DC30-128 RS-590 Service Manual - Rev F
DC30-128 RS-590 Service Manual - Rev F
RS-590-MAT
Tube Stand
Service Manual
This manual is copyrighted and all rights are reserved. No portion of this document
may be copied, photocopied, reproduced, translated, or reduced to any electronic
medium or machine readable form without prior consent in writing from Carestream
Health, Inc.
Page
Part Number Rev
Revision History 3 to 4 F
Table of Contents 5 to 6 F
Introduction / Overview 7 to 16 F
Getting Started 17 to 36 F
Lifecycle 37 to 42 F
Appendix 43 to 50 F
PART: INTRODUCTION
DOCUMENT INFORMATION ..................................................................... 9
MODELS AND CONFIGURATIONS ............................................................. 9
ACRONYMS AND ABBREVIATIONS ........................................................... 9
SAFETY .................................................................................................. 9
TECHNICAL SPECIFICATIONS .................................................................. 9
COMPATIBLE X-RAY TUBES AND COLLIMATORS ................................. 9
COLLIMATOR COMPATIBILITY* ....................................................... 10
POWER SUPPLY .................................................................................... 12
POSITIONING THE TUBE STAND ABOVE THE WALL STAND .................... 12
INSTALLATION PREPARATIONS ............................................................. 12
PART: LIFECYCLE
MAINTENANCE ..................................................................................... 39
PERIODIC INSPECTION ........................................................................ 39
ADJUSTMENT ....................................................................................... 39
REPLACEMENT of PARTS ....................................................................... 39
REPLACEMENT PARTS AND ORDERING INFORMATION ..................... 40
ORDERING INFORMATION .............................................................. 40
PART: APPENDIX
LOCATION OF LABELS ...........................................................................45
TABLE OF LABELS ...........................................................................46
COLLIMATOR INTERFACE DIAGRAMS .....................................................46
7
8
Part: Introduction/Overview
DOCUMENT INFORMATION
The RS-590 and RS-590-MAT Tube Stand Service Manual (this manual) is designed by
Carestream, to serve as a supplement to its companion document, the Siemens Tube
Stand Technical Manual (order No. 7042000 VFT 01S 00).
In general, this manual (hereinafter "Supplement,"), is written to closely correspond
with the Siemens Tube Stand Technical Manual (hereinafter "Technical Manual"), with
respect to topics and its organization. The Supplement defers (by reference) to the
descriptions found in the Technical Manual, with regard to the assembly of the struc-
tural system, and mechanical aspects of the positioning axes of the RS-590/RS-590-
MAT Tube Stand (hereinafter referred to as the Tube Stand). Service instructions
related to controls (depending on handgrip option), and Power Supply installation are
found in the Supplement, or when appropriate, in the ancillary documents, such as
TechVision Service Manual (DC30-049), or the Power Supply Installation Instructions
(DC30-014).
Assembly, installation and calibration issues related to the Tube Head, and associated
equipment integrated by Carestream are addressed in the Supplement. Topics, in the
Supplement, which are further discussed in the Technical Manual, are limited to situa-
tions where the equipment is integrated. The reader should understand that the Sup-
plement only provides service and assembly information for the portions of the Tube
Stand, where Carestream equipment or technology has replaced, altered or enhanced
that which is present in the Siemens system.
MODELS AND CONFIGURATIONS
Please refer to the Technical Manual.
ACRONYMS AND ABBREVIATIONS
Please refer to the Technical Manual.
SAFETY
Please refer in the Technical Manual for Safety considerations:
• General Information
• Proper Product Usage
• General Hazards
TECHNICAL SPECIFICATIONS
Except for the following diagrams, tables and descriptive information, this supple-
ment, defers to the Technical Specifications section found in the Siemens Technical
Manual.
COMPATIBLE X-RAY TUBES AND COLLIMATORS
The following table lists the part number and description for X-ray Tubes currently
Manufacturer Type/Model
Toshiba E7239
Toshiba E7242
Toshiba E7252
Toshiba E7254
Toshiba E7869
Varian RAD-14
Varian RAD-60
Varian RAD-92
Dunlee DA-1092
Dunlee DA-1036
COLLIMATOR COMPATIBILITY*
The following collimators are compatible for use with the Tube Stand:
The Dimensional Drawing below depicts the dimensional envelope of the Tube Stand.
118.90 [3020]
32.67 [830] MIN 87.76 [2229]
103.54 [2630] MAX RANGE OF TRANSVERSE
? COLUMN MOVEMENT
? COLUMN
POWER SUPPLY
The following power supply options are available for the Tube Stand, and electrical
input characteristics are specified in the documents referenced for each.
• QG-28VPS: Mounted internally to the Generator cabinet (see Generator
Service Manual DC30-011)
• PS-6 or PS-7: Enclosed in a separate wall-mounted cabinet
(see Power Supply Installation Instructions DC30-014)
POSITIONING THE TUBE STAND ABOVE THE WALL STAND
The mechanical system for the Tube Stand is designed to provide sufficient longitudi-
nal travel to allow positioning of the X-ray Tube directly above the focal spot of a Wall
Stand fitted with a tilting receptor cabinet as shown below.
16.39 12.17
2.55 2.55
9.56 9.56
108.00
12.59
12.59
INSTALLATION PREPARATIONS
Refer to the Technical Manual for technical specifications related to Installation Prepa-
rations and Usable Focus Range, and a 3D view of the assembled mechanical system.
PART: GETTING
STARTED
17
18
Part: Getting Started
INSTALLATION
This supplement defers to the Technical Manual for all installation instructions, except
for those pertaining to wiring and connectivity of the power supply, and integration of
the Carestream tube head assembly, including handgrip devices.
STARTUP
Procedures pertaining to checking and aligning the mechanical system, including
bearings, end stops, collimator adjustments, alignment of the table and wall stand,
are provided in the Technical Manual, or the collimator documentation.
TUBE HEAD ASSEMBLY
The RS-590 Tube Stand includes a Tube Head Assembly, designed by Carestream.
This consists of an X-ray tube, a collimator, and handgrip unit, which ultimately is
interfaced to the Siemens Tube Support Arm.
The X-ray tube, the collimator, and the handgrip unit are packaged separately, to
ensure that they are not damaged in transit. It is necessary for the installer to
physically mount these components using the interface bracket (with trunnion
rings), affixed at the factory, to the mounting plate of the Siemens Tube Support
Arm Assembly. The Tube Head assembly is built-up from the trunnion bracket,
once it is properly interfaced to the rotating element of the Siemens Tube Support
Arm.
When properly mounted, the tube can be oriented downward, or to the side, since
the rotating element serves as the mounting surface for the trunnion bracket
assembly.
Tube Head Assembly
The Tube Stand is integrated with Carestream ’s Tube Head Assembly,
and the following descriptions and figures show how the X-ray Tube is
attached to the Tube Support Arm of the Siemens Tube Stand (vertical
axis). Refer to the following figures, and proceed with assembly of the
tube:
Siemens Tube
Support Arm
Assembly
3. Place X-ray tube onto lower half of Trunnion Rings and re-mount
upper half.
TUBE SUPPORT
BRACKET
TRUNNION
ASSEMBLY
TRUNNION
HANDLE 1/4-20 X 綌
BUTTON HEAD
SOCKET SCREW
TRUNNION (2X)
CAM
Refer to the TechVision Service Manual (DC30-126) for setup of the optional Tech-
Vision handgrip unit (QG-TV-C, QG-TV-CM, QG-TV-IP-C or QG-TV-IP-C-M).
1 2
4
3
1) SID display
2) Longitudinal detent active
3) Transverse detent active
4) Active motion axis icon
5) Status indications (see a through d)
a. Indicates a system fault due to an attempt to move the axes while locked.
b. Indicates that internal communications has failed within TechVision.
c. Indicates that the CAN bus cable connection is faulty (automatic collimators only).
d. Indicates that the CAN bus cable connection is faulty between TechVision and RS-590.
T:\Storage\Manuals\RS-590\GRAPHICS\RS-590-LD Touch Scr Features.cdr
INSTALLATION INSTRUCTIONS
The Digital Touch Screen Handgrip is shipped fully assembled. It is designed as an
enhanced capability operator control option, specifically for the RS-590 Tube
Stand, and is easily installed.
The following paragraph contains installation procedures for Digital Touch Screen
Handgrip.
Digital Touch Screen Handgrip Installation Procedure
When supplied with the Tube Stand, the Digital Touch Screen Handgrip is
shipped separately, and must be mounted to the tube head assembly, fol-
lowing the instructions provided in "Tube Head Assembly".
Refer to the figure below, and the Interconnection Diagram for electrical
connections required.
NOT USED
CONNECTOR J7
(POWER FROM RS-590 CONNECTOR J6
CABLE D306-X1) (CANBUS DATA1)
TO
SU TU FEMALE PLUG
PP BE
OR
T
T:\Storage\Manuals\RS-590 Service Manual\ELEMENTS\Power Supply Interface Cables and Plug\Power Supply Interface Cable and Plug.cdr
SYSTEM CONFIGURATION/CALIBRATION
The Digital Touch Screen Handgrip has Service Modes that allow the X-ray Techni-
cian or the Field Service Technician to modify the operation of Tube Stand, and allow
it to be calibrated. The procedures that follow describe, in order, how to:
• Access either the Field Tech Service Mode or Operator Service Mode options
• Use each of the Field Tech Service Mode functions and utilities (including
model-specific utilities for the Tube Stand).
• Use each of the Operator Service functions and utilities.
ACCESSING SERVICE MODE
Calibration and configuration can only be performed by accessing the Digital
Touch Screen Handgrip Service Mode. To gain access to the Service Mode, pass-
word entry is required. This is to prevent unintended and/or unauthorized
changes to system settings. To access Service Mode, proceed as follows:
1. On the Digital Touch Screen Hand-
grip Display, tap twice within one
second on the Quantum Logo.
2. Once the Service Mode has been selected, the Password Screen will appear
(see figure below):
CURRENT
SOFTWARE
REVISION IS
DISPLAYED HERE
Password Screen
3. Use the numeric "touch" keypad to enter the Service Password (58623), then
press ENTER. To start again, push Clear.
NOTE: Digital Touch Screen Handgrip returns to
the previous screen when an incorrect password is
entered, or after 15 seconds with no keys pressed.
Upon successful entry of the Service Password, the Service Options screen is dis-
played, as described below.
SERVICE OPTIONS
The Digital Touch Screen Handgrip calibration process is contingent upon the fol-
lowing configuration characteristics:
1. Whether, or not, the X-ray System is integrated with a computer-based imag-
ing system.
2. Whether the system employs a Carestream image receptor or a Carestream
receptor.
The FieldTech Options screens provide access to all of the configurable parame-
ters in the System. Note that the FieldTech Options screen (below), shows options
available if the "Down" button is used to scroll the screen.
The following information describes the purpose of each button on the Screen.
• Device Configuration: Select this button to verify or modify the
Digital Touch Screen Handgrip’s IP Address, Subnet Mask, or Default
Gateway value.
• RS590 Calibration: Allows calibration of physical orientation
parameters for the handgrip (RS-590-LD/RS-590-LDM).
• Language Configuration: Touch this button to access a menu of
languages which can be enabled for all displays and nomenclature.
• Regional Configuration: Touch this button to set displays of
measurements to metric or standard values.
FieldTech Options
Down
Exit
FieldTech Options
Restart Application
Up Down
Exit
FTP Screen
• Restart Application: This feature allows for initialization without
effecting any configuration changes. The RS-590-LD display will cycle
through it normal boot-up sequence, ending with Main Display Screen.
In order to again access the FieldTech Options screen, the Technician will
have to log-on, using the keypad and log-on procedure described
Accessing Service Mode at the beginning of this section.
• Up: Touch this button to scroll the screen listings up to reveal additional
options.
• Exit: Touch this button to leave the Service Mode and return to
Operator Mode.
NOTE: Pressing the Back button in any of
the following screens returns the system
to the Service Options Menu.
Locking Speed:
Cancel
(1) (2)
(3)
Left-Facing 90° Right-Facing 90°
Horizontal Calibration Screen
Calibration Screen Calibration Screen
SELECT “CALIBRATE”
TO BEGIN SID HEIGHT
CALIBRATION PROCESS
OK
OK
4. Exit to the Main Screen, and verify the SID indication on the Digital
Touch Screen Handgrip screen correctly tracks the height of the tube.
This completes the tube stand specific calibrations, proceed to "Calibrate
Touch Screen" to continue.
37
38
Part: Lifecycle
MAINTENANCE
This supplement defers to the Technical Manual for instructions as to maintenance
procedures for:
• Longitudinal Rails and Transverse Track
• Telescoping Carriage
• Support Arm
PERIODIC INSPECTION
Ensure that collimator and hand grip mounting hardware is secure.
ADJUSTMENT
This supplement defers to the Technical Manual for instructions as to adjustment pro-
cedures for:
• Ball Bearings
• Central Beam Drift
• Parallelism of the X-ray Tube to the Table
• Light Field / Radiation Field of Collimator
• Wobble of the Support Arm
• Tube Unit Drift
• Aligning the Collimator
• Adjusting the Weight Compensation of the Spring-loaded Mechanism
• Vertical Axis Brake Gap (RVA)
• Horizontal Axis Brake Gap (RHA)
• Transverse and Longitudinal Carriage Brake Gap
• Detent Forces
• Stop Assembly
REPLACEMENT OF PARTS
Except for the following descriptions as to replacement of parts, this supplement
defers to the Technical Manual:
• Handles and Buttons: Refer to the information provided in the following
section.
• Tube Assembly: Refer to X-ray Tube manufacturer documentation.
• Collimator: Refer to Collimator manufacturer documentation.
Part Used
Item Description Number On
1 (1 ea.)Main PC Board AY40-093S RS-590-LD,
RS-590-LDM
ORDERING INFORMATION
To order replacement parts for the tube stand, contact the Service Department
at;
Carestream Health, Inc
150 Verona Street, Rochester,NY14608, USA
Phone: 800-328-2910
www.carestream.com
43
44
Part: Appendix
LOCATION OF LABELS
The following illustration shows the location of the identification label, and the serial
number label for the power supply options.
This supplement defers to the Technical Manual for warning label locations related to
the Siemens Tube Stand (see the Table of Labels in the Appendix of the Technical
Manual).
A
1
A
4
3
Rear Panel
B
2
POWER SUPPLY MODULE
T:\Storage\Manuals\RS-590\GRAPHICS\Label Locations.cdr
TABLE OF LABELS
3 Collimator (typical)