Professional Documents
Culture Documents
Electronics Lab Machines
Electronics Lab Machines
DC POWER SUPPLY
The power supplies that we use have three independent supplies in one casing, which are 5V
fixed, Master (0 - 30V variable) and Slave (0 - 30 V variable).
They have three modes of operation, independent, series and parallel.
The independent mode: The output voltage and current of each supply (master and
slave) are controlled separately and each supply can provide a maximum of 30V, 3A. Both
tracking mode switches must be disengaged.
The series mode:
In this mode, the master and the slave power supplies are internally connected in series,
resulting in one supply with double the voltage capability, a maximum of 60Volts as
opposed to a maximum of 30Volts that the individual supplies can provide. The voltage of
both supplies is set using the master voltage controls. The left tracking switch must be
engaged and the right one disengaged.
The parallel mode: In this mode, the master and the slave power supplies are internally
connected in parallel resulting in one supply with double the current capability, a maximum
of 6 Amps as opposed to a maximum of 3Amps that the individual supplies can provide.
The output current is double the value read from the MASTER CURRENT display because
each supply is providing the same amount of current.
If the voltages required are the same value like (+15V and –15V) use the series mode, the
negative of the slave supply and the positive of the master supply are connected to form the
common ground. Between the common ground and the +ve terminal of the master supply is
a positive voltage i.e. +15V and between the common ground and the -ve terminal of the
slave supply is a negative voltage ie. -15V.
If the voltages required are not the same value like (+10V and –10V), Use the independent
mode, then connect the negative terminal of the master to the positive terminal of the slave
manually. The operation will be the same as the one above except that the master and the
slave voltages can be controlled separately.
Always make sure that the tracking switches are set for the required mode.
Make sure that the current knobs are not at zero but set to the safe current of the circuit to be
powered.
Set the voltage to the proper level, referring to the meter on the supply before you connect
the power supply to the test circuit.
Make sure you connect to the correct output terminal
Make a good connection to the power supply output with the leads that are specifically made
for that not any wire, to avoid damaging the equipment and the shorts.
AC POWER SUPPLY
The AC power supplies are used for circuits that require certain voltages other than the 220
AC that we get from the main supply in the wall.
This can be achieved from either transforming the mains supply to the desired level through
or by means of a transformer (step down) or a variac (variable transformer).
The other type of AC power supply is inverter type where one can have the variable
frequency and amplitude. They can also give different forms of waveforms depending on the
type e.g. the function generator.
A multimeter is the instrument used to measure voltage, current, resistance, capacitance, and
frequency, perform the diode check and the continuity test.
The multi meter has the following controls and connectors:
Digital display: It displays the functions selected like AC/DC, Ω, REL.MAX.MIN. And
unit’s indicators.
Rotary key: It is used to select the function and the Range needed.
COM Input Terminal: Ground input connector.
VΩ Hz ┤├ Input terminal: Positive connector for Volts, Ohms, Diode, Capacitance
and frequency measurements.
mA Input terminal: Positive input connector for current measurements.
A Input terminal: Positive input connector for higher currents, up to 20A maximum for
only 30 seconds.
Function key (Blue): The key is used to change between AC and Dc for current
measurements or Ohm and continuity checking or frequency and ADP measurements.
MIN MAX KEY: This key is used to enter the recording mode. In this mode the
minimum or the maximum readings are stored in the memory.
RELATIVE KEY: This key is used to enter the relative mode. In this mode the meter
uses the value stored in the memory as the reference, as a result the value indicated on the
display will be the difference between the present reading and the value stored in the
memory.
HOLD KEY: This key is used to go in and out of the Data hold mode.
DELAY KEY: Turns on the Hold Function 10seconds late after the key has been
pressed.
RANGE KEY (Manual Range): This key is used to select between manual and auto
ranges. When this key is pressed the auto range goes off, and it has to be pressed for
2seconds to return to auto range.
BACK LIGHT (Yellow): The key is used to turn on and off the light at the back of the
LCD display.
STORE KEY: This key is used to enter the memory store mode. In this mode the last
reading displayed on the meter is stored and if the meter power goes down the stored data is
not erased.
RECALL KEY: It is used to recall the data that is saved in the memory.
Voltage measurements
1. Connect the black test lead to the COM and the RED test lead to VΩ Hz ┤├ terminal.
2. Set the rotary switch to AC or DC function.
3. Connect the test leads across the source or load to be measured.
*NB.When the input voltage is over range the meter buzzer is turned on and the most
significant digit of the LCD display flashes and the three list significant digits are set to
zero.
Current measurement
1. Connect the black test lead to the COM and the RED test lead to the mA terminal for a
maximum of 400mA and to A terminal for a maximum of 10A.
2. Set the rotary switch to the range needed (4mA-400mA) for the current to be measured.
3. Select AC or DC current measurement using the function(blue)key. If the range of the
current needed is not known, start at the highest range and work down.
4. Connect the test leads in series with the load to be measured.
Resistance measurement
1. Connect the black test lead to the COM and the RED test lead to VΩ Hz ┤├ terminal.
2. Set the rotary switch to Ω position to measure resistance.
3. Select the Ω measurement using the function (blue) key.
4. Connect the leads across the resistance to be measured. For accuracy in measurements of
low resistance, press “Range” to set the 400Ω range, short the test leads together and
press the “REL” key to compensate the resistance measurement for the resistance of the
test leads.
Continuity test
1. Connect the black test lead to the COM and the RED test lead to VΩ Hz ┤├
terminal.
2. Set the rotary switch to “Ω•﴿﴿﴿” position.
3. Select the continuity “•﴿﴿﴿”measurement using the function (blue) key.
4. Connect the leads across the resistance to be measured. The buzzer sounds if the
resistance of the circuit measured is lower than 50 ohms.
Diode check
1. Set the rotary switch at diode position.
Capacitance measurement
1. Connect the black test lead to the COM and the RED test lead to VΩ Hz ┤├ terminal.
2. Set the rotary switch to ┤├ position.
3. Observe polarity when measuring polarized capacitors.
4. Connect the test leads across the capacitor to be measured.
*NB. The capacitor under test should be discharged before the testing procedure.
It is best to use the “relative key” to zero offset for accurate measurement.
Frequency measurement
1. Connect the black test lead to the COM and the RED test lead to VΩ Hz ┤├ terminal.
2. Set the rotary switch to “Hz” position.
3. Select the “Hz” measurement using the function key.
*NB: The last one or two digits of the display could be unstable for the frequency below
1kHz and above 20kHz.
The basic function of the function generator is to produce sine, square, triangle waveforms
and a pulse output for testing digital logic circuits.
Function generators have certain controls that allow one to select the type of waveform and
the other controls to adjust the amplitude and the DC level.
Frequency indicator: Indicates the value of the current frequency.
Frequency range selector: Selects the required frequency range.
Function selector: One of the push buttons is pressed to select the desired output wave
form (sine, square, triangle waveform)
Duty function: This knob is pulled out and rotated to adjust the duty circle of the
waveform.
TTL/CMOS: This knob is pulled out and rotated to adjust the CMOS compatible output
from 5-15Vp-p.
DC Offset : This knob is pulled out to select any DC level of the waveform between +-
10V. It is turned clockwise to set a positive DC level waveform and anticlockwise for a
negative DC level waveform.
Amplitude: This knob is to adjust the output peak-to-peak voltage. The knob is pulled out
for an additional attenuation of the output.
TTL/CMOS: TTL/CMOS compatible signal output.
1. Select the range, then rotate the frequency knob to set the required frequency.
2. Connect BNC connector of TTL/CMOS to the oscilloscope to observe the output signal.
3. The output is a square wave fixed to TTL level, suitable for general TTL integrated circuit .
4. If the square waveform of the CMOS level is required, pull out the CMOS knob to adjust the
level.
The oscilloscope is the measuring instrument that lets you to see the graph of the voltage as a
function of time in the circuit.
The voltage to be measured is converted into a visible display by a cathode ray tube.
The oscilloscope contains four functional blocks, the vertical controls, the horizontal controls, the
trigger controls and the display controls.
Vertical controls
Horizontal controls
Trigger controls
The trigger section is the source of most operator problems when using an oscilloscope. These
controls determine the proper time for the sweep to begin in order to produce a stable display. The
trigger controls include:
The MODE switch that selects AUTO or NORMAL operation.
The SOURCE switch, which selects the trigger source, the internal signal, the external trigger
source or the AC, line.
The trigger LEVEL and SLOPE which are used to select a specific point on either the rising or
falling edge of the input signal for generating a trigger.
The variable HOLD OFF control, which allows you to exclude the invalid triggers until the hold off
time has elapsed.
COUPLING switch is used to select the coupling of the trigger circuit in accordance with the
characteristics of the measured signal.
The display system contains controls for adjusting the electron beam. The controls included are:
The power control, which is the main power switch for the scope, when it is on the green LED next
to it, must be ON.
INTENSITY, which controls the brightness of the trace.
FOCUS is used for focusing the trace to the sharpest image.
TRACE ROTATION, which is used to align the sweep with the horizontal graticule line.
Basic operation
Before operating the scope make sure the switches and controls of the instrument are set as follows.
Item Setting
Change the VERT MODE switch to the DUAL state so that both CH1 and CH2 traces can
be displayed.
Apply the calibrator signal to both CH1 and CH2.
CH1 or CH2 must be selected for the triggering source signal by means of the SOURCE
switch, If CH1 and CH2 are synchronised, both waveforms will be displayed stationery.
If one of the waveforms is not stationery, engage the TRIG/ALT.
ADD operation
Voltage (Vp-p) Readings : Count the number of the vertical blocks within the bottom and the
top peak of the signal on the CRT display and multiply it by the scale at which the VOTS/DIV
knob is set.
Frequency: Count the number of the blocks within one period of the signal and multiply it by the
scale at which the TIME/DIV knob is set to get the time value, then take the inverse of that value to
get the frequency value.
Resistors
They are electronic components used to control current flow; they are available in different types
and values. There are fixed and variable resistors.
Capacitors
They are electronic components consisting of two plates separated by an insulating layer and they
are capable of storing electric charge. They are also available in different types and values.
Diodes
They are semiconductor electronic components with two electrodes, an anode and cathode. They
allow current to flow in only one direction. They are found in different types like: the zener diodes,
the schotky diodes and the light emitting diodes (LED).
Transistors
They are three layer semiconductor devices, which can operate as a switch and as an amplifier.
They are also found in different types like PNP and NPN.
Integrated circuits
They are semiconductor devices, which contain a chip comprising many transistors in a complex
circuit.
Breadboard
It is the tool on which the circuits are built temporarily for test purposes. The breadboard consists of
a moulded plastic body, which has a number of holes in the top surface, through which component
pins may be easily inserted. Underneath each hole there is a clip, which holds the component lead
tight enough so it doesn’t fall out. These clips are interconnected, so that by pushing leads of
different components into two holes of one group you have made an electrical contact between the
two leads.