PHYSIOGARD TM / CS 910

TM 910 CS 910 Clinical Central Station

The Universal Monitoring System

T h e A r t o f D i a g n o s t i c s
PHYSIOGARD TM 910

PHYSIOGARD TM 910:

Universal Monitor
• Multiple vital signs
• Up to 5 on-screen traces

Compact in design the PHYSIOGARD TM 910


monitor combines the experience and
knowledge of SCHILLER as a major player in
the field of patient monitoring. The universal
concept allows use in many applications,
moving with the patient from surgery
through recovery to intensive care, internal
hospital or emergency transport. Providing
monitoring for all patients: adult, child and
neonate.

Models in accordance with configurations:

ECG Resp. T1, T2*) P1, P2*) FiO2*) SPO2 NIBP EtCO2 REC Pacer

TM 910 • • • • • • •
TM 910 S • • • • • • • •
TM 910 N • • • • • • • •
TM 910 C • • • • • • • •
TM 910 S/N • • • • • • • • •
TM 910 S/C • • • • • • • • •
TM 910 C/N • • • • • • • • •
TM 910 S/N/C • • • • • • • • • •

S=SPO2 • Standard *) Optional Sensor • Option


N=NIBP
C=EtCO2

Flexible Monitoring! Mains independent use


In addition to ECG and Respiration, you can differential, combined with a powerful back- No matter what application you use, the
choose to monitor SpO2, NIBP, EtCO2, two light, the LCD screen provides clear, bright PHYSIOGARD TM 910 removable batteries
temperatures, two invasive pressures and comprehensive patient data. Even in unfa- offer a minimum of 4 hours mains inde-
FiO2. The internal printer provides waveforms, vourable lighting conditions, the sharp bright pendent use. With the optional vehicle input
full annotation and 16 hour trends. The display can be viewed easily from a distance. and mains power supply connector, the
unique SCHILLER pacemaker module utilises monitor can be used while batteries charge
one of the battery compartments to provide Easy handling automatically.
yet another application, without expansion Parallel use of the rotary button and the soft-
in size. keys guides you quickly through clear and
logical menus, saving time in configuration.
Powerful LCD screen The removable handle allows fixation onto a
Thanks to an exclusive system of color bed rail during transport.
CS 910 Clinical Central Station
Reviewing up to 72 hours ECG of
every single patient

Key technical features:

• Ready for up to 16 monitors


• Real-time display of 16 beds
(curves, values)
• Sample frequency: 75 Hz
• Quick overview of all patients connected
to the Central Station
• Direct control of the acoustic alarm and
brightness of each monitor
• Instant viewing of any monitor with
complete bed status and the five color
curves of each monitor
• User-definable list of special events on Multi patient view:
Basic monitoring information, together with alarm status, can
the Central Station, automatically be viewed simultaneously on one screen from all beds
assigned to any monitor
• On screen indication of all monitors with
alarms (visual and audible)
• Patient database available at all times.
• Patient discharge, admission or bed
transfer on a single click
• Date and time display

Real Time Central Monitoring


SCHILLER's CS 910 is a real time central management including printing functions
station for control of up to 16 bedside and storage capabilities. The CS 910 is
monitors. The day and night function delivered with Ethernet interface, is easy
allows the user to individually set the to install and hardware compatible with
acoustic alarm for every patient. The existing hospital networks.
SCHILLER CS 910 allows flexibility in data
Reviewing trends and events of all
connected monitors

Central Station CS 910


Real Time
Central Monitoring

Single patient view:


Whilst showing the complete monitoring information
from a selected bed – the basic information and alarm
status of all other monitors is available on screen

Bed-to-bed communication allows real time data transfer from monitor to monitor,
as well as reviewing trends and events of all connected monitors.
Technical Data for TM / CS 910:
PHYSIOGARD TM 910: PHYSIOGARD CS 910:

Monitor: IBP: Resolution of: 7,5 sec.


Screen: Color LCD-screen, 211 x 158 mm Input CF type using standard sensors Display of:
Resolution: 640 x 480 pixels Band width: 0 to 15 Hz (withour sensor) - Patient`s name
Number of channels: 3 to 5 waveforms Measuting range: -5 to 30 / -10 to 60 / - 25 to 150 / - 50 to - Bed identification number
Parameters displayed: 2 to 10 parameters 300 mm Hg - 24-hour trends or 4 curves
Trends displayed: 2 to 10 parameters Limit for zero setting: +/- 80 mm Hg - 72-hours of real time ECG
Menus: Configuration and operation Measuring error for inputs: +/- 1 mm Hg - Time-base
- Alarm history
Controls: Temp: Controls:
Permanent keys for alarm management, freeze, recording Input CF type using YSI 400 sensors - Scrolling
menu management with softkeys or navigation button Test: 38,8° C +/- 0,1° C - Zoom display of trends with a resolution of
Measuring range: 20 to 45° C 7,5 Sec./pixel
Set Up: - Shrink display of trends down to 24-hours
User settings: Configuration, user recognition, navigation - ECG strip display when clicking on an alarm event
command FiO2: Printout:
Language: D, GB, F, I or NL; date, time and color Measurement using oxide-reduction method - Of any displayed screen of the Central Station
combinations Measuring range: Setting 0 to 21 % or 0 to 100 % - Of any selected monitor
Accuracy: +/- 2 % from measurement system - Of the stored ECG and trends
ECG: Time constant: < 10 s with 90 % O2 The Central Station is available in different languages.
Input: CF type with 4-wire or 10-wire patient cable Life span: 20 months with 50 % O2
Leads: I, II, III, aVR, aVL, aVF, or 12 leads
Frequency range: 0,05 to 30 Hz on screen Technical data are subject to change without notice.
Transthoracic Pacemaker:
Input impedance: >10 µΩ External pacing module Type EDOS
Heart Rate: 0 to 350 bpm Modes: Fixed, Demand or Overdrive
Signal return: 200 ms after saturation Stimulus: 20 ms, 0 to 150 mA
Pacemaker detection: Asystoly alarm Frequency: 40 to 160 P/min. and 120 to 480 P/min.
ST measurement
Arrhythmia detection Recorder:
Thermal array with 2 traces
Alarms: Resolution: 8 dots/mm
System alarms: Battery charge, recorder Recorded data: Date and time, measurements, waveforms,
Physiological alarms: All parameters events and trends
Technical alarms: All parameters Mode: Automatic or manual
Recording duration: 5, 10, 20 s or continuous
Limit values: Recorder history: 0 to 5 s
Upper and lower limits: For all parameters Paper: Paper roll, 50 mm x 30 m
Differentiation: Alarm and warning
Limit values: Back up Outputs:
RS-485 for SCHILLER monitor net
Trends: RS-232 and analogical output for VM 900 C
16 h – all parameters – zoom to 1 h with 7,5 s of resolution Nurse call, synchro QRS
8 min. – HR and Resp. – zoom to 30 sec. with 125 ms of
resolution Power supply:
Power line: 220 to 240 V AC, 50/60 Hz, 30 VA
Respiration: Emergency vehicle: 11 to 16 V DC, 1 A
Measurement via ECG electrodes (impedance) Battery: Removable, 1 to 2, NiCd, 1,9 Ah
Frequency range: 0,1 to 3 Hz Battery Life: 2 to 4 h
Resp. Rate: 0 to 150 Resp/min Battery recharge: Automatic, built-in TM 910 or seperate
Signal return: 3 s after saturation charger

SpO2: Physical environment:


Measurement using spectrophotometric method Extreme temperatures: -20 to + 60° C for transport/storage,
Measuring range: 50 to 100 % 0 to + 50° C for operation
Accuracy: 70 to 89 % with +/- 2 %, 50 to 69 % with Relative humidity: 10 to 90% non condensing
+/- 3 %, 90 to 100 % with +/- 1,5 % Atmospheric pressure: 500 to 1060 hPa
Plethysmograph: On screen
Dimensions / weight:
NIBP: Dimensions: 220 x 306 x 185 mm (w x d x h)
Measurement using oscilliometric method Weight: 5,2 kg
Measures: SYS, MAP, DIA
Measuring range: 12 to 245 mm Hg for adults and 12 to
155 mm Hg for infants
Measuring duration: < 30 s
64 NIBP measurements saved

EtCO2:
Measurement with sidestream attachment and IR
spectrophotometry Y
Flow: 50ml/min.
Measuring range: 0 to 76 mm Hg with 0 to 100 Resp/min.
Capnography with 12,5 mm /s on screen
Monitoring Brosch_engl_ND 02.10.2002 15:48 Uhr Seite 2
REICHL UND PARTNER

Answer fax
Asia
Yes, I am interested in the innovative products from SCHILLER. SCHILLER Asia-Pacific
Please send me more information about: 10 C, Jalan SS 3/33
Taman Universiti
47300 Petaling Jaya
Resting ECG: Monitoring: Selangor, Malaysia
Phone: +603 78 77 5336
❏ CARDIOVIT AT-1 ❏ ARGUS PRO System Fax: +603 78 77 5744
❏ CARDIOVIT AT-1 smartprint ❏ PHYSIOGARD TM/CS 910 e-mail: schiller@top.net.my

❏ CARDIOVIT AT-2 ❏ Pulse-Oximeter OX-2 Austria


SCHILLER Handelsgesellschaft mbH
❏ CARDIOVIT AT-2 plus ❏ Magnetic Resonance Monitoring (MRM) Kampmüllerweg 24
A-4040 Linz
❏ ARGUS LCM Monitor Phone: +43 732 70 99-0
Fax: +43 732 757 000
Exercise ECG (Multitasking Systems): e-mail: sales@schiller.at
❏ CardioLaptop® AT-110 Defibrillation/Monitoring: C.I.S.
ZAO SCHILLER RU
❏ CardioLaptop® PT-160 ❏ FRED® (First Responder External Defibrillator) Pokrovsky Blvd., House 3
❏ CARDIOVIT AT-102 ❏ FRED® easy Block 1, Office 2202
P.O. Box 27
❏ CARDIOVIT CS-200 ❏ DEFIGARD (1002, 3002 IH, 6002) 109028 Moscow, Russia
Phone: +70 95 956 29 10
❏ CARDIOVIT AT-104 PC Fax: +70 95 956 18 63
e-mail: service@schiller.ru
❏ CARDIOVIT AT/SP-10 ❏ Holter ECG MT-100/200
France
❏ Ergometry ❏ Telemedicine: MT-120/220 and AT-4 TELE SCHILLER Médical S.A.
4, rue Louis Pasteur
❏ SEMA-200 Data Management System F-67162 Wissembourg/Cedex
Spirometry: ❏ BR-102 Blood Pressure Recorder Phone: +33 3 88 63 36 00
Fax: +33 3 88 94 12 82
❏ SPIROVIT SP-1 ❏ MS-3 Pocket ECG e-mail: info@schiller.fr

❏ SPIROVIT SP-2 ❏ VAC-100 Vacuum Electrode System Germany


SCHILLER Medizintechnik GmbH
❏ CS-200 Ergo-Spiro ❏ ECG & Spiro Accessories Prospect Rudolf-Diesel-Strasse 14
D-85521 Ottobrunn
❏ PC Spirometry Phone: +49 89 62 99 81-0
Fax: +49 89 609 50 90
Name: e-mail: info@schillermed.de

India
Hospital: SCHILLER Healthcare India Pvt. Ltd.
D.C. Silk Mills Compound, 'A' Wing,
Position: 1st floor
5, Chunawala Estate, Kondivitta Lane
Andheri-Kurla Road, Andheri (E)
Address: Mumbai - 400 059, India
Phone: +91 22 826 35 20
Fax: +91 22 826 35 25
e-mail: schiller@vsnl.com
Phone (daytime): Spain
SCHILLER ESPAÑA, S.A.
Fax: Avenida Baviera, 12 - 14 Posterior
E-28028 Madrid
Phone: +34 91 713 01 76
e-mail: Fax: +34 91 355 79 33
e-mail: schiller@schiller-es.com

Switzerland
SCHILLER-Reomed AG
Riedstrasse 14
CH-8953 Dietikon
Phone: +41 1 744 30 00
Fax: +41 1 740 37 10
e-mail: sales@schiller-reomed.ch

Turkey
SCHILLER TÜRKIYE
10/02

Halil Rifat Pasa Mah.


Darülaceze Cadd.
Perpa Ticaret Merkezi
B/Blok No: 69 K: 2
Okmeydani-Sisli – Istanbul
Phone: +90 212 210 8681 (pbx)
Fax: +90 212 210 8684
e-mail: sales@schiller-turkiye.com
Part No.: 2.500366

USA
SCHILLER America Inc.
11300 NW 41st Street
Miami, Florida 33178
Phone: +1 786 845 06 20
Fax: +1 786 845 06 02
e-mail: sales@schilleramerica.com

Your Specialist:

T h e A r t o f D i a g n o s t i c s
SCHILLER Médical S.A., 4, rue Louis Pasteur, F-67162 Wissembourg/Cedex, Phone: +33 3 88 63 36 00, Fax: +33 3 88 94 12 82, e-mail: info@schiller.fr
Headquarters: SCHILLER AG, Altgasse 68, CH-6341 Baar, Switzerland, Phone: +41 41 766 42 42, Fax: +41 41 761 08 80, e-mail: sales@schiller.ch

You might also like