Professional Documents
Culture Documents
Ivent101 Operator'S Manual
Ivent101 Operator'S Manual
Operator’s Manual
Manufacturer’s Address
VersaMed Medical Systems, Hasharon Industrial Park, Ornat Building, Kadima Israel.
P.O.B 5011, Kadima 60920, Israel
Tel: (972) 9 891 54000
Fax: (972) 9 899 7776
EC REP
Obelis S.A.
Av.de Tervuren 34, bte 44
B-1040 Brussels
Belgium
Tel: +32-2-732-59.54
Fax: +32-2-732-60.03
Table of Contents
Chapter 1: Introduction 1
1.1. Overview 1
1.2. How to use this manual 2
1.3. Cautions and Warnings 4
1.4. Intended Use 7
1.5. Contraindications 7
1.6. The iVent101 Ventilator 8
1.7. iVent101 Indicator Lights 11
1.8. Symbols and Labels 11
1.8.1. Front Panel Labeling 12
1.8.2. Back Panel Labeling – Connectors 13
1.8.3. Filter Side 14
1.8.4. Nameplate 15
1.9. Ventilator Specifications 16
1.9.1. Size and Weight 16
1.9.2. Environmental Specifications 16
1.9.3. Power Supply 17
1.9.4. Ventilation Modes 17
1.9.5. O2 Supply 18
1.9.6. Inspiratory and Expiratory Resistance 18
1.9.7. Ventilation Setting and Measured Value Parameters 18
1.9.8. Default Settings 20
1.9.9. Additional Displayed Parameters 20
1.9.10. Displayed Waveforms 21
1.9.11. iVent101 Ventilator Alarms 21
1.9.12. User Adjustable Alarms 21
1.9.13. Non-Adjustable Alarms 21
1.9.14. Accessories 22
1.9.15. Software Open Source Information 23
1.10. Standards and Safety Requirements 23
Chapter 2: Setting Up 25
2.1. Understanding The Ventilator Power Sources 25
2.1.1. Integrated Battery 27
2.2. Patient Circuit 27
2.2.1. Circuit Accessories 30
2.3. Attaching Air Inlet Filter 31
2.4. Choosing a Location for the Ventilator 32
2.5. Connecting the Ventilator to an Oxygen Source 33
2.6. Connecting the Ventilator to an AC Power Source 35
2.7. Powering up the Ventilator 36
2.8. Shutting Down the Ventilator 38
Chapter 5: Alarms 73
5.1. Alarm Priorities 73
5.2. Setting the Alarm Parameters 74
5.2.1. Setting the Alarm Volume 78
5.3. The Alarm Display Screen 78
5.4. Alarm Definition and Troubleshooting 81
5.5. Alarm Tests 91
5.6. The Log Book 92
Chapter 6: Maintenance 95
6.1. Cleaning and Maintenance Procedures 95
6.1.1. Cleaning Procedures 95
6.1.2. Preventive Maintenance 96
6.1.3. Preventive Maintenance Service Parts 97
6.1.4. Cleaning the ventilator 98
6.1.5. Exhalation Valve Maintenance 98
6.1.6. Integrated Battery Maintenance 102
6.2. Ventilation Circuit Test 104
6.3. Service Screens 106
Index 173
iVent101Operator's Manual
Chapter 1: Introduction
This manual is intended for the operator and qualified personnel.
This manual contains information about the proper setup, function and
maintenance of your iVent101 ventilator.
It is important that you read and understand all of the information presented in
this manual.
1.1. Overview
The iVent101 is a microprocessor-controlled ventilator, intended for use in home,
institutional and portable environment. It is compact, portable and feature-rich,
supporting both invasive and non- invasive ventilation.
The iVent101 ventilator is easy to use and enables rapid control and continuous
real-time monitoring of the patient's ventilation. The ventilation parameters and
alarm settings are fully adjustable and may be saved for future use. With its
bright, well-organized and easy-to-read touch-screen, the iVent101 ventilator
displays pressure and flow waveform data in a full array of graphical and
numerical modes.
The iVent101 ventilator supports the following ventilation modes:
A/C volume control
A/C pressure control
A/C PRVC (Pressure regulated volume control)
SIMV volume control
SIMV pressure control
SIMV PRVC (Pressure regulated volume control)
Adaptive Bi-Level:
CPAP/PSV
1
Introduction
2
iVent101Operator's Manual
Chapter 5: Alarms- Explains the iVent101 ventilator alarms and the
appropriate response and how to set the alarm parameters.
Refer to page 73.
Chapter 6: Maintenance: Explains how to care for and maintain the
iVent101 ventilator as well as the necessary test procedures. Refer to page
95.
Chapter 7: Troubleshooting - Troubleshooting guide for common issues
you may encounter while operating the iVent101 ventilator. Refer to page
107.
Appendices: Glossary, operation, and breath delivery theory, explanation
of the ventilations modes and the Hold function.
Screens names and tapping areas appear in Bold.
3
Introduction
4
iVent101Operator's Manual
Do not connect the iVent101 ventilator to a high pressure oxygen
supply.
Close the oxygen valve on the flow meter prior to disconnecting the
oxygen from the iVent101 ventilator.
Do not cover the iVent101 ventilator while it is in use. Make sure that
the unit is positioned so that its inlet ports are open to freely
circulating air.
5
Introduction
If your ventilator causes interference with another device:
If the ventilation equipment causes harmful interference to other devices,
you are encouraged to correct the interference by one or more of the
following measures:
Reorient the antenna on the affected device, such as, radio, television,
or cordless phone.
Move the device away from the ventilator.
Connect the equipment to an outlet which is on a different circuit than
the affected device.
Consult the dealer or an experienced radio/TV technician for help.
6
iVent101Operator's Manual
1.5. Contraindications
Not for use with anesthetic gases.
Do not use or store in the presence of strong electromagnetic fields such as an
MRI environment.
7
Introduction
1.6. The iVent101 Ventilator
8
iVent101Operator's Manual
1 Exhalation Exhaust
2 Exhalation Valve Connector
(one limb configuration only)
3 Proximal Pressure Connector
(one limb configuration only)
4 Inspiratory Port
5 LCD Touch Screen
6 Activity LED
7 Battery LED
8 Alarm LED
9 Expiratory Port (Two limb configuration only)
10 Air Inlet Filter
9
Introduction
1 O2 Inlet Connector
2 External DC Connector
3 On/ Off Button
4 Remote Alarm Connector
5 MMC Connector
6 SpO2 Connector
7 LAN Connector
10
iVent101Operator's Manual
8 USB Connector
9 Fuse
10 RS232 Connector
11 AC Connector
12 Power Pack
13 Power pack mounting knobs
Note: The MMC, the serial and Ethernet ports, and the SpO2 connectors are not
applicable in the current version.
11
Introduction
1.8.1. Front Panel Labeling
The labeling on the front panel of the iVent101 ventilator includes the following:
Upper Panel Symbols
Lower Panel Symbols
Alarm indicator
12
iVent101Operator's Manual
Lower Panel Symbols
The symbols below appear on the lower panel of the iVent101 ventilator.
Proximal Pressure
13
Introduction
FUNCTION ICON/SYMBOL LOCATION
O2 inlet Left side of the O2 inlet
connector
Remote Alarm Above the Remote
Alarm connector
SpO2 Above the SpO2
connector
Fuse Below the Fuse
connector
14
iVent101Operator's Manual
1.8.4. Nameplate
The nameplate is attached underneath the ventilator and contains the following
symbols and information:
FUNCTION ICON/WORDING
Ventilator Name
Manufacturer
”
Date of manufacturing
Barcode
S/N
Read this manual prior to use.
CE Mark
15
Introduction
1.9. Ventilator Specifications
16
iVent101Operator's Manual
17
Introduction
1.9.5. O2 Supply
Pressure: 100 cmH20 maximum
Flow: 20 L/min maximum
18
iVent101Operator's Manual
VENTILATION PARAMETER UNIT OF ACCURACY
SETTING RANGE MEASURE
Tidal Volume 40 – 2500 ml + 10% or 10 ml
Delivered whichever is greater
Tidal Volume Limit 40 – 2500 ml + 10% or 10 ml
whichever is greater
Peak Flow 5 – 120 or L/min + 2 lpm or 10% of
(mandatory) Adaptive Peak setting whichever is
Flow greater
Respiratory Rate 1 – 80 bpm + 1 bpm or 10% of
setting, whichever is
greater
Inspiratory Time 0.3 – 3.00 seconds 0.1sec or 10% of
setting, whichever is
greater
FiO2 (monitoring 21-100 % + 3% percentage
only) points up to 30% FiO2,
+ 7% percentage
points between 30 -
60% FiO2 (See note
below)
Rise Time 0-9 N/A N/A
ESens 5 –80 % + 1% for 5% -10%
+10 of the settings for
10% or greater.
Sensitivity 1-9 N/A N/A
Note: Above 60% FiO2 it is recommended to use an external oxygen analyzer for
the O2 monitoring.
Note: All flow, volume and pressure measurements in iVent-101 are expressed at
STPD conditions (Standard Temperature and Pressure Dry -0 °C. and 101,3 kPa
(760 mm mercury column)
The following table lists the displayed measured values including the range of the
ventilation parameters.
19
Introduction
MEASURED VALUE MEAUSRED RANGE UNIT OF
MEASURE
Exhaled Tidal Volume 0 – 9999 ml
Minute Volume Exhaled 0 – 99 L/min L/min
Respiratory Rate 1 – 99 bpm
Inspiratory Time 0.3 – 48 seconds
Peak Inspiratory Pressure 0 – 99 cmH2O
Peak Flow 0 – 240 L/min
Measurement Oxygen Mix (FiO2) 21 – 100 %
20
iVent101Operator's Manual
1.9.10. Displayed Waveforms
Real Time Pressure and Flow Waveforms can be displayed on screen.
21
Introduction
1.9.14. Accessories
The following accessories are provided with the ventilator:
DESCRIPTION VERSAMED GE
PART NUMBER PART NUMBERS
Europe Ac cable (1.8m length) 404J1802-01 M1161938
US AC Cable (1.8m length) 404J21111-01 M1184228
O2 Sensor 130B0002-03 M1161179
Red Cap (Sealing Cap) 32500004-A0 M1161821
Operator Manual 385OM001-01 M1184166
One Limb Kit 504HC051-01 M1184155
Two Limb disposable Patient 620B0101-A0 M1184227
Circuit (1)
2L Test Lung (Reservoir Bag) 910V0005-01 M1162061
Rp20 Resistor 910V0004-A0 M1161188
Exhalation Membranes (3 each) 323MHC01-01 M1161820
(1) The GE part number of this item refers to kit containing 10 units.
22
iVent101Operator's Manual
1.9.15. Software Open Source Information
The Product is based on Software which is developed and owned by the Supplier
(the "Software"). The Software may include or use certain open source code
software and materials which are subject to certain license terms which are
located at websites listed in Appendix G: Software Open Sources hereto (on page
155). The Open Source Licenses are displayed in Appendix G: hereto. By
accepting the terms and conditions of this Agreement, you also accept the terms
and condition of the Open Source Licenses. In the event of any inconsistencies or
conflicting provisions between the provisions of the Open Source Licenses and
the provisions of this Agreement, the provisions of the Open Source Licenses
shall prevail. Without derogating the generality of the foregoing, it is clarified that
the Open Source Licenses contains no warranties and/or liabilities from any kind
and under any circumstances, and the rights granted under this Agreement are
subject to Copyright Notice, to the extent that such notice appear in the Open
Source Licenses. Appendix G: hereto listed and contains certain open source code
software and materials which are not part of the Product, but supports the
Software operation as an external components to the Product.
The Kernel source can be viewed at
http://code.google.com/p/vm101hc/downloads/list.
23
Introduction
24
iVent101Operator's Manual
Chapter 2: Setting Up
This chapter explains how to set up the iVent101 ventilator before starting
ventilation and comprises the following sections:
Understanding The Ventilator Power Sources, section 2.1 below.
Patient Circuit, refer to section 2.2 on page 27.
Choosing a Location for the Ventilator, section 2.4 on page 32.
Connecting the Ventilator to an Oxygen Source, section 2.5 on page 33.
Connecting the Ventilator to an AC Power Source, section 2.6 on page
35.
Powering up the Ventilator, section 2.7 on page 36.
Shutting Down the Ventilator, section 2.8 on page 38.
If used in accordance with the product labeling contained in this manual, and your
doctor’s prescription, the ventilator is intended to provide reliable ventilation for
patients with respiratory insufficiencies or failures.
25
Setting Up
AC power is disconnected.
26
iVent101Operator's Manual
2.1.1. Integrated Battery
When the iVent101 ventilator detects an external power loss, it switches to the
integrated battery. When fully charged, the integrated battery provides up to 4
hours of power, depending on the ventilation conditions and settings.
Disposable one limb and two limb patient circuits are intended for a
single patient use only.
Reusable one limb and two limb patient circuits should be handled
according to their manufacturer’s specifications.
Note: The ventilator was tested with Hudson (Universal Single-Limb Ventilator
Circuit, Cat. No. 1696) and MPV TRUMA single limb patient circuit
27
Setting Up
(Dahlhausen Ref. No. 8065G). Using other patient circuit may affect ventilation
performance.
Inspect the patient circuit every day to:
Make sure there are no cracks or holes in the hose.
Be certain all the connections are secure and free from leaks.
28
iVent101Operator's Manual
2. Connect the second tubing to the Exhalation Inlet Port (refer to Figure 1.1
on page 8) by twisting and pushing it until it is firmly in place.
29
Setting Up
5. Connect the Exhalation valve tube to the exhalation valve connector on
30
iVent101Operator's Manual
2.3. Attaching Air Inlet Filter
Do not operate the iVent101 ventilator without an air inlet filter.
The iVent101 utilizes an air inlet filter that prevents the entry of any substance
greater than 5 microns.
Note: The air inlet filter must be replaced every 1 month of operating.
To replace the air inlet filter:
1. Turn the air inlet filter (shown in Figure 2.3 above) counterclockwise and
release it from the ventilator.
2. Insert a new air inlet filter to the air inlet and turn it clockwise until it is
securely in place.
31
Setting Up
2.4. Choosing a Location for the Ventilator
It is recommended that the iVent101 should be placed on a solid
surface so that the cooling air can flow through the vents in the
lower chassis without obstruction.
If the placement test reveals areas of the home or noisy activities over
which the alarm cannot be heard, do not leave the patient unattended
while you are in those areas or engaging in those activities.
32
iVent101Operator's Manual
2.5. Connecting the Ventilator to an Oxygen
Source
The iVent101 ventilator uses oxygen from a low pressure oxygen source such as
an oxygen concentrator or a flow meter. The low pressure oxygen supply is
connected to the iVent101 ventilator through an O2 connector.
The low pressure O2 should not exceed the following parameters:
Pressure: 100 cmH20 maximum
Flow: 20 L/min maximum
For the monitoring accuracy refer to section 1.9.7: Ventilation Setting and
Measured Value Parameters, on page 18.
Validation of the oxygen measurements was performed up to 100% FiO2. Above
60% FiO2 it is recommend to use an external oxygen analyzer for the FiO2
monitoring. It is advised not to exceed oxygen flows above 20 L/min.
Figure 2.4: Christmas tree connector Figure 2.5: Compression nut connector
33
Setting Up
Verify that Oxygen source is closed and that there is no pressure in the
tube prior to connecting the oxygen tubing to ventilator.
34
iVent101Operator's Manual
2. Press the Thumb Latch on the side of connector and pull the connector
(shown in Figure 2.6 above).
35
Setting Up
Battery knobs
Figure 2.7: The battery knobs can be used to wrap the power cord.
36
iVent101Operator's Manual
37
Setting Up
Figure 2.9: iVent101 ventilator main screen
38
iVent101Operator's Manual
39
Main Screens
NAME PURPOSE VIEW ICON EXAMPLE
Clinical View Used for advanced monitoring and Figure 3.3
operation of the ventilator, designed for
clinicians and technicians to operate
the ventilator. Changing of ventilation
settings and alarms and performing
service operations can be done only
through this view
For additional viewing options, refer to
section 4.7: Changing to Graph Tab on
page 72.
40
iVent101Operator's Manual
Figure 3.2: The iVent101 Ventilator in Basic View and in Standby mode
The Main screen in Basic View and Basic Night View contains the following
areas and buttons:
41
Main Screens
1. Current Measurements area: Displays six of the current ventilation
monitored parameters in table format. Displayed area only. Located at
the center of the screen.
2. Ventilation Mode Summary: Displays the current ventilation mode,
default patient type, and respiratory indicator. Displayed area only.
Located at the top of the screen.
3. Alarms area: Displays the most recent alarms and their severity. See
Chapter 5: Alarms on page 73. Tapped area. Located at the right side of
the screen.
4. Power Status area: Displays the power supply status. Refer to section
2.1: Understanding The Ventilator Power Sources on page 25. Displayed
area only. Located at the lower right side of the screen.
5. Screen Lock: Initiates and releases the screen lock. Refer to section 4.4:
Initiating and Releasing the Screen Lock on page 63. Tapped area.
Located at the bottom of the screen.
6. Start/ Pause button: Starts the ventilation process when the ventilator is
in Standby mode. Once the ventilation is started the button changes to a
Pause button. Refer to section 4.3: Starting and Stopping Ventilation, on
page 61. Located at the bottom of the screen.
7. Patient Tube Testing button: Opens the Patient Tube Test screen.
Refer to section: 6.2: Ventilation Circuit Test on page 104. Located at the
bottom of the screen.
8. Information Button: Provides general information about the ventilator.
Refer to section 4.3: on page 61.
9. Side Bar: Tapped area. Located at the left side of the screen, and
includes the following buttons:
View: Changes the current view. Refer to section 3.1: View
Options on page 39.
Manual Breath: Delivers one manual breath to the patient.
Refer to section 4.3.1: Manual Breath on page 62. Note that
this button is displayed only when the device is ventilating.
42
iVent101Operator's Manual
Menu: Opens a menu bar to enable changing option
settings. Refer to section 4.1: Menu Options on page 48.
43
Main Screens
measurements in graph format. To change the view, refer to section 4.7:
Changing to Graph Tab on page 72. Displayed area only. Located at the
upper central area of the screen.
2. Current Settings area: Displays the ventilation settings in table format
(as shown above). Displayed area only. Located at the lower central area
of the screen.
3. Ventilation Mode Summary: Displays the current ventilation mode,
default patient type and respiratory indicator. Displayed area only.
Located at the top of the screen.
4. Alarms area: Displays the most recent alarms and their severity. See
Chapter 5: Alarms on page.73. Tapped area. Located at the right side of
the screen.
5. Power Status area: Displays the power supply status. Refer to section
2.1: Understanding The Ventilator Power Sources on page 25. Located at
the lower right side of the screen.
6. Screen Lock: initiates and releases the screen lock. Refer to section 4.4:
Initiating and Releasing the Screen Lock on page 63. Tapped area.
Located at the bottom of the screen.
7. Start/ Pause button: Starts the ventilation process when the ventilator is
in Standby mode. Once the ventilation is started the button changes in a
Pause button. Refer to section 4.3: Starting and Stopping Ventilation, on
page 61. Located at the bottom of the screen.
8. Patient Tube Testing button: Opens the Patient Tube Test screen.
Refer to section 6.2: Ventilation Circuit Test on page 104. Located at the
bottom of the screen.
9. Information Button: Provides general information about the ventilator.
Refer to section 4.3, on page 61.
10. Monitor tabs: Change the monitor display. By default, the current
parameters are displayed in a table format, but graphical waveform can
also be displayed. Refer to section 4.7: Changing to Graph Tab on page
72. Located below the Current Settings area.
44
iVent101Operator's Manual
11. Side Bar: Tapped area. Located at the left side of the screen, and
includes the following buttons.
View: Changes the current view. Refer to section3.1: View
Options on page 39.
Manual Breath: Delivers one manual breath to the patient.
Refer to section 4.3.1: Manual Breath on page 62. Note that
this button is displayed only when the device is ventilating
Settings: Opens the Settings screen in order to change the
ventilation settings. Refer to section 4.2.2:Setting the
Ventilation Parameters on page 51.
Menu: Opens a menu bar to enable changing option
settings. Refer to section 4.1: Menu Options on page 48.
45
Main Screens
Notes:
46
iVent101Operator's Manual
47
erating the VentilatorOp
4.1. Menu Options
The Menu bar provides advanced options for operating the iVent101. The Menu
bar in Clinical View is displayed below:
Log Book Log Book Open the Logbook. Refer to section: 5.6 The
Log Book, on page 92.
48
iVent101Operator's Manual
BASIC VIEW MENU CLINICIAN VIEW DESCRIPTION
MENU
Pause Ventilator! Pause Ventilator! Stops the ventilation. Refer to section 4.3
Starting and Stopping Ventilation on page 61.
Close Close Closes the Menu bar.
49
Operating the Ventilator
2. Set the ventilation mode according to the patient's needs. Refer to Setting
Ventilation Modes on page 51.
3. Set the ventilation parameters and alarms according to the patient’s needs.
Refer to Modifying the Parameters on page 54.
Child 25 to 45 kg
Adult Over 45 kg
For a complete list of the default settings associated with each patient type, refer
to Appendix C: Default ventilation Settings on page 121.
To select a patient type for a new patient:
1. On the Main screen (Figure 3.2), tap Menu - New Patient. The New
Patient screen appears.
50
iVent101Operator's Manual
Once a new patient has been tapped the operation cannot be aborted, and
restoring previous settings can be done only manually.
The Ventilation Settings screen is displayed with the default settings.
3. Tap Accept to start ventilation with the default settings, or modify the
ventilation settings (refer to section 4.2.3: Modifying the Parameters on
page 54).
52
iVent101Operator's Manual
53
Operating the Ventilator
Note: When changing settings on a patient ventilated using adaptive peak flow,
the peak flow calculation is restarted, and ventilation waveform shape may
change, and will gradually return to the previous state.
The following table lists the available ventilation parameters from all ventilation
modes and their respective range of values:
54
iVent101Operator's Manual
PARAMETER NAME MEANING RANGE OF VALUES
PEEP PEEP setting 0 – 45 cmH2O
Press. Limit (Limit) Pressure Limit ventilation 5 – 60 cmH2O
setting
Vt. Set Set Tidal Volume setting Adult: 200-2500ml.
Child: 40-700ml.
Infant: 40-400ml
Peak Flow Peak Flow setting 5-120 L/min
IPAP High Pressure setting in 5-60 cmH20 (actual desired
Adaptive Bi-Level pressure)
EPAP Low Pressure setting in 0-45 cmH20
Adaptive Bi-Level
Insp. Time Inspiratory Time setting 0.3 – 3 seconds
ESens Termination of pressure 5-80 %
support breath after flow drop
detection
Rise Time Speed of the engine 1-9
acceleration during the
inspiratory phase.
PSV Pressure Support setting above 3 – 60 cmH2O
PEEP
Apnea Apnea alarm settings
Alarms Alarms settings Refer to section 5.2: Setting the
Alarm Parameters on page 74.
55
Operating the Ventilator
Slider
Calculator
Icon
56
iVent101Operator's Manual
57
Operating the Ventilator
Apnea Settings
Apnea is a backup mode that is automatically activated when no breath
(mandatory or spontaneous) has been delivered during the set apnea time interval.
The ventilation will continue with an apnea rate set by the clinician with the same
mandatory breath settings of the ventilation mode that is currently chosen.
If the ventilation mode is CPAP/PSV at the time when apnea occurs, the
ventilation is delivered according to the mode chosen by the clinician or with
default apnea settings. The default mode in CPAP/PSV for all patients is A/C
pressure control ventilation to allow appropriate ventilation regardless of the
patient size. Refer to section C.1: Default Apnea Settings on page 126.
You can change the default apnea settings for CPAP/PSV mode from the Settings
screen.
For all other ventilation modes you can set the ventilation rate delivered during
apnea.
58
iVent101Operator's Manual
59
Operating the Ventilator
60
iVent101Operator's Manual
Tapping Discard all in the Settings screen will discard the changes in the
Apnea settings even if you tap Accept in the Apnea Settings screen. You
have to tap Accept all in the Settings screen to apply the alarm settings
you changed
Do not ventilate a patient before verifying that the current settings are
correct.
To start ventilation while in Standby:
On the Main screen, tap the Start button.
To stop ventilation:
1. Perform one of the following option:
On the Main screen tap Pause.
On the side bar, tap Menu - Pause Ventilator.
A confirmation box is displayed.
61
Operating the Ventilator
62
iVent101Operator's Manual
4.4. Initiating and Releasing the Screen Lock
The iVent101 screen locks after it has not been used for a set period of time.
When locked, the iVent101 does not respond to tapping to prevent unintentional
change to the ventilation settings.
Note: When tapping several keys in rapid succession even with the screen lock
activated it might be possible to perform an action on the ventilator.
The initiation time can be set from the General Settings screen. The default time
is 3 minutes. To change the default time refer to section 4.6.3: Setting the on
page 68.
You can manually initiate the screen lock by tapping the Screen Lock icon
located at the bottom of the screen.
To release the screen:
1. Tap anywhere on the screen. The Screen Lock box is displayed.
63
Operating the Ventilator
4.5. The Information Screen
Tap the Information ( ) button, located on the lower left side of the
screen.
64
iVent101Operator's Manual
Last VVT – Lists the date and time when the VVT procedure was
performed. On the right side the status of the last VVT is displayed:
succeeded or failed.
Last OVT – The last time that OVT procedure was performed. On the
right side the status of the last OVT is displayed: succeeded or failed.
65
Operating the Ventilator
4.6.1. Setting the Time and Date
You can set the date and the time of the iVent101. Setting the time and date will
change the date display in the Main screen and in the logbook.
To set the Date & Time:
1. In the General Configuration screen tap Date & Time. The Set Date
and Time panel is displayed on the right side of the screen.
66
iVent101Operator's Manual
4. Tap Accept to save the settings and to close the screen.
5. Examine the time and date display at lower right side of the Main screen
to verify that the date and time have been changed and that they are
correct.
Make sure that the alarm can still be heard easily where caregivers are
located when you lower the alarm sound level. If the Alarm is not
heard increase the sound level.
To set the Sound levels:
1. In the General Configuration screen tap Sound Levels. On the right
panel two sliders are displayed: Button Volume and Alarm volume.
67
Operating the Ventilator
2. Tap the down or up arrows on each slider to change the sound level. The
slider displays the sound level accordingly, and a click is heard at a
volume corresponding to the current displayed sound level.
3. Tap Accept to save the settings and to close the screen.
68
iVent101Operator's Manual
2. Tap the down or up arrows to change the screen lock time. The slider
displays the time accordingly.
Select the Disable Auto Lock to enable or disable the screen lock (works as a
toggle). When the box is selected the automatic activation of the lock screen is
disabled.
If you enabled the screen lock the screen saver options are displayed.
Note: the screen-saver options are not available when the Auto Lock is disabled.
3. Select the Enable Screen Saver box to enable or disable the screen-saver (works as a
4. Tap the relevant screen saver option you want: Dim or Screen Off.
5. Tap Accept to save the settings and close the screen.
69
Operating the Ventilator
70
iVent101Operator's Manual
71
Operating the Ventilator
72
iVent101Operator's Manual
Chapter 5: Alarms
When the iVent101 ventilator detects a condition that requires your intervention,
an alarm sounds and a large message box appears that describes the alarm, often
including recommendation for possible solutions as well.
This chapter includes:
Alarm Priorities: Explanation of alarm severities along with the
alarm sound and pop-up screen associated with each severity.
Refer to section 5.1, below.
Setting the Alarm Parameters: Lists the alarms parameters that
can be set. Refer to section 5.2 on page 74.
The Alarm Display Screen: describes the Alarm screen and how
to suspend the audio alarm. Refer to section 5.3 on page 78.
Alarm Definition and Troubleshooting: List of alarm definitions
and the corrective actions that should be performed for each alarm.
Refer to section 5.3 on page 78.
The Log Book: Describes the iVent101 ventilator logbook. Refer
to section 5.6 on page 92.
73
Alarms
PRIORITY AUDIBLE VISUAL WHEN THE ALARM IS WHEN THE
SIGNAL SIGNAL ACCEPTED: ALARM IS
CLEARED:
Low -- Yellow The Alarm screen is closed and The alarm is
pop-up the alarm is displayed at the cleared from the
Screen. bottom of Alarm list on the right Main screen.
side of the Main screen.
74
iVent101Operator's Manual
75
Alarms
ALARM MODIFYING PROCEDURES
the left side will both display the new settings.
Tap the Low button to set the lower alarm limit. Do
one of the following:
Tap the Up or Down arrows on the slider until the
setting is correct.
Tap the Calculator icon. The calculator appears (Figure
4.5). Tap the setting you want and tap Enter. The
enlarged slider on the right side and the smaller one on
the left side will both display the new settings.
76
iVent101Operator's Manual
Tapping Discard all in the Settings screen will discard the changes in the
Alarms settings even if you tap Accept in the Alarm Settings screen. You
have to tap Accept all in the Settings screen to apply the alarm settings
you changed.
77
Alarms
78
iVent101Operator's Manual
79
Alarms
80
iVent101Operator's Manual
4. The Alarm suspend can be turned off when the alarm condition is
corrected. Tapping the Alarm suspend icon again will turn off the Alarm
suspend icon on the Main screen and restore the audible alarm
functionality.
81
Alarms
The following table displays the iVent101 ventilator alarms. The value in
parenthesis, either in the Activated or Cleared column, represents the delay
between the alarm condition and the ventilator response.
82
iVent101Operator's Manual
ALARM PRIORITY ACTIVATED WHEN PROBABLE CAUSE CORRECTIVE ACTION CLEARED WHEN
AC POWER Medium AC power failure If not intentionally Connect the ventilator to an The AC power is
DISCONNECT occurs, and the disconnected the alternative AC outlet. restored.
ventilator switches alarm might indicate a Verify the connection. (Immediately)
automatically to the failure in the AC
internal battery or power source, a Replace the electrical cord.
external DC. failure in the electrical Change to DC external power
(Immediately) cord, or internal source.
failure
APNEA High No breath is detected The patient clinical Verify that the patient is Spontaneous
during the apnea situation – when the adequately ventilated. triggers are
interval (as set by the ventilator is not Check the changes in the detected.
operator). detecting the patient patient symptoms and vital (After one minute,
(Immediately) efforts or when the signs. when three patient
mandatory breath spontaneous
rate is set too low. Verify appropriate sensitivity
trigger settings breaths are
detected).
Set a higher rate.
BATTERY High The integrated battery The ventilator Connect the ventilator to When an external
EMPTY nearly depleted – only operates with the external power source (AC or power source (AC
a few minutes of integrated battery DC), and recharge the or DC) is
ventilation left. only and the battery is battery. connected.
(Immediately) nearly depleted. (Immediately)
EMERGENCY High System switches to Hardware malfunction Assess the patient, Remove The ventilator
BACKUP Open Loop mode. or invalid sensors from the ventilator and returns to normal
(Immediately) calibration. provide alternative means of ventilation
ventilation if not sufficient. (immediately)
Follow the service notice
instruction (which follow the
Open Loop message)
83
Alarms
ALARM PRIORITY ACTIVATED WHEN PROBABLE CAUSE CORRECTIVE ACTION CLEARED WHEN
FACTORY Low Factory calibration is Calibration has not Have a qualified technician Factory calibration
CALIBRATION missing or invalid been performed, or perform factory calibration passed
REQUIRED internal malfunction. successfully.
HIGH FIO2 Medium Oxygen concentration Improper alarm Verify acceptable patient FiO2 is below the
exceeds the upper setting, a change in condition. alarm setting.
limit. the patient breathing Adjust the alarm setting. (4 breaths)
(One minute - see pattern, O2 sensor is
Verify the O2 concentration
note 2 below) out of calibration.
with an external analyzer.
Refer the ventilator to
qualified service technician
for calibration.
HIGH Medium The ventilator internal Block in the cooling Replace Air Inlet Filter. Internal
TEMPERATURE temperature has vents of the iVent101 Verify that the cooling vents temperature
exceeded 65ºC ventilator, high are clear and clean. decreases below
external heat, or 65ºC.
internal malfunction. Protect the iVent101
ventilator from external heat (Immediately)
source.
Refer the iVent101 ventilator
to qualified service personnel.
HIGH MINUTE Medium Minute Volume The patient clinical Check the patient condition. Volume is below
VOLUME exceeds the upper condition, alarm Increase the upper limit of the the alarm setting.
limit. setting too low, or the alarm setting. (2 breaths or 30
(One minute - see breath rate too high. seconds, whichever
Decrease the trigger
note 1 below) sensitivity. occurs first).
84
iVent101Operator's Manual
ALARM PRIORITY ACTIVATED WHEN PROBABLE CAUSE CORRECTIVE ACTION CLEARED WHEN
HIGH High 2 continuous events Patient agitation or Adjust the alarm setting. The pressure is
PRESSURE of High Pressure coughing, poor Assure no tubing occlusions below the alarm
synchronization setting.
between the ventilator More actions are subject to
clinical judgment. (One breath).
settings and patient
breathing, incorrect
alarm setting, circuit
or trachea tubes
occluded.
HIGH Medium Respiration exceeds Escalation of the Check the patient condition. Rate is lower the
RESPIRATORY the upper rate limit. patient breath rate, or Increase the alarm setting. alarm setting.
RATE (One minute – see if the iVent101 (Two breaths)
ventilator is auto- Decrease the trigger
note 1 below) sensitivity or the set breath
triggering.
rate.
INTEGRATED Medium Integrated battery Integrated battery Replace the integrated Integrated battery
BATTERY readings are invalid malfunction or battery. readings are valid
FAILURE (Immediately) disconnection, burned If the problem continues with (Immediately)
fuse. a new battery refer the
ventilator to qualified service
technician.
LOST LAST Low The last APNEA Internal malfunction Turn the ventilator off and Apnea settings are
APNEA settings cannot be then turn it on again. valid.
SETTINGS read If the problem persists modify (Immediately)
(Immediately) the settings again to match
the previously prescribed
settings.
85
Alarms
ALARM PRIORITY ACTIVATED WHEN PROBABLE CAUSE CORRECTIVE ACTION CLEARED WHEN
LOST LAST Medium The last Clinical Internal malfunction Turn the ventilator off and Clinical settings are
CLINICAL (ventilation) settings then turn it on again. valid.
SETTINGS cannot be read If the problem persists modify (Immediately)
(Immediately) the settings again to match
the previously prescribed
settings.
LOST LAST Low The last Technical Internal malfunction Turn the ventilator off and Technical
TECHNICAL configuration cannot then turn it on again. configuration
CONFIGURATION be read settings are valid.
If the problem persists refer
(Immediately) the ventilator to qualified (Immediately)
service personnel.
LOST LAST USER Low The last user Internal malfunction Turn the ventilator off and Technical
CONFIGURATION configuration cannot then turn it on again. configuration
be read If the problem persists match settings are valid.
(Immediately) the previously prevailing (Immediately)
settings.
LOW BATTERY Medium Integrated battery Ventilator operates Connect the ventilator to an When an external
running low, about 20 without external external power source (AC or power source (AC
minutes left. power source (AC or DC). or DC) is
(Immediately) DC). connected
(Immediately)
LOW FIO2 High Oxygen concentration Alarm setting too Check the oxygen supply. FiO2 is above the
is below the lower high, low or no Decrease the lower limit of alarm setting.
alarm limit. oxygen flow, out of the alarm setting.
calibration, high (Four breaths)
(One minute - see Refer the ventilator to
note 2 below) variability in the
patient breathing. qualified service technician
for calibration.
86
iVent101Operator's Manual
ALARM PRIORITY ACTIVATED WHEN PROBABLE CAUSE CORRECTIVE ACTION CLEARED WHEN
LOW MINUTE High Minute volume is A big leak, decrease Check the patient condition. Volume exceeds
VOLUME below the lower in the patient minute Increase ventilation. the alarm setting.
minute volume limit. volume or the lower (Four breaths)
limit of the alarm is Decrease the lower limit of
(One minute - see the alarm setting.
note 1 below) set too high, or might
indicates internal
malfunction of the
ventilator.
LOW Medium Airway pressure Patient clinical Verify that the patient circuit Pressure exceeds
PRESSURE never reaches the condition, incorrect is connected. the alarm setting.
lower pressure limit alarm setting, patient Correct the alarm setting. (One breath)
during inspiration. disconnect, high leak.
More actions are subject to .
(Three breaths or 30 clinical judgment.
seconds, whichever
occurs first)
LOW High Respiration is below Patient clinical Check the patient condition. Rate is higher than
RESPIRATORY the lower rate limit. condition, trigger Decrease the alarm setting. the lower alarm
RATE (One minute - see sensitivity set too low, setting.
set rate too low, Increase the trigger sensitivity
note 1 below)) or the breath rate. (Two breaths or
alarm setting too one minute,
high. whichever is last).
87
Alarms
ALARM PRIORITY ACTIVATED WHEN PROBABLE CAUSE CORRECTIVE ACTION CLEARED WHEN
O2 SENSOR Low O2 sensor is O2 sensor is Disable the O2 monitoring O2 sensor is
DISCONNECT disconnected. disconnected. and have the O2 sensor connected, or O2
replaced by a qualified monitoring is
service personnel. disabled
O2 SENSOR Low O2 sensor Aged O2 sensor or Disable the O2 monitoring O2 calibration
WORN OUT parameters are out of malfunctioning. and have the O2 sensor passed
scale. replaced by qualified service successfully, or O2
personnel. sensor is
disconnected, or
O2 monitoring is
disabled
OCCLUSION High 5 breaths in which Occlusion in the Check patient circuit and After 1 breath
occlusion is detected patient circuit or remove occlusions.
- see note 2 below) exhalation valve Check and clean the
Exhalation valve.
OVT Low Patient Tube Testing Patient circuit Check Patient circuit OVT passes
RECOMMENDED failed, or no data of connections, connections, perform Patient successfully
valid test. Exhalation valve not Tube Testing, check the
properly assembled, exhalation valve assembly
or Patient tubing test
has failed.
PATIENT High Loss of patient Patient or tubing Reconnect the tubing Reconnection of
DISCONNECT resistance or loss of disconnected or a Reconnect the patient. the patient is
exhaled volume. very large leak in the detected.
patient circuit, Correct the leak or revise
settings. (Immediately).
improper settings.
SENSORS Low Sensors calibration is Calibration has not Have a qualified technician Sensors calibration
CALIBRATION missing or invalid been performed, or perform factory calibration passed
REQUIRED internal malfunction successfully
88
iVent101Operator's Manual
ALARM PRIORITY ACTIVATED WHEN PROBABLE CAUSE CORRECTIVE ACTION CLEARED WHEN
SERVICE High A technical Power surge, multiple Provide alternative The system failure
NOTICE irregularity occurs sensor disconnection, ventilation. Turn the that caused the
that requires a needed calibration, or ventilator off and then turn it Service notice is
qualified technical other malfunctions. on again. fixed.
service. If the problem persists refer (Immediately)
(Immediately) the ventilator to qualified
service personnel.
SET PRESSURE Medium In Pressure control, Decreased patient Check the patient. The set pressure is
NOT Adaptive Bi-level, or effort or lung Checks for leaks. achieved for 2
DELIVERED CPAP/PSV mode – compliance, pressure consecutive
the monitored mode settings too high, or Reduce the pressure setting if breaths.
is 5 cmH2O or 25% may indicate internal applicable.
below the set malfunction in the
pressure (whichever ventilator.
is greater)
(8 consecutive
breaths - see note 2
below)
VOLUME LIMIT Medium In pressure control Increased patient Check the patient. The limit volume
REACHED mode - the set effort or lung Reduce the inspiratory has not been
volume limit has been compliance, or too pressure or pressure limit reached.
reached. low of a limit setting. setting. (one breath)
(8 breaths, or 20 Increase the volume limit.
seconds, whichever
occurs first - see note
2 below)
89
Alarms
ALARM PRIORITY ACTIVATED WHEN PROBABLE CAUSE CORRECTIVE ACTION CLEARED WHEN
VOLUME NOT Medium In Volume control Decreased patient Check the patient. The volume has
DELIVERED mode – the set effort or lung Check for occlusions. been reached for 2
volume has not compliance, or conclusive breaths.
delivered. volume setting to Reduce the volume setting if
high. applicable.
(8 breaths or 20
seconds, whichever
come first - see note
2 below)
VVT Low VVT failed, or no data Internal malfunction Provide alternative VVT passes
RECOMMENDED of valid VVT. ventilation. Turn the successfully
ventilator off and then turn it
on again.
Refer the ventilator to
qualified service personnel.
Notes:
1. Alarms conditions are checked after 30 seconds have elapsed from ventilation initiation or patient
reconnection after disconnect or ventilation settings change.
2. Alarms conditions are checked after 60 seconds have elapsed from ventilation initiation or patient
reconnection after disconnect or ventilation settings change.
3. Alarms conditions are checked after 10 seconds have elapsed from ventilation initiation or patient
reconnection after disconnect or ventilation settings change.
90
iVent101Operator Manual
Do not test the Patient Disconnect alarm while the patient is connected
to the iVent101 ventilator.
4. Patient Disconnect: Disconnect the patient circuit tubing from the
Inspiratory Outlet Port (refer to Figure 1.1 on page 8). Verify that the
Patient Disconnect alarm is activated.
Reconnect the patient circuit tubing to the Inspiratory Outlet, and verify
that the Patient Disconnect alarm stops automatically when ventilating a
test lung or simulated patient resumes.
91
Alarms
92
iVent101Operator Manual
The events on the Logbook are ordered chronologically. The date and time of
the event are written on the left side of the screen, and the event description on
the right.
93
Alarms
Notes:
94
iVent101Operator Manual
Chapter 6: Maintenance
This chapter covers the maintenance and test procedures you perform on the
iVent101 ventilator.
These procedures include:
Cleaning and maintenance procedures, refer to section 6.1 below.
Performing the Ventilation Circuit Test, refer to section 6.2: Ventilation
Circuit Test , page 104.
Note: When moving the iVent101 ventilator, make sure that you hold it with both
hands and not just by the handle.
95
Maintenance
PART PROCEDURE COMMENTS
O2 Sensors Replace every 1year or as Refer to the Service Manual
needed.
96
iVent101Operator Manual
FREQUENCY PART / PROCEDURE MAINTENANCE
Ventilation Verification Test VVT procedures, performed by
(VVT) an authorized service
technician.
Calibration and alarm tests Perform all calibration
procedures, and all alarm
tests, performed by authorized
service technician.
Exhalation valve Replace the exhalation valve
with a new one. Refer to
Annually section 6.1.4:Exhalation Valve
Maintenance, page 98.
Integrated battery (2) Replace the integrated battery.
Refer to section 6.1.6:
Integrated Battery
Maintenance, page 102.
Every 10,000 operating hours Pneumatic unit Replace the unit by an
authorized service technician.
97
Maintenance
In order to obtain the full service parts list, please refer to the service manual.
98
iVent101Operator Manual
99
Maintenance
Lever
handle
Lever
handle
100
iVent101Operator Manual
5. Release the latch from the cover and detach the cover and
the membrane (see Figure 6.4).
6. Clean the exhalation valve or put it in the sterilization
device, as instructed above.
7. Replace the membrane if needed.
The Exhalation membrane has to be replaced every 3 months.
8. Re-assemble the exhalation membrane and cover. Note
the exhalation membrane asterisk should be facing the
cover asterisk. You do not have to push the membrane
into its place, just to place it inside the cover and the
membrane will seat when the cover is pressed on.
9. Open the cover. Verify that the membrane sits in its
place.
Note: you have to verify that the membrane is in its place before continuing
the assembly.
10. Close the cover.
11. Close the latch and verify that the cover is closed.
101
Maintenance
12. Insert the exhalation valve back into the exhalation
compartment ensuring the pneumatic connectors are
properly engaged.
13. Rotate the locking lever and turn the lever back through
90 degrees in order to lock the exhalation valve in
position.
14. Place the exhalation valve compartment cover back in
position.
Note: When the Empty Battery alarm message is displayed, you must
immediately recharge the battery. If not, it will lead to battery failure and affect
the battery charging capability.
This procedure requires as much as 24 hours to complete.
To recharge the battery:
1. Assure that the AC power cord has been plugged into the ventilator and
the battery is fully charged. On the front panel of the iVent101 ventilator
(Figure 1.1), the Charging light will be either yellow, indicating that the
battery is charging or blue if already fully charged.
2. Assure that the battery has been charged for at least 10 hours.
102
iVent101Operator Manual
3. If the ventilator was shut down then turn it on.
4. Connect the iVent101 ventilator to a patient circuit attached to an Rp20
resistor and reservoir bag. Refer to section 2.2: Patient Circuit, page 27.
Note: You may connect the iVent101 ventilator to the patient at this time, if
needed.
The Fully Charged icon is displayed on the Main screen. If the Fully
Charged icon is not displayed, you need to replace the battery.
103
Maintenance
1. From the Main screen tap the (Ventilation Circuit Test) button.
Note: This button is also displayed on both the Basic View and Basic Night
View screens.
The Patient Tube Testing screen is displayed.
104
iVent101Operator Manual
105
Maintenance
5. Remove the plastic cap from the patient wye.
Do not ventilate the patient with a ventilator that failed the Ventilation
Circuit Test.
106
iVent101Operator Manual
Chapter 7: Troubleshooting
The iVent101 has a built-in safety and internal monitoring mechanism whose
purpose is to ensure the patient is safely ventilated, detect internal faults, and
initiate recovery when possible.
In the table below is a list of fault scenarios, their possible causes, and suggested
solutions.
Note: when an alarm is activated, first and foremost check the patient’s status and
well-being. Verify that patient is well and being ventilated properly. If not, take
action to resolve the situation.
107
Troubleshooting
CONDITION POSSIBLE CAUSE SUGGESTED SOLUTION
1. Verify that the patient
is being ventilated.
2. The SW should auto
recover the problem in
The ventilator is up to 20 seconds.
ventilating and the 3. If the problem persists
display is visible, but the provide alternative
Internal malfunction
touch-screen does not means of ventilation.
respond to the operator’s 4. Turn the ventilator off
tapping and then on again.
5. If the problem persists
contact qualified
service personnel.
The ventilator is Power up the ventilator.
turned off.
The ventilator is not
ventilating, the touch- Connect the ventilator to an
screen is black without AC power supply and power
any display, and the There is no power it up.
green LED is off. supply to the device
Let the integrated battery
charge.
108
iVent101Operator Manual
CONDITION POSSIBLE CAUSE SUGGESTED SOLUTION
1. Check the ventilator
The ventilator is not
connection to power
ventilating, the touch-
and provide power if
screen is black without
found missing.
any display, but the
2. If the splash screen is
ventilator is turned on,
visible within 10
and a constant beep can
seconds, it is very
be heard.
likely that the
ventilation will resume
automatically within 20
No AC power to 30 seconds.
connected, and the 3. If the problem persists
integrated battery has Provide alternative
just been depleted, or means of ventilation.
internal malfunction 4. The device should
recover completely
within 50 to 60 seconds
from the onset of the
condition. If not turn
the ventilator off and
on again.
5. If the problem persists
refer the ventilator to
qualified service
personnel.
109
Troubleshooting
CONDITION POSSIBLE CAUSE SUGGESTED SOLUTION
1. Tap the touch-screen
The ventilator is
ventilating, but the touch- several times.
2. The device should
screen is black without
recover completely
any display, an
within 50 seconds from
alternating beep sound
the onset of the
may be heard. Screen saver is on, or condition. If not turn
internal malfunction the ventilator off and
on again.
3. If the problem persists
refer the ventilator to
qualified service
personnel.
1. If there is no visible
The ventilator is not
alarm it is very likely
ventilating, but the
display is visible, and a that the ventilation will
resume automatically
constant beep sound can
within 2 to 5 seconds.
be heard. Accompanying
alarms may be displayed. 2. If the problem persists
provide alternative
means of ventilation.
3. The device should
Internal malfunction recover completely
within 50 to 60 seconds
from the onset of the
condition. If not turn
the ventilator off and
on again.
1. If the problem persists
refer the ventilator to
qualified service
personnel.
110
iVent101Operator Manual
CONDITION POSSIBLE CAUSE SUGGESTED SOLUTION
1. Disconnect the
The exhalation valve
exhalation valve, and
was re-assembled
re-assemble it
incorrectly after
High leak from the following the
cleaning it.
exhalation valve. instructions on section
6.1.5: Exhalation Valve
The exhalation valve Maintenance, on page
was bumped when the 98.
ventilator was moved. 2. If the problem persists
contact qualified
service personnel.
111
Troubleshooting
CONDITION POSSIBLE CAUSE SUGGESTED SOLUTION
1. Provide an alternative
Repeated instances of Patient conditions
clinical alarms, with no changed, obstructed means of ventilation,
and check if the
apparent cause. inlet, internal
condition is solved, if
malfunction
not refer to a qualified
clinician.
2. Check that the inlet is
not obstructed.
3. Replace the inlet filter
if needed. Refer to
section 2.3, on page 31.
4. If the problem persists
contact qualified
service personnel.
112
iVent101Operator Manual
Appendix A: Glossary
TERM DEFINITION
Adaptive Bi-Level Combination of a facemask and invasive mode of ventilation,
which utilizes two levels of pressure for each breath cycle, IPAP
during inspiration and EPAP or PEEP during exhalation.
TM
Adaptive Peak Flow Inspiratory Peak Flow Rate determined and delivered to meet the
target mandatory tidal volume, while maintaining an I:E ratio of
1:2 or reasonable inspiratory time.
Adaptive I-TimeTM Inspiratory time determined by the iVent101 ventilator to maintain
a more physiologic I:E ratio.
Airway Pressure Measured pressure, estimating pressure, at the proximal airway
of the breathing circuit. Measurement unit- cmH2O.
Alarm Combination of an audible buzzer and message pop-up
generated when the iVent101 ventilator detects an operating
condition that requires intervention.
Apnea Cessation of breathing.
Apnea Mode Ventilation mode that automatically starts when apnea is
detected. The apnea period, which is the number of seconds
from the last breath until apnea is detected, is configurable.
Assist Mode Ventilator operation in which a machine breath is delivered only
when an inspiratory effort by the patient is sensed by the
ventilator.
Assist/Control Mode Ventilator operation in which a machine breath is delivered when
an inspiratory effort by the patient is sensed by the ventilator
(Assist), or at set time intervals if the rate of occurrence of
inspiratory efforts falls below a set rate (Control).
Auto PEEP Positive End-Expiratory Pressure in the lung that is higher than
the pressure seen at the airway opening.
Breath Period The length of time from the start of a ventilator-initiated breath to
the start of the next breath.
Breath Termination End of the inhalation phase of the breath.
CMV Mode Controlled Machine Ventilation. A ventilation mode in which the
patient receives only a fixed number of ventilator breaths per
minute.
Compliance A measure of the stiffness of the lung and chest wall.
113
Glossary
TERM DEFINITION
CPAP Continuous Positive Airway Pressure. A mode of ventilator
operation in which the airway pressure remains constant above
ambient during spontaneous breaths.
TM
Easy Exhale Pressure relief at the very beginning of the exhalation phase to
enhance exhalation.
EPAP The absolute low pressure setting for Adaptive Bi-Level.
Analogous to PEEP.
ESens A percentage of the peak flow used as a termination criteria for
pressure support breaths – a breath is terminated once the flow
has dropped to the ESens level. Also known as Exhalation
sensitivity.
Exhalation Phase The part of the breath cycle which starts with the end of
inspiration and ends with the start of the next breath.
Exhaled Tidal Volume The exhaled volume of a single breath measured by the flow
sensor for all breath types.
Flow Trigger Method of initiating breath in response to the patient effort, by
measuring an increase in the inspiratory flow.
High Pressure Alarm Pressure at the entrance of the patient airway exceeds the
defined pressure alarm limit. The usual response is the
immediate termination of the breath, and an alarm sounds.
I:E Ratio Ratio of the inspiratory phase time to the expiration phase time.
Inspiratory Phase The phase of the breath in which the patient inhales, or
inspiratory flow is delivered into the lungs.
Inspiratory Pressure Pressure above the PEEP level during the inspiratory phase of
the breath.
Inspiratory Time Duration of the inspiratory inhalation phase.
IPAP The absolute inspiratory pressure setting for Adaptive Bi-Level.
Low Pressure Alarm Pressure at the entrance of the patient's airway is lower than the
defined pressure alarm limit. An alarm sounds.
Mandatory Breath Any breath that is controlled and terminated by the ventilator, in
order to achieve either a preset tidal volume or target pressure
for a preset time.
Manual Breath An operator initiated ventilator breath.
Maximum Inspiratory Time A predetermined time limit for inspiratory time for all breath types.
Maximum Pressure In PRVC, maximum allowed pressure.
iVent101Operator Manual
TERM DEFINITION
Mean Airway Pressure The average airway pressure throughout the breath cycle.
Minimum Exhalation Time A period of time from the beginning of the exhalation phase,
during which no breath type may be initiated.
Minimum Pressure In PRVC, minimum allowed pressure.
Patient Breathing Circuit The tubing, valve, and flow sensor, which provides the ventilatory
interface between the patient and ventilator.
Patient Breath Any ventilator delivered breath that is initiated and terminated by
the patient. Could be either Spontaneous or Pressure Support.
The breath is initiated by flow or pressure trigger, and terminated
by decrease in flow or increase in pressure. Available in SIMV,
Adaptive Bi-Level, and CPAP modes.
Patient Effort Any inspiratory effort initiated by the patient.
PCV Pressure Controlled Ventilation. A type of ventilation in which the
ventilator breaths are controlled by pressure, terminated by time,
and can be limited by volume.
Pressure Limit In volume control ventilation, the patient pressure is limited to
this pressure level. The tidal volume delivery may decrease
when the pressure limit is reached which can cause an alarm.
PRVC Pressure Regulated Volume Control Breath. A type of ventilation
in which the ventilator adjusts the pressure for each successive
breath in order to target a set tidal volume.
PEEP Positive End Expiratory Pressure. Positive pressure in the lung
during expiration.
PIP Peak Inspiratory Pressure. The maximum airway pressure that
occurs during inspiration.
Pressure Control Mode A mode of ventilator operation in which the ventilator attempts to
deliver a preset pressure waveform during inspiration. Sufficient
flow is provided to achieve and maintain a set pressure in the
patient circuit during the inspiratory phase of the breath.
Pressure Support Breath A support mode in which the patient effort is accompanied with
an elevated pressure augmenting tidal volume. The patient
determines the inspiratory time. Available in SIMV and
CPAP/PSV modes.
Pressure Trigger A method of initiating a breath in response to patient efforts by
measuring a decrease in airway pressure below the baseline.
Rise Time The rate of acceleration of pressure to reach the set pressure
level.
115
Glossary
TERM DEFINITION
RR/Vt The ratio of spontaneous breathing rate to the exhaled tidal
volume. Typically used as a measurement of breathing stress.
SIMV Synchronized Intermittent Mandatory Ventilation. A mode of
ventilation in which both mandatory and spontaneous breath
types are allowed. The mandatory breaths are synchronized with
the patient efforts.
Spontaneous Breath A patient initiated breath in which inspiratory time is determined
by the patient
Static Compliance A measurement of stiffness of the lung and chest wall. Typically
measured as the volume exhaled after the hold maneuver
divided by the change in the pressure from plateau pressure to
PEEP.
Tidal Volume (VT) The amount of volume inhaled or exhaled in one breath.
Measured in milliliters.
Tidal Volume Limit In pressure control ventilation, the maximum volume allowed to
be delivered to the patient.
Tidal Volume Set In volume control ventilation, the volume preset by the clinician
and delivered to the patient.
Time Constant A measurement of the time it takes to produce a percentage of
breath exchange either into or out from the lung. The product of
the resistance and the static compliance.
Total Breath Rate Total number of breaths per minute, including both patient and
mandatory breaths.
Total Minute Volume The volume delivered to the patient for all breath types over one
minute. Measured in liters per minute.
Trigger Defines the patient effort that starts a breath.
Ventilator Breath A ventilator delivered breath that is initiated, controlled, and
terminated by the ventilator. Available in Assist Control and SIMV
modes.
Volume Controlled Ventilation A type of ventilation in which the ventilator breaths are controlled
by flow and terminated by volume, as long as the airway
pressure is lower than the limit pressure.
iVent101Operator Manual
Appendix B: Warranty
VersaMed (“the Supplier”) guarantees to the Purchaser the material and
workmanship used in the manufacture of the iVent101 ventilator (“the Product”)
sold to the Purchaser, and will repair any defects which may develop within 12
months from the delivery date to the Purchaser, which are not due to fire,
dampness, willful or accidental damage, or improper use or care, or other causes
beyond the control of the Supplier. This 12 months warranty does not extend to
expendable items such as membranes, hoses, and filters which are warranted to be
free from defects only at the time of the original delivery.
The express warranties set forth above specifically exclude defects in the Products
that are (1) caused through no fault of Supplier during shipment to or from the
Purchaser, (2) caused by the use of or operation of the Products in any application
or environment other than that instructed, intended or recommended by the
Supplier, (3) caused by modifications or alterations made to the Products by the
Purchaser or any third party, (4) caused by unauthorized maintenance performed
on the Products by the Purchaser or any third party, (5) caused by failure of the
Purchaser to comply with any the return procedures, (6) are the result of the
Products being subjected to unusual physical and/or electrical stress.
Except for the above express limited warranties or conditions, Supplier makes and
Purchaser receives no warranties in connection with the Products, express and/or
implied and/or statutory, and the Purchaser specifically disclaims any implied
warranty or condition or merchantability or fitness for a particular purpose. In no
event shall the Supplier be liable for any damages, including loss of profits,
incidental or consequential damages, arising out of or in connection with the use,
or inability to use the Products.
117
Warranty
Subject to the following, the Purchaser shall send the Products with defects that
are covered by this warranty to the Supplier repair center.
The Purchaser shall request a written return authorization from the Supplier prior
to the return of each defective Product for repair or replacement by the Supplier.
Upon request, Supplier will provide the Purchaser with a Return Goods
Authorization (RGA) number to be prominently displayed on the shipping
container for the defective Products.
Once the Supplier authorizes the return of defective Products, the Purchaser shall
ship such Products to the repair facility, freight and insurance expenses prepaid in
its original container and without relieving the Purchaser from its responsibility
for the shipment. If such defective Products are received by the Supplier during
the applicable warranty period, the Supplier shall, at its sole option and expense,
repair or replace such Products, to make such repair or replacement, and shall
ship the repaired or replaced Products to the Purchaser.
In any case and subject to the terms hereof the Supplier will be responsible for the
repair of or replacement of such defective Products only.
Items returned without such approval shall be returned to the Purchaser at its
expense. The Purchaser undertakes that the Products be shipped by a certified
carrier experienced in handling sensitive freight. The Products returned by the
Purchaser for repair or replacement must include a report indicating the type of
failure.
Supplier reserves the right at any time to change the specifications or design of
the Products. In the event of any such change in specifications or design, Supplier
shall be under no obligation to make the same or similar change in the Products
previously produced or sold by the Supplier, unless the change is made to correct
a safety or operational deficiency, or is required by the US Food and
Administration.
iVent101Operator Manual
Neither this Warranty nor any of the rights of the Purchaser under this Warranty
may be assigned, transferred, or conveyed by the operation of law or otherwise,
without the prior written consent of the Supplier nor shall this warranty or any
rights of the Purchaser inure to the benefit of any trustee in bankruptcy, receiver,
creditor, trustee or successor of the Purchaser’s business or its property, whether
by operation of the law or otherwise, or to a purchaser or successor of the
business or any of the assets of the Purchaser, without the written consent of the
Supplier.
This Warranty shall apply only to Products purchased by end users directly from
VersaMed and provided they are paid for in full. This warranty shall not apply to
distributors or resellers (“Distributor”). VersaMed may provide individual
warranty protection to its authorized Distributors, according to the terms and
conditions agreed upon between VersaMed and each Distributor.
119
Warranty
121
Default ventilation Settings
Parameter MEANING INFANT CHILD ADULT FULL RECCOM- LOW HIGH
RANGE ENDED CAUTION CAUTION
8-25 KG 25-45 KG >45KG RANGE ZONE ZONE
VtLimit (ml) Volume 300 N/A N/A 40 - 400 50 - 200 None 300
Limit
(Pressure N/A 500 N/A 40–700 200 - 400 100 600
Control)
Pinsp Inspiratory
(cmH20) Pressure
(Pressure
15 15 15 5 - 60 15 - 30 10 40
Control)
Plim Pressure
(cmH20) Limit
(Volume
35 35 35 5 - 60 20 - 35 10 40
Control)
Tinsp (sec) Inspiratory
Time
0.6 0.8 1.2 0.3 - 3 0.5-1.5 0.4 2.0
Thold Inspiratory
Hold Time
(Volume
0 0 0 Yes No setting recommendations
Control)
Peep Set Peep
5 5 5 0 - 45 3 - 20 3 20
(cmH20)
PSV Pressure
(cmH20) Support
Ventilation
5 5 5 3 - 60 5 - 20 N/A 25
(SIMV)
122
iVent101Operator Manual
Parameter MEANING INFANT CHILD ADULT FULL RECCOM- LOW HIGH
RANGE ENDED CAUTION CAUTION
8-25 KG 25-45 KG >45KG RANGE ZONE ZONE
Trigger The 3 3 3 1–9 1-9 N/A N/A
sensitivity Triggers’
sensitivity
level
Peak Flow The Peak Adaptive N/A N/A 5 – 120 8 - 20 N/A 30
(L/min) flow
N/A Adaptive N/A 5 – 120 20 - 40 10 60
N/A N/A Adaptive 5 - 120 30 - 50 20 80
Peak Flow Adaptive
Adaptive State of
(true/ false) Peak Flow
Adaptive Adaptive Adaptive Yes N/A N/A N/A
Adaptive or
Not
Adaptive
Peak Flow The peak 12 N/A N/A 5 – 120 8-20 N/A 30
not flow when
adaptive is N/A 20 N/A 5 – 120 20 – 40 10 60
Adaptive
(L/min) cleared
N/A N/A 30 5 – 120 30 – 50 20 80
Low Peak Low
Pressure Pressure
Alarm 5 5 5 1 - 55 N/A N/A N/A
Alarm
(cmH2o)
123
Default ventilation Settings
Parameter MEANING INFANT CHILD ADULT FULL RECCOM- LOW HIGH
RANGE ENDED CAUTION CAUTION
8-25 KG 25-45 KG >45KG RANGE ZONE ZONE
High Peak High
Pressure Pressure
Alarm 35 35 35 5 - 60 20 - 35 10 40
Alarm
(cmH2o)
Apnea Alert Apnea
Time Alarm 15 15 20 Yes N/A N/A N/A
(seconds)
High O2 High O2
Alarm 100 100 100 Yes 21-100 N/A N/A
Alarm (%)
Low O2 Low O2 21 21 21 Yes 21-100 N/A N/A
Alarm (%) Alarm
124
iVent101Operator Manual
Parameter MEANING INFANT CHILD ADULT FULL RECCOM- LOW HIGH
RANGE ENDED CAUTION CAUTION
8-25 KG 25-45 KG >45KG RANGE ZONE ZONE
Low Rate Low Rate
Alarm Alarm 10 5 3 Yes N/A N/A N/A
(bpm)
Patient Whether
Disconnect Patient
Alarm (true/ Disconnect
Alarm Is
True True True N/A N/A N/A N/A
false)
Enabled Or
Disabled.
Rise time The rate of
acceleratio
n of
pressure to
reach the
1 1 1 0-9 1-9 NA NA
set
pressure
level
ESens Termination
of pressure
support
breath after
25% 25% 25% 5-80 5-80 NA NA
flow drop
detection
125
Default ventilation Settings
RATE
(ALSO INITIAL INSP. INSP.
MODE
DEFAULT IN PRESSURE TIME
OTHER MODES)
INFANT A/C Pressure 30 15 0.6
CHILD A/C Pressure 20 15 0.8
ADULT A/C Pressure 12 15 1.2
126
iVent101Operator Manual
127
Pneumatic Unit Scheme
The POV – Proportional Outlet Valve is a flap unit, which controls the air volume
by adjusting between an open and closed position. The different positions allow
variable air volumes to be delivered to the patient, in order that the patient will be
ventilated at the set volume or pressure. The surplus air is returned to the turbine
inlet through a bypass.
The core of the pneumatic unit – the motor, impeller, and the POV - is surrounded
by the acoustic box. The acoustic box, built from metal or high density plastic and
covered, on the inside, by air foam, reducing the noise level generated by the
airflow and blower in the system. The airflow sensor measures the flow near the
acoustic box outlet, and the software uses this measurement in controlling the
POV.
From the ventilator, the air is delivered to the patient via the patient circuit. The
pressure-tap assists or measures the air pressure at the patient wye in the single
limb circuit or at the exhalation valve in a dual limb circuit. Pressure inside the
pneumatic unit also measures pressure on the inspiratory side and estimates wye
pressure. This estimated wye pressure is utilized in comparing pressure readings
for various functions and assuring maximum safety.
The exhalation valve is controlled by adjusting pneumatic unit pressures with the
POV valve. Rapid servo control with this valve allows regulation of both PEEP
and inspiratory pressure. The valve is constantly adjusting to prevent higher than
set target pressure during the inspiratory phase of the breath when there is patient
breathing or coughing efforts. This servo-regulation results in venting of pressure
during patient effort and free breathing during all phases of the breath.
128
iVent101Operator Manual
129
Pneumatic Unit Scheme
Notes:
130
iVent101Operator Manual
131
Theory of Breath Delivery
132
iVent101Operator Manual
combined flow and pressure triggering system that utilizes relative
sensitivity units from 1 – 9, 1 being the most sensitive. Clever algorithms
looks for the shape as well as level of pressure and flow changes to
maximize triggering sensitivity while simultaneously resisting auto
triggering from leaks.
The trigger sensitivity is set in the Settings screen (Refer to section 4.2.3:
Modifying the Parameters on page 54.)
133
Theory of Breath Delivery
by the patient (pressure or flow triggering) in Assist Control or SIMV. When
initiated it is controlled and terminated by the ventilator according to the settings.
134
iVent101Operator Manual
Pressure Control Breath
A pressure control breath is a mandatory breath in which the ventilator provides a
constant pressure during the inspiratory phase according to the settings. The
breath can be initiated by the ventilator or by the patient in Assist Control or
SIMV. The inspiratory phase is terminated by the ventilator when inspiration time
is elapsed or when the set tidal volume has been delivered (whichever occurs
first).
Figure E- 4 below demonstrates the pressure and flow behavior when the
ventilator delivers a pressure control breath. When the breath initiated, the
ventilator delivers the maximum possible flow until the patient airway pressure
exceeds the set level. Once the pressure level is exceeded, the ventilator adjusts
the flow to whatever rate is required to maintain the airway pressure between the
target pressure and the actual value, which about 2 cmH2O lower. At the end of
the inspiratory phase the ventilator terminates flow and opens the exhalation
valve allowing the patient to exhale.
135
Theory of Breath Delivery
Inspiratory Pressure
Inspiratory Time
Tidal Volume Limit
Rise Time
For setting the parameters, refer to section 4.2.3: Modifying the Parameters on
page 54.
PRVC Breath
A PRVC breath is a mandatory breath in which the ventilator provides a constant
pressure during the inspiratory phase according to the settings. The breath can be
initiated by the ventilator or by the patient in Assist Control or SIMV. The
inspiratory phase is terminated by the ventilator when inspiration time is elapsed
or if the tidal volume limit is reached at1.5 X set target volume (whichever occurs
first).
Pressure and flow behavior are similar to the pressure control breath above for
each individual breath. The difference is that the ventilator continually adjusts the
pressure target on subsequent breaths attempting to deliver the set target tidal
volume. If the volume delivered is lower than set, the pressure is raised on the
next breath to correct the discrepancy. If the volume is higher than set, the
subsequent breath is delivered with a lower pressure. The amount of change from
breath to breath is greater if the error is greater but never more than 4 cm H20
between breaths.
When PRVC is initiated, the patient receives three test breaths of a set tidal
volume with a short plateau. The ventilator determines the compliance and correct
target pressure and initiates the pressure breaths at the calculated pressure. The
pressure is then adjusted as described to maintain target volumes although the
delivered tidal volumes will change temporarily as demand or compliance
changes.
136
iVent101Operator Manual
PRVC Parameter Settings
The following parameters may be set:
Sensitivity
PEEP
Rate
Inspiratory Time
Target Tidal Volume
Rise Time
Max Inspiratory Pressure
For setting the parameters, refer to section 4.2.3: Modifying the Parameters on
page 54.
Manual Breath
Manual breath is a ventilator breath that is initiated by the operator (Refer to
Manual Breath on page 62.) This breath may be delivered in all ventilation
modes.
If the Manual Breath is initiated during inspiration or during minimum
exhalation time, the iVent101 ventilator waits and delivers the manual breath at
the end of the minimum exhalation time (see Figure E- 5 below).
137
Theory of Breath Delivery
138
iVent101Operator Manual
The airway pressure exceeds a value of 5 cmH2O above the target
pressure.
Three seconds have elapsed.
At that point the ventilator terminates flow, allowing the patient to exhale.
139
Theory of Breath Delivery
MANDATORY BREATH
VENTILATOR By Ventilator Flow Pressure or By the Ventilator,
VOLUME (time- Time when set VT is
CONTROL triggered) delivered.
ASSIST By Patient Flow Pressure or By the Ventilator,
VOLUME Time when set VT is
CONTROL delivered
VENTILATOR By Ventilator Pressure Level Volume By the Ventilator,
PRESSURE when Inspiratory
CONTROL time is elapsed.
ASSIST By Patient Pressure Level Volume By the Ventilator,
PRESSURE when Inspiratory
CONTROL time is elapsed.
VENTILATOR By Ventilator Pressure Level Pressure or By the Ventilator,
PRVC to achieve Volume when Inspiratory
target Vt set time is elapsed.
ASSIST PRVC By Patient Pressure Level Pressure or By the Ventilator,
to achieve Volume when Inspiratory
target Vt set time is elapsed.
MANUAL By Operator Volume or Pressure or By the Ventilator,
Pressure Level Time. when set VT is
delivered, or when
Inspiratory time is
elapsed.
PATIENT BREATH
PRESSURE By Patient Target Inspiratory By Patient (flow drop
SUPPORT Pressure Time or pressure
BREATH (above PEEP) increase).
140
iVent101Operator Manual
141
Theory of Breath Delivery
The speed of change for peak flows and inspiratory time is based upon the
difference between the target flow/time and the desired flow/time. The greater the
difference, the greater the steps changes in peak flow or inspiratory time during
the next breath. Most changes are gradual and may take eight to ten breaths to
make a full change to the desired patient condition. The iVent101 ventilator can
deliver up to 200 liters per minute, but there are other factors limiting the delivery
of air flow. Theses factors include the patient lung compliance, airway resistance,
and the patient circuit compliance and resistance. Inspiratory time is limited to
two seconds or 1:1 I:E ratio based upon the set rate.
Another feature of Adaptive Peak FlowTM is its ability to track and match the
mandatory breath inspiratory flow rate to the patient spontaneous inspiratory flow
demand during SIMV. The ventilator constantly monitors the patient spontaneous
inspiratory flow demand during SIMV. The ventilator constantly monitors the
patient spontaneous breath flow rate. If the mandatory inspiratory flow is less
than the patient average spontaneous inspiratory flow demand, the mandatory
inspiratory flow is increased to the same value to minimize the feeling of “air
hunger” in the patient.
The Adaptive Peak Flow™ value is displayed on Settings screen when the
Ventilation Mode is SIMV Volume Control or A/C Volume Control. Refer to
Setting Ventilation Modes on page 51.
Notes
1. When setting a manual inspiratory flow during low flow conditions, this
flow target may briefly be exceeded as the ventilator controls the flow to
the patient. However, the average flow will be maintained at the set value.
2. If the target flow is set higher than the capability of the ventilator, due to
high resistance and/or low compliance of the patient circuit, the ventilator
attempts to deliver the highest possible flow during the inspiratory phase.
Under these conditions, the inspiratory time is increased in order to deliver
the set tidal volume.
142
iVent101Operator Manual
Definition
Assist/Control mode combines two traditional modes of ventilation: Assist
Ventilation and Control Mechanical Ventilation (CMV). Unlike the standard
Control Ventilation mode, a patient may breathe more frequently than the set
respiratory rate by triggering a mandatory assist breath before the end of CMV
breath cycle.
iVent101 ventilator offers A/C Pressure Control, A/C Volume Control, and A/C
PRVC as three Assist/Control modes.
Description
Figure F- 1 illustrates breath delivery in Assist/Control mode.
143
Ventilation Modes
Parameter Settings
The following parameters may be set:
Breath Rate
Sensitivity
Inspiratory Pressure (A/C Pressure Control)
144
iVent101Operator Manual
Inspiratory time
Pressure Limit
Peak Flow, Adaptive or set value
PEEP
Tidal Volume (A/C Volume Control)
Tidal Volume Limit (A/C Pressure Control)
Volume Target (A/C PRVC)
Rise Time (A/C Pressure Control and A/C PRCV)
For setting the parameters, refer to section 4.2.3: Modifying the Parameters on
page 54.
F.2. SIMV Mode
Definition
Synchronized Intermittent Mandatory Ventilation Mode (SIMV)) ensures that in
spontaneously or partly spontaneously breathing the patient will receive a set
number of mandatory breaths. All breath types are available in this mode.
This mode is identical to Assist/Control mode except that in this mode patient
spontaneous breaths are allowed between mandatory breaths. Another difference
is that although mandatory breaths are synchronized with the patient inspiration,
the breath period is not reset when patient initiates an assist breath. Therefore the
average mandatory rate does not change.
Description
During an SIMV breath cycle, mandatory breath may be initiated and/or occur
during any breath period or at the end of a breath period when no patient efforts
are detected. Spontaneous or Pressure Support breaths may be initiated and/or
occur during breath periods between mandatory breaths.
Figure F-2 illustrates breath delivery in SIMV mode.
146
iVent101Operator Manual
147
Ventilation Modes
Definition
Continuous Positive Airways Pressure (CPAP) is a ventilation mode intended for
patients who are breathing spontaneously at a rate and depth sufficient to meet
their ventilation requirements. During CPAP, the airway pressure remains above
ambient all the times, reducing the work of breathing. A small amount of Pressure
Support is always present (minimum 3 cm H20) to maximize response time and
guarantee minimal imposed work of breathing at all CPAP levels. Many
ventilators deliver this small level of pressure support during spontaneous
breathing whether it is set or not.
Available Breath
Patient pressure support breaths
Manual breath
Description
Figure F-3 illustrates breath delivery in CPAP mode.
148
iVent101Operator Manual
Patient triggering results in the delivery of a patient breath. The iVent101
ventilator maintains airway (Figure F- 3) during the inspiratory phase at a PEEP
level. Breaths can be initiated only after inspiration has ended, and the minimum
exhalation time has elapsed.
Definition
Adaptive Bi-Level is a support mode of ventilation, intended for use in high leak
situations, such as when using a facemask or a nasal CPAP mask but can be used
in intubated patients as well. In this mode, the spontaneous breathing efforts are
supported at high (inspiratory) and lower (expiratory) pressure.
Adaptive Bi-Level is intended for patients who exhibit clinically appropriate
conditions for facemask ventilation. Such patients include those with acute or
chronic respiratory insufficiency secondary to acute exacerbation COPD,
hypercapnic acute respiratory failure, or decompensated heart failure.
Description
149
Ventilation Modes
Adaptive Bi-Level combines Pressure Control and pressure support modes.
Pressure Support Ventilation is delivered and ventilation rate and breath
termination is determined by the patient. If breathing slows or stops the iVent101
ventilator provides mandatory breaths at a rate indicated by the respiratory rate
settings. The default setting should be lower than the spontaneous breathing rate
to reduce dysynchrony with the patient breathing efforts.
Breaths are terminated by a reduction in peak flow to the ESens % setting of the
delivered peak flow.
150
iVent101Operator Manual
Detection
An excessive leak is indicated by detecting flow during the EPAP phase or by
visually examining mask fit on the patient's face.
Intervention
Ensure that an approved and suitable sized nasal or facemask is being used (check
the manufacturer instructions). Adjust the headgear and ensure that there is equal
tension among all the straps. If necessary, apply padding across the nasal bridge
and chin.
Detection
Failure to trigger the ventilator is detected by patient dysynchrony with
mechanical breaths. These are apparent as spontaneous efforts, on the flow
waveform, that do not result in a rise in pressure on the pressure waveform.
Intervention
Make sure that the enclosed volume of the facemask is not too large, or that the
facemask is too loose (Refer to Excessive Leak, page 152). Decrease the
Sensitivity level if appropriate.
152
iVent101Operator Manual
Detection
Auto cycling is detected by observation of a rapid auto cycling pattern,
independent of the patient breathing pattern.
Intervention
Increase the sensitivity setting. Observe the patient and confirm that auto cycling
persists.
Confirm that the facemask interface has a tight seal.
If the condition persists, recalibrate the ventilator, or refer it to a qualified service
technician.
F.8.5. Cycling Dysynchrony
Detection
Delayed I:E cycling is indicated by ventilator patient dysynchrony and breath
stacking.
Intervention
Observe the patient condition and change the settings to synchronized between
the patient breath and the iVent101 ventilator.
Optimizing expiratory synchrony is important for patients with obstructive
ventilatory disorder. These patients generally require prolonged expiratory
periods to optimize airway emptying, and to minimize breath stacking. In these
patients it may be necessary to switch to Pressure Control if managing inspiratory
time is deemed necessary.
153
Ventilation Modes
Notes:
154
iVent101Operator Manual
G.1. Introduction
G.1.1. Scope
To list the open sources software which are used in the iVent101 or support its
operation as external components, and provide their licenses.
G.1.2. Purpose
To provide the operator of the iVent101 a list of all the open source software that
have been used in programming the Software or are necessary to its operation,
and to provide the information required in order to comply with the licenses of
such open source software.
G.1.3. Overview
During the development of the iVent101 certain open source software were been
used. In addition, different open source software are support its operation as
external components, without being a part of the iVent101. The following
document provides the licenses which have to be distributed with any product that
combined or aggregated with such open source software. .
G.1.4. Applicable documents:
Kernel (Linux): GNU General Public License - GNU Project - Free
Software Foundation (FSF)
BusyBox: http://www.busybox.net/license.html
DSP Gateway:
http://dspgateway.sourceforge.net/pub/index.php?Page=License
155
Software Open Sources
G.2. Licenses of Open Sources Software which
Combined with the iVent101
The followings is the open sources software that has been used for developing the
iVent101 and is combined in it as integral part:
DSP Gateway, see section 2.1
G.2.1. DSP Gateway
The SW uses BSD for DSP, version 3. The ARM uses the Kernel, see section
G.3.1.
The license was taken on November 9, 2006.
156
iVent101Operator Manual
THIS SOFTWARE IS PROVIDED BY THE COPYRIGHT HOLDERS AND CONTRIBUTORS
"AS IS" AND ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, BUT NOT
LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR
A PARTICULAR PURPOSE ARE DISCLAIMED. IN NO EVENT SHALL THE COPYRIGHT
OWNER OR CONTRIBUTORS BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL,
SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES (INCLUDING, BUT NOT
LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES; LOSS OF USE,
DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER CAUSED AND ON
ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR TORT
(INCLUDING NEGLIGENCE OR OTHERWISE) ARISING IN ANY WAY OUT OF THE USE
OF THIS SOFTWARE, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGE.
157
Software Open Sources
Preamble
The licenses for most software are designed to take away your freedom to share and change it. By
contrast, the GNU General Public License is intended to guarantee your freedom to share and
change free software--to make sure the software is free for all its users. This General Public
License applies to most of the Free Software Foundation's software and to any other program
whose authors commit to using it. (Some other Free Software Foundation software is covered by
the GNU Library General Public License instead.) You can apply it to your programs, too.
When we speak of free software, we are referring to freedom, not price. Our General Public
Licenses are designed to make sure that you have the freedom to distribute copies of free software
(and charge for this service if you wish), that you receive source code or can get it if you want it,
that you can change the software or use pieces of it in new free programs; and that you know you
can do these things.
To protect your rights, we need to make restrictions that forbid anyone to deny you these rights or
to ask you to surrender the rights. These restrictions translate to certain responsibilities for you if
you distribute copies of the software, or if you modify it.
For example, if you distribute copies of such a program, whether gratis or for a fee, you must give
the recipients all the rights that you have. You must make sure that they, too, receive or can get
the source code. And you must show them these terms so they know their rights.
We protect your rights with two steps: (1) copyright the software, and (2) offer you this license
which gives you legal permission to copy, distribute and/or modify the software.
Also, for each author's protection and ours, we want to make certain that everyone understands
that there is no warranty for this free software. If the software is modified by someone else and
passed on, we want its recipients to know that what they have is not the original, so that any
problems introduced by others will not reflect on the original authors' reputations.
Finally, any free program is threatened constantly by software patents. We wish to avoid the
danger that redistributors of a free program will individually obtain patent licenses, in effect
making the program proprietary. To prevent this, we have made it clear that any patent must be
licensed for everyone's free use or not licensed at all.
The precise terms and conditions for copying, distribution and modification follow.
GNU GENERAL PUBLIC LICENSE
TERMS AND CONDITIONS FOR COPYING, DISTRIBUTION AND MODIFICATION
0. This License applies to any program or other work which contains a notice
placed by the copyright holder saying it may be distributed under the terms of
this General Public License. The "Program", below, refers to any such program
158
iVent101Operator Manual
or work, and a "work based on the Program" means either the Program or any
derivative work under copyright law: that is to say, a work containing the
Program or a portion of it, either verbatim or with modifications and/or
translated into another language. (Hereinafter, translation is included without
limitation in the term "modification".) Each licensee is addressed as "you".
Activities other than copying, distribution and modification are not covered by this
License; they are outside its scope. The act of running the Program is not restricted,
and the output from the Program is covered only if its contents constitute a work
based on the Program (independent of having been made by running the Program).
Whether that is true depends on what the Program does.
1. You may copy and distribute verbatim copies of the Program's source code as
you receive it, in any medium, provided that you conspicuously and
appropriately publish on each copy an appropriate copyright notice and
disclaimer of warranty; keep intact all the notices that refer to this License and
to the absence of any warranty; and give any other recipients of the Program a
copy of this License along with the Program.
You may charge a fee for the physical act of transferring a copy, and you may at
your option offer warranty protection in exchange for a fee.
2. You may modify your copy or copies of the Program or any portion of it, thus
forming a work based on the Program, and copy and distribute such
modifications or work under the terms of Section 1 above, provided that you
also meet all of these conditions:
a. You must cause the modified files to carry prominent notices stating that
you changed the files and the date of any change.
b. You must cause any work that you distribute or publish, that in whole or
in part contains or is derived from the Program or any part thereof, to be
licensed as a whole at no charge to all third parties under the terms of this
License.
c. If the modified program normally reads commands interactively when
run, you must cause it, when started running for such interactive use in
the most ordinary way, to print or display an announcement including an
appropriate copyright notice and a notice that there is no warranty (or
else, saying that you provide a warranty) and that users may redistribute
the program under these conditions, and telling the user how to view a
copy of this License. (Exception: if the Program itself is interactive but
does not normally print such an announcement, your work based on the
Program is not required to print an announcement.)
159
Software Open Sources
These requirements apply to the modified work as a whole. If identifiable sections of
that work are not derived from the Program, and can be reasonably considered
independent and separate works in themselves, then this License, and its terms, do
not apply to those sections when you distribute them as separate works. But when
you distribute the same sections as part of a whole which is a work based on the
Program, the distribution of the whole must be on the terms of this License, whose
permissions for other licensees extend to the entire whole, and thus to each and every
part regardless of who wrote it.
Thus, it is not the intent of this section to claim rights or contest your rights to work
written entirely by you; rather, the intent is to exercise the right to control the
distribution of derivative or collective works based on the Program.
In addition, mere aggregation of another work not based on the Program with the
Program (or with a work based on the Program) on a volume of a storage or
distribution medium does not bring the other work under the scope of this License.
3. You may copy and distribute the Program (or a work based on it, under Section
2) in object code or executable form under the terms of Sections 1 and 2 above
provided that you also do one of the following:
a. Accompany it with the complete corresponding machine-readable source
code, which must be distributed under the terms of Sections 1 and 2
above on a medium customarily used for software interchange; or,
b. Accompany it with a written offer, valid for at least three years, to give
any third party, for a charge no more than your cost of physically
performing source distribution, a complete machine-readable copy of the
corresponding source code, to be distributed under the terms of Sections
1 and 2 above on a medium customarily used for software interchange;
or,
c. Accompany it with the information you received as to the offer to
distribute corresponding source code. (This alternative is allowed only
for noncommercial distribution and only if you received the program in
object code or executable form with such an offer, in accord with
Subsection b above.)
The source code for a work means the preferred form of the work for making
modifications to it. For an executable work, complete source code means all the source
code for all modules it contains, plus any associated interface definition files, plus the
scripts used to control compilation and installation of the executable. However, as a
special exception, the source code distributed need not include anything that is
normally distributed (in either source or binary form) with the major components
(compiler, kernel, and so on) of the operating system on which the executable runs,
unless that component itself accompanies the executable.
160
iVent101Operator Manual
If distribution of executable or object code is made by offering access to copy from a
designated place, then offering equivalent access to copy the source code from the same
place counts as distribution of the source code, even though third parties are not
compelled to copy the source along with the object code.
4. You may not copy, modify, sublicense, or distribute the Program except as
expressly provided under this License. Any attempt otherwise to copy, modify,
sublicense or distribute the Program is void, and will automatically terminate
your rights under this License. However, parties who have received copies, or
rights, from you under this License will not have their licenses terminated so
long as such parties remain in full compliance.
5. You are not required to accept this License, since you have not signed it.
However, nothing else grants you permission to modify or distribute the
Program or its derivative works. These actions are prohibited by law if you do
not accept this License. Therefore, by modifying or distributing the Program (or
any work based on the Program), you indicate your acceptance of this License
to do so, and all its terms and conditions for copying, distributing or modifying
the Program or works based on it.
6. Each time you redistribute the Program (or any work based on the Program), the
recipient automatically receives a license from the original licensor to copy,
distribute or modify the Program subject to these terms and conditions. You
may not impose any further restrictions on the recipients' exercise of the rights
granted herein. You are not responsible for enforcing compliance by third
parties to this License.
7. If, as a consequence of a court judgment or allegation of patent infringement or
for any other reason (not limited to patent issues), conditions are imposed on
you (whether by court order, agreement or otherwise) that contradict the
conditions of this License, they do not excuse you from the conditions of this
License. If you cannot distribute so as to satisfy simultaneously your obligations
under this License and any other pertinent obligations, then as a consequence
you may not distribute the Program at all. For example, if a patent license would
not permit royalty-free redistribution of the Program by all those who receive
copies directly or indirectly through you, then the only way you could satisfy
both it and this License would be to refrain entirely from distribution of the
Program.
If any portion of this section is held invalid or unenforceable under any particular
circumstance, the balance of the section is intended to apply and the section as a
whole is intended to apply in other circumstances.
It is not the purpose of this section to induce you to infringe any patents or other
property right claims or to contest validity of any such claims; this section has the
161
Software Open Sources
sole purpose of protecting the integrity of the free software distribution system,
which is implemented by public license practices. Many people have made
generous contributions to the wide range of software distributed through that system
in reliance on consistent application of that system; it is up to the author/donor to
decide if he or she is willing to distribute software through any other system and a
licensee cannot impose that choice.
This section is intended to make thoroughly clear what is believed to be a
consequence of the rest of this License.
8. If the distribution and/or use of the Program is restricted in certain countries
either by patents or by copyrighted interfaces, the original copyright holder who
places the Program under this License may add an explicit geographical
distribution limitation excluding those countries, so that distribution is permitted
only in or among countries not thus excluded. In such case, this License
incorporates the limitation as if written in the body of this License.
9. The Free Software Foundation may publish revised and/or new versions of the
General Public License from time to time. Such new versions will be similar in
spirit to the present version, but may differ in detail to address new problems or
concerns.
Each version is given a distinguishing version number. If the Program specifies a
version number of this License which applies to it and "any later version", you have
the option of following the terms and conditions either of that version or of any later
version published by the Free Software Foundation. If the Program does not specify
a version number of this License, you may choose any version ever published by the
Free Software Foundation.
10. If you wish to incorporate parts of the Program into other free programs whose
distribution conditions are different, write to the author to ask for permission.
For software which is copyrighted by the Free Software Foundation, write to the
Free Software Foundation; we sometimes make exceptions for this. Our
decision will be guided by the two goals of preserving the free status of all
derivatives of our free software and of promoting the sharing and reuse of
software generally.
NO WARRANTY
11. BECAUSE THE PROGRAM IS LICENSED FREE OF CHARGE, THERE IS
NO WARRANTY FOR THE PROGRAM, TO THE EXTENT PERMITTED
BY APPLICABLE LAW. EXCEPT WHEN OTHERWISE STATED IN
WRITING THE COPYRIGHT HOLDERS AND/OR OTHER PARTIES
PROVIDE THE PROGRAM "AS IS" WITHOUT WARRANTY OF ANY
162
iVent101Operator Manual
KIND, EITHER EXPRESSED OR IMPLIED, INCLUDING, BUT NOT
LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY
AND FITNESS FOR A PARTICULAR PURPOSE. THE ENTIRE RISK AS
TO THE QUALITY AND PERFORMANCE OF THE PROGRAM IS WITH
YOU. SHOULD THE PROGRAM PROVE DEFECTIVE, YOU ASSUME
THE COST OF ALL NECESSARY SERVICING, REPAIR OR
CORRECTION.
12. IN NO EVENT UNLESS REQUIRED BY APPLICABLE LAW OR AGREED
TO IN WRITING WILL ANY COPYRIGHT HOLDER, OR ANY OTHER
PARTY WHO MAY MODIFY AND/OR REDISTRIBUTE THE PROGRAM
AS PERMITTED ABOVE, BE LIABLE TO YOU FOR DAMAGES,
INCLUDING ANY GENERAL, SPECIAL, INCIDENTAL OR
CONSEQUENTIAL DAMAGES ARISING OUT OF THE USE OR
INABILITY TO USE THE PROGRAM (INCLUDING BUT NOT LIMITED
TO LOSS OF DATA OR DATA BEING RENDERED INACCURATE OR
LOSSES SUSTAINED BY YOU OR THIRD PARTIES OR A FAILURE OF
THE PROGRAM TO OPERATE WITH ANY OTHER PROGRAMS), EVEN
IF SUCH HOLDER OR OTHER PARTY HAS BEEN ADVISED OF THE
POSSIBILITY OF SUCH DAMAGES.
END OF TERMS AND CONDITIONS
163
Software Open Sources
You should have received a copy of the GNU General Public License along with this
program; if not, write to the Free Software Foundation, Inc., 51 Franklin St, Fifth
Floor, Boston, MA 02110-1301 USA
Also add information on how to contact you by electronic and paper mail.
If the program is interactive, make it output a short notice like this when it starts in an interactive
mode:
Gnomovision version 69, Copyright (C) year name of author
Gnomovision comes with ABSOLUTELY NO WARRANTY; for details type `show
w'.
This is free software, and you are welcome to redistribute it under certain conditions;
type `show c' for details.
The hypothetical commands `show w' and `show c' should show the appropriate parts of the
General Public License. Of course, the commands you use may be called something other than
`show w' and `show c'; they could even be mouse-clicks or menu items--whatever suits your
program.
You should also get your employer (if you work as a programmer) or your school, if any, to sign a
"copyright disclaimer" for the program, if necessary. Here is a sample; alter the names:
Yoyodyne, Inc., hereby disclaims all copyright interest in the program `Gnomovision' (which
makes passes at compilers) written by James Hacker.
<signature of Ty Coon>, 1 April 1989
Ty Coon, President of Vice
This General Public License does not permit incorporating your program into proprietary
programs. If your program is a subroutine library, you may consider it more useful to permit
linking proprietary applications with the library. If this is what you want to do, use the GNU
Library General Public License instead of this License.
G.3.2. Busybox
The uses BusyBox version 1221
This license was taken on the 28 December 2006.
164
iVent101Operator Manual
BusyBox is licensed under the GNU General Public License, version 2
BusyBox is licensed under the GNU General Public License version 2, which is often abbreviated
as GPLv2. (This is the same license the Linux kernel is under, so you may be somewhat familiar
with it by now.)
A complete copy of the license text is included in the file LICENSE in the BusyBox source code.
Anyone thinking of shipping BusyBox as part of a product should be familiar with the licensing
terms under which they are allowed to use and distribute BusyBox. Read the full test of the GPL
(either through the above link, or in the file LICENSE in the busybox tarball), and also read the
Frequently Asked Questions about the GPL.
Basically, if you distribute GPL software the license requires that you also distribute the source
code to that GPL-licensed software. So if you distribute BusyBox without making the source code
to the version you distribute available, you violate the license terms, and thus infringe on the
copyrights of BusyBox. (This requirement applies whether or not you modified BusyBox; either
way the license terms still apply to you.) Read the license text for the details.
165
Software Open Sources
License enforcement
BusyBox's copyrights are enforced by the Software Freedom Law Center (you can contact them at
gpl@busybox.net), which "accepts primary responsibility for enforcement of US copyrights on the
software... and coordinates international copyright enforcement efforts for such works as
necessary." If you distribute BusyBox in a way that doesn't comply with the terms of the license
BusyBox is distributed under, expect to hear from these guys. Their entire reason for existing is to
do pro-bono legal work for free/open source software projects. (We used to list people who violate
the BusyBox license in The Hall of Shame, but these days we find it much more effective to hand
them over to the lawyers.)
Our enforcement efforts are aimed at bringing people into compliance with the BusyBox license.
Open source software is under a different license from proprietary software, but if you violate that
license you're still a software pirate and the law gives the vendor (us) some big sticks to play with.
We don't want monetary awards, injunctions, or to generate bad PR for a company, unless that's
the only way to get somebody that repeatedly ignores us to comply with the license on our code.
166
iVent101Operator Manual
d) The use of accessories, transducers and cables other than those specified, with the
exception of transducers and cables sold by the manufacturer of the iVent101 as
replacement parts for internal components, may result in increased emissions or decreased
immunity of the iVent101
e) The iVent101 should not be used adjacent to or stacked with other equipment and that if
adjacent or stacked use is necessary, the iVent101 should be observed to verify normal
operation in the configuration in which it will be used.
d) The use of the accessory, transducer or cable with equipment and systems other than those
specified may result in increased emissions or decreased immunity of the iVent101.
167
Index
Electro-magnetic emissions
The iVent101 is designed to be used in the electro-magnetic environment specified below. The
equipment’s customer or user must ensure that it is used in such an environment.
Emission test Compliance Recommended electro-magnetic environment
RF emissions The iVent101 uses RF energy only for its internal function.
Group 1 Therefore its RF emissions are very low and are not likely to cause
CISPR 11
any interference in nearby electronic equipment.
RF emissions
Class B
CISPR 11
Harmonic
emissions Class A The iVent101 is suitable for use in all establishments, including
IEC 61000-3-2 domestic establishments and those directly connected to the public
Voltage low-voltage power supply network that supplies buildings used for
fluctuations/flick domestic purposes.
Complies
er emissions
IEC 61000-3-3
168
iVent101Operator Manual
Electro-magnetic immunity
The iVent101 is designed to be used in the electro-magnetic environment specified below. The
equipment’s customer or user must ensure that it is used in such an environment.
Immunity IEC 60601 Compliance Recommended electro-magnetic
test test level level environment
Electrostatic
The floors should be wood, concrete or ceramic tile.
discharge ± 6 kV contact ± 6 kV contact
If the floors are covered with synthetic material, the
(ESD) IEC ± 8 kV air ± 8 kV air
relative humidity should be at least 30%.
61000-4-2
± 2 kV on
± 2 kV on
power supply
Electrical fast power supply
lines Mains power quality should be that of a typical
transient/ burst lines
± 1 kV on commercial or hospital environment.
IEC 61000-4-4 ± 1 kV on input
input / output
/ output lines
lines
± 1 kV ± 1 kV
differential differential
Surge Mains power quality should be that of a typical
mode mode ± 2 kV
IEC 61000-4-5 commercial or hospital environment.
± 2 kV common common
mode mode
< 5% UT < 5% UT
(>95% dip in (>95% dip in
Mains power quality should be that of a typical
UT) for 0.5 UT) for 0.5
commercial or hospital environment. If the user of
Voltage dips, cycles 40% UT cycles 40% UT
short (60% dip in UT) (60% dip in the iVent101 requires continued operation during
interruptions for 5 cycles UT) for 5 power mains interruptions, it is recommended that
and voltage 70% UT cycles the iVent101 be powered from an uninterruptible
variations on (30% dip in UT) 70% UT power supply or a battery.
power supply for 25 cycles (30% dip in
input lines < 5% UT UT) for 25 Note: Voltage reduction didn't influence the
IEC 61000-4-11 (>95% dip in cycles iVent101 which is always supplied with the back
UT) for 5 s < 5% UT
up battery
(>95% dip in
UT) for 5 s
Power
frequency Power frequency magnetic fields should be at levels
(50/60 Hz) 3 A/m 3 A/m characteristic of a typical location in a typical
magnetic field commercial or hospital environment.
IEC 61000-4-8
NOTE: UT is the AC mains voltage prior to application of the test level.
169
Index
170
iVent101Operator Manual
marked with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the highest frequency interval is applied NOTE 2: This guide is not
applicable in every situation. The issue of electro-magnetic waves is affected by the absorption and the
reflection of structures, objects and people.
(a) The ISM bands (Industrial, Scientific and Medical) between 150 kHz and 80 MHz are 6.765 MHz to 6.795
MHz, 13.553 MHz to 13.567 MHz, 26.957 MHz to 27.283 MHz and 40.66 to 40.70 MHz.
(b) Compliance levels in ISM frequency bands between 150 kHz and 80 MHz and between 80 MHz and 2.5
GHz are intended to decrease the likelihood that mobile/portable communication equipment could cause
interference if it is inadvertently brought into patient areas. This is why an additional 10/3 factor is used in
calculating the distance of separation for emitters in these frequency ranges.
(c) Fields from fixed emitters, such as cordless telephone bases, mobile radios, CB radios, AM and FM radio
emissions, TV emissions may not be accurately predicted in theory. In order to evaluate the electro-
magnetic environment from fixed emitters, an electro-magnetic surveillance site must be considered. If the
force field measured in the environment where the iVent101 has to be used, exceeds the above
applicable RF levels, theiVent101 must be watched to check if it is working normally. If abnormal
performance is observed, extra measures have to be taken such as redirecting or moving the iVent101.
(d) Above the 150 kHz to 80 MHz frequency range, force fields should be less than 10 V/m.
171
Index
NOTE 1: At 80 MHz and 800 MHz, the separation distance of the highest frequency range is applied.
NOTE 2: ISM bands (Industrial, Scientific and Medical) between 150 kHz and 80 MHz are 6.765 MHz to
6.795 MHz,
13.553 MHz to 13.567 MHz, 26.957 MHz to 27.283 MHz and 40.66 to 40.70 MHz.
NOTE 3: An additional factor of 10/3 is used in calculating the separation distance for emitters in the 150
kHz and 80 MHz ISM range as well as the frequency range between 80 MHz and 2.5 GHz in order to
reduce the probability that mobile/portable communication equipment causes interference if they are
unknowingly taken near the patient.
NOTE 4: This guide is not applicable in every situation. The issue of electro-magnetic waves is affected by
the absorption and the reflection of structures, objects and people.
172
iVent101Operator Manual
Index
PRVC breath 136
A
Button volume 67
A/C pressure control 52
A/C PRVC 52 C
A/C volume control 52 Cleaning Procedure 95
Adaptive Bi-Level 52, 149 Clinical view 40
Alarm Continuous Positive Airways
Parameters 74 Pressure 148
Priority 73 CPAP 148
Settings 74 CPAP/PSV 52
Severity 73
D
Silencing 78
Volume 78 Date 66
Alarm volume 67 Default settings 121
Apnea 58, 151 Dysynchrony 150, 151
Apnea Mode 60 E
Apnea Settings 58
Assist breath 133 Easy ExhaleTM 150
Assist/Control mode 143 Emergency Backup 151
Exhalation phase 131
B
F
Basic Night View 39
Basic View 39 Filter
Battery Air Inlet 31
Integrated 27, 102 Flow 72
Maintenance 102 G
Breath Styles 133
Mandatory 133 General configuration 65
Mandatory volume control 133 Graph tab 72
Manual Breath 137 H
Patient breath 138
Heated humidification 30
Pressure control 135
HME 30
Pressure support 138
173
Index
I Patient breath 138
I. Time 54 Patient triggering 132
Patient type 50
I:E 141, 153
Inspiratory phase 131 Pause ventilator 61
Peak flow 55, 141
Inspiratory time 54
PEEP 55, 131, 138, 144, 149, 150
Integrated Battery 27, 102
Pressure 72
Recharge 102
Pressure control breath 135
Intended Use 7
Pressure limit 55
L Pressure support breath 138
Leak 149, 152 Preventive maintenance procedures
Lock screen 68 96
Log book 92 PRVC breath 136
PSV 55
M
R
Maintenance 96
Battery 102 Rate 54
Mandatory breath 133 S
Manual breath 62, 137
Screen 43
MDI 30
Screen brightness 70
Menu 48
Screen lock 63
N Sensitivity 54, 132
New patient 50 Settings
Type 50
O Ventilation parameters 49
O2 69 Silencing the alarm 78
O2% SIMV 145
Low Oxygen Supply 33 SIMV pressure control 52
OVT 104 SIMV PRVC 52
SIMV volume control 52
P Sounds levels 67
P (Limit) 55 Starting ventilation 61
P. High 55 Stopping ventilation 61
P. Low 55 Synchronized Intermittent
Mandatory Ventilation Mode 145
174
iVent101Operator Manual
T Adaptive Bi-Level 149
Apnea 151
Tab
Assist/Control mode 143
Graph 72
CPAP 148
Settings 44
Easy ExhaleTM 150
Test
OVT 104 Emergency Backup 151
Tidal volume 133, 135, 136, 141 SIMV 145
Time 66 Ventilation parameters 49
Trigger 132 Volume 78
Volume control breath 133
V Vt. Limit 54
Ventilation Vt. Set 55
Pausing 61 W
Starting 61
Stopping 61 Weight selection 50
Ventilation Mode 17, 51
175
Index
Notes:
176