Professional Documents
Culture Documents
Mak 2002
Mak 2002
Abstract: The contact mechanics in ceramic-on-ceramic hip implants has been analysed in this study
using the nite element method. Only the ideal conditions where the contact occurs within the
acetabular cup were considered. It has been shown that the contact pressure distribution and the
contact area at the main articulating bearing surfaces depend largely on design parameters such as
the radial clearance between the femoral head and the acetabular cup, as well as the thickness of the
ceramic insert. For the ceramic-on-ceramic hip implants used in clinics today, with a minimum 5-mm-
thick ceramic insert, it has been shown that the radius of the contact area between the femoral head
and the acetabular cup is relatively small compared with that of the femoral head and the ceramic
insert thickness. Consequently, Hertz contact theory can be used to estimate the contact parameters
such as the maximum contact pressure and the contact area.
NOTATION being highly wettable, hard, smooth and inert [1]. The
alumina-on-alumina bearing couple was introduced in
a contact radius 1970 [2], while zirconia-on-zirconia or zirconia-on-
c radial clearance, R R alumina combinations have recently received signicant
2 1 attention [3, 4]. The alumina-on-alumina combination
E modulus of elasticity for the ceramic bearing
surface has been shown to produce extremely low friction [5, 6 ]
p pressure and low wear in articial hip joint replacements under
p maximum contact pressure normal conditions from both clinical [2] and simulator
0 studies [7, 8]. Furthermore, it has been shown that the
R equivalent radius
R femoral head radius ceramic wear particles can also cause osteolysis, but
1 cause less systemic eVects than the metallic wear particles
R acetabular socket radius
2 [9, 10]. Therefore, minimizing wear and the number of
w applied load in the vertical direction
x, y horizontal and vertical coordinates respectively ceramic wear debris particles in ceramic-on-ceramic hip
implants is critical. Extensive research and testing have
î Poisson’s ratio for the ceramic bearing surface been conducted in the literature to demonstrate the
intrinsically low wear and abrasion resistance of ceramic
materials.
1 INTRODUCTION In the meantime, a number of nite element studies
have been carried out to examine the stresses with the
Long-term concerns over ultra-high molecular weight ceramic components in order to reduce the likelihood of
polyethylene ( UHMWPE ) wear debris and the resulting fracture failure, in particular the xation of the ceramic
osteolysis have promoted renewed interest in alternative femoral head to the metallic stem employing a taper
bearing couples for total hip implants. Medical grade connection [11]. The importance of the design param-
ceramics such as alumina and zirconia have been used eters for the bearing surfaces such as the radial clearance
in ceramic-on-ceramic hip implants because of their between the femoral head and the acetabular cup, as
attractive tribological and biological features such as well as the bearing surface construct, has not been
adequately addressed. For example, matched pairs with
The MS was received on 19 October 2001 and was accepted after revision clearances of 7–10 ím were used from 1977 to 1993
for publication on 28 March 2002.
* Corresponding author: Department of Mechanical and Medical [12, 13]. However, relatively large clearances have been
Engineering, University of Bradford, Bradford BD7 1DP, UK. introduced recently in order to reduce manufacturing
H06201 © IMechE 2002 Proc Instn Mech Engrs Vol 216 Part H: J Engineering in Medicine
tionless to represent the well-lubricated condition. a, were transformed into the angular coordinate and the
Bilinear axisymmetric nite elements with four nodes contact half-angle, using the femoral head radius R as
1
were used to mesh all the components. A xed load of shown Fig. 3.
2500 N was applied through the pole of both the cup
and the head. The elastic modulus and Poisson’s ratio
were chosen to be 380 GPa and 0.26, 1 GPa and 0.4 and 3 RESULTS
210 GPa and 0.3 for alumina, the polyethylene backing
and the metallic stem respectively. The nite element
The eVect of mesh sensitivity was rstly examined on
model thus formulated was solved using ABAQUS
the predicted contact parameters at the bearing surface.
( Version 5.8–9). Both the contact area and the contact
For the nominal case, with a ceramic insert and an
stresses were readily obtained from the nite element
model. The Hertz contact theory, based on the assump- UHMWPE inlay, both of 5 mm thickness, it was found
tion of semi-innite solids for the ceramic bearing sur- that 36 and 8 elements, in the circumferential and the
radial directions respectively, for each component were
faces, was also used to predict the contact pressure
adequate. This also resulted in a satisfactory element
distribution, p, and the contact radius, a. This was done
aspect ratio of about unity for the majority of the
by means of an equivalent ball-on-plane model using the
elements, as shown in Fig. 2. The boundary conditions
equivalent radius R determined from the radii of the
femoral head and the cup, R and R , and the radial at the interface between the UHMWPE inlay and the
1 2 ceramic insert, as well as at the taper connection between
clearance, c:
the ceramic femoral head and the metallic stem, were
R R found to have a negligible eVect on the predicted contact
R= 1 2 (1)
c parameters at the articulating surfaces. Therefore, only
the results for the fully bonded condition at these
This enabled the contact radius, a, and the maximum interfaces are presented in this study.
contact pressure, p , to be determined under a given
0 Figure 4a shows the predicted contact pressure distri-
load, w: bution at the bearing surfaces for the nominal con-
[3wR(1 î2)] 1/3
G H
ditions, with a ceramic insert and an UHMWPE inlay,
a=
2E
(2) both of 5 mm thickness, and a radial clearance of 40 ím.
The contact pressure distribution predicted from Hertz
3w contact theory is also superimposed for the purpose of
p = (3)
0 2ða2 comparison. Figures 4b and c show the contour plot of
the von Mises and the larger principal stresses within
The corresponding contact pressure distribution was both the ceramic and the UHMWPE components.
given as follows: The eVect of varying the radial clearance between 20
and 80 ím on the predicted contact pressure distribution
C ABD
x 2 1/2
p=p 1 (4) is shown in Fig. 5 for a given ceramic insert and an
0 a
UHMWPE inlay, both of 5 mm thickness. Figure 6
For the purpose of comparison with the ball-in-socket shows the eVect of varying the thickness of the ceramic
geometry, the horizontal axis, x, and the contact radius, insert for a xed radial clearance of 40 ím, while the
H06201 © IMechE 2002 Proc Instn Mech Engrs Vol 216 Part H: J Engineering in Medicine
Fig. 4b Prediction of the von Mises stress distribution within Fig. 5 EVect of the radial clearance on the predicted contact
various components (ceramic insert thickness, 5 mm; pressure distribution (ceramic insert thickness, 5 mm;
UHMWPE inlay thickness, 5 mm; radial clearance, UHMWPE inlay thickness, 5 mm)
40 ím)
total thickness of the acetabular cup including the cer- design parameters such as the radial clearance between
amic insert and the polyethylene backing was xed at the femoral head and the acetabular cup, and the thick-
10 mm. Table 1 compares the predicted maximum con- ness of the acetabular insert. For the nominal conditions
tact pressure and the contact half-angle from the present considered, the predicted contact pressure distribution is
nite element model with the Hertz contact theory. only slightly less than that from the corresponding Hertz
contact theory as shown in Fig. 4a; the diVerence in the
predicted maximum value being less than 15 per cent.
4 DISCUSSION This is expected since the contact half-angle predicted
from the nite element model is approximately 17.5°,
It is clear from the results presented in Section 3 that corresponding to a contact radius of 4 mm, which is less
the contact pressure distributions depend largely on the than the thickness of the ceramic insert of 5 mm. This
Proc Instn Mech Engrs Vol 216 Part H: J Engineering in Medicine H06201 © IMechE 2002
Proc Instn Mech Engrs Vol 216 Part H: J Engineering in Medicine H06201 © IMechE 2002