Models RS-590,

RS-590-MAT
Tube Stand

Service Manual

Manual Part No. DC30-128 Revision F


The original version of this manual dated 2012-09-26 (September 26, 2012) was
drafted in the English Language by Carestream Health, Inc.

This manual is copyrighted and all rights are reserved. No portion of this document
may be copied, photocopied, reproduced, translated, or reduced to any electronic
medium or machine readable form without prior consent in writing from Carestream
Health, Inc.

© Carestream Health, Inc.,2016

Carestream Health, Inc.


150 Verona Street, Rochester,NY14608, USA
www.carestream.com
Revision History

REVISION DATE TYPE OF MODIFICATION

A 2012-10-25 Initial Release.

B 2013-03-22 Incorporated ECO 2736, 2780

C 2014-09-15 Revised Collimator Compatibility Table, added collima-


tor interface diagrams, deferred to collimator manufac-
turer’s installation instructions, and incorporated
warning as to security of collimator mounting hard-
ware.Updated service options and system calibration

D 2015-06-10 Incorporated ECO 3501, 3522 (motorized model RS-


590-MAT)

E 2016-04-04 Incorporated ECO 3612, 3646

F 2016-06-30 Updated manufacturer’s information

Page
Part Number Rev

Revision History 3 to 4 F

Table of Contents 5 to 6 F

Introduction / Overview 7 to 16 F

Getting Started 17 to 36 F

Lifecycle 37 to 42 F

Appendix 43 to 50 F

Model RS-590/RS-590-MAT Tube Stand - Service Manual Revision F


Carestream Health, Inc 3
Revision History

THIS PAGE INTENTIONALLY LEFT BLANK

Revision F Model RS-590/RS-590-MAT Tube Stand - Service Manual


4 Carestream Health, Inc
Table of Contents

PART: INTRODUCTION
DOCUMENT INFORMATION ..................................................................... 9
MODELS AND CONFIGURATIONS ............................................................. 9
ACRONYMS AND ABBREVIATIONS ........................................................... 9
SAFETY .................................................................................................. 9
TECHNICAL SPECIFICATIONS .................................................................. 9
COMPATIBLE X-RAY TUBES AND COLLIMATORS ................................. 9
COLLIMATOR COMPATIBILITY* ....................................................... 10
POWER SUPPLY .................................................................................... 12
POSITIONING THE TUBE STAND ABOVE THE WALL STAND .................... 12
INSTALLATION PREPARATIONS ............................................................. 12

PART: GETTING STARTED


INSTALLATION ..................................................................................... 19
STARTUP ............................................................................................. 19
TUBE HEAD ASSEMBLY ................................................................... 19
Tube Head Assembly ................................................................ 19
OPTIONAL DIGITAL TOUCH SCREEN HANDGRIP
(RS-590-LD/RS-590-LDM) .............................................................. 22
INSTALLATION INSTRUCTIONS ....................................................... 24
Digital Touch Screen Handgrip Installation Procedure .................. 24
Power Supply Installation .......................................................... 25
SYSTEM CONFIGURATION/CALIBRATION ............................................... 26
ACCESSING SERVICE MODE ............................................................ 26
SERVICE OPTIONS ......................................................................... 27
RS-590 System Calibration ........................................................ 31
Calibrate Touch Screen ............................................................. 36

PART: LIFECYCLE
MAINTENANCE ..................................................................................... 39
PERIODIC INSPECTION ........................................................................ 39
ADJUSTMENT ....................................................................................... 39
REPLACEMENT of PARTS ....................................................................... 39
REPLACEMENT PARTS AND ORDERING INFORMATION ..................... 40
ORDERING INFORMATION .............................................................. 40

Model RS-590/RS-590-MAT Tube Stand - Service Manual Revision F


Carestream Health, Inc 5
Table of Contents

PART: APPENDIX
LOCATION OF LABELS ...........................................................................45
TABLE OF LABELS ...........................................................................46
COLLIMATOR INTERFACE DIAGRAMS .....................................................46

Revision F Model RS-590/RS-590-MAT Tube Stand - Service Manual


6 Carestream Health, Inc
5
PART:
INTRODUCTION/
OVERVIEW

7
8
Part: Introduction/Overview

DOCUMENT INFORMATION
The RS-590 and RS-590-MAT Tube Stand Service Manual (this manual) is designed by
Carestream, to serve as a supplement to its companion document, the Siemens Tube
Stand Technical Manual (order No. 7042000 VFT 01S 00).
In general, this manual (hereinafter "Supplement,"), is written to closely correspond
with the Siemens Tube Stand Technical Manual (hereinafter "Technical Manual"), with
respect to topics and its organization. The Supplement defers (by reference) to the
descriptions found in the Technical Manual, with regard to the assembly of the struc-
tural system, and mechanical aspects of the positioning axes of the RS-590/RS-590-
MAT Tube Stand (hereinafter referred to as the Tube Stand). Service instructions
related to controls (depending on handgrip option), and Power Supply installation are
found in the Supplement, or when appropriate, in the ancillary documents, such as
TechVision Service Manual (DC30-049), or the Power Supply Installation Instructions
(DC30-014).
Assembly, installation and calibration issues related to the Tube Head, and associated
equipment integrated by Carestream are addressed in the Supplement. Topics, in the
Supplement, which are further discussed in the Technical Manual, are limited to situa-
tions where the equipment is integrated. The reader should understand that the Sup-
plement only provides service and assembly information for the portions of the Tube
Stand, where Carestream equipment or technology has replaced, altered or enhanced
that which is present in the Siemens system.
MODELS AND CONFIGURATIONS
Please refer to the Technical Manual.
ACRONYMS AND ABBREVIATIONS
Please refer to the Technical Manual.
SAFETY
Please refer in the Technical Manual for Safety considerations:
• General Information
• Proper Product Usage
• General Hazards
TECHNICAL SPECIFICATIONS
Except for the following diagrams, tables and descriptive information, this supple-
ment, defers to the Technical Specifications section found in the Siemens Technical
Manual.
COMPATIBLE X-RAY TUBES AND COLLIMATORS
The following table lists the part number and description for X-ray Tubes currently

Model RS-590/RS-590-MAT Tube Stand Service Manual Revision F


Carestream Health, Inc 9
Part: Introduction/Overview

available for integration on the Tube Stand.


The following X-ray tubes are compatible for use with the RS-590 Tube Stand.

Model RS-590/RS-590-MAT Compatible X-Ray Tube List

Manufacturer Type/Model
Toshiba E7239

Toshiba E7242

Toshiba E7252

Toshiba E7254

Toshiba E7869

Varian RAD-14

Varian RAD-60

Varian RAD-92

Dunlee DA-1092

Dunlee DA-1036

COLLIMATOR COMPATIBILITY*
The following collimators are compatible for use with the Tube Stand:

Model RS-590/RS-590-MAT Compatible Collimator List

Manufacturer Model/Part Number


Carestream/Progeny MC150 (P/N 70-63011-6)

Carestream/Progeny Linear II (P/N 70-43008-6)*

Carestream/Progeny MC150 Pinnacle (P/N 70-63009-6)

Ralco R 221/A DHHS

* Additional information as to implementation of the Progeny Linear II and


Linear IV Automatic Collimators is provided in this manual at Part Appendix.

Revision F Model RS-590/RS-590-MAT Tube Stand - Service Manual


10 Carestream Health, Inc
Part: Introduction/Overview

The Dimensional Drawing below depicts the dimensional envelope of the Tube Stand.

27.17 45.63 [1159]


[690]
7.17
[182] 20.55
[522]

118.90 [3020]
32.67 [830] MIN 87.76 [2229]
103.54 [2630] MAX RANGE OF TRANSVERSE
? COLUMN MOVEMENT

? COLUMN

COLLIMATOR 19.251? 20 4.74


[489? ]
[121]

RS-590/RS-590 Tube Stand Operating Envelope

Model RS-590/RS-590-MAT Tube Stand Service Manual Revision F


Carestream Health, Inc 11
Part: Introduction/Overview

POWER SUPPLY
The following power supply options are available for the Tube Stand, and electrical
input characteristics are specified in the documents referenced for each.
• QG-28VPS: Mounted internally to the Generator cabinet (see Generator
Service Manual DC30-011)
• PS-6 or PS-7: Enclosed in a separate wall-mounted cabinet
(see Power Supply Installation Instructions DC30-014)
POSITIONING THE TUBE STAND ABOVE THE WALL STAND
The mechanical system for the Tube Stand is designed to provide sufficient longitudi-
nal travel to allow positioning of the X-ray Tube directly above the focal spot of a Wall
Stand fitted with a tilting receptor cabinet as shown below.

16.39 12.17

2.55 2.55
9.56 9.56

108.00

12.59
12.59

Tube Stand Positioned Over Wall Stand

INSTALLATION PREPARATIONS
Refer to the Technical Manual for technical specifications related to Installation Prepa-
rations and Usable Focus Range, and a 3D view of the assembled mechanical system.

Revision F Model RS-590/RS-590-MAT Tube Stand - Service Manual


12 Carestream Health, Inc
Part: Introduction/Overview

Model RS-590/RS-590-MAT System Operating Envelope - QT-750 Table and


QW-420 Wall Stand with Standard Image Cabinet

Model RS-590/RS-590-MAT Tube Stand Service Manual Revision F


Carestream Health, Inc 13
Part: Introduction/Overview

Model RS-590/RS-590-MAT System Operating Envelope - QT-750 Table and


QW-420 Wall Stand with QW-420-T-D (Tilt Option)

Revision F Model RS-590/RS-590-MAT Tube Stand - Service Manual


14 Carestream Health, Inc
Part: Introduction/Overview

Model RS-590/RS-590-MAT System Operating Envelope - QT-750 Table,


QW-420-T-D (Manual Tilt) and FDR Fixed Receptor Option

Model RS-590/RS-590-MAT Tube Stand Service Manual Revision F


Carestream Health, Inc 15
Part: Introduction/Overview

Model RS-590/RS-590-MAT System Operating Envelope - QT-750 Table, QW-420-T-D


(Manual Tilt) and Q-DRT-R (Rotating Receptor Option

Revision F Model RS-590/RS-590-MAT Tube Stand - Service Manual


16 Carestream Health, Inc
4

PART: GETTING
STARTED

17
18
Part: Getting Started

INSTALLATION
This supplement defers to the Technical Manual for all installation instructions, except
for those pertaining to wiring and connectivity of the power supply, and integration of
the Carestream tube head assembly, including handgrip devices.

STARTUP
Procedures pertaining to checking and aligning the mechanical system, including
bearings, end stops, collimator adjustments, alignment of the table and wall stand,
are provided in the Technical Manual, or the collimator documentation.
TUBE HEAD ASSEMBLY
The RS-590 Tube Stand includes a Tube Head Assembly, designed by Carestream.
This consists of an X-ray tube, a collimator, and handgrip unit, which ultimately is
interfaced to the Siemens Tube Support Arm.
The X-ray tube, the collimator, and the handgrip unit are packaged separately, to
ensure that they are not damaged in transit. It is necessary for the installer to
physically mount these components using the interface bracket (with trunnion
rings), affixed at the factory, to the mounting plate of the Siemens Tube Support
Arm Assembly. The Tube Head assembly is built-up from the trunnion bracket,
once it is properly interfaced to the rotating element of the Siemens Tube Support
Arm.

NOTE: Additional information as to imple-


mentation of the Progeny Linear II and Linear IV
Automatic Collimators is provided in this manual
at Part Appendix.

When properly mounted, the tube can be oriented downward, or to the side, since
the rotating element serves as the mounting surface for the trunnion bracket
assembly.
Tube Head Assembly
The Tube Stand is integrated with Carestream ’s Tube Head Assembly,
and the following descriptions and figures show how the X-ray Tube is
attached to the Tube Support Arm of the Siemens Tube Stand (vertical
axis). Refer to the following figures, and proceed with assembly of the
tube:

Model RS-590/RS-590-MAT Tube Stand - Service Manual Revision F


Carestream Health, Inc 19
Part: Getting Started

Siemens Tube
Support Arm
Assembly

Interface Between Siemens Tube Support Arm Assembly


and X-ray Tube Trunnion Assembly
The Tube Head is assembled as follows:
1. Install the Trunnion Assembly onto the Siemens Tube Support
Arm Assembly, using four (4) 1/4-20 x 2" hex head bolts and lock-
washers (see Figures).
2. Remove top half of Trunnion Rings.

Typical Trunnion Bracket Assembly Components

Revision F Model RS-590/RS-590-MAT Tube Stand - Service Manual


20 Carestream Health, Inc
Part: Getting Started

3. Place X-ray tube onto lower half of Trunnion Rings and re-mount
upper half.

TUBE SUPPORT
BRACKET

TRUNNION
ASSEMBLY

1/4-20 X 2" HEX


COLLIMATOR
HEAD BOLT AND
MOUNTING
LOCK-WASHER
PLATE
(4X)

Tube Head Assembly Ready for HandGrip and Collimator Mounting


4. Install Trunnion Handle on Trunnion Assembly using two (2) 1/4-
20 x 1/2 in. button head socket screws.
5. Check trunnion handle locking. If too loose, remove Trunnion
Handle and rotate each Trunnion Cam clockwise a few turns. If
too tight, remove Trunnion Handle and rotate each Trunnion Cam
counterclockwise a few turns. Re-install Trunnion Handle and re-
check locking. Repeat until properly adjusted.

Model RS-590/RS-590-MAT Tube Stand - Service Manual Revision F


Carestream Health, Inc 21
Part: Getting Started

TRUNNION
HANDLE 1/4-20 X 綌
BUTTON HEAD
SOCKET SCREW
TRUNNION (2X)
CAM

T:\Storage\Manuals\RS-590\GRAPHICS\RS-590 Tube & Trunnion.cdr

Trunnion Handle Installed

6. Remove the Collimator Mounting Plate from the collimator.


7. Attach the Collimator Mounting Plate to the Tube Support Bracket
and X-ray tube. Install required screws and spacer(s) supplied
with collimator between X-ray tube and Tube Support Bracket
(refer to the installation instructions provided with the collimator
for complete mounting instructions).
8. Mount Tube Support Bracket onto Trunnion using four (4) 1/4-20
hex head bolts.

WARNING! Double-check collimator mounting


hardware for security, after installation in accor-
dance with instructions provided by collimator
manufacturer. Personal injury can result, if the
collimator separates from its mounting bracket

NOTE: THE REMAINING INSTRUC-


TIONS, IN THIS SECTION, PERTAIN TO
INSTALLATION OF OPERATOR DIGITAL
TOUCH SCREEN HANDGRIP (RS-590-LD
or RS-590-LD-MAT).

Refer to the TechVision Service Manual (DC30-126) for setup of the optional Tech-
Vision handgrip unit (QG-TV-C, QG-TV-CM, QG-TV-IP-C or QG-TV-IP-C-M).

Revision F Model RS-590/RS-590-MAT Tube Stand - Service Manual


22 Carestream Health, Inc
Part: Getting Started

OPTIONAL DIGITAL TOUCH SCREEN HANDGRIP (RS-590-LD/RS-590-LDM)


This section describes installation and startup of the enhanced capability handgrip
RS-590-LD or RS-590-LDM (auto-tracker version), which features large angulation
display, tactile motion control buttons, and digital touch screen with motion lock
capability, SID display, longitudinal and transverse detent indicators, and current
motion icon.

Tube Stand with Digital Touch Screen Handgrip Features Shown


Systems configured with the Digital Touch Screen Handgrip do not have
generator controls, but the touch screen provides the functionality shown
in the figure below. The Digital Touch Screen Handgrip also requires
setup and calibration, which is described in the next section.

Model RS-590/RS-590-MAT Tube Stand - Service Manual Revision F


Carestream Health, Inc 23
Part: Getting Started

1 2

4
3

(a) Collimator Disconnected (c)

(b) Internal Comm Fault RS-590 Disconnected (d)

1) SID display
2) Longitudinal detent active
3) Transverse detent active
4) Active motion axis icon
5) Status indications (see a through d)
a. Indicates a system fault due to an attempt to move the axes while locked.
b. Indicates that internal communications has failed within TechVision.
c. Indicates that the CAN bus cable connection is faulty (automatic collimators only).
d. Indicates that the CAN bus cable connection is faulty between TechVision and RS-590.
T:\Storage\Manuals\RS-590\GRAPHICS\RS-590-LD Touch Scr Features.cdr

Touch Screen Features for Digital Handgrip — RS-590-LD/RS-590-LDM

INSTALLATION INSTRUCTIONS
The Digital Touch Screen Handgrip is shipped fully assembled. It is designed as an
enhanced capability operator control option, specifically for the RS-590 Tube
Stand, and is easily installed.
The following paragraph contains installation procedures for Digital Touch Screen
Handgrip.
Digital Touch Screen Handgrip Installation Procedure
When supplied with the Tube Stand, the Digital Touch Screen Handgrip is
shipped separately, and must be mounted to the tube head assembly, fol-
lowing the instructions provided in "Tube Head Assembly".

Revision F Model RS-590/RS-590-MAT Tube Stand - Service Manual


24 Carestream Health, Inc
Part: Getting Started

Refer to the figure below, and the Interconnection Diagram for electrical
connections required.

DIGITAL TOUCH SCREEN HANDGRIP: CONNECTOR PANEL

NOT USED

CONNECTOR J7
(POWER FROM RS-590 CONNECTOR J6
CABLE D306-X1) (CANBUS DATA1)

Digital Touch Screen Handgrip Rear Panel Cable Connectors


1. Connect 8-pin plug attached to cable marked D306X1 (this cable exits
RS-590 Tube Stand corrugated hose) to POWER connector J7 on Dig-
ital Touch Screen Handgrip rear panel.
2. Connect RS-590 Tube Stand CANBUS cable D306 plug to CANBUS
Data 1 connector J6 on Digital Touch Screen Handgrip rear panel.
Power Supply Installation
Typically, the power supply (QG-28VPS) is installed within the Carestream
Generator, but when another generator is configured into the system, an
external power supply unit PS-7 is required, depending upon whether or
not a collimator power supply is available from the generator. These
devices are documented in the Power Supply Installation Instruction
(DC30-014).

Model RS-590/RS-590-MAT Tube Stand - Service Manual Revision F


Carestream Health, Inc 25
Part: Getting Started

RS-590 TUBE SUPP0RT


POWER CABLE LABEL:
W110
PSU.0802.X1 >
TS.D303.X1 <

TO
SU TU FEMALE PLUG
PP BE
OR
T

POWER CABLE ASSEMBLY


FOR TUBE SUPPORT
(QG-28VPS POWER
SUPPLY OPTION)
TO
GE
CA NE
BI RA
MALE NE TO
RECEPTACLE T R

T:\Storage\Manuals\RS-590 Service Manual\ELEMENTS\Power Supply Interface Cables and Plug\Power Supply Interface Cable and Plug.cdr

Power Supply Interface Cable For QG-28VPS Option

SYSTEM CONFIGURATION/CALIBRATION
The Digital Touch Screen Handgrip has Service Modes that allow the X-ray Techni-
cian or the Field Service Technician to modify the operation of Tube Stand, and allow
it to be calibrated. The procedures that follow describe, in order, how to:
• Access either the Field Tech Service Mode or Operator Service Mode options
• Use each of the Field Tech Service Mode functions and utilities (including
model-specific utilities for the Tube Stand).
• Use each of the Operator Service functions and utilities.
ACCESSING SERVICE MODE
Calibration and configuration can only be performed by accessing the Digital
Touch Screen Handgrip Service Mode. To gain access to the Service Mode, pass-
word entry is required. This is to prevent unintended and/or unauthorized
changes to system settings. To access Service Mode, proceed as follows:
1. On the Digital Touch Screen Hand-
grip Display, tap twice within one
second on the Quantum Logo.

Revision F Model RS-590/RS-590-MAT Tube Stand - Service Manual


26 Carestream Health, Inc
Part: Getting Started

2. Once the Service Mode has been selected, the Password Screen will appear
(see figure below):

CURRENT
SOFTWARE
REVISION IS
DISPLAYED HERE

Password Screen
3. Use the numeric "touch" keypad to enter the Service Password (58623), then
press ENTER. To start again, push Clear.
NOTE: Digital Touch Screen Handgrip returns to
the previous screen when an incorrect password is
entered, or after 15 seconds with no keys pressed.

Upon successful entry of the Service Password, the Service Options screen is dis-
played, as described below.
SERVICE OPTIONS
The Digital Touch Screen Handgrip calibration process is contingent upon the fol-
lowing configuration characteristics:
1. Whether, or not, the X-ray System is integrated with a computer-based imag-
ing system.
2. Whether the system employs a Carestream image receptor or a Carestream
receptor.
The FieldTech Options screens provide access to all of the configurable parame-
ters in the System. Note that the FieldTech Options screen (below), shows options
available if the "Down" button is used to scroll the screen.

Model RS-590/RS-590-MAT Tube Stand - Service Manual Revision F


Carestream Health, Inc 27
Part: Getting Started

NOTE: Configuration and calibration procedures


for TechVision used in Q-Rad QS-MAT (Auto-track)
applications are described in the Q-Rad Model QS-
MAT Service Manual (DC30-143)

The following information describes the purpose of each button on the Screen.
• Device Configuration: Select this button to verify or modify the
Digital Touch Screen Handgrip’s IP Address, Subnet Mask, or Default
Gateway value.
• RS590 Calibration: Allows calibration of physical orientation
parameters for the handgrip (RS-590-LD/RS-590-LDM).
• Language Configuration: Touch this button to access a menu of
languages which can be enabled for all displays and nomenclature.
• Regional Configuration: Touch this button to set displays of
measurements to metric or standard values.

FieldTech Options

Device Configuration RS590 Calibration

Language Configuration Regional Configuration

Change Date Time Calibrate Touch Screen

Down
Exit

T:\Storage\Manuals\RS-590\Service Manual\GRAPHICS\RS590 SERVICE OPTIONS SCR.cdr

FieldTech Options Screen

Revision F Model RS-590/RS-590-MAT Tube Stand - Service Manual


28 Carestream Health, Inc
Part: Getting Started

FieldTech Options

Change Date Time Calibrate Touch Screen

Reset Settings FTP

Restart Application

Up Down
Exit

T:\Storage\Manuals\RS-590\Service Manual\GRAPHICS\RS590 SERVICE OPTIONS SCR.cdr

FieldTech Option Screen - Continued


• Change Date Time: Select to modify the system date and/or time
settings.
• Calibrate Touch Screen: Begins a routine that matches the points
touched on the calibration setup screen to the internal position
interpretation.
• Down: Touch this button to scroll the screen listings down to reveal
additional options (see bottom of FieldTech Options screen).
• Reset Settings: Factory default settings can be applied via the "Reset
Settings" button. Upon activation of the "Reset Settings" button, a sub-
screen, is displayed, stating the default configuration, including a restart
of the application will occur if the "OK" button is selected. The screen
will revert to "FieldTech Options" screen
• FTP: Activation of the "FTP" causes the Digital Touch Screen Handgrip
unit to enter a state, where field upgrades can be implemented This

Model RS-590/RS-590-MAT Tube Stand - Service Manual Revision F


Carestream Health, Inc 29
Part: Getting Started

feature is used when installing updates. Selection of the "Close" button


on the FTP screen will initiate a restart of the application.

FTP Screen
• Restart Application: This feature allows for initialization without
effecting any configuration changes. The RS-590-LD display will cycle
through it normal boot-up sequence, ending with Main Display Screen.
In order to again access the FieldTech Options screen, the Technician will
have to log-on, using the keypad and log-on procedure described
Accessing Service Mode at the beginning of this section.
• Up: Touch this button to scroll the screen listings up to reveal additional
options.
• Exit: Touch this button to leave the Service Mode and return to
Operator Mode.
NOTE: Pressing the Back button in any of
the following screens returns the system
to the Service Options Menu.

Revision F Model RS-590/RS-590-MAT Tube Stand - Service Manual


30 Carestream Health, Inc
Part: Getting Started

RS-590 System Calibration


Selecting RS-590 Tubestand Calibration button in the Field Tech Options
menu, displays the Tubestand Calibration Menu:

Tubestand Calibration Menu


This screen offers five choices:
• Vertical 40" Detent - This feature provides a software-driven
detent, when the Tubestand Calibration item ("Vertical 40" Detent")
is set to "On". This functionality also requires that the X-ray tube is
pointed at the radiographic table, and the velocity of the vertical axis
is low and steady. When properly activated, the vertical axis will
stop at the 40" SID position.
• Handle Unlock Vertical Only - The brake release buttons ("pinky"
switches), imbedded in the handles of the Digital Touch Screen
Handgrip unit, can be configured to allow vertical movement only
when the Tubestand Calibration screen item is set to "On".
Otherwise, the buttons unlock all linear travel axes.
• Locking Speed Calibration - Adjusts the speed limit for lock
activation (200 mm/s, 175 mm/s, 150 mm/s, or 125 mm/s settings
are available). Note that the faster the limit setting, a more abrupt
"stop" will occur when the tube support is moved to either a
transverse or longitudinal detent.

Model RS-590/RS-590-MAT Tube Stand - Service Manual Revision F


Carestream Health, Inc 31
Part: Getting Started

• Plumb Angle Calibration - The Plumb Angle Calibration function


calibrates the Digital Touch Screen Handgrip’s tube plumb angulation
display.
• SID Height Calibration - The SID Calibration function is used to
set internal variables to match the Digital Touch Screen Handgrip
hardware that measures the SID.
Locking Speed Calibration
Proceed as follows:
1. When viewing the Tubestand Calibration Menu, the current locking
speed setting appears in the Locking Speed field, as shown in the
sample screen Figure 3-9.
2. To modify the setting, select the Modify button adjacent to the
"Locking Speed" field. The Locking Speed Setting Sub-screen
displays:

Locking Speed:

125 mm/s 150 mm/s

175 mm/s 200 mm/s

Cancel

Locking Speed Setting Sub-screen


3. Select the desired locking speed (125 mm/s, 150 mm/s, 175 mm/s, or
200 mm/s). The new setting now appears in the Locking Speed field.

Revision F Model RS-590/RS-590-MAT Tube Stand - Service Manual


32 Carestream Health, Inc
Part: Getting Started

Plumb Angle Calibration


Proceed as follows:
1. In the Tubestand Calibration Menu, select the Calibrate button
adjacent to the "Plumb Angle" field. Note that the first of three
plumb angle calibration screen appears.
2. Follow the instructions, and then select OK.
3. Follow the instructions, and then select OK for the next two plumb
angle calibration screens.
4. The Final Angle displayed on the bottom of the screen will change to
0.
5. Rotate the Digital Touch Screen Handgrip to both the +90° and -90°
detent positions and verify the is 90° in each position.

(1) (2)
(3)
Left-Facing 90° Right-Facing 90°
Horizontal Calibration Screen
Calibration Screen Calibration Screen

3-Point Plumb Angle Calibration


SID Height Calibration
Proceed as follows:
1. In the Tubestand Calibration Menu, select the Calibrate button
adjacent to the "SID Height" field. The sub-screen shown next, calls
for the Digital Touch Screen Handgrip to be set at 10".

Model RS-590/RS-590-MAT Tube Stand - Service Manual Revision F


Carestream Health, Inc 33
Part: Getting Started

SELECT “CALIBRATE”
TO BEGIN SID HEIGHT
CALIBRATION PROCESS

Tubestand Calibration Sub-screen

RS590 SID Height Calibration

Please place TechVision at 10?

OK

RS-590 SID 10" Height Sub-screen


2. Move the tube/handgrip upward until a measure of 10" SID is
reached. Press OK.

Revision F Model RS-590/RS-590-MAT Tube Stand - Service Manual


34 Carestream Health, Inc
Part: Getting Started

The following screen will appear:

RS590 SID Height Calibration

Please place TechVision at 40?

OK

RS-590 SID Calibration Sub-screen


3. Move the tube/hand grip so that a 40" SID is achieved (pull out the
cassette tray and measure 40" SID relative to the cassette). Press
OK.
NOTE: The displayed SID is the truncation of the dis-
tance to the nearest whole inch value. It may be nec-
essary to measure at a fraction of an inch added to
both 10" and 40" measurements so that the whole
inch display on the Digital Touch Screen Handgrip
more closely matches other SID measurement and dis-
play means. If acceptable precision cannot be
achieved in the display relative to another, contact
Carestream Service Department. If the SID display
does not move in concert with the tube/handgrip ver-
tical travel, check the connection path from the height
potentiometer on the tube stand and the Digital Touch
Screen Handgrip.

4. Exit to the Main Screen, and verify the SID indication on the Digital
Touch Screen Handgrip screen correctly tracks the height of the tube.
This completes the tube stand specific calibrations, proceed to "Calibrate
Touch Screen" to continue.

Model RS-590/RS-590-MAT Tube Stand - Service Manual Revision F


Carestream Health, Inc 35
Part: Getting Started

Calibrate Touch Screen


Pressing the Calibrate Touch Screen button in the Service Options
menu (see figure below) displays the Calibrate Touch Panel Screen:

Calibrate Touch Panel Screen


This procedure will reset the calibration of finger press and location on
the screen. When the above screen appears, follow the instructions; press
your finger on the cross as it appears in various positions on the screen,
each time holding it until the cross moves to a different location or the
calibration ends. There is a two-second pause after the calibration ends
while the calibration constants are permanently stored. Note that this
function is normally done at the factory and does not need to be
repeated, as the Digital Touch Screen Handgrip remembers the settings
while the power is off.

Revision F Model RS-590/RS-590-MAT Tube Stand - Service Manual


36 Carestream Health, Inc
PART: LIFECYCLE

37
38
Part: Lifecycle

MAINTENANCE
This supplement defers to the Technical Manual for instructions as to maintenance
procedures for:
• Longitudinal Rails and Transverse Track
• Telescoping Carriage
• Support Arm
PERIODIC INSPECTION
Ensure that collimator and hand grip mounting hardware is secure.
ADJUSTMENT
This supplement defers to the Technical Manual for instructions as to adjustment pro-
cedures for:
• Ball Bearings
• Central Beam Drift
• Parallelism of the X-ray Tube to the Table
• Light Field / Radiation Field of Collimator
• Wobble of the Support Arm
• Tube Unit Drift
• Aligning the Collimator
• Adjusting the Weight Compensation of the Spring-loaded Mechanism
• Vertical Axis Brake Gap (RVA)
• Horizontal Axis Brake Gap (RHA)
• Transverse and Longitudinal Carriage Brake Gap
• Detent Forces
• Stop Assembly
REPLACEMENT OF PARTS
Except for the following descriptions as to replacement of parts, this supplement
defers to the Technical Manual:
• Handles and Buttons: Refer to the information provided in the following
section.
• Tube Assembly: Refer to X-ray Tube manufacturer documentation.
• Collimator: Refer to Collimator manufacturer documentation.

Model RS-590/RS-590-MAT Tube Stand - Service Manual Revision F


Carestream Health, Inc 39
Part: Lifecycle

HANDLES AND BUTTONS


Replacement parts for the three handgrip options available with the Tube Stand, are
described as follows:
• Digital Touch Screen Handgrip (RS-590-LD/RS-590-LDM) — see this
Supplement.
• TechVision Control Panel (QG-TV-IP-C/QG-TV-IP-C-M) — see TechVision
Control Panel Service Manual (DC30-049).
REPLACEMENT PARTS AND ORDERING INFORMATION
The table below provides a list of replaceable parts for the RS-590-LD/RS-590-
LDM Handgrip. The annotated photograph of the Handgrip shows the locations
of the replaceable parts.

Digital Touch Screen Handgrip (RS-590-LD/RS-590-LDM) Replaceable Parts

Part Used
Item Description Number On
1 (1 ea.)Main PC Board AY40-093S RS-590-LD,
RS-590-LDM

2 Single Board Computer (1 ea.) AY40-092S RS-590-LD,


RS-590-LDM

3 Touch Screen Cable (1 ea.) AY51-334 RS-590-LD,


RS-590-LDM

4 SD card (Operating System) (1 ea.) EL85-199S RS-590-LD,


RS-590-LDM

5 Front Panel (1 ea.) AY20-392 RS-590-LD,


RS-590-LDM

Revision F Model RS-590/RS-590-MAT Tube Stand - Service Manual


40 Carestream Health, Inc
Part: Lifecycle

ORDERING INFORMATION
To order replacement parts for the tube stand, contact the Service Department
at;
Carestream Health, Inc
150 Verona Street, Rochester,NY14608, USA
Phone: 800-328-2910
www.carestream.com

When ordering replacement parts, supply the following information:


• Model and serial number of equipment
• Part number
• Part description
• Quantity required
• P.O. Number
• Shipping Instructions
When ordering components or parts not listed in the Handgrip Replaceable
Parts Table above, a complete description of the part, including its function and
location should be provided with the model number and serial number of the
unit. Use only original replacement parts, as supplied or recommended by Car-
estream Health, Inc.

Tube Stand Parts Location Diagram (RS-590-LD/RS-590-LDM Handgrip)

Model RS-590/RS-590-MAT Tube Stand - Service Manual Revision F


Carestream Health, Inc 41
Part: Lifecycle

Tube Stand Parts Location Diagram (RS-590-LD/RS-590-LDM Handgrip) Continued

Revision F Model RS-590/RS-590-MAT Tube Stand - Service Manual


42 Carestream Health, Inc
PART: APPENDIX

43
44
Part: Appendix

LOCATION OF LABELS
The following illustration shows the location of the identification label, and the serial
number label for the power supply options.
This supplement defers to the Technical Manual for warning label locations related to
the Siemens Tube Stand (see the Table of Labels in the Appendix of the Technical
Manual).

A
1

A
4

3
Rear Panel

B
2
POWER SUPPLY MODULE
T:\Storage\Manuals\RS-590\GRAPHICS\Label Locations.cdr

Serialization Label Locations —Tube Stand

Model RS-590/RS-590-MAT Tube Stand - Service Manual Revision F


Carestream Health, Inc 45
Part: Appendix

TABLE OF LABELS

No. Label Description


(1) RS-590/RS-590-MAT subsystem identifica-
tion label

(2) Power Supply (optional):


PS-7 (supplied with manual collimators)

3 Collimator (typical)

4 X-ray Tube (typical)

COLLIMATOR INTERFACE DIAGRAMS


The following diagrams provide information pertaining to the proper connection of the
manual, or automatic collimators shown in the Tube Stand Compatible Collimator List
found in Part: Introduction/Overview in this manual.

Revision F Model RS-590/RS-590-MAT Tube Stand - Service Manual


46 Carestream Health, Inc
Part: Appendix

Manual Collimator Interface to Tube Stand Diagram

Model RS-590/RS-590-MAT Tube Stand - Service Manual Revision F


Carestream Health, Inc 47
Part: Appendix

Auto-track System Cable Diagram

Revision F Model RS-590/RS-590-MAT Tube Stand - Service Manual


48 Carestream Health, Inc
Part: Appendix

Linear IV Collimator Interface to Tube Support Diagram

Model RS-590/RS-590-MAT Tube Stand - Service Manual Revision F


Carestream Health, Inc 49
Part: Appendix

THIS PAGE INTENTIONALLY LEFT BLANK

Revision F Model RS-590/RS-590-MAT Tube Stand - Service Manual


50 Carestream Health, Inc