Professional Documents
Culture Documents
Drystar 5300 Software Patch 2.1.1 - C1 Installation Instructions
Drystar 5300 Software Patch 2.1.1 - C1 Installation Instructions
2008
# Topic Refer to ….
1 Software Update via CF card Section 1
2 Software Update via browser Section 2
3 Solved bugs in SW 2.1.1_C1 Section 3
REFERENCES:
• Drystar 5300 Service Bulletin no. 13, DD+DIS388.07E:
Sporadic “Error System Failure” after Switch-off: Install SW Patch 2.1.1_C1
PREREQUISITE:
• Software 2.1.0_C1 has to be installed on the printer.
REQUIRED TIME:
• 30 min. for SW update 2.1.1_C1
TOOLS:
• Service PC to copy software patch to CF card
IMPORTANT:
Before starting any repair or software installation activities ……
(1) Confirm that the printer queue is empty
(2) Remove the network cable, to ensure, that nobody sends print-jobs to the printer
during your work.
Page 1 of 3
1 Software Update via CF Card
NOTE:
Preferably use the CF card which came with the printer or a CF card from Agfa stock.
For other CF cards it cannot be guaranteed that they can be read by the printer.
(1) Delete all files on the CF card except previous backup files if available
(2) Copy file SWR_DRYSTAR5300_R2.1.1_C01.zip to the root of the CF card (this means,
not in any sub-directory)
(3) In the key-operator menu select '9 Installation – 1 Installation from CF card'
(4) Follow the further instructions on the local display.
(5) When finished, re-connect the network cable
(6) Print an internal test image and a test image from the modality
(7) Check image quality at the lightbox
Page 2 of 3
3 Solved bugs in Software Rel. 2.1.1_C1
Symptom Sporadically, after switch-off the printer hangs with message “Error system failure”.
Unplugging and reconnecting the power cord solves the problem. The printer boots up normally
afterwards.
This problem may appear at all printers with SW version < 2.1.1_C1
(SN ≤ 5864 without 5831 and 5832)
Solution Install software patch 2.1.1_C1 on top of software 2.1.0_C1 at the next service intervention.
If software 2.1.0_C1 is not installed yet, first perform a software upgrade to 2.1.0_C1.
Page 3 of 3