Professional Documents
Culture Documents
FSMC Acquisition A
FSMC Acquisition A
FSMC Acquisition A
・ SSD is exchanged.
(1) Complete the X-ray tube alignment and the FMD alignment.
(4) Set L phantom at the center of scanning field. Since the patient couch
moves during data correction, do not place any objects that may prevent
patient couch movement.
(5) Select the utility button in the startup window. Then select [[Maintenance
Utility]] and then [[Calibration]] to open the calibration window.
(7) When [[Acquire]] is clicked, the screen for last scan time confirmation is
displayed as shown below. If the time is correct, click [[OK]].
Figure xx.2-3
(8) After the following window is displayed, click [[OK]] again.
Preparation for starting acquisition of the focal spot motion correction data
starts.
After preparation is complete, the scan start button on the keyboard is lit.
When the scan start button is pressed, acquisition of the focal spot motion
correction data starts.
Figure xx.2-4
(9) While correction data is being acquired, screen (A) shown below is
displayed. When heat-up scanning is being performed for changing the focal
spot, screen (B) shown below is displayed. It takes approximately 1 hour ( It
depends on the system ).
Figure xx.2-7
The focal spot motion correction function can be set to ON or OFF in the
following manner.
(1) Click the Utility button in the startup window and then select
[[Maintenance]] and [[Calibration data acquisition]] in that order. The
calibration window is displayed.
Figure xx.3-1
(2) When [[Maintenance]] at the lower left of the window is clicked, the
window changes to the setting window shown in figure xx-3-2.
Figure xx.3-2
When [[a Fov]] is clicked, the focal spot motion correction function
ON/OFF setting window appears.
For the "Acq Mode", select "Normal" to have scanning performed with the
DR mode set to OFF in the eXam Plan, or "Dose Reduction" to have
scanning performed with the DR mode set to ON in the eXam Plan.
For "Tube Voltage", set the desired tube voltage.
In the table, the slice thickness values are displayed in the left-most column
and the FOV sizes are displayed in the top row. The remainder of the table
indicates the function setting status for each set of conditions as described
below:
Figure xx.3-3
(4) Move the cursor to the field located to the left of [A Fov], change the value
to 1 to set the function to ON or 0 to set the function to OFF, and click
[[Change]] (located to the right of [A Fov]). The value in the
corresponding cell is changed to the entered value.
Figure xx.3-4
Be sure to perform this procedure after the setting has been changed.
Otherwise, the change is not accepted.
(6) Click [[QUIT]] to close the setting window and then click [[QUIT]] in the
calibration window.