Professional Documents
Culture Documents
DC30-049 TechVision Service Manual - Rev Q
DC30-049 TechVision Service Manual - Rev Q
QG-TV-C, QG-TV-IP-C,
QG-TV-F-M, QG-TV-IP-F-M,
QG-TV-C-M, QG-TV-IP-C-M
TechVision Remote
Control Panel
Service Manual
This manual is copyrighted and all rights are reserved. No portion of this document
may be copied, photocopied, reproduced, translated, or reduced to any electronic
medium or machine readable form without prior consent in writing from Carestream
Health, Inc.
This Service Manual is incomplete without the appropriate TechVision Operator’s Man-
ual:
CHAPTER 1 - SPECIFICATIONS
INTRODUCTION . ................................................................................. 1-3
PHYSICAL SPECIFICATIONS ................................................................. 1-3
DIMENSIONS . ............................................................................... 1-3
PUSH BUTTON SWITCHES: ............................................................ 1-4
INDICATORS: . ............................................................................... 1-4
DISPLAY SCREEN AND TOUCH PANEL: ............................................ 1-4
ELECTRICAL SPECIFICATIONS .............................................................. 1-5
SYSTEM OPERATING ENVIRONMENT .................................................... 1-5
NON-OPERATING ENVIRONMENT ......................................................... 1-5
CHAPTER 6 - DIAGRAMS
TechVision Dimension Diagram ................................................................ 6-3
TechVision Wiring Diagram ...................................................................... 6-4
TechVision Main PCB Component Location Diagram ................................... 6-5
Single Board Computer Connections (AY40-092S) ........... .......................... 6-6
TechVision QS-550 Interface Board (AY40-094S) ....................................... 6-7
SPECIFICATIONS
1-1
1-2
Chapter 1 Specifications
INTRODUCTION
The TechVision Remote Control Panel is a device that not only replaces a Care-
streamTube support’s "standard equipment" hand grip unit for positioning the X-ray
Tube, but it also serves as a convenient patient area control with respect to generator
exposure settings, and image preview, when Carestream DRX-1 inte-
grated system software is implemented. A membrane, tactile-switch pad, housing
with conveniently located buttons for control of tube positioning is integrated in Tech-
Vision’s faceplate.
The touch screen, which occupies the innermost part of the TechVision faceplate, is a
16-bit, 600 x 800 pixel, color touch screen, measuring 8.25 in. (20.6 cm) diagonally.
This screen features the capability of instantly switching the view between the Gener-
ator Control Panel, and the image field of the Study Management environment. Tech-
Vision models are designated for the two basic tube transport systems Carestream
offers:
• QG-TV-C for Overhead Tube Crane, Model RS-590
• QG-TV-F for Free Standing Tube Stand, Model QS-550
• QG-TV-IP-C or f verhead
O Tube Crane, Model RS-590
• QG-TV-IP-F for Free Standing Tube Stand, Model QS-550
• Model QG-TV-C-M: Replaces the standard equipment hand grips on
Carestream Model RS-590-MAT Overhead Tube Crane System (with Auto-
tracking option)
• Model QG-TV-IP-C-M: Similar to Model QG-TV-C-M, but also has image
preview capability, and is compatible with Model RS-590-MAT Overhead
Tube Crane System (with Auto-tracking option)
• Model QG-TV-F-M: Replaces the standard equipment hand grips on
Carestream Model QS-550-MAT Free-Standing Tube Stand (with Auto-
tracking option)
• Model QG-TV-IP-F-M: Similar to Model QG-TV-F-M, but also has image
preview capability, and is compatible with Model QS-550-MAT Free-Standing
Tube Stand (with Auto-tracking option)
PHYSICAL SPECIFICATIONS
DIMENSIONS
Refer to Figure 5-1 in Chapter 5, DIAGRAMS for TechVision Remote Control Hand
Grip dimensions.
• Height: 36 cm (14.17 in.)
• Width: 55.8 cm (21.97 in.)
• Depth: 13.51 cm (5.32 in.)
• Weight: 4.9 kg (10.8 lb)
ELECTRICAL SPECIFICATIONS
NON-OPERATING ENVIRONMENT
• mbient
A
Temperature: -18°C to +70°C (-0.4 °F to 158 °F)
• Relative Humidity: 20 to 95%, non-condensing
• Atmospheric
Pressure: 500 hPa to 1060 hPa
ASSEMBLY &
INSTALLATION
2-1
2-2
Chapter 2 Assembly & Installation
OVERVIEW
This chapter describes the steps required to assemble and install the TechVision
Remote Control, including making all required electrical connections, and performing
all required alignments and calibrations.
The TechVision Remote Control is delivered fully assembled. Hence, assembly and
installation requires little more than:
• Unpacking
• Mounting the TechVision Remote Control (Required for Models QG-TV-F,
QG-TV-IP-F, QG-TV-F-M, QG-TV-IP-F-M)
Any person who connects additional equipment to the signal input part or signal out-
put part configures a medical system.
The entity accountable for the use and maintenance of the Q-Rad System (Responsi-
ble Organization is responsible for understanding that the connection of the Q-Rad
System to a Network/Data Coupling, that includes other equipment could result in
previously unidentified risks to patients, operators, or third parties. The Responsible
Organization, should identify, analyze, evaluate and control these risks.
Subsequent changes to the Network/Data Coupling, could introduce new risks and
require additional analysis. Changes to the Network/Data Coupling include:
• Changes in Network/Data Coupling configuration
• Connection of additional items to the Network/Data Coupling
• Disconnecting items from the Network/Data Coupling
• Upgrade of equipment connected to the Network/Data Coupling
REQUIRED TOOLS AND MATERIALS
The following tools and materials are required to complete the assembly and
installation procedures:
• Standard hand tools
UNPACKING
TechVision Remote
Model Tube Support Device Used With
Control Model
TUBE SUPPORT
BRACKET
TECHVISION
CONNECTOR PANEL
CONNECTOR J7 CONNECTOR J2
CONNECTOR J6
(POWER FROM RS- (ETHERNET SWITCH)
(CANBUS DATA1)
590 CABLE D306)
TUBE MOUNTING
BRACKET
ME10-207
TECHVISION
MOUNTING HOLES
(4X)
CALIBRATION &
FDA CERTIFICATION
3-1
3-2
Chapter 3 Calibration & Certification
OVERVIEW
This chapter provides instructions for configuring, calibrating and certifying the Tech-
Vision Remote X-ray Control following installation. Certification must be performed to
assure compliance with applicable provisions of FDA Performance Standards 21 CFR
Sub-Chapter J, Part 1020.
SYSTEM CONFIGURATION/CALIBRATION
The TechVision Remote Control has Service Modes that allow the X-Ray Technician
or the Field Service Technician to modify the operation of TechVision and allow it to
be calibrated. The procedures that follow describe, in order, how to:
• Access either the Field Tech Service Mode or Operator Service Mode options
• Use each of the Field Tech Service Mode functions and utilities (including
model-specific utilities depending on the platform to which the TechVision
Remote Control is attached)
• Use each of the Operator Service functions and utilities.
ACCESSING SERVICE MODE
Calibration and configuration can only be performed by accessing the TechVision
Service Mode. To gain access to the Service Mode, password entry is required.
This is to prevent unintended and/or unauthorized changes to system settings.
To access Service Mode, proceed as follows:
1. On the TechVision Touch Screen Display, tap
twice within one second on the Logo at the
upper left of the screen (see logos below).
or or
2. Once the Service Mode has been selected, the Password Screen will appear
(see Figure 3-1):
TechVision : 3.7.0.0
RS590 : 1.1.0.0
Linear IV : 1.2.0.0
Upon successful entry of the Service Password, the FieldTech Options screen is
displayed, as described below.
SERVICE OPTIONS
The TechVision calibration process is contingent upon the following configuration
characteristics:
1. Whether, or not, the X-ray System is integrated with a computer-based imag-
ing system.
2. Whether the system employs a Carestream image receptor or a Carestream
receptor.
3. Whether the Tube Stand is a Model QS-550, QS-550-MAT, RS-590, or a Model
RS-590-MAT.
NOTE: Calibration procedures for TechVision used
in Auto-track applications are described in the
Q-Rad with Auto-Track option Service Manual
(DC30-143).
The FieldTech Options screens (see Figures 3-2 and 3-2a) provide access to all of
the configurable parameters in the System. Note that Figure 3-2a reflects the
FieldTech Options screen if the "Down" button is used to scroll the screen.
The following information describes the purpose of each button on the Screen.
Note that the menu selections (buttons) are dependent on whether or not the
TechVision is standalone (not working in conjunction with a generator control
panel) or is integrated with an X-ray generator control panel (such as Q-Vision or
DRX.
• FTP: Activation of the "FTP" causes the TechVision unit to enter a state,
where field upgrades can be implemented and log files can be
transferred (see Figure 3-2b.). This feature is used when installing
updates to the TechVision. Selection of the "Close" button on the FTP
screen will initiate a restart of the application..
will have to log-on, using the keypad and log-on procedure (see Figure
3-1).
FieldTech Options
Restart Application
Up
Exit
are all expressed in 32 bits divided into four 8-bit sections, each represented
by a decimal number from 0 to 255, separated by a period.
Study Management
The Study Management Configuration function allows settings that affect
what is displayed on the Study Management screens. When the Study Man-
agement Configuration button is selected from the FieldTech Options
menu (see Figure 3-2), the following screen appears:
Room Configuration
Selecting the Room Configuration button in the Field Tech Options menu
(see Figure 3-2), displays the Room Configuration Menu (below):
the limit setting, a more abrupt "stop" will occur when the tube
support is moved to either a transverse or longitudinal detent.
• Autotracker Enable - Activates/deactivates TechVision autotracker
functionality.
• Wall Stand Enable - Activates/deactivates TechVision wall stand
autotracker functionality.
• Table Enable - Activates/deactivates TechVision table autotracker
functionality.
Locking Speed Calibration (Overhead Systems Only)
Proceed as follows:
1. When viewing the Room Configuration Menu, the current locking
speed setting appears in the Locking Speed field, as shown in the
sample screen Figure 3-9.
2. To modify the setting, select the Modify button adjacent to the
"Locking Speed" field. The Locking Speed Setting Sub-screen
displays:
Locking Speed:
Modify
125 mm/s 150 mm/s
alibrate
0/0 Cancel
Down
Back
Room Calibration
NOTE: Performance of the following cali-
brations depends on the type tube sup-
port (i.e., Model QS-550/QS-550-MAT
Free-Standing Tube Support, or RS-590/
RS-590-MAT Overhead Tube Crane) on
which TechVision is installed. Refer to the
paragraph corresponding to your tube
support type.
Selection of the Room Calibration button on the Field Tech Options Menu
(see Figure 3-2) displays the Room Calibration menu (below).
(1) (2)
(3)
Left-Facing 90° Right-Facing 90°
Horizontal Calibration Screen
Calibration Screen Calibration Screen
2. Move the tube/hand grip so that a 40" SID is achieved (pull out the
cassette tray and measure 40" SID relative to the cassette). Press
OK.
3. Exit to the Main Screen, and verify the SID indication on the
TechVision screen correctly tracks the height of the tube.
Wall Stand Alignment
Proceed as follows:
1. In the Room Calibration Menu (see Figure 3-9a), select the Calibrate
button adjacent to the "Wallstand Alignment" field. The sub-screen
shown in Figure 3-12 appears, which calls for the TechVision to be
set at 40" SID relative to the wall stand.
Language / Regional
The Language / Regional screen includes settings that affect the language
and formatting to permit localization of the user interface to regional require-
ments. When the Language / Regional button is selected from the FieldTech
Options menu (see Figure 3-2a), the following screen appears:
Language
In the Language / Regional Screen (see Figure 3-14), select the Modify
button adjacent to the "Language" field. The Language sub-screen shown
in Figure 3-15 appears:
Region
In the Language / Regional Screen (see Figure 3-14), select the Modify
button adjacent to the "Region" field. The Region sub-screen shown in Fig-
ure 3-16 appears:
System of Measurement
In the Language / Regional Screen (see Figure 3-14), select the Modify
button adjacent to the "System of Measurement" field. The System of
Measurement sub-screen shown in Figure 3-17 appears:
Date Pattern
In the Language / Regional Screen (see Figure 3-14), select the Modify
button adjacent to the "Date Pattern" field. The Date Pattern sub-screen
shown in Figure 3-18 appears:
Date Separator
In the Language / Regional Screen (see Figure 3-14), select the Modify
button adjacent to the "Date Separator" field. The Date Separator sub-
screen shown in Figure 3-19 appears:
Time Separator
In the Language / Regional Screen (see Figure 3-14), select the Modify
button adjacent to the "Time Separator" field. The Time Separator sub-
screen shown in Figure 3-20 appears:
Decimal Separator
In the Language / Regional Screen (see Figure 3-14), select the Modify
button adjacent to the "#Decimal Separator" field. The Decimal Separa-
tor sub-screen shown in Figure 3-21 appears:
Group Separator
In the Language / Regional Screen (see Figure 3-14), select the Modify
button adjacent to the "#Group Separator" field. The Decimal Separator
sub-screen shown in Figure 3-22 appears:
Change Date/Time
When the Change Date Time key is selected from the FieldTech Options
Menu (see Figure 3-2a), the following screen appears:
MAXIMUM MAXIMUM
MODEL mA MODEL mA
QG-32-SE 400 QG-40-2 500
QG-40-SE 450 QG-40-3 500
QG-50-SE 500 QG-50 650
QG-32 500 QG-50-2 650
QG-32-2 500 QG-65 800
QG-32-3 500 QG-80 800
QG-40 500
4. After the tenth exposure, select (coefficient of variation) COEFF VAR on Dyn-
alyzer; reading must be less than or equal to 0.05. If a Dynalyzer is not used,
calculate the C.O.V. (coefficient of variation) according to the following:
1/2
s 1
n
(xi - x)2
C= =
X X i=1
n-1
Where
C= Coefficient of Variation
s = standard deviation
x = Mean value of mR readings
xi = ith observation sampled
n = number of observations sampled
Page 1 of 5
Page 2 of 5
SAMPLE 1 mA SAMPLE 2 mA
(From Table 3-2) = _______ (From Table 3-2) = _______
EXPOSURE mR mAs EXPOSURE mR mAs
1 1
2 2
3 3
4 4
5 5
6 6
7 7
8 8
9 9
10 10
SUM SUM
MEAN VALUES MEAN VALUES
= SUM ÷ 10 = SUM ÷ 10
X1 = _______ X2 = _______
= _______ X1 + X2 = _______
X1 - X2
X1 – X2
LINEARITY = ---------------------
- = ________________ must be < 0.1
X1 + X
2
Page 3 of 5
SAMPLE 1 mA SAMPLE 2 mA
(From Table 3-2) = _______ (From Table 3-2) = _______
EXPOSURE mR mAs EXPOSURE mR mAs
1 1
2 2
3 3
4 4
5 5
6 6
7 7
8 8
9 9
10 10
SUM SUM
MEAN VALUES MEAN VALUES
= SUM ÷ 10 = SUM ÷ 10
X1 = _______ X2 = _______
= _______ X1 + X2 = _______
X1 - X2
X1 – X2
LINEARITY = ---------------------
- = ________________ must be < 0.1
X1 + X
2
Page 4 of 5
Table 3-7. Maximum Deviation Data Sheet
SETTING 45 200L 50 mS 10
1
MEASURED
MEASURED
* Only required when generator is equipped with 150 KV option
Page 5 of 5
BACK UP Error
Indicated mAs Measured mAs
(Y/N)
THEORY
OF OPERATION
4-1
4-2
Chapter 4 Theory of Operation
OVERVIEW
This section describes various characteristics of the TechVision’s design, functionality,
and compatibility with available mechanical support systems. Figure 4-1 is a high-
level block diagram, which illustrates the four interconnected internal components
built into the TechVision unit, and its connectivity to external devices.
PLATFORM SYSTEM
TechVision is an option available on Carestream's X-ray platforms:
• Carestream Model QS-550 Free-Standing Tube Stand
• Carestream Model RS-590 OverHead Tube Crane
It replaces the standard-equipped hand-grip on the above units and provides the fol-
lowing value-added features:
• 10.4", 1024 x 768 dpi, feature rich XGA interactive color touch-screen
• Digital Tube Angle measurement and SID Display
• 10/100Mbps TCP/IP interface
• CAN-BUS interface for control of advanced platforms, Collimators, and DAP
subsystems
• Remote selection and display of generator exposure settings at TechVision
Hand grips via Ethernet communications with the X-ray generator.
• Digital X-Ray image preview/recall
The following paragraphs provide detailed theory of operation of the various
TechVision components. See Chapter 6, DIAGRAMS for system schematics.
THEORY OF OPERATION
OVERALL HARDWARE ARCHITECTURE
The TechVision platform electronics consists of 4 major interconnected
components:
1. TechVision Main Board (AY40-093S)
2. Single Board Computer (AY40-092S)
3. LCD Touchscreen Display (EL31-032)
4. TechVision QS-550 Interface PCB (AY40-094S), used on systems with QS-550
Tubestand
5. Membrane Keypad and Switches
ter (U1). This interface also provides galvanic isolation between the CANBUS
data I/O lines and the TechVision electronics. CAN-BUS interface is provided
on the TechVision QS-550 Interface PCB (connectors J5 and J6).
4. Isolated SID Detent Input Circuit: The Main Microcontroller (U2) reads all
detent switches (optically isolated by U7 and U8), and will relay these switch
closures to the TechVision Main PCB (AY40-093S), via high speed CAN-BUS.
5. Isolated Output Circuit: The QS-550 Interface PCB (AY40-093S), provides
relay-isolated, dry contact closures, to control up to four electromagnetic
brakes in the QS-550 Tubestand. Relay K1 will unlock the Vertical Brake, K2
unlocks the Tube Angle Brake, K3 unlocks transverse movement, and K4
unlocks Longitudinal movement. These relays are all designed to maintain all
brakes in a locked state (i.e., fail-safe), when they are not energized. The
relays are controlled by Main Microcontroller U2, via commands received over
the CAN-BUS interface from TechVision Main PCB (AY40-093S).
CAN
External INTERFACE
24-28V
Power
Input CAN Bus
TECHVISION
TECHVISION PCB
MEMBRANE KEYPAD &
AY40-093S HANDLE SWITCHES
SERVICE
INSTRUCTIONS
5-1
5-2
Chapter 5 Service Instructions
OVERVIEW
This chapter is designed to assist the service technician in maintaining the smooth
operation of TechVision Control Panel. The TechVision Control Panel will function reli-
ably when maintained according to the instructions provided in this chapter. Safe
equipment performance requires the attention of service personnel who are specifi-
cally trained and experienced with medical x-ray apparatus. Applicable preventive
maintenance or any repair service should be performed by these skilled individuals.
Failure to follow manufacturer’s or service personnel’s recommendations may result in
serious injury.
SERVICE MAINTENANCE
Cleaning
Painted surfaces should be cleaned using a clean cloth slightly moistened
in warm soapy water (use mild soap). Wipe the TechVision front display
screen surface using an absorbent cotton cloth or other soft material.
Wipe the motion control push button overlay with a clean wet cloth, then
dry. Never use strong cleaners, solvents, or abrasive polish on this equip-
ment.
Preventive Maintenance
The following inspection and adjustment procedures are recommended to
maintain the system in its original operating condition.
Perform Every Six Months:
• Visually inspect the entire TechVision unit for loose hardware,
loose cable connections
• Conduct a general inspection for worn or damaged parts
• Verify that all ground conductors are properly and securely
installed and free of corrosion or damage
• Check all electrical cabling and wiring for wear and fraying
• Verify normal operation of the TechVision front panel
(membrane) and hand grip "pinky" switches
TROUBLESHOOTING PROCEDURE
The TechVision troubleshooting issues covered in this manual are described in flow
diagram format on the following pages. Each diagram is designed to provide
instructions aimed at discovering the cause of the fault condition described in that
diagram’s figure caption, however the instructions in certain diagrams direct the
troubleshooting process to particular topics found in other diagrams. For the pur-
pose of assisting the Troubleshooter in quickly locating the proper path, each dia-
gram is assigned a number (Troubleshooting Condition #), and this number is
referenced in the instruction block, where the troubleshooting path transitions to
another diagram. The table below, correlates the troubleshooting conditions con-
tained in this manual with their respective Troubleshooting Condition Number.
Condition Condition
Troubleshooting Issue Troubleshooting Issue
No. No.
TROUBLESHOOTING CONDITION # 1
No
Power Possible Cause
No 28V input
Proceed to
troubleshooting step Yes Are all 3
揘o Display? illuminated?
(see Condition #2)
No
Which
Tubestand
Recheck/Replace
RJ45 Jumper
cable from Yes
AY40-093S to Replace
AY40-094S CAN AY40-094S
DATA1 on back
of TechVision
TROUBLESHOOTING CONDITION # 2
No
Display
Possible Cause:
No 28V input
No
Possible Cause:
Corrupted operating system.
Check backlight cable
Replace the SD Card (EL85-199S) (AY51-333), and replace
if damaged.
TROUBLESHOOTING CONDITION # 3
Not responsive
to touch input
Possible cause:
Damage to touchscreen digitizer
Yes Physical No
Damage?
Possible cause:
Replace (as necessary): Damage to touchscreen cable.
Possible cause:
Replace the cable (AY51-334) Fault on the Single Board Computer
TROUBLESHOOTING CONDITION # 4
Not
responsive to
pushbutton
inputs
Possible Cause:
Tubestand disconnected
Yes No
Disconnected?
Proceed to trouble-
shooting step
揟ubestand
Disconnected? Possible cause:
(see Condition #8) Button lock enabled
Unlock the
Locked or Do handle Replace
Tubestand by
Yes 揅 lick to Unlock No switches work? No
clicking the locked AY40-093S
icon
Yes
NOTE: Part Numbers for Membrane Keypad vary with TechVision application.
Contact Carestream Service for replacement part number.
No
Possible cause: Possible cause:
Alignment of indent 40” 72” Detent Switches
Yes
Update D303 Firmware
or Replace D303 Board
in RS-590
J7-5 to J7-2 > No
16 V dc only when Do NOT replace
at 72”? any boards
TROUBLESHOOTING CONDITION # 5
Yes
Check wiring from
detent switches to
Replace QS-550 DB15 connector J7
Interface PCB on AY40-094S PCB.
Revision Q
5-9
Chapter 5 Service Instructions
TROUBLESHOOTING CONDITION # 6
Inaccurate
Date Time
Possible Cause
Date Time Not Set Possible Cause
Battery Failed
Yes Keeping No
Accurate Time?
Yes
Resolved
This means No
everything is in
working order. If still unresoolved,
Replace the
SD Card replace the Single
(EL85-199S) Board Computer
(AY40-092S)
TROUBLESHOOTING CONDITION # 7
Internal
Comm
Fault
Possible Cause:
Communication error between TechVision
Main Board AY40-093S and Single Board
Computer AY40-092S.
Resolved?
If problem persists,
replace the Single Board
Computer (AY40-092S).
TROUBLESHOOTING CONDITION # 8
Tubestand
Disconnected
Possible Cause:
CAN Bus could not establish communications
with the tubestand
Is the
Replace: Yes No
blue LED
1. SD Card EL85-199S
illuminated
2. Main Board AY40-093S
Check the CanBus wiring and make sure the communications cable
from the RS-590 Control Board is connected to the jack marked
揅AN DATA 1?and on the back of the TechVision is NOT in the jack
marked 揘 ETWORK EHTERNET PORT? Note: They are both
RJ45 connectors but are not interchangeable.
Which
QS-550 Tubestand? RS-590
Yes Issue No
Replace the cable Resolved?
Figure 5-9.
Use pin on a connected PC to
check for presence of the
TechVision. Did ping succed?
Green LED
Yes
Revision Q
5-13
Chapter 5 Service Instructions
TROUBLESHOOTING CONDITION # 10
Screen rotated
incorrectly
Possible cause:
Tube Angle not
calibrated
Replace the
Main Board
(AY40-093S)
TROUBLESHOOTING CONDITION # 11
SID not
being
displayed
properly
Yes No
Tube Angle > ?
SID is *NOT* Degrees
displayed if
tube angle is
>5 degrees.
Yes No
Linear IV
system
Possible fault
Does Does
Yes No Yes with vertical SID
raw angle RS-590 magnetic potentiometer or
value change brake release D303 Board in
work RS-590
Proceed to troubleshooting
flowchart 揇isplay OK, no
response to buttons?
(see Condition #4)
Contact Carestream
Chapter 5 Service Instructions
Replace LCD
Module
(EL31-032)
ORDERING INFORMATION
To order replacement parts for the tube stand, contact the Service Department at;
Carestream Health,Inc
150 Verona Street, Rochester, NY14608, USA
www.carestream.com
4
3
SD Card
FRONT VIEW
11
10
DIAGRAMS
6-1
6-2
Chapter 6 Diagrams
13.51 cm
5.32 in.
55.80 cm
21.97 in.
35.99 cm
14.17 in.
55.80 cm
21.97 in.
T:\Storage\Manuals\Techvision\Operator抯Manual\TECHVISION DIMENSIONS.cdr
Figure 6-1. TechVision Dimension Diagram
5VOK LED
3V3 OK LED
Q1
C18
10/100 Ethernet
C32
C165
C31 C57
C33 R81
R48 R80
C27 C93 R47 R45 R84
R82
P8 SINGLE BOARD COMPUTER
C166 R46
R83
C157 C168
R85
C160 C167
R86
R87
R88
R89
R90
R91
TO J14 OF MAIN PCB ASSY. (AY40-093)
C169
C156
C163
R92
R93
R94
R75
C173
C153
R154 R150
R153 R155
R152 C154 R151
microSD Card
R52
C35
SD CARD SLOT
C171
R53
C76
C149
R148 R149
C150
T:\Storage\ManualsTechvision\Service Manual\DIAGRAMS\AY40-092S
C147 C148
C146 R70
R69
R68
C145 R67
C144
R66
R65
R64
C155 C172 R63
FB5 R62 C75
C95
C103 R61 R101
C170 R60 R105
C102
C162 R58 R102
C22 C101 C161 R57 R110
R54 R59
C21 C74 R106
R26 R55 R109
C68
R21 R164 C104 C164 R56
R163 C92
C105 C50 C43
C89 C47 C42 C56 C81
C91 R17
C88 C72 C49 C106
C90
C63 C94
C59
R20
R19
R18 R51 R167 C158
P2 C80
C159
20
R160
C130 C125
C129 R162
19
R111
C128 R104
C152
C127
R156
R107
C15 R1 C143
R142
R141
R143
R6 C10 C4 C142
R77
C36 R76
AY40-092S Rev. A
R79
R78
C20
Figure 6-4. Single Board Computer
R136
Connections (AY40-092S)
Chapter 6 Diagrams
Revision Q
6-6
Chapter 6 Diagrams
U2 MICRO-CONTROLLER
?LOGIC
?PERIPHERAL COMMUNICATIONS
T:\Storage\Manuals\Techvision\Service Manual\DIAGRAMS\AY40-094S.cdr