Bagull Ha Goad AutLas) dco pci!
UNIVERSAL INSPECTION CO.WLL.
CR.No 111877 VV AWW) Gage
INSTRUMENT & EQUIPMENT CALIBRATION DIVISION
Tee OY
e
CERTIFICATE NO: UIC /QAT/550 M
This certifies that the instrument below has been tested/ealibrated using
equipment traceable to National/International Standards or verified by
ralio type measurements using self calibrating techniques.
Information on the calibration process and the results are produced
below and given in the accompanying sheets of this document
All related traceable documents are on file at C7INIOWT SSA,
INSPECTION CO.WLL. and may be examined on request
Customer FERNAS CONSTRUCTION CO
Address MESAIEED,STATE OF QATAR
Description PIPE LINE & CABLE LOCATOR:
Manufacturer RADIODETECTION
Model cat ae
Serial No. 10/CAEN31-492 Receipt Date : 23duhI9
1D, No. NA Calibration Date
Location Calibration Lab (Qatar) Calibration Due Date
22am 2)
The instrument was calibrated using the following reference, maintained at UIC Insifument & I
Equipment Calibration division conforming to ISO 17025 requirements
Reference Standard Equipment Used
Description Cert. No. Serial No
Multi Function
Calibeator 413531 21-Feb-20
‘6-72 Digital 4
Multimeter x 16-Feb-20
Due Date
Calibrated By ved By
ineer ‘Quaiity/ Lab Manager
“The measurement results pertain only to the instrument calibrated.
“This cetiicate andthe attached measurement resul sheets shall nol be reproduced except in fl without the prio permission
Building No: 44, Unit Nos A2, Zone : 56 eT Rake AD Hing tH Wg
Street No 208, Vin Khalid St. Seva Row, Qutat Mig see gob Mb oe EAL Toe EE
Mud: 1978 86836251, Have Suararet dias
Email: qata uicom.ge www .aiconigatUNIVERSAL INSPECTION OOWLL.
Function Tested
CERTIFICATE NO: UIC/QAT/550
Performance Test Data
‘Physical Check, Inspection, Functional Check:
‘The Pipeline & Cable Locator has been tested for its functionality, and found satisfactory.
1. Cable Locator Performance Test
Power ON
Performing self test Display model and
Software version.
PIPELINE & CABLE LOCATOR APPLICATION
‘Transmitter:
“Transmitter Power
Pulsed Signal
‘Continuous Signal
Low Output Signa Strength
High Output signal Strength
Frequency (33H2)
Receiver Alarm
Receiver Battery
‘Temperature Rang;
Minimum Operating Temperature (-20°C)
Requirements
creek
check
heck
Cheek
Cheek
Cheek
Check
erreerrinireayy
Results
Pass
Pass
Pass,
Pass
a7< UNIVERSAL INSPECTION CO.WLL, Dagal acd Ladd ahlall 2S pil
Depth Accuracy:
= Une: 5% 0.1 mto'3m (into 10) eck Pass
= Sonde: 5% 0.1 mto7m (4 Into 16) ‘check Pass
General Information about the testing:
‘The Pipeline & Cable Locator has been tested on the functions of Frequency, Voltage & Putse.
During the testing the ambient temperature was 23.3 °C and the relative Humidity was 49:3 % RH.
Comments:
“The results reported in this certificate refer tothe condition of the instrument on the date of
‘testing and carry no implication regarding the long-term stabity ofthe instrument.
Traceability:
“The reported tests were undertaken using equipment detailed below for which the company
‘maintain traceability to the NIST Va SASSOO Certification.
Quality /Lab Manager