Professional Documents
Culture Documents
Operating Manual Density Meter LB 444
Operating Manual Density Meter LB 444
Table of Contents
Page
1. INTRODUCTION....................................................................................................... 1
1.1 Safety Summary ............................................................................................. 1
1.2 Overview ......................................................................................................... 2
2. SYSTEM DESCRIPTION.......................................................................................... 3
2.1 Use and Function............................................................................................ 3
2.2 The Principle of Measurement........................................................................ 4
2.3 Measuring Configuration................................................................................. 5
2.4 Instrument Description.................................................................................... 7
Radioactive Source......................................................................................... 7
Shieldings ....................................................................................................... 8
2.4.1.1 Shielding Types LB744............................................................ 8
2.4.1.2 Shieldings for Installation in a Container ................................. 9
Detector ........................................................................................................ 11
Evaluation Unit LB 444 ................................................................................. 12
2.4.1.3 General Description ............................................................... 12
2.4.1.4 Display ................................................................................... 13
2.4.1.5 Keypad Function.................................................................... 13
2.4.1.6 Menu Structure (Flow Chart) ................................................. 14
2.5 Software Functions and System Configuration ............................................ 15
3. INSTALLATION...................................................................................................... 26
3.1 General Safety Precautions.......................................................................... 26
3.2 Installation..................................................................................................... 27
Installation on Pipelines ................................................................................ 27
Installation in a Container or Tank ................................................................ 31
3.3 Installation of Water Cooling......................................................................... 33
3.4 Installation of Resistance Thermometer Pt 100............................................ 35
3.5 Installation of Evaluation Unit ....................................................................... 35
4. CONNECTIONS...................................................................................................... 36
4.1 Detector ........................................................................................................ 36
4.2 Evaluation Unit LB 444 ................................................................................. 37
i
Density Meter LB 444 Table of Contents
ii
Density Meter LB 444 Table of Contents
10. APPENDIX.............................................................................................................. 91
10.1 Configuration Checklist................................................................................. 91
10.2 Parameter Protocol ....................................................................................... 92
10.3 Absorption Coefficients ................................................................................. 93
10.4 Temperature Coefficients ............................................................................. 93
10.5 Density of Water as a Function of the Temperature..................................... 96
10.6 Setup Protocol .............................................................................................. 97
10.7 Wiring Diagram ........................................................................................... 100
10.8 Dimensional Drawings ................................................................................ 101
10.9 Index ............................................................................................................. 105
iii
Density Meter LB 444 List of Figures
List of Figures
Page
Figure 1: Principle of measurement .................................................................................... 4
Figure 2: Measuring system mounted on a straight pipeline .............................................. 5
Figure 3: Installation on a U-shaped measuring path ......................................................... 6
Figure 4: Shielding container type LB744 ........................................................................... 8
Figure 5: Shielding for installation in a container ................................................................ 9
Figure 6: Shutter switch for ex-protected area .................................................................. 10
Figure 7: Scintillation counter ............................................................................................ 11
Figure 8: Front Panel LB 444 ............................................................................................ 12
Figure 9: Softkeys functions .............................................................................................. 13
Figure 10: Live display ...................................................................................................... 13
Figure 11: Diagram illustrating the hysteresis ................................................................... 19
Figure 12: Installation on a horizontal pipeline.................................................................. 27
Figure 13: Outdoor installation .......................................................................................... 28
Figure 14: Installation on S or U-shaped measuring path................................................. 29
Figure 15: External installation with shielding and detector .............................................. 30
Figure 16: Installation in a container ................................................................................. 31
Figure 17: Installation in a container with horizontal flow.................................................. 32
Figure 18: Installation in a container with vertical flow...................................................... 32
Figure 19: Installation of water-cooling ............................................................................. 33
Figure 20: Required amount of water-cooling ................................................................... 34
Figure 21: Cable connections at detector ......................................................................... 36
Figure 22: Terminal connection evaluation unit (rear panel)............................................. 37
Figure 23: Rear view of shielding container type LB 744.................................................. 40
Figure 24: Locking mechanism with knurled nut ............................................................... 41
Figure 25: Locking mechanism with spring pin ................................................................. 42
Figure 26: Influence of absorption coefficients on one-point calibration ........................... 63
Figure 27: One-point calibration with additional calibration points.................................... 63
Figure 28: Example of multi-point calibration .................................................................... 64
Figure 29: Response ......................................................................................................... 70
Figure 30: Source holder................................................................................................... 86
iv
OPERATING MANUAL
DENSITY METER LB 444
Revision History
v
Density Meter LB 444 SECTION 1. INTRODUCTION
1. INTRODUCTION
GENERAL WARNINGS
SPECIFIC WARNINGS
SPECIFIC CAUTIONS
1
Density Meter LB 444 SECTION 1. INTRODUCTION
1.2 Overview
The Density Meter LB 444 is designed for density measurements of liquids, sus-
pensions, pulps and bulk materials. The measurement may be made directly in-
line or in a product container or tank.
The hardware and software of the LB 444 system makes it easy to adapt the
system to a variety of measurement geometries and measuring tasks. There-
fore, the settings and parameters of the measuring instrument must be defined
with care for each specific measuring task when commissioning the system. Im-
portant parameters may not easily be changed once the system is operational.
The system should commissioned and set-up only by persons who are familiar
with the instrument. Therefore, all users should read these operating instructions
carefully. In addition, we recommend documenting all settings in a setup proto-
col (see Appendix).
Before starting any work, please read this operating manual carefully!
2
Density Meter LB 444 SECTION 2. SYSTEM DESCRIPTION
2. SYSTEM DESCRIPTION
b) Special configuration
• Back-scatter measurement
• Continuous measurement
• Discontinuous measurement
- Batch operation via keyboard (press <run> button)
- Batch operation via digital input
3
Density Meter LB 444 SECTION 2. SYSTEM DESCRIPTION
I = I0 * e - µ' * ρ * d
Source Detector
ρ
I0 I
The intensity of the radiation picked up by the detector is also dependent on the
distance between source and detector. As in the case of light, the function in-
volved is a square function, i.e. doubling the distance reduces the radiation in-
tensity to ¼ if all other conditions remain the same.
Assuming a constant distance between source and detector and a fixed meas-
uring path, the radiation picked up by the detector is only dependent on the den-
sity of the material to be measured.
4
Density Meter LB 444 SECTION 2. SYSTEM DESCRIPTION
Detector power and output signals use a common two-wire cable connected
between detector and evaluation unit.
Different configurations and mounting devices may be required, depending on
the measurement tasks, the condition of the product being measured and the
pipe or container.
“Fehler! Verweisquelle konnte nicht gefunden werden.Shows a basic setup
in a pipeline with Pt-100 resistance-thermometer and a 90°-mounting device for
density, concentration and mass-flow measurement.
Detector
Pt 100
Shielding
Container
Pipeline
BERTHOLD
Evaluation unit
enter
clear run
5
Density Meter LB 444 SECTION 2. SYSTEM DESCRIPTION
45° and 30° mounting devices are available to extend the measuring path (see
enclosed Technical Information LB 444).
RADIOACTIVE
The system may have to be installed on external supports when the heat insu-
lating material of the pipeline does not allow safe installation (e.g. glass and rock
wool) or when strong vibrations are likely to damage the detector. For further
explanation see section 3. INSTALLATION.
Measurements in Containers
6
Density Meter LB 444 SECTION 2. SYSTEM DESCRIPTION
Radioactive Source
Radioactive sources for industrial applications are always „encapsulated radio-
active substances“ which are tightly welded into a sturdy stainless steel capsule,
so that the radioactive substance cannot leak out. Contamination is therefore not
possible. Moreover activation of the product being measured by the source used
is also not possible. (Laws of Physics).
The following point-shaped Gamma sources can be used for these measuring
241 60 137 244 85 90
configurations. Am, Co, Cs, Cm, Kr, Sr.
7
Density Meter LB 444 SECTION 2. SYSTEM DESCRIPTION
Shieldings
Except for a few special cases, the source is firmly installed into the working
shielding which includes a radiation exit channel to release the active beam to-
wards the detector. The active beam can be shielded during transport and in-
stallation, and when performing service work. This is accomplished by a shutter
mechanism.
The operation of the shutter mechanism and shielding function must be checked
every six months!
Pb
8
Density Meter LB 444 SECTION 2. SYSTEM DESCRIPTION
After taking off the covering cap which is secured by a lock you may open the
knurled screw and, using the stay bar, move the radioactive source forward
(OPEN) or back (CLOSED).
Locking
mechanism
Covering cap
Lead
Source position
Radiation exit CLOSED
channel
45°
9
Density Meter LB 444 SECTION 2. SYSTEM DESCRIPTION
Connection for
pressurized air electr. signal
OPEN/CLOSED
Cover cap
Proximity
switch
Closed
Switch flag
Open
See the drawings that are part of the documentation for detailed information on
the construction and function of the shielding type used.
10
Density Meter LB 444 SECTION 2. SYSTEM DESCRIPTION
Detector
Scintillation counters are used as detectors, since only these detector systems
provide the required high sensitivity to Gamma radiation and the radiation field
does not limit their service life.
The detector comprises
• a sodium iodide crystal doped with Thallium
• a photomultiplier
• an electronics unit.
The built-in microprocessor tracks the configuration, counts the pulse rates,
controls the high voltage, measures the temperature of the probe and an exter-
nal Pt 100, transfers data and performs calibration and control functions.
..
LIGHT FLASH
PHOTO CATHODE − +
The detector probe is mounted in a sturdy stainless steel housing which pro-
tects the instrument against normal environmental stress.
To ensure reliable operation and a long service life, the detector probe
must not be subject to heavy mechanical stress or vibrations.
Cooling must be provided for ambient temperatures above 50°C. (see sec-
tion 3.3 Installation of Water Cooling).
11
Density Meter LB 444 SECTION 2. SYSTEM DESCRIPTION
The illuminated display field on the front panel shows the relevant instrument
function on four lines. The front panel also includes an RS 232 interface.
BERTHOLD
Softkeys
enter
..
..
..
..
clear run
..
..
Funktion keys
The terminal strip on the instrument rear panel includes all terminals for power
for the detector and for the analog and digital output signals. The current output
is isolated and the built-in relays for max.-min.-indication and for error messages
include an isolated contact.
The system automatically corrects for the decrease of source activity with age.
12
Density Meter LB 444 SECTION 2. SYSTEM DESCRIPTION
2.4.1.4 Display
The illuminated display field of the instrument comprises 4 lines; the first three
lines show the menu titles, the currently selected parameters or the current
measurement value. The bottom line shows the current function of the softkey
button located below or, when a measurement is running, the status “run“.
Softkeys
Product select
b) [extern/intern] Softkeys are used to select different menu groups and operating lev-
intern els within the menu structure. Depending on the current position in
^^^ done the menu structure, functions are assigned to these keys, as shown
on the display above the respective key.
<sk1> and
<sk2> switch to the indicated menu (Figure 9 a).
time constant
c) <more> switch to the next menu group.
value: 20.000 <done> shows the end of the menu and takes you back to the
∧∧∧ ←←← more menu group.
<^^^> text: scrolls through the various selection options (Figure
9 b)
numerical values: increments the number marked by the
cursor by 1.
HV (500): 112 cps
d) HV (560): 116 cps <!!!> moves the cursor to the left and at the end of the input
HV (620): 122 cps field again to the start position (Figure 9 c).
+ - exit <+> and
<-> scrolls forward / back in the submenus (Figure 9 d).
Function Keys
LB 444
<enter> accepts the entry and moves the cursor to the next input
Density = 1.5g/cm³
field or toggles between two input fields.
+ - RUN more
<clear> clears the numerical value.
Figure 10: Live display <run> starts or ends a measurement or leads directly back to
the display and at the start of a measurement automati-
cally changes to the measurement value display.
13
Density Meter LB 444 SECTION 2. SYSTEM DESCRIPTION
↓
sk1 sk2 more
↓ ↓
General Data Oper. Mode
Password Config Instrument
Date & Time Density / TC
System/ Version Mass flow
Language Rad./Backscatter
Print Parameter Error Mode
Factory Setting Interfaces
↓
sk1 sk2 more
↓ ↓
Parameter Product Data
Productselect Product Selection
Isotope
Curr. output Limits
Time Constant
Current outp. error
Rapid Switch-Over
Relay 2 + 3 Setup
Rad. Interference
Solids density
Maximum Rate
Minimum Rate Liquid density
↓
sk1 sk2 more
↓ ↓
Calibrate Live Display
Product Selection Density display
Countrate
Data Input
HV Detector
Calculate
Massflow display
Cal. Mode Integr. M.
Resuls Flow
↓
sk1 sk2 more
↓ ↓
Service menu Massflow
Test calculation Unit
HV Setting/Status Range Input
Pt100 Adj./Plateau Flow
Adj. Current Out Counter Interval
14
Density Meter LB 444 SECTION 2. SYSTEM DESCRIPTION
The factory settings are listed in the Configuration Checklist in the Appendix.
Enter the final operating settings in this list. Please remember that the entry in all
operating levels is made with the <^^^> key and the new value is accepted with
<enter>.
General Data
Password:
One may enter a number comprising up to 6 digits. Upon confirmation with <en-
ter>, the system is protected against unauthorized manipulation of the parame-
ters. All parameters can still be viewed, however. The password protection is
canceled when you enter the correct numerical value and confirm your entry
with <enter>.
Date:
Enter the current date in the format TT.MM.YY. The correct date is important for
automatic correction of the activity decay of the source.
Time:
Enter the current time in the format HH.MM. Time deviations have hardly any
effect on the correction of the activity decay; checking the time can help to check
the instrument function.
System/Version:
Display of instrument type and software program version.
Language:
You can choose German, English or French as dialog language by pressing the
<^^^> key.
Print Parameter:
A printer with serial interface can be connected on the front panel. All parame-
ters can be printed for documentation. The printout in the appendix shows an
example. A PC may also be connected and the data transferred to the computer
and processed there using a terminal program.
Factory Setting:
All parameters can be reset to the factory setting with <sk1>. However, a meas-
urement with this data only is not possible.
15
Density Meter LB 444 SECTION 2. SYSTEM DESCRIPTION
Operating Mode
Config Instrument :
The instrument configuration is defined by the required measuring configuration
and the measuring task and can be selected by pressing <^^^> and <enter>.
a) Density without TC
Density measurement without temperature compensation
Radiation/Backscatter measure
Generally, irradiation is selected. In this case, the product is irradiated with
source and detector installed on opposite sides.
In case of a backscatter measurement, source and detector (shielded against
each other) are installed next to each other.
Error mode:
Select if the measurement is to be aborted or continued if an error is detected. If
you select “continue measurement” the error is merely displayed, but the meas-
urement continues. Press <enter> to delete the error message.
RS232 interface:
Select the baud rate by pressing <^^^> and confirm with <enter>.
Select the device to which the data is to be transferred: select either PC/Printer
or Modem.
The RS232 interface is located on the front panel of the evaluation unit.
RS 485 interface:
Interface. (Presently not used.)
16
Density Meter LB 444 SECTION 2. SYSTEM DESCRIPTION
Parameter
Isotope: Select the isotope used for the measuring configuration so that the nu-
clide-specific factor for the automatic half-life correction becomes effective.
Time constant:
Enter the time constant of the instrument system. It determines the average cal-
culation of the counts supplied by the detector. In order to reduce statistical
variations, select the highest time constant that is permissible. It is dependent
upon the permissible error during the maximum possible density rate-of –
change. A time constant of less than 20s will therefore be required only in
exceptional cases. Typical input values are between 30 and 300 s.
17
Density Meter LB 444 SECTION 2. SYSTEM DESCRIPTION
Rapid switch-over:
The function "Rapid switch-over" (=1/10 of the defined time constant) is needed
to adjust the output signal quickly to the new value in the presence of sudden
density changes. This function is enabled and disabled via <ON> and <OFF>.
Sigma defines a range (window) which the display change has to exceed before
the selected time constant automatically switches over to a value that is smaller
by a factor of 10 (max. to 0.8 sec).
When using the rapid switch-over function, enter "4" or "5" as Sigma value to
make sure that a switch-over of the time constant is not triggered merely by sta-
tistical variations or minor density changes. We recommend the determination of
optimum entry value during routine operation (see also section 6.4 Automatic
Measuring Time Switchover).
Radiation interference:
The function "Radiation interference" is needed only when unforeseeable influ-
ences due to interfering radiation are to be expected. This function is enabled
and disabled via <ON> and <OFF>. Sigma defines a limit value which the dis-
play change has to exceed before the measurement is aborted (see also section
6.3 Interference Radiation Detection). Enabling the radiation interference de-
tection disables the rapid switchover function.
Maximum Rate:
If the countrate supplied by the detector exceeds the entered value, the meas-
urement is stopped automatically. The output current signal is held at the last
value. The error message "Measurement halted" appears on the display. If the
countrate drops below the entered value again, the measurement continues
automatically.
The entered values are permanently corrected during decay compensation.
The functions „Maximum Rate“ and „Minimum Rate“ can be used to “freeze” the
measured value in case of problems (e.g. empty measuring path, empty con-
tainer, increased radiation due to weld seam testing, etc) and to avoid long re-
sponse times of the control devices. Select the input values such that that they
will not be reached during regular operation.
Minimum Rate:
If the countrate supplied by the detector falls below the entered value, the
measurement is automatically halted. If this function is not important for your
operation, you can disable it by entering a pulse rate of „0“.
Current Input:
Temperature compensation via current input. Select the current input range 0-20
mA or 4-20 mA. For mass flow measurements the range defined here is used in
the menu sk2 Volume Flow.
18
Density Meter LB 444 SECTION 2. SYSTEM DESCRIPTION
Product Data
Enter the product data for calibration separately for each product.
Product Selection
No. 1/2/3/4:
Enter the product number for calibration.
Current Output :
Select the start value 0 or 4 mA of the current output signals (26a/26c).
de-energized
energized
energized
1.6 1.6
Hysteresis
3
3
Density in g/cm
Density in g/cm
Signal
Signal
1.4 1.4
falling
rising
Hysteresis
1.2 1.2
de-energized
de-energized
energized
energized
1.0 1.0
19
Density Meter LB 444 SECTION 2. SYSTEM DESCRIPTION
Reference temperature
Enter the reference temperature for the density, concentration or mass flow
measurement. If the actual temperature deviates from the reference tempera-
ture entered here, the measured density or concentration values are tempera-
ture corrected.
20
Density Meter LB 444 SECTION 2. SYSTEM DESCRIPTION
Calibrate
Product Selection
No. 1/2/3/4:
Enter the product number for the calibration product and confirm with <enter>.
Select Unit
g/cm³, t/m³, Bx, Weig.%, Be, g/l
Select the unit of measure for density and concentration measurements. For
calculation of the calibration curve, the values are converted into g/cm³. How-
ever, the results are always displayed in the selected unit.
21
Density Meter LB 444 SECTION 2. SYSTEM DESCRIPTION
sk2: Calculate
Calibration
Calibrate mode
none/lin/squ/cub/auto
Select the calibration mode.
none: One-point calibration. For this you only need one value pair (countrate
and density or concentration value) and the absorption coefficient. For
common products it can be taken from the table in the Appendix and en-
tered at Coefficient a1 (see also section 6.2 Calibration Method).
lin: Linear curve fit . This mode is used when at least two value pairs are
available. It should also be used when several calibration points are avail-
able which are very close together, so that not the entire measuring range
is covered by samples. For more information see section 6.2 Calibration
Method.
squ Square curve fit can be selected when at least 3 calibration points are
available which are distributed fairly evenly over the measuring range or
when the measurement is carried out in %-concentration, °Bx or in an-
other unit of measure which is not in linear correlation with the density.
cub Cubic curve fit should be applied when the same conditions exist as under
squ, but at least 4 value pairs are available and it becomes apparent dur-
ing operation that in some parts the value deviate when using squ.
auto The program automatically selects the best curve fit . Square or cubic
curve fit are used only when it can be ruled out with 95% probability that
the values correspond to the curve only by chance (check by means of f-
test).
At least 4 value pairs must be available for this function.
Upon selection of the required calibration mode and pressing of the <en-
ter> button, the calibration factors are calculated.
Exception: one-point calibration
Coefficient a1
Linear absorption coefficient (range -10 to +10).
22
Density Meter LB 444 SECTION 2. SYSTEM DESCRIPTION
Zero Countrate I0
The zero countrate is automatically calculated for calibration. It corresponds to
the zero point of the unit of measure, i.e. the density or concentration value “0”.
The zero countrate is automatically corrected at midnight, compensating for
source decay. Do not change it manually.
Coefficient a1
Linear absorption coefficient. This value is automatically calculated for calibra-
tion or has to be entered for one-point calibration.
Coefficient a2
Square absorption coefficient. This value is automatically calculated for two- or
multi-point calibration.
Coefficient a3
Cubic absorption coefficient. This value is automatically calculated for multi-point
calibration.
Coefficient sq
Square error. For two- or multi-point calibration this value is calculated automati-
cally. It indicates the quality of the calibration curve for the selected calibration
mode. The smaller the numerical value, the better the curve fit (see section 6.2
Calibration Method). This coefficient is not a criterion for selection of the best
calibration mode (see calibration mode auto).
Factor
Option: You can enter a multiplication factor (0 to 10) to correct the measured
values. Each measured value is multiplied by this factor. It allows you to change
the slope of the calibration curve. When performing a new calibration, the factor
is automatically set to "1" (see also section 6.6 Correcting the Results: Addition
and Multiplication).
Offset
Option: Offset (-9999 to +9999) for additive correction of the measured values,
allowing a parallel offset of the curve. The offset entered here is added to each
measured value. When performing a new calibration, the value is automatically
set to "0" (see also section 0 Additive Constant).
Density TC
Use only for temperature compensation via Pt 100 or current input. The tem-
perature compensated lab values are displayed which were entered at rates 1.
to 10. (g/cm³). Scroll with <+> and <->.
23
Density Meter LB 444 SECTION 2. SYSTEM DESCRIPTION
Live Display
Start measurement
A measurement can be started from any menu item by pressing the <run> but-
ton (exception: Service menu). Then the measured values are displayed con-
tinuously in the selected unit for the selected product. RUN appears on the dis-
play.
With <+> and <-> you can cycle through the displays below.
Density display:
The actual density can be displayed in the defaulted range, e.g. 1.234 g/cm³.
I-mean
The average countrate is displayed. The entered basic time constant is used as
time constant.
I-actual
The actual countrate is displayed. The entered basic time constant is used as
time constant.
HV auto:
Display of adjusted high voltage (HV).
The actual concentration [%] and the mass flow [kg/h] are displayed in the de-
faulted range.
Integr. M=[t]
The integrated mass is displayed.
Flow =[m³/h]
The actual volume flow is displayed.
24
Density Meter LB 444 SECTION 2. SYSTEM DESCRIPTION
Service Menu
Mass Flow
Flow
Displays the flow in m³/h.
Counter Interval
Enter a volume per hour that is used as counting unit. Every time this value is
reached relay 1 adds one unit.
25
Density Meter LB 444 SECTION 3. INSTALLATION
3. INSTALLATION
The shielding with the radioactive source is delivered in a box in compliance with
the regulations concerning the transportation of radioactive devices.
Take the shielding out of the box just prior to installation. Up to that time,
store it in a location that is guarded against unauthorized access.
Using the drawings of the shielding and taking into account the situation at the
measuring site, carefully install the mounting brackets and fixtures. Make sure
that the mechanical rigidity of the mounting hardware matches the weight of the
shielding. The shielding should be mounted just prior to taking the system into
operation.
Tighten all screws and fixing parts, so that they cannot come undone dur-
ing operation.
26
Density Meter LB 444 SECTION 3. INSTALLATION
3.2 Installation
Installation on Pipelines
When selecting the installation site, please keep the following criteria in mind:
a) Selection of measuring site. At the measuring point, the pipeline must al-
ways be completely filled with the product being measured. If the pipeline is
only partially filled, this may result in incorrect measurements.
c) Gas bubbles in the product falsify the results. This error can be avoided or
at least reduced by installing the measurement system at a point on the
pipeline where the pressure is fairly high (e.g. installation in pressure pipes,
at the foot of a standpipe).
Gas bubbles
RADIOACTIVE
Deposits
27
Density Meter LB 444 SECTION 3. INSTALLATION
RADIOACTIVE
28
Density Meter LB 444 SECTION 3. INSTALLATION
RADIOACTIVE
29
Density Meter LB 444 SECTION 3. INSTALLATION
Cover sheet
RADIOACTIVE
Ensure that no persons can insert any part of the human body in the radia-
tion beam!
k) The pipeline must not be subject to excessive vibration as this could damage
the detector. If this cannot be avoided, the shielding container and the de-
tector must be mechanically isolated from the pipe or source of vibration.
(See Figure 15).
l) Since the calibration of the measurement requires that samples of the prod-
uct be taken, a sampling point should be provided near the installation.
30
Density Meter LB 444 SECTION 3. INSTALLATION
Brace
The installation has to be done such that the radiation path lies diagonally to the
flow direction. With horizontal flow the radiation exit channel should point toward
the bottom to prevent deposit formation in the path of radiation.
31
Density Meter LB 444 SECTION 3. INSTALLATION
a) Selection of measuring point. At the measuring point, the pipeline must al-
ways be completely filled with the product to be measured. If the pipeline is
only partially filled, this may result in incorrect measurements.
b) For outdoor installation, the measuring system must be protected from rain
and direct sunshine (e.g. by a thin sheet metal).
c) Do not expose the cables to temperatures exceeding 70°C.
d) Protect the detector against temperatures > 50°C by a water-cooling
jacket.
e) Prevent water entry via cable bushings.
f) Avoid vibration at the detector.
g) The product temperature for temperature compensation must be measured
directly next to the density measurement.
32
Density Meter LB 444 SECTION 3. INSTALLATION
The water must enter from the bottom and exit at the top. Thus, the cooling
jacket is always filled with water.
Water cooling
Inlet
33
Density Meter LB 444 SECTION 3. INSTALLATION
Cooling Medium
Water should preferably be used as the cooling medium, because air does not
ensure adequate cooling with high temperatures. The water must be clean to
avoid dirt deposits in the cooling jacket.
The enclosed diagram shows the quantity of cooling water required for the de-
tectors LB 44... / LB 54.... The ambient temperature was taken into considera-
tion when calculating the required quantities, but not heat emission by the sur-
rounding components, e.g. containers, pipelines, etc.
220
40°C
200
150
30°C
100
20°C
50 10°C
Flow rate l/h
10
34
Density Meter LB 444 SECTION 3. INSTALLATION
This type of installation requires that the pipeline including the resistance ther-
mometer must be provided with a temperature insulation over a length of 1 - 2
m, ensuring that the surface temperature of the pipeline at the temperature
measuring point is practically equal to the product temperature. Nevertheless, it
may happen, particularly with plastic or coated pipelines, that very sudden tem-
perature changes in the product to be measured lead to measuring errors
caused time lag. Correction is possible by operating the system with a large time
constant.
Before connecting the resistance thermometer to the detector, a Pt 100 adjust-
ment has to be performed in the Service menu (see section 5 GETTING
STARTED) .
a) According to the Ex protection criteria, the evaluation unit must only be used
outside the Ex-area. An equi-potential bonding must be established between
detector and evaluation unit.
b) Please note the maximum cable lengths between evaluation unit and detec-
tor. See section 7 TECHNICAL DATA.
35
Density Meter LB 444 SECTION 4. CONNECTIONS
4. CONNECTIONS
4.1 Detector
Switch off the evaluation unit before connecting the scintillation counter,
Open the screws and remove the connection box cover to expose the connec-
tion area.
If a resistance thermometer is connected, the cables coming from the Pt 100 are
passed through the second cable bushing to terminals 3 and 4.
For equipment installed in the Ex-area, please observe the special re-
quirements for the cable and the preparation of the connection wires.
The detector LB 44... must be used for installation in the Ex-area. The de-
tector LB 54.. may only be used for non Ex-applications.
The evaluation unit must always be installed outside the Ex-area.
36
Density Meter LB 444 SECTION 4. CONNECTIONS
Detector
LB 444 2(-) 1(+)
c a
2
4
6
8
10
12 Relay 2
14 Relay 3
16 Relay 1
18 Dig. In 1
20 Dig. In 2
22 Dig. In 3
Voltage 24
selector 26 Current Out Density (-/+)
28 Current In (+/-)
30
Fuse 32
Mains supply
Carefully note the power supply for the instrument and observe all safety
provisions regarding the power supply. Separate fuse protection and an
easy to access power switch must be provided as part of the installation,
since the evaluation unit does not include its own mains switch.
Refer to the wiring diagram in the appendix to this manual for the connections.
The terminals are described as follows:
Connection is made via 2-wire technique; the detector protection type is EEx de
IIC T6 or EEx de (ib) IIC T6 or EEx de (ib) IIB T6. For inherently safe installation,
the cable ends on the strip terminal must be protected by a 10 mm long shrink
tube.
The relay can be used as Max. or Min. relay depending on the software configu-
ration and the setting of the break-over point.
The relay can be used as Max. or Min. relay depending on the software configu-
ration and the setting of the break-over point.
37
Density Meter LB 444 SECTION 4. CONNECTIONS
0/4-20mA (26a+/26c-)
RS 232
Terminal for data transfer from evaluation units to a printer or PC (front panel of
evaluation unit).
Power supply 115V/230V, depending on power supply unit (see label on instru-
ment rear panel!).
Fuses
To replace fuses, you have to open the turn-lock using a coin. Note the fuse
type and power!
Before turning on the power supply, carefully check all connections once
more to prevent damage to the instruments.
38
Density Meter LB 444 SECTION 5. GETTING STARTED
5. GETTING STARTED
7. Only for mass flow measurement: The volume current signal can be
fed in via terminals 28a+ and 28c-. Temperature compensation is 40
possible only via Pt 100 connected to the detector.
39
Density Meter LB 444 SECTION 5. GETTING STARTED
Getting Started
Turn on the power supply of the instrument. Then the manufacturer’s name,
versions number and instrument type is displayed (Figure 8)
• Insert key in cylinder lock and turn it to the right to release the catch of lock-
ing handle.
Lock
Protection
cap
ZU OFFEN
CLOSED OPEN
Locking handle
• Remove key and arrest locking using the bolt provided. Note: the shielding
MAY NOT BE LOCKED IN THE OPEN CONDITION!
40
Density Meter LB 444 SECTION 5. GETTING STARTED
• Open padlock and turn protective cap until it can be taken off.
• Turn counter nut completely back toward the screw head. Firmly tighten the
knurled nut.
• Firmly turn counter nut toward the front, in the direction of the protection
pipe and tighten it firmly.
Knurled nut
CLOSED
Counter nut
OPEN
41
Density Meter LB 444 SECTION 5. GETTING STARTED
• Open padlock and turn protective cap until it can be taken off. Open fasten-
ing screw.
• Pull out spring pin and push source-locking lever toward the front in position
“OPEN” until the spring pin engages.
Spring pin
Locking screw
CLOSED
OPEN
42
Density Meter LB 444 SECTION 5. GETTING STARTED
The Pt 100 input is calibrated in the factory. A new calibration is required only
when the detector boards or the detector microprocessor are replaced.
1. Apply 100 Ω (with lowest tolerance) to Pt 100 input at the detector (terminals
3 and 4, Figure 21).
2. Press <enter> until the menu Service Menu/Mass Flow is displayed.
3. Select Service menu with <sk1>.
4. Press <more> twice and then select Pt 100 calibration with <sk2>.
nd
5. Press <sk2> (2 softkey) and <clear> at the same time. When the display
„Pt 100 calibration running...“ appears, press <more>.
6. Then connect resistance thermometer to detector, terminal 3 and 4, (see
Figure 21).
Basic Settings
These steps describe the procedure for taking the system into operation
and changing the system setting.
Note: To enter numbers, select the entry position with <←←←> and the number
you want with <^^^>. Confirm your selection or new entry with <enter>.
Press <run> to return to the display mode directly any time (except in the Serv-
ice menu).
1. Select menu General Data / Operating Mode via the key <more>.
1.1 Select General Data menu with <sk1> and with <more> call the parame-
ters sequentially.
• Enter password and unlock or lock the system with <enter>
• Check date & time and correct it, if necessary
• Check instrument version (e.g. LB 444, version 1.0)
• Select language
• Print Parameter, only if needed
Factory Setting, only if needed
1.2 Select Operating Mode menu with <sk2> and the operating modes with
<^^^>.
• Select Config Instrument with <^^^> and confirm with <enter>:
Density measurement with or without TC (Pt 100 or current input)
Mass flow measurement with or without TC (Pt 100)
• Select radiation measure to irradiate the product. Select backscat-
ter measure if source and detector are installed parallel.
• Define Error mode
• Select parameters for RS232 interface
Return to the menu group with <done>.
43
Density Meter LB 444 SECTION 5. GETTING STARTED
2.1 Select Parameter menu with <sk1> and with <more> call the parameters
sequentially.
• Select product
• Select detector and isotope
• Measuring path in product (e.g. pipe diameter) in cm
• Select Measure mode (i.e. continuous)
• Enter time constant (e.g. 60 s)
• Disable or enable rapid switch-over
• Disable or enable Interference Radiation
• Default max. countrate (e.g. 100000 I/s)
• Default min. countrate (e.g. 0 I/s)
• For TC via current input:
Select current input range 0-20mA or 4-20mA
Enter temperature value for 0/4mA and 20mA
Return to the menu group with <done>.
2.2 Select Product Data menu with <sk2> and with <more> call the parameters
sequentially
• Select product
• Select current output 0 –20 or 4 – 20 mA
• Define current output limit values 0/4 and 20 mA
• Define current output error message
• Define Relay 2: Function (Min)
3
• Switchpoint (e.g. 1.2 g/cm )
• Hysteresis (e.g. 5 %)
• Define Relay 3: Function (Max)
• Switchpoint (e.g. 1.6 g/cm³)
• Hysteresis (e.g. 5 %)
• Only for suspension measurements
Enter solids density in g/cm³ for the selected product
Enter liquid density in g/cm³
Activate Water TC when solution is to be measured with water as
carrier liquid.
• Enter reference temperature
44
Density Meter LB 444 SECTION 5. GETTING STARTED
45
Density Meter LB 444 SECTION 5. GETTING STARTED
Calibration
One-point calibration is sufficient for preliminary adjustment of the measurement
ranges. Two- or multi-point calibrations should be performed for accurate and fi-
nal adjustment. Different procedures are required for each calibration method.
For more information on the calibration methods see section 6.2 Calibration
Method.
Please read section 6.2 Calibration Method in any case before performing
calibration!
46
Density Meter LB 444 SECTION 5. GETTING STARTED
47
Density Meter LB 444 SECTION 5. GETTING STARTED
48
Density Meter LB 444 SECTION 5. GETTING STARTED
5.2 Measurement
49
Density Meter LB 444 SECTION 5. GETTING STARTED
c) All errors are signaled via the “Error” output (Relay 16a/16c).
50
Density Meter LB 444 SECTION 5. GETTING STARTED
*) When these error messages are displayed, the measurement is stopped automatically
and must be restarted by pressing the <run> button
51
Density Meter LB 444 SECTION 5. GETTING STARTED
52
Density Meter LB 444 SECTION 6. BASICS OF DENSITY MEASUREMENTS
The data for temperature compensation (TC1, TC2) and the reference tem-
perature have to be entered before starting calibration. The temperature
compensation is then turned on automatically.
Temperature Measurement
The product temperature can be measured either
a) with the resistance thermometer Pt 100 (see also section 3.4 Installation of
Resistance Thermometer Pt 100)
or
b) by entering the product temperature as current signal 0 - 20 mA.
53
Density Meter LB 444 SECTION 6. BASICS OF DENSITY MEASUREMENTS
(The terms in Italics refer to the parameters that have to be entered in the Prod-
uct Data menu).
Example:
Product: HCl - water mixture
Measuring range: 20 - 40 weight % HCl
Temperature range: 10° - 30° C
3
Unit of measure: g/cm
ρ1 - ρ 2 1.1551 - 1.1433
TC = - = - = 0.00059
ζ1 - ζ 2 10 − 30
3
ρ1 = Density in g/cm at ζ1
3
ρ2 = Density in g/cm at ζ2
54
Density Meter LB 444 SECTION 6. BASICS OF DENSITY MEASUREMENTS
For other units of measure (e.g. °Bx, %-concentration, g/l, etc.), the temperature
coefficient has to be converted accordingly. To do this, one first has to calculate
the density change/unit of measure obtained for the measuring range (e.g. %):
Density change ∆ρ ρ C1 − ρ C2
= =
Concentration change ∆C C1 − C 2
Example:
3
ρC1 at 20° C and 20 % HCl = 1.0979 g/cm
3
ρC2 at 20° C and 30 % HCl = 1.1493 g/cm
∆ρ 1.0979 - 1.1493 g
= = 0.00514
∆C 20 % - 30 % 3
cm • %
TC1 0.00059
TC' = = = 0.11478
∆ρ 0.00514
∆C
55
Density Meter LB 444 SECTION 6. BASICS OF DENSITY MEASUREMENTS
If significant temperature variations are likely to occur in the product (approx. > ±
20°C), it is advisable to enter the square temperature coefficient as well.
Calculation:
ρ2 = ρ1 + (ζ1 - ζ2 ) · TC1
'
d) Calculate TC2 :
ρ2 - ρ2
'
TC 2 = -
(ζ1 - ζ2 )
2
Example:
3
Ethanol, concentration 30 %, reading in g/cm , relative to 10°C.
st
1 temperature coefficient in range 10 - 20° C:
0.9599 - 0.9540
TC1 = - = 0.59 ⋅ 10-3
10 − 20
-3
Input value in code XX47: 0.59 * 10
ρ2
' -3
= 0.9599 + (10 - 100) · 0.59 · 10 = 0.9068
0.9068 - 0.8936
TC 2 = - = 1.6296 ⋅ 10 -6
(10 - 100)
2
-6
Input in code XX48: 1.6296 * 10
56
Density Meter LB 444 SECTION 6. BASICS OF DENSITY MEASUREMENTS
Reference Temperature
a) With linear temperature coefficient:
If you use only TC1, enter the average product temperature as reference
temperature, rather than the Lab temperature of, say, 20°C.
Add a correction value (ρP - ρRef) · TC1 to the measured value.
A slightly incorrect value entered as TC1 is the more significant, the greater
the difference ρP - ρRef. If one correlates the countrates entered at operating
temperature with the density values obtained at constant lab temperature,
the reading will always be equal to the density value calculated at lab tem-
perature.
a) Fill the measuring path with the product to be measured. The density should
be about the average value of the requested measuring range.
b) Disable temperature compensation (instrument configuration without TC).
c) Start measurement and wait for a short time until the reading shows the
value of the product in the measuring path.
d) Write down the density or concentration value ρ1.
e) Read off product temperature ζ1, write it down and enter it in reference tem-
perature.
f) Change the temperature of the product in the measuring path through heat-
ing or cooling by approx. 10 to 15°C.
g) Write down this density or concentration value ρ2, and also the associated
temperature ζ2.
h) Calculate the temperature coefficient as follows:
ρ1 - ρ 2
TC1 = -
ζ1 - ζ 2
i) Enter this temperature coefficient in TC1. TC2 must contain "0".
j) Enable temperature compensation in the Config Instrument / Operating Mode
menu. The same density or concentration value as noted under d) must now
be available again.
57
Density Meter LB 444 SECTION 6. BASICS OF DENSITY MEASUREMENTS
a) To calculate the square coefficients, one first has to calculate the linear coef-
ficient, starting from the reference temperature (here: lowest temperature), as
described above. The initial temperature should match the lowest product
temperature.
c) From the resulting change in reading and the associated temperature change
one then calculates the square coefficient TC2:
ρA - ρD
TC 2 = -
(ζ1 - ζ2 )
2
d) Enter the calculated value at TC2. Now the originally noted density or con-
centration value must be obtained again.
58
Density Meter LB 444 SECTION 6. BASICS OF DENSITY MEASUREMENTS
The counts supplied by the detector are read into memory at various product
densities or concentrations. At the same time the product temperature is auto-
matically stored. The density or concentration values (analysis values) deter-
mined in the lab are entered in another memory. The diagram below illustrates
the correlation of memories.
The calibration points should be distributed fairly evenly over the entire
measuring range. Some calibration points may lie outside the measuring range
(e.g. water). However, calibration with all point lying outside the measuring
range is not advisable.
Calibration Modes
The calibration mode has to be entered in the Calibrate menu, submenu Calcu-
late. Five calibration modes are available.
or
59
Density Meter LB 444 SECTION 6. BASICS OF DENSITY MEASUREMENTS
Select lin
or
• sampling can only be done very inaccurately, so that the samples are
subject to errors.
c) Calibrate mode 3: Curve fit with linear and square coefficients (squ) can
be selected when
• the entire measuring range has been covered evenly by several (at
least 3) calibration points.
• the measurement is run in % concentration, °Bx or another unit which
is not in linear proportion to the density.
d) Calibrate mode 4: Curve fit with linear, square and cubic coefficients
(cub) can be selected
• when the same conditions exist as under b), but at least 4 value pairs
are available and the actual measurement shows that with calibration
mode 3 deviations occur in parts of the measuring range.
60
Density Meter LB 444 SECTION 6. BASICS OF DENSITY MEASUREMENTS
One-Point Calibration
For one-point calibration one needs a value pair (countrate and density or con-
centration value) and the absorption coefficient. The absorption coefficient for
common products is listed in the enclosed table (see Appendix). For products
3
not listed there, enter the following values for the unit g/cm .
Cs-137:
Absorption coefficient µ
- 0.0664 - 0.06 -0.057
Co-60:
∆ρ ρ - ρmin
µx = µ ⋅ ⋅ max
∆C C ρmax - C min
Example 1:
Isotope: Cs-137
Crystal: 40/35; frontal irradiation
ρmax = 1.098: Cρmax = 20 %
ρmin = 1.045: Cρmin = 10 %
Cρ = Concentration in % at respective density value
1.098 - 1.045
µ% = - 0.0664 = - 0.000352
20 − 10
Example 2:
61
Density Meter LB 444 SECTION 6. BASICS OF DENSITY MEASUREMENTS
0.95 - 0.91
µ% = - 0.0664 = - 0.000133
30 − 50
62
Density Meter LB 444 SECTION 6. BASICS OF DENSITY MEASUREMENTS
c a
ln I b
Density/Concentration
The curve obtained with one-point calibration passes through the calibration
point. The gradient of the curve is determined by the entered absorption coeffi-
cient and may therefore be somewhat inaccurate.
One-point calibration is also advisable if only calibration points with very minor
density differences can be recorded, for in this case minor errors in the lab or
during sampling will significantly affect the curve.
a
a
cps
cps
3 3
Density in g/cm Density in g/cm
63
Density Meter LB 444 SECTION 6. BASICS OF DENSITY MEASUREMENTS
For multi-point calibration, the respective values should be entered in the Prod-
uct Data menu at the queries Current Output Limit Values (see also page 19).
Current output 26a/26c.
6.000
7.000
cps
8.000
9.000
10.000
11.000
0,9 1 1,1 1,2 1,3 1,4
Density in g/ccm
You need not observe a rising or falling order when entering the sample values.
However, you must observe the proper correlation countrate and density value.
Remedy:
The entered data can be corrected as needed since the calibration procedure is
performed only after selection of the calibration mode.
64
Density Meter LB 444 SECTION 6. BASICS OF DENSITY MEASUREMENTS
The countrate read into the Data input menu is dependent upon the density of
the product during reading-in. In the laboratory the density is usually determined
at constant temperatures. With constant concentrations but different tempera-
tures during reading in of the pulses one may get:
• the same lab values
• different density values and therefore
• different countrates.
To avoid calibration errors, the product temperature is stored as well. When cal-
culating the calibration curve, the entered laboratory value is corrected by
means of the temperature coefficient. This correction is carried out only when
• temperature compensation has been enabled in Config. Instrument
• a temperature coefficient has been entered.
The corrected lab values from which the calibration curve is calculated are dis-
played after the coefficients and the square error.
The following two points are relevant only when measuring the density. For con-
centration measurements the temperature compensation is always enabled prior
to performing calibration.
65
Density Meter LB 444 SECTION 6. BASICS OF DENSITY MEASUREMENTS
ρM ζP ζM ρ
C
in g/cm
3 in °C in °C 3
in g/cm
1.30 75 70 1.2975
1.32 70 70 1.3200
1.35 65 70 1.3750
Since in this case the correction has already been done manually, the tempera-
ture compensation must be turned off and then on again in the instrument con-
figuration.
66
Density Meter LB 444 SECTION 6. BASICS OF DENSITY MEASUREMENTS
Display Value
Coefficient a2 0
Coefficient a3 0
Factor 1
Offset 0
Square error 0
Display Value
Coefficient a2 0
Coefficient a3 0
Factor 1
Offset 0
Square error 0
If other values should be displayed, check your calibration or the data you have
entered.
67
Density Meter LB 444 SECTION 6. BASICS OF DENSITY MEASUREMENTS
With larger values you should check your calibration particularly with re-
gard to incorrect entries.
68
Density Meter LB 444 SECTION 6. BASICS OF DENSITY MEASUREMENTS
Select the function Radiation Interference (Parameter menu, page 18) to sup-
press interfering radiation (e.g. weld seam testing).
Caution:
When this function is on (Interference Radiation Detection <On>)
• the automatic switchover of the time constant (function Rapid switch-over) is
always off.
When turning this function on, you have to define a Sigma value which defines a
window for the measuring signal. If the detector signal suddenly exceeds this
threshold,
• the measurement is aborted
• the error message "Interference Radiation" appears on the display.
The measured value and the current outputs are held on the last value.
Reset the error message and start the measurement via the keyboard by ac-
knowledging the error message with <enter> and start the measurement with
<run> or externally via the digital input.
Note:
To rule out false alarms with sufficient statistical safety, you should enter n > 5.
The mathematical correlation shows that the distance from the alarm threshold
is dependent upon the respective mean pulse rate Im.
For calculation it holds: Sigma = I/s
Example:
IS = Im + n * Im
Note:
Due to the dynamic behavior of the interfering radiation detection, a quick in-
crease of the pulse rate due to operative factors (e.g. very fast emptying of the
vessel or major short-term density changes caused by stirrers) can be inter-
preted as interfering radiation.
For example, opening the active beam channel on the shielding can also cause
a quick increase of the pulse rate. You have to reset the alarm that is then trig-
gered, or better, do not enable the interfering radiation detection at first. Enable
the interfering radiation detection only after the calibration has been performed.
69
Density Meter LB 444 SECTION 6. BASICS OF DENSITY MEASUREMENTS
If a sudden density change occurs which is higher than the entered threshold
(Sigma), the entered time constant is reduced to 1/10 of the original value, but
not less than 0.8 s. Thus, the output signal adjusts much quicker to the new
density value. The changed time response of the output signal is illustrated in
the diagram below.
Response
100
Density
Display and output signal in %
80 Output signal
with fast
response
40
20
0
-3 2 7 12 17 22 27 32 37 42 47 52 57 62 67 72 77 82 87 92 97
T in s
When the rapid switchover is in effect, the measurement first works with the de-
fined basic time constant. If sudden density changes occur which exceed the
Sigma band, the system switches to the 10-times lower time constant and the
reading and the output signal can follow the density change much quicker. At
the same time, the statistical variations increase by a factor of about 3. When
the density value becomes stable again, the system automatically switches back
to the longer basic time constant after a period of time that corresponds to the
basic time constant.
70
Density Meter LB 444 SECTION 6. BASICS OF DENSITY MEASUREMENTS
Measurement
Density or concentration measurements of suspensions can be carried out in
these units:
3
a) Density in g/cm
b) Concentration in g/l
c) Concentration in weight-%
d) °Be
Data input, calibration and display are done in the selected unit of measure. For
calculation of the calibration curve, the values entered under b) through d) will
3
be converted into g/cm .
Concentration in g/l:
S S
ρ = (1 − ) * ρL +
ρ * 1000 1000
s
ρ
3
= Density of suspension in g/cm
S = Solids content in g/l
ρS
3
= Solids density in g/cm
ρL
3
= Liquid density in g/cm
Concentration in %:
ρL
ρ=
C * ρL
,
+ 1 − C,
ρS
o
Concentration in Be
144,3 144,3
ρ= ρ= n = °Be
144,3 − n 144,3 + n
Strictly speaking, the formulae for g/l and concentration in % apply only to suspensions. For solutions or
liquid mixtures, large deviations may occur when the measurement involves larger ranges.
The diagram on the next page shows how to convert units of measure in the
density measuring system:
71
Calibration
Measurement
Density Meter LB 444
Calibration
°Be °Be
or or
Conversion Conversion
g/l into in selected g/l
g/cm 3 unit
72
or or
% %
or or
Calculate Measured
3 Temperature values in Temperature 3
g/cm calibration g/cm
correction g/cm3 correction
curve
SECTION 6. BASICS OF DENSITY MEASUREMENTS
Density Meter LB 444 SECTION 6. BASICS OF DENSITY MEASUREMENTS
With suspensions the solids density is usually known and water is used as car-
rier liquid (select Water TC and enter the solids density); with solutions, on the
other hand, the density of the relevant components often has to be calculated
from table values.
Table 1 in the Appendix lists the values for common products. If your product is
not included, just use the formula below for calculation.
Using the values listed in the table, you can calculate the density of the compo-
nents. Enter the density of the carrier liquid or the density of the attendant com-
ponent (mostly water) at average temperature (reference temperature) as liquid
density.
C, * ρL
ρS =
ρL
− 1 + C,
ρ
Example:
0.2 * 0.99823
ρS = = 18294
.
0.99823
− 1 + 0.2
10980
.
73
Density Meter LB 444 SECTION 6. BASICS OF DENSITY MEASUREMENTS
ρL ⋅ S
ρS =
ρL + S − ρ
Example:
0.99823 * 0.2196
ρS = = 18294
. g / cm 3
0.99823 + 0.2196 − 10980
.
74
Density Meter LB 444 SECTION 6. BASICS OF DENSITY MEASUREMENTS
Additive Constant
The value stored under Offset is added to the density value calculated on the
basis of the countrate. This allows a parallel shift of the calibration curve.
Example:
3
Measuring range 1.1 - 1.3 g/cm
3
It is found that the measuring values are too low by 0.05 g/cm .
Remedy:
Enter “0.05” at Offset.
3
All measured values will be increased by 0.05 g/cm .
3 3 3
Instead of 1.1 g/cm , 1.15 g/cm is now indicated and instead of 1.3 g/cm , 1.35
3 3
g/cm . If the measured values would be too high by 0.05 g/cm , you would have
to enter “-0.05”.
6.6.2 Multiplication Factor
Each measuring value is multiplied by the value stored as Factor. This allows
you to change the gradient of the calibration curve.
Example:
3
Measuring range 1.1 - 1.3 g/cm
3 3
If you enter 1.1 at Factor, 1.21 g/cm will be displayed instead of 1.1 g/cm . In-
3 3
stead of 1.3 g/cm the value 1.43 g/cm is now displayed. Thus even the low
point of the measuring range is increased, provided it is not at 0.
Remedy:
This can be corrected by entering the appropriate data at Offset at Factor.
Example:
3 3
For 1.1 g/cm the indicated value is correct; instead of 1.2 g/cm the reading
3
should show 1.25 g/cm .
75
Density Meter LB 444 SECTION 6. BASICS OF DENSITY MEASUREMENTS
The same is true if the upper and lower point of the measuring range is to be
changed by different values.
Example:
3
Lactual = 1.12 g/cm
3
Lnominal = 1.15 g/cm
3
Hactual = 1.25 g/cm
3
Hnominal = 1.3 g/cm
1.30 g / cm 3 - 1.15 g / cm 3
F = = 1.1538
1.25 g / cm 3 - 1.12 g / cm 3
3 3
K = 1.15 g/cm – 1.12 g/cm · 1.1538 = 0.1423
76
Density Meter LB 444 SECTION 7. TECHNICAL DATA
7. TECHNICAL DATA
Weight: approx. 2 kg
Time constant: 0,5 – 9999 s with automatic reduction to 1/10 of the values in
case of sudden density changes (can be turned off)
Adjustable response threshold for time constant switchover.
Automatic decay
241 137 60 244 85 90
compensation: For Am, Cs, Co, Cm, Kr and Sr .
77
Density Meter LB 444 SECTION 7. TECHNICAL DATA
Inputs/Outputs:
Detector terminal: EEx iB IIB
(2a/2c) For inherently safe installation, the cable ends on the strip
terminal must be protected by a 10 mm long shrink plastic
tubing (see also the connection diagram)
3 digital inputs:
2 Dig. In. for external product selection (18a/18c and
20a/20c)
1 Dig. In. for external Start/Stop of the measurement:
Start/Stop in batch mode
Stop of continuous measurement
Restart after interruption of measurement due to interfer-
ing radiation.
Limit value outputs: 2 relay outputs for max. / min. (12a/12c and 14a/14c)
1 relay output for collective failure message (16a/16c)
Loading capacity: 30 V AC, 1 A
78
Density Meter LB 444 SECTION 7. TECHNICAL DATA
7.2 Detector
The probe is designed for industrial application. It includes a scintillation counter
for Gamma radiation, which is equipped with a NaI crystal. The built-in micro-
processor tracks the configuration, counts the pulse rates, controls the high volt-
age, measures the temperature of the probe and an external Pt 100, transfers
data and performs calibration and control functions. Power supply and commu-
nication take place via 2-wire technology. The probe is ex-protected and is li-
censed for:
Protection type EEx de IIc T6 / EEx de [ib] II C / IP 65, stainless steel housing.
Cable entrance: PG 7 for external cable diameters 4 .. 6.5 mm.
Temperature
measurement probe: via NTC, measuring range -30 to +80°C
Accuracy: ±3°C
Temperature range:
Operating temperature:
-30 to +50°C. A water-cooling device is available for higher tem-
peratures.
Monitoring of detector temperature and alarm when exceeding the
max. permissible temperature.
Storage temperature:
-50 to +70°C.
79
Density Meter LB 444 SECTION 7. TECHNICAL DATA
Power Supply
Input voltage: +11 to 16 V
Permissible
cable resistance: 29 Ohm (for wire and return wire)
Data Transfer
Transfer procedure: ASK (Amplitude-Shift-Keying = high-frequency carrier turned
on and off by data bits)
Detector List
80
Density Meter LB 444 SECTION 8. Service Instructions
8. Service Instructions
81
Density Meter LB 444 SECTION 9. RADIATION PROTECTION
9. RADIATION PROTECTION
The radioactive sources used for density measurements are sealed sources, i.e.
the actual radioactive substance is surrounded by at least one, often several
sealed layers made of stainless steel, each of which is checked individually for
leaks. Another check ensures that no radioactive particles are deposited on the
surface of the capsule. The user receives a certificate specifying these features
of the radioactive source.
These may stipulate that access to the vessel shall only be permitted after the
active beam is shielded. Radiation protection zones outside the shielding must
be - if they are accessible - marked and guarded.
These instructions should also include checks of the shutter device of the
shielding and measures for serious operational trouble - such as fire or explo-
sion.
Any special event has to be reported to the Radiation Safety Officer imme-
diately. He/she will immediately take suitable precautions if he/she detects de-
fects that may adversely affect the operation or safety of the system.
82
Density Meter LB 444 SECTION 9. RADIATION PROTECTION
The Radiation Safety Officer must ensure that the provisions of the Radiation
Protection Regulations are observed. In particular, his duties include instructing
the staff on the proper handling of radioactive substances.
Radioactive sources that are no longer in use or have reached the end of
their service life must be returned to the manufacturer.
Generally, every member of staff should endeavor to minimize any radiation ex-
posure - even within the permissible limits - by careful and responsible action
and by observing certain safety standards.
The total sum of the radiation dose absorbed by a body is dependent upon three
factors. On the basis of these factors, certain fundamental radiation protection
rules can be derived:
Distance
This means the distance between the radioactive source
and the human body. The radiation intensity (doserate)
decreases - like light - in proportion to the square of the
distance, i.e., doubling the distance to the source reduces
DISTANCE the doserate to one quarter.
" Conclusion:
When handling radioactive substances, maximum distance
to the source should be maintained. This is especially true
for persons that are not directly involved in this work.
Time
The total time a person stays in the vicinity of a radiomet-
ric measuring system and the body is exposed to radia-
tion. The effect is cumulative and increases therefore with
the duration of the radiation exposure.
#
TIME
$ Conclusion:
Any work in the vicinity of radiometric measuring system
has to be prepared carefully and organized such that it
can be carried out in the shortest time possible. Having
the proper tools handy is of particular importance.
83
Density Meter LB 444 SECTION 9. RADIATION PROTECTION
Shielding
The shielding material surrounding the source pro-
vides the shielding effect. As the shielding effect de-
pends, following an exponential function, on the prod-
uct of thickness multiplied by the density, it follows
SHIELDING that material with a high specific weight will normally
be used for shielding. The supplier usually calculates
% suitable dimensions for the shielding.
Conclusion:
Before installing or dismantling the shielding, make
sure that the radiation exit channel is locked-closed.
The source must not be removed from the shielding.
84
Density Meter LB 444 SECTION 9. RADIATION PROTECTION
Safety Instructions
Shielding Installation
ONLY persons specifically licensed by the US Nuclear Regulatory Commission
or an Agreement State may install, move, repair or relocate a device or shielding
containing a radioactive source.
The work must be performed according to the instructions and under the super-
vision of the Radiation Safety Officer.
& the lock of the shielding is closed and secured, so that no unshielded radia-
tion can exit
& the shielding is not modified or damaged
& the function of the locking mechanism is checked following installation of
the shielding
& the function check is repeated every six months
& a new source number has to be attached on the shielding or the type label
replaced when replacing a source.
Preparation
All necessary work has to be prepared such that it can be carried out quickly, so
that exposure to the unshielded source is kept to a minimum. Using a drawing of
the shielding, you should plan the best procedure and have the following tools
handy: You must wear a personal dosimeter.
85
Density Meter LB 444 SECTION 9. RADIATION PROTECTION
Depending on the construction, you either have to open the lock on the shielding
and turn the lever to center position between ON and OFF until the hexagon
head bolt of the source holders becomes visible, or remove the locking plate, so
that you can unscrew the source holder.
Source Replacement
Proceed as follows for standard point source shieldings (Figure 30):
1. Open the lock of the shielding (1) halfway, so that you can unscrew the
source holder (2) together with the source (3) using a socket wrench.
2. You may touch the source holder at the far end of the threaded part with your
hands. Hold it far away from your body and put it down behind an auxiliary
shielding.
3. Unscrew the source from the source holder using a socket wrench (SW 10).
Hold the source holder using a second socket wrench (SW 12). For this work,
you should use the shielding housing as auxiliary shielding between source
and body.
3
2
1
4. Take hold of the source using a pair of pliers and put it into the transport
shielding or another shielding immediately.
Do not touch the source to prevent a high partial body dose. Make sure the
source is not mixed up with the new or another source.
86
Density Meter LB 444 SECTION 9. RADIATION PROTECTION
5. If necessary, clean and grease the threading on the source holder and the
shielding.
6. Using a pair of pliers, take the new source out of the transport shielding and
firmly fix it onto the source holder together with the locking washer.
7. Put the source holder with the source again into the shielding and fix it using
the socket wrench.
8. Check the proper OPEN/CLOSED function. Survey and wipe-test the shield-
ing. Test shutter operation. Lock in the closed position. Update the labels and
complete the records.
9. Carefully survey the transport shielding again, after you have placed the old
source into the transport shielding. The special regulations regarding labeling
and transport of the shielding back to the manufacturer must be observed.
The expected radiation exposure can be calculated quite easily and with suffi-
cient accuracy, provided you know the isotope and the activity of the source
used. You can take this information from the source documentation or from the
type label on the shielding.
A ⋅ k ⋅ t
Dose D =
r2 ⋅ s
A is the activity of the source and k the respective specific Gamma radiation
constant (see the table below). The distance from the measuring point to the
source is r and the duration of stay at this point is t. s is the shielding factor of
the shielding used; it is listed in the shielding brochure or can be calculated. s =
1 when calculating the doserate for work with an unshielded source.
NUCLIDE k DIMENSIONS
2
Co-60 0.35 µSv * m
Examples of calculations:
87
Density Meter LB 444 SECTION 9. RADIATION PROTECTION
This calculation can be simplified, and the accuracy is still OK, by using the
given doserate value in 1 meter distance from the shielding.
Shieldings for measuring systems are usually designed such that, regardless of
the activity and the type of source (point or rod source), the limit of the controlled
area with 7.5 µSv/h is in a distance of max. 1 meter around the shielding. This
value can be taken for granted when the mean distance for mounting the
shielding or operating the lock is 0.5 m. As a result, we get a doserate of max.
30 µSv at this point. Provided that any work in the vicinity of the shielding is pre-
pared well, it will not take more than 20 minutes (= 1/3 hour) and from this we
can calculate a dose of 10 µSv for working a shorter time in the vicinity of
the shielding.
If you compare this dose with the permissible annual doses of 5 mSv for oper-
ating staff, this work can be carried out 500 times per year by one and the
same person. Due to the very low radiation exposure it is not necessary to carry
a person dosimeter for this work, since the lower detection limit of these meas-
uring systems lies just above this dose.
The anticipated working hours should be spilt up in work in the direct vicinity of
the shielding during mounting and dismantling the source holders and work with
the unshielded source while fixing and dismantling the source and the source
holder. The dose obtained while working in the vicinity of the shielding and the
dose obtained while working with the unshielded source have to be calculated
separately and added up.
Again, a rather simplified estimation is possible when the work is prepared well.
Based on the assumptions of a mean distance of 0.5 m for the whole body ra-
diation and the time you are working with the unshielded source of 6 minutes (=
1/10 hour), the radiation exposure can be calculated for different activities (A) as
follows:
88
Density Meter LB 444 SECTION 9. RADIATION PROTECTION
Using a pocket dosimeter with direct reading, measure the accurate radiation
exposure during this work, even if the radiation exposure lies below the detec-
tion limit of dosimeters.
89
Density Meter LB 444 SECTION 9. RADIATION PROTECTION
If you suspect such a severe malfunction, the Radiation Safety Officer has
to be notified immediately. He/she will prevent further more exposure of per-
sons to radiation.
The Radiation Safety Officer has to make sure that the measuring system is no
longer in operation and then take appropriate steps. He/she may have to inform
the authorities or contact the manufacturer or supplier of the measuring system.
Emergency measures may be taken when a properly trained and licensed per-
son is available:
1. Locate the shielding. Cordon off a safe distance, label and prevent persons
entering the designated area.
4. Secure the shielding. Document the event and estimate the possible radia-
tion level to which the persons involved were exposed.
If you suspect any damage to the source capsule, the following points must be
observed as well:
1. Avoid contamination.
2. Take hold of source using a tool (a pair of pliers or a pair of tweezers) and
put both (source and tool) into a plastic bag.
3. Secure them behind an auxiliary shielding (concrete wall, steel or lead plate).
5. Make sure the radioactive waste is secured and disposed off in compliance
with the pertinent regulatory requirements.
90
Density Meter LB 444 SECTION 10. APPENDIX
10. APPENDIX
Source
91
Density Meter LB 444 SECTION 10. APPENDIX
LB 444 D E N S I T Y - M E T E R V 1.00 a
Parameters
Probe type : No. 0
Isotope : Am241
Current input
Temp. limit values : 0 - 20 mA
0/4 mA : 0.0 deg C
20 mA : 100.0 deg C
Product selection : internal
Product data
Product : 1
Current output
Current output range : 0 - 20 mA
0/4 mA : 3.0 g/cm³
20 mA : 5.0 g/cm³
Error current : -1 mA
Calibration
Unit : g/cm³
Suspension measurement : Yes
92
Density Meter LB 444 SECTION 10. APPENDIX
1. When using a 50/50 crystal, multiply the values listed below by 0.9. With ra-
dial irradiation, these values must be multiplied by 0.85.
2. For measurements with Co-60 sources the listed absorption coefficients must
be multiplied by 0.6.
2
Absorption coefficients in cm /g
3
Product g/cm % conc. Conc. in g/l
Whole milk - .0737 - .00018 - .000017
Skim milk - .0737 - .000027 - .000295
Whey - .0737 - .000294 - .0000254
Sugar solution - .0657 - .00044 - .000021
Hydrochloric acid (HCl) - H2O - .0608 - .0003 - .000025
93
Density Meter LB 444 SECTION 10. APPENDIX
94
Density Meter LB 444 SECTION 10. APPENDIX
Note:
The temperature coefficients can be used for many low concentration products, such as milk, beer, whey,
starch suspension, etc.
For upgraded milk, whey, cottage cheese, etc. the values given for the sugar solution can be used with a
good approximation.
95
Density Meter LB 444 SECTION 10. APPENDIX
3 3
Temp. in °C Density in g/cm Temp. in °C Density in g/cm
10 0.99973 55 0.98673
11 0.99963 56 0.98524
12 0.99951 57 0.98475
13 0.99939 58 0.98425
14 0.99926 59 0.98375
15 0.99911 60 0.98324
16 0.99896 61 0.98272
17 0.99879 62 0.98220
18 0.99861 63 0.98167
19 0.99843 64 0.98113
20 0.99823 65 0.98059
21 0.99801 66 0.98005
22 0.99779 67 0.97950
23 0.99755 68 0.97894
24 0.99731 69 0.97838
25 0.99706 70 0.97781
26 0.99680 71 0.97723
27 0.99653 72 0.97665
28 0.99625 73 0.97607
29 0.99597 74 0.97548
30 0.99567 75 0.97488
31 0.99536 76 0.97428
32 0.99504 77 0.97368
33 0.99472 78 0.97307
34 0.99439 79 0.97425
35 0.99405 80 0.97183
36 0.99370 81 0.97120
37 0.99335 82 0.97057
38 0.99299 83 0.96993
39 0.99262 84 0.96929
40 0.99224 85 0.96864
41 0.99185 86 0.96799
42 0.99146 87 0.96734
43 0.99106 88 0.96668
44 0.99065 89 0.96601
45 0.99024 90 0.96534
46 0.98982 91 0.96467
47 0.98939 92 0.96399
48 0.98896 93 0.96330
49 0.98852 94 0.96261
50 0.98807 95 0.96192
51 0.98761 96 0.96122
52 0.98715 97 0.96052
53 0.98668 98 0.95981
54 0.98621 99 0.95910
96
Density Meter LB 444 SECTION 10. APPENDIX
Product .........................................................................
97
Density Meter LB 444 SECTION 10. APPENDIX
Data input:
Calculation:
98
Density Meter LB 444 SECTION 10. APPENDIX
7.
8.
9.
10.
99
Density Meter LB 444 SECTION 10. APPENDIX
100
Density Meter LB 444 SECTION 10. APPENDIX
101
Density Meter LB 444 SECTION 10. APPENDIX
102
Density Meter LB 444 SECTION 10. APPENDIX
103
Density Meter LB 444 SECTION 10. APPENDIX
104
Density Meter LB 444 SECTION 10. APPENDIX
10.9 Index
105
Density Meter LB 444 SECTION 10. APPENDIX
2
Density Meter LB 444 SECTION 10. APPENDIX