vdc
22 kV 307 Vdc
23 kV 320 Vdc
24 kV 334 Vdc
25 kV 348 Vdc
26 kV 362 Vdc
27 kV 376 Vdc
28 kV 390 Vdc
29 kV 404 Vdc
30 kV 418 Vdc
31 kV 432 Vdc
32 kV 446 Vdc
33 kV 460 Vdc
34 kV 474 Vdc
35 kV 488 Vdc
The table shows the kV values and the corresponding values measured on TP1/TP2.
NOTE
1. The values in the table are valid if the reference voltage of the kV DAC measured on TP8 is
5.00Vdc (standard calibration).
2. In exceptional instances kV values can be calibrated with the RV1 trimmer (PCB 94-132). In
this instance the reference values on TP1/TP2 are no longer valid. RV1 has influence on the
whole kV range, adjust kV at 28 kV as pivot point.
3. The kV values are not affected by the mains voltage within the specified limits
4. The kV values are not affected by the anodic load within the normal operating range.
5. DIP4 SW1 on board 01-170 must be in the ON position to exclude ray interruption during
measurement. The same effect can be obtained by pressing the [RESET alarm]+[kV+] button, a
valid function only for the next exposure
PCB 94-132
TP1 kV check according to conversion table
TP2 ground
TP3 reference grounding
TP4 +5 Vdc
TP5 exposure time control
TP6 reference frequency 50 kHz or 100 kHz (optional)
TP7 kV control from CPU 1/100 VTP1 (valid in PREP and EXP )
TP8 +5.00 Vdc of DAC reference (standard calibration)
TP9 not used
TP10 internal input voltage +15Vdc
TP11 output current limitation reference
TP12 output current 1Vdc=10A
TP13 reference grounding
7 – CONTROL LED
The green LED1 indicates the soundness of the F2 fuse and the functioning of the internal input at
+15Vdc.
The yellow LED1 turns on during exposure and can be used to verify continuous or pulsated X-ray
emission.
8 – TROUBLE SHOOTING
There are no on-site reparable parts inside the inverter and the HV transformer.
Since damage to the inverter could have been caused by damage to the HV transformer, it is advisable
to replace both of them to avoid the risk of damaging the new Inverter.
On-site checking, to detect if the HV transformer works, is not possible with the normal instruments
provided with the unit.
The Mosfet inverter and the HV transformer can both be repaired at the factory.
If the following message appears:
And the Vdc in is much less than 550 V, check the input power of the inverter and associated fuses.
Check the voltage and the control circuit of the power input.
The control circuit can be excluded during testing by short-circuiting the JP1 bridge on the CPU 92-
082-6 motherboard.
Carry out the kV check and if the voltage on TP1/TP2 is zero, the inverter must be replaced.
2 - PROCEDURE PREPARATION
2.1 Software installation
2.2 Developer
2.3 X-ray cassette for tests
2.4 Film pack for tests
2.5 Test phantom
2.6 O.D. reference X-ray
2.7 Time for carrying out procedure
2.8 Anode filter combination
2.9 Detector calibration
2.10 Offset [o] Calibration
2.11 Detector [r] Gain Calibration
2.12 Connecting PC for calibration
1.13 Selection of test layers
1.14 Other Techniques and operating modes
1.15 Automatic Molibdenum Rhodium filter
1.16 Modification of the reference Density after calibration
1.17 Maximum breast density limit
2.18 Calibration concept
2.19 Linearity check
2.20 Unused Operating Technique and Film Screen
2.21 Disabling specific Operating Technique and/or Film Screen
3 - CALIBRATION PROCEDURE
3.1 Zero Point calibration
3.2 Manual Density steps
3.3 One point calibration
3.4 Rhodium filter
3.5 Maximum breast density limit
3.6 AEC calibration files backup
The figure above shows a time schematization of the exposure concept in ZERO POINT technique,
with completely automatic kV and mAs.
The parameter [p] derived directly from the [r], the density of the breast being examined, is used as a
pointer on two look-up tables containing the kV and mAs necessary for obtaining an image with the
right blackening and contrast to fully satisfy any Quality Protocol in the Mammography field today.
A high value [p] corresponds to a LOW Density Breast and low values of kV and mAs are delivered.
A low value [p] corresponds to a HIGH Density Breast and high values of kV and mAs are delivered.
kV values can’t be modified, but four different ranges are made available to the operator, with
different kV profiles suited to different requirements that go beyond the possibilities offered by the
well-known High Contrast (HC) and Low Dose (LD) techniques.
From the control panel, the operator can not only select the technique that he feels most suitable but he
can also make fine-tuning in steps of 0.5kV on the kV range selected.
The figure above is a diagram of the operating possibilities of the DEATA PLUS software with Mo
target / Mo filter or Mo target with automatic Mo/Rh filters.
The large number of Film Screen and techniques available have to be intended to give the possibility
to tune the system to the specific requirements of any single or multiple user and or different film that
can be momentarily found in critical markets without loosing the possibility to go back to the
previously used when newly available.
For maximum image quality, it is advisable to use only film/screen combinations recommended by the
manufacturer.
The United States regulations have made this recommendation a must.
The X-ray cassettes for mammography use have only one intensifying screen that converts the X-rays
to light that impresses the image on film.
Every screen has its own x-ray/light conversion factor that must be suited to the film to be used.
Films for mammography use are single emulsion and have various sensitivities and contrasts.
An appropriate choice of these products is reflected in image quality.
We leave this responsibility to the user, dutifully pointing out the difference that it can represent in
terms of x-ray Dose delivered to the patient.
2 - PROCEDURE PREPARATION
2.2 - Developer
The developer must be in perfect conditions for use with developing and fixing replaced a day
beforehand.
This configuration was selected in order to eliminate all the variable components that could be placed
between the beam source and the detector.
Carry out an exposure and then check that [r] value is within ±10 of the value from the phantom label.
At the same time check that [o] value is higher than 8 and lower than 12.
If out of range adjust according to the following points 10 and 11.
Normal operating condition is restored when cassette is inserted in the table.
The value must be checked after at least 5 minutes from turning the unit on and it should be the same
as the value shown on the test phantom +/-10.
O Î offset; G Î Gain ([r] on the display).
Service
socket
If the mammography unit has not a service socket on the back side, insert the dedicated cable into the
socket on the PCB of the mammo (PCB 92-076 or PCB 01-170).
LS-DWG729-00
Use the 5mm Plexiglass layer or even remove/insert the compression plate in order to accommodate p
(plog)
Repeat the test for all thicknesses that are to be used.
3 - CALIBRATION PROCEDURE
Two different online Help are available
♦ Verify that appropriate Deata Type is defined according to the configuration of the specific unit you
have to calibrate.
Press Download
Reference O.D. can also be selected later from inside Zero point calibration curve procedure.
and start procedure from one of the phantom thicknesses having Manual Density set to ZERO on the
control panel of Mammo unit.
Automatic
exposure parameters
To accept click Save to File and transfer new curve to the unit.
Values in the table are saved to file and can be recalled by means of Reload Last button
Important NOTE
If default curve used for calibration is too much different from characteristics of Film/screen to be
calibrated, may happen that manual density adjustment is not wide enough to obtain High and Low
densities for each phantom thickness.
In such circumstance it’s necessary to modify the default curve and store it again in the unit before to
proceed with next points/thickness.
Calibration curve can be modified editing points manually or by means of %O.D. correction
to obtain coordinates Plog (p) and Pulse (i) to build up the calibration curve.
In Manual calibration procedure curve shape must be manually adjusted for plog = 0 and plog = 159
being the extremities of the curve.
♦ For each phantom thickness used for calibration, make a test exposure and check that O.D. is now
within requested tolerance.
select automatic
Check with exposures that step O.D. changes according to selected value.
< than kV in zero point mode (dkV between -14 and -1), or
> than kV in zero point mode (dkV between +1 and +10).
If they are > than kV in zero point mode, they can also be:
< than 30 kV
>= of 30kV.
For each specific condition correction factors % are defined to compensate O.D. in function of kV
difference between Zero (CPU) and One point (operator).
Select % value to change in the appropriate column. If after exposure, d parameter from
communication window was +2 , One point was 2 kV higher than Zero point kV
If measured O.D. was outside tolerance, % can be adjusted with the same concept of High and Low
O.D. as seen during the Zero point calibration.
values must be filled manually, when clicking OK, % in the selected cell will be automatically
adjusted.
Repeat the same procedure for all points with: fix phantom and variable kV, fix kV and variable
thickness according to the next forms.
PHANTOM 45 mm 26 kV 28 kV 30 kV
Mas
plog(p)
pulse(i)
d(kV)
O.D.
28 kV 20 mm 30 mm 40 mm 50 mm
Mas
plog(p)
pulse(i)
D(kV)
O.D.
If number is decreased RH filter will come in for larger breasts and vice versa.
The concept of this paragraph is extended if a double trace tube is present.
If [p] value will be lower than the defined limit, “ Breast too dense” alarm will come on and exposure
is stopped releasing only the dose of pre-exposure x-ray pulse.
The variation of percentage correction value is done with keys mAs+ and mAs- and the relative value
appears on the display in the place of the Manual Density value.
Regulate the Optical Density on a Medium (4cm) phantom thickness, repeat the sequence correction /
exposure / Density check until obtaining the desired value.
After adjusting the Density to 4 cm, make a Film with 2 cm and one with 6 cm of Plexiglas.
Compare the three Films and if the darkness is insufficiently constant, repeat the test with another
Film/Screen combination on control panel until obtaining the best compromise.
In this configuration the unit has 3 preset curves, corresponding to film/screen combination 1, 2 and 3,
in order to have different slants that allow a limited and empirical linearization of the Optical Density.
When the O.D. is sufficiently constant, exit the procedure putting the DIP6 SW1 OFF
Putting DIP 5 SW1 on PCB 01-170 to ON, it is possible to freeze the selection from control panel to
only one film-screen (for example the number 1) and only one operating technique (for example the
STD). This is useful if only one film-screen and operating technique have been calibrated (both with
DEATA PLUS, and with REDUCED DEATA PLUS).
DOSE CALCULATOR
1 - INTRODUCTION
Note: for the Average Glandular Dose calculator the calibration is the same.
WARNING!
If for any reason, proper operation of Dosimeter can’t be guaranteed, IT’S RECOMMENDED TO
TOTALLY DISABLE it
1 - INTRODUCTION
Dose calculator evaluates the skin entrance dose knowing :
1. kV of exposure
2. mAs supplied
3. distance from Focus to skin (resolution 5mm)
4. measured dose with dosemeter at a distance of 55 cm from Focus for each kV value and for
each combination of anode and filter
The dose is calculated at the end of an exposure.
The obtained dose value is shown on screen. It can be printed on Film by ID printer and recorded in
the unit for statistical purposes.
The Dose is visualized in mGy with a resolution of 0.01 from 0 to 9.99 mGy, and 0.1 from 10 to 99.9
mGy. Beyond a 99.9 value, HH.H will be displayed.
The ID print function on Film can be enabled or not by the user or by the Health Physicist. Calibration
and periodic maintenance are based on external calibrated reference Dosimeter.
So that you don’t need to create a new skin dose file for a mammo unit and you cannot delete it.
The Deata Plus program does this for you, with a default set of values at creation, and deleting the skin
dose file when a mammo unit is deleted using the Deata Plus program.
All modifications that have been made to configuration file are either immediately
saved to file or retained in memory. At any point in the program you can save all
this modifications, regardless partial savings occurred before. Press "File"-"Save".
To view and edit the radiation table for the selected mammo
press the "Edit"-"Radiation table" menu.
Data are presented in a "kV" versus reference dose grid. Any row correspond
to a point of the curve. The kV steps are fixed, so this column is unselectable.
Data are shown (and accepted) in µGy or mR, depending on the radiobutton
selected in the "Radiation unit" radiogroup, and for 1mAs or 50mAs as selected
in the corresponding radiogroup, while the graph is always constructed with
µGy for 1mAs. To view and edit data for a target/filter combination just click
on the corresponding tab. The radiation unit and mAs are separately selectable
for each target/filter.
You can specify the dosemeter used in measures. With this information
(and related tuning parameters, see "Dosemeter data") the program can
tune dose values inserted by user to obtain effective values to download
to mammo.
The choosen instrument name fills the corresponding edit field. You
can specify a dosemeter for each target/filter combination.
With this option you can set some operational parameters. Press the "Edit"-
"Dosemeter configuration" menu.
DOSE FUNCTIONS
Dose shown on the LCD Display as well stored in the
memory for statistic purposes and print to the ID
printer can be enabled/disabled in the specific boxes.
Data are presented in a "kV" versus "Correction factor" grid. Any row
corresponds to a point of the curve.
Finally, once all changes are made, you can save modifications to file...
4.2 - Options
COMPRESSION SYSTEM
1 – INTRODUCTION
1 - INTRODUCTION
The compression device consists of a drive motor with gear box coupled to an electromagnetic
clutch.To access drive motor, remove tube cover.
To access electromagnetic clutch remove tube cover and C-arm front plate.
Compression force is adjusted by means of current flowing in the electromagnetic clutch that also has
safety function.
Drive motor has current sensing circuit for speed reduction when compression plate gets in contact
with the breast.
For function description refer to Operator’s manual.
SW1
DIP7 = ON Compression force calibration
Insert shifted compression plate, position a bathroom balance on the table and a soft cushion between
plate and balance as alternate method, spring balance can be used.
Compression force can be adjusted changing the correction factor from 1 to 20 that will appear on the
display in place of compression force.
Correction factor can be changed by means of mAs [+] [-] pushbuttons.
To the lowest correction factor corresponds the highest compression force and vice versa.
MOTOR WITH
GEAR BOX
MICRO
CALIPER SWITCHES
D
Electrical Connections
Because of the various electrical connections, it is not recommended to replace a tube by another
having different characteristics.
Firmware
Each tube needs its own specific Firmware to comply with load values.
Check of Position
To check the position of the field, use a film inserted into the cassette of the 18x24 potter and another
film inserted in a film envelope without screen located diagonally on the potter plane, so that its four
corners can record the radiation external to the edges of the first film.
Also position two coins above the envelope to create two reference points on radiographs allowing
their superposition.
RING
Note: the picture refers to LAMBDA E, for LAMBDA the same concept applies.
SWITCHING ON CIRCUIT
1 – POWER SUPPLY
2 – ON/OFF PHASES
1 – POWER SUPPLY
A completely independent circuit, equipped with its own microprocessor, controls all on/off phases of
the unit.
The circuit is located on the 92-082 board and is supplied by the T1 transformer connected to the
mains supply.
The transformer is equipped with a 220/230/240 voltage selector.
For the voltage selection refer to chapter 2.
2 – ON/OFF PHASES
The IC20 microprocessor controls a complex sequence of events at switching on and off.
SWITCHING ON
a) to avoid a high current inrush, RY2 is switched on and after some seconds RY1 excludes the
damping resistance.
b) to avoid random characters on screen, LCD voltage is excluded.
SWITCHING OFF
To avoid stripes on screen, LCD voltage is removed before +5 voltage of CPU.
RYS2
+5
-15LCD
RYS1
ON OFF
FILTROMAMM
Mo/Rh AUTOMATIC FILTER
1 – DESCRIPTION
2 – ELECTRICAL COMPONENTS
3 – INSTALLATION
1 – DESCRIPTION
The Mo/Rh automatic filter is an optional accessory and it can be easily installed further on.
Filtromamm is used to have a double ANODE/FILTER combination (Mo/Mo Mo/Rh) for a better
optimization of the X-ray beam spectrum, in function of the real density of the examined breast.
The breast density is quickly determined by a short exposure in ZERO POINTS technique and the
more appropriate ANODE/FILTER combination will be automatically selected according to the breast
density.
The Mo/Mo combination is used for exams of normal density breasts, whereas the Mo/Rh combination
is used for high density breasts.
The intervention threshold beyond which the Mo/Rh combination is selected, corresponds to
approximately 6 cm of RMI Equivalent Tissue and [r] = 62 for large focus (default).
After the sampling pulse, the value [r] is compared with the reference level. If [r] is higher than the
reference value, Mo/Mo remains positioned, if [r] is lower, the RY1 / RY 5 circuit on the PCB 92-082,
change the filter in the dead time between the sampling impulse and exposure.
The filter change is temporized and it does not depend on the status of MW12.
At the end of this time, MW12 is checked and if the filter has not reached the final position, an alarm
is generated.
2 – ELECTRICAL COMPONENTS
Motor MT4 24 Vdc
Micro-switch MW10/11 in/out detectors
MW12 FILTER IN POSITION (detected by the CPU)
!!Mo/Rh FILTER VERSION ONLY
!!SOLTANTO SU VERSIONE CON FILTRO Mo/Rh
1
2
+Mo MT4 3
D5 D7
4
Mo Rh
MW12
FILTER OK
D6 D8
3 – INSTALLATION
Filtromamm has the same dimensions as the previously existing aluminum tube support.
Before installation, remove the existing support, the Lead ring and the fixed filter in Molybdenum.
Then position the Tube again as described in relevant manual section.
Electrical connections are already prepared and to activate it move DIP 8 SW1 on PCB 01-170 to the
ON position to enable the FILTROMAMM.
2 – CONNECTIONS
1 – DESCRIPTION
The unit, configured for a three-dimensional Biopsy with BYM-3D, can be equipped with an interface
to the 50x50mm small field digital image acquisition system with a digital box having the same
dimensions as those of a 18x24 cassette.
The digital interface is an optional of the standard configuration of the unit and it can also be added
after installation.
2 – CONNECTIONS
Connections of the whole mammography system fully equipped with Digital Cassette, BYM-3D and
PC with Digital Imaging software can be seen in the picture in the following page.
Before connection, you need to know some important issues to maintain the electrical safety level of
the system.
This system consists of the below listed and classified elements:
As you can see in this picture, I, H, L signal connections, are all optically isolated to avoid any direct
electrical connection between Medical Devices complying with IEC 601-1 and the others.
Connections shall be in compliance with these instructions not to modify the electrical safety level of
the system.
NOTE
Connections from PC to the Mammography unit and the Stereotaxis device are made by two CV0131
(short) and CV0132 (long) cables between which the optic isolator is inserted.
The CV0131 cable SHALL be positioned on the PC side for both connections, otherwise it will not
operate
LAMBDA/ EASYLABEL
COMMUNICATION INTERFACE
1 >
2 5 Line number on EASY LABEL screen
3
4 X Kv tens
5 X kV units
6 .
7 X kV tenths
8 k
9 V
10
11 X mAs hundredths
12 X mAs tents
13 X mAs units
14 .
15 X mAs tenths
16 m
17 A
18 s
19
20 . .
21 X Dimension of focal spot
22 X
23
24 X type of ANODE Mo, W
25 X
26 /
27 X type of FILTER Mo, Rh
28 X
29 Y tecnique kV ZERO POINTS s=Standard h=HIGH CONTRAST
l=LOW DOSE e=EXTENDED
30
31 X G = with grid space = without grid
32
33 X 0P = ZERO POINTS 1P = ONE POINT
34 X
35
36 D manual density
37 X +/-
38 X number of step from –5 a +5
39 DOSE IF ENABLED
40 F film/screen
41 X film screen number from 1 to 15
42 X
43 $5C
44 $0D
CHAPTER 4
PERIODICAL MAINTENANCE
TROUBLE SHOOTING
PERIODICAL INSPECTIONS ...................................................................................148
LIST OF FUSES ........................................................................................................154
DIP SWITCH CONFIGURATION ..............................................................................154
CONTROL LEDS ......................................................................................................155
CPU RESET SEQUENCE .........................................................................................157
BOARDS CONFIGURATION (JUMPERS, VARIABLE RESISTORS AND TEST
POINTS)....................................................................................................................157
SUBASSEMBLIES AND PCB REPLACEMENT ........................................................159
SWITCHING ON CIRCUIT AND SEQUENCE ...........................................................163
TABLE DETECTION AND GRID TEST .....................................................................164
FUNCTIONAL BLOCKS ............................................................................................165
SWITCHING ON PROBLEMS...................................................................................167
OPERATING ERRORS ON DISPLAY .......................................................................168
ERRORS IN THE TECHNICAL MENU ......................................................................175
TECHNICAL ERRORS IN LOG FILE.........................................................................178
MEMORY STRING....................................................................................................179
LIST OF ERROR CODES..........................................................................................180
WARNING
Maintenance and handling of the device must be performed by
qualified personnel only
PERIODICAL INSPECTIONS
SUGGESTED EVERY 6 MONTHS, RECOMMENDED EVERY YEAR
LAMBDA s/n_________________
PLACE OF INSTALLATION _____________________________
DATE OF INSTALLATION _____________________________
INSTALLATION COMPANY _____________________________
RESPONSIBLE CONTACT PERSON _____________________________
DATE______________ SIGNATURE_________________________
8) Set hour/date
If the date and time are not correct, set it.
Refer to Chapter 2, Date And Time Setting
10) Read out the last 1300 exposures and delete memory
Connect the Service PC to the mammography unit and download the last 1300 exposures on the PC.
Verify that there are not many errors, especially of the same type. If everything is OK, delete memory.
Refer to related paragraph of this manual.
At the end of inspection, clean the unit using the materials recommended in the Instructions for
Use and repair any damaged paint.
LIST OF FUSES
All the fuses of the mammography unit, both in fuse holders and on the PCBs, are listed in the chapter
5.
Refer to chapter 5.
CONTROL LEDS
CPU 01-170
Color Function
LED1 YELLOW Don’t care
LED2 YELLOW blinking during x-ray
LED3 YELLOW bright during x-ray ( mAs meter pulses)
LED D1 RED IRQ activity monitor, blinks in normal condition; if bright fix CPU
hanged UP
LED DU1 YELLOW internal +2,5Vdc
LED DU2 YELLOW internal +3,3Vdc
LED DU6 RED ON at start UP during FPGA initialization
LED DU8 RED ON during CPU initial reset
LED DU11 YELLOW +5Vdc
Motherboard 92-082-6
Color Function
DL1 GREEN - 15 Vdc
DL2 GREEN + 15 Vdc
DL3 GREEN + 24 Vdc
DL4 GREEN + 5 Vdc
DL5 RED X rays ON
DL6 GREEN - 15 Vdc delayed
DL7 GREEN + 5 Vdc delayed
DL8 RED Grid movement request (from CPU)
Color Function
LED1 YELLOW Bright during tube running
LED2 YELLOW Bright during brake or alarm
LED3 YELLOW Bright during brake
LED5 GREEN Bright in Stand by or rotation
LED6 GREEN Bright during rotation
Color Function
LED1 GREEN Bright fix in normal condition
LED2 RED Anode W
LED3 RED Bright when Small focal spot is selected
LED4 GREEN Bright with filament ON
LED5 YELLOW bright during exposure. Brightness is directly proportional to anode
current
Color Function
DL1 GREEN +15Vdc
DL2 YELLOW bright or blinking during exposure depending on continuous or
pulsed mode
Motherboard 92-082-6
Function
JP1 Close to exclude the inverter power supply monitor/alarm during tests
JP5 If closed stand-by LED is ON with unit OFF. If Open stand by LED is OFF when
unit is OFF
JP6 Open for LAMBDA SE and LAMBDA with stand-by Lamp.
Closed for LAMBDA Ewith stand-by LED.
JP7
JP8
JP9 Reserved. Not for service
JP10 Reserved. Not for service
JP11 Reserved. Not for service
JP12 Reserved. Not for service
JP13 Reserved. Not for service
JP14 Reserved. Not for service
JP15 Factory preset closed (Bucky driver)
JP16 Open for Skin Dose display. Closed for AGD display
JP17 Closed for proper operation of 92-082-6
JP18 Closed to limit compression force to 15
JP19 Not in use
JP20 Factory preset (closed disables the delay of the + 5 V dc delayed)
JP21 (3 PIN) Pin 2-3 closed with IC20 PCB 92-082-6 = MOTOR14 $BAD4 for Bucky with
double LED yellow and green. Breast thickness indication not correct with 4-
INFL-XX (1,8x magnification device without grid).
Pin 1-2 closed with IC20 PCB 92-082-6 = MOTOR15A $8B81 for Bucky with
single green LED. Breast thickness indication is correct with 4-INFL-XX (1,8x
magnification device without grid).
JP22 o Open for LAMBDA E(electromagnetic brake).
o Closed for LAMBDA SE and LAMBDA (disk brake) (if open brake
doesn’t position properly)
RV1 Calibration of DVM for inverter power supply voltage located on technical menu .
RV2 Reserved. Not for service
RV3 Factory preset (reference voltage for compression plate speed reduction)
Setting Function
R16 Anodic current sensor resistor 1Vdc=10mA (1Vdc=5mA options
5kW or bimetallic tube)
RV3 Calibration of filament safety (voltage)
RV5 Reserved. Not for service
RV6 Calibration of filament safety (current)
RV7 Reserved. Not for service
Setting Function
RV1 Calibration of capacitive phase safety circuit
RV2 Calibration of direct phase safety circuit
TP1 test-point of capacitive phase current
TP3 VTP1 – 45%
TP2 test-point of direct phase current
TP4 VTP2 – 45%
JP1 OPEN braking enabled
JP3 OPEN braking time 6”
JP4 OPEN safety enabled
Setting Function
TP1/TP2 Tension test during exposure
RV1 Regulation kV
RV2 Regulation maximum output current
TUBE
When replacing a tube, with another one of same trademark and model, the following operations are
necessary:
kV check and eventual calibration
Rotating anode safety circuit check and eventual calibration
Filament safety circuit and eventual calibration
Also take care of the alignment of the tube.
We advice against the replacing of the tube with a different model. In this case also the firmware
version of the EEPROM on the CPU board 01-170 must be changed.
H.V. GENERATOR
When replacing a H.V. generator it is necessary to check that the kV used for the exposure correspond
to the kV set on the control panel. Refer to the related chapter.
INVERTER MODULE
When replacing a H.V. generator it is necessary to check that the kV used for the exposure correspond
to the kV set on the control panel. Refer to the related chapter.
Replacing the board it is possible to take the three chip from the old board and insert them on the new
one. So the firmware, calibration and other information are maintained.
If you change the EEPROM with a new one having a different firmware version, also take care of the
display firmware version: they must be compatible
If you change the battery RAM containing the calibration data with a new one, you need to calibrate
the filament current, the optical density (DEATA PLUS calibration or REDUCED DEATA) and the
dose calculator.
Also take care of the jumpers configuration.
KEYBOARD (PCB/92-083-X)
On this board the language and the contrast of the display are set. Take care of this two aspects. Refer
to related chapters.
Pay attention
Pay attention to
Code Description to Firmware
Calibrations
versions
X- ray tube Yes, several Yes, CPU
Yes, High
KTFLI0009 H.V. Transformer NO
Voltage
Yes, High
KTFLA0206 INVERTER module NO
Voltage
DS/INVC132 Display SP14N003 power supply NO NO
DS/LMG7401 SP14N003 DISPLAY NO NO
Depending on
PCB/01-170-X CPU Board replacement of YES
the battery RAM
PCB/02-171-0 TUBE Temperature sensor NO NO
Automatic Collimator and C-ARM Service
PCB/02-179-X NO YES
board
PCB/96-153-X Mirror motor driver NO NO
PCB/03-181-X CPU Keyboard NO YES
YES (two
PCB/92-082-6 CPU Mother board YES
firmware)
PCB/92-073-X Filament power supply YES NO
PCB/92-074-X Anode starter YES YES
PCB/92-083-X Keyboard NO NO
PHTM9000 AEC Detector YES NO
Switch On sequence
Main contactor CT1 switches ON at first and dumping resistor R1 smooths inrush current of TF203
and capacitors C1,C2,C3,C4.
After few seconds RY1, with zero cross drive, shorts R1 and full power is available for H.V. generator
and other high power services.
For further details refer to chapter 3, “switching on circuit”
Emergency shutdown
Emergency push buttons located on both sides of the unit.
Operating Emergency push button will switch OFF immediately main contactor CT1 and the whole
unit is switched OFF.
Table driver
Table driver is onboard of PCB 92-082 motherboard, it contains both cassette sensing + leds driver for
all tables and stepper motor driver for grid tables.
Table detection
The following tables are detected by the CPU in order to guarantee the appropriate functionality.
Model Code
18x24 gridless PMX 1500-XX
18x24 grid PMX 2500-XX
24x30 gridless PMX 4500-XX
24x30 grid PMX 3500-XX
18x24 gridless 1,8 X 4-INFL-XX
magnification device
Cassette detection functions are the same for all the tables.
Green LED is bright fix when unexposed cassette is in. Unit could not be ready for exposure
depending on Door Open and/or no collimation plate.
Yellow LED is bright fix if cassette is not inserted or not properly inserted or exposed.
Green and yellow LED blink alternatively if cassette is already in at power on.
If cassette notch sensing switch is enabled and cassette has no notch, yellow Led is bright even if
cassette is properly inserted.
Note: the color and number of LEDs on the table can vary depending on the firmware version.
Grid tables.
If grid table is recognized, x-ray beam is pulsed. Both large and small focus can be selected from the
control panel.
Gridless tables.
If gridless table, any kind, is recognized, x-ray beam is continuous, small focus is selected
automatically and deselected when gridless table is removed.
FUNCTIONAL BLOCKS
For a better comprehension of the unit, you must know that this unit is subdivided into several
operational blocks:
1. Main CPU, mother board
2. Control keyboard and display
3. Rotating Anode Starter
4. Filament power supply
5. Inverter of High Voltage generator
The main CPU (1) and the control keyboard (2) are equipped with independent microprocessors,
connected among them by a serial interface.
Blocks (3), (4) and (5) are connected to the main CPU by Optoisolated digital interfaces.
Supervision circuit
A general supervision circuit checks the integrity of various fuses and complete functional blocks
whose state is indicated in the TECHNICAL MENU.
Blocks simulation
Blocks (3), (4) and (5) can be isolated from the main CPU for test purposes.
To isolate a block, it is necessary to close the supervision circuits from the CPU side so that it detects a
normal condition.
Here follow the necessary connections to simulate:
Inverter
Connections can be realized directly on the board 01-170 excluding the LAMBDA connection cables
or directly at the end of the LAMBDA cable to also include it in the test.
Under these conditions, it is NOT possible any exposure.
SWITCHING ON PROBLEMS
If unit doesn’t switch ON check if one or both emergency buttons are active.
Check the main fuse of the switch or fuses referred to board 92-082 and relevant switching on circuit.
Refer to the special section in Chap. 3 on this manual.
If unit can be switched ON but no Frame and no Icons appear on the screen and CPU reset sequence is
correct check keyboard 92-083, KEYBOARD CPU 03-181, cable CV0052 and supply voltages
onboard of KEYBOARD 92-083
If unit can be switched ON, Frame appears on the screen, CPU reset sequence is correct and no Icons
appear on the screen, press the x-ray pushbutton and if intermittent beep is generated, serial
communication between CPU and Keyboard is not working.
Check connection cable, line drivers, replace 92-083 and/or 01-170
.
If the display is empty wait few seconds for the start up sequence and after check if the CPU is alive
by means of C-arm functions.
If C-ARM can be rotated or moved up/down, CPU and Keyboard are working and the problem can be
due to the display itself.
Why Pushbutton or foot pedal switch of mentioned functions remains activated for more than
20”
Action Check pushbuttons or foot pedal switches and eventually disconnect them from the boards
for easier investigation.
Note Alarm goes OFF by itself if normal condition is restored releasing pushbutton or foot
pedal switch.
Note Alarm goes OFF by itself if normal condition is restored removing cassette
Why Door of Examination room was open when operator pressed x-ray push button
Action Close door or check door contact and / or wiring to PCB 03-186.
Note Alarm goes OFF by itself if normal condition is restored closing the door.
Exposure is not interrupted if door is open after that exposure sequence started.
Action Insert collimator or check sensing microswitch and / or PCB 02-179 and its connection
cable to main CPU CV0158
Why Cassette is not inserted or already exposed or cassette sensing switches are defective
Action Check if mirror doesn’t move totally or partially for mechanical or electrical problems.
Mirror drive motor has position sensing switch inside that must switch (audible click )
when is at 60 degrees with respect to the base plate.
Mirror driver is onboard of PCB 02-179 connected to the main CPU by CV0158.
Note Available only with PCB 02-179, C-arm service board (instead of PCB 96-153 mirror
motor driver)
Action Select manual mode, select filters several time to investigate if it’s mechanical or electric
problem.
Why Film-screen or technique STD, EXT, HC, LD selected on control panel is not
programmed.
Action Select programmed film-screen and technique STD, EXT, HC, LD If only one is
programmed, freeze it by means of DIP 5 SW1 PCB 01-170.
If problem persists it’s probably due to memory corruption, transfer again AEC calibration
curve to CPU memory.
Action Inspect pushbutton at sight, select high mAs value and during exposure force: cable,
connector and pushbutton to locate possible false contact.
Replace pushbutton.
Action Check anode current calibration according to specific section of this manual.
Also check kV value according to specific section of this manual.
Alarm can also depend on malfunction of V/F converter onboard of PCB 92-073, event
counter on board of PCB 01-170 and cable CV0072.
To investigate mAs integrator, check LED5 PCB 92-073 and LED3 PCB 01-170
For further investigation, external Dose meter or radiographic cassette with film can be
used to detect if x-ray are generated or not.
LOW ANODE CURRENT could also be due to H.V. generator defect or failure, check it
as described in dedicated section of this manual.
Why Breast and / or detector position is/ are wrong, part of detector was exposed to direct x-ray
beam.
Breast or phantom density was too low.
Action Check for proper breast / detector positioning or thin phantom.
Why Breast density is too high and [p] value lower than limit selected during AEC calibration
Breast could have silicone prosthesis.
Refer to AEC calibration procedure for [p] lower limit calibration.
Action Reduce breast density and check for AEC calibration with reference phantom.
Why The exposure has been terminated by the timer because exposure requested a mAs value
greater than the tube can handle. Repeat exposure in ONE POINT mode by raising KV or
by using a more sensitive screen-film combination.
Could happen with magnification and small focus if grid table is used and breast is large
or very dense.
Can also happen for calibration problems or system malfunction.
Action Evaluate operating conditions and eventually use One point mode with higher kV.
Note Quite impossible with large focus, if happens check very carefully mA, kV, [r] and [o]
calibrations.
Why Broken Fuse or defective block has been located by CPU diagnostic functions
Fault INVERTER
Action 1) if problem is LV=NOK and LED DL1 is Dark replace F2
2) if problem is LV=NOK and LED DL1 is Bright check LAMBDA cable CV0074 and
eventually simulate inverter block by means of Jumpers on CN3 PCB 01-170
3) If technical menu appears with Vdc (in)= 000 then F3 of PCB 94-132 is broken,
inverter block 6 and H.V. transformer must be replaced.
4) If technical menu appears with Vdc (in) = 270 V F6 or F9 is broken.
For investigation of INVERTER Vdc problems, sensing circuit and alarm can be disabled
by means of JP1 closed on PCB 92-082-6.
Vdc voltage sensing is connected at CN4 PCB 94-132 and CN9 PCB 92-082.
Normal voltage across CN9 with cable connected is about 19,5Vdc.
Voltage across CN4 PCB 94-132 is about 47 Vdc with cable disconnected.
Vdc (in) drop is about 100Vdc during exposure and ripple in the same condition is
typically 50 Vpp.
If Vdc drop and ripple voltage are much higher capacitors C1,C2,C3,C4 must be replaced
Note For kV problems refer to High Voltage generator section in this manual
CPU of Mammography unit has a memory with the last 1300 events including errors and fault
events. This 1300 events can be uploaded on the PC.
insert a comment, press Start and press the two buttons Alarm reset and Optical density plus
on the mammo at the same time.
Now you can see, on the “Uploading statistical info” window, the data being uploaded. When
the mammo finishes to transmit data, it emits three quick tones to advise you. If you want to
stop the transmission push the two buttons Alarm reset and Optical density minus
contemporarily. If you want to delete all the 1300 events in the mammo, press the three
buttons alarm reset, optical density plus and optical density minus contemporarily.
DEATA Plus software creates a LOG file that can be opened clicking on View Stat button.
MEMORY STRING
The following string is an example of data available for each exposure or fault.
----------------------------------------------------------------
>5 26.5kV 126.6mAs .30 Mo/Mos G 0P 27.2mGy\
>6 r0110 p076 i0472 d 0 c100 %100 o011 C1 \
>7 kS206 r0110/0110/0110/0110 38°C Af [12/03/03 15:20:02]\
----------------------------------------------------------------
It’s referred to exposure in Zero Point mode with code Af ( no error in automatic mode)
CHAPTER 5
ELECTRIC DIAGRAMS
LIST OF FUSES
GENERAL SCHEMATIC DIAGRAM .........................................................................182
92-082-6 MOTHERBOARD.......................................................................................184
01-170 CPU BOARD .................................................................................................195
92-083-8 KEYBOARD / CONTROL PANEL...............................................................198
92-073-5 FILAMENT POWER SUPPLY ....................................................................203
92-074-3 ROTATING ANODE STARTER..................................................................206
96-153-0 MIRROR MOTOR DRIVER ........................................................................209
02-179-1 AUTOMATIC COLLIMATOR MIRROR ACTUATOR, BEAM LIGHT AND
FILTROMAMM ..........................................................................................................212
LIST OF FUSES ........................................................................................................215
CABLES ....................................................................................................................218
92-082-6 MOTHERBOARD
Main functions:
Unit ON/OFF drive
C-ARM up/down motor
Compression motor drive, magnetic clutch, foot pedal switches
Collimator light, collimator presence sensing switch
Rotation brake
Mo/Rh filter
Exposure counter
X-rays esternal lamp
Door safety switch
Potter-bucky circuit
X-rays pushbutton
Calibration:
For Led, Jumpers, variable resistor and Test points refer to chapter 4
Calibration:
no calibration is requested
Replacement:
EEPROM contains specific firmware
2 k RAM contains panel settings and CLOCK
512 k RAM contains calibrations
Calibration:
RV1 LCD contrast
Calibration:
refer to chapter 3
Replacement:
anode current calibration
filament safety circuit calibration:
Refer to chapter 3
Calibration:
refer to chapter 3.
Replacement:
safety circuit calibration (RV1, RV2)
voltage selection (220V - 230V - 240V)
Refer to chapter 3.
Warning:
Circuit ground is floating with respect to other blocks in the unit having the same
symbol.
Be careful making measurements on this PCB, use only insulated instruments.
Calibration:
no calibration is requested
Replacement:
no adjustment
Main functions:
Mirror actuator
Centering lamp
Automatic collimator
Automatic Mo/Rh filter FILTROMAMM
Calibration:
no calibration is requested
Replacement:
no adjustment
Note:
this board is present when the unit has automatic collimator.
LIST OF FUSES
FUSE HOLDERS
Numeration is referred to general schematics
F1 = mains power
20A 10×38 WEBER CH10 CEI 32-1-4 IEC239
F2 = Main contactor
0.1A slow blow 6.2×30
F3 = c-arm motor3.15 delayed 6.2x30
F4 = toroidal transformer primary TF205 – 3.15A delayed 6.2 X 30
F5 = Power on circuit on PCB 92-082 40mA delayed 5×20; F8 is in series to F10 on PCB 92-
082.
F6 = toroidal transformer secondary TF203 – 10A 10X38 weber CH10 CEI 32-
1-4 IEC 239
F7 = toroidal transformer secondary TF203 – 10A 10X38 weber CH10 CEI 32-
1-4 IEC 239
F8 = External lamp 500mA delayed 6x30
TF 203 Protected by termoswitch that acts on mains breaker turning off the device and
avoiding self power-on at the termoswitch restoration.
Further protection against transient overload by fuses on primary and secondary
circuits.
CABLES
CN 14 CN 13
CN15
CHAPTER 6
SPARE PARTS LIST
WARNING
♦ G.M.I. recommends the use of original spare parts,
otherwise declines its responsibility for any damage to the
device and/or to third party.
CODE DESCRIPTION
Electronic PCBs & subassemblies
DS/INVC132 LMG7401 INVERTER POWER DISPLAY
DS/LMG7401 LMG7401 DISPLAY –PLGE (BLUE)
FL/4192209500 COGEMA FILTER 823.020
KTFLA0009 H.V. GENERATOR
KTFLA0206 INVERTER MODULE
PCB/01-170-1 CPU BOARD
PCB/02-171-0 TEMPERATURE SENSOR RX TUBE
PCB/02-172-X POTTER BUCKY DRIVER
PCB/03-181-X CONTROL PANEL CPU BOARD
PCB/92-073-X FILALMENT BOARD POWER SUPPLY 3F LAMBDA
PCB/92-074-X ROTATING ANODE BOARD
PCB/92-082-X MOTHERBOARD
PCB/92-083-X CONTROL PANEL (PUSHBUTTONS) BOARD
PCB/94-117-X LED ON/OFF BOARD
PCB/94-127-0A EXPOSURE METER BOARD 1
PCB/94-127-0B EXPOSURE METER BOARD 2
PCB/94-133-0 C-ARM AUXILIARY BUTTONS
PCB/96-153-X MIRROR DRIVER BOARD
-TRF/1PL HIGH SPEED STARTER
PHTM9000 AEC DETECTOR (3 FIELD)
Lamps LEDs components
DI/BYV96E REC VEL 1A5 1000V DIODE(10pcs set)
LD/QD5GL YELLOW SQUARE LED 5mm
LD/QD5VR GREEN SQUARE LED 5mm
LD/TN5BL ROUND BLUE LED 5mm
LM/AL24V-100W ALOGEN LAMP 24V 100W
LMPL/AL24-100 ALOGEN LAMP 24V 100W SUPPORT
PD/SKB30-12 SINGLE PHASE BRIDGE 30 A –1200 V
PT/5K0-1/2W5% POTENTIOMETER 5K0 10 TURNS 1/2W
RLST/2450 D2450 SOLID STATE RELAY
RSFL82RO-50W SRH 82 Ω 50 W RESISTOR
RSFL1KO-50W SRH 1K Ω 50 W RESISTOR
Fuses
FS/B1000F20 INVERTER FUSE (5 pcs set)
FSADV0.5/250 SLOW BLOW FUSE D05L20 0.5 A 250V (10pcs set)
FSADV-4.0/250 FUSE D06L32 4.0A 250V (10pcs set)
FSAVD-0.1/250 SLOW BLOW FUSE D06L32 0.1A 250V (10pcs set)
FSAVD-0.5/250 SLOW BLOW FUSE D06L32 0.5A 250V (10pcs set)
FSAVD10.0/250 SLOW BLOW FUSE D06L32 10.0A 250V (10pcs set)
FSDFSM-10/500 FUSE D10 L38 10.0A 500V (10pcs set)
FSDFSM-20/500 FUSE D10 L38 20.0A 500V (10pcs set)
FST/T.04A/250 SLOW BLOW FUSE D05L20 40mA 250V (10pcs set)
FST/T1.0/250 SLOW BLOW FUSE D05L20 1.0A 250V (10pcs set)
FST/T-1.6/250 SLOW BLOW FUSE D05L20 1.6A 250V (10pcs set)
FST/T2.0/250 SLOW BLOW FUSE D05L20 2.0A 250V (10pcs set)
FST/T-2.5/250 SLOW BLOW FUSE D05 L20 2.5A 250V (10pz)
FST/T-3.15A SLOW BLOW FUSE D05L20 3.15A 250V (10pcs set)
FST/T3.15A/250 SLOW BLOW FUSE D06L32 3.15A 250V (10pcs set)
FST/-T8A/250 SLOW BLOW FUSE D06L32 8A 250V (10pcs set)
KT/FUSE-MAMM FUSES KIT (for one unit)
Motors
AT/37000002 MIRROR LINEAR ATTUATOR
ATT/33X100C24A COMPRESSION PADDLE SUPPORT ATTUATOR (BYM MODEL)
GENERAL MEDICAL ITALIA CHAPTER 1 Page 232 of 236
LAMBDA TECHNICAL MANUAL
CODE DESCRIPTION
EM/40X30X150N DELTI 24V BRAKE
EM/D-45S43--F EM F8M3W CLUTCH/COMPRESSION TROLLEY
MR—2338S012S CONTROL PANEL MOTOR
MR61.50.465 COMPRESSION MOTOR GEAR BOX BUHLER 24VCC50:1
MT/HP-0,33-4P C-ARM MOTOR
MTPP/7220R040 5VDL-15V-10R LINEAR ATTUATOR (POTTER)
Transformers
TF/205-T SERVICE TRANSFORMER
TF/211-TT POWER TRANSFORMER
TF/212-TT POWER TRANSFORMER
Pushbuttons
KT/-RX-BUTTON RX PUSHBUTTON WITH PROTECTION + CONNECTOR
KTFLA0205 EXTERNAL PEDALS PAIR
PL/1.5108451 1 RAFI 1.15108.451 ON/OFF PUSH BUTTON
PL-080BF11 EMERGENCY MUSHROOM PUSHBUTTON CONTACT LOCK
PL-080-RER PULSANTE A FUNGO ROSSO
Microswitches Relay Contactor
MW/AH34669 MICROSWITCH LEVETTA ROTELLA (POTTER)
MW/AVL34253 WEEL LEVER MICROSWIT (POTTER 24X30)
MW/AVL34553J WEEL LEVER MICROSWITCH (POTTER)
MW/C35ZP WEEL LEVER MICROSWITCH (C-ARM)
PL/83505.IW2 MICROSWITCH CROUZET (COLLIMATOR)
RLCS/MZP00245 RELAY PCB2SC 12VCC MZPA 0024505
TL/3RT1034-1A CONTACTOR 220 VAC 5NO+2NC
Lexan Plates
MEEQ3142-XX MAGNIFICATION DEVICE PLATE
MEEQ3214-XX PLATE FOR 12X8 COMPRESSOR PADDLE
MEEQ3215-XX PLATE FOR 10X22 COMPRESSOR PADDLE
MEEQ3216-XX PLATE FOR 12X22 COMPRESSOR PADDLE
MEEQ3217-XX PLATE FOR 16X22 COMPRESSOR PADDLE
MEEQ3613-XX PLATE FOR 24X30 COMPRESSOR PADDLE
MEEQ3646-XX PLATE FOR 16X22 FOR 2D BIOPSY COMPRESSOR PADDLE
MEEQ3692-XX PLATE FOR 10X8 COMPRESSOR PADDLE
Carbon Fiber covers
MEEQ2840-10 CARBON FIBER COVER FOR PMX2500
MEEQ3806-00 CARBON FIBER COVER FOR PMX3500
Grids
MEEQ1896-03 PMX2500 BUCKY GRID
MEEQ3815-00 PMX3500 BUCKY GRID
X-ray Tube
-TB/M147 VARIAN M147
-TB/P40W-100G PANTIX
Miscellaneous
CN/8140-1 POTTER FEMALE CONNECTOR
CN/8140-2 POTTER MALE CONNECTOR
CN/PAN3PFEM PANEL FEMALE CONNECTOR for X-RAY Push Button
CN/VOL3POLIM MALE CONNECTOR for X-RAY Push Button
CNTV10U-K45 10.0µF 450V CAPACITOR
CNTV14U-K45 14.0 µF 450 V CAPACITOR
CNTV16U-K45 16.0 µF 450 V CAPACITOR
CNVT2M2-K35 2200 µF 350 V CAPACITOR
ET108 ON/OFF LABEL (LAMBDA E/BYM)
ETLX0106-02 KEYBOARD LABEL
ETLX0147-XX AUXILIARY PUSHBUTTONS C-ARM UP/DOWN LABEL
Via Basecchia 2a
83024-Monteforte Irpino (AV)
Tel 0825683921
Cell. 0825683308
www.gmitalia.eu
Manufactured upon OEM agreement for GMI srl, By Metaltronica via della Pisana 431- 00163 Roma