CPM 223 / 253 Transmitter For PH and Redox: Liquisys M
CPM 223 / 253 Transmitter For PH and Redox: Liquisys M
BA 194C/07/en/07.02
No. 51500268
Software version 2.50 or later
Operating Instructions
ENDRESS+HAUSER
LIQUISYS S
ENDRESS+HAUSER
LIQUISYS S
Quality made by
slp 8.0 deen 220802
Endress+Hauser
Endress Hauser
The Power of Know How
ISO 9001
Need information on the instrument?
Please read the following chapters:
General information
Safety
Installation
Operation
Instrument configuration
Interfaces
Diagnosis Maintenance
Ω
% °C
Accessories Technical data
Appendix Index
Liquisys M CPM 223 / 253 Table of contents
PM253E00.CHP
Table of contents
1 General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.1 Symbols used . . . . . . . .... ........ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.2 Storage and transport . .... ........ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.3 Unpacking . . . . . . . . . . .... ........ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.4 Dismantling, packaging and disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.5 Product structure . . . . . .... ........ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.1 Intended application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.2 General safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.3 Installation, start-up, operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.4 Monitoring and safety features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.5 Immunity to interference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.6 Declaration of conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3 Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3.1 Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3.2 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.3 Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.4 Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.5 Electrode installation and measuring cable connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.1 Operator interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.2 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.3 Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
4.4 Auto / manual mode of operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
4.5 Operating concept. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
5 Instrument configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.1 Start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
5.2 System configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
5.3 Current input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
5.4 Current outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.5 Monitoring functions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
5.6 Relay contact configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
5.7 Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
5.8 E+H Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
5.9 Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
5.10 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
5.11 Offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
6 Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
9 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
10 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
11 Appendix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
12 Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Endress+Hauser 1
General information Liquisys M CPM 223 / 253
1 General information
Caution:
1.3 Unpacking
Verify that the packaging and the contents Keep the original packaging materials for
are undamaged! Inform the post office or future storage or shipping of the instrument.
freight carrier of any damage.
Damaged merchandise must be retained until If you have any questions, consult your
the matter has been settled. supplier or the Endress+Hauser sales agency
in your area (see back cover of these
Check that the delivery is complete and operation instructions for addresses).
agrees with the shipping documents and your
order (refer to nameplate for type and
version).
2 Endress+Hauser
Liquisys M CPM 223 / 253 General information
PM253E01.CHP
,l
Version
IS pH/redox measurement with IsFET or glass electrode
PR pH/redox measurement with glass electrode
PS pH/redox measurement with glass electrode, with additional functions
(Plus package)
Power supply
0 Power supply 230 V AC
1 Power supply 115 V AC
2 Power supply 230 V AC, CSA Gen. Purp.
3 Power supply 115 V AC, CSA Gen. Purp.
5 Power supply 100 AC
7 Power supply 24 AC, CSA Gen. Purp.
8 Power supply 24 V AC/DC
Measurement output
0 1 output signal: pH / redox
1 2 output signals: pH / redox and temp. / pH or redox / set value
3 1 output signal: PROFIBUS-PA
4 1 output signal: PROFIBUS-DP
5 1 output signal: pH / redox with HART®
6 2 output signals: pH / redox, HART® and temperature
Contacts
05 No additional contacts
10 2 contacts (limit values / PID / timer)
15 4 contacts (limit values / PID / timer / ChemoClean)
16 4 contacts (limit values / PID / timer)
20 2 contacts with current input (limits / PID / timer)
25 4 contacts with cleaning, current input
(limit / PID / ChemoClean)
26 4 contacts with timer, current input
(limits / PID / timer)
CPM253-
complete order code
CPM223-
Endress+Hauser 3
Safety Liquisys M CPM 223 / 253
2 Safety
• This device may only be installed, • Check that all connections have
connected electrically, been properly made before
commissioned, operated and powering up the system!
serviced by properly trained
personnel authorized by the • Damaged equipment that may be
system operator. dangerous must not be operated
and should be clearly identified as
• Technical personnel must be being defective.
familiar with these operating
instructions and must adhere to • Any troubleshooting of the
the instructions described therein. measuring system is to be
performed exclusively by
• Make sure that the power supply authorized, trained personnel.
ratings match the data specified
on the nameplate before you • If faults cannot be remedied, the
connect the instrument to a power instrument must be removed from
source. service and secured to prevent
accidental start-up.
• A clearly identified mains
disconnecting device must be • Repairs not described in these
installed close to the instrument. operating instructions may only be
performed at the manufacturer’s
• Live components can be touched works or by the Endress+Hauser
through the vent slots in the Service Organization.
housing and the openings on the
rear of the housing. Do not insert
tools, wires, etc., in these slots
(CPM 223 only).
4 Endress+Hauser
Liquisys M CPM 223 / 253 Safety
PM253E02.CHP
The transmitter is protected against external In the event of a system error or power failure,
influences and damage by the following an alarm condition is signalled via a fault-
design measures: signalling contact.
• Rugged housing
• Degree of protection provided by
enclosure: IP 65 (CPM 253)
• UV resistance
Endress+Hauser 5
Installation Liquisys M CPM 223 / 253
3 Installation
The following procedure should be followed for a complete measuring system installation:
ENDRESS+HAUSER
LIQUISYS
REL1
E REL2
Flow assembly
CPA 250
Measuring cable
Immersion assembly
e. g. CPK 9
CPA 111
CleanFit W
REL2
CAL REL
REL1
E REL2
CPA 450
6 Endress+Hauser
Liquisys M CPM 223 / 253 Installation
PM253E03.CHP
3.2 Dimensions
Pg 13.5
Pg 7
Pg 16
11
ENDRESS+HAUSER
LIQUISYS
157
REL1
ALARM
REL2
CAL REL
REL1
E REL2
∅6
247
70
115 70
154
170
Dimensions of Liquisys M
FELD1.CDR
Fig. 3.2 CPM 253
Note:
Inside of housing of
Liquisys M CPM 253:
1 Removable
3 electronics box
2 Partition plate
3 Terminal blocks
FELD2.CDR Fig. 3.3 4 Fuse
Endress+Hauser 7
Installation Liquisys M CPM 223 / 253
96
90 6
139
3.3 Mounting
270
320
300
8 Endress+Hauser
Liquisys M CPM 223 / 253 Installation
PM253E03.CHP
Mounting kit for installation of field housing on Square tube for mounting of measuring
horizontal and vertical pipes (max. ∅ 60 mm); transmitters; SS 304 material;
material: stainless steel SS 304 (AISI 304); order no.: CYY102-A
order no. 50086842
90
70
∅8
∅ max. 60
1495
60 x 60
M6
70
90
60
∅6
70
90
150
left:
Mounting kit
for post mounting on
cylindrical pipes
120
right:
MONT5.CDR Fig. 3.6 Square mounting post
Endress+Hauser 9
Installation Liquisys M CPM 223 / 253
Pipe mounting of
MONT10.CDR
Fig. 3.7 Liquisys M field instrument
left:
Wall mounting
right:
Mounting with universal
post and weather
MONT2.CDR
Fig. 3.8 protection cover
10 Endress+Hauser
Liquisys M CPM 223 / 253 Installation
PM253E03.CHP
92+0.5 92+0,5
3
92
3
90
96
3
57
33
Attachment of panel-
mounted instrument
max. 45
1 Wall of control
approx. 139 6
cabinet
25
2 Gasket
MONT9.CDR Fig. 3.9 3 Tensioning screws
Endress+Hauser 11
Installation Liquisys M CPM 223 / 253
The connection diagram depicted in Fig. 3.10 shows the connections of an instrument
equipped with all the options. The connection of the various electrodes with the measuring
cables is shown in more detail in the subsequent figures.
pH/SRC 41
42
Electrode Alarm
Ref. 43 (current-free
contact
position)
47
Ground glass electrodes S
48
Ground IsFET PA/
PM 49
Relay 1
(current-free
contact
position
GN 11 57
WH 12 58
Relay 2
Temperature
YE 13 59 (current-free
contact
position)
51
31
mA
52
Relay 3
Signal output 1 32
53 (current-free
pH / redox pH contact
position)
33 54
mA
55
Signal output 2 34 Relay 4
temperature 56 (current-free
contact
position)
81
Digital 23 +
input 1 10-50 V 24 -
(Hold) 82
Current input
4...20 mA
mA
∼ =
93 L1 L+
Digital
input 2 N L-
10-50 V Mains
(Chemoclean) 94
85
Aux. voltage
15 V
output 86
Electrical connection of
Liquisys M CPM 223/253
Fig. 3.10 with full wiring ANPH01.CDR
12 Endress+Hauser
Liquisys M CPM 223 / 253 Installation
PM253E03.CHP
Note: Caution:
REF (Glas/IsFET)
AC: 250V/500VA
DC: 30V/60W
RD WH BN
54 55 56 51 52 53 57 58 59 47 48 49 41 42 43
PA/PM
Mains
Hilfsenergie
DRN 12
NC
NC 31 32 33 34 NC 85 86 93 94 81 82 23 24 NC L1 N (AC)
+ - + - + - + - + - + - + - (DC)
4..20mA
Digital 2
Digital 1
SRC 11 13 S
GN YE
Redox
pH
(opt.)
Temp.
10mA
+15V
12
(IsFET)
11 13 131094-4G
Connection compartment
131094F.CDR Fig. 3.12 sticker of field instrument
>>
131084-4C 131081-4E
Sensor
BNC: pH/Redox + REF (Glass)
Contacts: max. 2A, AC:250V/500VA, DC:30V/60W
49
59 33 + 13 YE
Temp.
REL2 57 Ground 34 -
(opt.)
12 WH
terminal
COM 58 for IS NC 11 GN
53 85 + SRC (IsFET)
+15V
REL3 51 86 - 10mA DRN RD
COM 52 93 + S
Digital 2
56 94 - 42
REL4 54 81 + 43
Digital 1
COM 55 82 - 41
- 24 NC AC DC
Mains
4..20mA L1 +
+ 23 NC N -
Connection sticker of
131081F.CDR Fig. 3.11 panel-mounted instrument
Endress+Hauser 13
Installation Liquisys M CPM 223 / 253
Construction of special
Fig. 3.13 measuring cable CPK1_12e.EPS
14 Endress+Hauser
Liquisys M CPM 223 / 253 Installation
PM253E03.CHP
Cable termination for panel-mounted instrument CPM 223 with solder-free BNC plug
(for glass electrodes only)
Connection examples ➀ ➁ 16
➅
➆
➇
Installation of terminated
pH connecting line
BNC2NEU.CDR Fig. 3.17 in BNC elbow plug
Endress+Hauser 15
Installation Liquisys M CPM 223 / 253
Note:
pH electrode connection
to Liquisys M
CPM 223 / 253
left:
Connection of CPS 11
with CPK 9
right:
Connection of CPS 401
Fig. 3.18 with CPK12 PH-AN01E.CDR
Redox electrode
connection to Liquisys M
CPM 223 / 253
left:
Asymmetrical (without PM)
right:
Fig. 3.19 Symmetrical (with PM) RE-AN01E.CDR
16 Endress+Hauser
Liquisys M CPM 223 / 253 Operation
PM253ED4.CHP
4 Operation
ENDRESS+HAUSER
LIQUISYS M Display of active
Liquid crystal display
contact and key for
for display of measured
relay switching in
values and onfiguration
manual mode
data pH
705 R262
Setpoint
LED indicator for
Field for user labelling alarm function
4.2 Display
LED indicators
Indication of current operating mode “Auto” (green LED) or “Manual” (yellow LED)
Alarm indication e.g. for continuous limit violation, temperature sensor failure or system error
(see error list in chapter 7)
Setpoint
offset
In measuring mode:
Main measured value.
In setup / calibr. mode
Error In measuring mode: e.g. parameter
Manual Automatic Secondary measured value.
indicator
temperature temperature In setup / calibr. mode
compensation compensation e.g. setting value
DISPLAY3.CDR Fig. 4.2 Liquid crystal display
Endress+Hauser 17
Operation Liquisys M CPM 223 / 253
4.3 Keys
CAL key
When the CAL key is pressed, the instrument prompts for the calibration
access code:
Code 22 for calibration
Code 0 or any code for calibration data checking.
Press the CAL key to acknowledge the calibration data or to proceed
within the calibration menu.
ENTER key
The ENTER key has serveral functions:
Opens the Setup menu in measuring mode
Stores (acknowledges) data entered in Setup mode
Moving on within the function groups.
REL key
The REL key toggles between the relay and manual cleaning start in
manual mode. In automatic mode you can output the corresponding
switch-on points (limit contactor) ot set points (PID controller) when
pressing the REL key. Pressing the PLUS key allows you to display the
settings of the following relay. Press the REL key to return to measuring
mode (automatic return after 30s).
AUTO key
The AUTO key is used to toggle between the automatic and manual
modes of operation.
Escape function
Press the PLUS and MINUS keys simultaneously to return to the main
menu. Press the PLUS and MINUS keys again to return to measuring
mode.
18 Endress+Hauser
Liquisys M CPM 223 / 253 Operation
PM253ED4.CHP
Auto mode
In this mode of operation, the relays are controlled by the transmitter.
REL key
In manual mode, the REL key is used to select one of the relays or the
cleaning function present in the instrument.
Select relay or function. Press the REL key to toggle between the relays.
The display shows the selected relay and the switching status (ON /
OFF) in the second line. In manual mode, the measuring value is
continuously displayed(e.g. for monitoring during dosage).
Set the relays. Switch on with PLUS, switch off with MINUS. The relay
state remains in effect until it is actively reset.
Note:
Endress+Hauser 19
Operation Liquisys M CPM 223 / 253
Measuring mode:
Standard mode of operation,
display of current
measured values
Code
Description of possible
SET-MOD.CDR
Fig. 4.3 operating modes
Note:
20 Endress+Hauser
Liquisys M CPM 223 / 253 Operation
PM253ED4.CHP
• CAL key + Code 22: Access to Calibration • PLUS + ENTER + Code keys 9999:
and Offset menus. Locks the keypad.
• ENTER + Code keys 22: Access to the • CAL + MINUS + Code keys 0:
Configuration menus, allowing configurati- Unlocks the keypad.
on and user-specific settings. • CAL or ENTER + Code keys: access to
Read mode, i.e. all settings can be read
but not changed.
Functions
(parameter selection, value entry)
Function groups
Move forward and backward with
the PLUS and MINUS keys
E E E E
Schematic representation
of Liquisys M menu
BEDSCHE.CDR Fig. 4.4 structure
Endress+Hauser 21
Operation Liquisys M CPM 223 / 253
Note:
22 Endress+Hauser
Liquisys M CPM 223 / 253 Instrument configuration
PM253E5A.CHP
5 Instrument configuration
After power-up the instrument performs a Setup mode
self-test and then enters to measuring mode.
➜ SETUP 1 (A) see chap. 5.2.1
Now it can be configured and calibrated for ➜ SETUP 2 (B) see chap. 5.2.2
the first time. The values set by the user are ➜ CURRENT INPUT (Z)
kept even in the event of a power failure. see chap. 5.3
➜ CURRENT OUTPUT (O)
The following function groups are available on
see chap. 5.4
the Liquisys M (the function groups that are
only available in the Plus package are marked ➜ ALARM (F) see chap. 5.5.1
accordingly in the functional descriptions): ➜ CHECK (P) see chap. 5.5.2
➜ RELAY (R) see chap. 5.6
➜ SERVICE (S) see chap. 5.7
➜ E+H SERVICE (E) see chap. 5.8
➜ INTERFACE (I) see chap. 5.9
Endress+Hauser 23
Instrument configuration Liquisys M CPM 223 / 253
Factory settings
When the instrument is switched on for the first time, the factory settings are in effect. The
following table provides an overview of all major settings. Please refer to the description of the
individual functions in chapter 5 for all other factory settings (the factory settings are printed in
bold face).
Alarm contact
41 41
42 42
43 43
Recommended fail-safe
Fig. 5.3 circuit for an alarm contact RELAY2.CDR
24 Endress+Hauser
Liquisys M CPM 223 / 253 Instrument configuration
PM253E5A.CHP
5.1 Start-up
After switching the instruement on, make the following settings to the specified function groups:
Selection or range
Coding Field Display Info
Factory setting (bold)
Function group A
Initial display in function group
A SETUP 1.
SETUP 1
SETUP 1
pH
A2 Select display unit mV pH A2
%
Unit
A3
Select connection
mode
sym = symmetrical sym A3 See chapter 3.
asym = asymmetrical
wiring
Measured value damping causes
averaging over the specified
number of individual measured
A4
Enter measured
value damping
1 1 A3 values and is used e. g. to
1 ... 60 stabilise the display and the
Damping signal output. There is no
damping if “1" is entered.
Endress+Hauser 25
Instrument configuration Liquisys M CPM 223 / 253
Selection or range
Coding Field Display Info
Factory setting (bold)
Selection or range
Coding Field Display Info
Factory setting (bold)
Function group B
Initial display in function group
B SETUP 2.
SETUP 2
SETUP 2
– For pH operating
mode:
Selection of type ATC At B1 = ATC: Move to B3.
B1
of temperature
compensation
MTC
– For ORP operating
ATC B1
At B1 = MTC: Enter the process
temperature in B2, which is to be
(during process) mode: °C-Proces used for compensation.
Off
On
Only at B1 = ATC.
Temperature °C
B5 difference (offset) 0.0 °C 0.0 B5 The difference between the
is displayed –5.0 ... 5.0 °C temperatures measured and
Temp.Offs entered is displayed.
Factory settings are printed in bold face;
base version does not include functions in italic.
26 Endress+Hauser
Liquisys M CPM 223 / 253 Instrument configuration
PM253E5A.CHP
ON
Flow in
main stream
OFF
ON
Relay contacts
see field Z2 see field Z3
of PID controller
OFF
ON
Alarm relay
OFF
Flow alarm
Flow below Flow
switch-off limit Z4 restored
Alarm signalling and
or flow failure
dosing switch-off by the
Fig. 5.4 main stream
Endress+Hauser 27
Instrument configuration Liquisys M CPM 223 / 253
Measuring water
extraction point Static mixer Injection points Flow meter
mg/l
0.42
CPS 11 25.0 °C
REL1
ALARM
REL2
CAL REL
REL1
E REL2
Acid Lye
Feedforward control is a multiplying function as depicted in the below figure (factory setting as
example):
Gain
Kinflu
1.5
0.5
0 20 40 Z7 60 80 100
Current input
Multiplying feedforward signal [%]
Fig. 5.6 control
28 Endress+Hauser
Liquisys M CPM 223 / 253 Instrument configuration
PM253E5A.CHP
Selection or range
Coding Field Display Info
Factory setting (bold)
Initial display in
Function group Z function group
Z
CURRENT INPUT CURRENT INPUT.
CUR.INPUT
Select flow rate Only switch on when flow meter
monitoring of main is connected in main stream.
Off Z1
Z1 stream (with Off
controller Input When Z1 = Off, fields Z2 to Z5 do
switch-off) Cont.stop not exist.
0 … 100% corresponds to
Enter switch-off % 4 … 20 mA at current input. Note
Z4 threshold for 50% 50 Z4 the measured value allocation to
current input 0 ... 100% the current output of the flow
A.Thresh meter.
Endress+Hauser 29
Instrument configuration Liquisys M CPM 223 / 253
pH
12
10
6
0 4 8 12 16 20
Current [mA]
∆ current
User-defined current output
Fig. 5.7 characteristic
The distance ∆ signal per mA between two table value pairs must exceed:
• pH: 0.03
• Redox: 5 mV
• Temperature: 0.25 °C
First enter the current output configuration you require in the following blank table. Ensure the
required minimum distance by calculating the resulting signal distance per mA. Then enter the
result in the instrument.
pH / mV / pH / mV /
Current Distance Current Distance
Value pair % / °C % / °C
[mA] per mA [mA] per mA
[ ] [ ]
1
2
3
4
5
6
7
8
9
10
30 Endress+Hauser
Liquisys M CPM 223 / 253 Instrument configuration
PM253E5A.CHP
Selection or range
Coding Field Display Info
Factory setting (bold)
Function group
O
Initial display in function group
O CURRENT CURRENT OUTPUT.
OUTPUT
OUTPUT
O1
Selection of
current output
Out1 Out1 O1 A different characteristic can be
selected for each output.
Out2
Sel.Out
O311
Selection of
current range
4–20 mA 4-20 O311
0–20 mA
Sel.Range
pH 2.00
pH –2.00 ... 16.00
0/4 mA value; Enter the measured value
–1500 mV pH
enter corresponding the the minimum
O312 corresponding pH
–1500 ... 1500 mV 2.00 O312 current value (0/4 mA) at the
(redox) or 0.0 % transmitter output. (Spreading
temperature value 0.0 ... 100.0 % 0/4 mA see Technical data.)
0.0 °C
–20.0 ... 150.0 °C
pH 12.00
pH –2.00 ... 16.00
20 mA value; Enter the measured value
1500 mV pH
enter corresponding to the maximum
O313 corresponding pH
–1500 ... 1500 mV 12.00 O313 current value (20 mA) at the
(redox) or 100.0 % transmitter output. (Spreading
temperature value 0.0 ... 100.0 % 20 mA see Technical data.)
100.0 °C
–20.0 ... 150.0 °C
Endress+Hauser 31
Instrument configuration Liquisys M CPM 223 / 253
Selection or range
Coding Field Display Info
Factory setting (bold)
mA
O321
Enter simulation
value
current value 10.20 O321 The current value entered here is
output through the current output.
0.00 ... 22.00 mA
Simulat.
Values may also be added or
changed at a later point in time.
Enter current lin = linear (1) The values entered are automatic-
O3 (3) output table sim = simulation (2) table O2 ally sorted in ascending order by
(only for S version) Tab = Table (3) current value.
Sel.Type See O2 (1), O2 (2) for other
characteristics.
O331
Selection of table
option
read read O331
edit
Sel.Table
1
O333
Selection of table
value pair 1 ... No. Elem. 1 O333
asign
Sel.Elem.
pH 0.00
pH –2.00 ... 16.00
pH
O334 Enter x value 0 mV 0.00 O334 x value = Measured value
determined by user.
–1500 ... 1500 mV
0.0 % Meas.val.
0.0 ... 100.0 %
32 Endress+Hauser
Liquisys M CPM 223 / 253 Instrument configuration
PM253E5A.CHP
5.5.1 Alarm
Selection or range
Coding Field Display Info
Factory setting (bold)
Function group
F F Alarm function settings.
ALARM
ALARM
Stead = steady
F1
Selection of
contact type
contact Stead F1 The contact type selected here
only applies to the alarm contact.
Fleet = fleeting
contact Cont.Type
F2
Selection of time
unit
s s F2
min
Time Unit
Endress+Hauser 33
Instrument configuration Liquisys M CPM 223 / 253
Selection or range
Coding Field Display Info
Factory setting (bold)
If set “no” all the other alarm
settings (e. g. alarm delay) are
also deactivated. The settings
Set alarm contact
F6 to be effective for yes yes F6 themselves are retained.
This setting only applies to the
selected error no
Rel.Assg error selected in F5.
Factory setting is no starting with
E080.
F8
Automatic start of
cleaning function?
no no F8 This field only exists for some
errors, see chapter 7.1.
yes
CleanTrig
34 Endress+Hauser
Liquisys M CPM 223 / 253 Instrument configuration
PM253E5A.CHP
5.5.2 Check
The function group CHECK is only accessible for instruments equipped with the Plus packet.
Two different monitoring functions can be selected for the measurement in the function group
CHECK:
The sensor check system monitors the pH and reference electrode for inaccurate measurement
and total failure.
Function AC is used to examine the measuring signal for deviations. If the measuring signal is
constant for a specific period of time (several measured values), an alarm is issued. This type
of behaviour may be caused by soiling, blocking, etc.
Function CC monitors the relay activity. Due to freely selectable monitoring periods for the limit
functions, a relay failure will be recognised and an alarm will be triggered.
Endress+Hauser 35
Instrument configuration Liquisys M CPM 223 / 253
Limit monitoring
This function monitors the measuring value for exceeding monitoring limits and triggers an
alarm.
Selection or range
Coding Field Display Info
Factory setting (bold)
Function group P
Settings for measuring electrode
P and process monitoring
CHECK
CHECK
The glass electrode is monitored
for glass breakage (Error no.
Switch SCS alarm
P1 for measuring Off Off P1 E071.) Response time approx. 30s.
electrode on or off On
SCS Glass SCS monitoring is inactive during
calibration.
The reference electrode is
monitored for soiling and
Switch SCS alarm
P2 for reference Off Off P2 blocking. (Error no. E152.)
Response time approx. 60s.
electrode on or off On
SCS Ref
Only at A2 = sym.
The measuring value contains
also the impedance of the
Enter SCS alarm
k@ medium.
threshold for 50.0 kΩ 50.0 P3
P3
reference 0.5 ... 50.0 kΩ The impedance of the reference
electrode SCSThresh electrode increases with the
degree of soiling.
Only if A4 = IsFET and
P1 = on.
P4
Leakage current
display for IsFET 0.0 ... 9.9 µA 0.0 P4
Display only.
pH
P7
Enter lower alarm
threshold
-2.00 pH -2.00 P7 Omitted when P5 = Off.
-2.00 ... 16.00 pH
LowAlarm
36 Endress+Hauser
Liquisys M CPM 223 / 253 Instrument configuration
PM253E5A.CHP
Selection or range
Coding Field Display Info
Factory setting (bold)
pH
P8
Enter upper alarm
threshold
16.00 pH 16.00 P8 Omitted when P5 = Off.
-2.00 ... 16.00 pH
HighAlarm
AC = Sensor alternation check,
Off CC = Controller check.
AC
Select process CC Alarm signalling optionally with or
P9 monitoring AC+CC Off P9 without simultaneous controller
(PCS alarm) AC! switch-off.
CC! ProcMonit xxxxx = without controller
ACCC! switch-off,
xxxx! = with controller switch-off.
Enter maximum
min
P10
permissible
period for lower
60 min 60 P10 Only when P9 = CC or AC+CC.
0 ... 2000 min
alarm threshold Tmax Low
Enter maximum
min
P11
permissible
period for upper
120 min 120 P11 Only when P9 = CC or AC+CC.
0 ... 2000 min
alarm threshold Tmax High
Endress+Hauser 37
Instrument configuration Liquisys M CPM 223 / 253
The relay contacts described below can be selected and configured as required (max. four
contacts depending on options installed):
Switch-on point
Switch-off point
Contact OFF
t1 t2 t3 t4 t5 t6 t7 t
Contact OFF
Switch-on point
Switch-off point
Contact ON
Alarm OFF
Alarm threshold
Alarm ON
t1 t2 t3 t4 t5 t6 t7 t
Depiction of alarm and
Fig. 5.10 limit functions GRENZ-E.CDR
38 Endress+Hauser
Liquisys M CPM 223 / 253 Instrument configuration
PM253E5B.CHP
• P controller: Used for simple linear control purposes with small system deviations. Where
major changes are to be controlled, overshooting may occur. A control offset is to be
expected.
• PI controller: Used for processes where overshooting is to be avoided and permanent
offsets are not allowed.
• PD controller: Used for processes that require quick response and where peaks are to be
corrected.
• PID controller: Used for processes for which the type of control provided by a P, PI or PD
controller is inadequate.
Start-up
If there are no empirical values available for setting the control parameters, use values that
provide the greatest possible stability of the control loop. To optimise the control loop further:
• Increase the control gain Kp until the control variable just starts to overswing.
• Decrease Kp again slightly and shorten the integral action time Tn to achieve the shortest
possible correction time without overswing.
• In order to shorten the response time of the controller, you also have to set the derivative
action time Tv.
Endress+Hauser 39
Instrument configuration Liquisys M CPM 223 / 253
Actual Actual
value
Tn too large
Tn too small
Time Time
Actual Actual
value
Kp too small
Kp too large
Time Time
Actual
Optimal setting
The control contact in question outputs a switched signal. The intensity of this signal is
proportional to the controller’s control output. A distinction is made according to the type of
signal output:
Contact IMPULS.CDR
Contact
tON
tOFF tON
On On
40 Endress+Hauser
Liquisys M CPM 223 / 253 Instrument configuration
PM253E5B.CHP
Field R236 offers two control characteristics for selection which have the effects shown in the
following diagram.
100 %
direct =
max. function
– Xw Xw
Y
100 %
inverted =
min. function
Control characteristic of a
Xw proportional-action
–Xw controller with direct and
REGEL03.CDR Fig. 5.13 inverted control action
inactive
inactive
Time Relationship among
cleaning time, pause time
DIAGR-TI.CDR Fig. 5.14 and hold dwell period
Endress+Hauser 41
Instrument configuration Liquisys M CPM 223 / 253
Hold
t1 cleaning start
Water t2 – t1 pre-rinse time
t3 – t2 cleaning time
t4 – t3 post-rinse time
t5 – t4 hold dwell period
Cleaner
t0 t1 t2 t3 t4 t5
Fig. 5.15 Cleaning cycle sequence CHEMO.CDR
Y [%]
NEUTRA.CDR Neutralisation control means that the pH value
of a medium is held constant by adding acid
100 or alkali as required. This task requires two
neutral separate actuating signals – one for acid and
zone one for alkali.
50
42 Endress+Hauser
Liquisys M CPM 223 / 253 Instrument configuration
PM253E5B.CHP
Selection or range
Coding Field Display Info
Factory setting (bold)
Function group R
Relay contacts can be selected
R and adjusted.
RELAIS
RELAY
Rel3 (water) and Rel4 (cleaner)
Rel1 are only available on instrument
Selection of
R1 contact to be Rel2 Rel1 R1 equipped accordingly. If
ChemoClean selected as the
configured Rel3
Rel4 Sel.Relay cleaning type, Rel4 is not
available.
R211
Switch function of
R2 (1) off or on
off off R211 All settings are retained.
on
Function
pH 16.00
pH –2.00 ... 16.00 Never set switch-on point and
Enter switch-on pH
R212 point of the 1500 mV 16.00 R212
switch-off point to the same value.
(Only the operating mode
contact –1500 ... 1500 mV
100.0 % On value selected in A1 appears.)
0.0 ... 100.0 %
The switch-off point entry selects
pH 16.00 a max contact (switch-off point
pH –2.00 ... 16.00 < switch-on point) or a min
Enter switch-off pH
R213 point of the 1500 mV 16.00 R213 contact (switch-off point
> switch-on point), thereby
contact –1500 ... 1500 mV
100.0 % Off value implementing an always required
0.0 ... 100.0 % hysteresis function (see Fig.
5.10).
s
R214 Enter pickup delay 0s 0 R214
0 ... 2000 s
On Delay
s
R215
Enter dropout
delay
0s 0 R215
0 ... 2000 s
Off Delay
Endress+Hauser 43
Instrument configuration Liquisys M CPM 223 / 253
Selection or range
Coding Field Display Info
Factory setting (bold)
When the alarm threshold is
violated, the measuring
pH 16.00 transmitter issues an alarm with
pH –2,00 ... 16.00 an error message and error
Enter alarm
pH current (note alarm delay in
R216
threshold
(as an absolute
1500 mV 16.00 R216 field F3).
–1500 ... 1500 mV
value) 100.0 % A.Thresh When defining the min contact,
0.0 ... 100.0 % the alarm threshold must be set
to a lower value than the
switch-off point.
R217
Show status for
limit contact
MAX MAX R217 Only display.
MIN
LC State
LC PV = limit
contactor pH/redox (1)
Configure limit LC °C = limit Confirmation with ENTER
R2 (2)
contactor for
temperature
contactor T (2) LC °C R2 switches off a different, already
switched-on function and its
PID-controller (3)
measurement Timer (4) Sel.Type settings are reset to the default.
Clean = ChemoClean (5)
Neutra controller (6)
R221
Switch function of
R2 (2) off or on
off off R221
on
Function
150.0 R222
°C
Enter switch-on 150.0 °C Never set switch-on point and
R222
temperature –50.0 ... 150.0 °C switch-off point to the same value.
On value
The switch-off point entry selects
a max contact (switch-off point
< switch-on point) or a min
150.0 R223
°C
Enter switch-off 150.0 °C contact (switch-off point
R223
temperature –50.0 ... 150.0 °C > switch-on point), thereby
Off value implementing an always required
hysteresis function (see Fig.
5.10).
0 R224
s
R224 Enter pickup delay 0s
0 ... 2000 s
On Delay
0 R225
s
Enter dropout 0s
R225
delay 0 ... 2000 s
Off Delay
44 Endress+Hauser
Liquisys M CPM 223 / 253 Instrument configuration
PM253E5B.CHP
Selection or range
Coding Field Display Info
Factory setting (bold)
When the alarm threshold is
violated, the measuring
transmitter issues an alarm with
an error message and error
Enter alarm
current (note alarm delay in
150.0 R226
°C
threshold 150.0 °C
R226 field F3).
(as an absolute –50.0 ... 150.0 °C
value)
A.Thresh When defining the min contact,
the alarm threshold must be set
to a lower value than the
switch-off point.
R227
Show status for
limit contact
MAX MAX R227 Only display.
MIN
LC State
LC PV = limit
contactor pH/redox (1)
LC °C = limit
contactor T (2) Confirmation with ENTER
R2 (3)
Configure
P(ID) controller PID controller (3) PID R2 switches off a different, already
switched-on function and its
Timer (4)
Clean = ChemoClean (5) Sel.Type settings are reset to the default.
Neutr =
Neutra controller (6)
Off
On = PID control
R231
Switch function
R2 (3) off or on
On off R231 Basic = only basic load dosage
Basic
PID+B = PID control with basic
PID+B
Function load dosage
pH 16.00
pH –2.00 ... 16.00 The set point is the value to be
pH maintained by the control. This
R232 Enter set point 1500 mV 16.00 R232 value is to be re-established by
–1500 ... 1500 mV the control in the event of any
0.0 % Setpoint deviation (up or down).
0.0 ... 100.0 %
R233
Enter control gain
Kp
1.00 1.00 R233 See chapt. 5.5.2
0.01 ... 20.00
Kp
See chap. 5.5.2
Enter integral min
At every hold the I component is
R234
action time Tn
(0.0 = no I
0.0 min 0.0 R234
set to zero.
Hold can be deactivated in S2,
0.0 ... 999.9 min
component)
Time Tn but not for ChemoClean and
timer!
Endress+Hauser 45
Instrument configuration Liquisys M CPM 223 / 253
Selection or range
Coding Field Display Info
Factory setting (bold)
s
R2310 Enter minimum
ON time tON
0.3 s 0.3 R2310 This field only appears if pulse
length is selected in R237.
0.1 ... 5.0 s
Min.PTime
Selecting the basic load, you
choose the desired dosage
% quantity.
R2311
Enter basic load 0% 0 R2311 100% basic load corresponds to:
0 ... 40% steadily on at R237 = on
BasicLoad Fmax at R237 = freq
20 mA at R237 = curr
Batch = discontinuous process
Inlne = continuous process
46 Endress+Hauser
Liquisys M CPM 223 / 253 Instrument configuration
PM253E5B.CHP
Selection or range
Coding Field Display Info
Factory setting (bold)
LC PV = Limit
contactor pH/redox (1) Cleaning is performed using only
LC °C = Limit one cleaning agent (usually
Configure water); see Fig. 5.14.
R2 (4) cleaning function
contactor T (2)
PID-controller (3)
Timer R2 Confirmation with ENTER
(timer) switches off a different, already
Timer (4) Sel.Type switched-on function and its
Clean = ChemoClean (5) settings are reset to the default.
Neutra controller (6)
R241
Switch function of
R2 (4) off or on
off off R241
on
Function
min
The minimum pause time
R244
Enter minimum
pause time
120 min 120 R244 prevents continuous cleaning
when the cleaning trigger is
1 ... R243 min
Min.Pause present.
LC PV = Limit
Configure contactor pH/redox (1)
cleaning with LC °C = Limit See chap. 5.5.4, p.37.
ChemoClean contactor T (2) Confirmation with ENTER
R2 (5)
(on version with
four contacts and
PID controller (3) Clean R2 switches off a different, already
Timer (4) switched-on function and its
appropriate
assignment of
Clean = Sel.Type settings are reset to the default.
contacts 3 and 4) ChemoClean (5)
Neutra controller (6)
R251
Switch function of
R2 (5) off or on
off off R251
on
Function
int = internal
(timer-contr.) The “int” cycle is triggered by the
ext = external end of the pause time (R257).
R252
Selection of start
pulse
(digital input 2) int R252
There is no real-time clock.
External suppression is required
i+ext = intern. + extern.
i+stp = internal, CleanTrig for irregular time intervals (e.g.
weekends).
suppressed
by external
Endress+Hauser 47
Instrument configuration Liquisys M CPM 223 / 253
Selection or range
Coding Field Display Info
Factory setting (bold)
s
R253
Enter pre-rinse
time
20 s 20 R253 Water is used for rinsing.
0 ... 999 s
PreRinse
s
R254
Enter cleaning
time
10 s 10 R254 Cleaning agent and water are
used for cleaning
0 ... 999 s
CleanTime
s
R255 Enter rinse time 20 s 20 R255 Water is used for rinsing.
0 ... 999 s
PostRinse
R256
Enter number of
repeat cycles
0 0 R256 R253 ... R255 is repeated.
0 ... 5
Rep.Rate
R261
Switch function of
R2 (6) off or on
off off R261
on
Function
48 Endress+Hauser
Liquisys M CPM 223 / 253 Instrument configuration
PM253E5B.CHP
Selection or range
Coding Field Display Info
Factory setting (bold)
s
R269
Enter minimum
ON time tON
0.3 s 0.3 R269
0.1 ... 5.0 s
Min.PTime
Batch = discontinuous process
Inlne = continuous process
Endress+Hauser 49
Instrument configuration Liquisys M CPM 223 / 253
5.7 Service
Selection or range
Coding Field Display Info
Factory setting (bold)
Function group
S S
SERVICE
SERVICE
ENG = English
GER = Deutsch This field must be set once
S1
Selection of
language
FRA = Francais ENG S1 during start-up. Then you can
ITA = Italiano exit S1 and continue.
NEL = Nederlands Language
ESP = Espanol
S+C
CAL = during S = setup,
S2 Hold configuration calibrating S2
C = calibration.
Setup = during
configuring Auto HOLD
none = no hold
s
S4
Enter hold dwell
period
10 s 10 S4
0 ... 999 s
Cont.Time
Entering an incorrect code
returns you to the measurement
Enter software
menu. The number is edited with
S5
upgrade release
code (Plus
0000 0000 S5 the PLUS or MINUS key and
0000 ... 9999 confirmed with the ENTER key.
package)
PlusCode “1" is displayed if the code is
active.
Entering an incorrect code
returns you to the measurement
Enter software
menu. The number is edited with
S6
upgrade
release code
0000 0000 S6 the PLUS or MINUS key and
0000 ... 9999 confirmed with the ENTER key.
ChemoClean
CleanCode “1" is displayed if the code is
active.
S8
Serial number is
displayed
SerNo S8
12345678
Factory settings are printed in bold face;
base version does not include functions in italic.
50 Endress+Hauser
Liquisys M CPM 223 / 253 Instrument configuration
PM253E5B.CHP
Selection or range
Coding Field Display Info
Factory setting (bold)
Reset of
instrument Facty= All data (except A1 and S1)
(restore default no are erased and reset to
S9 values) Sens = sensor data no S9
the factory settings!
Sens = Last calibration will
Facty= factory
settings
S.Default be erased and reset to the
defaults.
S10
Perform
instrument test
no no S10
displ = display test
Test
Is used for checking the
reference potential. Values
25
mV
Reference voltage S11
S11 current value in mV >50 mV indicate galvanic voltage
is displayed
in the medium. Values >1000 mV
uncomp Ref may falsify the measuring value.
Endress+Hauser 51
Instrument configuration Liquisys M CPM 223 / 253
Function group E
E
E+H SERVICE
E+H SERV
E112
E122
E132
Hardware version
is displayed
xx.xx E112 Cannot be edited.
E142
HW-Vers.
E113
E123
E133
Serial number is
displayed
SerNo E113 Cannot be edited.
E143
12345678
E114
E124
E134
Module name is
displayed
LSG E114 Cannot be edited.
E144
Modul-ID
5.9 Interfaces
Selection or range
Coding Field Display Info
Factory setting
Function group I
I
INTERFACE
INTERFACE
Address
I1 Enter address
HART: 0 ... 15
or
126 I1 Only for communication.
PROFIBUS: 1 ... 126 Address
52 Endress+Hauser
Liquisys M CPM 223 / 253 Instrument configuration
PM253E5B.CHP
Selection or range
Coding Field Display Info
Factory setting
5.10 Calibration
This function group is used to calibrate the Note:
transmitter. Two types of calibration are
possible:
• If the calibration procedure is
• Calibration by means of measurement in aborted by pressing the PLUS and
two calibration solutions with a known pH MINUS keys at the same time
value. (return to C19, C25 or C36) or if
• Calibration by entering daa for slope and the calibration is faulty, then the
zero point previous calibration data is
• For redox measurement by entering the reinstated. A calibration error is
mV value or two different % values. indicated by the “ERR” message
and flashing of the sensor symbol
on the display.
Repeat the calibration!
Endress+Hauser 53
Instrument configuration Liquisys M CPM 223 / 253
Selection or range
Coding Field Display Info
Factory setting (bold)
Only if A1 = pH.
Function group
C ( 1) CALIBRATION: pH calibration C
Calibration with two different
pH calibration
CALIBRAT buffer solutions
°C
C11
Enter calibration
temperature
25.0 °C 25.0 C11 Only if B1 = MTC.
–50.0 ... 150.0 °C
MTC temp.
Enter the pH value Buffer value of the pH The buffer must have a different
C14 of the second last calibration 4.00 C14 pH value to buffer 1. Plausibility
buffer solution pH 0.00 ... 14.00 check takes place.
Buffer 2
Buffer 2: proceed as for buffer 1.
54 Endress+Hauser
Liquisys M CPM 223 / 253 Instrument configuration
PM253E5B.CHP
Selection or range
Coding Field Display Info
Factory setting (bold)
Glass: pH 7.00
pH 5.00 ... 9.00
pH
C17
Zero point (U_is)
is displayed
Antimony: pH 1.00 7.00 C17 The zero point of the IsFET
sensor is displayed in mV.
pH –1.00 ... 3.00
IsFET: present value Zero
–500...+500 mV
C18
Calibration status
is displayed
o.k.
E xxx
o.k. C18
Status
If C18 = E xxx, then only
yes no or new.
C19
Store calibration
result no yes C19 If new, return to C.
new If yes/no, return to
Store “Measurement”.
Function group
CALIBRATION:
C (2) Calibration C Only if A1 = ORP (mV).
Calibration for
Redox mV
redox mV CALIBRAT
The transmitter has a calibrated mV display range. An
absolute mV value with a single buffer solution is entered The maximum calibration offset is
(measuring chain offset adaptation). Use a buffer solution ±100 mV.
preferably with 225 or 475 mV.
mV
C23 Zero is displayed –100 ... 100 mV 0 C23
Zero
C24
Calibration status
is displayed
o.k.
E xxx
o.k. C24
Status
Endress+Hauser 55
Instrument configuration Liquisys M CPM 223 / 253
Selection or range
Coding Field Display Info
Factory setting (bold)
Function group
CALIBRATION: Calibration
C (3) C Only at A1 = ORP (%).
Calibration for Redox %
redox % CALIBRAT
For calibration, a medium sample is drawn and place in two
buckets. The contents of the first bucket are detoxicated. The Default values:
contents of the second bucket remains unchanged. A relative 0 % = –1000 mV
value of 80 % is enterd for the “toxic” sample. The “non-toxic” 100 % = +1000 mV
sample is used for a relative value 20 %.
Start the calibration with the
“toxic” sample by pressing the
Determine 80% %
C31 value of the 80% 80 C31 CAL key. The value is accepted,
as soon as it is stable or is
“toxic” sample 0 ... 100%
Sample 1 confirmed with the CAL key (see
pH calibration).
C35
Calibration status
is displayed
o.k.
E xxx
o.k. C35
Status
If C35 = E xxx, then only
yes no or new.
C36
Store calibration
result no yes C36 If new, return to C.
new If yes/no, return to
Store “Measurement”.
56 Endress+Hauser
Liquisys M CPM 223 / 253 Instrument configuration
PM253E5B.CHP
Selection or range
Coding Field Display Info
Factory setting (bold)
Funktionsgruppe
N NUMERIC N
CALIBRATION
NUMERIC
N1
Enter reference
temperature
25 °C 25°C N1
-50.0 ... 150.0 °C
T°C Ref
Glass 59.18 mV/pH
38.00 ... 65.00 mV/pH
mv/pH
N2 Enter slope
Antimony 59.18 mV/pH
25.00 ... 65.00 mV/pH
59.18 N2 If A5 = IsFET: Enter slope from
quality certificate
IsFET 59.18 mV/pH Slope
38.00 ... 65.00 mV/pH
Glass 7.00 pH
5.00 ... 9.00 pH
pH
N3 Enter zero point Antimony 1.00 pH 7.00 N3 If A5 = IsFET: Enter voltage UIS
from quality certificate
-1.00 ... 3.00 pH
IsFET 0 mV Zero
-500 ... + 500 mV
pH
N4
Calibration status
is displayed
o.k.
o.k. N4
E-
Status
Endress+Hauser 57
Instrument configuration Liquisys M CPM 223 / 253
5.11 Offset
The settings in the OFFSET function group can be used to calibrate the measurement to a reference
measurement. This requires a linear shift of all values measured, i.e. the adjustment is determined
for one measured value, and all others are calculated using the same offset.
Note:
Time t
Selection or range
Coding Field Display Info
Factory setting (bold)
Current measured
0.00 V1
Enter desired value pH
V1 pH –2.00 ... 16.00 Display can be edited.
value
–1500 ... 1500 mV
0.0 ... 100.0 % Real PV
pH 0.00
pH –2.00 ... 2.00
0.00 V2
pH
Current offset is 0 mV
V2
displayed –120 ... 120 mV
0.0 % PV Offset
–50.0 ... 50.0 %
V3
Calibration status
is displayed
o.k.
E xxx
o.k. V3
Status
If V3 = E xxx, then only
yes no or new.
V4
Store calibration
result no yes V4 If new, return to V.
new If yes/no, return to
Store “Measurement”.
58 Endress+Hauser
Liquisys M CPM 223 / 253 Interfaces
PM253E06.CHP
6 Interfaces
For instruments equipped with a communication interface, please refer to separate operating
instructions BA 208C/07/en (HART®) or BA 209C/07/en (PROFIBUS®).
Endress+Hauser 59
Maintenance and troubleshooting Liquisys M CPM 223 / 253
• Maintenance means that all measures which will guarantee the safety of operation and
reliability of the entire measuring system are taken in due time.
Maintenance of CPM 223 / 253 includes:
• Calibration (see chapter 5.9)
• Cleaning of assembly and sensor
• Checking of cables and connections
Troubleshooting:
Determination and elimination of the cause of the problem in the case of an operating fault.
Troubleshooting refers to measures that can be performed without intervention in the instrument
(for instrument defects, see chapter 8, Corrective maintenance).
Troubleshooting of the CPM 223/253 and the measuring system is performed with the aid of the
troubleshooting table in chapter 7.1.
Warning: Note:
60 Endress+Hauser
Liquisys M CPM 223 / 253 Maintenance and troubleshooting
PM253E07.CHP
Endress+Hauser 61
Maintenance and troubleshooting Liquisys M CPM 223 / 253
62 Endress+Hauser
Liquisys M CPM 223 / 253 Maintenance and troubleshooting
PM253E07.CHP
Automatic start
Error Contakt Error current
Display Measures of cleaning
no.
Fact User Fact User Fact User
E001 EEPROM memory error Switch instrument off and back on, yes no — —*
return instrument to your local
Endress+Hauser sales agency for repair yes no — —*
Instrument not calibrated,
or replace instrument.
calibration data invalid,
no user data invalid
Load software compatible with hardware.
E002 (EEPROM error.
Instrument software not
Load measuring parameter specific
suitable for hardware
instrument software.
(controller).
Endress+Hauser 63
Maintenance and troubleshooting Liquisys M CPM 223 / 253
Automatic start
Error Contakt Error current
Display Measures of cleaning
no.
Fact User Fact User Fact User
Repeat calibration and renew buffer yes no — —*
Stability requirement for solution; replace electrode if necessary,
E044
calibration not fulfilled and check instrument and measuring
cable with simulator.
Repeat calibration and renew buffer yes no — —*
solution; replace electrode if necessary,
E045 Calibration aborted
and check instrument and measuring
cable with simulator.
Check measurement and connections; yes no no
Below main parameter
E055 check instrument and measuring cable
measuring range
with simulator if necessary.
Main parameter Check measurement and connections; yes no no
E057 measuring range check instrument and measuring cable
exceeded with simulator if necessary.
Below temperature yes no no
E059
measuring range
Temperature measuring yes no no
E061
range exceeded
Below current output Check configuration in “current output” yes no no
E063
range 1 menu; check measurement and
connections; check instrument and
Current output range 1 yes no no
E064 measuring cable with simulator if
exceeded
necessary.
Below current output yes no no
E065
range 2
Current output range 2 yes no no
E066
exceeded
Set point exceeded Check configuration. yes no no
E067
controller 1
Set point exceeded yes no no
E068
controller 2
Set point exceeded yes no no
E069
controller 3
Set point exceeded yes no no
E070
controller 4
Current output 1 range Expand range in “current output” menu. no no — —*
E080
too small
Current output 2 range Expand range in “current output” menu. no no — —*
E081
too small
PM253E07.CHP
Automatic start
Error Contakt Error current
Display Measures of cleaning
no.
Fact User Fact User Fact User
Below lower alarm If necessary perform manual reference yes no no
E154 threshold for period measurement. Repair sensor and
exceeding alarm delay recalibrate.
Above upper alarm yes no no
E155 threshold for period
exceeding alarm delay
Actual value undershoots yes no no
alarm threshold for longer
E156
than the set permissible
maximum period
Actual value exceeds yes no no
alarm threshold for longer
E157
than the set permissible
maximum period
Check settings in CURRENT INPUT or yes no no
E162 Dosage stop
CHECK function group.
Dynamic range of pH Check cable and sensor. yes no -
E164
convertor exceeded
Dynamic range of Check cable and sensor. yes no -
E165 reference convertor
exceeded
Warning: IsFET leakage Check IsFET for abrasion and tightness, no no no
E168
current > 200 nA replace soon.
Flow in main stream loo Restore flow. yes no no
E171
low or zero
Switch-off limit for current Check process variables at sending yes no no
E172
iput exceeded measuring instrument.
Check process variables at sending yes no no
E173 Current input < 4 mA measuring instrument.
Change range assignment if required.
Check process variables at sending yes no no
E174 Current input > 20 mA measuring instrument.
Change range assignment if required.
Endress+Hauser 65
Diagnosis and corrective maintenance Liquisys M CPM 223 / 253
Corrective maintenance:
• replacement of parts diagnosed to be defective;
• testing of instrument and measuring system function;
• restoration of complete functionality.
Diagnosis based on the error table below and depending on difficulty and measuring
equipment at hand is to be performed by:
• trained operator personnel
• operator’s electricians
• company responsible for system installation/operation
• E+H Service.
Please refer to the tables in chap. 8.2 and 8.3 for identification of spare parts required.
8.1 Diagnosis
The table below will help you diagnose problems and specifies the spare parts required. Please
refer to chap. 8.2.3 and 8.3.3 for information on the exact designations of the spare parts and
their installation.
Error Possible cause Tests and / or remedial measures Equipment, spare parts, personnel
Display dark, no LEDs – No mains voltage Check if mains voltage is available Electrician / e.g. multimeter
active – Wrong supply voltage / voltage Compare mains voltage and rating Operator (utility company
too low on nameplate specification or multimeter)
– Connection fault Terminal not tightened; Electrician
insulation clamped in terminal;
wrong terminal used
– Instrument fuse defective Replace fuse, Electrician /
first compare mains voltage and see drawings in chap. 8.2.1 and
rating on nameplate 8.3.1 for corrected fuse
– Power supply unit defective Replace power supply unit, On-site diagnosis by E+H Service
using correct variant (test module required)
– Controller defective Replace controller, On-site diagnosis by E+H Service
using correct variant (test module required)
– CPM 253: ribbon cable Check ribbon cable, See spare parts for CPM 253
Pos. 310 loose or defetive replace if necessary
Display dark, but LED(s) Controller defective Replace controller On-site diagnosis by E+H Service
active (Module: LSCH/LSCP) (test module required)
66 Endress+Hauser
Liquisys M CPM 223 / 253 Diagnosis and corrective maintenance
PM253E08.CHP
Error Possible cause Tests and / or remedial measures Equipment, spare parts, personnel
Display shows – Instrument or module not CPM 223: Reinstall module Refer to assembly drawings in
measured values but properly installed in instrument CPM 253: Reinstall display module chapters 8.2.1 and 8.3.1
– value does not change – Impermissible operating system Switch instrument off and back on Possible EMC problem: if problem
and / or state persists, call E+H Service to have
– instrument cannot be installation checked
operated
Instrument gets hot – Incorrect voltage / too high Compare mains voltage and rating
of nameplate
– Power supply unit defective Replace power supply unit Can only be diagnosed by E+H
Service
Incorrect meas. pH / – Transmitter module defective Test measuring inputs: If test fails: replace module (using
mV and / or (module: MKP1), please perform – Connectc pH, Ref and PA correct variant), refer to exploded
temperature value tests and take measures directly on instrument with wire views in chap. 8.2.1 and 8.3.1
according to chap. 7.1 jumpers
= display pH 7
– Connect 100 Ω resistance to
terminals 11 / 12 + 13
= display 0 °C
Current output, – Not calibrated correctly Test with built-in current simulation, If simulation value is incorrect:
incorrect current – Load excessive connecting mA meter directly to recalibration at factory or new
current output LSCxx module required.
– Shunt / short-circuit to frame in
current loop
If simulation value is correct:
– Incorrect mode of operation Check whether 0–20 mA or 4–20 check current loop for load and
mA has been selected shunts.
No current output signal – Current output stage defective Test with built-in current simulation, If test fails:
(module: LSCH/LSCP) connecting mA meter directly to Replace controller
current output (using correct variant)
Additional relays do not – CPM 253: ribbon cable Make sure ribbon cable is properly See Spare parts COM 253 F
work Pos. 320 loose or defective connected, replace cable if nec.
Only 2 additional relays – Relay module LSR1-2 with 2 Convert to LSR1-4 with 4 relays Operator or E+H Service
can be used relays installed
Enhanced functions – No or incorrect release code If upgraded: verify that correct Handled by E+H Sales
(S package) not used serial number has been used to
available order S package
– Incorrect serial number of Check whether serial number on Serial number of instrument is
instrument stored in LSCH/LSCP) nameplate matches serial number required to enable S package.
module of LSCH/LSCP (field S8).
Enhanced functions – Serial number entered for For LSCH / LSCP with SNR 0000 A detailed description can be
(S package and / or replacement module LSCH or an instrument serial number can found in chap. 8.3.5.
Chemoclean) are not LSCP at factory is 0000. be entered once in fields E114 to
available after S package or Chemoclean has E116.
replacement of not been enabled with release Next step is to enter release codes
LSCH-/LSCP module codes. for S package and/or Chemoclean.
No HART or PROFIBUS – Wrong controller HART: LSCH-H1 or -H2 Replace controller;
interface function PROFIBUS-PA/-DP: LSCP-PA/-DP operator or E+H Service
module, see field E112
– Wrong software SW version see field E111
– Bus problem Remove some instruments and Contact E+H Service for support
repeat testing
Endress+Hauser 67
Diagnosis and corrective maintenance Liquisys M CPM 223 / 253
Pos. 230
pH transmitter MKP1
Pos. 10 / 20
Power supply unit LSGA / LSGD
Pos. 40 – 80
Controller LSCH / LSCP
Pos. 330
End frame Housing
latches assembly
Pos. 320
Back panel coupling
Pos. 310
Side panel for
equipment
without relay
module
Pos. 430
Push-on
terminal strip for
LSR1-x module
Module rotated
180° for
representation.
Pos.3030/ jumper
Item / Steckbrücke
Fuse
Sicherung
230 V
Netzspannungs-
Mains voltage
wahl
selection
115 V
100 V
68 Endress+Hauser
Liquisys M CPM 223 / 253 Diagnosis and corrective maintenance
PM253E08.CHP
Endress+Hauser 69
Diagnosis and corrective maintenance Liquisys M CPM 223 / 253
Back panel
340 Back panel
for PROFIBUS-DP
70 Endress+Hauser
Liquisys M CPM 223 / 253 Diagnosis and corrective maintenance
PM253E08.CHP
Pos. 10
Pos.
Netzteil 100 10 V AC
– 230
Power supply unit 100 - 230 V AC
Pos. 450a
Torx screws K4x10
Pos. 460a Pos. 310 / 320
Ribbon cables Pos. 400
Cover
Bottom of housing
screws
Pos. 230
Transmitter Top
module MKP1 Bottom
Pos. 430
Hinge
Pos. 420
Connection
compartment
cover Pos. 10 / 20
Power supply Pos. 999
module pH Terminal Pos. 290 / 300
LSGA / LSGD compartment Relay module
module LSR1-x
Pos. 340
Docking module
LSK1
Pos. 480
Cable glands
Endress+Hauser 71
Diagnosis and corrective maintenance Liquisys M CPM 223 / 253
72 Endress+Hauser
Liquisys M CPM 223 / 253 Diagnosis and corrective maintenance
PM253E08.CHP
Endress+Hauser 73
Diagnosis and corrective maintenance Liquisys M CPM 223 / 253
74 Endress+Hauser
Liquisys M CPM 223 / 253 Diagnosis and corrective maintenance
PM253E08.CHP
Clean soilings on the glass electrodes as • Coatings containing protein (food industry)
follows: Use mixture of hydrochlorid acid (0.5 %)
and pepsin (1 %), then rinse thoroughly.
• Clean oily and greasy coatings
Clean with detergent (fat solvent, Blocked diaphragms can under certain
e.g. alcohol, acetone, poss. detergent). circumstances be mechanically cleaned
(does not apply for teflon diaphragms and
Warning: open ring junction electrodes):
• Use a small key file
Hands, eyes and clothes are to be • File in one direction only
protected when using the cleaning Air bubbles in the electrode
agents described below. • Air bubbles may indicate an assembly
problem; check installation position
• Limestone deposits or metal hydroxide • The range from 15° to 165° from the
coatings horizontal is permissible
Loosen coatings with diluted hydrochlorid • Horizontal installation or installation with the
acid (3 %), then rinse thoroughly. plug-in head pointing down is not allowed
8.6.3 Assembly
Please refer to the operating instructions for assembly and dismantling, sensor and gasket
the assembly in question for maintenance and replacement as well as information on
troubleshooting instructions. The operating resistance, spare parts and accessories.
instructions also include instructions for
Endress+Hauser 75
Diagnosis and corrective maintenance Liquisys M CPM 223 / 253
76 Endress+Hauser
Liquisys M CPM 223 / 253 Accessories
PM253E09.CHP
9 Accessories
Mounting accessories Connection accessories
• Weather protection cover CYY 101 Technical information about all cable types:
For mounting on field housing, TI 118C/07/en
for outdoor installation. Order no.: 50068526
Dimensions (H × W × D):
320 × 300 × 270 mm • CPK1
Material: Stainless steel SS 304 Measuring cable with auxiliary core and
Order no.: CYY 101-A additional outer screen, in PVC sheath,
• Universal upright post CYY 102 ∅ 7.2 mm; extension with cable CYK 71
Square tube for mounting of
field housing. • CPK 2
Dimensions (H × W × D): Coaxial measuring cable with six auxiliary
1495 × 60 × 60 mm cores and additional outer screen, in PVC
Material: Stainless steel SS 304 sheath, ∅ 12 mm; extension with cable
Order no.: CYY 102-A PMK
Electrodes • CPK 9
Coaxial measuring cable with four auxiliary
• OrbiSint W CPS 11/12/13 cores and additional outer screen, in PVC
pH / redox combination electrode sheath, ∅ 7.2 mm; with waterproof TOP 68
with PTFE diaphragms, with gel filling, plug-in head, IP68 ingress protection,
applicable for up to 6 bar. extension with cable CYK 71
Technical Information TI 028C/07/en
Order no.: 50054649
• CPK 12
Coaxial measuring cable with four auxiliary
• OrbiTex W CPS 21 cores and additional outer screen, in PVC
pH combination electrode
with open ring junction, with gel filling, sheath, ∅ 7.2 mm; for IsFET sensors, with
applicable for up to 6 bar. waterproof TOP68 plug-in head, IP 68
Technical Information TI 029C/07/de ingress protection, extension with cable
Order no.: 50054650 CYK 71
Endress+Hauser 77
Accessories Liquisys M CPM 223 / 253
Buffer solutions
• CPY 3
for redox calibration
+220 mV (pH 7,0)
+ 468 mV (pH 0,0)
Content: 100 / 5000 ml
• CPY 4
KCl electrolyte solution
for liquid filled electrodes
3 mol (-10 °C ... 100 °C)
1.5 mol (-30 °C ... 100 °C)
Content: 100 / 1000 ml
78 Endress+Hauser
Liquisys M CPM 223 / 253 Technical data % °C
Ω
PM253E10.CHP
10 Technical data
pH measurement
Measuring range pH –2 ... 16
Display range pH –2 ... 16
with indication of measuring range violation
pH offset range ±pH 2
Slope Glass: 38.00 ... 65.00 mV/pH (nominal 59.16 mV/pH)
Antimon: 25.00 ... 65.00 mV/pH (nominal 59.16 mV/pH)
IsFET: 38.00 ... 65.00 mV/pH (nominal 59.16 mV/pH)
Zero point Glass: pH 5.00 ... 9.00 (nominal pH 7.00)
Antimon: pH –1.00 ... 3.00 (nominal pH 1.00)
IsFET: –500 ... +500 mV
Terminal cross-section 2.5 mm2
Required conductor cross-section CPM 223 0.75 mm2 / AWG 18
CPM 253 1.50 mm2 / AWG 14
Maximum cable length to pH electrode 50 m
pH signal input
Input resistance for nominal operating conditions > 1 × 1012 Ω
Redox measurement
Display and measuring range –1500 ... +1500 mV / 0 ... 100 %
Redox offset range ±120 mV / ±50 %
Terminal cross-section 2.5 mm2
Required conductor cross-section CPM 223 0.75 mm2 / AWG 18
CPM 253 1.50 mm2 / AWG 14
Maximum cable length to pH electrode 50 m
Temperature measurement
Temperature sensor Pt 100, Pt 1000, NTC 30K
Measuring range Pt 100, Pt 1000 –50 ... +150 °C
NTC 30K –20 ... +100 °C
Temperature offset range ±5 °C
Current input
Current range 4 ... 20 mA, galvanically separated
Load 260 Ω at 20 mA (voltage drop 5.2 V)
Endress+Hauser 79
% °C Technical data Liquisys M CPM 223 / 253
Ω
Limit contactor
Pickup / dropout delay 0 ... 2000 s
Controller
Function (adjustable) pulse length / pulse frequency controller
Controller response PID
Control gain Kp 0.01 ... 20.00
Integral action time Tn 0.0 ... 999.9 min
Derivative action time Tv 0.0 ... 999.9 min
Period for pulse-length controller 0.5 ... 999.9 s
Frequency for pulse frequency controller 60 ... 180 min–1
Basic load 0 ... 40% of max. set value
Alarm
Function (switchable) steady / fleeting contact
Alarm threshold adjustment range pH / temperature: entire measuring range
Alarm delay 0 ... 2000 s (min)
80 Endress+Hauser
Liquisys M CPM 223 / 253 Technical data % °C
Ω
PM253E10.CHP
Accuracy pH measurement
Reference temperature +25 °C
Measured value resolution 0.01 pH
Display deviation1 max. 0.5 % of measuring range
Repeatability1 max. 0.2 % of measuring range
1
Measurement deviation pH signal output max. 0.75 % of measuring range
Redox measurement
Measured value resolution 1 mV / 0.1 %
1
Display deviation max. 0.5 % of measuring range
Repeatability1 max. 0.2 % of measuring range
Measurement deviation1 redox signal output max. 0.75 % of measuring range
Temperature measurement
Measured value resolution. 01 °C
Display deviation1 max. 1.0 % of measuring range
Measurement deviation1 temperature signal output max. 1.25 % of current output range
Ambient conditions Ambient temperature (nominal operating conditions) –10 ... +55 °C
Ambient temperature (limit operating conditions) –20 ... +60 °C
Storage and transportation temperature –25 ... +65 °C
Relative humidity (nominal operating conditions) 10 ... 95 %, non-condensing
Ingress protection of panel-mounted instrument IP 54 (front panel), IP 30 (enclosure)
Ingress protection of field instrument IP 65
Pollution degree 2 acc. to IEC 61010-1
Installation category II
Maximum altitude 2000 m above sea level
Electromagnetic compatibility Interference emission and interference immunity acc. to
EN 613261:1998 / A1:1998
Materials
Housing of panel-mounted unit Polycarbonate
Front membrane Polyester, UV-resistant
Field instrument standard versions / CSA GP version ABS PC Fr / Polycarbonate
Power supply Supply voltage 100 / 115 / 230 V AC +10 / –15 %, 48 ... 62 Hz
24 V AC/DC +20 / –15 %
Power consumption max. 7.5 VA
Mains protection microfuse, medium slow 250 V / 3.15 A
1
According to IEC 746-1, for nominal operating conditions
Subject to modification.
Endress+Hauser 81
Appendix Liquisys M CPM 223 / 253
11 Appendix
Function group Entry of absolute value Current offset is Calibration status Offset results
OFFSET displayed is displayed
current measured value
0.00 pH, -2.00...2.00 pH o.k. yes; no; new
-2.00...16 pH
0 mV, -120...120 mV E--
-1500...1500 mV
V 0.0...100.0 % V1 0.0 %, -50.0...50.0 % V2 V3 V4
Function group Enter reference Enter slope Enter zero point Calibration status Store calibration
temperature Glass 59.18 mV/pH Glass 7.00 pH is displayed results
NUMERIC 5.00… 9.00 pH
38.00… 65.00 mV/pH
CALIBRATION Antimon 59.16 mV/ pH Antimon 1.00 pH o.k. yes; no; new
25 °C –1.00 … 3.00 pH
25.00 … 65.00 pH E---
-20.0...150.0 °C IsFET
N N1 IsFET N2 N3 N4 N4
Calibration of 80% Calibration Calibration of 20% Calibration Calibration status Store calibration
Function group is displayed
CALIBRATION value (toxic sample) value (non-toxic results
Acceptance when stable sample) Acceptance when stable
-1500...1500 mV ≤ ± 5 mV for more than 5 s ≤ ± 5 mV for more than 5 s o.k. yes; no; new
-1500...1500 mV E--
Redox %
calibration C31 C32 C33 C34 C35 C36
Enter value of Calibration Zero point is displayed Calibration status Store calibration
redox buffer is displayed results
Redox mV Acceptance when stable -100...100 mV
calibration current measured value ≤ ± 1 mV for more than 5 s o.k. yes; no; new
-1500 mV ... 1500 mV E--
C21 C22 C23 C24 C25
pH
Enter calibration Entry of pH value of Calibration Entry of pH value of Calibration Display of slope
calibration
temperature first buffer solution second buffer solution Glass 59.16 mV/pH
C 38.00...65.00 mV/pH
(calibration type options (if B3 = MTC) Acceptance when stable Acceptance when stable:
Buffer value from last Antimon 59.16 mV/pH
displayed depend on Buffer value from last ≤ ± 0,05 pH for more ≤ ± 0,05 pH for more 25.00...65.00 mV/pH
selection in field A1) 25.0°C calibration than 10 s calibration than 10 s
0.00 pH...14.00 pH IsFET 59.16 mV/pH
-20.0...150.0°C C11 0.00 pH...14.00 pH C12 C13 C14 C15 38.00 … 65.00 mV/pH C16
any code
Read mode:
code 22
Edit mode:
Temperature display Temperature display Measured value display Measured value display Measured value display
MEAS. VALUE DISPLAY in °F suppressed in mV
Current input in % Current input in mA
with
Function group Selection of Selection of display unit Selection of connection Entry of measured Selection of electrode Selection of
operation mode type value damping temperature sensor
SETUP 1 Pt 100
pH Glass (E0 = 7,0)
pH; ORP (mV); ORP sym = symmetrical 1 (no damping) Antim = Antimon Pt 1k
A ORP (%) A1 A2 asym = asymmetrical A3 1-60 A4 IsFET A5 NTC 30K A6
Function group Temperature compensation Entry of MTC Temperature compensation Entry of correct Display of
SETUP 2 selection (for process) temperature selection process temperature temperature difference
(if B1=MTC (for calibration) (if B1=ATC) (offset)
pH: Redox: and A1=pH)
ATC; on ATC; MTC 25.0°C 0.0°C
B MTC off B1 25.0 °C B2 B3 -50.0°C ... +150.0°C -5.0...5.0 °C B5
B4
-50...+150°C
Function group Cont. switch-off Delay for cont. Delay for cont. Switch-off limit value Switch-off direction Feedforward control
CURRENT INPUT by current input switch-off current input switch-on current input for current input for current input to PID controller
Off; Input 0s 0 ... 2000 s 0s 0 ... 2000 s 50% 0 ... 100% Low; High Off; lin = linear
Z Z1 Z2 Z3 Z4 Z5 Z6
O321
Function group Current output selection Current output selection Current range selection 0/4 mA value entry 20 mA value entry
CURRENT OUTPUT
Out1; Out2; Out1; Out2; 4-20 mA; 0-20 mA +2.00 pH; -2.00...16.00 pH 12.00 pH; -2.00...16.00 pH
lin = linear O3 (1)
-1500 mV; -1500...1500 mV 1500 mV; -1500...1500 mV
0.0 %; 0.0...100.0 % 100.0 %; 0.0...100 %
O 0.0°C; -50...150.0°C 100.0°C; -50...150.0°C
O1 O2 O311 O312 O313
Function group Select contact type Select alarm Alarm delay Error current Error number Set alarm contact
ALARM delay unit setting selection to be effective
Stead = steady contact; 0 s (min)
Fleet = fleeting contact s; min 0 s... 2000 s (min) 22 mA; 1 yes; no
F F1 F2 F3 F6
(depends on F2) 2,4 mA F4 1... 255 F5
82 Endress+Hauser
Liquisys M CPM 223 / 253 Appendix
PM253E11.CHP
Feedforward control =1
at
50% 0 … 100%
Z7
Endress+Hauser 83
Appendix Liquisys M CPM 223 / 253
Function group SCS alarm SCS alarm for SCS alarm threshold Display of Select process Alarm delay
CHECK measuring electrode reference electrode leakage current monitoring
(if A2=sym) (only with IsFET sensor)
50 kΩ 0,0 … 9,9 µA Off; Low; High; 0 min (s)
P off; on P1 off; on P2 1.5...50 kΩ P3 P4 LoHi; Lo!; Hi!; LoHi! P5 0 … 2000 min (s) P6
Limit contactor Function R2 (6) Set point 1 (or 2) Entry of control Entry of integral action time
configuration Switch off or on gain Kp1 (or Kp2) Tn (0.0 = no I component)
6.00 pH
Neutr = R2 (6) off; on -2.00...16.00 pH 1.00 0.0 min
neutralisation controller 0.01...20.00 0.0...999.9 min
(only with rel1 and rel2 R261 R262 R263 R264
and with A1 = pH)
Function R2 (5) Start pulse selection Entry of pre-rinse time Entry of cleaning time
Switch off or on int = internal;
ext = external; 30 s 10 s
Clean = R2 (5) off; on i+ext = internal +external; 0...999 s 0...999 s
Chemoclean i+stp = internal,
(only with rel3) R251 suppressed by ext R252 R253 R254
Function R2 (4) Rinse time setting Pause time setting Entry of minimum
Switch off or on pause time
30 s 360 min
Timer R2 (4) off; on 0...999 s 1...7200 min 120 min
1...3600 min
R241 R242 R243 R244
Function group Select contact to be Function R2 (1) Select contact switch-on Select contact switch-off Pickup delay
RELAY configurd LC PV = R2 (1) Switch off or on point point entry
pH/redox
Rel1; Rel2; Rel3; Rel4; limit contactor off; on 16.00 pH; -2.00...16.00 pH pH 16.00; pH -2.00...16.00 0s
1500 mV; -1500...1500 mV 1500 mV; -1500...1500 mV 0...2000 s
100.0 %; 0...100.0 % 100.0 %; 0...100.0 %
R R1 R211 R212 R213 R214
Function group Language selection Hold configuration Manual hold Entry of hold Entry of SW upgrade Entry of SW upgrade
SERVICE s+c = during setup dwell period release code release code
ENG; GER and calibration off; on (Plus package) Chemoclean
ITA; FRA CAL = during calibration 10 s
ESP; NEL Setup = during setup 0...999 s 0000 0000
S S1 none = no hold S2 S3 S4 0000...9999 S5 0000...9999 S6
84 Endress+Hauser
Liquisys M CPM 223 / 253 Appendix
PM253E11.CHP
Set lower Set upper Select process Set max. perm. period Set max. perm. period Set monitoring
alarm threshold monitoring for lower limit exceeded for upper limit exceeded value
alarm threshold
Off; AC; CC; AC CC 60 min 0 ... 2000 min 120 min 0 ... 2000 min pH 1.00 pH -2 ... 16
pH -2.00 pH -2 ... 16 P7 pH 16.00 pH -2 ... 16 P8 AC!; CC!; ACCC! P9 P10 P11 P12
Entry of derivative action Selection Entry of Entry of max. Entry of min. ON time Enter process type
time Tv pulse interval pulse frequency ton
(0.0 = no D component) len = pulse length 120 1/min
freq = pulse frequency 10.0 s 60...180 1/min 0.3 s Batch
0.0 min 0.5...999.9 s 0.1...5.0 s Inlne
0.0...999.9 min R266
R265 R267 R268 R269 R2610
Entry of post-rinse time Number of repeat cycles Set interval between Entry of minimum Number of cleaning
two cleaning cycles pause time cycles without
20 s 0 (pause time) cleaning agent
0...999 s 0...5 120 min
360 min 1...R357 min 0
R255 R256 1...7200 min 0...9
R257 R258 R259
Entry of derivative action Selection of control Selection Entry of Entry of max. Entry of min. ON time Enter basic load Enter process type
time Tv characteristic pulse interval pulse frequency ton
(0.0 = no D component) Ien = pulse length
dir = direct; freq = pulse frequency 10.0 s 120 1/min 0.3 s 0% Batch
0.0 min inv = inverted; 0.5...999.9 s 60...180 1/min 0.1...5.0 s 0 … 40% Inlne
0.0...999.9 min R236 R237
R235 R238 R239 R2310 R2311 R2312
0s 150.0 °C
0...2000 s -20.0...+150.0 °C
R225 R226
Order number Serial number Reset intrument Perform instrument test Reference voltage Select
is displayed is displayed (restore default values) is displayed AC frequency
no; no; display
Sens = sensor data;
Facty = factory settings
S7 S8 S9 S10 S11 S12
Endress+Hauser 85
Index Liquisys M CPM 223 / 253
12 Index
! D
0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18, 21 Damage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
22 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 Declaration of conformity . . . . . . . . . . . . . . . 5
9999 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18, 21 Definition of terms . . . . . . . . . . . . . . . . . . . . 66
Derivative action time . . . . . . . . . . . . . . . . . 45
A Diagnosis. . . . . . . . . . . . . . . . . . . . . . . . . 66-76
A-Function group . . . . . . . . . . . . . . . . . . . . . 25 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Access codes . . . . . . . . . . . . . . . . . . . . . . . . 21 Disassembly of CPM 223 . . . . . . . . . . . . . . 69
Accessories . . . . . . . . . . . . . . . . . . . . . . 77, 78 Disassembly of CPM 253 . . . . . . . . . . . . . . 72
Actuating signal outputs . . . . . . . . . . . . . . . . 40 Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
ALARM (P) . . . . . . . . . . . . . . . . . . . . . . . . . . 33 Disposal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Alarm contact . . . . . . . . . . . . . . . . . . 24, 33, 34 Dropout delay . . . . . . . . . . . . . . . . . . . . 43, 44
Alarm delay . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Alarm threshold. . . . . . . . . . . . . . . . . . . . 44, 45 E
Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . 78 E+H Service. . . . . . . . . . . . . . . . . . . . . . . . . 52
Assembly of CPM 223 . . . . . . . . . . . . . . . . . 69 E-Function group. . . . . . . . . . . . . . . . . . . . . 52
Assembly of CPM 253 . . . . . . . . . . . . . . . . . 72 Electrical connection . . . . . . . . . . . . . . . . . . 12
Asymmetrical . . . . . . . . . . . . . . . . . . . . . . . . 15 Electrode installation . . . . . . . . . . . . . . . . . . 14
AUTO key . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 Electrodes . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Electronics box . . . . . . . . . . . . . . . . . . . . . . . 7
B Entry of hold dwell period . . . . . . . . . . . . . . 50
B-Function group . . . . . . . . . . . . . . . . . . . . . 26 Error codes . . . . . . . . . . . . . . . . . . . . . . . . . 63
Basic load . . . . . . . . . . . . . . . . . . . . . . . . . . . 46 Error current. . . . . . . . . . . . . . . . . . . . . . 33, 34
C F
C-Function group . . . . . . . . . . . . . . . 54, 55, 56 F-Function group . . . . . . . . . . . . . . . . . . 33, 34
Calibration. . . . . . . . . . . . . . . . . . . . . . . . 53, 54 Factory settings . . . . . . . . . . . . . . . . . . . . . . 24
CALIBRATION (C) . . . . . . . . . . . . . . . . . . . . 23 Freezing of outputs . . . . . . . . . . . . . . . . . . . 22
Calibration mode . . . . . . . . . . . . . . . . . . 20, 23 Function coding. . . . . . . . . . . . . . . . . . . . . . 23
Characteristic controller . . . . . . . . . . . . . . . . 46 Function group . . . . . . . . . . . . . . . . . . . . . . 21
CHECK (P) . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Chemoclean function . . . . . . . . . . . . . . . 38, 42 G
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66 General . . . . . . . . . . . . . . . . . . . . . . . . . . . 2, 3
Cleaning function . . . . . . . . . . . . . . . . . . . . . 47 General safety notes . . . . . . . . . . . . . . . . . . . 4
Cleaning time . . . . . . . . . . . . . . . . . . . . . 42, 48
Code 0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 H
Code 22 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 Hardware version . . . . . . . . . . . . . . . . . . . . 52
Code 9999 . . . . . . . . . . . . . . . . . . . . . . . . . . 18 HART . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Coding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 Hazards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Complete Installation . . . . . . . . . . . . . . . . . . . 6 Hold configuration . . . . . . . . . . . . . . . . . . . . 50
Complete measuring system . . . . . . . . . . . . . 6 Hold dwell period . . . . . . . . . . . . . . . . . . . . 42
Connection diagram . . . . . . . . . . . . . . . . . . . 12 Hold function . . . . . . . . . . . . . . . . . . . . . . . . 22
Connection examples . . . . . . . . . . . . . . . . . . 15
Connections of field instrument . . . . . . . . . . 13 I
Contact selection . . . . . . . . . . . . . . . . . . . . . 33 I-Function group . . . . . . . . . . . . . . . . . . . . . 52
Control characteristic . . . . . . . . . . . . . . . . . . 41 Immunity to interference . . . . . . . . . . . . . . . . 5
CPK 1, CPK 2, CPK 7 . . . . . . . . . . . . . . . . . . 14 Installation . . . . . . . . . . . . . . . . . . . . . . . 4, 6-16
Current input . . . . . . . . . . . . . . . . . . . . . . . . . 27 Instrument configuration . . . . . . . . . . . . . 23-37
CURRENT INPUT (Z) . . . . . . . . . . . . . . . . . . 29 Instrument upgrade. . . . . . . . . . . . . . . . . . . 78
CURRENT OUTPUT (O) . . . . . . . . . . . . . 23, 31 Integral action time . . . . . . . . . . . . . . . . . . . 45
Current output table . . . . . . . . . . . . . . . . . . . 32 Intended application . . . . . . . . . . . . . . . . . . . 4
Current outputs . . . . . . . . . . . . . . . . . . . . . . . 30 INTERFACE (I) . . . . . . . . . . . . . . . . . . . . . . . 52
Interfaces. . . . . . . . . . . . . . . . . . . . . . . . . . . 59
K
Key functions . . . . . . . . . . . . . . . . . . . . . . . . 18
Kp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
86 Endress+Hauser
Liquisys M CPM 223 / 253 Index
PM253E12.CHP
L P
LED indicator . . . . . . . . . . . . . . . . . . . . . . . . 17 P controller . . . . . . . . . . . . . . . . . . . . . . . . . 39
Limit contactor . . . . . . . . . . . . . . . . . . . . 38, 43 P(ID) controller . . . . . . . . . . . . . . . . . . . 38, 39
Limit contactor for conductivity measured P-Function group. . . . . . . . . . . . . . . . . . . . . 36
value. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 Packaging . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Limit contactor for temperature . . . . . . . 38, 44 Panel-mounted instrument . . . . . . . . . . . . . 11
Liquid crystal display . . . . . . . . . . . . . . . . . . 17 Partition plate. . . . . . . . . . . . . . . . . . . . . . . . . 7
Locking the keypad . . . . . . . . . . . . . . . . 18, 21 Pause time. . . . . . . . . . . . . . . . . . . . . . . 47, 48
PCS alarm . . . . . . . . . . . . . . . . . . . . . . . . . . 35
M PD controller . . . . . . . . . . . . . . . . . . . . . . . . 39
Maintenance and troubleshooting . . . . . 60-65 PI controller . . . . . . . . . . . . . . . . . . . . . . . . . 39
Maintenance of CPM 223 . . . . . . . . . . . . . . 68 Pickup delay . . . . . . . . . . . . . . . . . . . . . 43, 44
Maintenance of CPM 253 . . . . . . . . . . . . . . 71 PID controller. . . . . . . . . . . . . . . . . . . . . 39, 45
Maintenance of measuring system . . . . . . . 75 PLUS key . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Manual Hold . . . . . . . . . . . . . . . . . . . . . . . . . 50 Post mounting kit . . . . . . . . . . . . . . . . . . . . . . 9
Measuring cable . . . . . . . . . . . . . . . . . . . . . 14 Post mounting on a square post . . . . . . . . . . 8
Measuring cable connection . . . . . . . . . 14, 16 Post mounting on cylindrical pipes. . . . . . . . 8
Measuring mode . . . . . . . . . . . . . . . . . . . . . 23 Post-rinse time. . . . . . . . . . . . . . . . . . . . . . . 42
Measuring system . . . . . . . . . . . . . . . . . . . . . 6 Potential matching connection . . . . . . . . . . 15
Menu structure . . . . . . . . . . . . . . . . . . . . . . . 21 Pre-rinse time . . . . . . . . . . . . . . . . . . . . 42, 48
Mimimum pause time. . . . . . . . . . . . . . . . . . 48 Product structure. . . . . . . . . . . . . . . . . . . . . . 3
Minimum pause time . . . . . . . . . . . . . . . . . . 47 PROFIBUS . . . . . . . . . . . . . . . . . . . . . . . . . . 52
MINUS key . . . . . . . . . . . . . . . . . . . . . . . . . . 18 Pulse frequency. . . . . . . . . . . . . . . . . . . 46, 49
Module name . . . . . . . . . . . . . . . . . . . . . . . . 52 Pulse frequency modulation . . . . . . . . . . . . 40
Monitoring features . . . . . . . . . . . . . . . . . . . . 5 Pulse length. . . . . . . . . . . . . . . . . . . . . . 46, 49
Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Pulse length modulation . . . . . . . . . . . . . . . 40
Mounting examples . . . . . . . . . . . . . . . . . . . 10
Mounting of field instrument . . . . . . . . . . . . . 8 Q
Questions . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
N
N-Function group . . . . . . . . . . . . . . . . . . . . . 57 R
Neutralisation controller . . . . . . . . . . . . . 42, 48 R-Function group. . . . . . . . . . . . . . . . . . . 43-49
Numeric calibration . . . . . . . . . . . . . . . . . . . 57 REL key . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
NUMERIC CALIBRATION (N) . . . . . . . . . . . 57 RELAY (R) . . . . . . . . . . . . . . . . . . . . . . . 23, 43
Relay contact configuration . . . . . . . . . . . . 38
O Repeat cycles . . . . . . . . . . . . . . . . . . . . . . . 48
O-Function group. . . . . . . . . . . . . . . . . . 31, 32 Replacement of processor module. . . . . . . 74
Offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 Rinse time . . . . . . . . . . . . . . . . . . . . . . . 47, 48
Offset mode . . . . . . . . . . . . . . . . . . . . . . . . . 23
ON time . . . . . . . . . . . . . . . . . . . . . . . . . 46, 49
Operating concept. . . . . . . . . . . . . . . . . . . . 20
Operating modes . . . . . . . . . . . . . . . . . . . . . 20
Operation . . . . . . . . . . . . . . . . . . . . . . 4, 17-22
Operator interface . . . . . . . . . . . . . . . . . . . . 17
Optoscope . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Order number . . . . . . . . . . . . . . . . . . . . . . . 50
Endress+Hauser 87
Index Liquisys M CPM 223 / 253
S W
S-Function group . . . . . . . . . . . . . . . . . . . . . 50 Wall mounting . . . . . . . . . . . . . . . . . . . . . . . . 8
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4, 5 Weather protection cover CYY 101. . . . . 8, 77
Safety features . . . . . . . . . . . . . . . . . . . . . . . . 5
safety notes. . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Z
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . 2 Z-Function group . . . . . . . . . . . . . . . . . . . . . 29
Scopeware . . . . . . . . . . . . . . . . . . . . . . . . . . 74
SCS electrode monitoring. . . . . . . . . . . . . . . 35
Selection of error. . . . . . . . . . . . . . . . . . . . . . 33
Selection of language . . . . . . . . . . . . . . . . . . 50
Self-test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Serial number . . . . . . . . . . . . . . . . . . . . . . . . 52
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
SERVICE (S) . . . . . . . . . . . . . . . . . . . . . . 25, 50
SETUP 1 (A) . . . . . . . . . . . . . . . . . . . . . . . . . 25
SETUP 2 (B) . . . . . . . . . . . . . . . . . . . . . . . . . 25
Setup mode. . . . . . . . . . . . . . . . . . . . . . . . . . 23
Setup-Modus . . . . . . . . . . . . . . . . . . . . . . . . 20
Shipping documents. . . . . . . . . . . . . . . . . . . . 2
Simulation . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Software version instrument . . . . . . . . . . . . . 52
Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Spare parts for CPM 223 . . . . . . . . . . . . . . . 70
Spare parts for CPM 253 . . . . . . . . . . . . . . . 73
Spare parts orders . . . . . . . . . . . . . . . . . . . . 74
Start pulse . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Start-up menu . . . . . . . . . . . . . . . . . . . . . . . . 25
Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Structure of coding . . . . . . . . . . . . . . . . . . . . 23
Structure of measuring cables . . . . . . . . . . . 14
SW upgrade . . . . . . . . . . . . . . . . . . . . . . . . . 50
Switch-off point of contact . . . . . . . . . . . . . . 43
Switch-off temperature . . . . . . . . . . . . . . . . . 44
Switch-on point of contact . . . . . . . . . . . . . . 43
Switch-on temperature . . . . . . . . . . . . . . . . . 44
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Symbols used . . . . . . . . . . . . . . . . . . . . . . . . . 2
Symmetrical . . . . . . . . . . . . . . . . . . . . . . . . . 15
System configuration . . . . . . . . . . . . . . . . . . 25
T
Technical data . . . . . . . . . . . . . . . . . . . . 79-81
Terminal blocks. . . . . . . . . . . . . . . . . . . . . . . . 7
Termination of measuring cables . . . . . . . . . 14
Timer for cleaning function . . . . . . . . . . . 38, 41
Tn . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Transport. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Troubleshooting common problems . . . . . . 60
Tv . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
U
Universal mounting post CYY 102 . . . . . . . . . 9
Unlocking the keypad. . . . . . . . . . . . . . . 18, 21
Unpacking. . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Upright post CYY 102 . . . . . . . . . . . . . . . . . . 77
V
V-Function group . . . . . . . . . . . . . . . . . . . . . 58
VBA junction box . . . . . . . . . . . . . . . . . . 14, 77
VBM junction box . . . . . . . . . . . . . . . . . . 14, 77
88 Endress+Hauser
Declaration of contamination
Dear customer,
Because of legal determinations and for the safety of our employes and operating equipment we need this
“Declaration of contamination” with your signature before your order can be handled. Please put the completely
filled in declaration to the instrument and to the shipping documents in any case. Add also safety sheets and/or
specific handling instructions if necessary.
:DUQLQJKLQWVIRUPHGLXPXVHG
SAFE
o o o o o o o o
radioactive explosive caustic poisonous harmful of biological inflammable safe
health hazardous
Please mark the appropriate warning hints.
5HDVRQIRUUHWXUQ
BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB
BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB
&RPSDQ\GDWD
I hereby certify that the returned equipment has been cleaned and decontaminated acc. to good industrial prac-
tices and is in compliance with all regulations. This equipment poses no health or safety risks due to contamination.
_______________________________ ___________________________________
(Date) (company stamp and legally binding signature)