Summary of Revision Status History for F-70H Compressor

Sumitomo Cryogenics P/N 267046D23L (Siemens P/N 10121435)


Customer Effectivity Siemens Siemens SCAI ICR # Summary of Change S/N Service Depot
Revision Date CR # Change # CRSP# Effectivity Disposition
Status
st FS00145 --
00 N/A N/A N/A N/A N/A Original Design – 1 Production Unit

01 February 2007 VCR07-002 172789 4429 SCAI-07-02 Change voltage setting to 400V FS00174 No action

March 2007 VCR07-003 172790 4438 SCAI-07-05 Specify use of Bussman 1.0 A fuse (and delete FS00181 Upgrade & change
02
reference to Littelfuse alternative). revision status label

April 2007 VCR07-004 172791 4439 SCAI-07-07 Firmware upgrade to V1.5 & add labels (revision FS00222 Upgrade & change
03
status, drain & voltage tap tag) SC207001B revision status label to 03

04 July 2007 VCR07-001 172792 4431 SCAI-07-03 Add M6 Earth Stud FS00334 No action
4482 SC207102C
04 October 2007 VCR07-007 514068 4522 SCAI-07-15 Relocate Voltage Tag from main switch to earth FS00502 No action
(no change) stud SC207296C
04 March 2008 VCR07-008 514483 4527 SCAI-07-16 Add mode switch handle extension. FS00651 No action
(no change) SC207690C
August 2008 VCR08-003 522022 4628 SCAI-08-03 Increase the Oil Charge qty from 3300cc to FS00769 Only add oil if required by
05
3400cc SC208057D performance test; do not
change revision status
05 June 2009 VCR08-004A N/A 4722 SCAI-09-01 Change internal motor protector in Copeland FS01110 No action
(no change) capsule. Capsule model number changed from SC209035E
ZCH72C2G-TFD to ZCH72C3G-TFD.
05 December VCR09-001 538165 4874 SCAI-09-13 For final oil separator, change supplier for final FS01443 No action
(no change) 2009 assembly manufacture to SHIMs. Oil separator SC209335E
design does not change. Continue to use Kaydon
internal oil separator element. P/N changed from
267013A to 267600D.
06 May 2010 VCR09-002 538165 4954 Tanashi 10- For adsorber, add alternate materials of FS01626 No action
01 construction and SHIMs as an alternate SC209716
manufacturer. Fit and function remain
unchanged. P/N 300138D remains the same
07 January VCR10-002 551286 4999 SCAI 10-08 Use F70 FH FOS, P/N 267785D as alternative to current FS02080 No action
FOS P/N 267600D. Top and bottom heads were
2011 redesigned from cast domed heads to machined flat end SC210851F
caps. Alternative design is a drop-in replacement
(backwards compatible). Fit and function remain
unchanged. The coalescing element remains the same.
RBS
23 April 2015
Revision Status History for F-70H Compressor

Customer Effectivity Siemens Siemens SCAI ICR # Summary of Change S/N Service Depot
Revision Date CR # Change # CRSP# Effectivity Disposition
Status
07 January FS02591
N/A N/A TR12-01 N/A RoHS Compliance No action
(no change) 2012 SC210918F
08 April 2012 VCR11-002B 568146 5137 SCAI 11-06 Upgrade firmware to V1.7 Upgrade firmware to
nd
Replace oil level switch with 2 sight glass SC212021G V1.7 per SAN12-01 and
5148
Change thread sealant from Loctite 577 to where applicable
Loctite 554 on threads of sight glass, ARV, IRV FS02777 change out A” and “B”
and pressure gauge. Change flathead panel style FOS per SAN11-
screws from mild carbon steel to stainless steel 02.
09 October VCR12-001 572917 5224 SCAI 12-02 Change HV electrical chassis from P/N 267109D SC212978H No action
to 268205D. New HV electrical chassis uses
2012
60601-1 compliant transformer. P/N 268180A
having increased creepage and clearance gaps.
10 August 2013 VCR13-001 590968 5336 SCAI 13-03 Change HV electrical chassis from P/N 268205D FS04001 No action
to 268481D. New HV electrical chassis uses
60601-1 compliant transformer, P/N 268419A SC213516J
having lower operating temperatures and SC313189J
increased creepage and clearance gaps.
10 December VCR12-002 573989 TR12-05 SCAI13-06 Fuji circuit breaker rotary handle has been re- FS04155 No action
designed internally. Internal mounting of rotary SC213882J
(no change) 2013
handle is changed from snap fit to screw type. SC314001J
Revised circuit breaker is functionally equivalent
to current design and was found to be
acceptable for use on F70 series compressors.
Changing the handle from snap fit to screw type
has improved strength, durability and reliability of
circuit breaker.
10 March 2014 VCR13-003 600418 5351 SCAI14-01 On side panel control board, 267024A used in FS04283 Upgrade PCB per
electrical chassis of F70-H and F70-L SC214310
(no change) TR13-44 SAN14-01
compressors, revise C2, C3, C5 and C7 SC314156
electrolytic capacitors with upgraded alternative
capacitor components having higher rated
temperature and better rated electrical
properties.
10 March 2015 VCR14-002 N/A 5476 SCAI14-02 Three port connection fittings on the F70 FS04771 No action
compressor BOS were changed from copper SC215252J
(no change) TR14-40 couplings hand brazed by the BOS manufacturer SC315068J
QAR14- to nickel plated couplings furnace brazed by
04A supplier. BOS P/N changed from P/N 266946A
to 268600A. Reduces potential risk of helium
Leaks and improves process throughput.

2
RBS
23 April 2015
Revision Status History for F-70H Compressor
Customer Effectivity Siemens Siemens SCAI ICR # Summary of Change S/N Service Depot
Revision Date CR # Change # CRSP# Effectivity Disposition
Status
10 June 2015 VCR14-005 0117 5513 SCAI14-03 Change inlet connection on panel mounted (PG) No action
pressure gauge from NPT pipe threads to FS04893
(no change) 5514
O-Ring face seal. Change IRV housing (ARV) TBD
5515 (IRV) TBD
from NPT pipe threads to straight threads &
O-Ring seal. Change ARV from NPT pipe
threads to straight threads & O-Ring seal. SC215346K
SC315078K

3
RBS
23 April 2015